FLUXUS G601 Technical specification
Transmitter
Technical data
FLUXUS G601** FLUXUS G601ST (steam measurement)
design portable
measurement
test measurement to validate the application required in advan-
ce
measurement transit time difference correlation principle
principle
flow velocity m/s 0.01...35, depending on pipe diameter depending on pipe diameter and transducer, see diagrams
repeatability 0.15 % of reading ±0.005 m/s
fluid all acoustically conductive gases, saturated steam, superheated steam
e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene,
propane
fluid pressure see transducers 3...10
fluid temperature °C see transducers 135...180
temperature com- corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
pensation
measurement uncertainty (volumetric flow rate)
measurement uncer- ±0.3 % of reading ±0.005 m/s ±0.3 % of reading ±0.005 m/s
tainty of measuring
system1
measurement uncer- ±1...3 % of reading ±0.005 m/s, depending on application ±1...3 % of reading ±0.005 m/s, depending on application
tainty at the measu-
ring point
transmitter
power supply • 100...230 V/50...60 Hz (power supply unit: IP40, 0...40 °C)
• 10.5...15 V DC (socket at transmitter)
• integrated battery
integrated battery Li-Ion, 7.2 V/6.2 Ah
• operating time h • > 14 h (without outputs, inputs and backlight)3
• > 25 h (1 measuring channel, ambient temperature > 10 °C, without outputs, inputs and backlight)3
power consumption W < 6 (with outputs, inputs and backlight), charging: 18
number of measuring 2
channels
damping s 0...100 (adjustable)
measuring cycle Hz 100...1000 (1 channel)
response time s 1 (1 channel), option: 0.07
housing material PA, TPE, AutoTex, stainless steel
degree of protection IP65
dimensions mm see dimensional drawing
weight kg 2.1
fixation QuickFix pipe mounting fixture
ambient temperature °C -10...+60
display 2 x 16 characters, dot matrix, backlight
menu language English, German, French, Dutch, Spanish
measuring functions
physical quantities operating volumetric flow rate, standard volumetric flow rate, operating volumetric flow rate, mass flow rate, flow velocity
mass flow rate, flow velocity
totalizer volume, mass
calculation functions average, difference, sum
diagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times
communication interfaces
service interfaces • RS232
• USB (with adapter)
process interfaces • Modbus RTU (optional)
accessories
serial data kit
• сable RS232
• adapter RS232 - USB
software • FluxDiagReader: download of measured values and parameters, graphical presentation
• FluxDiag (optional): download of measurement data, graphical presentation, report generation
• FluxSubstanceLoader: upload of fluid data sets
adapter AO5, AO6, AO7, AO8, AI1, AI2
transport case dimensions: 500 x 400 x 190 mm
1
with aperture calibration of the transducers
2 for transit time difference principle and reference conditions
3
operating time extension using the power pack PP0026NN (optional, order code: ACC-PO-#601-/B6)
For the technical data in the flow measurement of liquids mode see Technical specification TSFLUXUS_F601V*-*.
6 2019-11-01, TSFLUXUS_G601V2-4EN_Leu