Mark V

User Manual

Please read this manual carefully before using Alcovisor® Mark V.

Alcovisor® Mark V uses an electrochemical fuel cell to detect and measure the
concentration of alcohol vapor in expired breath.

Although the Alcovisor® Mark V is extremely simple to operate, it is important that the
following procedure is used each time a breath is given. Deviation from the proper
procedure will not generally affect the result of a test. However, the integrity and
capability of an operator, and even the legality of any resulting action, could be questioned
and brought into doubt if it is found later that the operator did not follow the proper testing
procedure.

PRETEST QUESTIONING
Some foods and even “non-alcoholic” drinks may contain traces of alcohol, which will
affect the test result through a “mouth alcohol” effect. To prevent this, wherever possible,
insure that a delay of about 20 minutes has elapsed since taking anything by mouth –
even medicines which may contain alcohol. Even a glass of water prior to the test will
cool the mouth and dilute the saliva, temporarily reducing the amount of alcohol in the
breath, and, consequently, the instrument reading. A delay of at least two minutes
should take place between the time of last smoked and the test.
STRUCTURE

Sample Port

Screen

Function Button
On/Off and Enter Button

ICON DESCRIPTION

1 Weekday Indicator

2 Time Indicator

3 Battery Indicator

4 Standard Test

5 Test Records

6 Setting

7 Date Indicator

Previous Screen/Cancel Next Screen/Confirm Main Screen

Print Repeat Test


DETAILED PROCEDURE

Standard Test

1. Attach a disposable mouthpiece onto the sample port.


2. Select the “Standard Test” icon by “Function Button ” and “Enter Button”.
3. When displays “Please Blow”, take a deep breath and blow until BEEP tone stops.
4. Test result displays on the display.
5. If printer is using, connect the printer with the device by the cable provided, switch on

the printer. After test result displays, select and the test result will be printed
automatically.

6. Select to start another test.


7. If testee did not blow within 30 seconds, tester can select “Refuse” to confirm the
testee refused to do the test; or select “Test Again” to allow testee to do the test
again.
8. If testee failed to give a valid breath sample, tester can select “Discontinue” to
confirm; or select “Test Again” to allow testee to do the test again.
9. A passive test can be performed by select “Manual” during the testee is blowing.

Test Record

1. 2,000 test records can be stored.

2. Use or to move to the previous or next record.


3. For printing an individual test record, select the test record which needs to be printed,
connect the printer with the device by the cable provided, switch on the printer, select

and the test result will be printed automatically.


Setting

1. Date & Time


i. Set correct Date and Time.
ii. Select “Update” to confirm.

2. Auto Off
i. Select time limit to turn off the device automatically.

ii. Select to confirm

3. Print
i. Select “Auto Print” to print the result automatically after each test.
ii. Select number of printout copies needed.

iii. Select to confirm.


iv. If last calibration date needs to be shown on the printout, select “Cal.”.

4. Cali
i. When the screen displays “Calibration Needed”, send the device for calibration.
ii. A password is needed for calibration to ensure calibration is performed by
eligible persons.
iii. Please refer to Calibration Instruction.

5. Advanced Setting
i. Only legitimated persons are allowed for advanced setting.
ii. A Password is needed.
iii. Please refer to Advanced Setting Instruction.