Genutech CCK
Genutech CCK
SURGICAL TECHNIQUE
1. TIBIAL TIME (I) 2
1.1 Intramedullary fixation 2
1.2 Assembly and fitting of the tibial cutting guide 4
1.3 Tibial resection 7
1.4 Measuring of the tibial tray 8
1.5 Measuring of the interline 9
2. FEMORAL TIME 10
2.1 Measuring of the femoral component 10
2.2 Intramedullary fixation 11
2.3 Distal cut 14
2.4 Femoral cuts 15
2.5 Verification of stability during flexo-extension 20
2.6 Reaming to house the femoral post 21
2.7 Assembly of the trial femoral component 23
3. TIBIAL TIME (II) 24
4. MOBILITY / STABILITY CHECK 29
5. PATELLA TIME 30
6. FINAL IMPLANT 31
ANNEXES 32
IMPLANTS AND INSTRUMENTS 44
Surgical Technique
Diameter of the reaming: The recommended diameters for this reaming are normally around Ø9-10 or 11mm, although the suitability of the final
diameter for the tibial diaphysis to be used is always at the discretion of the surgeon and, through gradual reaming, it may even be close to that
of the inner cortex diameter fitting. This is the case when a 200 mm stem must be inserted.
Depth: Generally, the diaphyseal reamer is inserted up to the second 70-120-200 mark, because at this depth greater stability, accuracy, tighte-
ning and safety are achieved for the installation of the tibial cutting guide. If it is not possible to penetrate this deeply (short tibia, fracture callus
which closes the medullary cavity…), at a minimum it must penetrate as far as the 70 mark in order to house a stem which is 70 mm long.
The diameter and depth are marked on the axis of the diaphyseal reamer, and as the tibial time is being performed, depths must be adjusted according
to the marks close to the initial T, depending on the length of the tibial stem (70, 120 or 200 mm) which you intend to implant.
2
Set 0. Lower tray
Depth: In order to install the 70 mm stem, ream until the first 70 mark is level with the tibia surface and if, on the other hand, the stem must be
102 or 200 mm, ream until the second 120-200 mark.
Diameter: Carry out gradual reaming in order to carve the cavity which will house the tibial stem.
Cavity reamed
to house
tibial stem
Note:
When using a 200 m stem, the diameters of the reamers (diaphyseal and cannulated) will be the same, because in order to reach a reaming depth of
200 mm, you use part of the length of the reaming which has already been completed to secure the diaphyseal reamer.
(See other examples of intramedullary reaming according to the length of the stem being implanted in Annex I)
3
Surgical Technique
The diameter of the sleeve being used is the same as the diameter of the last cannulated reamer which was used, installing a short metaphyseal sleeve
if the reaming was performed in order to insert a 70 mm stem, or a long one if reaming has reached the second mark (120-200) to insert a 120 or 200
mm stem.
The sleeves must be inserted with the revision sleeve inserter/extractor at a minimum depth, as indicated on the axis of the inserter, according to the
readings which correspond to the surgical time we are performing (tibia or femur) and the type of surgery being carried out (Semi-Constrained Revision
“REV” or Primary Semi-Constrained “PRIM”).
4
Graduation for
displacement control
Transverse connector
Fixation
screws
Tibial cutting
guide
Telescopic
guide system
Tibial
Tibial clamp cutting guide
In this position tighten the blocking screws or fixation screws of the transverse connector and of the telescopic guide, and then install the revision probe.
Transverse connector
Blocking
screws
Slot “0”
Reference “0”
Probe hook
5
Surgical Technique
In the most favourable cases, where you only need to make a thin cut to sanitise the tibial
surface, with the “0” end (short arm) of the probe, look for the most depressed area of the
tibial plateau to ensure that the cut being made is always below said point. To do this you
must use the threaded displacement measuring system to move it down until it rests on said
area. Once the most depressed area has been reached, fix the tibial cutting guide and then
make the cut.
In extreme cases, in which the tibia is extremely damaged, a prosthesis has been replaced
with supplements or it is necessary to sanitise a bone which is significantly (8 mm, 12 mm,
etc.). Using supplements, in the medial, lateral or even the bilateral area, proceed as follows:
The ”0” end (short arm of the tibial probe) must be directed towards the most depressed
area of the most prominent tibial plateau. To do this you must use the threaded displacement
measuring system to move it down until it is rests on said area.
Then note down the amount of millimetres which the system has gone down (see measure-
ment on the graduated marks on the telescopic guide).
With the arm of the rod in position “8” or “12” probe the most depressed area of the plateau
with the most wear, and the following situations may arise:
When performing primary tibial surgery by implanting a semi-constrained prosthesis tibial component, install the “0-10” rod on the tibial cutting guide.
Firstly, the “10” end of the probe must be directed towards the most depressed area of the most prominent plateau and then, with the arm of the rod
in position “0”, probe the most depressed area of the plateau with the most wear. If in order to probe on the most depressed plateau it is necessary to
move the system down, this measurement must be noted down.
Note:
For the “0-10” rod, which is only used for primary semi-constrained surgery, the “0” probe coincides with the deepest point of the rod, unlike the “0-8”
and “0-12” rods used for revision surgery.
Once the tibial cutting guide is in the correct position, it is fixed to the tibia using pins. Then the telescopic intramedullary system is removed (loose-
ning all of the fixation screws, moving to the end of the stroke of the thread of the tibial cutting guide and pulling up the telescopic system), and the
intramedullary system is removed.
This is when the tibial cut is made.
First line of
parallel pins
Oblique pin
7
Surgical Technique
Note:
· Should there be any difficulties when removing the intramedullary system because of an obstructing metaphyseal sleeve, this can be reamed with the
tip of the sleeve inserter/extractor and it can be released by tapping it outwards.
· It is recommended to insert the parallel pins on the first line of pins, using at least three pins and one oblique pin to ensure effective fixation.
· It is recommended that you first put in place the pins at both ends, so that the central pins do not impede the removal of the intramedullary compo-
nents and then, once the intramedullary system components have been removed (diaphyseal reamer and metaphyseal sleeve), install the central pins.
· The parallel pins can be kept, by way of a reference point, to make it possible to make the 2 mm recut.
· If after removing the pins we want to recover this reference point, it is recommended to put back the cutting guide by inserting the probe hook through
the corresponding cutting slot, supporting it on the resected surface and making use of the drill holes which have already been made in order to reinsert
the pins and thus fix the cutting guide in the right position. For greater accuracy when putting back the cutting guide it is recommended to perform the
above action by reassembling the extramedullary telescopic system on the intramedullary system.
After making the cut, when placing the trial templates on the tibial cut, select the size which best suits the bone morphometry of the tibial cut which
has been made.
Support handle
8
Set 2. Upper tray
Example:
If “a” is the height from the insertion of the patellar ten-
don with the ATT to the tibial osteotomy, and knowing that
the thickness of the tibial tray is approximately 4 mm and
the interline is approximately 25 mm from the insertion of
the patellar tendon with the ATT, we can conclude that the
thickness of tibial insert “I” to be used must be as follows
in order to maintain the interline:
(see image)
Tibial Tray
Tibial Cut
Note:
When it is necessary to perform an osteotomy of the TTA to make it possible to dislocate the patella, the height of this insertion will be marked on the
front side of the metaphysis so that measurements can be taken from this mark.
It is essential that the interline is located in the right place in order to ensure proper knee function.
9
Surgical Technique
Primary surgeries: using the primary femoral sizing guide which will indicate the correct size, supporting the posterior condyles and tapping with the
rod on the anterior cortex. In this case the sizing guide is inserted with the diaphyseal reamer through the centring sleeve. So for primary surgeries,
ream with the diaphyseal reamer and then measure the size of the femoral component.
Note:
· At this moment, check and assess the size which was
previously selected for the tibia (see 1.4.)
· When the sizing gauge is positioned between two sizes,
the smaller size must be selected.
10
Set 0. Upper tray
Diameter of the reaming: The recommended diameters for this reaming are normally around Ø9-10 or 11 mm, although the suitability of the final
diameter for the femoral diaphysis to be used is always at the discretion of the surgeon and, through gradual reaming, it may even be close to that
of the inner cortex diameter fitting. This is the case when a 200 mm stem must be inserted.
Depth: Generally, the diaphyseal reamer is inserted up to the second 70-120-200 mark, because at this depth greater stability, accuracy, tighte-
ning and safety are achieved for the installation of the femoral cutting guide. If it is not possible to penetrate this deeply (short femur, fracture
callus which closes the medullary cavity…), at a minimum it must penetrate as far as the 70 mark in order to house a stem which is 70 mm long.
As the femoral time is being performed, the depths must be adjusted to the marks close to the initial F, as indicated in the image below.
Depth: To install a 70 mm stem, penetrate as far as the first 70 mark. If, on the other hand, the stem must be 102 or 200 mm, ream until the second
120-200 mark. As this is the femoral time, pay attention to the depth marks closest to the initial F.
11
Surgical Technique
Diameter: Carry out gradual reaming in order to carve the cavity which will house the femoral stem.
F-70 F-70
Cavity to house
stem
Note:
When using a 200 m stem, the diameters of the reamers (diaphyseal and cannulated) will be
the same, because in order to reach a reaming depth of 200 mm, you use part of the length of
the reaming which has already been completed to secure the diaphyseal reamer.
(See other examples of intramedullary reaming according to the length of the stem being
implanted in Annex I)
12
Set 0. Lower tray
The diameter of the sleeve being used is the same as the diameter of the last cannulated reamer which was used, inserting a short metaphyseal sleeve
if the reaming was performed up to the 70 mark, or a long one if reaming has reached the 120-200 mark.
The sleeves must be inserted with the revision sleeve inserter/extractor, with its threaded cap which protects the extracting thread, at a minimum
depth, as indicated on the axis of the inserter, according to the readings which correspond to the surgical time we are performing (femur) and the type
of surgery being carried out: Semi-Constrained Revision or Primary Semi-Constrained.
Note:
It is recommended to insert the sleeve with its threaded cap (to protect the extracting thread).
When inserting a sleeve of Ø10 mm the inserter must be used without the cap because if this protective cap is used it will exceed the 10 mm
diameter of the sleeve.
13
Surgical Technique
On the axis of the diaphyseal reamer insert the distal cutting mask, which has been installed on the 5° angular guide which crosses the axis.
Then insert the mask until it comes into contact with the distal part of the femur, supporting the mask’s upper visor on its front and secure with threa-
ded pins of Ø 4.4 mm.
Two options can be seen on the cutting mask, which are marked on the cutting slots. Through these you will take into account the readings on the left
to perform “Primary Semi-Constrained” surgery and those on the right for “Semi-Constrained Revision” surgery.
In the case of “Primary” surgery, cut the support area to 8 mm, and for “Revision” surgery cut to 1 mm. We will also find slots marked for additional
recuts of 4, 8 and 12 mm, to enable sanitising or on condylar cuts. In these cases we will use distal supplements with the same thickness as the distal
cut which has been made.
Note:
Before fitting the mask and making the cuts it is advisable to take into account its degree of rotation, perpendicularly aligning the mask with
the tibial axis. This is performed by inserting the “Alignment Bar” through the hole of the “5° Angular Guide”, aligning it with the “tibial crest”.
Alternatively, the mask can also be guided using the femoral epicondyles.
14
2.4 Femoral cuts
If it is being supported offsetting is not necessary. If it is not being supported remove the concentric positioner and put the eccentric positioner in its
place, rotating it to the left or right until the rod is resting on the frontal area of the femur.
15
Surgical Technique
Set 3.
Once the rod is touching it, externally rotate the cutting mask by 3°.
Note I:
The medial-lateral displacement resulting from this off-
setting must be carried out as deemed appropriate by
the physician, seeking to centre the femoral component,
although it must be noted that rotating towards the ex-
ternal part has the advantage of reducing the tension on
the patella and is conducive to it being centred on the
femoral component.
Note II:
The cutting mask has housings for 4, 8 and 12 mm dis-
tal supplements, which must be installed to provide pro-
per distal support to the cutting mask, mainly in those
cases where recuts have been necessary on the distal
condyles.
3° of external rotation
With the knee at 90° flexion, and the posterior condyles resting on the surface of the tibia which has previously been cut, insert the 3° external rotation
system into the posterior cutting slots of the cutting guide.
Fit the alignment bar into the hole 3° to the left or right. When this bar is aligned with the axis of the tibia, externally rotate the cutting guide by 3°.
This support also has a hole at 0° which may be used at the discretion of the surgeon.
16
Alignment Alignment of bar with
bar tibial crest
Tibial axis
17
Surgical Technique
Note:
It is recommended to insert the thread as far as possible, only leaving one or two threads outside the cortex to ensure that the system has optimal
fixation and stability. The pins have a reference slot which indicates the maximum depth to which the pins should be inserted to avoid problems
when extracting the diaphyseal reamer and the sleeve from inside the intramedullary axis.
If we are making a 12 mm distal recut, the fixation of this condyle will be unstable if we only insert the pin as far as the slot, because only the tip
of the pin will reach the bone. If it is inserted deeper, this will present an obstacle when removing the diaphyseal reamer and sleeve. Therefore, it
is recommended to initially insert the pin as far as the slot and, once the intramedullary components have been removed, finish inserting the pin
until it penetrates the threaded area of the bone.
Set 3.
Making femoral cuts
Fit the appropriate cutting insert for the size of the femur and use it to make frontal, rear and bevel cuts.
Note:
The cutting mask makes it possible to make subsequent
additional cuts at 4 and 8 mm.
To facilitate the identification and correspondence between
sizes of masks and their corresponding cutting inserts, the-
re are colour markers which also coincide with the colour
which will subsequently be used for the trial tibial insert for
the flexo-extension test.
18
Intercondylar cut
When the intercondylar cut template has been fitted on the slots of the appropriate cutting insert for to the size of the femur, the cuts are made from
the front of the femur.
Intercondylar cut
template
In order to guide the saws better there is the option of installing the intercondylar insert of the appropriate size and frontally finishing the sides of the
intercondylar box.
With all of the femoral cuts made, remove the cutting mask.
19
Surgical Technique
Once the cuts have been made, spacers are used to check the looseness of the system and the alignment of the limb using the corresponding supple-
ments in extension and flexion.
To achieve stability, bilateral supplements are placed on the distal femur at 4, 8 or 12 mm on the femoral spacer (yellow piece) until it stabilises in
extension, or by increasing the thickness of the tibial spacer (blue piece) 2 by 2 mm, bearing in mind that, in both situations, whenever possible you
must compensate the increased thickness with the supplements on the femur to avoid increasing the thickness of the tibial polyethylene inserts, as
this thickness will have been established in order to maintain the position of the interline.
For example:
· If the necessary increase is 2 mm, the thickness of the tibial insert must be increased by 2 mm, because the Genutech CCK system does not have
2 mm femoral supplements.
· If the increase is 4 mm, 4 mm bilateral distal supplements are placed on the femur.
· If the increase is 6 mm, a tibial insert which is 2 mm bigger and 4 mm bilateral distal supplements are placed on the femur.
· Etc…
20
This femoral recutting mask is secured using pins of Ø 3.4 mm (the holes marked with arrows indicate the entry position for the inclined pins, which
are necessary to ensure the mask is properly secured).
The necessary distal and posterior supplements are fitted onto the intercondylar mask in order to impact it onto the femoral cuts which have been made,
securing it with 2 pins located on the front edge.
21
Surgical Technique
With the intercondylar mask fitted, create the housing for the femoral post by reaming up to the maximum depth possible with the Ø 16 mm reamer
designed for that purpose.
Note:
During this step you must be careful with the potential obstacles for the posterior cortex, above all at small sizes when the offset has been per-
formed close to the lower polar positions.
22
Set 4. Lower tray Set 5
Stem: it is recommended to go a size smaller than the reaming diameter, but this will be at the sole discretion of the surgeon. For 70 mm stems
it is recommended to use the same diameter, and for 120 and 200 mm stems, to use a size which is smaller than the reaming diameter to avoid
the “tip effect”. It is assembled by screwing the stem into the distal part of the femoral component, using the screws provided in Box 7 for that
purpose.
Supplements: its pre-mounted screw is screwed into the threaded holes of the femoral component (see Annex VI).
Position between the stem and the femoral component: if it has been necessary to perform an offset, fit the stem while ensuring that the mark of
the stem coincides with the number displayed on the femoral component which originates from the reading which was last taken with the eccentric
positioner on the insert positioner (see 2.4 Reading of the Offset).
Using the femoral component inserter/extractor, insert the femoral component into the femur, making the final adjustments to adapt it to the femoral
cuts. Once inserted, check that it has been adjusted correctly.
Then remove the femoral component, ensuring that the marked number has not changed position, and said offset position will be transferred to the
final prosthesis.
23
Surgical Technique
Optionally, at the discretion of the surgeon, before removing the femoral component, the trial insert of a suitable size is fitted on the tibial tray and
flexion and extension movements are performed to adapt the rotation of the tibial component to the femoral component.
In a suitable place, mark the tibia with an electric scalpel, a mark in the centre of the tibial tray which will help to select the rotational positioning of
the tibial component.
On its axis install the concentric chisel, and on this place the tibial template of the established size (see 1.4). Furthermore, it must be ascertained
whether the template’s support on the surface of the tibial cut is well contained and centred or, on the contrary, it is off-centre.
In the first case a stem with an offset is not necessary. In the second case it is, as can be seen in the image, the tray is of a suitable size but it is jutting
out, so it would be advisable to move it forwards.
24
In this situation observe which number is showing on the measuring system opposite the longitudinal mark on the chisel. This is the position in which
the stem must be fixed to the tibial tray.
In this position the tibial template is fixed to the tibia with pins. At this moment you can optionally perform a flexo-extension check, by fitting the trial
insert onto the tray, taking care to use the housings marked on the template for pins.
25
Surgical Technique
Then create the housings for the fins of the keel through a keel impactor which will be placed on the eccentric or concentric keel.
26
Then, remove the entire system.
In the figure below it is possible to see that the intramedullary axis and the housing for the post of the tibial tray are not properly centred, due to the
offset.
If tibial supplements have been used, use the hemi-keel (right or left) to assist with the anterior-posterior cut which is made around the two plateaus
so that there is no obstruction for the tibial supplements.
27
Surgical Technique
It is then impacted and the corresponding tibial insert is fitted (see annexes VII and VIII).
28
4. MOBILITY / STABILITY CHECK
Trial components
The check is carried out with the trial femoral and tibial components fitted to assess the mobility, stability and alignment of the limb.
29
Surgical Technique
Leave a minimum bone thickness to avoid the risk of a fracture (12-14 mm).
Avoid the hyper pressure which would be caused by a prosthetic knee which is thicker than the original. Therefore, when choosing the size of the
prosthetic patella, you do not only have to consider the diameter which is best suited to the resected bone surface, but also the height of each
patella component measurement:
Diameter 32 mm 34 mm 36 mm 38 mm 40 mm
Height 7 mm 8 mm 9 mm 10 mm 11 mm
The mobile arm of the clamp is positioned so that the perforation of the prosthetic pivot is slightly medialised in relation to the centre of the patella.
The perforation is performed with the specific drill bit.
When the trial prosthesis is in place, its travel along the femoral intercondylar channel is assessed with flexo-extension movements.
Note:
If the trial (or prosthetic) patella increases the tension of the extensor apparatus, and therefore the chances of dislocation and wear, a smaller
patella component will be selected to reduce its height.
30
6. FINAL IMPLANT
Final check
Fit the final prosthesis (tibial, femoral and patella component) but, prior to fixing the polyethylene insert to the tibial tray with bolts, perform the mobility
and stability test again.
After performing the check and ensuring that it is to be used as the final insert, it is finally fixed to the tibial tray, inserting the metal-reinforced post
of the tibial insert and anterior safety bolt.
Example:
31
Surgical Technique
ANNEX I
EXAMPLES OF DIAPHYSEAL AND CANNULATED REAMING ACCORDING TO THE STEM TO BE
IMPLANTED
TIBIA
Insertion of 70 mm stems in tibia.
T-70 T-70
T-70 T-70
T-70 T-70
T-70 T-70
32
Insertion of 120 mm stems in tibia.
T-120 T-120
T-120 T-120
T-200 T-200
øC
øC= øD
T-200 T-200
øD
The diameters of the diaphyseal reamer and cannulated reamer must be the same, because the 200 mm stem uses part of the depth created by the
diaphyseal reamer.
33
Surgical Technique
FEMUR
Insertion of 70 mm stems in femur.
F-70
F-70
34
Insertion of 70 mm stems in the femur (greater intramedullary stability).
F-70
35
Surgical Technique
F-120
F-120
36
Insertion of 200 mm stems in femur.
F-200
F-200
øC= øD
øC
øD
The diameters of the diaphyseal reamer and cannulated reamer must be the same, because the 200 mm stem uses part of the depth created by the
diaphyseal reamer. 37
Surgical Technique
If the surgeon believes that there is a possible risk of the diaphyseal reamer being fully inserted into the intramedullary cavity of the bone when can-
nulated reaming is performed, he or she has a retainer cable for these diaphyseal reamers.
Given that the instruments used for the following steps are cannulated, using it will not alter the surgical steps described in these instructions.
ANNEX III
T-HANDLE FOR MANUAL REAMING
The surgeon has a T-handle with quick coupling for “AO hexagonal” connectors to enable manual diaphyseal and cannulated reaming.
Note:
Diaphyseal reamers require the “Tri-lobular > AO-HEX” Adaptor
38
ANNEX IV
WRENCH (14 mm wide)
If vertical movement of the tibial cutting guide is harder than usual, there is a wrench which enables you to easily rotate the sheave of the tibial
cutting guide.
ANNEX V
PIN IMPACTOR
The pin impactor can also be used as an extractor rod, passing it through holes made for that purpose for instruments which, because of the way in
which they are operated, can become stuck or difficult to extract (keel impactors, femoral component inserter/extractor, etc.).
39
Surgical Technique
ANNEX VI
BALL-HEAD SCREWDRIVER OF Ø 2.5 mm
To facilitate the screwing of the posterior supplements into the femoral components, the Genutech CCK set of instruments has a ball-head screwdriver
(Box 6 – Upper tray) to enable oblique screwing/unscrewing.
40
ANNEX VII
CLIPPING AND EXTRACTION OF TIBIAL INSERTS
Clipping is performed manually, by first resting the back of the insert on the tibial tray and then pressing both pieces until they clip together.
To extract the tibial insert, insert the flat tip of the extractor in one of the slots which is left in the tibial insert when it has been clipped into the tibial
tray, turning it slightly (do not pry open).
41
Surgical Technique
ANNEX VIII
TIBIAL IMPACTORS
The Genutech CCK set of instruments (Box 6 Upper tray) contains a tibial tray impactor and a second impactor which is designed to enable the fully
assembled tibial component to be impacted, protecting both the tibial post and the surface of the joint. This impactor of the tibial component can also
help to clip the tibial insert into its tray in the (unlikely) event of it being impossible to do so manually.
The Genutech CCK set of instruments (Box 7 Lower tray) contains a patellar compression clamp to cause bone cement overflow and maintain pressure
during the necessary setting time so that the patella is properly fitted.
43
Implants
Cemented femoral
revision component
Ref. D8033100E 1
Ref. D8033200E 2
Ref. D8033300E 3
Ref. D8033400E 4
Ref. D8033500E 5
44
Revision tibial insert
10 mm 18 mm
12 mm 20 mm
14 mm 22 mm
16 mm 24 mm
18 mm 10 mm
20 mm 12 mm
22 mm 14 mm
24 mm 16 mm
10 mm 18 mm
12 mm 20 mm
14 mm 22 mm
16 mm 24 mm
18 mm
20 mm
22 mm
24 mm
10 mm
12 mm
14 mm
16 mm
Distal femoral
supplement
Posterior femoral Ref. D8026150E 1 4 mm
supplement Ref. D8026190E 1 8 mm
Ref. D8026010E 1 4 mm Ref. D8026195E 1 12 mm
Ref. D8026015E 1 8 mm Ref. D8026250E 2 4 mm
Ref. D8026020E 2 4 mm Ref. D8026290E 2 8 mm
Ref. D8026025E 2 8 mm Ref. D8026295E 2 12 mm
Ref. D8026030E 3 4 mm Ref. D8026350E 3 4 mm
Ref. D8026035E 3 8 mm Ref. D8026390E 3 8 mm
Ref. D8026040E 4 4 mm Ref. D8026395E 3 12 mm
Ref. D8026045E 4 8 mm Ref. D8026450E 4 4 mm
Ref. D8026490E 4 8 mm
Ref. D8026495E 4 12 mm
Tibial supplement
Ref. D8032710E 1 8 mm
Ref. D8032715E 1 12 mm Patellar component
Ref. D8032720E 2 8 mm
Ref. D8030140E ø 32 mm
Ref. D8032725E 2 12 mm
Ref. D8030150E ø 34 mm
Ref. D8032730E 3 8 mm
Ref. D8030160E ø 36 mm
Ref. D8032735E 3 12 mm
Ref. D8030170E ø 38 mm
Ref. D8032740E 4 8 mm
Ref. D8030180E ø 40 mm
Ref. D8032745E 4 12 mm
Ref. D8032750E 5 8 mm
Ref. D8032755E 5 12 mm
45
Set of Instruments
Set of Instruments
*
Complete Set of Genutech ® CCK Revision Instruments
Ref. D8501000
6 Revision diaphyseal
Ref. D8271420 reamer
5
Diameter
2 IM awl 1 Ref. D8171009 ø 9 mm
Ref. D8171010 ø 10 mm
Ref. D8210010S
Ref. D8171011 ø 11 mm
3 Quick coupling T-handle key 2 Ref. D8171012 ø 12 mm
AO-HEXAGONAL 6
Ref. D8171013 ø 13 mm
Ref. D8271475 3
4 Ref. D8171014 ø 14 mm
4 Tri-lobular adaptor Ref. D8171015 ø 15 mm
Ref. D8171016 ø 16 mm
Ref. D8271410
Ref. D8171017 ø 17 mm
5 9 mm drill bit
Ref. D8171018 ø 18 mm
Ref. D8210020
Lower tray
1 Revision cannulated
reamer
Diameter 3 Inserter/extractor of revision sleeves
Ref. D8171110 ø 10 mm 1
2
Ref. D8171112 ø 12 mm Ref. D8271405
Ref. D8171114 ø 14 mm
4 Plug of inserter/extractor of revision
Ref. D8171116 ø 16 mm sleeves
Ref. D8171118 ø 18 mm Ref. D8271406
3 4
Ref. D8171120 ø 20 mm
2 Revision metaphyseal
sleeves
Diameter Length Diameter Length
Ref. D8271010 ø 10 mm 70 mm Ref. D8271110 ø 10 mm 110 mm
Ref. D8271012 ø 12 mm 70 mm Ref. D8271112 ø 12 mm 110 mm
Ref. D8271014 ø 14 mm 70 mm Ref. D8271114 ø 14 mm 110 mm
Ref. D8271016 ø 16 mm 70 mm Ref. D8271116 ø 16 mm 110 mm
Ref. D8271018 ø 18 mm 70 mm Ref. D8271118 ø 18 mm 110 mm
Ref. D8271020 ø 20 mm 70 mm Ref. D8271120 ø 20 mm 110 mm
* This complete set can be supplied in water and airtight containers to maintain their sterility, at the customer's request. Contact your distributor or manufacturer.
8
46
Case 1: Ref. D8307200 Genutech ® CCK Set 1. of tibial instruments
Upper tray
2 3 Ref. D8220140
Ref. D8277124
11 Revision tibial
4 Revision tibial cutting guide 5 Revision probe axle 6 Primary semi-constrained tibial probe template Size
Ref. D8270301 1
Ref. D8270510 Derecho Ref. D8270608 0-8 mm Ref. D8220270
Ref. D8270302 2
Ref. D8270520 Izquierdo Ref. D8270612 0-12 mm
7 Probe hook Ref. D8270303 3
Ref. D8270304 4
Ref. D8220290 Ref. D8270305 5
Lower tray
47
Set of Instruments
Set of Instruments
Upper tray
1 Revision interline sizing guide
Ref. D8277140
2 Revision 5° angular guide 4 4 Primary femoral sizing guide
1
Ref. D8271055 Ref. D8271450
3 Revision femoral 3
sizing guide 5 Distal cutting mask
Size
5
Ref. D8271441 1
Ref. D8271050
Ref. D8271442 2 2
Ref. D8271443 3
Ref. D8271444 4
Lower tray
2 Revision femoral spacer
1 Revision tibial spacer
Length
Ref. D8274008
Ref. D8274110 10 mm
Ref. D8274112 12 mm 3 Supplement for spacer
1
Ref. D8274114 14 mm Length
Ref. D8274204 4 mm
Ref. D8274116 16 mm
Ref. D8274208 8 mm
Ref. D8274118 18 mm
Ref. D8274212 12 mm
Ref. D8274120 20 mm
3
2 4
Ref. D8274122 22 mm 4 Revision distal recut mask
Ref. D8274124 24 mm
Ref. D8271455
Ref. D8271208 8 mm 12
Ref. D8271091 1
Ref. D8271212 12 mm
5 Ref. D8271092 2
10
6 Ref. D8271093 3
13
3 Revision Ref. D8271094 4
semi-constrained rod Size
7
14
Ref. D8271301 1
9
Ref. D8271302 2 8 13 Revision Intercondylar
cutting insert Size
Ref. D8271303 3 Ref. D8271081 1
Ref. D8271304 4 Ref. D8271082 2
6 Revision concentric chisel 9 Revision horizontal reference rod mask
4 Primary Ref. D8271083 3
semi-constrained rod Size Ref. D8271210 Ref. D8271435 Ref. D8271084 4
Ref. D8271251 1
Ref. D8271252 2
7 Revision eccentric chisel 10 3° rotation system 14 Chisel
Ref. D8271253 3
Ref. D8271254 4 Ref. D8271215 Ref. D8271310 Ref. D8210220
10
48
Case 4: Ref. D8307500 Genutech ® CCK Set 4. of femoral trial instruments
Upper tray
Ref. D8126030 3 4 mm 3
Ref. D8126290 2 8 mm
Lower tray
49
Set of Instruments
Set of Instruments
Upper tray
Lower tray
12
50
Case 6: Ref. D8307700 Genutech ® CCK Set 6. of tibial trial instruments
Upper tray
1 Revision trial tibial tray
Size
Ref. D8133100 1
3 Tibial tray impactor
Ref. D8133200 2
1
Ref. D8133300 3 Ref. D8220615
Ref. D8133400 4
Ref. D8133500 5 4 Tibial component impactor
2
13
51
Set of Instruments
Set of Instruments
1 Pin extractor 5 Headless pin 7 Fixation screw for trial 9 Tornillo suplemento tibial
Diameter Length tibial tray de prueba
Ref. D8220280 Ref. D8210166 ø 3,4 mm 80 mm Ref. D8232600 Ref. D8132760 8 mm
Ref. D8132762 12 mm
10 AO drill adaptor
3 4
3 Pin with head 5 6
Ref. A1700390 AOHEX
Diameter Length
8
Ref. D8210161 ø 3,4 mm 30 mm 9
6 Revision threaded
pin Diameter Length *
4 Headless support 8 Support handle for fixing bolt 12 Torque wrench 9N
pin Diameter Length Ref. D8210172 ø 4,4 mm 75 mm
Ref. D8210165 ø 3,4 mm 55 mm Ref. D8210171 ø 4,4 mm 100 mm Ref. D8220530 Ref. D8277250 4,5 mm
*The reference doesn´t correspond to
the image.
Lower tray
2 3
1 Drill bit for patella 2 Clamp for patellar resection 3 Clamp for patella 4 Trial patella
Diameter
Ref. D8230120 Ref. D8230110 Ref. D8230130 Ref. D8130140 ø 32 mm
Ref. D8130150 ø 34 mm
Ref. D8130160 ø 36 mm
Ref. D8130170 ø 38 mm
Ref. D8130180 ø 40 mm
52
0086
0318
IROGEKCA11 / 07-2015 © Rev. 1
El marcado CE es válido únicamente si también está impreso en la etiqueta del producto.