Aliments Des Animaux - Détermination de La Teneur en Vitamine E
Aliments Des Animaux - Détermination de La Teneur en Vitamine E
STANDARD 6867
First edition
2000-12-01
--`,,```,,,,````-`-`,,`,,`,`,,`---
Reference number
ISO 6867:2000(E)
© ISO 2000
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall not
be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In downloading this
file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat accepts no liability in this
area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation parameters
were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the unlikely event
that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2000
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic
or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO's member body
in the country of the requester.
ISO copyright office
Case postale 56 · CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail [email protected]
Web www.iso.ch
Printed in Switzerland
--`,,```,,,,````-`-`,,`,,`,`,,`---
ii
Copyright International Organization for Standardization © ISO 2000 – All rights reserved
Provided by IHS under license with ISO
No reproduction or networking permitted without license from IHS Not for Resale
ISO 6867:2000(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO
member bodies). The work of preparing International Standards is normally carried out through ISO technical
committees. Each member body interested in a subject for which a technical committee has been established has
the right to be represented on that committee. International organizations, governmental and non-governmental, in
liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical
Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 3.
Draft International Standards adopted by the technical committees are circulated to the member bodies for voting.
Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this International Standard may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights.
International Standard ISO 6867 was prepared by Technical Committee ISO/TC 34, Food products, Subcommittee
SC 10, Animal feeding stuffs.
--`,,```,,,,````-`-`,,`,,`,`,,`---
Annex A of this International Standard is for information only.
1 Scope
This International Standard specifies a method for the determination of the vitamin E (DL-a-tocopherol) content of
animal feeding stuffs and pet foods using high performance liquid chromatography.
2 Normative references
The following normative documents contain provisions which, through reference in this text, constitute provisions of
this International Standard. For dated references, subsequent amendments to, or revisions of, any of these
publications do not apply. However, parties to agreements based on this International Standard are encouraged to
investigate the possibility of applying the most recent editions of the normative documents indicated below. For
undated references, the latest edition of the normative document referred to applies. Members of ISO and IEC
maintain registers of currently valid International Standards.
ISO 3696:1987, Water for analytical laboratory use — Specifications and test methods.
3 Principle
A test portion of the sample is saponified with ethanolic potassium hydroxide solution and the vitamin E is extracted
into light petroleum. The light petroleum is removed by evaporation and the residue is dissolved in hexane. The
vitamin E concentration in the hexane extract is determined by normal-phase liquid chromatography using
conditions that separate DL-a-tocopherol from other tocopherols.
4 Reagents
Use only reagents of recognized analytical grade, unless otherwise stated.
4.1 Water, complying with at least grade 3 in accordance with ISO 3696.
4.5 Light petroleum, boiling range 40 °C to 60 °C; the residue on evaporation shall be less than 20 mg/l.
4.6 Vitamin E standard substance: DL-a-tocopherol, minimum purity not less than 96,0 %.
The purity of the standard substance should be checked spectrophotometrically (see 8.5.2).
--`,,```,,,,````-`-`,,`,,`,`,,`---
5 Apparatus
--`,,```,,,,````-`-`,,`,,`,`,,`---
Using laboratory apparatus and, in particular, the following.
5.1.1 Pump, set to deliver a constant eluent volume flow rate of 1,5 ml/min.
5.1.3 Column, length 250 mm, internal diameter 4,6 mm, packed with a stationary phase consisting of silica.
A column with at least 5 000 theoretical plates and a k¢ value of 0,8 m, both with respect to DL-a-tocopherol, has
been found to be satisfactory. The particle size should not be smaller than 5 µm and not greater than 10 µm. Other
systems may be used provided that a satisfactory separation of vitamin E from other co-extractives is achieved.
5.1.4 Detector, allowing the measurement of fluorescence emitted at a wavelength of 326 nm when the column
eluent is irradiated with ultraviolet light at a wavelength of 293 nm, with integrator/recorder.
¾ a cylinder of 1 litre capacity fitted with a ground glass neck and stopper;
¾ a ground glass joint, fitting the cylinder and equipped with an adjustable tube passing through the centre; and
¾ a side-arm.
The adjustable tube should have a U-shaped lower end and a jet at the opposite end so that the upper liquid layer
in the cylinder may be transferred to a separating funnel of 1 litre capacity.
Other extraction equipment such as conical flasks and separating funnels may be used in place of the apparatus
shown in Figure 1, provided that satisfactory recoveries of vitamin E are achieved.
5.5 Membrane filter, 0,45 m pore size, for filtration of mobile phase (4.11) and sample test solutions.
2
Copyright International Organization for Standardization © ISO 2000 – All rights reserved
Provided by IHS under license with ISO
No reproduction or networking permitted without license from IHS Not for Resale
ISO 6867:2000(E)
5.6 Grinding apparatus, capable of grinding the sample so that it passes through a sieve with 1 mm apertures.
5.7 UV (or UV/Visible) spectrometer, capable of measuring absorbance at the wavelengths defined in 8.5.2,
equipped with quartz cells of 10 mm path length.
Key
1 Cylinder, of capacity 1 litre, with ground-glass neck 5 Bottle, of capacity 1 litre, with ground-glass joint
2 Light petroleum layer 6 Side-arm
3 Aqueous layer + saponified feed 7 Adjustable tube
4 Jet
--`,,```,,,,````-`-`,,`,,`,`,,`---
6 Sampling
It is important that the laboratory receive a sample which is truly representative and has not been damaged or
changed during transport or storage.
Sampling is not part of the method specified in this International Standard. A recommended sampling method is
given in ISO 6497 [1].
Store the sample in such a way that deterioration and change in its composition are prevented.
Just prior to starting the analysis, grind a portion of the well-mixed laboratory sample so that it passes through a
sieve with 1 mm apertures. Mix thoroughly.
Homogenize canned pet foods. Pass semi-moist pet foods through a mincer with 4-mm apertures.
8 Procedure
8.1 General
Because of the sensitivity of vitamin E to UV radiation and air, perform all operations away from natural and strong
fluorescent light and as rapidly as is consistent with accurate working. Use amber glassware where possible.
Complete each assay within one working day.
8.2 Saponification
--`,,```,,,,````-`-`,,`,,`,`,,`---
Weigh, to the nearest 0,1 g, approximately 50 g of the prepared sample (see clause 7) into a 1 litre conical flask.
Add to the test portion 200 ml of ethanol (4.3) whilst swirling the flask to disperse the sample. Add 2 ml of sodium
ascorbate solution (4.9), mix by swirling and then add 50 ml of potassium hydroxide solution (4.2) and swirl again.
Fit a reflux condenser to the flask and immerse the flask in the boiling water bath (5.2).
Allow the contents of the flask to reflux for 30 min, swirling occasionally.
NOTE In exceptional cases some products may require a longer saponification time.
Transfer the contents of the flask into the extraction cylinder (see 5.4).
Rinse the saponification flask with two 25 ml portions of ethanol (4.3) and transfer the rinsings to the cylinder.
Repeat the rinsing of the flask with two 125 ml portions of light petroleum (4.5) and one 250 ml portion of water
(4.1), each time transferring the rinsings to the cylinder.
Stopper the cylinder and shake well for 1 min, releasing the pressure from time to time.
Cool the cylinder under a stream of cold water while waiting for the two liquid phases to separate, before removing
the stopper.
4
Copyright International Organization for Standardization © ISO 2000 – All rights reserved
Provided by IHS under license with ISO
No reproduction or networking permitted without license from IHS Not for Resale
ISO 6867:2000(E)
When the layers have separated, remove the stopper, wash the sides of the stopper with a few millilitres of light
petroleum (4.5) and insert the adjustable tube (see 5.4), positioning the lower open end so that it is just above the
level of the interface.
By application of a slight pressure of inert gas (4.10) to the side arm tube, transfer the upper, light petroleum layer
to a 1 litre separating funnel (see 5.4).
Add 125 ml of light petroleum (4.5) to the cylinder, stopper and shake well for 1 min.
Allow the layers to separate and transfer the upper layer to the separating funnel using the adjustable tube
(see 5.4) as before.
Again, add 125 ml of light petroleum (4.5) to the cylinder, stopper and shake well for 1 min.
Again, allow the layers to separate and transfer the upper layer to the separating funnel using the adjustable tube
as before.
Wash the combined light petroleum extracts with four 100 ml portions of water using at first only gentle inversion
then only gentle shaking in order to keep emulsion formation to a minimum.
Transfer the washed extract through a medium/fast filter paper containing 30 g of anhydrous sodium sulfate (4.8)
into a 1 litre flask suitable for vacuum evaporation (5.3).
Rinse the separating funnel with two 20 ml portions of light petroleum (4.5) and add the rinsings through the filter to
the evaporation flask.
Wash the filter further with two 25 ml portions of light petroleum (4.5) and collect the washings in the evaporation
flask.
Evaporate the light petroleum extract to dryness under vacuum at a temperature not exceeding 40 °C.
Care should be taken to ensure that the flask is removed from the rotary evaporator immediately after reaching the
point of dryness; prolonged drying may lead to loss of vitamin E from the extract residue.
If the vitamin E concentration of the light petroleum extract is sufficiently high, the extract may be made up to a
fixed volume with light petroleum and an aliquot part taken for the rotary evaporation stage.
8.4 Determination
8.4.1 Dissolve the residue from 8.3 in a minimum volume of hexane (4.4) and transfer quantitatively to a 25 ml
volumetric flask.
Rinse the evaporation flask with three small portions of hexane (4.4), transferring the rinsings to the volumetric
flask. Dilute to volume with hexane and mix.
If necessary, filter the sample extract through a membrane filter (5.5) or centrifuge.
8.4.2 Inject 20 µl of the sample extract onto the column of the liquid chromatograph (5.1) and measure the area
of the DL-a-tocopherol peak. The following HPLC conditions are offered for guidance; other conditions may be used
provided that they give equivalent results:
¾ liquid chromatographic column (5.1.3): 250 mm ´ 4,6 mm, silica 5 µm or 10 µm packing, or equivalent;
¾ mobile phase (4.11): mixture of hexane (4.4) and 1,4-dioxan (4.7), 970:30 (by volume);
¾ detector (5.1.4): fluorescence detector (excitation 295 nm, emission 330 nm).
--`,,```,,,,````-`-`,,`,,`,`,,`---
Reverse-phase chromatography may also be used provided that the efficiency of the column is sufficient to allow
the separation of DL-a-tocopherol from other tocopherols and sample co-extractives. If reverse-phase
chromatography is used, sample and standard solutions should be made up in an appropriate solvent, for example
methanol (4.13).
8.4.3 Calculate the mean peak area from replicate injections of the sample extract and determine the
DL-a-tocopherol concentration in micrograms per millilitre of the extract, either
a) by reference to the mean peak areas obtained from replicate injections of DL-a-tocopherol standard solution of
concentration within 5 % of the concentration in the sample extract, or
8.5 Calibration
8.5.1.1 =-tocopherol
DL-= stock standard solution
Dissolve approximately 100 mg, weighed to the nearest 0,1 mg, of DL-a-tocopherol (4.6) in 100 ml of hexane (4.4).
The stock standard solution is stable for 1 week when stored at u 4 °C in an airtight amber glass flask.
8.5.1.2 =-tocopherol
DL-= working standard solution: single-point calibration
Prepare a working standard by diluting the stock standard (8.5.1.1) with hexane (4.4) to give a concentration
approximately equal to that expected in the sample extract. Alternatively, proceed as in 8.5.1.3.
8.5.1.3 =-tocopherol
DL-= working standard solution: multi-point calibration
Prepare a range of calibration working standards containing 2 µg/ml, 4 µg/ml, 6 µg/ml, 8 µg/ml and 10 µg/ml of
DL-a-tocopherol by diluting the stock standard solution (8.5.1.1) with hexane (4.4).
Weigh, to the nearest 0,1 mg, 100 mg of DL-a-tocopherol (4.6) in a 100 ml volumetric flask. Dissolve in ethanol
(4.12). Dilute to the mark with the same solvent and mix.
Dilute 2,0 ml of this solution to 25,0 ml with ethanol (4.12 ) and measure the UV spectrum of the resulting solution
against ethanol (4.12) in the spectrometer (5.7) at wavelengths of between 250 nm and 320 nm. The absorption
maximum should be at 292 nm:
1%
E 1cm = 75,8 at 292 nm in ethanol.
--`,,```,,,,````-`-`,,`,,`,`,,`---
6
Copyright International Organization for Standardization © ISO 2000 – All rights reserved
Provided by IHS under license with ISO
No reproduction or networking permitted without license from IHS Not for Resale
ISO 6867:2000(E)
9 Expression of results
wE = 25 ´ c ´ 1,1
m
where
wE is the numerical value of the vitamin E content of the test sample, in International Units per kilogram;
c is the numerical value of the DL-a-tocopherol concentration of the extract, in micrograms per millilitre;
10 Precision
Details of a laboratory test on the precision of the method are summarized in annex A. The values derived from
these tests may not be applicable to concentration ranges and matrices other than those given.
--`,,```,,,,````-`-`,,`,,`,`,,`---
10.2 Repeatability
The absolute difference between two independent single test results, obtained using the same method on identical
test material in the same laboratory by the same operator using the same equipment within a short interval of time,
will in not more than 5 % of cases exceed the repeatability limit (r) mentioned in or derived from Table 1.
10.3 Reproducibility
The absolute difference between two single test results, obtained using the same method on identical test material
in different laboratories by different operators using different equipment, will in not more than 5 % of cases exceed
the reproducibility limit (R) mentioned in or derived from Table 1.
11 Test report
¾ all operating details not specified in this International Standard, or regarded as optional, together with details of
any incidents which may have influenced the test result(s);
¾ the test results obtained, or the two test results obtained if the repeatability has been checked.
--`,,```,,,,````-`-`,,`,,`,`,,`---
8
Copyright International Organization for Standardization © ISO 2000 – All rights reserved
Provided by IHS under license with ISO
No reproduction or networking permitted without license from IHS Not for Resale
ISO 6867:2000(E)
Annex A
(informative)
The precision of the method was established by interlaboratory tests carried out in accordance with ISO 5725 [2]1).
The results of these tests have been published (see reference [5]). In the tests 10 to 12 laboratories participated
and samples of pet foods, cattle, poultry and pig feeds were investigated.
Table A.1 — Statistical results of an interlaboratory test for determination of vitamin E in feeding stuffs
Parameter Samplea
1 2 3 4b 5b
Number of laboratories 10 10 10 12 12
Number of individual results 20 20 20 24 24
Number of accepted results 14 16 16 18 18
Mean vitamin E content, IU/kg 23,11 29,66 64,94 78,03 140,63
Repeatability standard deviation (sr), IU/kg 0,44 0,69 1,35 1,66 5,27
Repeatability coefficient of variation, % 1,92 2,32 2,07 2,12 3,75
Repeatability limit (r) [r = 2,8 ´ sr], IU/kg 1,36 2,07 4,06 4,94 15,73
Reproducibility standard deviation (sR), IU/kg 1,15 3,33 5,79 6,13 14,72
Reproducibility coefficient of variation, % 4,97 11,21 8,91 7,86 10,46
Reproducibility limit (R) [R= 2,8 ´ sR], IU/kg 3,51 10,02 17,44 18,31 43,91
a 1: cattle ration
2: poultry ration
3: pig ration
4: animal feed A (ration 722)
5: animal feed B (ration 748)
b Calculated on dry matter basis.
--`,,```,,,,````-`-`,,`,,`,`,,`---
1) ISO 5725:1986 (now withdrawn) was used to obtain the precision data.
Table A.2 — Statistical results of an interlaboratory test for the determination of vitamin E in pet foods
Parameter Samplea
6 7 8
Number of laboratories 11 11 11
Number of individual results 22 22 22
Number of accepted results 14 14 14
Mean vitamin E content, IU/kg 20,55 180,39 78,66
Repeatability standard deviation (sr), IU/kg 0,32 4,98 1,38
--`,,```,,,,````-`-`,,`,,`,`,,`---
10
Copyright International Organization for Standardization © ISO 2000 – All rights reserved
Provided by IHS under license with ISO
No reproduction or networking permitted without license from IHS Not for Resale
ISO 6867:2000(E)
Bibliography
[2] ISO 5725:1986, Precision of test methods — Determination of repeatability and reproducibility for a
standard test method by inter-laboratory tests.
[3] ISO 5725-1:1994, Accuracy (trueness and precision) of measurement methods and results — Part 1:
General principles and definitions.
[4] ISO 5725-2:1994, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic
method for the determination of repeatability and reproducibility of a standard measurement method.
--`,,```,,,,````-`-`,,`,,`,`,,`---
ICS 65.120
Price based on 11 pages