9.2.1.2 Certificado Datalogger - Encardio Rite

The document provides test results for a data logger including firmware version, battery voltage, calibration frequencies and temperatures with errors, modem and sensor functions were OK. Notes provide equations to enter sensor coefficients into the data logger.

Uploaded by

italodex
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
29 views1 page

9.2.1.2 Certificado Datalogger - Encardio Rite

The document provides test results for a data logger including firmware version, battery voltage, calibration frequencies and temperatures with errors, modem and sensor functions were OK. Notes provide equations to enter sensor coefficients into the data logger.

Uploaded by

italodex
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

TEST CERTIFICATE

MODEL ES DL-30-7V2-4G-(US) DATALOGGER

Mfg. Sl. No. : 22J0478 Date : 15.11.2022


Firmware version : 01.60
Battery Voltage : 7.2V
License provided : 05

Calibration Data
The below mentioned data was taken with calibrated vibrating wire interfac e Model ESVI-01-XX.
Logger frequency calibration data

Standard Frequency read by


Error
Frequency data logger
% of reading
(Hz) (Hz)
1250 1249.92 0.006
2500 2499.95 0.002
5000 4999.90 0.002
Logger temperature calibration data
Calculat ed
Observed 0
Input Resistance (Ohm) temperature 0 Error ( C)
0 temperature ( C)
( C)
29130 -20 -19.8 0.2
16600 -10 -9.8 0.2
9796 0 0.1 0.1
1598 40 40.0 0.0
1081 50 50.0 0.0
274.9 90 90.0 0.0
216.1 98 98.1 0.1

GSM/GPRS modem : 4G (US) Scanning function : OK


Functioning of SDI-12 Sensors : OK

Note :

1) Data Logger Model ES DL-30 has the capability to calculate the paramet er values using second order
2
polynomial equation : Y = A2 X + A 1 X + A 0 .
where Y is paramet er value in engineering units and X is a function of the sensor frequency output.
2
In V/W sensors supplied by Encardio -Rite , value of X is in terms of Digits (Frequency / 1000)
Replace A 1 with the value B / 1000 and A 2 with A / 10,00,000 when entering these coefficients in
data logger . A 0 will remain the same as value of C.
A,B,C values are mentioned in all sensors test certificate.

2) In case of Linear calculation : Feed A 2 = 0 , A 1 = G / 1000 and A 0 = -(G X R0) ,


G is linear gage factor, value of G is mentioned in sensors test certificate.

Tested By:

ENCARDIO-RITE GROUP