Heat of Reaction Sum 2011
Heat of Reaction Sum 2011
HEATS OF REACTION
LEARNING GOALS
INTRODUCTION
Chemical and physical changes usually involve the absorption or liberation of heat, given the
symbol q. When heat is absorbed, the change is said to be endothermic, and the numerical value
of the heat is given a positive sign (q > 0). When heat is liberated, the change is said to be
exothermic, and the numerical value of the heat is given a negative sign (q < 0). Under standard
conditions of temperature and constant pressure (exactly 25 oC and 1 atmosphere) with all
reactants in their usual physical states, the measured heat of a reaction is called the standard
enthalpy, given the symbol ΔHo.
Whether under standard conditions or not, the heats of reactions can often be measured using
an apparatus called a calorimeter. One of the most useful kinds of heats of reaction to measure
and tabulate is the standard enthalpy of formation of a substance, ΔHof, the enthalpy involved in
the formation of one mole of a substance from its elements under standard conditions. For
example, ΔHof for H2O(l) is defined by the following thermochemical reaction:
This indicates that under standard conditions the formation of one mole of liquid water from its
elements liberates 285.83 kJ in an exothermic reaction. By definition, the standard enthalpies of
formation of all elements in their standard states are taken to be zero. Therefore, ΔHof = 0 for
both H2(g) and O2(g). The reason for tabulating standard enthalpies of formation is that the data
can be used to calculate the enthalpy of any reaction under standard conditions, if we know the
ΔHof values for all the reactants and products. For any reaction under standard conditions, the
enthalpy of the reaction, ΔHorxn, is the sum of the standard enthalpies of formation of the
products, multiplied by their stoichiometric coefficients in the reaction equation, minus the sum
1
of the standard enthalpies of formation of the reactants, multiplied by their stoichiometric
coefficients in the reaction equation. Mathematically, this can be stated as
where np and nr are the stoichiometric coefficients of each of the products and each of the
reactants, respectively. For example, for a reaction aA + bB → cC + dD, we would have
One of the heat changes we will measure in this experiment results from the reaction of
metallic magnesium metal with dilute hydrochloric acid to form hydrogen gas and a solution of
magnesium chloride:
The appropriate data from the short table of standard enthalpies of formation shown below can
be used to calculate the value of ΔHorxn for this reaction.
HCl(g) –92.30
HCl(aq) –167.2
MgCl2(s) –641.6
MgCl2(aq) –796.9
H2O(l) –285.83
Your calculated value of ΔHorxn for this reaction is -462.5 kJ per mole of Mg(s) reacted.
This reaction is quite exothermic.
PROCEDURE
continuous stirring of any solution, using the spinning magnet inside the motorized stirrer
underneath the cups. The thermometer probe should fit snugly in the hole in the cover, and its
tip should be just high enough above the bottom of the cup to avoid being hit by the spinning stir
bar.
qcal = CcalΔT
where qcal is the heat change for the calorimeter, Ccal is the heat capacity of the calorimeter (the
apparatus and its contents), and ΔT is the temperature change of the calorimeter, defined as the
final temperature (Tf, after the heat change) minus the initial temperature (Ti, before the heat
change); i.e.,
ΔT = Tf – Ti
In precise calorimeter studies the heat capacity of the apparatus is carefully obtained as an
instrument constant. To arrive at a working value of the heat capacity of our coffee-cup
calorimeter, we will need to make some simplifying assumptions. Expanded polystyrene is a
good insulator, which is why it keeps hot or cold liquids from changing temperature rapidly.
However, polystyrene itself has a relatively low heat capacity, so very little heat is needed to
change the temperature of the inside of the cup to that of the liquid it contains. Therefore, we
will assume that the heat capacity of the apparatus is essentially that of the liquid in it.
3
Heats of Reaction
Moreover, in this experiment we will be creating dilute solutions, whose heat capacities are not
significantly different from pure water, which has a specific heat of 4.184 J/oC·g. In each case,
then, we will obtain the heat capacity of the calorimeter as the product of the mass of the
solution in the cup times the specific heat of water:
In a precision laboratory calorimeter, the temperature change, ΔT, is measured as the change
for the water used to surround the reaction chamber and fill the apparatus. In our experiments,
we can obtain the ΔT value by measuring the temperatures of the liquid in the cup before and
after the chemical reaction or physical process we are studying, using the digital thermometer. If
the temperature of the liquid in our coffee-cup calorimeter rises as a result of a chemical reaction
or physical change, the calorimeter itself has absorbed heat, an endothermic heat change. By the
First Law of Thermodynamics, heat is not lost, but rather is transferred from one body to
another. This means that the endothermic heat change of the calorimeter is the result of an
exothermic reaction or process, which gives off the same amount of heat as the calorimeter takes
in. In the same manner, if the temperature of the calorimeter falls as a result of some chemical
reaction or physical process, the calorimeter has experienced an exothermic change in which it
has given an equal amount of heat to the reaction or process. In this case, the chemical reaction
or physical process is endothermic. In general, then, the heat change for the calorimeter is equal
but opposite in sign to the heat change for the chemical reaction or physical process taking place
in it; i.e.,
qcal = –qrxn.
4
Heats of Reaction
When an ionic compound is dissolved in water, the resulting solution may be warmer or
cooler than the initial temperature of the pure water, depending on the particular ionic compound
dissolved. The heat of solution of a substance is defined as the heat absorbed or liberated when
one mole of solute is dissolved in an infinite amount of water. The phrase “infinite amount of
water” simply means an amount of solvent so large that further addition of solvent causes no
measurable change in the observed heat of dissolving the solute. The conditions in the following
procedure only approximate infinite dilution. As a result, the measured heat of solution will be
slightly lower than the actual value at infinite dilution.
1. On a triple-beam balance, determine the mass of one polystyrene cup with the stir bar placed
inside it. Record this mass to ±0.01 g on the data sheet.
2. With a graduated cylinder, measure 100 mL of deionized water, and transfer it to the
previously weighed polystyrene cup. On the same triple-beam balance used in step 1,
determine the mass of the cup, stir bar, and water. Record this mass to ±0.01 g on the data
sheet. Determine the mass of water by subtraction.
3. On a triple-beam balance, determine the mass of an aluminum weighing dish. Add to the
weighing dish a sample of potassium nitrate whose mass is in the range of 2.0 to 2.5 g.
Record the mass of the weighing dish and sample to ±0.01 gram on the data sheet.
Determine the mass of the sample by subtraction.
4. Assemble the coffee-cup calorimeter as shown in Fig. 1, being careful not to spill any of the
previously weighed water. Start the stirrer motor, and adjust the speed to obtain vigorous
stirring without splashing. Record the temperature of the water in the polystyrene cup just
prior to adding the sample of potassium nitrate (next step). This is the initial temperature, Ti.
5. Quickly add all of the potassium nitrate sample to the water (tap the bottom of the weighing
dish, if necessary), and immediately cover the cup with the lid. Continuously stir the mixture
for about one minute until the temperature remains constant for 15 seconds. Record this as
the final temperature, Tf, on the data sheet.
6. Check to be sure that all the solid potassium nitrate dissolved. If it did, carry out the
remaining calculations on the data sheet. If it did not, repeat the experiment.
5
Heats of Reaction
Partner(s)_______________________________________
DATA SHEET
6
Heats of Reaction
B. Heat of Neutralization
Whenever a strong acid and strong base are mixed, a neutralization reaction occurs in which
hydronium ion, H3O+, from the acid reacts with hydroxide ion, OH–, from the base to form
water:
The heat of neutralization for this reaction is a measure of the heat change when one mole of
hydronium ion reacts with one mole of hydroxide ion.
1. Empty your calorimeter, and rinse the inner polystyrene cup, stir bar, and thermometer probe
with deionized water. Allow the excess water to drain from the cup, then dry the cup, stir
bar, and thermometer probe with a paper towel.
2. Assemble the coffee-cup calorimeter as shown in Fig. 1, and place the stir bar in the inner
cup.
4. With another graduated cylinder (not previously used for hydrochloric acid solution),
measure between 30 and 40 mL of 1.00 M sodium hydroxide solution, and record the exact
volume to 0.1 mL on the data sheet.
5. Start the stirrer motor, and adjust speed to obtain vigorous stirring without splashing.
Measure and record the temperature of the hydrochloric acid solution in the cup just prior to
adding the sodium hydroxide solution (next step). This is the initial temperature, Ti. (The
temperature of the sodium hydroxide solution can be assumed to be the same, because both
solutions have been standing in the laboratory for the same amount of time.)
6. Add all of the sodium hydroxide solution to the cup containing the hydrochloric acid
solution, and place the lid over the cup. Continue stirring until the temperature becomes
constant. Record this as the final temperature, Tf, on the data sheet.
7
Heats of Reaction
Partner(s)_______________________________________
DATA SHEET
B. Heat of Neutralization
8
Heats of Reaction
In this experiment you will measure the heat of the reaction of metallic magnesium metal
with 1.00 M hydrochloric acid solution to form hydrogen gas and a solution of magnesium
chloride:
Mg(s) + 2HCl(aq) → H2(g) + MgCl2(aq)
1. Empty the previous solution in your calorimeter into the appropriate chemical waste
container. Rinse the inner polystyrene cup, stir bar, and thermometer probe with deionized
water. Allow the excess water to drain from the cup, then dry the cup, stir bar, and
thermometer probe with a paper towel.
2. On a triple-beam balance, determine the mass of the inner polystyrene cup with the stir bar
placed inside it. Record this mass to ±0.01 g on the data sheet.
3. With a graduated cylinder, measure 75 mL of 1.00 M hydrochloric acid solution, and transfer
it to the previously weighed polystyrene cup. On the same triple-beam balance used in step
2, determine the mass of the cup, stir bar, and hydrochloric acid solution. Record this mass
to ±0.01 g on the data sheet. Determine the mass of the hydrochloric acid solution by
subtraction.
4. With sandpaper, clean a 15-cm strip of magnesium ribbon weighing about 0.1 g until it is
shiny. Shape the strip into a zig-zag pattern, so that it will fit the bottom of the cup when you
eventually add it to the hydrochloric acid solution (step 6, below). Then determine the mass
of the strip on an analytical balance to ±0.0001 g. Record this mass on the data sheet.
5. Assemble the coffee-cup calorimeter as shown in Fig. 1, being careful not to spill any of the
previously weighed hydrochloric acid solution. Start the stirrer motor, and adjust the speed
to obtain vigorous stirring without splashing. Record the temperature of the hydrochloric
acid solution in the polystyrene cup just prior to adding the sample of magnesium (step 6,
below). This is the initial temperature, Ti.
6. Add the previously weighed magnesium strip to the cup containing the hydrochloric acid
solution, and place the lid over the cup. Continue stirring until the temperature becomes
constant. Record this as the final temperature, Tf, on the data sheet.
8. Clean up by emptying the contents of your calorimeter into the appropriate chemical waste
container. Rinse the cup, stir bar, and thermometer probe with deionized water. Allow the
excess water to drain from the cup, then dry the cup, stir bar, and thermometer probe with a
paper towel. Reassemble the apparatus so it is ready for the next laboratory section.
9
Heats of Reaction
Partner(s)_______________________________________
DATA SHEET
10
Heats of Reaction
Lab Report:
Your lab report will consist of your data sheets (pgs 6, 8 and 10) and a discussion of your results
in paragraph form. Points to be addressed in your discussion.
11