UN1250
UN1250
Instruction
Manual
MODEL 7010
GAUSS / TESLA METER
Instruction Manual
Manual UN-01-250
Item 359935
Rev. D, ECO 14211
All Rights Reserved.
F.W. BELL Model 7010 Gauss/Tesla Meter Instruction Manual
This symbol appears on the instrument and probe. It refers the operator
to additional information contained in this instruction manual, also
identified by the same symbol.
NOTICE:
Table of Contents
Section – 1 Introduction
Overview…………………………………………………………………………………….……….. 1-1
General Description…………………………………………………………………………………. 1-3
Applications…………………………………………………………………………………..…….… 1-8
Section – 2 Specifications
Instrument…………………………………………………………………………………………….. 2-1
Calibration Service…………………………………………………………………………………… 2-5
Zero Flux Chamber………………………………………………………………………………….. 2-6
Section – 3 Probes
Overview………………………………………………………………………………………...……. 3-1
Probe Variations………………………………………………………………………………….…. 3-2
Probe Memory……………………………………………………………………………………...… 3-3
Probe Stem…………………………………………………………………………………..……….. 3-3
Temperature Effects…………………………………………………………………………..…….. 3-3
Fixturing………………………………………………………………………………………………. 3-3
Section – 4 Setup
Safety………………………………………………………………………………………………… 4-1
Line Voltage Settings / Fuse Panel………………………………………………………………… 4-3
Adjusting the Handle………………………………………………………………………………… 4-4
Probe Installation…………………………………………………………………………………….. 4-5
Power Up……………………………………………………………………………………………… 4-6
i
F.W. BELL Model 7010 Gauss/Tesla Meter Instruction Manual
ii
F.W. BELL Model 7010 Gauss/Tesla Meter Instruction Manual
Warranty
List of Tables
iii
F.W. BELL Model 7010 Gauss/Tesla Meter Instruction Manual
List of Illustrations
iv
F.W. BELL Model 7010 Gauss/Tesla Meter Instruction Manual
Section 1 Introduction
OVERVIEW The F.W. BELL model 7010 gauss/tesla meter incorporates the latest
developments in magnetic flux density measurement technology with a
modern user interface. It features a large display for easy viewing, a
comprehensive keypad for control of common functions, and an easy-to-use
menu system. The model 7010 has high accuracy suitable for use in the
laboratory, and enough features to be versatile in a manufacturing
environment. Table 1-1 on the following page provides a list of features.
GENERAL The Model 7010 gauss/tesla meter utilizes a Hall effect probe to
DESCRIPTION measure magnetic flux density in units of gauss (G), tesla (T),
amp/meter (A/m), or oersted (Oe). Either steady-state (dc) or
alternating (ac) fields may be measured. Fields as low as 10 µGauss
(0.001 µT) or as high as 300k gauss (30 tesla), at frequencies up to
50 kHz, can be measured with extreme accuracy and 5-3/4 digit
resolution. The instrument is calibrated and linearized from data
stored within its probe. With a temperature compensated Hall probe,
the instrument can compensate for errors due to temperature
variations.
User Interface
Auto Range
GENERAL Zero
DESCRIPTION
(Continued) The “zero” function allows the user to remove undesirable readings
from nearby magnetic fields (including earth’s) as well as to remove
initial electrical offsets in the probe and instrument. A “zero flux
chamber” is included as an accessory which shields the probe from
external magnetic fields during this operation.
Hold
When the hold function is enabled the instrument will “hold” and
display the highest and/or lowest flux density readings that have been
measured. Hold features include capturing peaks and valleys of rapid
changing pulses as well as arithmetically calculated max and min of
slow changing signals.
Relative
Update Interval
Analog Output
GENERAL Standard full scale output ranges are 3V, 10V, 3VRMS, and 10VRMS. An
DESCRIPTION adjustable full scale up to 9.9V or 9.9VRMS, in increments of 0.1V, is
(Continued) also available. These outputs may be connected to a voltmeter,
oscilloscope, recorder, or external analog-to-digital converter.
Analog Filters
A low pass filter is available that may be set to pass frequencies only
below 50kHz, 5kHz, or 500Hz. The instrument can automatically
select the best filter setting based on the present flux density being
measured. The filters affect both the displayed reading and the
analog outputs.
Field Classifier
The “Classifier” function allows the user to define a lower and upper
limit of flux density that can be used to quickly determine the status of
a magnetic field. The instrument will indicate visually whether the
field is below, within, or above the pre-defined limits. The same
information is provided in the form of general purpose switch closures
available at a standard 15 pin “D” type female connector.
Remote Operation
Accessories
(1) Display 320 x 240 Pixel Electro-luminescent ¼ VGA display, Multi-Shade Amber Color
(2) Power Switch Push Button Type Power Switch
(3) Menu Key Used to Enter and Leave the Menu System
(4) Direction Keys Used to Navigate the Menu System
(5) Enter Key Activates / De-activates Selections in the Menu System
(6) Range Key Selects Fixed Ranges or Autorange
(7) AC/DC Key Selects ac or dc Field Measurement
(8) Hold Key Activates / Deactivates the Hold Feature
(9) Reset Key Resets the Min/Max and Peak/Valley Detectors Used with the Hold Feature
(10) Relative Key Activates / De-activates the Relative Function
(11) Manual Adjust Adjusts the Relative or Zero Setting Up and Down
(12) Zero Key Starts the Zeroing Process
(13) Probe Connector 12 Pin Non-Magnetic Female Connector for Hall Effect Probes
GENERAL Cabinet
DESCRIPTION
The cabinet is equipped with a rotating handle that also serves to
(Continued) adjust the tilt angle of the instrument. If necessary, the handle is easily
removed with a standard Phillips screwdriver. The rear of the housing
features protruding feet that allow the instrument to be placed down in
a vertical position, without damage to the rear panel. Figure 1-3
shows the instrument in various positions.
Section 2 Specifications
INSTRUMENT
Full-scale ranges are shown in the Tables 2-1a – 2-1c below; listed by probe type. In all cases, the
resolution is 1 part in 300,000.
3V range: 0.25% of
±0.05% of reading and reading ±40mV
±0.15% of 3V or 10V ranges
dc accuracy ±0.01% of range
10V range: 0.25% of
reading±120mV
ac accuracy 2% of range 2% of range
In dc mode
N/A
dc to 100 hz dc to 100Hz
2.0% of 3V or 10V ranges See Figure 2-1 for Graph
ac accuracy 2.0% of reading ± 0.15% of
(ac 20 to 500 Hz) (Typical)
in ac mode range (20 Hz to 50 kHz)
(ac rms (dc) 20Hz-50kHz) ac and ac rms (dc) output
ac peak accuracy 5.00% of Reading N/A N/A
*Probes Errors Not Included
5.00
0.00
1 10 100 1000 10000 100000
-5.00
PERCENT OF READING ERROR
-10.00
-15.00
-20.00
-25.00
-30.00
-35.00
FREQUENCY (Hz)
Communications
Serial
Format: RS-232C
Connector type: 9-pin “D” female
Cable length: 3 m (9.8 ft.) maximum
Receive input resistance: 3 kΩ minimum
Receive voltage limit: ± 30 V maximum
Transmit output voltage: ± 5 V min, ± 8 V typical
Baud rate: 300, 600, 1200, 2400, 4800, 9600, 19200, 38400
Stop bits: 1, 2
Character length: 7,8
Parity: None, Odd, Even
Handshaking None, Hardware, Software
Standards supported: IEEE-1987.2, SCPI-1999
Digital I/O
Classifier Connections
Low 3,11
High 4,12
Regulatory Information:
Compliance was demonstrated to the following specifications as listed in the official Journal of the European
Communities:
OECO, LLC
4607 SE International Way
Milwaukie, OR 97222
Phone: 503-659-5999
Model Number:
YA-111
Cavity Dimensions:
Length: 50.8 mm (2”)
Diameter: 8.7 mm (0.343”)
Attenuation:
80 dB to 30 mT (300 G)
Purpose: Figure 2- 3
To shield the probe from external magnetic fields Zero Flux Chamber
during the ZERO or RELATIVE operations.
Section 3 Probes
OVERVIEW F.W. Bell’s 7000 series gauss/tesla meter probes are designed to meet
the electrical and mechanical requirements of virtually any application.
Models are available for transverse, axial, and very low field
measurements. The probe style is dependent upon the measurement
environment. The standard fiberglass stem is recommended for
laboratory or light handling environments, while the heavy duty aluminum
stem is recommended for harsher environments. The probe’s length,
outside diameter (axial probes) or thickness and width (transverse
probes) are important if there are physical constraints where the probe
will be used.
Handle Hall probes with care. Do not bend the stem or apply
pressure to the probe tip as damage may result.
PROBE A wide variety of probes are available for use with the 7010 gauss/tesla meters.
VARIATIONS The types include heavy-duty transverse and axial, standard transverse and axial,
standard transverse with exposed element, flexible transverse and axial with
exposed element and low field probes. All of these probes are available with or
without temperature compensation. All probes are available with 5, 15 or 30 foot
(1.5, 4.5 and 9 meter) cable lengths and most are available with various stem
lengths. Table 3-1 lists the maximum field measurement capabilities and
resolutions.
Figure 3-1 serves as an ordering guide for F.W. BELL 7000 series probes.
Full electrical and mechanical specifications of all probes are available on
request.
PROBE The connector of each probe contains a memory device which stores
MEMORY registration information (model number, serial number, date calibrated, etc.)
as well as performance information for Hall generator sensitivity, linearity,
frequency response and temperature response. Each probe is physically
identified with model number, serial number and a maximum voltage rating of
“30VRMS / 60Vdc MAX” on a durable polyester label wrapped around the
cable jacket.
PROBE STEM All probes except the low field probe are supplied with a rigid stem cover to
protect the probe when not in use. It is strongly recommended to use the
stem protector when storing the probe or when the probe will not be used for
any length of time. If a probe stem becomes damaged it can not be repaired.
TEMPERATURE All Hall probes have an initial electrical offset that will affect the accuracy of
EFFECTS static (dc) field measurements. This offset should be canceled using the
instrument’s “zero” function. However, the probe’s offset and sensitivity will
change with temperature. Using temperature-compensated probes will
minimize these effects.
Section 4 Setup
SAFETY
INSTRUCTIONS GENERAL:
For safe and correct use of this instrument it is necessary that both
operating and servicing personnel follow generally accepted safety
procedures plus the safety cautions and warnings specified.
CAUTION:
All input and output voltages, except line (mains), are less than 20V.
WARNING:
The opening of covers or removal of parts might expose live parts
and accessible terminals which can be dangerous.
WARNING:
The display generates voltages capable of causing personal injury
(high voltage pulses up to 195 Vac ). Do not touch the display
electronics during operation.
WARNING:
Any interruption of protective earth conductors or disconnection of
the protective earth terminals inside or outside of the instrument can
create a dangerous condition.
CAUTION:
For continued protection replace the fuse with the same type (IEC
127 type T).
SAFETY
INSTRUCTIONS
WARNING:
(Continued) The Hall probe is a non-contact measuring device. The probe is not to
contact a surface which exceeds a voltage of 30Vrms (42.4V peak) or
60V d.c.
CAUTION:
This instrument may contain ferrous components which will exhibit
attraction to a magnetic field. Care should be utilized when operating
the instrument near large magnetic fields, as pull-in may occur.
Extension cables are available to increase the probe cable length, so
that the instrument can remain in a safe position with respect to the
field being measured with the probe.
The model 7010 gauss/tesla meter has been factory-configured to operate on a line voltage of 120
Vac. See the following heading LINE VOLTAGE SETTINGS / FUSE PANEL if operation at 100,
220, or 240 Vac is required. Operating the instrument at 220 or 240 Vac without setting the
line voltage switch properly may result in damage to the instrument.
Follow these instructions to replace a fuse or change the line voltage settings.
1) See figure 4-2a. This is the power cord receptacle. It is designed to accept an international
instrumentation power cord. This receptacle also contains the fuse and voltage selector.
Remove the power cord. Near the top of the receptacle is a slot. Insert a narrow screwdriver
and gently release the access door (the door will not open unless power cord is removed). The
door will swing down.
2) See figure 4-2b. Remove the voltage selector drum by pulling straight out. Remove both fuse
holders by pulling straight out. The left-hand holder is used to hold a spare fuse. The right
hand fuse is the actual in circuit fuse. The instrument is shipped with a 220/240 Vac 0.5 Amp
fuse in the spare location, and a 100/120 Vac 1.0 Amp fuse in the active location.
3) See figure 4-2c. Reverse the positions of the holders and reinstall, making sure the “→” points
to the right. Rotate the voltage selector drum to read “220 or 240 (your countries nominal
mains voltage) Vac” and reinstall.
4) See figure 4-2d. Close the access door. The “220 Vac” legend should be visible through the
access door window. Reinstall the power cord.
See Figure 4-3. To adjust the position of the handle, press and hold the buttons on the sides. Then
Rotate the handle to the desired position and release the buttons. The handle will lock into
position every 30 degrees.
To remove the connector, grab at the collar and slide back. The probe
connector cannot be removed by pulling only on the connector body.
Note: You may install or remove a probe at any time, although it is not
recommended to do so during the zeroing operation.
POWER - UP Locate the power switch on the front panel. The On/Off positions are labeled
below the switch and are shown in Figure 4-5. To turn on the instrument
press the switch and release. The switch will remain depressed and the boot
up procedure will begin.
To shut the instrument off, press the power switch again and release.
Note: Allow three seconds for the instrument to be off before turning it
back on.
POWER – UP Initialization
(Continued)
During Boot up, the F.W. BELL logo will appear and a status bar indicates
boot up progress. See Figure 4-6. Internal diagnostics are performed first,
followed by the initial formatting of the display.
BIOS VERSION
CPU TYPE
STATUS BAR
Calibration data stored in the instruments memory and its probe are
retrieved next. The message “Reading Probe” is displayed during this
process. If no probe is connected, the message “No Probe” will be
displayed.
If any of the internal diagnostic tests fail, an error message will appear and
the instrument will halt the power-up procedure.
OVERVIEW Most of the features and functions of the 7010 are activated directly from
the front panel keypad. Other functions are accessed through the menu
system.
FRONT PANEL There are two sets of keys on the 7010 front panel. One set is associated
KEYPAD with the menu system. This set contains the MENU, ENTER, and
DIRECTIONAL keys. The other set is associated with flux density
measurement operations. This set contains the RANGE, AC/DC, HOLD,
RESET, RELATIVE, ZERO, and two MANUAL ADJUST keys. See Figure
5-1.
Each key features a back-light that will illuminate indicating that it is active.
In some instances, certain keys will flash indicating that a particular function
is enabled or that some condition requires attention.
MENU SYSTEM The menu system of the 7010 is very intuitive to use. It features many items
that are similar to those found in the operating systems of personal
computers. These items include the “pointer”, “buttons”, “checkboxes”,
“selections”, and simple text. The menu items are shown in Figure 5-2.
Button
Checkbox
Text
Selection
Pointer
Pointer
Buttons
“Buttons” are used to initiate a particular action such as entering into a sub-
menu. To activate a “button”, position the “pointer” over the “button” and
press the ENTER key.
Checkboxes
Numbers
Numbers act as groups of “selections” where the possible choices for each
digit in the number are the numerals 0-9.
Pressing the MENU key from the measurement screen will access the main
menu. The main menu is shown in Figure 5-4. To return to the
measurement screen, press the MENU key again.
To enter a sub-menu, position the “pointer” over a “button” and press the
ENTER key. To return to the main menu, press the MENU key.
USING THE As an option, the 7000 series supports the use of a Microsoft® compatible
MOUSE serial mouse. The mouse must use a female DB9 female connector. A 9-
pin male to 9-pin male gender changer and null modem adapter is also
required.
Note: Since the mouse uses the RS-232C serial communications port,
it is not possible to use the mouse and support RS-232C
communications at the same time.
The 7010 can be completely operated from the mouse. Pressing the right
mouse button is the same as pressing the MENU key on the front panel.
Moving the mouse will cause the “pointer” to move about the screen. In the
menu system, pressing the left mouse button is the same as pressing the
ENTER key.
Pressing the left mouse button from the measurement view will display an
image of the front panel channel keys. At this point moving the “pointer”
over one of the images of a front panel key and pressing the left mouse
button will have the same effect as pressing the key itself. Pressing the
right mouse button returns to the measurement view.
HELP SYSTEM The 7010 features a comprehensive help system. Each sub-menu has a
help screen that can be accessed by “pressing” the “?” “button” in the
upper right corner of the screen. These help screens contain information
about the functions of the sub-menu from which the help screen was
accessed. More general information about the instrument is accessed by
pressing the “help” “button” from the main menu.
To return to a sub-menu from a help screen, press the MENU key on the
front panel.
SYSTEM MENU The SYSTEM menu provides information on the firmware version and
probe connected to the instrument. It is also from the SYSTEM menu that
configuration setups may be saved and recalled, See the next heading
SETUP SAVE-LOAD.
Firmware Version
The firmware version is provided to assist in servicing or updating the
instrument.
Probe Information
Probe information includes the following:
1) Whether or not a Probe is Connected.
2) Probe Model Number.
3) Probe Serial Number.
4) Probe Calibration Date.
SETUP Four (4) configuration setups may be saved in flash memory. These
SAVE-LOAD setups may be recalled at anytime, including after the instrument has
been shut off. All settings from all menus are saved including those
enabled from the front panel keypad.
DISPLAY FORMAT The instrument will automatically format the screen for the best
possible view based on the number of parameters enabled.
Range Always
AC/DC Always
Enabled from DISPLAY Menu
Temperature Optional
(Temperature Compensated Probes only)
Minimum Optional
Maximum Optional Display of these parameters are selected from the HOLD menu.
Hold However the selected parameters are displayed only when the HOLD
Peak Optional key on the front panel is activated.
Valley Optional
Date Optional
Enabled and Set from DISPLAY Menu
Time Optional
SETTING THE The date and time is set and enabled from the display menu. When
DATE AND TIME enabled, the date and time is displayed at the bottom of the measurement
screen.
OVERVIEW This section describes field measurement operations for the model 7010
gauss/tesla meter. Each of the instrument’s functions and features
relating to field measurement is described in detail. This includes
general considerations, technical information, and instructions on how to
use each feature or function.
MEASUREMENT The measurement units selected are displayed at the bottom of the
UNITS measurement screen. Table 6-1 lists the available units for each parameter.
PRESENT FLUX The flux density reading contains five or six digits of information and
DENSITY READING includes the decimal point, polarity information, and the units of
measurement; see Figure 6-1. The number of digits depends on the
update interval setting. The update interval is discussed later in this
section. The present flux density reading is always displayed with a
larger text size than any other parameter. In the dc mode of
operation, the polarity is indicated with a “+” or a “-“. The polarity
information is absent when in ac mode, unless relative mode is
enabled. Relative mode operation is discussed later in this section.
+ 12.345 KG
Polarity Indicator
Flux Density Reading
Units
Auto 30G DC
Autorange mode
Present Range
ac or dc Mode
Pressing the RANGE key will advance the range setting. The four ranges
are followed by the autorange selection.
Autorange
In the autorange mode, the range is advanced if the reading reaches 90% full
scale of the present range. The range is lowered if the present reading falls
below 8% full scale of the present range.
NOTE: When the word “Auto” does not appear on the “measurement
mode indicator” line, the channel is in manual ranging mode.
Overrange Condition
When an overrange condition occurs the instrument will display the message
“OverRange” in place of the flux density reading. The next highest range
should be selected. If already on the highest range, then the flux density is
too great to be measured with this instrument and its probe.
Note: For dc mode operation, the polarity of the flux density that caused
the overrange condition will be displayed to the left of the
“OverRange” message.
AC MODE It is possible for the flux density signal to contain both a dc component and
OPERATION an ac component. In the ac mode the value displayed is the true rms value
of the waveform with its dc component removed.
Frequency Measurement
When used in the ac mode, the instrument can display the frequency of an
ac field. The frequency indicator line displays the measured frequency and
the selected analog filter setting, see Figure 6-3. Analog filters are
discussed later in this section.
------Hz f5k
Note: Although the flux density reading will be displayed when the
frequency is indeterminate, it is not specified at rated
accuracy.
AC MODE A low pass filter is available that may be set to pass frequencies only below
ANALOG 50kHz, 5kHz or 500 Hz. The instrument can automatically select the best filter
FILTERING setting based on the current flux density being measured. The filters affect the
displayed reading, and analog outputs.
The autofilter uses hysteresis and will enable the filters according to Table 6-3.
ZEROING Overview
“Zeroing” the probe and instrument is one of the most important steps to obtaining
accurate flux density measurements. An ideal Hall effect sensor produces zero
output in the absence of a magnetic field, but actual devices are subject to variations
in materials, construction and temperature. Therefore, most Hall effect sensors
produce some output even in a zero field. This will be interpreted by the instrument
as a magnetic field signal.
Also, the circuits within the instrument can produce a signal even when there is no
signal present at the input. This will also be interpreted as a magnetic field by the
instrument. Lastly, magnetic sources close to the actual field being measured, such
as those from electric motors, permanent magnets and the earth’s magnetic field
(roughly 0.5 gauss or 50 µT), can introduce errors in the final reading.
The zero operation is performed for both ac and dc modes of operation. While in dc
mode, the zero point can be adjusted manually. When the zeroing process is
initiated, the instrument performs separate zeroing sequences for ac and dc mode
operation. Both sequences occur regardless of whether a channel is in ac or dc
mode, and each has its own considerations.
NOTE: The process of zeroing also affects other functions such as corrected analog
output signals, the hold and relative, and the field classifiers. Uncorrected analog
outputs are not affected.
ac Zeroing
There are two modes of ac zeroing, basic and advanced. For most purposes, basic
ac mode zeroing is sufficient and is set as the default. Basic ac mode zeroing is
performed internal to the instrument and is not affected by the presence of magnetic
fields. The use of a “zero flux chamber” is not required for basic ac mode zeroing.
This is advantageous for situations when the probe is to be purposely exposed to a
magnetic field during zeroing. This situation is described under the next sub-
heading dc Zeroing.
Advanced ac zeroing is only required when measuring ac fields on the lower ranges
of the instrument. Advanced ac mode zeroing is enabled from a “checkbox” in the
SYSTEM menu. This mode requires a substantially longer zeroing period, but is
required for rated ac accuracy on the lower ranges. The probe must be shielded in a
“zero flux chamber” prior to initiating the zeroing procedure when using advanced ac
zeroing.
Typically it is only necessary to use advanced ac zeroing once when a new probe is
installed. Advanced ac zeroing data for the probe is stored, including after the
instrument has been shut off. The advanced ac zeroing mode may then be
disabled, and basic ac zeroing used thereafter. The only exception is when
continued measurement of low range ac fields are being made.
ZEROING dc Zeroing
(Continued)
For most situations it is preferable to shield the probe from all external magnetic
fields prior to zeroing. Provided with the instrument is a “zero flux chamber” which is
capable of shielding against fields as high as 30 mT (300 G or 23.88 kA/m). The
probe is simply inserted into the chamber before the zeroing process begins.
Handle the Hall probe with care. Do not bend the stem or apply pressure to
the probe tip as damage may result.
In other situations the user may want the probe to be exposed to a specific magnetic
field during the zeroing process. As an example, consider zeroing the probe when it
is exposed to the earth’s magnetic field. This will cancel the effect of the earth’s
magnetic field for all future readings. If a range is exceeded by the level of field
while zeroing, the lower range cannot be entered by the instrument.
Zero Prompt
The instrument will inform the user to zero a channel by flashing the ZERO key on
the front panel. The following are the conditions that will cause the instrument to
prompt the user to initiate the zeroing process:
Initiating Zeroing
The instrument will display the message “Zeroing” and a countdown indicates the
approximate duration of the zeroing process. Once zeroing begins it must be
allowed to complete. During this time, all controls are disabled except for the power
switch. The zeroing process is approximately 60 seconds when basic ac mode
zeroing is enabled and is approximately 2 minutes for advanced ac mode zeroing.
UPDATE The update interval of the reading may be adjusted. Shorter update intervals
INTERVAL allow rapid fluctuations in flux density levels to be observed. Longer update
intervals provide higher resolution, accuracy, and stability in the flux density
reading. The number of samples used for determining the flux density
reading for each setting is shown in Table 6-4.
The MIN and MAX hold functions are useful in determining the maximum or
minimum value of magnetic events that occur over a period of time. The
MIN and MAX use readings that are arithmetically calculated by the
instrument for slow changing signals.
The MAX hold function holds the reading that is arithmetically greater than
all previous readings. For instance, a reading of +125.0 is greater than
+99.0 or –150.0.
The MIN hold function holds the reading that is arithmetically less than all
previous readings. For instance, a reading of -125.0 is less than -99.0 or
+150.0.
If the reading exceeds the limit of the selected range the MIN and MAX
functions will display “OverRange” in place of minimum and maximum flux
density readings.
PEAK and VALLEY hold modes are used to track rapid events such as
magnetizing pulses. Unlike MIN and MAX, PEAK and VALLEY are not
arithmetically processed by the instrument. They represent the extremes of
a flux density waveform that have been captured within the response times
of the instrument. The response times of these functions are listed in
Section 2 – Specifications.
Numerically, The PEAK hold function holds it’s reading the same as the
MAX hold function. The PEAK hold function holds the reading greater than
all previous readings, within the response time of the instrument. For
instance, a reading of +125.0 is greater than +99.0 or –150.0.
Numerically, The VALLEY hold function holds it’s reading the same as the
MIN hold function. The VALLEY hold function holds the reading that is
lower than all previous readings, within the response time of the instrument.
For instance, a reading of -125.0 is less than -99.0 or +150.0.
Note: The hold functions are reset automatically when they are
activated. (i.e. the “held” readings are not saved when the
hold functions are de-activated).
RELATIVE MODE The relative mode allows a specific flux density value to be subtracted
from all future readings. Thus all future readings will be “relative” to that
value. For instance if the relative value is +100.0 gauss, and the
present flux density is +112.0 gauss, the actual displayed value will be
+12.0 gauss. If the flux density drops to +77.0 gauss, the actual
displayed value will be -23.0. Thus the relative mode allows for the
direct readout of variations around a given field, whether static (dc) or
alternating (ac).
There are two ways to generate a relative value. When the RELATIVE
key on the front panel is pressed, the instrument uses the present flux
density reading from the probe as the relative value. Alternatively, you
may specify a pre-determined value from the RELATIVE menu.
There may be situations when the user may prefer to shield the probe
from all external magnetic fields prior to performing a relative operation.
Provided with the instrument is a “zero flux chamber” which is capable of
shielding against fields as high as 30 mT (300 G or 23.88 kA/m). The
probe is simply inserted into the chamber before the relative operation
begins.
Handle the Hall probe with care. Do not bend the stem or apply
pressure to the probe tip as damage may result.
1) Select ac or dc mode.
2) Press the RELATIVE key.
The RELATIVE key will flash to indicate that the relative function is
enabled. The “relative value indicator” displays a value that was the
present flux density reading before the RELATIVE key was pressed , see
Figure 6-5. The present flux density reading should then display a value of
zero. All readings from this point forward are “relative” and not absolute.
The relative value is subtracted from all future readings. In the relative
mode, a reading of zero indicates that the actual flux density being
measured is equal to the relative value.
Rel +1.23456G
The RELATIVE key will discontinue to flash and the relative value
indicator will disappear. All readings from this point forward are
now absolute.
1) Select ac or dc mode.
2) From the main menu, choose RELATIVE.
3) Use the “checkbox” to the relative mode.
4) Position the “pointer” over the word “scale” and press the
ENTER key to advance the scale of the relative setting. As you
press the ENTER key, the decimal point and/or unit multiplier in
the relative setting will advance.
5) For dc mode operation, position the “pointer” over the polarity
indicator and press the ENTER key to change the setting.
6) Position the “pointer” over each digit in the relative setting and use
the ENTER key to advance the number.
7) Exit the menu system.
Once the relative value is set, either from the front panel or from the
RELATIVE menu, its value may be adjusted by using the MANUAL
ADJUST keys. The right MANUAL ADJUST key will add to the relative
value, and the left MANUAL ADJUST key will subtract from the relative
value
ANALOG A corrected and uncorrected analog output voltage signal available from standard
OUTPUTS BNC connectors. The uncorrected output signal is representative of the magnetic
flux density as measured by the Hall probe. The corrected output signal is
compensated for influences of temperature and frequency, as well as non-
linearities inherent in the Hall probe and instrument. The corrected output is
specified with a higher accuracy than the uncorrected output, with a bandwidth up
to 250 Hz. The uncorrected output is less accurate, but has a bandwidth of 50kHz.
Standard full scale output ranges are 3VRAW, 10VRAW, 3VRMS, and 10VRMS.
Adjustable full scale ranges up to 9.9 VRAW or 9.9VRMS, in increments of 0.1V, are
also available. The raw output settings provide voltage signals that are replicas of
the magnetic flux density waveforms being measured. The rms settings provide dc
voltage signals that are proportional to the rms value of the ac component of a flux
density signal. These outputs may be connected to a voltmeter, oscilloscope,
recorder, or external analog-to-digital converter.
Note: With dc mode operation and rms output settings, only the ac
component of the flux density is represented at the analog outputs.
See Section 2 – Specifications, for bandwidth and accuracy of the analog
outputs.
Full scale of the present flux density range always corresponds to the full scale
setting for the analog output voltage range. For example, if the full scale flux
density range is 30 gauss and the full scale analog output voltage range is 3 volts;
then a reading of 30 gauss will produce a voltage of 3 volts at the output. A reading
of -20 gauss will produce a voltage of -2 volts at the output.
CLASSIFIERS The classifier function allows a lower and upper limit of flux density to be
defined that can be quickly used to determine the status of a magnetic field.
The instrument will indicate visually whether the field is above, within, or
below these limits. The same information is presented in the form of general
purpose switch closures available at standard 15 pin “D” type female
connector. The connector is located at the rear of the instrument.
The pin assignments for the classifier connector (digital I/O) are indicated in
Section 2 – Specifications.
When the classifier function is enabled the word “Class” will be displayed.
If the measured field is below the pre-defined limits, the word “Low” will be
displayed next to the word “Class”. If the measured field is above the pre-
defined limits, the word “High” will be displayed next to the word “Class”.
If the field is within the pre-defined limits, the word “Class” will be displayed
by itself.
Two flux density values are set. One value sets the upper limit and the other
sets the lower limit. It does not matter which value is chosen to be the upper
or lower limit, the instrument will automatically make this determination. For
dc mode operation, polarity information needs to be specified. With ac
mode, only positive values may be set. The instrument will retain dc polarity
information when switching between ac and dc modes.
SOURCES OF When making flux density measurements there are several conditions
MEASUREMENT that can introduce errors:
ERROR
1) Failure to zero the error signals from the instrument, probe, and
nearby sources of magnetic interference.
Note: The probe is calibrated and specified with flux lines passing
perpendicularly through the Hall effect sensor.
INTRODUCTION In 1975 the IEEE-488 standard defined the electrical and mechanical
aspect for cables and connectors, as well as the handshaking,
addressing and general protocol for transmitting individual bytes of
information between instruments and computers. The standard was later
updated and is now called IEEE 488.1-1987.
Figure 7-1
IEEE-488 Connector
Pin-6: Data Available (DAV) This line is asserted when a talker puts a
data byte on the bus.
Pin-7: Not Ready for Data This line is asserted by a listener until it is
(NRFD) ready to accept a data byte.
IEEE-488 BUS Pin-8: Not Data Accepted This line is asserted by a listener until it has
FUNCTIONAL (NDAC) accepted a data byte.
DESCRIPTION
(Continued) Pin-9: Interface Clear Asserted by the system controller to force all
(IFC) other devices to go idle.
Pin-10: Service Request This line can be asserted by any device requiring
(SRQ) service.
Pin-18,19, 20. 21, 22, 23. These are the common return lines for all other
24 Ground (GND) signals except earth ground (Pin-12).
Each device on the system bus must be assigned a unique address ranging
from 00 to 31 decimal.
RS-232 The instrument has been designed to interface directly to a 9-pin “D” serial
FUNCTIONAL connector on most personal computers using a straight-through cable. This
DESCRIPTION means Pin-1 on the meter is connected to Pin-1 on the computer, Pin-2 to
Pin-2, etc. The serial transmission characteristics such as baud rate, parity
setting, etc. are defined in the COMMUNICATIONS menu. These settings
must match those of the computer.
A set of controls lines are used to indicate the overall status of the instrument
and computer. The instrument asserts the Data-Terminal-Ready (DTR) to
inform the computer that it is powered up, initialized and ready for
communications. Likewise the computer asserts Data-Set-Ready (DSR) to
indicate that it is ready for communications.
RS-232 Data is transmitted to the instrument on the receive (RX) line. Data is
FUNCTIONAL transmitted to the computer on the transmit (TX) line. In most cases these
DESCRIPTION are the only signals needed for communications.
(Continued) However, there may be a need to control the flow of data between the
instrument and the computer. For example, if the computer is receiving
data from the instrument but is interrupted by another task, it may want to
halt the instrument’s transmission until it has completed that task. This can
be accomplished in one of two ways.
Hardware Handshaking
Two lines called Clear-To-Send (CTS) and Request-To-Send (RTS) are
used to control the flow of data. This is often referred to as “hardware
handshaking”. The instrument will not transmit data until the computer
asserts the CTS line. Likewise the computer should not transmit data until
the instrument asserts the RTS line.
Hardware handshaking can be enabled or disabled from the
COMMUNICATIONS menu. It must also be enabled or disabled on the
computer.
Software Handshaking
Another way to control the flow of data is called “software handshaking”. In
this method two special characters are used to halt or resume the flow of
data. The XOFF character is decimal 19 (hex 13). If the instrument
receives the XOFF character at any time it will halt transmission until it
receives the XON character, which is decimal 17 (hex 11). Likewise if the
instrument transmits an XOFF character to the computer the computer
should halt transmission until it receives the XON character.
Software handshaking can be enabled or disabled from the
COMMUNICATIONS menu. It must also be enabled or disabled on the
computer. If disabled the instrument will treat the XON and XOFF
characters as any other character.
COMMUNICATIONS Settings for the IEEE 488 (GPIB) and RS-232 formats are set in the
SETUP COMMUNICATIONS menu.
The only setting that may be configured for the IEEE 488 bus is the
instruments address. This may be any value between 0 and 31.
RS-232 Configuration
The 9 pin connector on the rear panel of the instrument may be used for
RS-232 serial communications or may be used to interface to the
mouse. If RS-232 is enabled, the mouse is disabled and vice versa. See
Section 5 – User Interface for information regarding the use of the
mouse.
IEEE-488 General bus commands have the same meaning to any instrument on the
GENERAL BUS IEEE-488 bus. Often the user will be using an IEEE-488 controller card,
COMMANDS such as a National Instruments PCI-GPIB. Along with the card a software
package is supplied, commonly called a Universal Language Driver (ULD)
or a Universal Language Interface (ULI). This software translates the
complicated lower level operations into easy-to-use commands.
The following commands assume the instrument’s address has been set to
15:
This command cancels the remote mode and places the bus in an
idle state.
All front panel controls are disabled except for the power switch.
GTL - GO TO LOCAL
Same as the DCL command except only for the addressed instrument.
This command obtains the serial poll status byte, which contains
important status information about the instrument. Often a serial poll is
initiated when one or more devices are asserting an interface line called
SRQ (service request).
ERROR QUEUE AND Internally there are two buffers that accumulate messages. Errors are
OUTPUT QUEUE generated by a variety of sources, such as hardware errors or errors in
the command syntax. As errors occur messages are stored in an
ERROR QUEUE. They can be read by specific commands discussed
later in this section.
STATUS There are four register sets that indicate the status of the instrument,
such as errors or the present state of the machine. These are 16-bit
registers, but in many cases not all of the bits are used. The four
register sets are called:
There is also an 8-bit register that provides a 1-bit summary for each of
the four register sets. This is called the STATUS BYTE.
STATUS Each register set consists of three individual registers, as depicted in Figure 7-3.
(Continued)
The CONDITION register is a real time, read-only register that is constantly
updated to reflect current operating conditions.
The EVENT register is fed by the CONDITION register, but operates as a latch.
Whenever any bit in the CONDITION register goes to “1”, a corresponding “1” is
latched into the EVENT register and remains that way until cleared by a specific
command.
The ENABLE register is a mask register that is used to generate the single status
bit in the STATUS BYTE. Setting any bit in the ENABLE register to “1” will allow
a corresponding “1” in the EVENT register to set the summary bit in the STATUS
BYTE.
STATUS BYTE The IEEE-488 bus is a common bus linking a master controller (your
AND computer) to one or more instruments. If an instrument requires action from
SERVICE the controller it must somehow get the controller’s attention. A hardware line
REQUEST (SRQ) on the bus called SRQ (service request) is used to signal the controller that
one or more instruments on the bus require attention. A bit in the STATUS
BYTE called RQS (request for service) also sets. The master then reads the
STATUS BYTE from each instrument to determine which instrument(s)
require service.
The RQS bit can set if any of the summary bits from the MEASUREMENT
EVENT, OPERATION EVENT, STANDARD EVENT or QUESTIONABLE
EVENT registers are set, or if an error or output message is ready to be
transmitted to the master. The SRQ ENABLE register is a mask register that
is used to allow any of these conditions to set the RQS bit, and thus the SRQ
line. Setting any bit in the SRQ ENABLE register to “1” will allow a
corresponding “1” in the STATUS BYTE register to set the RQS bit.
If any of the bits in the STANDARD EVENT register set, and their
respective enable bits are set, the Event Summary Bit (ESB) will set.
If any of the bits in the QUESTIONABLE EVENT register set, and their
respective enable bits are set, the Questionable Summary Bit (QSB) will
set.
This bit sets any time there is a message available in the output queue.
This bit sets any time there is an error message available in the error
queue.
If any of the other bits in the STATUS BYTE are set, and their
respective enable bits are set in the SRQ ENABLE register, the Request
For Service (RQS) will set, causing the SRQ interface line to be
asserted if using the IEEE-488 bus.
STANDARD EVENT If any of these bits set, and their respective enable bits are set, the Event
REGISTER Summary Bit (ESB) will set in the STATUS BYTE.
Indicates that the instrument was turned off and on since the last
communication.
Indicates that there was a syntax or spelling error in the command, or the
command received is not supported.
Indicates that the instrument did not operate properly due to some
internal error.
MEASUREMENT If any of these bits set, and their respective enable bits are set, the
EVENT REGISTER Measurement Summary Bit (MSB) will set in the STATUS BYTE.
When the classifier function is enabled these bits indicate that the
present reading is higher than the upper classifier limit.
When the classifier function is enabled these bits indicate that the
present reading is below the lower classifier limit.
Indicates that the present reading exceeds the measurement range for
the instrument.
OPERATION EVENT If any of these bits set, and their respective enable bits are set, the
REGISTER Operation Summary Bit (OSB) will set in the STATUS BYTE.
QUESTIONABLE If any of these bits set, and their respective enable bits are set, the
EVENT REGISTER Questionable Summary Bit (QSB) will set in the STATUS BYTE.
IEEE-488.2 The “common” commands are recognized and acted upon in a similar
“COMMON” manner by all IEEE-488.2 instruments, whether a DVM, scope, frequency
COMMAND meter, tesla meter, etc. These are the syntax rules:
SYNTAX
1) A common command always begins with an asterisk character (*) followed
by a three or four character acronym and possibly one other parameter. For
instance a command to reset the instrument is *RST.
2) The commands are not case sensitive. For instance the *RST, *rst and
*rSt commands are valid.
A set bit in the STANDARD EVENT ENABLE register allows its corresponding event to
set the ESB (event summary bit) in the STATUS BYTE register. <NRf> is an ASCII
string representing an integer mask. For instance a value of 45 decimal is the same as
binary 00101101, thus setting bits 5, 3, 2 and 0 in the enable register.
Read the contents of the STANDARD EVENT ENABLE register. The results are
placed in the output queue of the instrument.
Places an ASCII “1” in the output queue once all commands have been
executed. This is particularly useful when using the RS-232 Port.
*GTL - GO TO LOCAL
This command cancels remote mode and restores front panel operation. This
command was primarily intended for use with the RS-232 port, but can be used
on the GPIB bus.
SCPI The SCPI commands provide a language protocol and define a standard set of
COMMAND commands to program every aspect of the instrument. These are the syntax
SYNTAX rules:
2) The commands are not case sensitive. For instance the :MEASURE,
:measure and :MEASure commands are valid.
4) For any command there is a short and long spelling of the command. Use
the following rules for the short version:
b) If the command has more than four letters and the fourth letter is a vowel,
drop it and all letters that follow it. For instance the command :INPUT
can be shortened to :INP.
c) If the command has more than four letters and the fourth letter is a
consonant, drop all letters that follow it. For instance the command
:MEASURE can be shortened to :MEAS
f) The use of anything other than the short or long version of a command is
invalid. For instance both the :MEASURE and :MEAS commands are
valid, but :MEASU is not.
SCPI COMMANDS- In the following discussion the commands are written such that the short
GENERAL form of the command is written in UPPER CASE letters and the remainder
of the command is written in lower case letters. Either form can be used.
If parameters are required they will appear within <> brackets. A <b>
parameter indicates a Boolean function, either “0” or “OFF”, or “1” or “ON”.
A <n> parameter is a single ASCII digit. <NRf> is usually a multiple digit
number and means Numeric Representation format
SCPI COMMANDS- As error messages occur, they are placed in the error queue. Each
ERROR QUEUE message will contain a number, a comma (,) and a brief description of the
MESSAGES AND error. Negative (-) numbers are used for SCPI defined messages while
COMMANDS positive (+) numbers relate specifically to the instrument.
The error queue can hold up to 10 messages. Each time the queue is
read the oldest message is presented and removed from the queue. If
no errors exist the message “0, No error” will be returned. If the queue is
full the message “-350, Queue Overflow” will occupy the last queue
location. It is an indication that at least one, perhaps more error
message(s) were lost.
There are certain error status bits that will set in the STANDARD EVENT
register. These bits provide general error indications. The error queue
will provide more detailed information about the errors.
:SYSTem:ERRor?
:SYSTem:ERRor:NEXT?
This command places the oldest error message in the output queue and
removes it from the error queue.
The instrument’s real time clock can be set from the DISPLAY menu or
with this command using the following format:
SCPI COMMANDS – The clock supports 12-hour or 24-hour format. However, the format can
SYSTEM only be set using the DISPLAY menu. The :SYSTem:TIMe command
INFORMATION AND supports only a 24-hour format. If the clock has been set for 12-hour
CONFIGURATION format <hours> will be converted internally to 12-hour format. For
instance if <hours> is specified as 17 the time will be converted internally
COMMANDS
to 5 pm. Here is a simple conversion chart:
(Continued)
:SYSTem:TIMe?
The contents of the instrument’s real time clock will be returned using the
following format:
<hour>, <minute>, <second>
The clock supports 12-hour or 24-hour format. However, the format can
only be set using the DISPLAY menu. The :SYSTem:TIMe? command
supports only a 24-hour format. If the clock has been set for 12-hour
format <hours> will be converted to 24-hour format. For instance if the
present hours is 5 pm then <hours> will be returned as 17. See Table 7-5.
SCPI COMMANDS – A <year> less than 2000 will not be accepted because the instrument
SYSTEM did not exist prior to the year 2000. The instrument does not check the
INFORMATION AND validity of the day of the month. For instance, a date of February 31 will
CONFIGURATION not generate an error.
COMMANDS
(Continued) :SYSTem:DATe?
The contents of the instrument’s real time calendar will be returned
using the following format:
<year>, <month>, <day>
where: <year> is in the range of 2000 – 9999
<month> is in the range of 1 – 12
(1 = January, 2 = February, etc)
<day> is in the range of 1 – 31 (day of the month)
:SYSTem:AZERo<b>
This command enables advanced ac zeroing when <b> is “1” or “ON”, or
disables advanced ac zeroing when <b> is “0” or “OFF”. See Section 6
for more information about advanced ac zeroing.
:SYSTem:AZERo?
The on/off state of advanced ac zeroing is placed in the output queue.
:INPut:STATe <b>
This command activates the channel when <b> is “1” or “ON”, or
deactivates the channel when <b> is “0” or “OFF”.
:INPut:STATe?
The on/off state of the channel (“OFF” or “ON”) is placed in the output
queue.
:DISPlay:ENABle <b>
When <b> is “1” or “ON” the display will be present normal flux density
and related information. When <b> is “0” or “OFF” the display will
contain the phrase “Display turned off via comm port.
The process of updating the display on a continuous basis takes time
away from the measurement process. Turning the display off can
increase the speed at which readings can be acquired from a
communication port.
:DISPlay:ENABle?
The on/off state of the display (“OFF” or “ON”) is placed in the output
queue.
SCPI COMMANDS- The STATUS commands control and query the MEASUREMENT EVENT,
STATUS OPERATION EVENT and QUESTIONABLE EVENT registers as well as
COMMANDS the error queue.
:STATus:MEASurement:EVENt?
:STATus:OPERation:EVENt?
:STATus:QUEStionable:EVENt?
Reads the contents of the specified EVENT register and places it in the
output queue, then clears the register contents.
:STATus:MEASurement:ENABle <NRf>
:STATus:OPERation:ENABle <NRf>
:STATus:QUEStionable:ENABle <NRf>
Programs the specified EVENT ENABLE register with the value <NRf>.
<NRf> is an ASCII string representing an integer mask. For instance a
value of 45 decimal is the same as binary 00101101, thus setting bits 5, 3,
2 and 0 in the enable register.
:STATus:MEASurement:ENABle?
:STATus:OPERation:ENABle?
:STATus:QUEStionable:ENABle?
Reads the contents of the specified EVENT ENABLE register and places it
in the output queue.
:STATus:MEASurement:CONDition?
:STATus:OPERation:CONDition?
:STATus:QUEStionable:CONDition?
:STATus:PRESet
SCPI COMMANDS- These commands select the measurement units for temperature, flux
UNIT COMMANDS density and time-based measurements.
:UNIT:TEMPerature C|CEL|F|FAR|K
Specifies that the temperature units are to be in Celsius (C or CEL),
Fahrenheit (F or FAR) or Kelvin (K).
:UNIT:TEMPerature?
Places an ASCII string in the output queue representing the temperature
units. The ASCII string will be CEL for Celsius, FAR for Fahrenheit or K
for Kelvin.
:UNIT:FLUX GAUSs|TESLa|AM|OERSted
Specifies that the flux density units are to be in gauss (GAUSs), tesla
(TESLa), amps per meter (AM) or oersted (OERSted).
:UNIT:FLUX?
Places an ASCII string in the output queue representing the flux density
units. The ASCII string will be GAUSS, TESLA, AM or OERSTED.
:UNIT:TIME SEC|HZ
:UNIT:TIME?
Places an ASCII string in the output queue representing the time-based
measurement units. The ASCII string will be SEC for period
measurements or HZ for frequency measurements.
SCPI COMMANDS- These commands select either a fixed range or AUTO range, and specify
RANGE COMMANDS if the measured flux density will be static (dc) or alternating (ac).
:SENSe:FLUX:AC
:SENSe:FLUX:DC
Specifies that the measured flux density will be static (dc) or alternating
(ac). These commands do not alter the present range setting.
:SENSe:FLUX:RANGe:FIXed <n>
:SENSe:FLUX:DC:RANGe:FIXed <n>
:SENSe:FLUX:AC:RANGe:FIXed <n>
Specifies that the measured flux density will be static (dc) or alternating
(ac). If the “AC” or “DC” key word is omitted, then static (dc) operation is
selected. Also specifies one of four fixed operating ranges for the
instrument, as defined in the following table:
:SENSe:FLUX:RANGe:AUTo <b>
:SENSe:FLUX:DC:RANGe:AUTo <b>
:SENSe:FLUX:AC:RANGe:AUTo <b>
Specifies that the measured flux density will be static (dc) or alternating
(ac). If the “AC” or “DC” key word is omitted, then static (dc) operation is
selected.
AUTO ranging is turned off when <b> is “0” or “OFF”, or on when <b> is
“1” or “ON”. When AUTO range is turned on the instrument determines
the range that will offer the best resolution. When AUTOrange is turned
off the range will be fixed to the last range that was in use when the
command was issued.
The string will start with the phrase “AC” or “DC, followed by a comma(,).
The next single digit will be the range setting, as shown in Table 7-6,
and will be in the range of 1 to 4, followed by a comma(,). Next will be
the phrase “OFF” or “ON, indicating the state of the AUTO range
feature.
:SENSe:FLUX:FILTer:LPASs:FREQuency <NRf>
There are two choices for the cutoff frequency for the low pass filter; 500
Hz or 5 kHz. These are two-pole filters with an attenuation of –12 db per
octave. <NRf> specifies the cutoff frequency in hertz, either 500 or
5000.
:SENSe:FLUX:FILTer:LPASs:FREQuency?
Places the low-pass filter’s cutoff frequency in hertz in the output queue,
either 500 or 5000.
:SENSe:FLUX:FILTer:LPASs:STATe <b>|AUTo
The low-pass filter is turned off when <b> is “0” or “OFF”, on when <b>
is “1” or “ON”, or placed in AUTO mode then <b> is “AUTo”.
When the low pass filter is turned off, the instrument has an inherent
cutoff of 50kHz.
:SENSe:FLUX:FILTer:LPASs:STATe?
The state of the low-pass filter (”OFF”, “ON” or “AUTO”) is placed in the
output queue.
:CALCulate:AVERage:COUNt?
A new measurement is acquired approximately every 0.032 seconds.
The system requires a minimum of 6 measurements to be averaged,
which results in a new flux density reading every 0.2 seconds. For
greater stability the number of measurements to average can be
increased to 15, 30 or 60, resulting in new flux density readings every
0.5, 1 or 2 seconds, respectively.
This command returns the number of measurements being used for the
average, either “6”, “15”, “30” or “60”.
SCPI COMMANDS- These commands control the CLASSIFIER, or limit checking function.
CLASSIFIER (LIMIT) Two individual flux density limits can be specified. If the present flux
COMMANDS density reading becomes arithmetically less than the lower limit, or
arithmetically greater than the upper limit, the situation is considered to
be a “failure”.
:CALCulate:LIMit:LOWer:DATa?
:CALCulate:LIMit:UPPer:DATa?
The requested classifier limit is placed in the output queue. The limit
will be between 0.0000000000 and ± 29999999. The meaning of the
limit depends upon the currently selected unit of measure (gauss,
tesla, etc.). For instance the value 1.0203 could mean 1.0203 G or
1.0203 T.
:CALCulate:LIMit:STATe <b>
The classifier function is turned on when <b> is “1” or “ON”, or turned
off when <b> is “0“ or “OFF”.
:CALCulate:LIMit:STATe?
The state of the classifier function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate:LIMit:FAIL?
The pass/fail status of the classifier test is placed in the output queue.
A “0” indicates that the measured value falls between the lower and
upper limit and is considered a “pass” situation. A “1” indicates that
the measured value either falls below the lower limit or above the
higher limit.
SCPI COMMANDS- These commands control zeroing of the Hall probe and the instrument.
ZEROING
COMMANDS
:CALibration:ZERo:HSENsor:INITiate
This command initiates the same zeroing process as the front-panel
ZERO push button. The Hall probe should be placed in a stable, low-
flux environment such as a zero flux chamber prior to issuing this
command. While zeroing is in progress the “ZERO” status bit in the
OPERATION CONDITION REGISTER will be set.
:CALibration:ZERo:HSENsor:INITiate?
Same as the “:CALibration:ZERo:HSENsor:INITiate” command, but a
response is returned upon completion of the zeroing process. A “0”
indicates successful completion, otherwise a non-zero number is
returned indicating a failure.
SCPI COMMANDS- These commands control the RELATIVE function. The RELATIVE
RELATIVE OFFSET function subtracts a fixed flux density value from all measured values.
COMMANDS The user can define the offset or it can be generated using the present
value of flux density as measured by the Hall probe.
There are some limitations when using the RELATIVE function. Refer to
Section 6 - Flux Density Measurement for details.
:INPut:OFFSet?
The relative offset value is placed in the output queue and will be
between 0.0000000000 and ± 29999999. The meaning of the offset
depends upon the currently selected unit of measure (gauss, tesla,
etc.). For instance the value 1.0203 could mean 1.0203 G or 1.0203 T.
:INPut:OFFSet:STATe <b>|ONCe
When <b> is “0” or “OFF” the RELATIVE function is turned off. The
present flux density readings will not be altered. The previously defined
relative offset value will be preserved.
:INPut:OFFSet:STATe?
The state of the RELATIVE function (“OFF” or “ON”) is placed in the
output queue.
SCPI COMMANDS- The instrument provides an uncorrected analog output signal that
ANALOG OUTPUT originates from the amplifier section. This signal is not corrected for errors
COMMANDS in the Hall probe or instrument but is capable of high frequency operation.
Also provided is a corrected output signal that is generated digitally. This
signal is corrected for most errors in the Hall probe and instrument, but is
only capable of low-frequency reproduction.
You can define the type of signal that appears at each output and the scale
of that signal, from the ANALOG OUT menu or by issuing these commands
:OUTPut:COUPling1 AC|DC
When the channel is configured for alternating (ac) flux density
measurement you can select one of two signals to appear at the
uncorrected analog output. When “AC” is specified the signal will be the
real-time waveform as measured by the Hall probe. When “DC” is
specified the output signal will be a dc level representing the true rms value
of the waveform.
When the channel is configured for static (dc) flux density measurement
you can select one of two signals to appear at the uncorrected analog
output. When “AC” is specified the signal will be the real-time, low
frequency waveform as measured by the Hall probe, from 0 – 400 Hz.
When “DC” is specified the output signal will be a dc level representing the
true rms value of the low frequency waveform.
:OUTPut:COUPling1?
The type of signal selected for the uncorrected analog output (“AC” or
“DC”) is placed in the output queue.
:OUTPut:ATTenuation1 <NRf>
The full scale output voltage of the uncorrected analog output can be
specified to any level between 0 and 10 volts in 0.1 volt steps. Normally a
full-scale level of 3.0 volts is used because it directly corresponds to the
full-scale ranges of the instrument. For instance, if on the 300 mT range, a
reading of -189.3 mT will result in an analog output of -1.893 volts. If on
the 3 kG range, a reading of +2.6 kG will result in an analog output of +2.6
volts.
SCPI COMMANDS- Some data acquisition systems may require different analog input
ANALOG OUTPUT ranges. For example, assume such a system cannot accept signals
COMMANDS greater than ± 2.0 volts. In this case you can specify 2.0V as the full-
(Continued) scale range of the analog output (<NRf> = 2.0).
The relationship between the output voltage and the flux density reading
is:
Full scale output voltage / full scale flux density range
For example, assume the full scale output voltage has been set to 2.0
volts and the present range setting is 300 mT. The relationship will be:
Thus, when the flux density reaches the maximum for the present range,
the analog output voltage will reach a maximum of ± 2.0 volts.
:OUTPut:ATTenuation1?
The full scale voltage setting of the uncorrected analog output (0.0 to
10.0) is placed in the output queue.
:OUTPut:COUPling2 AC|DC
When the instrument is configured for alternating (ac) flux density
measurement you can select one of two signals to appear at the
corrected analog output. When “AC” is specified the signal will be the
real-time waveform as measured by the Hall probe, within the low
frequency limits of the corrected output. When “DC” is specified the
output signal will be a dc level representing the true rms value of the low
frequency waveform.
:OUTPut:ATTenuation2 <NRf>
The full scale output voltage of the corrected analog output can be
specified to any level between 0 and 10 volts in 0.1 volt steps. Normally a
full-scale level of 3.0 volts is used because it directly corresponds to the
full-scale ranges of the instrument. For instance, if on the 300 mT range, a
reading of -189.3 mT will result in an analog output of -1.893 volts. If on
the 3 kG range, a reading of +2.6 kG will result in an analog output of +2.6
volts.
Some data acquisition systems may require different analog input ranges.
For example, assume such a system cannot accept signals greater than ±
2.0 volts. In this case you can specify 2.0V as the full-scale range of the
analog output (<NRf> = 2.0).
The relationship between the output voltage and the flux density reading is:
:OUTPut:ATTenuation2?
The full scale voltage setting of the corrected analog output (0.0 to 10.0) is
placed in the output queue.
:CALCulate:HOLD:MINimum:STATe?
The state of the MIN HOLD function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate:HOLD:MINimum:CLEar
The value being held by the MIN HOLD function is reset to 0.
:CALCulate:HOLD:MINimum?
The value being held by the MIN HOLD function is placed in the output
queue and will be between 0.0000000000 and ± 29999999. The
meaning of this value depends upon the currently selected unit of
measure (gauss, tesla, etc.). For instance the value 1.0203 could mean
1.0203 G or 1.0203 T.
:CALCulate:HOLD:MAXimum:STATe <b>
The MAX HOLD function is turned on when <b> is “1” or “ON”, or turned
off when <b> is “0“ or “OFF”.
:CALCulate:HOLD:MAXimum:STATe?
The state of the MAX HOLD function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate:HOLD:MAXimum:CLEar
The value being held by the MAX HOLD function is reset to 0.
:CALCulate:HOLD:MAXimum?
The value being held by the MAX HOLD function is placed in the output
queue and will be between 0.0000000000 and ± 29999999. The
meaning of this value depends upon the currently selected unit of
measure (gauss, tesla, etc.). For instance the value 1.0203 could mean
1.0203 G or 1.0203 T.
:CALCulate:HOLD:PEAK:STATe?
The state of the PEAK HOLD function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate:HOLD:PEAK:CLEar
The value being held by the PEAK HOLD function is reset to 0.
:CALCulate:HOLD:PEAK?
The value being held by the PEAK HOLD function is placed in the output
queue and will be between 0.0000000000 and ± 29999999. The
meaning of this value depends upon the currently selected unit of
measure (gauss, tesla, etc.). For instance the value 1.0203 could mean
1.0203 G or 1.0203 T.
:CALCulate:HOLD:VALLey:STATe <b>
The VALLEY HOLD function is turned on when <b> is “1” or “ON”, or
turned off when <b> is “0“ or “OFF”.
:CALCulate:HOLD:VALLey:STATe?
The state of the VALLEY HOLD function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate:HOLD:VALLey:CLEar
The value being held by the VALLEY HOLD function is reset to 0.
:CALCulate:HOLD:VALLey?
The value being held by the VALLEY HOLD function is placed in the
output queue and will be between 0.0000000000 and ± 29999999. The
meaning of this value depends upon the currently selected unit of
measure (gauss, tesla, etc.). For instance the value -1.0203 could mean
-1.0203 G or -1.0203 T.
:MEASure:FFLux?
This command is similar to the MEASure:FLUX? command but can be
used to acquire new flux density readings at a much faster rate. However,
the returned values may not be as precise or accurate. For example, if
the presently displayed reading is 2.87347 T, this command may return
2.872 T.
This command uses the same information that is used to construct the
corrected analog output, therefore the accuracy is the same as specified
for the corrected analog output. The corrected analog output must be
turned on to use this command. Use the ANALOG OUT menu or OUTPut
commands to enable the output. The analog output voltage range has no
affect on the returned value, so it can be set to any desired level.
However, the type of output selected, whether “raw” or “rms”, will affect
the value returned by this command.
SCPI COMMANDS- The rate at which new flux density readings can be acquired using this
MEASUREMENT command is roughly 100 per second, but the speed of the
COMMANDS communications port can degrade this. For example, suppose the
(Continued) abbreviated version of the command is issued, such as “MEAS:FFL?<lf>.
If the RS-232 port is used at a baud rate of 9600, a character length of 7,
one stop bit and no parity, it would take at minimum about 0.011 seconds
to transmit the command. If the meter returns a value of “–22.34;<lf>”,
this would take about 0.008 seconds. There is also a small amount of
time needed for the meter to interpret the command and acquire a
reading. So, let’s assume a total of 0.02 seconds for one transaction.
This would limit the number of readings to 50 per second, best case. Use
a higher baud rate or the GPIB interface to increase the speed.
:MEASure:TEMPerature?
The latest temperature reading from a temperature-compensated Hall
probe is placed in the output queue. The meaning of this value depends
upon the currently selected unit of measure (celsius, kelvin or fahrenheit).
For instance the value 45.3 could mean 45.3 °F or 45.3 °C.
:MEASure:TIMe?
The latest frequency or period measurement of the flux density signal is
placed in the output queue. The meaning of this value depends upon the
currently selected unit of measure (seconds or hertz).
INTERMIXING As mentioned earlier a string sent to the instrument can contain more than
“COMMON” AND one command as long as the commands are separated by semicolons(;).
SCPI COMMANDS “Common” and SCPI commands can be intermixed.
When using the RS-232 port the instrument will always send the LF
character every time it transmits a message to the system controller.
When using the IEEE-488 bus the instrument will send the LF character
and assert the EOI line every time it transmits a message to the system
controller.
EXAMPLE: If you are overwhelmed by the use of the EVENT, CONDITION and ENABLE
USING THE registers, this example should help.
EVENT, ENABLE,
AND CONDITION The goal is to get the attention of the computer when a reading is available
from the instrument. This means we have to assert the SRQ interface line on
REGISTERS
the GPIB bus, or set some status condition that can be read via the RS-232
port.
The “RQS” bit in the STATUS BYTE directly affects the SRQ interface line.
When it becomes “1” the line is asserted. “RQS” can be affected by six
conditions, four of which are controlled by event registers. In one of these,
the MEASUREMENT EVENT register, there is a bit called “RAV”, which sets
to “1” when a reading is available. So, when “RAV” goes to “1” we want
“RQS” to go to “1”.
1) To allow the “RAV” event to reach the STATUS BYTE we must set the
corresponding bit in the MEASUREMENT ENABLE register. “RAV” is
bit location D4. The decimal value that will set bit location D4 is 16. The
SCPI command is
:STATUS:MEASUREMENT:ENABLE 16
2) The “RAV” event will now set the “MSB” (Measurement Summary Bit) in
the STATUS BYTE. But this will not set the “RQS” bit unless the “MSB”
event is enabled in the SRQ ENABLE register. “MSB” is bit location D0.
The decimal value that will set bit location D0 is 1. The “common”
command is
*SRE 1
Now as soon as a reading is available the “RQS” bit will set and the SRQ
interface line will be asserted. For GPIB users the system controller, not
knowing which instrument is requesting service, will sequentially read each
instrument’s serial poll register to determine who is asserting SRQ. For
RS-232 users there is no dedicated interface line that can be used to get the
attention of the computer. However, the STATUS BYTE can be inspected
from time to time to determine when an event has occurred. This is done
using the *STB? Command.
Normally the STATUS BYTE would be read first to determine the general
cause for the request, then the EVENT register would be read to determine
the exact nature of the request. The latter step is very important because
the act of reading the EVENT register is what clears the event. Simply
reading the STATUS BYTE or asking for a flux density reading will not reset
the “RAV” bit. You must read the MEASUREMENT EVENT register to reset
“RAV”. The SCPI command is
:STATUS:MEASUREMENT:EVENT?
One line of flux in the cgs measurement system is called a maxwell (Mx), but the weber (Wb),
which is 108 lines, is more commonly used.
Flux density, also called magnetic induction, is the number of flux lines passing through a given
area. It is commonly assigned the symbol “B” in scientific documents. In the cgs system a gauss
(G) is one line of flux passing through a 1 cm2 area. The more commonly used term is the tesla (T),
which is 10,000 lines per cm2 . Thus
It is important to know that magnetic field strength and magnetic flux density are not the same. The
only time the two are considered equal is in free space. Only in free space is the following
relationship true:
A device commonly used to measure flux density is the Hall effect sensor. A Hall effect sensor is a
thin slice of a semiconductor material to which four leads are attached at the midpoint of each edge,
as shown in Figure A-2.
Hall Effect
Sensor
A constant current (Ic) is forced through the material. In a zero magnetic field there is no voltage
difference between the other two edges. When flux lines pass through the material the path of the
current bends closer to one edge, creating a voltage difference known as the Hall voltage (Vh). In
an ideal Hall generator there is a linear relationship between the number of flux lines passing
through the material (flux density) and the Hall voltage.
The Hall voltage is also a function of the direction in which the flux lines pass through the material,
producing a positive voltage in one direction and a negative voltage in the other. If the same
number of flux lines pass through the material in either direction, the net result is zero volts. This
sensitivity to flux direction makes it possible to measure both static (dc) and alternating (ac)
magnetic fields.
The Hall voltage is also a function of the angle at which the flux lines pass through the material.
The greatest Hall voltage occurs when the flux lines pass perpendicularly through the material.
Otherwise the output is related to the cosine of the difference between 90° and the actual angle.
The sensitive area of the Hall effect sensor is generally defined as the largest circular area within
the actual slice of the material. This active area can range in size from 0.2 mm (0.008”) to 19 mm
(0.75”) in diameter. Often the Hall effect sensor assembly is too fragile to use by itself so it is often
mounted in a protective tube and terminated with a flexible cable and a connector. This assembly
is known as a Hall probe. For more information on Hall Probes see Section 3 – Probes.
OECO reserves the right to make changes in design at any time without incurring
any obligation to install same on units previously purchased.
This warranty gives you specific legal rights, and you may also have other rights
that vary from state to state.
THIS SIDE BLANK !
(Front of Rear Cover)
4607 SE International Way
Milwaukie, OR 97222
Phone: 503-659-5999
www.oeco.com