IPA-1000 User Manual
IPA-1000 User Manual
IPA-1000 SYSTEM
USER MANUAL
BC BIOMEDICAL
IPA-1000 SYSTEM
TABLE OF CONTENTS
DESCRIPTION ............................................................................................................. 1
OVERVIEW .................................................................................................................. 2
KEYS............................................................................................................................ 4
SCREENS .................................................................................................................... 5
MESSAGES ................................................................................................................. 7
SETUP ........................................................................................................................... 9
OPERATIONS.............................................................................................................. 10
SPECIFICATIONS ....................................................................................................... 15
NOTES ......................................................................................................................... 17
IPA-1000
MC3.5-1000
MC35-1000
i
WARNING - USERS
The IPA-1000 System is for use by skilled
technical personnel only.
WARNING - USE
The IPA-1000 System is intended for testing only
and should never be used in diagnostics,
treatment or any other capacity where they would
come in contact with a patient.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the IPA-1000
System. A serious hazard may occur if the
patient is connected when testing with the unit.
Do not connect any leads from the patient
directly to the unit or DUT while it is powered by
the System.
WARNING - LIQUIDS
Do not submerge or spill liquids on the IPA-1000
System. Do not operate the IPA-1000 System if
internal components not intended for use with
fluids may have been exposed to fluid, as the
internal leakage may have caused corrosion and
be a potential hazard.
WARNING - MODIFICATIONS
The IPA-1000 System is intended for use within
the published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
ii
CAUTION - FLUID
Only Distilled Water should be used in the
chambers with the IPA-1000 System. Do not use
tap water, glucose or any other fluid; this will
cause the tubing to become contaminated.
CAUTION - SERVICE
The IPA-1000 System is intended to be serviced
only by authorized service personnel.
Troubleshooting and service procedures should
only be performed by qualified technical
personnel.
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the IPA-1000 System. Allow the
IPA-1000 System to acclimate to specified
conditions for at least 30 minutes before
attempting to operate it.
CAUTION - CLEANING
Do not immerse. The IPA-1000 System should be
cleaned by wiping gently with a damp, lint-free
cloth. A mild detergent can be used if desired.
CAUTION - INSPECTION
The IPA-1000 System should be inspected before
each use for wear and should be serviced if any
parts are in question.
iii
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
Center Negative
NOTICE – ABBREVIATIONS
ANSI American National Standards Institute
C Celsius
° degree(s)
DUT Device Under Test
Euro European
F Fahrenheit
FS Full Scale
hr hour(s)
Hz hertz
kg kilogram(s)
µA microampere(s)
mA milliampere(s)
mL milliliter(s)
mm millimeter(s)
min minute(s)
NEDA National Electronic Distributors Association
Lbs Pounds
RH Relative Humidity
USA United States of America
VDC Volts Direct Current
iv
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
v
This Page Intentionally Left Blank
vi
BC BIOMEDICAL
IPA-1000 SYSTEM
INFUSION PUMP ANALYZER
The Model IPA-1000 System is a Microprocessor based High Precision Infusion Pump
Analyzer (IPA). It tests the flow rate of Intravenous (I.V.) Infusion Pumps. The flow rates
are displayed in milliliters per hour. The unit can test two volumetric pumps for output flow
rate simultaneously using two volumetric chambers.
AVAILABLE MODELS:
The base unit plus one of the volumetric chambers are required to test Infusion Pumps.
OPTIONAL ACCESSORIES:
BC20-21100 BATTERY ELIMINATOR, USA version
BC20-21101 BATTERY ELIMINATOR, EUROPE version
BC20-30109 SOFT CARRYING CASE
BC20-40631 OVERFLOW ASSEMBLY KIT 3.5 mL
BC20-40632 OVERFLOW ASSEMBLY KIT 35 mL
BC20-40607 CHAMBER INTERCONNECT CABLE
1
OVERVIEW
This section looks at the layout of the IPA-1000 System, including the Base Unit and the
Chambers and gives descriptions of the elements that are present.
9V Battery
Compartment
(Back)
2
3.5 mL Chamber 35 mL Chamber
MC3.5-1000 MC35-1000
8 pin Mini-Din
connector for
connection to
IPA-1000
8 pin Mini-Din
connector for
connection to
IPA-1000
Luer Lock
Connector for
Fluid In
3
KEYS
– This key turns the unit off and on. The unit will return to the screen that was
– In the DISPLAY MODE, these keys toggle the display through the
available main screens. In the SELECT MODE, if a parameter has been highlighted, these
– On the Setup screen, there are a number of parameters that may be selected
and changed. This key sequences the cursor (Highlight) through those parameters.
– This key is used to STOP any of the running flow tests or to manually reset the
system if Auto Test Reset is set to NO. (See Setup Section for more details.)
– This key toggles the unit into and out of the Setup Mode. Depressing this key
will enter the Setup screen where the configuration can be viewed and adjusted.
Depressing the key again will exit the Setup Mode and return to the previously viewed
main screen.
4
SCREENS
MAIN SCREENS – There are four main screens: Chamber 1, Chamber 2, DUAL and
CHAMBER SCREENS – The Chamber 1 and Chamber 2 screens have a large time and
flow rate display, as shown below. The arrow at the top of the screen indicates the
connector used for that Chamber. The level indicator at the bottom of the screen identifies
LEVEL
INDICATOR
See Page 11 for
special graphics
DATA
UNITS
• Seconds
• mL/hr
5
DUAL SCREEN – The DUAL screen shows the status of both Chambers at the same time.
MESSAGE DISPLAY
OR DATA
CONNECTOR
• Chamber Not Detected LOCATION
• Ready For Flow • Chamber 1
• Flow Rate Too High • Chamber 2
• Invalid Sensor Sequence
• Chamber Needs Calibration
• Low Battery LEVEL
INDICATOR
See Page 11
for special
DATA graphics
UNITS
• Seconds
• mL/hr
CONNECTORS SCREEN – The connectors screen indicates the layout of the connectors
at the top of the IPA-1000. Also shown is the volume capacity of any connected
Chambers.
VOLUME
CAPACITY
• 3.5 mL
• 35 mL
• Not Detected
6
MESSAGES
Several status messages are available to indicate the present state of the system. The
CHAMBER NOT DETECTED – This message indicates that a Chamber has not been
detected.
READY FOR FLOW– This message indicates that the Chamber is reset and waiting for
fluid to be detected at the bottom sensor. As soon as fluid is detected, the timer will begin
running and the test will begin. The test ends when the fluid reaches the top sensor or
FLOW RATE TOO HIGH – This message indicates that the flow rate is higher than the
INVALID SENSOR SEQUENCE– This message indicates that Distilled Water is detected
at the top sensor but not at the bottom sensor. This could be caused by either a bad
CHAMBER NEEDS CALIBRATION– This message indicates that a Chamber has been
detected, but the calibration data stored in the Chamber is invalid. The Chamber should
be serviced.
7
LOW BATTERY– This message indicates that the batteries are low and should be
replaced.
LEVEL INDICATOR
A special graphic has been incorporated into the display to identify the level of Distilled
Water in each Chamber. The graphic is located at the lower right corner of the display for
each Chamber.
This graphic indicates that Distilled Water is not present at the lower or upper
sensor.
This graphic indicates that Distilled Water is detected at the top and bottom sensor.
This graphic indicates that Distilled Water is present at the top sensor, but not the
8
SETUP
The Setup Mode allows the user to adjust the configuration of the meter. The Setup
screen can be entered using the key. The parameters can be changed by using
The following is a breakdown of the parameters available in the configuration of the unit
9
OPERATIONS
Chambers are connected to the base unit using the Chamber Interconnect Cable
available chambers. For example, if a test is running on Channel 1 and a second Chamber
is plugged into Channel 2, the display will automatically change to the dual display mode.
If both chambers are connected and Chamber 1 is removed, the display will automatically
NOTE: The screen can always be manually changed by using the keys.
10
THEORY OF OPERATIONS
Infrared sensors are used to detect the presence of Distilled Water in the chamber.
There are two sensors in each chamber. As Distilled Water is detected at the bottom
sensor, an internal timer is started. The timer runs internally in hundredths of seconds but
is only displayed in tenths of seconds. The test automatically ends when Distilled Water is
detected at the top sensor. A buzzer will sound to indicate that the test is complete. The
buzzer can be programmed to sound for a specific number of beeps (0-15 beeps, see
setup).
The flow rate is calculated based on the volume of the chamber and the time
required to fill the tube. The chamber rating (3.5 mL or 35 mL) represents the nominal
volume between the two sensors. For maximum accuracy, the precise volume of each
individual chamber is calibrated and digitally stored in the chamber memory. Since the
calibration data is stored in the chamber, they are interchangeable with any base unit.
the chamber or by Distilled Water not being detected at the bottom sensor.
11
RUNNING A TEST
To run a flow test, the chamber must be Ready for Flow, as indicated in the base
unit display (see Messages). When the base unit is setup for Manual Reset, the
key must be used to manually reset the unit after each test. When the base unit is setup
for Automatic Reset, the chamber will be automatically reset when the Distilled Water level
A flow test can be started whenever the display indicates “Ready for Flow”. The flow
test is started by Distilled Water being detected at the bottom sensor. Before initiating the
flow on the Infusion Pump under test, ensure that the Infusion Pump is programmed to
deliver a sufficient amount of Distilled Water to fill the chamber to the top sensor.
NOTE: The chamber volumes of 3.5mL and 35mL are nominal; it is advised that delivery
CAUTION - FLUID
Only Distilled Water should be used in the
chambers with the IPA-1000 System. Do not use
tap water, glucose or any other fluid; this will
cause the tubing to become contaminated.
12
OPTIMIZING MEASUREMENTS
wetting agent be used to minimize the surface tension of the test solution. The
(www.coleparmer.com).
Water. If foaming becomes a problem, change the ratio to 0.5 mL of “MICRO-90 ®” per 1
The sheeting action of the test solution has a minor effect on test results. It is
recommended to discard the results of the first test and only record data from the second
Any droplets remaining in the chamber after a test run will affect the accuracy of the
13
MANUAL REVISIONS
LIMITED WARRANTY
WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\MANUALS\BCGroup\…\IPA-1000_UM_Rev06.doc
14
SPECIFICATIONS
ELECTRICAL
9V Alkaline Battery
BATTERY (ANSI/NEDA 1604A or
equivalent)
POWER 9 VDC, 200 mA
BATTERY
ELIMINATOR BC20-21100 (USA Version)
BC20-21101 (Euro Version)
8 mA (No Chambers)
ON 12 mA (1 Chamber)
POWER CONSUMPTION 16 mA (2 Chambers)
OFF < 60 µA
CONTINUOUS > 24 Hours
BATTERY LIFE
OFF 12 Months
15
MC3.5-1000 & MC35-1000 CHAMBERS
FLOW
3.5 mL Chamber 0 to 999.9 mL/hr
RANGE
35 mL Chamber 15.0 to 9999.9 mL/hr
ELECTRICAL
POWER Powered by IPA-1000 Unit
8-pin Female Mini-DIN Receptacle
16
NOTES
17
NOTES
18
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
Copyright © 2012
Made in the USA