173028ug en MK
173028ug en MK
1 Spectrophotometers................................................... 8
1.1 Photometry.............................................................................................................. 8
3
1.2 The Spectrophotometers............................................................................................ 9
3 Sample Preparation.................................................. 16 7
3.1 Taking Samples ..................................................................................................... 16
3.2 Preliminary Tests ................................................................................................... 16
3.3 Dilution................................................................................................................. 17
3.4 Filtration ............................................................................................................... 18 8
3.5 Homogenization...................................................................................................... 19
3.6 Decomposition ....................................................................................................... 19
4 Pipetting System....................................................... 21 9
5 Analytical Quality Assurance (AQA).......................... 22
5.1 Quality Control at the Manufacturer .......................................................................... 22
5.2 Quality Control for the User ..................................................................................... 24 10
5.2.1 Checking the Spectrophotometer................................................................... 25
5.2.2 Checking the Overall System........................................................................ 26
5.2.3 Checking the Pipettes.................................................................................. 27
5.2.4 Checking Thermoreactors ............................................................................ 27 11
5.2.5 Testing for Handling Errors .......................................................................... 28
5.3 Determination of Sample Influences (Matrix Effects) ................................................... 28
5.4 Definition of Errors.................................................................................................. 28
12
6 Overview.................................................................. 30
6.1 Scope of Delivery.................................................................................................... 30
6.2 Overview of the Instrument...................................................................................... 30
6.3 Display and User Interface....................................................................................... 31 13
7 Safety....................................................................... 39
7.1 Intended Use......................................................................................................... 40
7.2 General Safety Instructions...................................................................................... 40 14
7.2.1 Function and Operational Safety ................................................................... 40
7.3 Target Group and User Qualification.......................................................................... 40
7.4 Handling Hazardous Substances................................................................................ 41
15
16
Version 1.0 – 01/2024 3
1
8 Getting Started......................................................... 42
8.1 General Notes on Handling....................................................................................... 42
3
8.2 Initial Setup........................................................................................................... 42
8.2.1 Connecting the Power Supply........................................................................ 42
8.2.2 First Power-on............................................................................................ 43
8.2.3 Language Setup.......................................................................................... 44
4
8.2.4 Date, Time and Country-specific Settings........................................................ 44
8.2.5 Self-test..................................................................................................... 45
8.3 Connecting Optional Peripheral Devices...................................................................... 46
5 8.3.1 Communication Ports................................................................................... 46
8.3.2 Printer....................................................................................................... 46
8.3.3 USB Memory Device.................................................................................... 47
8.3.4 Barcode Reader........................................................................................... 47
6
9 Operation.................................................................. 48
9.1 Switching the Spectrophotometer On or Off................................................................ 48
9.2 System Setup ........................................................................................................ 50
7 9.2.1 Information................................................................................................ 51
9.2.2 Interface (Setup 1)...................................................................................... 52
9.2.3 Region (Setup 2)......................................................................................... 52
9.2.4 Quality (Setup 3)........................................................................................ 55
8 9.2.5 Automation (Setup 4).................................................................................. 57
9.2.6 User Management....................................................................................... 59
9.2.7 Service...................................................................................................... 60
9.2.8 Updates..................................................................................................... 62
9 9.2.9 Network and Prove Connect.......................................................................... 63
9.3 Measurements........................................................................................................ 64
9.3.1 Performing a Measurement........................................................................... 65
9.4 Zero Adjustment .................................................................................................... 67
10 9.4.1 Notes on Zero Adjustment ........................................................................... 67
9.4.2 When to Repeat the Zero Adjustment? .......................................................... 68
9.4.3 Zero Adjustment for Concentration Measurement Methods................................ 69
9.4.4 Zero Adjustment for Absorbance/Transmission Measurements (Ad hoc menu)...... 69
11 9.4.5 Zero Adjustment for Spectrum Measurements ................................................ 70
9.4.6 Zero Adjustment for Kinetic Measurements .................................................... 70
9.5 Method List............................................................................................................ 70
9.5.1 Selecting a Method Manually ........................................................................ 70
12 9.5.2 Searching and Filtering the Method List ......................................................... 71
13
14
15
16
4 Version 1.0 – 01/2024
1
12
13
14
15
16
Version 1.0 – 01/2024 5
1
9.10 Kinetics................................................................................................................128
9.10.1 General Information...................................................................................128
3 9.10.2 Recording Kinetics......................................................................................128
9.10.3 Evaluating a Kinetic....................................................................................130
9.11 AQA (Analytical Quality Assurance)..........................................................................131
9.11.1 Spectrophotometer Monitoring (AQA1)..........................................................131
4 9.11.2 Total System Monitoring (AQA2)...................................................................133
9.11.3 AQA Overview............................................................................................134
9.11.4 Perform AQA1 Status Check.........................................................................135
9.11.5 Perform AQA2 Status Check.........................................................................136
5 9.11.6 AQA1 Selection List....................................................................................137
9.11.7 Activating and Deactivating an AQA1 Test......................................................138
9.11.8 Editing an AQA1 Test..................................................................................138
9.11.9 Performing an AQA1 Test.............................................................................141
6 9.11.10 Creating a User-defined AQA1 Test ..............................................................143
9.11.11 AQA2 Selection List....................................................................................144
9.11.12 Activating and Deactivating an AQA2 Test......................................................145
9.11.13 Editing an AQA2 Test..................................................................................145
7
9.11.14 Performing an AQA2 Test.............................................................................148
9.11.15 Perform PipeCheck.....................................................................................150
9.12 Timer...................................................................................................................153
9.13 Results and Measurement Datasets............................................................................154
8
9.13.1 Displaying the Results ................................................................................155
9.13.2 Show Details of One Result..........................................................................156
9.13.3 Filtering Results to Further Process Measurement Datasets..............................157
9 9.13.4 Panorama – Value Control Card ...................................................................158
9.13.5 Print Results and Measurement Datasets.......................................................159
9.13.6 Deleting results .........................................................................................160
9.13.7 Export Results and Measurement Datasets.....................................................160
10 Data Transfer from Spectroquant® Prove plus with USB stick............................160
9.14 User Management..................................................................................................162
9.14.1 Activating/Deactivating the User Management Function...................................163
9.14.2 Create an Administrator/User Account...........................................................164
11 9.14.3 Edit or Delete a User...................................................................................165
9.15 Login/Logout.........................................................................................................166
9.15.1 Change Password with User Rights Only........................................................167
12
13
14
15
16
6 Version 1.0 – 01/2024
1
14 Appendix............................................................... 186 8
15 List of Smart Icons on Display.............................. 188
16 Contents of log files.............................................. 195
16.1 Error log file..........................................................................................................195 9
16.2 User log file..........................................................................................................195
16.3 Service log file.......................................................................................................196
10
11
12
13
14
15
16
Version 1.0 – 01/2024 7
1
1 Spectrophotometers
1.1 Photometry
When a beam of light is transmitted through a ized using the transmittance T (or, respectively,
3
colored solution, then this beam loses its intensi- T in percent).
ty, in other words a part of the light is absorbed
by the solution. Depending on the substance T = I/I0
in question, this absorption occurs at specific
4
wavelengths. I0 = Initial intensity of the light
I = Intensity of the transmitted light
Monochromators (e. g. narrow-band interference
5 filters, lattices) are used to select the wave- If the light is not absorbed at all by a solution,
length from the total spectrum of a tungsten- then this solution has a transmittance of 100%;
halogen lamp (VIS spectrum), a deuterium lamp a complete absorption of the light in the solution
(UV spectrum) or, respectively, a xenon lamp. means 0% transmittance.
6 The intensity of the absorption can be character-
The measure generally used for the absorption
of light is the absorbance (A), since this cor-
relates directly with the concentration of the
7 absorbing substance. The following connection
exists between absorbance and transmittance:
A = –log T
8
Experiments by BOUGUER (1698–1758) and
LAMBERT (1728–1777) showed that the ab-
sorbance is dependent on the thickness of the
9 absorbing layer of the cell used. The relationship
between the absorbance and the concentration
of the analyte in question was discovered by
BEER (1825–1863). The combination of these two
10 l natural laws led to the derivation of Lambert-
c Beer’s law, which can be described in the form of
l0 the following equation:
11 A = ε λ · c · d
13
14
15
16
8 Version 1.0 – 01/2024
1 Spectrophotometers – 1.2 The Spectrophotometers 1
11
12
13
14
15
16
Version 1.0 – 01/2024 9
1
2 Photometric Test Kits
Additional reagent(s)
13
Certain cell tests, e. g. COD or nitrite, already
contain all necessary reagents in the cells, and
the sample must merely be added with a pi-
pette. In other tests, however, for reasons of
14
chemical compatibility it is necessary to sepa-
rate the test into two or three different reagent
mixtures. In such cases, besides the sample a
15 metered reagent must also be added.
16
10 Version 1.0 – 01/2024
2 Photometric Test Kits – 2.1 Basic Principle 1
11
12
13
14
15
16
Version 1.0 – 01/2024 11
1 2 Photometric Test Kits – 2.2 Notes for Practial Use
15
16
12 Version 1.0 – 01/2024
2 Photometric Test Kits – 2.2 Notes for Practial Use 1
3
2.2.2 Influence of pH 2.2.3 Influence of Temperature
Chemical reactions follow an optimal course
only within a certain pH range. The reagents
contained in the test kits produce an adequate
4
buffering of the sample solutions and ensure
that the pH optimal for the reaction in question
is obtained.
5
Absorbance
Strongly acidic (pH <2) and strongly alkaline (pH
>12) sample solutions can prevent the pH from
being adjusted to an optimal range, since under
6
certain circumstances the buffering capacity
of the test-kit reagents may not be sufficient.
Any necessary correction is made by the drop-
wise addition of diluted acid (reduces the pH) or 7
diluted lye (raises the pH), testing the pH with
suitable indicator strips after each drop is added.
The addition of the acid or lye results in a dilu-
tion of the test solution. When up to five drops 10 20 30 40
8
are added to 10ml of sample, the change in the Temperature (°C)
volume can be neglected, since the resultant er-
ror is lower than 2%.
The temperature of the sample solution and the
The addition of larger quantities should be duly
reagents may have an effect on the color reac- 9
considered by adjusting the sample volume ac-
tion and thus on the measurement result. The
cordingly.
typical temperature course is illustrated in the
figure.
The specified pH values for the sample solution 10
and, wherever applicable, for the measurement
If the sample temperature is lower than 15 °C,
solution are defined in the respective package
false-low results must be reckoned with. Tem-
inserts and in the Analytical Procedures and Ap-
peratures exceeding 30 °C generally influence
pendices. 11
the stability of the compound that is formed in
the reaction. The optimal temperature for the
color reaction is stated in the package inserts of
the respective Spectroquant® test kits.
12
CAUTION
After thermic decomposition procedures, the de
termination of COD or total contents of nitrogen, 13
phosphorus, or metal, a sufficient waiting time
must be allowed for to permit the solution cool to
room temperature.
14
15
16
Version 1.0 – 01/2024 13
1 2 Photometric Test Kits – 2.2 Notes for Practial Use
13
14
15
16
14 Version 1.0 – 01/2024
2 Photometric Test Kits – 2.2 Notes for Practial Use 1
CAUTION
5
Reclose the reagent bottle with the black screw
cap at the end of the measurement series, since
the function of the reagent is impaired by the
absorption of atmospheric moisture.
6
CAUTION
When using dropper bottles it is extremely im-
portant that the bottle be held vertically and 7
that the drops be added slowly (approx. 1 drop 2.2.7 Shelf-life of the Reagents
per second). If this is not observed, the correct
The Spectroquant® test kits are in most cases
drop size and thus the correct amount of reagent
stable for three years when stored in a cool, dry
are not achieved. 8
place. A few test kits have a lower shelf-life of 18
or 24 months or must else be stored in a refrig-
A positive-displacement pipette should be used erator.
for larger quantities of liquid or for the exact
dosage of smaller reagent quantities. In these COD Cell Tests must be stored protected from 9
cases the reagent bottles are not fitted light. The expiry date of the package unit is
with a dropper insert. printed on the outer label. The shelf-life may be-
come reduced when the reagent bottles are not
reclosed tightly after use or when the test kit is
10
stored at temperatures higher than those speci-
fied.
11
12
16
Version 1.0 – 01/2024 15
1
3 Sample Preparation
15
16
16 Version 1.0 – 01/2024
3 Sample Preparation – 3.3 Dilution 1
3.3 Dilution
Dilution of samples is necessary for two reasons: Example
3
Step 1: ake up 2 ml of sample to 200 ml
M
• The
concentration of the parameter under with distilled water;
investigation is too high, i. e. it lies outside the DF = 100, dilution number 1+99.
measuring range
4
• Other substances contained in the sample Step 2: ake 5 ml of the above solution
T
interfere with the determination (matrix inter- and make up to 100 ml;
ference); false-high or false-low results may DF = 20, dilution number 1+19.
ensue 5
The dilution factor for the total dilution is calcu
The following auxiliaries are absolute prerequi- lated by multiplying the individual dilutions:
sites for the dilution of the sample:
DFtotal = DF1 × DF2 = 100 × 20 = 2000, 6
•V olumetric flasks of varying sizes (e. g. 50, 100 dilution number 1+1999
and 200 ml)
• Positive-displacement pipette Simplified procedure
• Distilled or DI water Dilutions up to 1:10 can also be prepared without 7
volumetric flasks in a glass beaker, measuring
Only dilutions carried out with these auxiliary the volumes of the sample and the dilution water
products are of sufficient reliability in the area of using a previously calibrated positive-displace-
trace analysis, to which photometry belongs (for ment pipette (see table for instructions). 8
the simplified procedure see page 17).
An important aspect here is that once the volu- Desired Volume Volume Dilution Dilution
metric flask has been filled up to the mark with dilution of of dis- factor number
distilled water the flask must be closed and the sample tilled 9
contents thoroughly mixed. (ml) water
The dilution factor (DF) resulting from the (ml)
16
Version 1.0 – 01/2024 17
1 3 Sample Preparation – 3.4 Filtration
3.4 Filtration
Strongly turbid samples require pretreatment Procedure for Microfiltration
3
before they can be determined in a spectropho-
tometer, since the effect of turbidity can result 1 2
in considerable variations in the measurement
values and in false-high readings. Care must be
4
taken here to ensure that the substance to be
determined is not contained in the suspended
material, in which case a sample decomposition
5 must be carried out.
12
13
14
15
16
18 Version 1.0 – 01/2024
3 Sample Preparation – 3.5 Homogenization – 3.6 Decomposition 1
3.5 Homogenization
As a measure to ensure that a representative
3
sample can be taken in the presence of sus-
pended matter in the water sample in question,
for certain parameters – e. g. COD and the total
content of heavy metals – the sample must be
4
homogenized. This must be carried out using a
high-speed blender (2 minutes at 5000 – 20000
rpm and taking the sample while stirring.
5
3.6 Decomposition
Water-borne substances can be present in the The manner in which the sample is pretreated 6
sample for investigation in a variety of forms: as enables the three proportions to be distinguished
the ion, bound more or less solidly in a complex, from each other. This can be illustrated using
or as a solid substance. a copper-containing wastewater sample as an
example. 7
Ion Complex
Example
8
Filtration
9
Proportion:
Ionogenic =C
Complex = B–C 12
Solid
substances = A–B
Total content = A
13
Decomposition converts the substance to be
determined into an analyzable form. In most
cases, decomposition agents take the form of
acids in combination with oxidizing agents; in 14
exceptional cases (e. g. in the determination of
total nitrogen) an alkaline decomposition is more
effective. The type of decomposition procedure
used depends on the analyte to be determined 15
and the sample matrix.
16
Version 1.0 – 01/2024 19
1 3 Sample Preparation – 3.6 Decomposition
The ready-to-use sample-decomposition prod- The necessity for decomposition can be checked
3
ucts Spectroquant® Crack Set 10 and 20 are according to the following diagram.
suited for the preparation of the sample materi-
als for the determinations stated in the table.
Decomposition
4
Determination of Sample preparation with Procedure Procedure
Total phosphorus * Crack Set 10/10C **
Measurement Measurement
5 Total chromium * Crack Set 10/10C
[= sum of chromate and
Result A Result B
chromium(III)]
13
14
15
16
20 Version 1.0 – 01/2024
1
4 Pipetting System
11
12
13
14
15
16
Version 1.0 – 01/2024 21
1
5 Analytical Quality Assurance (AQA)
3
The objective of analysis must always be to de-
termine the true content of the analyte in ques- Certificate of Final Inspection (full details)
tion as accurately and precisely as possible.
Device Name: Spectroquant® Prove 600
quality of the user's own work can be assessed, Equipment Nominal value Tolerance limit ⃰ ⃰ Actual value Result
and the comparability with the results obtained 640.55 nm 639.4 - 641.8 nm 641.1 nm P
≤0.10 nm
0.01 nm P
P
Details regarding the necessity of AQA can be
361.30 nm 0.02 nm
Hellma 667-UV5
640.55 nm ≤0.10 nm 0.06 nm P
7 Waste Materials (Deutsche Vereinigung für Was- 440 nm 1.093 A 1.082 - 1.104 A 1.092 A P
itoring regulations of the German federal states Neutral Density 2.0 Abs.
Hellma 666-F203
546 nm 1.994 A 1.982 - 2.005 A 1.995 A P
8 (available in English).
635 nm 1.928 A 1.916 - 1.940 A 1.931 A P
Causes for errors can include: Equipment Wavelength Nominal value Actual value Result
9
Hellma 666-F4
Page 1 of 2
11
5.1 Quality Control at the
Stray Light ⃰
Manufacturer Equipment Wavelength Nominal value Actual value Result
12 possess specifications that are adhered to and Sodium Nitrite 340.00 nm ≤0.10 %T 0.00 %T P
14 Date: 29/01/2020
Inspector: gberg
- This document has been generated using electronic data processing and is valid without signature. -
16
22 Version 1.0 – 01/2024
5 Analytical Quality Assurance (AQA) – 5.1 Quality Control at the Manufacturer 1
3
The certificate for the test kit, available for Calibration function:
each lot produced, documents the quality of the The calculated function must agree, within speci-
reagents contained in the test kit. fied tolerances, with the function electronically
stored in the spectrophotometer. 4
Confidence interval:
Lot Certi fi cate Maximum deviation from the desired value over
Chargenzertifikat / Certificado del lote
Spectroquant® COD Cell Test
the entire m
easuring range; every measurement 5
®
Spectroquant CSB-Küvettentest value can be affected by this deviation; this pa-
®
Spectroquant Test en cubetas DQO
Cat.No. / Art.Nr. / Art. Nro. 1.14541.0001 n = 10 rameter is a measure for the accuracy.
Target value Sollwert Result Messergebnis /
Measuring Range / Messbereich / Intervalo de Valor nominal Resultado
25 - 1500 mg/l CSB / COD / DQO
6
medida (Standard / Patrón) (Standard / Patrón)
8
Standard Deviation of the Method
± 4,9 mg/l
Verfahrensstandardabweichung / Desviación estándar del procedimiento
900
600
300
10
0
0 300 600 900 1200 1500
Target value (mg/l)
12
13
14
15
16
Version 1.0 – 01/2024 23
1 5 Analytical Quality Assurance (AQA) – 5.2 Quality Control for the User
8
Test for
recovery
mg/l CSB 14 mm
2 hours. test-tube rack to cool.
2 ml Probelösung
in ein Reaktions-
küvette geben
Caution, the cell
Mischen
Küvette wird heiß,
am Verschluss
anfassen
im Thermoreaktor
erhitzen
148 C, 120 min
mind. 10 min
abkühlen
Mischen
becomes very hot!
Abkühlen auf
Raumtemperatur
(mind. 30 min)
Messen
10 1 2 3
4 5 6
7 8 9
. 0 C
Photometers
12
13 Thermoreactor
Suspend the bottom Carefully pipette 3.0 ml Heat the reaction cell in Remove the reaction Swirl the
Swirl the cellcell
after after
sediment in the cell by of the sample into a the thermoreactor at cell from the thermo- 10 minutes.
swirling. reaction cell, close tight- 148 °C for 2 hours. reactor and place in a
10 minutes.
ly with the screw cap, test-tube rack to cool.
and mix vigorously.
Caution, the cell
becomes very hot!
14
=
Test for the
15 overall system
16
24 Version 1.0 – 01/2024
5 Analytical Quality Assurance (AQA) – 5.2 Quality Control for the User 1
3
5.2.1 Checking the Spectrophotometer
As soon as the spectrophotometer is activated
it runs a Self-Check. This means the hardware
and the software of the spectrophotometer are 4
checked and compared with internal standards.
14
15
16
Version 1.0 – 01/2024 25
1 5 Analytical Quality Assurance (AQA) – 5.2 Quality Control for the User
12
13
14
15
16
26 Version 1.0 – 01/2024
5 Analytical Quality Assurance (AQA) – 5.2 Quality Control for the User 1
12
13
14
15
16
Version 1.0 – 01/2024 27
1 5 Analytical Q
uality Assurance (AQA) – 5.2 Quality Control for the User – 5.3 Determination of
Sample Influences (Matrix Effects) – 5.4 Definition of Errors
15
16
28 Version 1.0 – 01/2024
5 Analytical Quality Assurance (AQA) – 5.4 Definition of Errors 1
3
The following diagram illustrates the aspects of
accuracy and precision:
1 3
6
2 4 10
13
14
15
16
Version 1.0 – 01/2024 29
1
6 Overview
10
11
12
13
14
3 Shaft for rectangular 4 Shaft for round cells
cells
15
16
30 Version 1.0 – 01/2024
6 Overview – 6.2 Overview of the Instrument – 1
6.3 Display and User Interface
11
12
13
14
15
16
Version 1.0 – 01/2024 31
1 6 Overview – 6.3 Display and User Interface
5 System
Methods Setup
Settings
6 Ad hoc Login/Log-
out
AQA
7 Timer
NOTE
10 The menu selected is always outlined in
blue.
NOTE
11
Action buttons like "Start", "Save", "Print" give
the following touch feedback:
12 Normal
Remains static
16
32 Version 1.0 – 01/2024
6 Overview – 6.3 Display and User Interface 1
NOTE
The main menus "Settings (Method Settings)", 3
"Ad hoc", "AQA", "System (Instrument Settings)",
"Login/Logout", "Timer" open up a submenu.
Example "Settings":
4
To leave these, the submenu has to be closed
by t ouching the main menu button again, in this
case: 5
The main menu "Methods" comprises two main overview panels arranged as shown below: 8
the Concentration Measurement Overview and the Method List Overview.
1 Menu 3 Time-
title stamp 10
4 Measure-
ment range
5 Results 11
12 Main
menu but-
tons 6 Citation
12
7 Action
buttons
13
11 Main
menu
selection
button: 10 Brief
infor- 9 otification symbols/
N 8 Info bar with
14
Switch be- mation settings Information
tween the button
two main
menu 15
overviews
16
Version 1.0 – 01/2024 33
1 6 Overview – 6.3 Display and User Interface
3 2 ropdown
D 3 ropdown
D
box box
(closed) (closed)
1 Menu
4 title
7 ocuses
F 4 Selection
selected
buttons
main menu
5 button
6 Name of
method 5 Scrollbar
7
10
Screen layout results overview
2 Filter menu
11 1 Menu
title
7 ocuses
F 3 Selection
selected buttons
12
main menu
button
6 Result 4 Scrollbar
13 list
14
5 Action
buttons
15
16
34 Version 1.0 – 01/2024
6 Overview – 6.3 Display and User Interface 1
6
Method list Settings
List of all methods, irrespective of This button is used to activate meth-
mode od-specific settings (e. g. sample dilu-
tion, turbidity correction, zero adjust-
7
ment, sample blank, reagent blank)
10
11
Ad hoc AQA
or performing measurements (ab-
F Overview and list of all Analytical
sorbance/transmission, spectrum, Quality Assurance (AQA) modes
kinetics) 12
Allows measurements to be per-
formed without the need to create
methods
13
14
15
16
Version 1.0 – 01/2024 35
1 6 Overview – 6.3 Display and User Interface
6
Results list System setup
List of all stored results This button is for optional instrument
settings (e. g. date, time, updates
7 etc.)
10
12
13
14
15
16
36 Version 1.0 – 01/2024
6 Overview – 6.3 Display and User Interface 1
3
Buttons Description
Method list
List of all methods, irrespective of mode
4
Settings
This button is used to activate method-specific settings
(e. g. sample dilution, turbidity correction, zero adjustment, sample blank, reagent blank)
Ad hoc
5
For performing measurements (absorbance/transmission, spectrum, kinetics)
Allows measurements to be performed without the need to create methods
Absorbance/Transmission Mode
Ad hoc submenu: perform absorbance or transmission measurements
6
Spectrum Mode
Ad hoc submenu: record spectrum
Method list: create methods -> Spectrum Mode
7
Kinetic Mode
Ad hoc submenu: perform kinetic measurement
Method list: create methods -> Kinetic Mode
8
AQA
Overview and list of all Analytical Quality Assurance (AQA) modes
AQA1
AQA submenu: List of AQA1 methods 10
AQA2
AQA submenu: List of AQA2 methods
11
Pipette check
AQA submenu: List of pipette-checking methods
Result list
12
List of all stored results
System setup
This button is for optional instrument settings (e. g. date, time, updates etc.)
13
Login/logout
Log users in and out
14
Timer list
List of stopwatch functions
15
16
Version 1.0 – 01/2024 37
1 6 Overview – 6.3 Display and User Interface
3
Action & Selection Buttons Description
Start button
Start an action (e. g. measurement)
4
Start zero
Start zero adjustment for a method
5 Apply
Save
6
Stop
7
Close
8 Logout
User logout
Search method
9
Search/results list
Search function, search criterion: method name, method number or item number
10
Filter cancellation button
Cancel all set filter options
11 Edit
For editing parameters
Create method
12
Print
Print to pdf (USB device) or printer
13
Export button
All selected results are exported to an external memory device as .csv file
14 Import button
Updates/Methods are imported from an external memory device into the instru-
ment
Delete
15 The selected items are deleted
16
38 Version 1.0 – 01/2024
1
7 Safety
1
2
2
WARNING
Hazardous area (general). The xenon lamp (UV/VIS) emits radiation in the ultraviolet 7
region, which may cause damage to the eyes. Never look directly in the radiation of this
light source without wearing proper eye protection. Protect your skin from the direct ex-
posure to UV light.
WARNING
8
Dangerous electrical voltage.
8
WARNING
9
WARNING
Signifies instructions that must be followed precisely in order to prevent serious dangers
9
to personnel.
10
CAUTION CAUTION
Signifies instructions that must be followed precisely in order to avoid minor injuries to
10
personnel or damage to the instrument or the environment.
CAUTION
11
CAUTION
This is a cautionary notice with a warning symbol drawing your attention to the risk of
11
(limited) harm to personnel.
12
NOTE NOTE
Signifies a notice drawing your attention to special characteristics. 12
REFERENCE
13
Used to indicate references to other documents. 13
14
Please pay attention to the separate safety instructions leaflet (part of delivery scope) and read it 14
carefully.
15
15
16
Version 1.0 – 01/2024 39
1 7 Safety – 7
.1 Intended Use – 7.2 General Safety Instructions –
7.3 Target Group and User Qualification
15
16
40 Version 1.0 – 01/2024
7 Safety – 7.4 Handling Hazardous Substances 1
WARNING
6
Improper handling of certain reagents can ad-
versely affect your health. Always follow the
safety labels on the packaging and the safety in-
structions on the package insert. The protective 7
measures specified there must be carefully fol-
lowed.
10
11
12
13
14
15
16
Version 1.0 – 01/2024 41
1
8 Getting Started
11
12
13
14
15
16
42 Version 1.0 – 01/2024
8 Getting Started – 8.2 Initial Setup 1
1 5
7
1. Press the ON/OFF button 1 . The spectro-
photometer gives an audible signal (beep)
2
and starts booting for approximately 2 min-
utes. You will see the following display: 8
9
1
10
Connecting the power adapter:
1. Connect the miniplug 1 of the power adapter
to the socket 2 of the spectrophotometer. 11
2. Connect the power adapter to a wall socket.
2. T
he display switches to language setup (see
12
section 8.2.3).
13
14
15
16
Version 1.0 – 01/2024 43
1 8 Getting Started – 8.2 Initial Setup
5 1
6
2
10 4
3
11
12 3. Tap on OK 3 to confirm.
4. Tap on the Arrow button 4 to choose the
country-specific basic date setting. The date
format can be set and displayed for EU and
13 US.
14
15
16
44 Version 1.0 – 01/2024
8 Getting Started – 8.2 Initial Setup 1
8.2.5 Self-test
Following language, date and time setup the
3
spectrophotometer performs a self-test.
6 4
8 5
10
9
5
7
1
6
5. Tap on the Time format button 5 . The nu-
meric key panel 6 pops up. Now you can
enter the time.
6. Tap on OK 7 to confirm. 1. emove all cells and close the cell compart-
R 7
7. Tap on the Arrow button 8 to choose the ment cover.
country-specific basic time setting. The time 2. Start the self-test with the Start button 1 .
format can be set and displayed for EU and 3. The spectrophotometer performs the self-
US. test. 8
8. Tap on the Arrow button 9 to choose the
decimal separator "."/"," used in your coun- Self-test
try. The self-test covers:
9. Tap on the Save button 10 to confirm. • Checks on memory, processor, internal inter- 9
faces, filter and lamp
After initial setup is complete, you can change • A calibration of the wavelengths
the date and time at any time in the setup/date/
time menu (see section 9.2.3). When the self-test has ended, the display shows 10
the main menu.
11
12
13
14
15
16
Version 1.0 – 01/2024 45
1 8 Getting Started – 8.3 Connecting Optional Peripheral Devices
7
You can connect the following peripheral devices
to the spectrophotometer:
• Printer (see section 8.3.2)
8 • USB mass storage device (see section 8.3.3)
• Barcode reader (see section 8.3.4)
NOTE
9 If you want to connect several USB devices such
as a USB PC keyboard and a USB memory de-
vice to the instrument, you can increase the
number of USB-A sockets by attaching a com-
10 mercially available USB-2 hub with separate
power supply.
11
8.3.2 Printer
Printers can be connected to the spectrophoto-
12 meter as follows:
NOTE
14 All postscript printers can be used with
Spectroquant® Prove plus.
15
16
46 Version 1.0 – 01/2024
8 Getting Started – 8.3 Connecting Optional Peripheral Devices 1
3
8.3.3 USB Memory Device
Using a USB memory device (e. g. a USB mass
storage device), you can 4
• Update the instrument firmware and method
data (see section 9.2.8)
• Transfer data to the USB memory device
(see section 9.13.7) 5
8
8.3.4 Barcode Reader
The barcode reader enables the simplified ac-
quisition of alphanumerical character strings and
can be used in all operating situations that re- 9
quire the entry of text or numerals. The barcode
reader is connected to the USB-A port.
NOTE 10
The barcode reader must support USB/HID.
11
12
13
14
15
16
Version 1.0 – 01/2024 47
1
9 Operation
2
6
14
15
16
48 Version 1.0 – 01/2024
9 Operation – 9.1 Switching the Spectrophotometer On or Off 1
3
NOTE NOTE
After the self-test, the system automatically You can set a user-defined time for this function
checks the status of activated AQA tests. An (see section 9.2.5). 4
overview of any failed and/or expired AQA tests
automatically opens (see section 9.11).
Switching off
5
6
4
12
13
16
Version 1.0 – 01/2024 49
1 9 Operation – 9.2 System Setup
6
Buttons Description
Information
This submenu displays the following information about the device:
7 Software/method versions, device class, lamp counter and serial number
Interface
This submenu displays the following settings options and standard settings:
8 Audible signals – ON, Backlight – 100%, Print to pdf – ON
Region
9 This submenu displays the following settings options and standard settings:
Language, date, time and country zone EU/US, decimal separator – "."/"," (dot or comma)
Quality
10 This submenu displays the following settings options and standard settings:
Quick zero – OFF, AQA1 and AQA2 lock – OFF, Zero Adjustment expiry – ON (interval:
7 days), Use expired reagents – OFF, Service reminder – ON
Automation (Setup 4)
11 This submenu displays the following settings options and standard settings:
Energy saving mode – ON (10 minutes), Auto Power off – OFF, Auto log off – OFF,
Auto store – ON, Auto print – OFF, Sample ID popup – OFF
User management
12 This submenu displays the following settings options and standard settings:
Activation of user management and administrator settings, User login required – OFF
13 Service
This submenu displays the following settings options:
Various service functions such as backup, restore, export of log or system data and
import of methods
14 Update
This submenu displays the option for performing software and method updates
15
16
50 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
Buttons Description 3
Network
This submenu displays the setting options for connecting the Prove plus device with a network
4
Prove Connect
This submenu displays the settings options for connecting the Prove plus device with the Prove
Connect software (the Prove Connect software is optionally available, order No. Prove Connect
to LIMS Y110860001) 5
9.2.1 Information 6
This submenu displays the following
information and settings options:
7
2 8
8
3
5
9
6
7
8 eset button for lamp counter in
R
Prove 100 plus 10
Press this button after the change of the halogen
lamp to reset the lamp counter to zero.
1 Update Version
Instrument version number
11
2 MMI Version
Man-machine interfaces version number
3 MCS Version
Measurement and control software version 12
number
4 Methods Version
Method version currently in use
5 Device
13
Class of device being used
(Prove 100 plus | 300 plus | 600 plus)
6 Lamp counter
Lifetime/service performance of the lamp 14
7 MCS Serial
Instrument serial number
15
16
Version 1.0 – 01/2024 51
1 9 Operation – 9.2 System Setup
4
9
1
11
5 10 3 2
3 2
4
6
10 1 2
12
11
12 13
1. Select System 1 .
2. Select Region (Setup 2) 2 .
13
1. Select System.
2. Select Interface (Setup 1).
3. Set the backlight contrast 11 to the desired
14 level using the drop-down menu 12 .
4. Tap on the Save button 13 to store and close
the settings.
15
16
52 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
4
5
3
5
3. elect the desired language in the drop-
S 6
down menu 3 .
4. Tap on the Save button 4 to store your
changes and close the settings by touching
the System button again. 7
• Date display US/EU 3
• Time display US/EU 4
Date/time
• Decimal separator "."/"," (dot or comma) 5
Depending on country, the date format is pre-
sented in the sequence e. g. Day.Month.Year
8
(DD.MM.YY) or Month/Day/Year (MM/DD/YY or NOTE
MM.DD.YY). To reset or change the figures, pro- Make sure that the decimal separator used is the
ceed as follows: same which is used in your Excel software to
9
avoid any problems with the format of your csv
files.
10
1 2
7
11
6
8 12
1. Select System 1 .
2. Select Region (Setup 2) 2 . 13
3. apping on the Date field 6 pops up a calen-
T
dar 7 . Here you can set the date and tap on
OK 8 to confirm your selection.
14
15
16
Version 1.0 – 01/2024 53
1 9 Operation – 9.2 System Setup
10
5 11
9 12
10
5. Tap on the Save button 12 to store all of the
changed settings.
11
12
13
14
15
16
54 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
9.2.4 Quality
This submenu displays the following
3
information and settings options:
9 4 3
10 6
7
Possible settings (items 5 to 12 ):
5 QuickZero Performance and direct saving of the zero Performance and saving of the zero adjust- 8
(Possible only for adjustment for: ment only for the currently selected con-
concentration meth- • all 12 wavelengths that are used for mea- centration method.
ods) suring Spectroquant® test kits
• the wavelength currently being used in 9
the Concentration mode
6 AQA1 – Lock device If an invalid AQA1 test is present, a warning If an invalid AQA1 test is present, a warning
message appears. The instrument is locked message appears. Measurements can still
for all measurements except AQA1 tests. be performed. The instrument is not locked. 10
7 AQA2 – Lock If an invalid AQA2 test is present for a If an invalid AQA2 test is present for a
method concentration method, a warning message concentration method, a warning message
appears when this method is selected. The appears when this method is selected.
performance of a concentration measure- Measurements can still be performed. The
11
ment using this method is locked. instrument is not locked.
8 Zero adjustment When the pre-set expiry date in days 13 An already performed zero adjustment
EXP is reached, the zero adjustment must be does not have to be repeated.
repeated.
(Possible only for
12
concentration meth-
ods)
9 Use expired re- The use of Spectroquant® test kits beyond The use of Spectroquant® test kits beyond
agents the date of expiry is allowed. A warn- the date of expiry is not allowed. A warn-
13
ing message appears when the barcoded ing message appears when the barcoded
Spectroquant® cell or the AutoSelector Spectroquant® cell or the AutoSelector is
is inserted into the instrument; after the inserted into the instrument. No measure-
warning message is confirmed, however, ment can be made.
the measurement is still carried out. 14
10 Service reminder The instrument automatically signals the The instrument does not signal the need to
need to service the instrument. service the instrument.
15
16
Version 1.0 – 01/2024 55
1 9 Operation – 9.2 System Setup
11 Deactivate message When results are higher than the measuring Results beyond the measuring range of
indicating results range of the selected method, the display the selected method are also shown as a
beyond measuring of the device shows “HI”. numerical value. Negative measurement
4 range When results are lower than the measuring results may also be shown here.
range of the selected method, the display (This function can be of help in calculating
of the device shows “LO”. statistical parameters, e. g. in determining
limits of detection.)
5 12 Allow deletion of When the user management is activated, No results in the result list can be deleted.
results in the result results in the result list can be deleted.
list (This function can be activated only when
the user management is activated by a user
with administrator status.)
6
10
11
12
13
14
15
16
56 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
9.2.5 Automation
This submenu displays the following 3
information and settings options:
7 Auto logoff
The instrument logs the active user out after the
4
1 1
5 10
user-defined time has elapsed and no entry has
2
6 11 been made. The screen changes to the login/
logout screen.
7 12 5
When the function is activated, a field pops
8
up for setting the time. Here you can enter
9 4 3 your individual time in minutes and confirm by
tapping the OK button. Accept the time setting 6
by tapping on the Save button.
8 Prefix S-ID
This function enables you to prefix a sample ID
with a recurrent character string (e. g. Well), 7
which is then saved together with the sample ID.
1 Energy-saving mode
When the function is activated, a field pops up
To activate a function, tap on the input field 1 .
to enter data. Here you can enter the individual
A tick 2 appears when the setting option is acti-
vated. To cancel settings, tap on the C button
string of characters or description, confirming 8
your entry by tapping the OK button. Accept the
3 . This works only for changes that have been
entry by tapping on the Save button.
made but not yet stored. To accept the settings,
9 SampleID popup
tap on the Save button 4 .
5 Energy-saving mode After each measurement, an input window 9
automatically pops up to enter the sampleID.
After a user-defined time has elapsed, the back-
The data can be entered via the virtual keyboard
light of the display switches off automatically.
of the display, a keyboard connected via the USB
The backlight is switched on again with the next
port, or a hand-held scanner connected via the 10
tap.
USB port.
When the function is activated, a field pops
up for setting the time. Here you can enter
your individual time in minutes and confirm by
11
tapping the OK button. Accept the time setting
by tapping on the Save button.
6 Auto power off
After a user-defined time has elapsed, the 12
instrument switches off automatically.
When the function is activated, a field pops
up for setting the time. Here you can enter
your individual time in minutes and confirm by 13
tapping the OK button. Accept the time setting
by tapping on the Save button.
14
15
16
Version 1.0 – 01/2024 57
1 9 Operation – 9.2 System Setup
3 10 Auto store 13
11 Auto print
10
The instrument automatically starts a print
operation after the measurement has been
completed. Precondition: a USB memory device
(print as pdf) or PostScript printer (print on
11
paper) is connected.
12 Barcode scanner
This function enables the barcode scanner
12 for decoding round cells and AutoSelectors to
be deactivated. When the barcode scanner is
deactivated, the automatic method recognition
of barcoded round cells and AutoSelectors is
13 switched off and the method must be selected
manually.
This function comes in handy when e. g. you
wish to run your own methods in round cells.
14
15
16
58 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
6
1
7
3
Activate user management:
4
When the "Login required" box 1 is ticked, us-
2 5
ers have to log in in the "Login/Logout" menu
8
with their name and password to get certain ac-
cess user rights (see section 9.14.1).
NOTE
Edit user, e. g. change password:
9
Once "Login required" is ticked, it can only be 1. Select the user 1 whose password is to be
deactivated by an administrator. changed.
2. Tap on edit 2 to edit the user.
10
3. Enter the new user password 3 .
7 2 4. Confirm the password 4 .
5. Tap on OK 5 to confirm.
3
11
4
5
1 6
12
Create user: 13
1. Tap on the Add new user field 1 .
2. Assign the user rights by tapping on the but-
ton 2 : full rights = administrator; restricted
rights = user (see section 9.14). 14
3. Enter the user name 3 .
15
16
Version 1.0 – 01/2024 59
1 9 Operation – 9.2 System Setup
9.2.7 Service
3
This submenu offers the following
setting options:
4
4 6
1 1
7 7 4
2
5 3
6
5
3
2
6
14
15
16
60 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
11
12
15
16
Version 1.0 – 01/2024 61
1 9 Operation – 9.2 System Setup
9.2.8 Updates
3 Firmware and method updates
ensure your s
pectrophotometer is
always up-to-date.
NOTE
To keep the instrument always up-to-date, we
9
recommend always installing the latest update.
The relevant updates are available on
www.sigmaaldrich.com/photometer-service.
10
Firmware and method update spectro-
photometer 6. T
ap on the Import field 3 starts the search
11 for update files on USB device. Search needs
some time (approx. 1 minute).
7. You will be asked whether you want to install
the update version on the Prove plus. Con-
12 firm with OK.
1
8. The installation process is reported in the in-
formation window and confirmed by the final
2 message that the data have been success-
13 fully transferred.
9. Confirm with OK.
14
15
16
62 Version 1.0 – 01/2024
9 Operation – 9.2 System Setup 1
7
10. T
he instrument then shuts down and re-
boots. The boot screen appears in the dis-
play. Depending on the data volume, this
procedure may take several minutes. 8
10
4
11
14
15
16
Version 1.0 – 01/2024 63
1 9 Operation – 9.3 Measurements
9.3 Measurements
The spectrophotometer can be used to perform the measurements listed below.
3
Type of Measurement Description
Concentration re-programmed methods that can be executed using Spectroquant® test kits or self-
• P
4 prepared reagents
• User-programmed methods
Spectrum • P
rogrammed methods for establishing the absorbance or transmission of solutions over a
defined wavelength range
6
Kinetics • P
rogrammed methods for establishing the absorbance or transmission of solutions over a
defined period
10
11
12
13
14
15
16
64 Version 1.0 – 01/2024
9 Operation – 9.3 Measurements 1
6
6
2
4
7
10
5
7
1 11
3
15
16
Version 1.0 – 01/2024 65
1 9 Operation – 9.3 Measurements
3
Measuring with rectangular cells with open Measuring with rectangular cells with open
lid: lid:
Insert rectangular cells (10, 20, 50 mm) Insert 100 mm rectangular cells
4 (Prove 600 plus)
10
9
11
10
15
16
66 Version 1.0 – 01/2024
9 Operation – 9.3 Measurements – 9.4 Zero Adjustment 1
10
2
1
16
Version 1.0 – 01/2024 67
1 9 Operation – 9.4 Zero Adjustment
3
Zero adjustment with rectangular cells 9.4.2 When to Repeat the Zero
•W ith rectangular cells, zero adjustment must Adjustment?
be carried out using the same cell type (manu- We recommend that you repeat the zero
facturer and cell material [e. g. optical glass, adjustment in the following cases:
4 quartz glass, plastic]) as the one that will be • If the spectrophotometer was subject to me-
used for measurement. This is important be- chanical stress such as strong shock or trans-
cause cells of different manufacturers have dif- port
ferent absorption characteristics. When chang- • If the ambient temperature has changed by
5 ing the cell type, repeat the zero adjustment more than 5 °C since the last zero adjustment
with the new type • At least once a week. The interval to repeat a
• Prior to zero adjustment, clean the rectangular zero adjustment is set in the instrument to 7
cell and fill it with distilled water. The minimum days. You can change this under "System (In-
6 filling level is 20 mm strument Settings)"
• Rectangular cells always have to be inserted in • If a new cell type (different manufacturer, dif-
the cell compartment with the same orienta- ferent glass type) is used
tion for measurement and zero adjustment • Basically, each time you want to measure with
7 (e. g. cell inscription always on the left side) the highest possible accuracy
• Insert the rectangular cell until it touches the
bottom and left edge of the holder. The opaque
NOTE
sides of the rectangular cell must point to the
8 front and rear. The spectrophotometer detects If an interval to repeat a zero adjustment is set
stray light. If there is too much stray light, a you will be prompted to repeat it after the inter-
message prompts you to close the cell com- val has passed. You can also repeat a zero ad-
partment cover justment by selecting a method, then touching
9 the "Settings" icon. Choose "Zero adjustment"
and insert a zero cell to start the measurement.
NOTE
Ordering information for cells is provided in sec-
10 tion 13. The cells listed in section 13 are special-
ly intended for use with the Spectroquant® test
kit system. Note that the spectral transparency
of the cell must be suitable for the intended ap-
11 plication (example, quartz cell for UV range).
12
13
14
15
16
68 Version 1.0 – 01/2024
9 Operation – 9.4 Zero Adjustment 1
AutoSelector
• By manually selecting the concentration meth-
od via the method list (see section 9.5) 5
6
1
2
4. I nsert the zero cell according to cell type.
Zero adjustment starts automatically and, 7
if the zero adjustment is passed, a tick 4
appears in the status display field for the
Zero adjustment 3 .
In the case of a method which only mea- 8
sures the sample at a single wavelength
the absorbance of the Zero value 5 is also
1. nce the concentration method is selected
O displayed.
touch Settings 1 . 5. With a cell inserted, the zero adjustment 9
2. Tap on the Zero adjustment button 2 . can be repeated manually by tapping on the
Start zero button 6 .
6. Tapping on the OK button accepts the zero
adjustment value for the method.
10
3 7. The screen changes to show the concentra-
tion measurement screen.
8. The instrument is ready to start measuring
11
the sample.
16
Version 1.0 – 01/2024 69
1 9 Operation – 9.4 Zero Adjustment – 9.5 Method List
NOTE
The method designation contains the following
11 information
4 Method number
5 Method name
6 Measuring range + unit (if the method is
12 suited for several cell formats, the measuring
range over all cell formats is displayed)
7 Citation (convertible)
8 Cat. No. (6 digits) or reference to application
13 (“App” for test-kit-free methods)
14
15
16
70 Version 1.0 – 01/2024
9 Operation – 9.5 Method List 1
3
9.5.2 earching and Filtering the
S
Method List
You can search and filter the method list in order
to make it easier to find the method you are 10
4
looking for:
5
9
7
2. Set filter by criteria 10 :
• All methods
• L ast used methods: last six used in alphabetic
order 8
1. Filter by method type 9 : • Frequently used methods: six most frequently
• All used in alphabetic order
• Concentration • Factory pre-programmed methods only
• Kinetic • User-specific methods only 9
• Spectrum • Area of application (e. g. brewery, color, oils,
sugar)
10
11
12
13
14
15
16
Version 1.0 – 01/2024 71
1 9 Operation – 9.5 Method List – 9.6 Programming a User-defined Method
11 14 1 2
4
12
13
14
15
16
72 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
2
11
Absorbance
12
13
Nominal value (e. g. concentration)
NOTE 14
In measuring operation, the reverse calibration
function is used to output a measured absor-
bance as a concentration value. 15
16
Version 1.0 – 01/2024 73
1 9 Operation – 9.6 Programming a User-defined Method
5
1
5
3
4
6
2
6
9
7
7 8
14 9
Calibration func- (See examples on the fol-
tion lowing pages)
* Optional
15
16
74 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
4
11
10
5
10
11
12
13
14
15
16
Version 1.0 – 01/2024 75
1 9 Operation – 9.6 Programming a User-defined Method
4 7
6
8 1
2
3
5 9
5
10
11
6
6
7
4 15 14 13 12
Enter the value pairs, nominal value (concentra- 3. " Fit E0" 5 can be activated as an additional
8 tion)/measured absorbance einer of an already option. With "Fit E0" activated, concentra-
available test series with the following value tion 0 (= reagent blank value) intercepts the
pairs: absorbance axis at the associated E0 value.
• E0 2 = reagent blank (see section 9.7.8) 4. Once all values are available, tapping on the
9 • At least one, max. eleven value pairs in various field “Function” 9 calls up an overview of
concentrations the calculated coefficients. Tapping on the
field “Graph” 10 shows the calibration curve.
1. Enter E0 2 , concentration of the standard
10 solution 3 and associated absorbance 4 us-
ing the keypad 1 . Tapping on the + button NOTE
5 allows further (up to eleven) value pairs The calculated function maps the calculation of a
to be entered. The UP&DOWN buttons 6 are result (e. g. concentration) via a measured ab-
11 activated if more than four value pairs are sorbance in the form of a polynomial equation of
entered. the following type:
2. Activate the field “Linear” 7 to calculate a
linear function. When “Linear” is not activat- C = A0 + A1 x (Abs – E0) + A2 x (Abs – E0)2
12
ed, a non-linear function of the second order
is automatically calculated (square function). where:
C = measurement result (e. g. concentration)
NOTE A0, A1, A2 = coefficients (polynomial)
13
If you wish to calculate a linear function, you Abs = measured absorbance
require at least the E0 value and 2 value pairs. If E0 = absorbance of the reagent blank value
you wish to calculate a non-linear function, you
14 require at least the E0 value and 3 value pairs. 5. ou can now enter an ID or a specific batch
Y
code for the calibration. Tapping on the field
“Batch ID” 11 opens a virtual keyboard.
Enter the code and confirm by tapping on
15 the OK button.
16
76 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
6. T o close the determination of the You now have the following options: 3
coefficients, confirm the data by tapping on
the OK button 12 . • C lose the programming/processing of the
7. Tap on the Export button 14 to transfer method by tapping on the Save button 17 .
the data in the CSV format to an external All entered data are accepted. A method 4
storage medium. number appears 18 .
8. Tap on the Print button 15 to print out the Close the screen by tapping on the X
data. button 19 . The screen changes to the
9. To close the operation without accepting the method list.
5
data, tap on the X button 13 . All entered • To call up coefficients, value pairs, or the
data are deleted. graph again, tap on the field “Calibration”
16 .
6
• Cancel the programming/processing of the
method by tapping on the X button 19 .
All entered data are deleted. The screen
changes to the method list. 7
16
8
17 19
18
10
11
12
14
15
16
Version 1.0 – 01/2024 77
1 9 Operation – 9.6 Programming a User-defined Method
4 7
6
8 1
2
3
5 9
5
10
11
6
6
7
4 15 14 13 12
NOTE
9 16
appear.
16
78 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
16
Version 1.0 – 01/2024 79
1 9 Operation – 9.6 Programming a User-defined Method
Enter a function:
3 You now have the following options: This option can be used when an evaluation
function is already available or has already been
• Close the programming/processing of the calculated previously by a calculation pro-
method by tapping on the Save button 24 . All gramme on the basis of available data.
4 entered data are accepted. A method number Enter a function to calculate the concentration
appears 25 . from the absorbance (reverse calibration func-
Close the screen by tapping on the X button tion). You can enter on the spectrophotometer
26 . The screen changes to the method list. the coefficients of a polynomial equation
5 • To call up coefficients, value pairs, or the of the following type:
graph again, tap on the field “Calibration” 23 .
• Cancel the programming/processing of the C = A0 + A1 x (Abs – E0) + A2 x (Abs – E0)2 +
method by tapping on the X button 26 . A3 x (Abs – E0)3 + A4 x (Abs – E0)4 + A5 x
6
All entered data are deleted. The screen (Abs – E0)5
changes to the method list.
where:
7 C = measurement result (e. g. concentration)
A0, A1, A2, A3, A4, A5 = coefficients
(polynomial)
Abs = measured absorbance
8 E0 = absorbance of the reagent blank value
10
11
12
13
14
15
16
80 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
Enter coefficients:
1 4
3
5
2
5
4
6
7
9 8 7 6
calibration curve.
4. To close the determination of the coefficients,
confirm the data by tapping on the OK button 11
6 . 12
5. Tap on the Export button 8 to transfer
the data in the CSV format to an external
storage medium. 12
6. Tap on the Print button 9 to print out the
data.
7. To close the operation without accepting the
data, tap on the X button 7 . All entered 13
data are deleted.
14
8. Tapping on the OK button 6 confirms the
entered data, and the screen changes to the
method screen. A check appears in the field 15
“Calibration” 10 .
16
Version 1.0 – 01/2024 81
1 9 Operation – 9.6 Programming a User-defined Method
3 You now have the following options: Example 2 (non-linear calibration function)
The coefficients of the reverse calibration func-
• Close the programming/processing of the tion are determined by multiple regression.
method by tapping on the Save button 11 . All When doing so, the concentration has to be on
4 entered data are accepted. A method number the Y axis and the absorbance on the X axis. The
appears 12 . absorbance of the individual value pairs must
Close the screen by tapping on the X button always be corrected by the reagent blank.
13 . The screen changes to the method list.
5 • To call up coefficients, value pairs, or the X value Y value
graph again, tap on the field “Calibration” 11 .
Absorbance Absorbance - RB * Concentration
• Cancel the programming/processing of the
method by tapping on the X button 13 . 0.010 0.000 0.0 mg/l
6
All entered data are deleted. The screen 0.020 0.010 0.1 mg/l
changes to the method list.
0.070 0.060 0.2 mg/l
8 tion has to be on the Y axis and the absorbance 0.655 0.645 4.0 mg/l
on the X axis. The absorbance of the individual
0.825 0.815 5.0 mg/l
value pairs must always be corrected by the
reagent blank. * = reagent blank
9
X value Y value Calculated calibration function (third order
polynomial):
Absorbance Absorbance - RB * Concentration
C = -0.044983 + 7.4807 × A -4.5229 × A2 +
10 0.050 0.000 0.0 mg/l * 3.8305 × A3
0.250 0.200 1.0 mg/l
or
0.451 0.401 2.0 mg/l
14
15
16
82 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
5
1
2
6 5 4 7 3 9 8 10
9
1. The calibration function 1 appears in the 2. T he X axis shows the absorbance values,
graph. When the function has been calculat- the Y axis the corresponding results (e. g.
ed using value pairs, the coefficient of deter- concentration). Tapping on the X/Y button 10
3 changes the presentation of the axes
mination “R2” 2 is also shown.
around.
NOTE The formula of the calibration function 1
The function determined in this way is used for
that is shown remains unchanged. 11
3. When the function has been calculated
calculating a result (e. g. concentration) via a
via value pairs, these pairs are shown in a
measured absorbance in the form of a polynomi-
separate field 4 . The forward and back keys
al equation of the following type:
5 , 6 can be used to call up the next value 12
pair. Tap on the C button 7 to reset the
C = Polynomial (Abs)
view.
4. Tap on the Export button 8 to transfer
where:
the data in the CSV format to an external 13
C = measurement result (e. g. concentration)
storage medium.
Abs = [measured absorbance of the sample or
5. Tap on the Print button 9 to print out the
standard] minus [absorbance of the reagent
data.
blank value (E0)] 14
6. Tap on the X button 10 to close the view
of the graphic presentation. The screen
changes.
15
16
Version 1.0 – 01/2024 83
1 9 Operation – 9.6 Programming a User-defined Method
10
3
11
14
15
16
84 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
8 Unit * e. g. mg/l
10
9 Resolution 0.001, 0.01, 0.1, 0.25, 0.5 or 1
Lower and upper limit of the mea- Any value between zero and the highest concentration of the standard
10
suring range solutions used
11
11 Procedure variable * Procedure variables provide a value that has to be entered prior to each
measurement on the spectrophotometer (e. g. volume, pH value or tem-
perature)
12 Formula function Formula editor for convenient programming of any user-defined methods 12
13 Condition * With a condition you can check a value for validity (e. g. absorbance value,
procedure variable or the result of a formula).
* Optional 13
14
15
16
Version 1.0 – 01/2024 85
1 9 Operation – 9.6 Programming a User-defined Method
4 19
15
5 16 14 17 18 21 20 22 23
6
6. I nput field 5 – wavelength: Up to five Additional entry fields 19 are added by tapping
wavelengths can be set. Add further entry on the + button 20 .
fields for wavelengths 15 by tapping on the Tapping the Delete button 21 removes the most
7 + button 14 . Tapping the Delete button 16 recently programmed input row 19 . Tapping on
removes the most recently programmed the Save button 22 accepts the input. Tapping on
entry field 15 . Tapping on the Save b
utton 17 the X button 23 closes the screen and opens the
accepts the input. Tapping on the X button previous input mask.
8 18 closes the screen and opens the previous
input mask.
10 5
11 11 24
11
16
86 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
30
29
4
25 26 31 32 5
6
11. T
apping on the OK button 25 accepts the 13. O
ptional tapping on the Condition button 28
input. Tapping on the X button 26 closes the changes the screen. Here you can define a
screen and opens the previous input mask. condition for valid measurements (e. g. absor-
bance wavelength 1 ≤ 2.50). Tapping on the 7
Variable and/or wavelength button 29 accepts
the selection and displays it in the pre-select-
ed output field 30 .
8
NOTE
When methods are used that are subject to a
27
condition, the measurement result is calculated
only when the condition is met. 9
28
14. T
apping on the OK button 31 accepts the
input. Tapping on the X button 32 closes the 10
screen and opens the previous input mask.
12. A
n excerpt of the formula you have defined
appears in the display field 27 .
11
12
13
14
15
16
Version 1.0 – 01/2024 87
1 9 Operation – 9.6 Programming a User-defined Method
8
36
1. Define the method type (see section 9.6).
2. Enter a name for the method 1 . This name
9 is displayed in the method list.
10
11
17. The screen changes to show the method list
36 .
18. The method is now created and stored in the
12
instrument.
NOTE
13 To find the method more quickly, use the filter
function (see section 9.5.2).
14
15
16
88 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
3
8
6
4
2
4
5 5
9 7 9 7
Item Input Possible input 7. Tap on the Save button 7 . The method is
field automatically given a number generated by
the system 8 . All values are stored. 7
1 Name Any name
8. Tap on the X button 9 to leave the edit
2 Start & Wavelength range Prove 300 plus | method screen.
3 Stop 600 plus: 190 – 1100 nm
Wavelength range Prove 100 plus: 8
320 – 1100 nm
11 10
Peak
4 Threshold value for peak detection
detection
NOTE
A spectrum can comprise up to a maximum of 14
1000 measurement points. If the entry is invalid,
it is shown in red and cannot be accepted.
15
16
Version 1.0 – 01/2024 89
1 9 Operation – 9.6 Programming a User-defined Method
4
1
2 5
5
6
3
7
6
8
7 9
8 4 11
For Kinetic Mode you can develop and store 4. Create the user-defined parameters
your own user-defined methods under method • Wavelength 3
9 numbers 1201 to 1220. The s pectrophotometer • Unit 4
software supports you in creating the methods. • Interval 5
To create a user-defined kinetic method, proceed Prove 100 plus Minimum: 00:00:10 (hh:mm:ss)
as follows: Prove 300/600 plus Minimum: 00:00:05
10 (hh:mm:ss)
1. Define the method type (see section 9.6). • Delay 6
2. Enter a name for the method 1 . This name • Duration 7
is displayed in the method list. • Slope factor 8
11 3. Select type of measurement: absorbance or 5. Tap on the Save button 9 . The method is
transmission measurement 2 by tapping on automatically given a number generated by
the button for the required measurement the system 10 . All values are stored.
(the activated selection is light grey). 6. Tap on the X button 11 to leave the edit
12 method screen.
NOTE
Invalid entries are shown in red and cannot be
13
accepted.
14
15
16
90 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
3
Item Input field Possible input
4
2 Selection button Choice between absorbance or transmission
3 Wavelength Wavelength range Prove 300 plus | 600 plus: 190 – 1100 nm
Wavelength range Prove 100 plus: 320 – 1100 nm
5
4 Unit * User-defined input when an end result is to be calculated (e. g. enzyme activity U/ml)
8 Slope factor * User-defined input for calculating a result = "slope factor" × "∆ A/min" (instrument automati- 7
cally calculates the difference in absorbance/minute = ∆ A/min)
8
13 12
10
11
7. The screen changes to show the method list
12 .
8. The kinetic method is now created and 12
stored in the instrument.
NOTE
13
To find the method more quickly, use the filter
function 13 (see section 9.5.2).
14
15
16
Version 1.0 – 01/2024 91
1 9 Operation – 9.6 Programming a User-defined Method
9.6.8 opying/Duplicating a
C
User-defined Method
3
1. earch and select the method (see section
S
9.5.2).
5 4
7
5. he method is created under the method
T
name with the addition "name-copy" 4 and
appears in the method list.
8 1 6. Modify the duplicated method as needed
3
2
(see section 9.6.3).
10
2. A ctivate the method by tapping on the arrow
in the right column of the method list 1 .
11 3. A selection of various method editing possi-
bilities shows up 2 .
4. To copy the method, tap on the Copy button
3 .
12
13
14
15
16
92 Version 1.0 – 01/2024
9 Operation – 9.6 Programming a User-defined Method 1
1
3
2
1
5
3
2
7
2. A ctivate the method by tapping on the arrow 2. A ctivate the method by tapping on the arrow
in the right column of the method list 1 . in the right column of the method list 1 .
3. A selection of various method modification 3. A selection of various method modification
possibilities shows up 2 . possibilities shows up 2 . 8
4. To modify the method, tap on the Edit but- 4. To delete the method, tap on the Delete but-
ton 3 . ton 3 .
5. Now follow the description in the sections on
programming a: 9
• User-defined concentration method (see sec-
tion 9.6.1)
• User-defined spectrum method (see section
9.6.6) 10
• User-defined kinetic method (see section 9.6.7)
5 4
11
12
5. he confirmation box asking if you really
T
want to delete the selected method pops up.
6. To delete the method, tap on the OK button
4 to confirm or on the X button 5 to cancel 13
the deletion process.
CAUTION
14
Following your confirmation the method is per-
manently deleted. Before deleting any method,
we recommend you export a backup copy of it to
an external memory device. 15
16
Version 1.0 – 01/2024 93
1 9 Operation – 9.6 Programming a User-defined Method
3
9.6.11 E
xport User-defined Methods to
a USB Memory Device
4
1
5 2
10
3
11 4
14
15
16
94 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
4
2
3 5
1 4
5 7
3. Further options: 11
NOTE
• Select a different citation form by tapping on
If the barcode cannot be read, a message the citation form display field 4 (e. g. NH4 <–>
appears. In this case you can make a new NH4-N)
attempt by using the Spectroquant® round cell • Select a different measuring unit by tapping on 12
bearing the barcode or the AutoSelector as the unit display field 3 (e. g. mg/l <–> mmol/l)
described. Care must be taken to ensure that • Make further settings such as dilution or blank
the positioning mark on the round cell/ value measurements with "Settings" (see sec-
AutoSelector is aligned with the position mark on tion 9.7.5)
13
the spectrophotometer.
Alternatively, you can close the message and
then select the required method manually from
the method list. 14
15
16
Version 1.0 – 01/2024 95
1 9 Operation – 9.7 Measuring in Concentration Mode
3
4. Content of the information bar in the
concentration mode:
7
5
8
Tapping on the Information bar button 5
opens the extended information bar. In the
concentration mode the following information is
9 displayed:
13
14
15
16
96 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
13
14
15
16
Version 1.0 – 01/2024 97
1 9 Operation – 9.7 Measuring in Concentration Mode
NOTE
10 Measurement results beyond the measuring
range are marked separately in the display (see
section 9.7.4).
11
12
13
14
15
16
98 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
4
NOTE 13
For methods involving a special measurement
procedure, e. g. the chlorophyll analysis, results
beyond the measuring range are shown by “ --- ”.
14
15
16
Version 1.0 – 01/2024 99
1 9 Operation – 9.7 Measuring in Concentration Mode
1 9
5
2 10
3 11
6
4
5
7
6
8 7
13 NOTE
The individual settings are greyed out if they are
not available for the selected method. Switching
a gobal setting (turbidity correction, show absor-
14 bance) ON activates it for ALL methods where it
is applicable.
15
16
100 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
NOTE
Optimum measurement results are achieved if
the concentration of the diluted sample is in the 6
middle of the measuring range of the method af- 2. Select Dilution 2 and confirm. The input
ter diluting. screen 3 for Dilution appears.
3. Tap on the dilution value in the display fields
4 , enter the
factor on the keypad and tap 7
Having selected the method, enter the dilution
on OK button 5 .
as follows:
8
2
7
1
9
10
11
1. Open the Settings menu 1 . 4. he spectrophotometer is ready to start
T
measuring.
The display 6 changes to measuring mode.
12
The dilution that was just entered is used for the
next measurement.
13
14
15
16
Version 1.0 – 01/2024 101
1 9 Operation – 9.7 Measuring in Concentration Mode
10 1
11
12
13
14
15
16
102 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
NOTE
3
The use of the sample blank value is indicated by
the symbol 8 in the display being lit. The sam-
4
ple blank value is also shown with prefix "SB" in
the information bar at the bottom (see section 4
5 9.7.1).
6 5
6
2. Select Sample Blank Value 2 .
3. Insert the cell with a suitable sample blank.
The sample blank measurement starts auto-
matically. The value is used only for the next 7
measurement.
4. The first single measurement for the sample
blank value is performed. The following data
is displayed as the result: 8
• The measured absorbance from the (last)
single measurement 4
• The median from all single measurements
carried out so far 5 9
5. If necessary, carry out further single mea-
surements to calculate the median.
10
8
11
12
13
6. Tap on OK 6 to accept the measurement.
7. The screen changes to the measuring mode
7 . 14
8. The spectrophotometer is ready to start
measuring.
15
16
Version 1.0 – 01/2024 103
1 9 Operation – 9.7 Measuring in Concentration Mode
16
104 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
NOTE
3
In the event that a Lot ID different from the one
9
used to measure the reagent blank value is used
in a subsequent measurement, this is recognized
by the barcode of the inserted round cell or of 4
the AutoSelector. The activated user reagent
8 blank value is then automatically deactivated
and a corresponding message appears in the
display. Tapping “OK” clears the message and 5
the measurement is performed automatically
without taking a user reagent blank value into
account.
6
9. The screen changes to the measuring mode
8 .
10. The spectrophotometer is ready to start
measuring. 7
NOTE
The use of the reagent blank value is indicated
by the symbol 9 in the display being lit. The re- 8
agent blank value and its date of measurement
are also shown with prefix "RB" in the informa-
tion bar at the bottom (see section 9.7.1).
9
10
11
12
13
14
15
16
Version 1.0 – 01/2024 105
1 9 Operation – 9.7 Measuring in Concentration Mode
NOTE
6 The turbidity correction function is not active
when the spectrophotometer leaves the factory. 3. Select turbidity correction 3 with 0 = deac-
The setting for automatic turbidity correction is tivated and 1 = activated (in light grey).
possible with all methods where the turbidity 4. Accept the settings with OK 4 .
7 correction makes sense. Where a method does
not permit turbidity correction, the button 2 is
greyed out.
9 5
2
1
10
NOTE
12
The use of the turbidity correction is indicated
1. Open settings 1 .
by the symbol 6 in the display.
2. Select turbidity correction 2 .
13
14
15
16
106 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
9.7.10 U
ser Recalibration (Standard
Adjustment)
3
A few pre-programmed methods and the user- Calibration for user-defined methods
defined methods for concentration measurement
provide the option to optimize the original cali- 1. elect the method manually (see section
S
bration stored with the method by means of a 9.5.1).
4
user recalibration. When creating a user-defined
method you can also allow a user recalibration NOTE
(see section 9.6). If a zero adjustment has not been performed,
When a method is called up for which a user the spectrophotometer informs you that you 5
recalibration is required, measurement is only need to perform a zero adjustment.
possible with a valid user calibration. The use of
the user recalibration is documented along with
the measured value and indicated in by the cor- 6
responding symbol in the display. Activation
of user recalibration and its date are also shown
with prefix "U-CAL" in the information bar at the
bottom (see section 9.7.1). 7
NOTE
A user calibration is always stored for the meth- 1
od presently called up. A user calibration is 8
erased only if a new user recalibration is carried
out.
9
2. Tap on "Recalibration" 1 .
3. The screen changes.
10
11
12
13
14
15
16
Version 1.0 – 01/2024 107
1 9 Operation – 9.7 Measuring in Concentration Mode
4
1. S
elect the method manually (see section
9.5.1), or by inserting a barcoded cell/
5 AutoSelector.
6 5
NOTE
6 If a zero adjustment has not been performed,
4. he data 2 of the current calibration are
T the spectrophotometer informs you that you
shown. need to perform a zero adjustment.
5. Tap on the “Value pair” button 3 to switch
7 to the view of the value pairs.
6. T
ap on the “Graph” button 4 to switch to
the graphic presentation of the calibration
curve. 1
8 7. ap on the Export button 5 to transfer
T
the data in the CSV format to an external
storage medium.
9 8. Tap on the Print button 6 to print out the 2
data.
12
13
14
15
16
108 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
10
4
7
4
3 5 5
12
13
11 6 8 16 15 17 14 7
14
15
16
Version 1.0 – 01/2024 109
1 9 Operation – 9.7 Measuring in Concentration Mode
NOTE
The calculated function maps the calculation of a
6 result (e. g. concentration) via a measured 16. T
o close the determination of the
absorbance in the form of a polynomial equation coefficients, confirm the data by tapping on
of the following type: the OK button 14 . If the field “U-CAL on” has
been activated, an icon 18 pops up in the
7 C = A0 + A1 x (Abs – E0) + A2 x (Abs – E0)2 measurement display.
17. T
o call up the calibration data again, select
where: Settings 1 and Recalibration 2 again.
C = measurement result (e. g. concentration) 18. T
ap on the Export button 15 to transfer
8
A0, A1, A2 = coefficients (polynomial) the data in the CSV format to an external
Abs = measured absorbance storage medium.
E0 = absorbance of the reagent blank value
19. Tap on the Print button 16 to print out the
9 data.
15. You can now enter an ID or a specific batch
20. To close the operation without accepting the
code for the calibration. Tapping on the field
data, tap on the X button 17 . All entered
“Batch ID” 13 öopens a virtual keyboard.
data are deleted.
10 Enter the code and confirm by tapping on the
OK button 14 .
11
12
13
14
15
16
110 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
9.7.11 MatrixCheck
The MatrixCheck function is used to check if Practical instructions
3
the photometric determination is disturbed by • After evaluating the measured value of the
other substances present in the sample (sample sample, the
matrix). The MatrixCheck can be carried out by spectrophotometer suggests for the Matrix-
spiking or by diluting. The spectrophotometer Check to spike
4
enables a simplified MatrixCheck with the aid of or dilute the sample and standard with suitable
the Spectroquant® CombiCheck R-2 addition so- volumes.
lution or a pre-programmed ready-to-use stan- For each spiking or dilution, the relevant nomi-
dard. The MatrixCheck can be carried out im- nal concentration value is displayed 5
mediately. The volumes required for the sample • To be able to reliably recognize matrix effects
and standards are displayed on the screen. The by spiking, the increase in volume after spiking
MatrixCheck is then carried out with a single should be small
spiking. For the MatrixCheck with a standard of • To be able to reliably recognize matrix effects 6
your own, however, you can enter the number of by diluting, the dilution factor should be high
spikings or dilutions yourself (max. 3). • You can carry out the MatrixCheck as a series
of measurements, consisting of up to three
MatrixCheck by spiking determinations with different spiking volumes 7
For the MatrixCheck by spiking, the photomet- or dilutions respectively
ric determination is repeated after a defined • Prepare all test sample solutions simultane-
amount of analyte has been added to the test ously at the
sample in the form of standard solutions. Recov- beginning of the series of measurements 8
ery of the addition is automatically calculated as
follows: NOTE
The spectrophotometer suggests the optimal
Recovery of addition [%] = 100 x 9
version for the MatrixCheck. Depending on the
{measurement value (sample + standard solution) – concentration of the sample in relation to the
measurement value (sample)}/{nominal value measuring range the spectrophotometer sets
(sample + standard solution) – measurement spiking or dilution. If both are possible you can
value (sample)} 10
make your own choice.
15
16
Version 1.0 – 01/2024 111
1 9 Operation – 9.7 Measuring in Concentration Mode
5 9
7
8 4 3
10
10
11
11
6 7 8
12
13
14
15
16
112 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
3 Toggle switch Dilution/Spiking Dilution/Spiking option. (Acceptance of the instrument pre-sets is recommend-
ed. Switching the option is possible only as long as this does not cause values 4
to be outside the measuring range.)
Figure for the standard volume, in case of dilution the volume of distilled water 6
7 Standard (ml)
is displayed
7
9 Concentration Only active for user-defined standard (concentration definable).
8
If Spiking is Active:
9
1 2
3 4
10
11
15
16
Version 1.0 – 01/2024 113
1 9 Operation – 9.7 Measuring in Concentration Mode
12
1 2
13
14
3. Activate the input field 3 for the lower limit.
15
16
114 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
5
5
8
6
4. virtual numerical keypad 4 pops up.
A
Now enter the numerical value for the lower
limit and tap the OK button 5 to confirm.
5. Repeat this procedure to enter the upper 10. After the measurement, the position of 7
limit. the measurement result is shown on the
measurement-range screen 8 .
11. To deactivate the user-defined range, open
the “Settings” menu 1 and select “User- 8
defined range” 2 .
6 10
9
11
12. T
ap on the “C” button 9 . The user-defined 12
range is reset. The upper and lower limits
are shown as zero values.
7
13
14
16
Version 1.0 – 01/2024 115
1 9 Operation – 9.7 Measuring in Concentration Mode
9.7.13 Differentiation
For certain methods there is a special
3
“Differentiation” function that enables the
user to differentiate between varying chemical
formulas of the analyte when interpreting
measurement results. In some methods for the
4
determination of chlorine, for example, total
chlorine can be distinguished from free chlorine
and the proportion of bound chlorine can be
5 10 automatically calculated and displayed.
After selecting the method, you can activate the
differentiation mode in the following manner:
6
13. Accept settings by tapping on the OK button
10 .
2
14. The screen changes to the Measurement
1
7 mode.
15. The spectrophotometer is now ready to start
the measurement.
9
1. Open the menu “Method settings” 1 .
2. Select “Differentiation” 2 .
10
11
3
12 4
13
3. ctivate the differentiation function by
A
tapping on the shift key 3 . The function is
activated when “I” appears against a light
14 gray background 3 .
4. Accept the settings by tapping on the OK
button 4 .
5. The screen changes to the mode “Measure-
15 ment with differentiation”.
6. The spectrophotometer is now ready to start
the measurement.
16
116 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode 1
9.7.14 Plausibility
NOTE The function “Plausibility” is available for the
3
For those methods for which the differentiation ammonium methods.
option is available, the procedures for activation Once it has been activated, the function remains
and the measurement are individually described permanently switched on for all ammonium
in detail in the manual section “Analyical methods.
4
Procedures and Appendices”.
NOTE
In the practical environment it has emerged that
when the sample contains very high ammonium 5
concentrations a different color occurs and the
measurement signal in the photometric
5 6 measurement is no longer proportional to the
ammonium concentration. 6
In these cases, the reaction solution is no longer
yellow-green to green in color, the yellow
component becomes diminished, and the
solution turns turquoise to blue in color. 7
In this case a reliable statement on the
ammonium content is no longer possible, in the
worst case resulting in a substantial
7. o deactivate the differentiation function,
T misinterpretation of the ammonium content with 8
open the menu “Method settings” 5 . serious consequences for the environment.
The submenu shows the available settings When the optional plausibility check in the
options. Prove plus spectrophotometer is activated,
8. Select “Differentiation” 6 . the measurement is carried out at several 9
9. Deactivate the differentiation function by wavelengths and the yellow component of the
tapping on the shift key 3 . The function is reaction solution is also measured. The Prove plus
deactivated when “0” appears against a light spectrophotometers are thus capable of
gray background. recognizing this deviation in the color and giving 10
10. Accept the settings by tapping on the OK a corresponding warning signal.
button 4 . This plausibility check helps the user to avoid
11. The screen changes to the Measurement misinterpreting the results of the ammonium
mode. concentration.
11
12. The spectrophotometer is now ready to start
the measurement.
12
13
14
15
16
Version 1.0 – 01/2024 117
1 9 Operation – 9.7 Measuring in Concentration Mode
1
2
6
NOTE
The active status of the plausibility check func-
tion is shown on the display by the symbol 5 .
7 1. Open the menu “Method settings” 1 .
2. Select “Plausibility” 2 .
9
3
10 4
11
3. ctivate/deactivate 3 the plausibility
A
check function with 0 = deactivated or 1 =
activated (light gray).
12 4. Accept the settings by tapping on the OK 7
button 4 .
5. The screen changes.
6. The spectrophotometer is now ready to start
13 the measurement.
NOTE
14 If the plausibility check func¬tion is active and a
very high ammonium concentration is present in
the sample (concentration considerably beyond
the measuring range of the selected method), a
15 window with a warning message 6 pops up.
After confirming by tapping on the OK button,
the measurement result is shown as “--- ” 7 .
16
118 Version 1.0 – 01/2024
9 Operation – 9.7 Measuring in Concentration Mode – 9.8 Ad hoc Measurement 1
(without selecting a specific method)
NOTE
3
If a zero cell (cell with distilled water) is
measured as the sample when the plausibility
check func-tion is active, the warning message
and the result “---” 7 also appear. 4
The reason for this is that the zero cell does not
show the yellow color described above.
10
11
1 2 3
12
13
14
15
16
Version 1.0 – 01/2024 119
1 9 Operation – 9.8 Ad hoc Measurement (without selecting a specific method)
4
8
1
2 5 3
5
6 4 7 6 9
7
1. elect type of measurement: absorbance or
S 4. The screen changes 8 .
transmission measurement by tapping on 5. Perform zero adjustment by inserting a cell
the button 1 for the required measurement with distilled water or tapping on the Start
8 (the activated selection is light grey). zero button 9 .
2. Define the wavelength(s) for the measure-
ment(s). Do this by tapping on the button 2 .
A blue frame appears. Now enter the re- 10
9 quired wavelength by tapping on the keypad
3 .
NOTE
10 Tapping on the + button 4 brings up a new input
field. Different wavelengths can be programmed
11
for the measurement. The selection can be delet-
ed by tapping on the Delete symbol 5 .
11
NOTE
Invalid entries are shown in red and cannot be 6. he Start zero button changes to the Start
T
12 accepted. button. The instrument is ready to start mea-
suring.
7. Start the measurement by inserting a cell
3. onfirm your input by tapping on the Start
C
with sample or t apping on the Start button
button 6 .
13 To cancel, tap on the X button 7 .
11 . The measurement results are displayed
in the absorbance column 10 .
14
15
16
120 Version 1.0 – 01/2024
9 Operation – 9.8 Ad hoc Measurement (without selecting a specific method) 1
4
1
2
5
2
3
4
5 6
7
16 mm S-ID 0125 NOTE
Path length of inserted cell Sample ID with prefix "S-ID" A spectrum can comprise up to a maximum of
empty
1000 measurement points. If the entry is invalid,
it is shown in red and cannot be accepted. 8
empty
12
13
14
15
16
Version 1.0 – 01/2024 121
1 9 Operation – 9.8 Ad hoc Measurement (without selecting a specific method)
4
6
6 7
7
5. Tap on the Start button 5 . ADHOC SPECTRUM S-ID 0033
6. The screen changes 6 . Measurement mode Sample ID
7. Perform zero adjustment by inserting a cell with prefix
8 with distilled water or tapping on the Start "S-ID"
10
8
11
14
15
16
122 Version 1.0 – 01/2024
9 Operation – 9.8 Ad hoc Measurement (without selecting a specific method) 1
4
9
1
3 4
2 5 5
6
7
8
10 6
7
4. The screen changes 9 .
5. Perform zero adjustment by inserting a cell
with distilled water or tapping on the Start
zero button 10 . 8
10
1. elect type of measurement: absorbance or
S
11
transmission measurement by tapping on
the button 1 for the required measurement 11
(the activated selection is light grey).
2. Set the measuring range and the duration.
• Wavelength 2 6. Start the measurement by inserting a cell
• Unit 3 with sample or t apping on the Start button 12
• Interval 4 11 .
• Delay 5 7. The screen changes, the instrument is ready
• Duration 6 to start measuring.
• Slope factor 7 13
NOTE
Invalid entries are shown in red and cannot be
accepted. 14
16
Version 1.0 – 01/2024 123
1 9 Operation – 9.8 Ad hoc Measurement (without selecting a specific method)
7
ADHOC KINETIC S-ID 0034
Measurement mode Sample ID
with prefix
8 "S-ID"
10 mm 00:00:30 00:00:05
Path length of Duration Time
inserted cell interval
9 empty
10
11
12
13
14
15
16
124 Version 1.0 – 01/2024
9 Operation – 9.9 Spectrum 1
9.9 Spectrum
Baseline
A baseline has to be recorded before a spectrum
can be recorded. The baseline has to cover at 7
least the wavelength range of the spectrum be-
ing recorded. Once the baseline is measured, it
remains stored in the spectrophotometer until
• A new baseline is recorded 8
• Ad hoc Spectrum Mode is exited
• The loaded Spectrum method is exited
• The spectrophotometer is switched off
9
1
10
15
16
Version 1.0 – 01/2024 125
1 9 Operation – 9.9 Spectrum
6
4. Insert the cell with the sample vertically
until it touches the bottom (rectangular cells
should touch the left edge of the cell com-
7
partment; the opaque sides of the rectangu- 1103 user spectrum S-ID 0187 S1
lar cell must point to the front and rear). Sample ID with prefix
Method number Method name
"S-ID"
8 16 mm 400 – 500 nm 1 nm
Path length of Scan range Scan interval
inserted cell
empty
9
10 2
11
5. Recording of the spectrum starts automati-
cally.
6. Once the spectrum of your sample has been
12 recorded you have the following options:
• Evaluate the spectrum on the display immedi-
ately (see section 9.9.3)
• Print the spectrum by tapping on the Printer
13 button 2 either as a graph on a connected
printer – or as a pdf file, if Print to pdf is acti-
vated and a USB device is connected
• Save the spectrum to the result list. If
14 AutoStore is on, this is done automatically
15
16
126 Version 1.0 – 01/2024
9 Operation – 9.9 Spectrum 1
4
15 9
14 8
5
13 7
12 6 6
3
11 10
2 4 1 5 2
9
A spectrum can be evaluated immediately after 5. The following mathematical functions for
measurement. Stored spectra can be loaded and various evaluating and calculating operations
evaluated from the results list as well. The fol- are available for selection: 10
lowing tools are available for editing: •
Derivate 12 : calculates the derivative of the
total spectrum. To calculate the second and
1. Moving to individual measurement points: third derivative, the function can be carried
• Activate action icon 1 and use the next left/ out several times 11
next right icons 2 to move to all individual • Compare spectrum 13 : loads a second spec-
measurement points. The coordinates (wave- trum into the same diagram for direct com-
length and absorbance) of the respective mea- parison
surement point appear in the Info box 3 • Subtract spectrum 14 : subtracts a stored 12
• Activate icon 4 to move to maximum values spectrum from the current spectrum
and icon 5 to move to minimum values • Add spectrum 15 : adds a stored spectrum to
2. Switch between graph view 6 and table the current
view 7 . spectrum 13
3. In the graph view use icons 8 to zoom out
and 9 to zoom in. Use navigation icon 10 to NOTE
optimize the position of sections of the graph The addition and subtraction of two spectra can
in the display. be applied only to the common wavelength
14
4. Tap icon 11 to return to the original spectrum. range of both spectra.
15
16
Version 1.0 – 01/2024 127
1 9 Operation – 9.10 Kinetics
9.10 Kinetics
3
9.10.1 General Information
The kinetics function enables the temporal trac-
ing of the absorbance or transmission of a sam-
4 ple at a certain wavelength.
The spectrophotometer automatically calcu-
lates the slope between two adjacent measuring
points from the available measurement data.
5 The catalytic activity can also be determined and
displayed if required. To record the kinetics, the
spectrophotometer carries out single measure- 2
ments at regular intervals (measuring interval)
6 and stores the measured values as a time func-
tion.
NOTE
9 1 The zero adjustment can also be performed
without a cell being inserted (measurement
against air). Tap on the Start zero button to start
the procedure.
10
11
13
14
15
16
128 Version 1.0 – 01/2024
9 Operation – 9.10 Kinetics 1
10 mm 00:01:00 00:00:05
Path length of Duration Time interval 5
inserted cell
empty
3 6
7
4. ollowing successful zero adjustment the
F
Start zero button 2 becomes the Start but-
ton 3 .
5. The instrument is ready to start measuring 8
the sample.
6. Insert the cell vertically until it touches
the bottom (rectangular cells should touch
the left edge of the cell compartment; the 9
opaque sides of the rectangular cell must
point to the front and rear).
7. Recording of kinetics starts automatically.
10
11
12
13
14
15
16
Version 1.0 – 01/2024 129
1 9 Operation – 9.10 Kinetics
4
5
5 6
6
8
10 9
7
4 4
8
13
14
15
16
130 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
14
15
16
Version 1.0 – 01/2024 131
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
NOTE
Wavelength accuracy
The value of the tolerance consists of the toler-
Normally test items of known absorbance max-
7 ance of the standard (listed in the lot-specific
ima at defined wavelengths are used to monitor
certificate) and the specified tolerance of the
wavelength accuracy. The instrument is pre-pro-
spectrophotometer (see section 12).
grammed with standard AQA1 tests that can be
8 performed using Spectroquant® test items. Such
test items are, for instance: Certipur® UV/VIS- Spectral resolution
Standard 6. Spectral resolution can be monitored using the
Each pack is provided with a lot-dependent test spectrum of a 0.02% solution of toluene in hex-
certificate with all nominal values (wavelengths ane. The minimum ratio between the absorbance
9
with absorbance maxima) and tolerances of in the maximum at 269 nm and the absorbance
the test standards. These nominal values and in the minimum at 266 nm is a measure for
tolerances are already pre-programmed in the the spectral resolution. The instrument is pre-
10 spectrophotometer. Compare them with the programmed with standard AQA1 tests that can
lot-dependent values and adjust them if needed be performed using Spectroquant® test items.
(see section 9.11.8). The test item used here is: Certipur® UV/VIS-
Standard 5.
11 NOTE
The value of the tolerance consists of the toler-
ance of the standard (listed in the lot-specific
certificate) and the specified tolerance of the
12 spectrophotometer (see section 12).
13
14
15
16
132 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
NOTE
Only persons in the administrator user group 4
have access to the settings for AQA checks. Any
registered user can perform the AQA2 test.
Spectroquant® CombiCheck standards are ready-
to-use multiparameter standards, i.e. they can 5
be used for several test kits (methods). Standard
solutions are ready-to-use single-parameter
standards, i.e. they can be used for single test
kits (methods). In addition to the above solu- 6
tions, single-parameter standard solutions (z.B.
Certipur®) can also be used. They are adjusted
by dilution to the respective end concentration.
The end concentration should be approximately 7
in the middle of the measuring range.
NOTE 8
Suitable CombiCheck standards and single-pa-
rameter standards are listed in the catalog
"Water, Food and Enviromental Analytics" or on
the Internet. 9
10
11
12
13
14
15
16
Version 1.0 – 01/2024 133
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
6
4
7 1 2 3 4 5
2 AQA2 status Summary of the status of all activated AQA2 tests. (OK, fail, expired, next test in xx days)
10
3 AQA1 Activate, edit, perform and create AQA1 tests
Activated AQA1 test(s) and/or AQS2 test(s) are failed or expired. Tapping the symbol
6 Attention
changes the display and opens an overview of the AQA tests involved.
12
13
14
15
16
134 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
1
6
7
1. Tap on the AQA1 status button 1 .
2. The screen changes and a status overview of
the activated AQA1 tests is shown.
8
9
4
5
10
2
11
12
3
3. The screen shows the following information: • Three different status symbols 5 : 13
• AQA1 method number 2 ! = expired/invalid; ü = test OK; - = test
• Name of the AQA1 test 3 failed
• Number of days for which the AQA1 test is still 4. For quality or documentation purposes we
valid before a new test has to be performed 4 recommend printing the list.
14
15
16
Version 1.0 – 01/2024 135
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
6 1
7
1. Tap on the AQA2 status button 1 .
2. The screen changes and a status overview of
the activated AQA2 tests is shown.
8
7
10
11
12
3 4 5 6
13 3. The screen shows the following information: • Three different status symbols 7 :
• AQA2 method number 2 ! = expired/invalid; ü = test OK; - = test
• Name of the AQA2 test 3 failed
• Measuring range of the AQA2 check function 4 4. For quality or documentation purposes we
14 • Number of days for which the AQA2 test is still recommend printing the list.
valid before a new test has to be performed
5 , or number of measurements before a new
test has to be performed 6
15
16
136 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
1
6
7
1. Tap on the AQA1 button 1 .
2. The screen changes and a list of all AQA1
tests stored in the instrument is shown.
8
4 9
2 6
10
11
12
3 5
3. The screen shows the following information: • Three different status symbols 5 to remind 13
• AQA1 method number 2 you that a test is due. ! = expired/invalid;
• Name of the AQA1 test 3 ü = test OK; - = test failed
• Number of days for which the AQA1 test is still • An empty circle means that the AQA1 test is
valid before a new test has to be performed 4 not activated 14
• Input buttons for editing 6 and creating 7
AQA1 tests
15
16
Version 1.0 – 01/2024 137
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
9.11.7 A
ctivating and Deactivating an 9.11.8 Editing an AQA1 Test
AQA1 Test
3
Depending on the test type selected
• Photometric accuracy
4
6 • Wavelength accuracy
• Stray light test
• Spectral resolution (Prove 600 plus only)
5
the particular display screen changes. The input
values should be taken from the lot-specific cer-
tificates for the test items and adjusted to suit
6 as follows:
8
8
12 NOTE
Before activating the lot-specific values for the
current test item, compare with the existing en-
13 tries for the input fields and make changes ac-
cordingly (see section 9.11.8).
14
15
16
138 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
1 1
6 7 8
2 3 4 2 3 4
4
5 5
13
12 5
12
9 10 11 9 10 11
Wavelength accuracy (W) Spectral resolution (Prove 600 plus only) (R)
7
1 1
6 7 8
2 3 4 2 3 4 8
5 5
14
12 9
12
9 10 11 9 10 11
10
11
12
13
14
15
16
Version 1.0 – 01/2024 139
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
3
Item Display field Description Test
type
P, W,
1 Name Name of the test item
S, R
4 P, W,
2 EXP Expiry of the test item as shown in the certificate
S, R
P, W,
3 Lot number Lot number of the test item as shown in the certificate
S, R
P, W,
5 4 "0"/"I" (OFF/ON) Deactivate/activate the test
S, R
Test interval (input in weeks). For an active test the instrument reminds you when an P, W,
5 Interval
AQA1 test is due. S, R
6 Cell Pre-set name of the cell P, W
6 7 Target values Lot-specific nominal values P, W
8 Tolerance Tolerance range for the measurement value. P, W
(Tolerance = measurement uncertainty of the test item + specification of the spectro-
photometer)
7 9 Save Adopt values into the instrument
P, W,
S, R
10 Delete Delete user-defined AQA1 tests P, S
P, W,
11 Close Close display field
S, R
8 P, W,
12 Numeric field Touch the individual field to enter values
S, R
13 Transmission Pre-set instrument-specific value (in% transmission) S
Nominal (Amax/ Pre-set instrument-specific value (minimum ratio between absorbance in the maximum
9 14
Amin) and absorbance in the minimum)
R
11
12
13
14
15
16
140 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
1
5
6
NOTE 4. he screen changes. A type-specific screen
T
In order to perform an AQA1 test it must be ac- appears.
tivated (see section 9.11.7). 5. For further actions follow the on-screen
7
commands 3 .
6. Insert the corresponding reference cell (e. g.
1. Tap on the AQA1 button 1 .
the zero cell as reference cell in the Photo-
Check). 8
7. Insert following test cell as prompted in the
command line.
9
5
2 4
10
11
12
2. T he screen changes and you are presented 8. A tick appears following a successful test 4 .
with a list of all AQA1 tests stored in the 9. If all test steps are successful and the AQA1
instrument. test is passed, a tick appears in the com-
3. Select an AQA1 test by tapping on its name mand line 5 . 13
2 .
14
15
16
Version 1.0 – 01/2024 141
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
7 114693 PhotoCheck
Article number Method name
(first 6 digits of order
no.)
11
12
13
14
15
16
142 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
9.11.10 C
reating a User-defined AQA1
Test
3
3 4 4
5
1
10
11
2. T
he screen changes and you are presented 6. For editing (see section 9.11.8).
with a list of all AQA1 tests stored in the 12
instrument. NOTE
3. Tap on the Add button 2 . In the case of the user-defined AQA1 test for
photometric accuracy the input fields for the test
conditions (cell designation, 13
testing wavelength, nominal values for absor-
bance and absorbance tolerance) are added to
the screen by tapping on the
+ button 5 . 14
15
16
Version 1.0 – 01/2024 143
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
1
6
7 5 6
9 3 7
4 8
10
11
3. The screen shows the following information: • Number of measurements before a new AQA2
12 • Number of the method being tested 2 test has to be performed 6
• Name of the method being tested 3 • Four different status symbols 7 : ! = test
• Display of the measuring range of the method expired/invalid; ü = test OK; - = test failed;
being tested 4 = not activated
13 • Number of days for which the AQA2 test is still • Buttons to edit 8 AQA2 tests
valid before a new test has to be performed 5
14
15
16
144 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
9.11.12 A
ctivating and Deactivating an 9.11.13 Editing an AQA2 Test
AQA2 Test
3
2
4
2
1
5
3 4
1. Tap on the Edit button 1 . The input values should be taken from the lot-
specific certificates for the AQA2 test items and
2. The screen changes. 7
3. Tapping on the ON/OFF button 2 activates/ adjusted to suit as follows.
deactivates the AQA2 test. (I = on, 0 = off, 1. Tap on the Edit button 1 .
the light grey background shows which state 2. The screen changes. A method-specific
is active). screen appears. 8
See example screen on the next page.
NOTE 3. This submenu offers the following setting
options:
Before activating the lot-specific values for the
See description in the table below the ex- 9
current test item, compare with the existing en-
ample screens on the next page.
tries for the input fields and make changes ac-
4. Make your individual changes in the type-
cordingly (see section 9.11.13).
specific screens and store them by the Save
button 3 . 10
5. To close the editing screen touch the Close
button 4 .
11
12
13
14
15
16
Version 1.0 – 01/2024 145
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
1
4
3 4
8 12
5
9
6 11
7
13
2 5
8
9 7
10
10
11
12
13
14
15
16
146 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
3
Item Display field Description
Selection of Selection of the AQA2 standard (a choice can be made between Spectroquant®
2 4
standards pre-programmed standards such as
CombiCheck and a freely definable standard)
3 EXP Expiry of the standard as shown in the certificate for the standard
4 Lot number Lot number of the standard as shown in the certificate for the standard 5
5 "0"/"I" (ON/OFF) Deactivate/activate the test
6 Interval mode Test interval choice between weeks and and number of measurements
6
7 Interval (values) Enter test interval. For an active test the instrument reminds you when an AQA2 test is due.
First 6 digits of the order number of the pre-programmed test kit corresponding to the se-
13 Article number
lected method
* The user must edit the tolerance range according to his/her requirements. The tolerance range should include typical errors 9
(measurement uncertainty) of the test medium used and of the method being checked. The typical error of the test medium
used can be taken from the batch-specific certificate of the test medium. The typical error of the method being checked
must be calculated by the user under his/her conditions.
10
11
12
13
14
15
16
Version 1.0 – 01/2024 147
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
4 3
5 1
6
NOTE 5. erform an analysis as per the method de-
P
In order to perform an AQA2 test it must be ac- scription using the selected AQA2 standard
tivated (see section 9.11.12). as sample and insert the cell.
7 6. Measurement starts automatically.
7. The measurement result appears in the dis-
1. Tap on the AQA2 button 1 .
play.
8. A tick appears following a successful test 4 .
8
9
2
4
10
11
14
15
16
148 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
7
114942 Nitrate CombiCheck 20
Article number Method name AQA2 Standard Name
(first 6 digits of order
no.)
8
HC123456 EXP 12/31/2015 AQA2 EXP 08/28/2015 10 mm
Lot number test kit Expiration date Interval Path length
test kit with Prefix of inserted
"EXP" cell 9
ZA 08/21/2015 RB 0,100 A 08/21/2015
Date of Zero Adjust- Date + value user reagent
ment with Prefix "ZA" blank with Prefix "RB"
10
11
12
13
14
15
16
Version 1.0 – 01/2024 149
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
7 7
4
5 5
6
1
6
7
1. Tap on the PipeCheck button 1 . 5. Insert the reference cell.
2. The screen changes and a list of all Pipe- 6. After the reference cell has been successful-
Check tests stored in the instrument is ly measured, tick 4 appears on the display.
8 shown.
9
2
10
11
3. Select a PipeCheck by tapping on the name 7. window pops up in which a name or code
A
12 2 . can be entered for the pipet being tested
(e. g. manufacturer, serial number or similar).
Tap on the input field 8 and enter the name
or code. Tap the OK button to accept the
13 3
entry.
8. The screen changes.
14
15
16
150 Version 1.0 – 01/2024
9 Operation – 9.11 AQA (Analytical Quality Assurance) 1
3
9. Insert the "Check" cell.
10. After the "Check" cell has been successfully
measured, tick 5 appears on the display.
11. The difference between reference and Check 4
cell is calculated automatically. If this value
6 is within the pre-programmed tolerance,
the PipeCheck is passed shown by text and
tick 7 . 5
6
9
NOTE
9
If this value of the difference is outside the tol-
erance, the test has failed and a - 9 appears.
10
11
12
13
14
15
16
Version 1.0 – 01/2024 151
1 9 Operation – 9.11 AQA (Analytical Quality Assurance)
10
11
12
13
14
15
16
152 Version 1.0 – 01/2024
9 Operation – 9.12 Timer 1
9.12 Timer
You can use the timers to remind you
3
by an acoustic signal of a time inter-
val that has expired.
6
2
8
6
3 11
6. Tap on the Start button 7 to start the timer.
Tap on the Stop button 8 to stop the timer.
The count down is interrupted.
Tap on Start button 7 again will continue 12
4
the count down. When the count down is
finished an acoustic signal sounds. The
insertion of a barcoded round cell enables
a measurement to be started immediately. 13
The insertion of an AutoSelector enables a
method to be selected immediately. Tapping
on the Start button 7 is active again and the
timer can be started once more. 14
2. The input window 3 opens. 7. Tap on the X button 9 to close the timer.
3. Enter the required time by tapping on the The display changes to timer list.
corresponding field and confirm with OK 4 .
15
16
Version 1.0 – 01/2024 153
1 9 Operation – 9.13 Results and Measurement Datasets
4
11
14
15
16
154 Version 1.0 – 01/2024
9 Operation – 9.13 Results and Measurement Datasets 1
1 3 3 3 3 3 4 5
5
8
2 2 2
10
6
8
7 6 9
10
By tapping on the Results button 1 you can ac- • Export
selected results 6 . To export results
cess the result list via the main menu. they have to be selected by ticking the box 10
• Print selected results 7 . To print results they
The menu offers you the following choices: have to be selected by ticking the box 10 11
• Sort ascending/descending 2 • Scrollbar 8 . Scroll through the result list by
• Filter result list 3 – (see section 9.13.3) tapping on the arrows of the scrollbar
• Clear filter 4 • Panoramic view of selected results from one
• Select/deselect all results 5 . specific method 9 (see section 9.13.4) 12
• Results are selected by ticking the box 10 .
Remove the selection by tapping on the tick
symbol and the box will be emptied
13
14
15
16
Version 1.0 – 01/2024 155
1 9 Operation – 9.13 Results and Measurement Datasets
7
1 2 3 4 5 6 7
The individual result line shows the following • Result (e. g. in units or passed/failed, depend-
8 information: ing on measurement mode) 4
• Measurement mode (e. g. concentration, spec- • Analyte tested 5
trum, kinetic, AQA) 1 • Sample ID 6
• Method number 2 • Tick box to select a result line to print or ex-
9 • Date and time of measurement 3 port the results 7
11
12 1
13
1. Tap on an individual result line 1 2. A screen showing all details of one result or
dataset opens.
14 3. Tap on the print button 2 to print the indi-
vidual dataset on a printer or as a pdf.
15
16
156 Version 1.0 – 01/2024
9 Operation – 9.13 Results and Measurement Datasets 1
9.13.3 F
iltering Results to Further
Process M easurement Datasets
3
1 2 3 4 5
5
6
Specified filter criteria can be set to select cer- • Filter by date. Tapping on the Date button
tain results and datasets to be exported, print- 3 brings up an input mask for the date range
7
ed, displayed or deleted. from … to ...
The following criteria can be set: A calendar screen pops up.
• Filter using the individual measuring
mode 1 . The list shows the following modes: 8
8 8
concentration, kinetic, spectrum, ad hoc,
AQA1, AQA2, PipeCheck and MatrixCheck
9
2
9
10
11
• Use the forward and back buttons 8 to scroll
through the calendar. Select the required date
and confirm by tapping on the OK button 9 .
The calendar closes. 12
•F
ilter using character strings. Tapping on • Tap on the OK button 6 to adopt the selected
the Search icon 2 brings up the keypad dis- period for filtering.
play field. Enter search criteria such as method
name, method number or article number (first NOTE 13
six digits of order number) no decimal point.
Keeping your finger on the “Forward“ and “Back“
Tap on OK to activate the search filter
buttons 8 scrolls through the calendar forwards
and backwards year by year.
14
15
16
Version 1.0 – 01/2024 157
1 9 Operation – 9.13 Results and Measurement Datasets
5
1
6
• Filter by sample ID 4 . Tapping on the but-
2
ton brings up a s election bar 10 which you can
use to filter according to sample ID
7
11
1. se the filter option to create a result list for
U
5
one single method (see section 9.13.3).
8 2. Select the results to be included into the
panorama value control card by ticking the
boxes 1 .
3. The Panorama button is activated 2 .
9
10 12
15
16
158 Version 1.0 – 01/2024
9 Operation – 9.13 Results and Measurement Datasets 1
9.13.5 P
rint Results and Measurement
Datasets
3
If a postscript printer is connected to the spec-
trophotometer, results and measurement datas-
ets can be printed on paper by touching the Print
button (see section 8.3.2). 2 4
In addition, you can print the results and mea-
surement datasets as a pdf. Connect an USB
memory device and activate "Print to pdf" in the
submenu "Interface" in the "System" menu (see 5
section 9.2.2). To print to pdf touch the Print but-
3
ton.
11
1
12
13
15
16
Version 1.0 – 01/2024 159
1 9 Operation – 9.13 Results and Measurement Datasets
1
Data Transfer from Spectroquant®
Prove plus with USB Memory Device
7
NOTE
The reliability of data stored on USB memory
When the function is activated, the symbol for
devices depends on the quality of the memory
8 the function “Delete” 1 is shown in the screen
device and the data transmission.
of the results list.
15
16
160 Version 1.0 – 01/2024
9 Operation – 9.13 Results and Measurement Datasets 1
3
NOTE
The data are exported as CSV files. To open
them in a spreadsheet make sure that your PC
calculation program is set to the same decimal 4
separator as the spectrophotometer (see section
8.2.4).
1
6
8
1. elect the results you want to export by
S
ticking the box 1 .
9
10
11
2
12
2. E xport the selected results by tapping on the
Export button 2 .
3. The in-progress icon appears during the 13
data-export procedure.
4. When the in-progress icon disappears, wait
for some more time before removing the
USB device to make sure all data are trans- 14
ferred.
15
16
Version 1.0 – 01/2024 161
1 9 Operation – 9.14 User Management
6
User rights in detail
Firmware update X
8
Language update X
Method update X
Run AQA1/AQA2 X X
Create/edit AQA1/AQA2 X
11 Run measurements X X
Evaluate measurements X X
Backup/restore X
13
14
15
16
162 Version 1.0 – 01/2024
9 Operation – 9.14 User Management 1
9.14.1 A
ctivating/Deactivating the User
Management Function
3
If the user management function is activated by
ticking 1 , each user has to log in to the spec-
trophotometer. After the login, the user has
certain rights depending on the user group. Only 4
an administrator can deactivate the user man-
agement function. If the function is deactivated, 1
every user has full instrument rights.
2
5
10
11
12
13
14
15
16
Version 1.0 – 01/2024 163
1 9 Operation – 9.14 User Management
1
6
3
4
4
4
5 2
On the right part 1 of the screen "User Manage- 5. I f the new user has been created success-
ment" administrators or users can be created. fully, the user name appears on the left 6 .
7 1. Tap on the Add button 2 to start to create
an administrator/user.
2. Select whether you want to create an admin- 7
istrator or user by tapping on field 3 .
8
NOTE
The first person who is created is automatically
assigned administrator rights. Here it is not
9
possible to select “Administrator” or “User”.
7
11
12
5
13
15
16
164 Version 1.0 – 01/2024
9 Operation – 9.14 User Management 1
7 6
5
1
2
6
3. ap on the Edit button 3 .
T
4. You can edit the access rights (Administrator
3
or User) 4 and change the password 5 .
5. Confirm your changes with the OK button 6 . 7
6. To delete a user proceed as before and tap
on the Delete b utton 7 .
9 8
10
11
7. onfirm your changes with the OK button 8 .
C
You can quit your decision with the X button
9 .
12
13
14
15
16
Version 1.0 – 01/2024 165
1 9 Operation – 9.15 Login/Logout
9.15 Login/Logout
I f the user management is activated (see sec-
3
tion 9.14.1), login is mandatory to receive user
or administrator rights.
To login into the spectrophotometer proceed
as follows:
4
5 2
5
4
1
6
1. Select Login/Logout 1 from the main menu. 5. nter the user password by using the key
E
8 2. Tap on the arrow to open the user list 2 . pad 5 and confirm it with OK 6 .
10
11
3. Select the user name from the user list 3 . 6. The screen changes to the start screen. The
12 4. Tap on the entry field 4 . user name of the logged in user appears in
the upper status bar 7 .
13
14
15
16
166 Version 1.0 – 01/2024
9 Operation – 9.15 Login/Logout 1
9.15.1 C
hange Password with User 2
Rights Only
If you are registered with user rights only, you
have to change your password in the Login/ 3
8 Logout mode.
1
4
3
5
2
1
9
4
10
NOTE
2. S elect Login/Logout 2 from the main menu.
To continue as a different administrator or user, 3. Tap on the lock icon 3 .
open the user list by tapping on the arrow 3
12
4. Enter the new password and confirm it in the
and select the user from the list. Enter the cor- entry fields 4 .
responding password as described above. 5. Store your changes with the Save button 5 .
6. Login with your new password (see section 13
9.15).
7. The screen changes back to the Login/
Logout mode.
Press Start again to continue. 14
15
16
Version 1.0 – 01/2024 167
1
10 Maintenance and Cleaning
7 1
Battery service life
The power consumption of the clock is very low.
The service life of high-quality batteries is at
8
least five years.
4. Close the battery compartment cover 1 .
Disposal of batteries
Dispose of the batteries at a suitable facility
9 NOTE
according to local legal requirements. Please
do not dispose of the batteries with household If you leave the spectrophotometer switched on
refuse. during the
Within the European Union, the batteries are exchange or insert the new batteries within a
10 minute after taking out the old ones, the date
removed at specialized treatment centers at the
end of the instrument's life. Instruments are tak- and time are retained in the spectrophotometer.
en to one of these specialized treatment centers
11 via the recycling system set up for this purpose.
12
13
2
14
16
168 Version 1.0 – 01/2024
10 Maintenance and Cleaning – 10.2 Exchanging the Lamp (Prove 100 plus) 1
CAUTION
There is a risk of explosion if unsuitable lamps
6
are used.
Use only the lamp that is intended for use with
your instrument.
7
Disposal of the lamp
Dispose of the lamp at a suitable facility accord-
ing to local legal requirements. Please do not
dispose of the lamp along with household refuse. 8
Within the European Union, the lamp is removed
at specialized treatment centers at the end of
the instrument's life. Instruments are taken to
one of these specialized treatment centers via 9
the recycling system set up for this purpose.
11
12
1
13
1. Place the spectrophotometer bottom-side up 4. arefully remove the lamp module from the
C
14
on a soft surface. lamp compartment. Do not touch the lamp.
2. Remove the screws of the lamp compartment Use the mount 1 to take it off.
cover with an appropriate screw driver.
15
16
Version 1.0 – 01/2024 169
1 10 Maintenance and Cleaning – 10.2 Exchanging the Lamp (Prove 100 plus) – 10.3 Cleaning
10.3 Cleaning
If a cell has broken or in the event of a reagent
3
accident, the spectrophotometer must be cleaned
immediately.
4 WARNING
Cells can contain dangerous substances. If the
contents are released, follow the safety instruc-
tions of the Material Safety Data Sheet (MSDS).
5 If necessary, apply appropriate protective mea-
1 sures (protective goggles, protective gloves
etc.).
6 CAUTION
5. I nstall the new lamp module in the lamp
compartment. Use the mount 1 to insert Do not turn the spectrophotometer upside down
the lamp module. to remove the liquid. Doing so can cause the liq-
7 Do not touch the lamp to ensure that the uid could come into contact with electronic com-
service life of the lamp is not affected! ponents and damage the spectrophotometer.
6. Close the lamp compartment cover with an
appropriate screw driver. CAUTION
8 The spectrophotometer has two drains at the
NOTE bottom through which the c ontents of broken
Upon renewed use of the spectrophotometer, re- cells or spilled liquid can drain off without caus-
boot the instrument and reset the lamp counter ing any damage to the instrument.
9 to zero (see section 9.2.1). If the self-test is
passed the instrument is ready for further mea-
surements.
10.3.1 Cleaning the Housing and Display
10
CAUTION
The housing components are made of synthetic
materials. Avoid any contact with acetone, simi-
11
lar solvents and detergents containing such sol-
vents. Wipe off any splashes immediately.
Displays: Avoid contact with concentrated miner-
al acids, concentrated caustic solutions, benzyl
12
alcohol and methylene chloride. Wipe off any
splashes immediately.
16
170 Version 1.0 – 01/2024
10 Maintenance and Cleaning – 10.3 Cleaning 1
1
12
13
14
15
16
Version 1.0 – 01/2024 171
1 10 Maintenance and Cleaning – 10.3 Cleaning
NOTE
9
Upon renewed use of the spectrophotometer,
reboot the instrument. If the self-test is passed
the instrument is ready for further measure-
10 ments. If the self-test fails, check whether the
detector lens is dirty and clean it (see section
10.3.4).
11
12
13
14
15
16
172 Version 1.0 – 01/2024
10 Maintenance and Cleaning – 10.3 Cleaning 1
10
1
11
12
13
1. Open the cell compartment cover 1 .
2. Press the two fittings inside of the cover 2
and rotate the cover slightly. It will turn and
you can take it out. 14
15
16
Version 1.0 – 01/2024 173
1 10 Maintenance and Cleaning – 10.3 Cleaning
5 NOTE
If the lens is often smudged make sure to pro-
tect the instrument from dirt, dust and evapo-
rates of chemicals. Check operation conditions of
6 temperature and humidity. They have to be in
accordance with the values specified in the tech-
nical data sheet (see section 12).
9
2
10
Proceed as follows to clean the detector lens: 2. C
ut off the end (approx. 2 cm) of a Dacron®
The detector lens is situated on the front left swab, e. g. HY-LiTE® sampling pen, Cat. No.
side of the 1.30102.0021.
11 rectangular cell compartment 3 . 3. Grasp the cut-off end with the tip of a pair
of tweezers or small pliers. Clean the lens 4
1. witch off the spectrophotometer
S 1 and with the dry head of the swab. To do
disconnect the power plug 2 . so, move the head from the center of the
12 lens outward in
circles. If there is persistent contamination,
moisten the swab with a little deionized wa-
ter or isopropanol.
13
NOTE
Upon renewed use of the spectrophotometer,
14 reboot the instrument. If the self-test is passed
the instrument is ready for further measure-
ments.
15
16
174 Version 1.0 – 01/2024
1
10
11
12
13
14
15
16
Version 1.0 – 01/2024 175
1
11 Error Causes and Trouble-shooting
3 Self-test does not start. • A cell is inserted in one of the cell • Remove the cell
Start button is inactive compartments • Then tap on the Start button
•
A foreign object is inserted in one of the • Remove the foreign object
two cell compartment • Then tap on the Start button
4 •
The cell compartment is dirty •
Clean the cell compartment (see section
10.3.2)
• Restart the instrument
• Instrument defective v Contact service department
5
Self-test failed • System check: Instrument defective • Contact service department
• Lamp check: Lamp defective • Prove 100 plus: Change the lamp
•
Prove 300 plus | 600 plus: Contact service
6 department
• Wavelength check: • Remove the foreign object
• Foreign bodies in the cell compartment •
Clean the lens (see section 10.3.4)
• Lens dirty If this happens repeatedly, check the oper-
• Instrument defective ating conditions (see section 8.1)
7 • Contact service department
12
13
14
15
16
176 Version 1.0 – 01/2024
11 Error Causes and Trouble-shooting 1
Data transfer to USB • Power supply of USB is interrupted • Connect power supply 3
not working • USB has been disconnected while • Wait for one more minute before
data transfer was still running disconnecting the USB device from the
instrument
10
11
12
13
14
15
16
Version 1.0 – 01/2024 177
1
12 Technical Data
The serial number of the spectrophotometer is printed on the type plate at the rear of the instru-
ment starting with "SN". The serial number is also stored in the instrument and can be looked up
3 under "System" and submenu "Information", the last line MCS serial contains the serial number of the
instrument (see section 9.2.1).
Measuring modes Concentration, absorbance, transmission, multi wavelengths, spectra and kinetics in
6 absorbance and transmission mode
Spectral bandwidth 4 nm
Wavelength accuracy ± 1 nm
Cell size 16 mm round cells, 10, 20 and 50 mm rectangular cells with automatic recognition
12
Minimum filling volumes 16-mm round cells: 4 ml
10-mm rectangular cells (Standard): 2 ml (Semimicro): 1 ml
20-mm rectangular cells (Standard): 4 ml (Semimicro): 2 ml
50-mm rectangular cells (Standard): 8 ml (Semimicro): 4 ml
13
Cell holder Removable for easy cleaning
Methods Programmed methods of all Spectroquant® cell and reagent tests, additional user-defined
ethods:
m
14 99 concentration mode, 20 kinetic mode, 20 wavelength scans
15
16
178 Version 1.0 – 01/2024
12 Technical Data – 12.1 Spectroquant® Prove 100 plus 1
3
Spectroquant® Prove 100 plus
Software and method Free updates on our website (www.sigmaaldrich.com/photometer-service) via internet 6
update and USB stick
Communication interfaces USB: 2 × USB-A (for printer, USB memory devices, keyboard or bar code reader),
1 × USB-mini-B
Ethernet: LAN connection 7
Data storage 7,000 single measured values from the measuring modes concentration, absorbance/%
transmission and multi wavelengths. 500 measurement result records of spectra,
kinetics, AQA1, and AQA2 methods each
Languages English, German, Spanish, French, Italian, Brazilian-Portuguese, Chinese (simplified and 8
traditional), Japanese, Russian, Bulgarian, Czech, Danish, Dutch, Greek, Hungarian,
Indonesian, Malay, Macedonian, Norwegian, Polish, Romanian, Serbian, Slovene,
Swedish, Thai, Turkish, Vietnamese, Korean
Warranty 24 months
14
15
16
Version 1.0 – 01/2024 179
1 12 Technical Data – 12.2 Spectroquant® Prove 300 plus
Measuring modes Concentration, absorbance, transmission, multi wavelengths, spectra and kinetics in
absorbance and transmission mode
5 Spectral bandwidth 4 nm
10 Live ID barcode Automatic 2-D barcode reading system for all Spectroquant® cell and reagent tests
Barcode contains lot, expiry, and calibration data. Data stored with each measurement
Cell size 16 mm round cells, 10, 20 and 50 mm rectangular cells with automatic recognition
13
14
15
16
180 Version 1.0 – 01/2024
12 Technical Data – 12.2 Spectroquant® Prove 300 plus 1
3
Spectroquant® Prove 300 plus
Ambient light protection Measurement with open shaft possible due to proprietary solution (patent pending) 4
AQA prime Individual settings for all methods in
AQA1 mode: instrument check using PhotoCheck and/or Certipur® standards
AQA2 mode: system check using CombiCheck or standard solutions
5
Monitoring functions Instrument-supported pipette check and sample matrix check
Software and method Free updates on our website (www.sigmaaldrich.com/photometer-service) via internet
update and USB stick
6
Communication interfaces USB: 2 × USB-A (for printer, USB memory devices, keyboard or bar code reader),
1 × USB-mini-B
Ethernet: LAN connection
7
Data storage 7,000 single measured values from the measuring modes concentration, absorbance/%
transmission and multi wavelengths. 500 measurement result records of spectra,
kinetics, AQA1, and AQA2 methods each
Languages English, German, Spanish, French, Italian, Brazilian-Portuguese, Chinese (simplified and 8
traditional), Japanese, Russian, Bulgarian, Czech, Danish, Dutch, Greek, Hungarian,
Indonesian, Malay, Macedonian, Norwegian, Polish, Romanian, Serbian, Slovene,
Swedish, Thai, Turkish, Vietnamese, Korean
14
15
16
Version 1.0 – 01/2024 181
1 12 Technical Data – 12.3 Spectroquant® Prove 600 plus
Measuring modes Concentration, absorbance, transmission, multi wavelengths, spectra and kinetics in
absorbance and transmission mode
5 Spectral bandwidth 1.8 nm
Toluene/hexane ratio > 1.4 – the correlation of spectral bandwidth to resolution for a toluene in hexane
solution standard measured at ambient temperature 25 °C
Wavelength accuracy ± 1 nm
7 Stray light ≤ 0.1% transmission at 340 nm; ≤ 1% transmission at 198 nm
Live ID barcode Automatic 2-D barcode reading system for all Spectroquant® cell and reagent tests
Barcode contains lot, expiry, and calibration data. Data stored with each measurement
11 Cell size 16 mm round cells, 10, 20, 50 and 100 mm rectangular cells with automatic recognition
13 Methods Programmed methods of all Spectroquant® cell and reagent tests, additional user-defined
methods:
99 concentration mode, 20 kinetic mode, 20 wavelength scans
14
15
16
182 Version 1.0 – 01/2024
12 Technical Data – 12.3 Spectroquant® Prove 600 plus 1
3
Spectroquant® Prove 600 plus
Ambient light protection Measurement with open shaft possible due to proprietary solution (patent pending) 4
AQA prime Individual settings for all methods in
AQA1 mode: instrument check using PhotoCheck and/or Certipur® standards
AQA2 mode: system check using CombiCheck or standard solutions
5
Monitoring functions Instrument-supported pipette check and sample matrix check
Software and method Free updates on our website (www.sigmaaldrich.com/photometer-service) via internet
update and USB stick
6
Communication interfaces USB: 2 × USB-A (for printer, USB memory devices, keyboard or bar code reader),
1 × USB-mini-B
Ethernet: LAN connection
7
Data storage 7,000 single measured values from the measuring modes concentration, absorbance/%
transmission and multi wavelengths. 500 measurement result records of spectra,
kinetics, AQA1, and AQA2 methods each
Languages English, German, Spanish, French, Italian, Brazilian-Portuguese, Chinese (simplified and 8
traditional), Japanese, Russian, Bulgarian, Czech, Danish, Dutch, Greek, Hungarian,
Indonesian, Malay, Macedonian, Norwegian, Polish, Romanian, Serbian, Slovene,
Swedish, Thai, Turkish, Vietnamese, Korean
14
15
16
Version 1.0 – 01/2024 183
1
13 Accessories and Test Media
13.1 Accessories
3
Description Order No.
Case for Spectroquant spectrophotometer Prove 100 plus | 300 plus | 600 plus
®
1.73020.0001
4
Rectangular cells 10 mm (1 pack = 2 pcs) 1.14946.0001
10
11
12
13
14
15
16
184 Version 1.0 – 01/2024
13 Accessories and Test Media – 13.3 Test Media 1
Test media for system check Spectroquant® CombiCheck, standard solutions, and Certipur® standard
(AQA2) and MatrixCheck (AQA3) solutions are listed in the catalog "Water, Food and Enviromental Analytics" and on
the Internet under www.sigmaaldrich.com
8
Test equipment for pipet volume Spectroquant® PipeCheck 1.14962.0001
10
11
12
13
14
15
16
Version 1.0 – 01/2024 185
1
14 Appendix
Absorbance Logarithmic dimension for the absorption of the sample; negative decadal logarithm of the
transmission.
3
Analysis instructions The exact workflow for carrying out the detection procedure is described in the analysis
instructions.
AQA2 2nd step of analytical quality assurance: monitoring of the total system.
AQA2 labeling In the documentation, measured values are given an AQA2 labeling if the measurement was
5 carried out with AQA2.
AutoSelector Plastic cylinder with barcode. It transmits the code for a reagent test kit to the spectropho-
tometer. To insert it into the spectrophotometer open the lid and place it into the round cell
compartment.
6 Barcode 2-D barcode containing information on method number, expiry and lot number. If needed, it
also contains data for a calibration update. Barcode is read by the inbuilt barcode reader.
Baseline Reference value for the spectrum of reference absorbances or reference transmissions.
7 Cell Vessel to take a liquid sample for measurement in a spectrophotometer. The cell material
(mostly glass) must have certain optical features to be suitable for photometry.
Citation forms Different forms of representing a measured concentration value that can be derived from
each other. The method for the determination of phosphate, for example, provides a mea-
8 sured value for phosphorus P. This measured value can alternatively be given in the citation
forms PO4, PO4-P or P2O5.
CombiCheck Multiparameter standards used to check the total system for a method and for the Matrix-
Check.
9 Concentration Mass or amount of a dissolved substance per volume, e. g. in g/l or mol/l.
Detection procedure The detection procedure designates the general principle of how a sample is brought into
a form suitable for measurement. Different methods can be based on the same detection
procedure.
10
Kinetics Measurement over a period of time.
Log files These contain the automatically recorded log of all or specific actions of processes in the
device.
11
MatrixCheck Check on whether the photometric determination is disturbed by other sample ingredients
(sample matrix). The MatrixCheck can be carried out by spiking or diluting.
Measuring solution Name for the sample ready to be measured. A measurement sample is obtained from the
12 analytical sample (primary sample), usually through work-up. Measuring solution and analytical
sample are identical when there has been no work-up.
Measured value The measured value is the special value of a measured parameter to be determined. It is ex-
pressed as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
13 Method A method comprises a chemical detection procedure and special method data (calibration
line) that is required to evaluate the measurement results. How to carry out the method up
to measuring with the spectrophotometer is described in the analysis instructions. All
Spectroquant® Prove plus spectrophotometers contain a database with methods. Further-
more, user-defined methods can be entered in the database as well.
14
PhotoCheck standard Stable color solution with defined absorbance values for the check of AQA1 on the spectro-
photometer.
15
16
186 Version 1.0 – 01/2024
14 Appendix 1
Reagent blank The evaluation of the photometric measurement always refers to the comparison value of a
test solution without the substance to be determined (reagent blank value). Thus the influ-
ence of the basic absorbance of the reagents on photometric measurement is compensated 3
for. For all measurements with Spectroquant® test kits (concentration mode) there is an
exactly determined reagent blank value stored in the spectrophotometer. This value can,
however, be overwritten by a reagent blank value measured by yourself. If needed the 2-D
barcode on the cell tests and the AutoSelector may also contain an updated reagent blank
4
which overwrites the pre-programmed reagent blank in the instrument.
Recovery Recovery is the found measured value divided by the default value (percentage).
Example: Default value 20 mg/l; Found 19.7 mg/l => recovery 98.5%.
Recovery in MatrixCheck Recovery in MatrixCheck means recovery of the addition. Example of calculation: value with- 5
out addition = 100; addition of 20 = 120 theoretical value. Measured value = 115, only 15
from the addition of 20 were found, recovery = (115-100)/(120-100) = 75%
Sample blank The sample blank value is a characteristic of the sample (coloration) to be currently deter-
mined. 6
The blank is diluted depending on the method being used, but does not contain any color
reagents. The pH is the same as that of the test sample.
Test kit (test) A test kit contains all reagents that are required for the photometric determination of the
sample according to the analysis instructions.
10
11
12
13
14
15
16
Version 1.0 – 01/2024 187
1
15 List of Smart Icons on Display
3 Method list
List of all methods, irrespective of mode
Settings
4 This button is used to activate method-specific settings
(e. g. sample dilution, turbidity correction, zero adjustment, sample blank, reagent blank)
Ad hoc
For performing measurements (absorbance/transmission, spectrum, kinetics)
5 Allows measurements to be performed without the need to create methods
AQA
Overview and list of all Analytical Quality Assurance (AQA) modes
6 Results list
List of all stored results
Login/logout
Check on users in and out
8
Timer list
List of stopwatch functions
9
Info Buttons Description
Methods information
10
Main menu selection button – switches between 2 main menu overviews
13 Concentration
Method list: create methods -> Concentration Mode
Result list: filter Ad Hoc ABS/Trans measurements
Spectrum Mode
14 Ad hoc submenu: record spectrum
Method list: create methods -> Spectrum Mode
Result list: filter spectrum mode
15
16
188 Version 1.0 – 01/2024
15 List of Smart Icons on Display 1
Kinetic Mode 3
Ad hoc submenu: perform kinetic measurement
Method list: create methods -> Kinetic Mode
Result list: filter kinetic mode
AQA1
AQA submenu: List of AQA1 methods 5
AQA2
AQA submenu: List of AQA2 methods
6
Pipette check
AQA submenu: List of pipette checking methods
Information 7
System submenu displays the following information about the device:
Software/method versions, device class, lamp counter and serial number
Interface
System submenu displays the following settings options – and standard settings: 8
Audible signals – ON, Backlight – 100%, Print to pdf – ON
Region
System submenu displays the following settings options – and standard settings:
Language, date, time and country zone EU/US, decimal separator – "." (dot) 9
Quality
System submenu displays the following settings options – and standard settings:
Quick zero – OFF, AQA1 and AQA2 lock – OFF, Zero Adjustment expiry – ON (interval:
7 days), 10
Use expired reagents – OFF, Service reminder – ON
Automation
System submenu displays the following settings options – and standard settings:
Energy saving mode – ON (10 minutes), Auto Power off – OFF, Auto log off – OFF, 11
Auto store – ON, Auto print – OFF, Sample ID popup – OFF
User management
System submenu displays the following settings options – and standard settings:
Activation of user management and administrator settings, User login required – OFF 12
Service
System submenu displays the following settings options:
Various service functions such as backup, restore, export of log or system data and
import of methods 13
Update
System submenu displays the option for performing software and method updates
14
15
16
Version 1.0 – 01/2024 189
1 15 List of Smart Icons on Display
3 Network
This submenu of the “Instrument settings” menu displays the setting options for
connecting the Prove plus device with a network
Prove Connect
4 This submenu of the “Instrument settings” menu displays the setting options for
connecting the Prove plus device with Prove Connect
Start zero
6 Start zero adjustment for a method
Apply
7
Save
8
Stop
9 Close
Logout
10 User logout
Search method
11
Reset resp. clear filter options
12 Edit
For editing parameters
Create method
13
Duplicate/copy
The selected method is duplicated/copied
14
Print
15
16
190 Version 1.0 – 01/2024
15 List of Smart Icons on Display 1
Import button 4
Updates/methods are imported from an external memory device into the instrument
Delete
The selected items are deleted 5
Dilution
6
Activate and notify predilution
Turbidity on
Activate and notify the turbidity correction
7
Show absorbance
Activate and notify display of absorbance value in the result screen
8
Zero adjustment
Perform zero adjustment
9
Sample blank value
Activate and notify Sample blank value
Reagent blank 10
Activate and notify user-defined Reagent blank
Recalibrate
Activate and notify user-defined recalibration 11
MatrixCheck
Activate MatrixCheck
12
User defined measurement range
Activate user-defined lower and upper limit of the measuring range
13
14
15
16
Version 1.0 – 01/2024 191
1 15 List of Smart Icons on Display
OFF/ON Button
0 = Off, I = On – the part displayed in light grey is active – here: 0 = OFF
3
Date/measurement
Switch between date or measurement interval (AQA2); active here: measurement interval
4 Absorbance/transmission
Switch between absorbance or transmission mode; active here: transmission mode
Spiking/dilution
5 Switch between spiking and dilution (MatrixCheck); active here: dilution
Back
7
Close
Clear
8
Delete
9
Apply
10
Add
Warning
Warnig symbol check info box
12
Barcode scanner deactivated
The barcode scanner for reading out the Live ID barcode on round cells and AutoSelectors
has been deactivated
13
Locked
Change password
14 Choosen
Check mark
15
16
192 Version 1.0 – 01/2024
15 List of Smart Icons on Display 1
Search list
Search function, search criterion: method number, method name or article number (first 6
3
digits)
4
Sample ID
Search/results list. Search function, search criterion: sample ID
5
Select all/select none
Panorama view 6
Graphic representation of measurement series (control card, control chart, for trend
analyses)
8
Set formula for method calibration
9
Show graphic view
11
Return to the recorded spectrum
Zoom out
13
Zoom in
14
View of peak max of a spectrum
15
16
Version 1.0 – 01/2024 193
1 15 List of Smart Icons on Display
Overlay of spectra
6
First-order derivative of a spectrum
9 Time
Time Stamp
Passed
10 Status of a check; ü = passed
Off
Status of a check; = inactive
11
Failed
Status of a check; - = failed
12 Expired
Status of a check; ! = overdue
Progress
13 The instrument is in progress
Progress
The instrument is in progress
14
15
16
194 Version 1.0 – 01/2024
1
16 Contents of log files
The information documented in the user log file are coded in the following way:
7
Group Action code Additional information
0 = General 1 = Power On -
2 = Power Off (by User) - 8
3 = Auto Power Off -
10 = Self-Test passed -
11 = Self-Test failed 1 = System-, 2 = Lamp-, 3 = WLCheck
20 = Login -
9
21 = Logoff (by User) -
16
Version 1.0 – 01/2024 195
1 16 Contents of log files
5 10 = Backup
11 = Restore
Explanation of an example:
10 201208_1001;pm;3;20;9002;0;0;0
Timestamp = 201208_1001
UserId = pm
11 Actiongroup = 3 = General AQA
Actioncode = 20 = Result deleted
Info1 = 9002 = Method No
Info2 = 0 = no further details
12 Info3 = 0 = no further details
Info4 = 0 = no further details
14
15
16
196 Version 1.0 – 01/2024
We provide information and advice to our customers on application technologies and regulatory matters to the best of our knowledge and ability,
but without obligation or liability. Existing laws and regulations are to be observed in all cases by our customers. This also applies in respect to any
rights of third parties. Our information and advice do not relieve our customers of their own responsibility for checking the suitability of our pro-
ducts for the envisaged purpose.
The life science business of Merck KGaA, Darmstadt, Germany operates as MilliporeSigma in the U.S. and Canada.
Merck Life Science KGaA, 64271 Darmstadt, Germany, Tel. +49(0)6151 72-2440
www.sigmaaldrich.com
© 2023 Merck KGaA, Darmstadt, Germany and/or its affiliates. All Rights Reserved.
Merck, Supelco, Certipur, MQuant, and Spectroquant are trademarks of Merck
KGaA, Darmstadt, Germany or its affiliates. All other trademarks are the property of
their respective owners. Detailed information on trademarks is available via publicly
accessible resources.