D-FL 100
Volume flow
measuring system
Measuring system to measure flow
rate in dry emissions with a probe
using the differential pressure
principle
Features Measuring principle System components
●● Reliable measurement of the gas velocity The D-FL 100 measuring system operates ●● Flow probe
even at high temperatures according to the differential pressure principle. ●● Mounting flange
●● Calculation of volume flow at standard The probe has two separate chambers between ●● Differential pressure transducer
conditions which the flow builds up a differential pressure. ●● Cross-over cock
●● Automatic zero check option The evaluation unit determines the gas velocity ●● Probe adapter
●● Certified, cost effective measuring system and the volume flow (norm conditions or ●● Evaluation unit D-FL100-20
●● Versions with or without counter-support standard conditions) from the differential ●● Absolute pressure transducer
and for point measurement pressure, taking into account gas temperature ●● Temperature transducer
●● Extremely low maintenance, maintenance and gas pressure. ●● Counter-support (option)
interval 6 months
●● Convenient operation via remote access Models Options
with web interface
●● D-FL 100 probe assembly ●● Universal operating unit D-ISC 100
Applications with assembly of measuring transducer ●● Service software D-ESI 100
on the probe ●● Weather protection covers
●● Volume flow measurement at high ●● D-FL 100 hose assembly ●● Automatic back purging device
temperatures with the measuring transducer connection ●● Special designs in other materials for
●● Plants with large or small flue cross-sections via hose line applications with particularly aggressive
●● Volume flow measurement at high pressure exhaust gases or high gas temperatures
Probes ●● Differential pressure transducer in
Approvals Ex-version
●● Probe 1: 0.4 ... 2 m
●● Suitability-tested by the TÜV Cologne, ●● Probe 2: 2 ... 4 m
test report 936/21218492/A ●● Probe 3: 4 ... 8 m differential pressure transducer
●● Approved and certified acc. to EN 15267-3 flow 4 ... 20 mA
●● MCERTS
evaluation
4 ... 20 mA unit
4 ... 20 mA
cross-over cock
absolute temperature
pressure transducer
transducer
flow probe
©DURAG GROUP 07/2016 · Subject to change without notice
flow direction
measurements flue gas velocity, volume flow digital outputs 2 relay outputs,
permissible load 48 V / 0.5 A
measuring ranges 0 ... 3000000 m3/h / 2 ... 50 m/s measuring outputs 0/ 4 ... 20 mA/ 500 Ohm, Modbus
RTU, RS485
measuring differential pressure zero point drift <0.5% of measuring range
principle
flue gas above dew point, power supply Sensor power supply 24 VDC ±10%,
temperature -20 ... 450 °C 0,5 A, 90 ... 264 VAC, 48 ... 62 Hz
(option)
flue gas pressure ±200 hPa
duct diameter 0.4...8 m dimensions probe: 380 x 160 x (300 + probe
(h x w x d) length) mm
ambient -20 ... +50 °C
temperature
protection IP65, Ex optional weight 32 kg + 6,8 kg/m probe length
DURAG GmbH Tel. +49 (40) 554 218-0 · Fax +49 (40) 584 154
Kollaustrasse 105 · 22453 Hamburg · Germany E-Mail: [email protected] www.durag.com