0% found this document useful (0 votes)
34 views

Lab - 2C - Additional User Manuals For Oscilloscope

The document describes how to use an oscilloscope to measure voltages and waveforms. It provides instructions on setting up the oscilloscope channels and inputs, taking measurements of DC and AC circuits using probes, and using features like cursors and math functions. Tables are included to record measurement results.

Uploaded by

robiulislammohor
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
34 views

Lab - 2C - Additional User Manuals For Oscilloscope

The document describes how to use an oscilloscope to measure voltages and waveforms. It provides instructions on setting up the oscilloscope channels and inputs, taking measurements of DC and AC circuits using probes, and using features like cursors and math functions. Tables are included to record measurement results.

Uploaded by

robiulislammohor
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

3A

The Oscilloscope (Tektronix MDO 3000 series)

Objective
This exercise is of a particularly practical nature, namely, introducing the use of the oscilloscope. The
various input scaling, coupling, and triggering settings are examined along with a few specialty features.

Theory Overview
The oscilloscope (or simply scope, for short) is arguably the single most useful piece of test equipment in
an electronics laboratory. The primary purpose of the oscilloscope is to plot a voltage versus time
although it can also be used to plot one voltage versus another voltage, and in some cases, to plot voltage
versus frequency. Oscilloscopes are capable of measuring both AC and DC waveforms, and unlike typical
DMMs, can measure AC waveforms of very high frequency (typically 100 MHz or more versus an upper
limit of around 1 kHz for a general purpose DMM). It is also worth noting that a DMM will measure the
RMS value of an AC sinusoidal voltage, not its peak value.

While the modern digital oscilloscope on the surface appears much like its analog ancestors, the internal
circuitry is far more complicated and the instrument affords much greater flexibility in measurement.
Modern digital oscilloscopes typically include measurement aides such as horizontal and vertical cursors
or bars, as well as direct readouts of characteristics such as waveform amplitude and frequency. At a
minimum, modern oscilloscopes offer two input measurement channels although four and eight channel
instruments are increasing in popularity.

Unlike handheld DMMs, most oscilloscopes measure voltages with respect to ground, that is, the inputs
are not floating and thus the black, or ground, lead is always connected to the circuit ground or common
node. This is an extremely important point as failure to remember this may lead to the inadvertent short
circuiting of components during measurement. The standard accepted method of measuring a non-ground
referenced potential is to use two probes, one tied to each node of interest, and then setting the
oscilloscope to subtract the two channels rather than display each separately. Note that this technique is
not required if the oscilloscope has floating inputs (for example, in a handheld oscilloscope). Further,
while it is possible to measure non-ground referenced signals by floating the oscilloscope itself through
defeating the ground pin on the power cord, this is a safety violation and should not be done.

18 Laboratory Manual for AC Electrical Circuit Analysis


Equipment
(1) DC power supply model:________________ srn:__________________
(1) AC function generator model:________________ srn:__________________
(1) Digital multimeter model:________________ srn:__________________
(1) Oscilloscope, Tektronix MDO 3000 series model:________________ srn:__________________

Components
(1) 10 k actual:__________________
(1) 33 k actual:__________________

Schematics and Diagrams

Figure 3A.1

Laboratory Manual for AC Electrical Circuit Analysis 19


Figure 3A.2

Procedure

1. Figure 3A.1 is a photo of the face of a Tektronix MDO 3000 series oscilloscope. Compare this to the
bench oscilloscope and identify the following elements:
 Channel one through four BNC input connectors.
 RF input connector and settings section.
 Channel one through four select buttons.
 Horizontal Scale (i.e., Sensitivity) and Position knobs.
 Four Vertical Scale (i.e., Sensitivity) and Position knobs.
 Trigger Level knob.
 Math and Measure (in Wave Inspector) buttons.
 Save button (below display).
 Autoset button.
 Menu Off button.

2. Note the numerous buttons along the bottom and side of the display screen. These menu buttons are
context-sensitive and their function will depend on the most recently selected button or knob. Menus
may be removed from the display by pressing the Menu Off button (multiple times for nested menus).
Power up the oscilloscope. Note that the main display is similar to a sheet of graph paper. Each
square will have an appropriate scaling factor or weighting, for example, 1 volt per division vertically
or 2 milliseconds per division horizontally. Waveform voltages and timings may be determined
directly from the display by using these scales.

3. Select the channel one and two buttons (yellow and blue) and also press the Autoset button. (Autoset
tries to create reasonable settings based on the input signal and is useful as a sort of “panic button”).
There should now be two horizontal lines on the display, one yellow and one blue. These traces may
be moved vertically on the display via the associated Position knobs. Also, a trace can be removed by
deselecting the corresponding channel button. The Vertical and Horizontal Scale knobs behave in a
similar fashion and do not include calibration markings. That is because the settings for these knobs
show up on the main display. Adjust the Scale knobs and note how the corresponding values at the
bottom of the display change. Voltages are in a 1/2/5 scale sequence while Time is in a 1/2/4 scale
sequence.

20 Laboratory Manual for AC Electrical Circuit Analysis


4. When an input is selected, a menu will pop up allowing control over that input's basic settings. One of
the more important fundamental settings on an oscilloscope channel is the Input Coupling. This is
controlled via one of the bottom row buttons. There are two choices: AC allows only AC signals
through thus blocking DC, and DC allows all signals through (it does not prevent AC).

5. Set the channel one Vertical Scale to 5 volts per division. Set the channel two Scale to 2 volts per
division. Set the Time (Horizontal) Scale to 1 millisecond per division. Finally, set the input Coupling
to DC for both input channels and align the blue and yellow display lines to the center line of the
display via the Vertical Position knob (note that pushing the vertical Position knobs will
automatically center the trace).

6. Build the circuit of figure 3A.2 using E=10 V, R1=10 k and R2= 33k. Connect a probe from the
channel one input to the power supply (red or tip to the positive terminal, black clip to ground).
Connect a second probe from channel two to R2 (again, red or tip to the high side of the resistor and
the black clip to ground).

7. The yellow and blue lines should have deflected upward. Channel one should be raised two divisions
(2 divisions at 5 volts per division yields the 10 volt source). Using this method, determine the
voltage across R2 (remember, input two should have been set for 2 volts per division). Calculate the
expected voltage across R2 using measured resistor values and compare the two in Table 3A.1. Note
that it is not possible to achieve extremely high precision using this method (e.g., four or more digits).
Indeed, a DMM is often more useful for direct measurement of DC potentials. Double check the
results using a DMM and the final column of Table 3A.1.

8. Select AC Coupling for the two inputs. The flat DC lines should drop back to zero. This is because AC
Coupling blocks DC. This will be useful for measuring the AC component of a combined AC/DC
signal, such as might be seen in an audio amplifier. Set the input coupling for both channels back to
DC.

9. Replace the DC power supply with the function generator. Set the function generator for a one volt
peak sine wave at 1 kHz and apply it to the resistor network. The display should now show two small
sine waves. Adjust the Vertical Scale settings for the two inputs so that the waves take up the
majority of the display. If the display is very blurry with the sine waves appearing to jump about side
to side, the Trigger Level may need to be adjusted. Also, adjust the Time Scale so that only one or
two cycles of the wave may be seen. Using the Scale settings, determine the two voltages (following
the method of step 7) as well as the waveform’s period and compare them to the values expected via
theory, recording the results in Tables 3A.2 and 3A.3. Also crosscheck the results using a DMM to
measure the RMS voltages.

Laboratory Manual for AC Electrical Circuit Analysis 21


10. To find the voltage across R1, the channel two voltage (VR2) may be subtracted from channel one (E
source) via the Math function. Use the red button to select the Math function and create the
appropriate expression from the menu (ch1 – ch2). This display shows up in red. To remove a
waveform, press its button again. Remove the math waveform before proceeding to the next step.

11. One of the more useful aspects of the oscilloscope is the ability to show the actual waveshape. This
may be used, for example, as a means of determining distortion in an amplifier. Change the
waveshape on the function generator to a square wave, triangle, or other shape and note how the
oscilloscope responds. Note that the oscilloscope will also show a DC component, if any, as the AC
signal being offset or “riding on the DC”. Adjust the function generator to add a DC offset to the
signal and note how the oscilloscope display shifts. Return the function generator back to a sine wave
and remove any DC offset.

12. It is often useful to take precise differential measurement on a waveform. For this, the bars or cursors
are useful. Select the Cursors button toward the top of the oscilloscope. From the menu on the
display, select Vertical Bars. Two vertical bars will appear on the display (it is possible that one or
both could be positioned off the main display). They may be moved left and right via the
Multipurpose knobs (next to the Cursors button). The Select button toggles between independent and
tandem cursor movement. A read out of the bar values will appear in the upper portion of the display.
They indicate the positions of the cursors, i.e., the location where they cross the waveform. Vertical
Bars are very useful for obtaining time information as well as amplitudes at specific points along the
wave. A similar function is the Horizontal Bars which are particularly useful for determining
amplitudes. Try the Horizontal Bars by selecting them via the Cursors menu again (holding the
Cursors button will bring up the menu).

13. For some waveform parameters, automatic readings are available. These are accessed via the
Measure button. Press Measure, select Add Measurement, and page through the various options using
the Multipurpose b knob. Select Frequency. Note that a small readout of the frequency will now
appear on the display. Multiple measurements are possible simultaneously. Important: There are
specific limits on the proper usage of these measurements. If the guidelines are not followed,
erroneous values may result. Always perform an approximation via the Scale factor and divisions
method even when using an automatic measurement!

14. Finally, a snap-shot of the screen may be saved for future work using the USB port and a USB
memory stick via the Save Menu button. The pop up menu has options for saving the image as well as
the trace data or setup info. Select Save Screen Image to save a bit mapped graphics file that can be
used as is or processed further in a graphics program (for example, inverting the colors for printing).
The .PNG format is recommended.

22 Laboratory Manual for AC Electrical Circuit Analysis


Data Tables

VR2 Scale (V/Div) Number of Divisions Voltage Scope Voltage DMM

Oscilloscope

Theory X X

Table 3.1

Scale (V/Div) Number of Divisions Voltage Peak Voltage RMS


E Oscilloscope

E Theory X X

VR2 Oscilloscope

VR2 Theory X X

Table 3A.2

Scale (S/Div) Number of Divisions Period Frequency

E Oscilloscope
E Theory X X

Table 3A.3

Laboratory Manual for AC Electrical Circuit Analysis 23


3B
The Oscilloscope (Tektronix TDS 3000 series)

Objective
This exercise is of a particularly practical nature, namely, introducing the use of the oscilloscope. The
various input scaling, coupling, and triggering settings are examined along with a few specialty features.

Theory Overview
The oscilloscope (or simply scope, for short) is arguably the single most useful piece of test equipment in
an electronics laboratory. The primary purpose of the oscilloscope is to plot a voltage versus time
although it can also be used to plot one voltage versus another voltage, and in some cases, to plot voltage
versus frequency. Oscilloscopes are capable of measuring both AC and DC waveforms, and unlike typical
DMMs, can measure AC waveforms of very high frequency (typically 100 MHz or more versus an upper
limit of around 1 kHz for a general purpose DMM). It is also worth noting that a DMM will measure the
RMS value of an AC sinusoidal voltage, not its peak value.

While the modern digital oscilloscope on the surface appears much like its analog ancestors, the internal
circuitry is far more complicated and the instrument affords much greater flexibility in measurement.
Modern digital oscilloscopes typically include measurement aides such as horizontal and vertical cursors
or bars, as well as direct readouts of characteristics such as waveform amplitude and frequency. At a
minimum, modern oscilloscopes offer two input measurement channels although four and eight channel
instruments are increasing in popularity.

Unlike handheld DMMs, most oscilloscopes measure voltages with respect to ground, that is, the inputs
are not floating and thus the black, or ground, lead is always connected to the circuit ground or common
node. This is an extremely important point as failure to remember this may lead to the inadvertent short
circuiting of components during measurement. The standard accepted method of measuring a non-ground
referenced potential is to use two probes, one tied to each node of interest, and then setting the
oscilloscope to subtract the two channels rather than display each separately. Note that this technique is
not required if the oscilloscope has floating inputs (for example, in a handheld oscilloscope). Further,
while it is possible to measure non-ground referenced signals by floating the oscilloscope itself through
defeating the ground pin on the power cord, this is a safety violation and should not be done.

24 Laboratory Manual for AC Electrical Circuit Analysis


Equipment
(1) DC power supply model:________________ srn:__________________
(1) AC function generator model:________________ srn:__________________
(1) Digital multimeter model:________________ srn:__________________
(1) Oscilloscope, Tektronix TDS 3000 series model:________________ srn:__________________

Components
(1) 10 k actual:__________________
(1) 33 k actual:__________________

Schematics and Diagrams

Figure 3B.1

Figure 3B.2

Laboratory Manual for AC Electrical Circuit Analysis 25


Procedure

1. Figure 3B.1 is an outline of the main face of a Tektronix TDS 3000 series oscilloscope. Compare this
to the bench oscilloscope and identify the following elements:
 Channel one and two BNC input connectors.
 Trigger BNC input connector.
 Channel one and two select buttons.
 Horizontal Sensitivity (or Scale) and Position knobs.
 Vertical Sensitivity (or Scale) and Position knobs.
 Trigger Level knob.
 Quick Menu button.
 Print/Save button.
 Autoset button.

2. Note the numerous buttons along the bottom and side of the display screen. These buttons are
context-sensitive and their function will depend on the mode of operation of the oscilloscope. Power
up the oscilloscope and select the Quick Menu button. Notice that the functions are listed next to the
buttons. This is a very useful menu and serves as a good starting point for most experiment setups.
Note that the main display is similar to a sheet of graph paper. Each square will have an appropriate
scaling factor or weighting, for example, 1 volt per division vertically or 2 milliseconds per division
horizontally. Waveform voltages and timings may be determined directly from the display by using
these scales.

3. Select the channel one and two buttons (yellow and blue) and also select the Autoset button. (Autoset
tries to create reasonable settings based on the input signal and is useful as a sort of “panic button”).
There should now be two horizontal lines on the display, one yellow and one blue. They may be
moved via the Position knob. The Position knob moves the currently selected input (select the
channel buttons alternately to toggle back and forth between the two inputs). The Vertical and
Horizontal Scale knobs behave in a similar fashion and do not include calibration markings. That is
because the settings for these knobs show up on the main display. Adjust the Scale knobs and note
how the corresponding values in the display change. Voltages are in a 1/2/5 scale sequence while
Time is in a 1/2/4 scale sequence.

4. One of the more important fundamental settings on an oscilloscope is the Input Coupling. This is
controlled via one of the bottom row buttons. There are three choices: Ground removes the input thus
showing a zero reference, AC allows only AC signals through thus blocking DC, and DC allows all
signals through (it does not prevent AC).

5. Set the channel one Vertical Scale to 5 volts per division. Set the channel two Scale to 2 volts per
division. Set the Time (Horizontal) Scale to 1 millisecond per division. Finally, set the input Coupling

26 Laboratory Manual for AC Electrical Circuit Analysis


to Ground for both input channels and align the blue and yellow display lines to the center line of the
display via the Vertical Position knob.

6. Build the circuit of figure 3B.2 using E=10 V, R1=10 k and R2= 33k. Connect a probe from the
channel one input to the power supply (red or tip to the positive terminal, black clip to ground).
Connect a second probe from channel two to R2 (again, red or tip to the high side of the resistor and
the black clip to ground).

7. Switch both inputs to DC coupling. The yellow and blue lines should have deflected upward. Channel
one should be raised two divisions (2 divisions at 5 volts per division yields the 10 volt source).
Using this method, determine the voltage across R2 (remember, input two should have been set for 2
volts per division). Calculate the expected voltage across R2 using measured resistor values and
compare the two in Table 3B.1. Note that it is not possible to achieve extremely high precision using
this method (e.g., four or more digits). Indeed, a DMM is often more useful for direct measurement of
DC potentials. Double check the results using a DMM and the final column of Table 3B.1.

8. Select AC Coupling for the two inputs. The flat DC lines should drop back to zero. This is because AC
Coupling blocks DC. This will be useful for measuring the AC component of a combined AC/DC
signal, such as might be seen in an audio amplifier. Set the input coupling for both channels back to
DC.

9. Replace the DC power supply with the function generator. Set the function generator for a one volt
peak sine wave at 1 kHz and apply it to the resistor network. The display should now show two small
sine waves. Adjust the Vertical Scale settings for the two inputs so that the waves take up the
majority of the display. If the display is very blurry with the sine waves appearing to jump about side
to side, the Trigger Level may need to be adjusted. Also, adjust the Time Scale so that only one or
two cycles of the wave may be seen. Using the Scale settings, determine the two voltages (following
the method of step 7) as well as the waveform’s period and compare them to the values expected via
theory, recording the results in Tables 3B.2 and 3B.3. Also crosscheck the results using a DMM to
measure the RMS voltages.

10. To find the voltage across R1, the channel two voltage (VR2) may be subtracted from channel one
(E source) via the Math function. Use the red button to select the Math function and create the
appropriate expression from the menu (ch1 – ch2). This display shows up in red. To remove a
waveform, select it and then select Off. Remove the math waveform before proceeding to the next
step.

11. One of the more useful aspects of the oscilloscope is the ability to show the actual waveshape. This
may be used, for example, as a means of determining distortion in an amplifier. Change the
waveshape on the function generator to a square wave, triangle, or other shape and note how the
oscilloscope responds. Note that the oscilloscope will also show a DC component, if any, as the AC

Laboratory Manual for AC Electrical Circuit Analysis 27


signal being offset or “riding on the DC”. Adjust the function generator to add a DC offset to the
signal and note how the oscilloscope display shifts. Return the function generator back to a sine wave
and remove any DC offset.

12. It is often useful to take precise differential measurement on a waveform. For this, the bars or cursors
are useful. Select the Cursor button toward the top of the oscilloscope. From the menu on the display,
select Vertical Bars. Two vertical bars will appear on the display (it is possible that one or both could
be positioned off the main display). They may be moved left and right via the function knob (next to
the Cursor button). The Select button toggles between the two cursors. A read out of the bar values
will appear in the upper portion of the display. They indicate the positions of the cursors, i.e. the
location where they cross the waveform. Vertical Bars are very useful for obtaining time information
as well as amplitudes at specific points along the wave. A similar function is the Horizontal Bars
which are particularly useful for determining amplitudes. Try the Horizontal Bars by selecting them
via the Cursor button again.

13. For some waveform parameters, automatic readings are available. These are accessed via the Meas
(Measurement) button. Select Meas and page through the various options. Select Frequency. Note
that a small readout of the frequency will now appear on the display. Up to four measurements are
possible simultaneously. Important: There are specific limits on the proper usage of these
measurements. If the guidelines are not followed, erroneous values may result. Always perform an
approximation via the Scale factor and divisions method even when using an automatic measurement!

28 Laboratory Manual for AC Electrical Circuit Analysis


Data Tables

VR2 Scale (V/Div) Number of Divisions Voltage Scope Voltage DMM


Oscilloscope
Theory X X

Table 3B.1

Scale (V/Div) Number of Divisions Voltage Peak Voltage RMS


E Oscilloscope
E Theory X X
VR2 Oscilloscope
VR2 Theory X X

Table 3B.2

Scale (S/Div) Number of Divisions Period Frequency


E Oscilloscope
E Theory X X

Table 3B.3

Laboratory Manual for AC Electrical Circuit Analysis 29


3C
The Oscilloscope (GWInstek 2000)
Objective
This exercise is of a particularly practical nature, namely, introducing the use of the oscilloscope. The
various input scaling, coupling, and triggering settings are examined along with a few specialty features.

Theory Overview
The oscilloscope (or simply scope, for short) is arguably the single most useful piece of test equipment in
an electronics laboratory. The primary purpose of the oscilloscope is to plot a voltage versus time
although it can also be used to plot one voltage versus another voltage, and in some cases, to plot voltage
versus frequency. Oscilloscopes are capable of measuring both AC and DC waveforms, and unlike typical
DMMs, can measure AC waveforms of very high frequency (typically 100 MHz or more versus an upper
limit of around 1 kHz for a general purpose DMM). It is also worth noting that a DMM will measure the
RMS value of an AC sinusoidal voltage, not its peak value.

While the modern digital oscilloscope on the surface appears much like its analog ancestors, the internal
circuitry is far more complicated and the instrument affords much greater flexibility in measurement.
Modern digital oscilloscopes typically include measurement aides such as horizontal and vertical cursors
or bars, as well as direct readouts of characteristics such as waveform amplitude and frequency. At a
minimum, modern oscilloscopes offer two input measurement channels although four and eight channel
instruments are increasing in popularity.

Unlike handheld DMMs, most oscilloscopes measure voltages with respect to ground, that is, the inputs
are not floating and thus the black, or ground, lead is always connected to the circuit ground or common
node. This is an extremely important point as failure to remember this may lead to the inadvertent short
circuiting of components during measurement. The standard accepted method of measuring a non-ground
referenced potential is to use two probes, one tied to each node of interest, and then setting the
oscilloscope to subtract the two channels rather than display each separately. Note that this technique is
not required if the oscilloscope has floating inputs (for example, in a handheld oscilloscope). Further,
while it is possible to measure non-ground referenced signals by floating the oscilloscope itself through
defeating the ground pin on the power cord, this is a safety violation and should not be done.

30 Laboratory Manual for AC Electrical Circuit Analysis


Equipment
(1) DC power supply model:________________ srn:__________________
(1) AC function generator model:________________ srn:__________________
(1) Digital multimeter model:________________ srn:__________________
(1) Oscilloscope, GWInstek 2000 series model:________________ srn:__________________

Components
(1) 10 k actual:__________________
(1) 33 k actual:__________________

Schematics and Diagrams

Figure 3C.1A

Laboratory Manual for AC Electrical Circuit Analysis 31


Figure 3C.1B

Figure 3C.2

32 Laboratory Manual for AC Electrical Circuit Analysis


Procedure
1. Figure 3C.1 is an outline of the main face of a GWInstek 2000 series oscilloscope. Compare this to
the bench oscilloscope and identify the following elements:
 Channel one and two BNC input connectors.
 Trigger BNC input connector.
 Horizontal sensitivity (Time/Div) and Position knobs.
 Channel one and two Select buttons (lighted style).
 Channel one and two Vertical sensitivity (Volts/Div) and Position knobs.
 Trigger Level knob.
 Function keys, including Cursor and Measure.
 Hardcopy button.
 Autoset button.
 Math button.
 USB port.

2. Note the numerous buttons along the bottom and side of the display screen. These buttons are
context-sensitive and their function will depend on the mode of operation of the oscilloscope. Power
up the oscilloscope. Notice that the functions are listed next to the buttons. Note that the main display
is similar to a sheet of graph paper. Each square will have an appropriate scaling factor or weighting,
for example, 1 volt per division vertically or 2 milliseconds per division horizontally. Waveform
voltages and timings may be determined directly from the display by using these scales.

3. Depress the channel one and two Select buttons (they should light) and also select the Autoset button.
(Autoset tries to create reasonable settings based on the input signal and is useful as a sort of “panic
button”). There should now be two horizontal lines on the display, one yellow and one blue. They
may be moved via the Position knob. The Position knobs move the associated input. The Vertical and
Horizontal Scale knobs behave in a similar fashion and do not include calibration markings. That is
because the settings for these knobs show up on the main display. Adjust the Scale knobs and note
how the corresponding values in the display change. Voltages and Time base use a 1/2/5 scale
sequence.

4. One of the more important fundamental settings on an oscilloscope is the Input Coupling. This is
controlled via one of the bottom row buttons. There are three choices: Ground removes the input thus
showing a zero reference, AC allows only AC signals through thus blocking DC, and DC allows all
signals through (it does not prevent AC).

5. Set the channel one Vertical Scale to 5 volts per division. Set the channel two Scale to 2 volts per
division. Set the Time (Horizontal) Scale to 1 millisecond per division. Finally, set the input Coupling
to Ground for both input channels and align the blue and yellow display lines to the center line of the
display via the Vertical Position knobs.

Laboratory Manual for AC Electrical Circuit Analysis 33


6. Build the circuit of Figure 3C.2 using E=10 V, R1=10 k and R2= 33k. Connect a probe from the
channel one input to the power supply (red or tip to plus, black clip to ground). Connect a second
probe from channel two to R2 (again, red or tip to the high side of the resistor and the black clip to
ground).

7. Switch both inputs to DC coupling. The yellow and blue lines should have deflected upward. Channel
one should be raised two divisions (2 divisions times 5 volts per division yields the 10 volt source).
Using this method, determine the voltage across R2 (remember, input two should have been set for 2
volts per division). Calculate the expected voltage across R2 using measured resistor values and
compare the two in Table 3C.1. Note that it is not possible to achieve extremely high precision using
this method (e.g., four or more digits). Indeed, a DMM is often more useful for direct measurement of
DC potentials. Double check the results using a DMM and the final column of Table 3C.1.

8. Select AC Coupling for the two inputs. The flat DC lines should drop back to zero. This is because AC
Coupling blocks DC. This will be useful for measuring the AC component of a combined AC/DC
signal, such as might be seen in an audio amplifier. Set the input coupling for both channels back to
DC.

9. Replace the DC power supply with the function generator. Set the function generator for a one volt
peak sine wave at 1 kHz and apply it to the resistor network. The display should now show two small
sine waves. Adjust the Vertical Scale settings for the two inputs so that the waves take up the
majority of the display. If the display is very blurry with the sine waves appearing to jump about side
to side, the Trigger Level may need to be adjusted. Also, adjust the Time Scale so that only one or
two cycles of the wave may be seen. Using the Scale settings, determine the two voltages (following
the method of step 7) as well as the waveform’s period and compare them to the values expected via
theory, recording the results in Tables 3C.2 and 3C.3. Also crosscheck the results using a DMM to
measure the RMS voltages.

10. To find the voltage across R1, the channel two voltage (VR2) may be subtracted from channel one
(E source) via the Math function. Use the red button to select the Math function and create the
appropriate expression from the menu (ch1 – ch2). This display shows up in red. To remove a
waveform, simply deselect it (depress the associated button). Remove the math waveform before
proceeding to the next step.

11. One of the more useful aspects of the oscilloscope is the ability to show the actual waveshape. This
may be used, for example, as a means of determining distortion in an amplifier. Change the
waveshape on the function generator to a square wave, triangle, or other shape and note how the
oscilloscope responds. Note that the oscilloscope will also show a DC component, if any, as the AC
signal being offset or “riding on the DC”. Adjust the function generator to add a DC offset to the

34 Laboratory Manual for AC Electrical Circuit Analysis


signal and note how the oscilloscope display shifts. Return the function generator back to a sine wave
and remove any DC offset.

12. It is often useful to take precise differential measurement on a waveform. For this, the bars or cursors
are useful. Select the Cursor button toward the top of the oscilloscope. From the menu on the display,
select Vertical. Two vertical bars will appear on the display (it is possible that one or both could be
positioned off the main display). They may be moved left and right via the Variable knob (next to the
Cursor button). The Select button toggles between the two cursors. A read out of the bar values will
appear in the upper portion of the display. They indicate the positions of the cursors, i.e. the location
where they cross the waveform. Vertical Bars are very useful for obtaining time information as well
as amplitudes at specific points along the wave. A similar function is the Horizontal Bars which are
particularly useful for determining amplitudes. Try the Horizontal Bars by selecting them via the
Cursor button again.

13. For some waveforms parameters, automatic readings are available. These are accessed via the
Measure button. Select Measure and page through the various options. Select Frequency. Note that a
small readout of the frequency will now appear on the display. Now try RMS and compare the result
to that given by the DMM earlier. Note that several measurements are possible simultaneously.
Important: There are specific limits on the proper usage of these measurements. If the guidelines are
not followed, erroneous values may result. Always perform an approximation via the Scale factor and
divisions method even when using an automatic measurement!

14. Finally, a snap-shot of the screen may be saved for future work using the USB port and a USB
memory stick via the Hardcopy button. The result will be a bit mapped graphics file that can be used
as is (see below) or processed further in a graphics program (for example, inverting the colors for
printing).

Figure 3

Laboratory Manual for AC Electrical Circuit Analysis 35


Data Tables

VR2 Scale (V/Div) Number of Divisions Voltage Scope Voltage DMM

Oscilloscope

Theory X X

Table 1

Scale (V/Div) Number of Divisions Voltage Peak Voltage RMS

E Oscilloscope

E Theory X X

VR2 Oscilloscope

VR2 Theory X X

Table 2

Scale (S/Div) Number of Divisions Period Frequency

E Oscilloscope

E Theory X X

Table 3

36 Laboratory Manual for AC Electrical Circuit Analysis

You might also like