Omd 2008 Allversions e 20151013 240425 075243
Omd 2008 Allversions e 20151013 240425 075243
IMPORTANT NOTICE
Replacement components for 15ppm Bilge Alarms.
General
All monitors in our range are inspected and tested to the related I.M.O.
requirements at our factories prior to delivery. In normal use the units should
operate correctly and without fault over a long period of time requiring only small
amounts of maintenance to be carried out as outlined in the instruction manuals.
It is very important to correctly identify the Type and Version of the instrument for
technical support, or for ordering spare parts. Please include instrument type,
Version, and serial number, and if possible instrument date of commissioning into
any communication.
This procedure is by far the easiest and most cost effective way of ensuring the
monitor on board conforms to I.M.O. resolution MEPC.107(49).
Remark:
If for some reasons the computer unit needs to be changed, it has to be made
sure, that the Memory Card will remain on board for at least 18 month. The new
computer unit will carry its own memory card. The old card can be insert into the
new unit only for reading. Writing is only possible with the card delivered with the
new computer unit. For details see section 13.1.
Our warranty terms are12 months after installation but maximum 18 months after
delivery ex works. The maker undertakes to remedy any defect resulting from
faulty materials of workmanship except wearing parts.
The maker's obligation is limited to the repairs or replacement of such defective
parts by his own plant or one of his authorized service stations.
The purchaser shall bear the cost and risk of transport of defective parts and
repaired parts supplied in replacement of such defective parts.
Contents
1 Introduction ............................................................................................................. 4
1.1 General .....................................................................................................................4
1.2 Instrument Versions and Options ..............................................................................4
2 Important Notes ...................................................................................................... 6
3 Principle of operation ............................................................................................. 7
3.1 Measuring Principle ...................................................................................................7
3.2 Adjustment ................................................................................................................7
3.3 Displays and Alarms .................................................................................................7
3.4 Electronic record .......................................................................................................8
4 Specifications ......................................................................................................... 9
5 Construction ......................................................................................................... 11
5.1 Overview .................................................................................................................11
5.2 Computer Unit .........................................................................................................11
5.3 Measuring Cell ........................................................................................................ 11
5.4 Other devices .......................................................................................................... 12
6 Installation ............................................................................................................. 14
7 Piping ..................................................................................................................... 16
8 Wiring..................................................................................................................... 18
8.1 Typical Control System ...........................................................................................20
8.2 Remote EV Electric Valve control ............................................................................ 20
9 Power Supply ........................................................................................................ 20
10 Commissioning.................................................................................................. 21
10.1 Electrical and Piping ................................................................................................ 21
10.2 Flow rate adjustment ...............................................................................................21
10.3 Functional Tests ...................................................................................................... 22
10.4 Programming Mode .................................................................................................24
11 Operating instructions ...................................................................................... 29
11.1 Operator Notes........................................................................................................ 29
12 Operator Maintenance ....................................................................................... 30
12.1 Manual Cell Clean Unit (MCU) ................................................................................ 31
13 Fault finding ....................................................................................................... 34
13.1 Memory Card .......................................................................................................... 37
14 Calibration .......................................................................................................... 37
14.1 Calibration and repeatability check .......................................................................... 37
14.2 Function Test at Classification Survey and Port State Control .................................38
15 Spare Parts......................................................................................................... 39
15.1 Recommended On Board Spares ...........................................................................39
16 REMARKS .......................................................................................................... 40
1 Introduction
1.1 General
The OMD-2008 Bilge Alarm Unit has been designed specifically for use in
conjunction with 15 ppm oil-water separator units and has a specification and
performance which exceeds the requirements of the International Maritime
Organization specifications for 15ppm Bilge Alarms contained in Resolution
MEPC. 107 (49).
The unit is supplied with 2 works-adjusted alarms at 15 ppm. Other set points (10
ppm or 5 ppm) are possible and can be adjusted on site at any time by using the
buttons at the front panel.
If an alarm set point is exceed, the alarms are visible at the front panel and the
appropriate relays are switched. In case of malfunction the System LED at the
front panel will change from blinking green to permanent red.
For the data logging function the unit requires a status input from the separator.
It is possible to have a Flow Switch (similar to the Flow Sensor, but with a simple
ON/OFF output) connected to all versions of OMD-2008.
2 Important Notes
a) This equipment must be installed and operated in strict accordance with
the instructions contained in this manual. Failure to do so will impair the
protection provided.
d) The unit must be isolated from the electrical supply before any
maintenance of the equipment is attempted.
3 Principle of operation
If an alarm (works set point 15 ppm, or 10 ppm or 5 ppm respectively) occurs, the
two oil alarm relays are activated after the adjusted time delay.
3.2 Adjustment
The unit is delivered with a works calibration according the IMO-requirements.
The alarm points are set to 15 ppm (or 10 ppm, or 5 ppm for certain versions).
The "Zero" point is also works calibrated and can be re-adjusted on site by using
the programming mode and clean water. See Section 10.4 “Settings-Offset”.
Calibration of the instrument is not permitted. According to IMO Regulations a
calibration check has to be done by the manufacturer or persons authorized by
the manufacturer. The recommended way to ensure a properly calibrated
Measuring Cell on board is to exchange the Measuring Cell.
Both alarm circuits are also related to an alarm LED on the front panel.
Additional to the alarm LEDs each alarm circuit is equipped with a relay with
potential free switchover alarm contacts. These contacts can be used for external
processing of the signal or for control of further functions.
If a malfunction or failure of the power supply occurs, all three relays (both alarm
relays and the SYSTEMFAULT relay) will switch to alarm condition.
4 Specifications
Range: 0 – 30 ppm, Trend indication 50ppm
Accuracy According IMO MEPC. 107(49)
Linearity Up to 30 ppm better than ± 2 %
Display Yellow Graphic Display
Power Supply: 24 V - 240V AC or DC
Automatic Voltage selection
Consumption: < 10 VA
Alarm Points 1 + 2: Adjustable between 1 - 15 ppm
(Works adjustment 15 ppm)
Alarm 1 Operating Delay: Adjustable between 1 – 540 sec.
(for annunciation purpose) (Works adjustment 2 sec)
Alarm 2 Operating Delay: Adjustable between 1 – 10 sec.
(for control purposes) (Works adjustment 10 sec)
System Fault Alarm: Red LED
Alarm Contact Rating: Potential free 1 pole change over
contacts, 3 A / 240 V
Alarm Indication: Red LEDs
Output Signal: 0 – 20 mA or 4 – 20 selectable
ext. Load < 150
Sample Water Pressure: 0,1 – 10 bar
Sample Flow: Approx. 0,1 - 3 l/min depending on
pressure
Ambient Temperature: + 1 to + 55° C
Sample Water + 1 to + 65° C
Temperature:
Roll: Up to 45°
Degree of Protection: IP 65
Pipe Connections: for 6mm OD Cu or SS pipe
OMD-2008 EV
Sample pressure max. 6 bar
Clean water pressure max. 6 bar
Size (Computer Unit) 200 mm W x 200 mm H x 100 mm D
Size (EV-FC-Arrangement) 150 mm W x 290 mm H x 140 mm D
Weight: 5,9 kg
OMD-2008 EVFC
Sample pressure max. 6 bar
Clean water pressure max. 6 bar
Size (Computer Unit) 200 mm W x 200 mm H x 100 mm D
Size (EV-FC-Arrangement) 150 mm W x 290 mm H x 140 mm D
Weight: 5,9 kg
OMD-2008 EVFCMCU
Sample pressure max. 6 bar
Clean water pressure max. 6 bar
Size (Computer Unit) 200 mm W x 200 mm H x 100 mm D
Size (EV-FC-Arrangement) 150 mm W x 290 mm H x 140 mm D
Weight: 5,9 kg
5 Construction
5.1 Overview
OMD-2008 instruments consist of two main components:
The Computer Unit contains the display PCB with the data logger and the
terminals for external connections.
The Measuring Cell assembly consists of the Measuring Cell, a valve to select
sample water stream, or clean water stream, and, depending on the version of
the instrument, additional devices like a flow sensor.
Fig. 2a:
Dimensions of Computer Unit
The Computer Unit contains Display, Power Supply, all external electrical
connection terminals, and Memory Card for electronic record.
The Computer Unit is enclosed in a sheet metal housing, with several openings
for metric cable glands underneath. The Computer Unit should be mounted so
that the display and touch button area is easily accessible.
The Measuring Cell has all parameters and calibration data permanently stored,
and the calibration certificate is tied to the Measuring Cell. The recommended
way to renew a calibration is to exchange the Measuring Cell. Under no
conditions should the Measuring Cell be opened or dismantled, doing so would
void both warranty and calibration.
5.4 FW valve
The OMD-2008 uses a switchover valve to control the usage of clean water for
cleaning, or zeroing purposes. Whenever the FW valve is switched to Clean
water position, and not to sample stream position (normal operating position), the
instrument will be in alarm condition.
Fig. 1
6 Installation
See Section 2 for important notes concerning installation.
The OMD-2008 should be located as close as possible to the oily water separator
to minimize response delays. According MEPC.107(49) the layout of the
installation should be arranged so that the overall response time (including the
response time of the 15 ppm Bilge Alarm, which is less than 5 seconds) between
an effluent discharge from the 15 ppm Bilge Separator exceeding 15 ppm, and
the operation of the Automatic Stopping Device preventing overboard discharge,
should be as short as possible and in any case not more than 20 seconds.
Computer Unit and Measuring Cell assembly should be mounted close to each
other.
Pipe connections are prepared for 6mm OD stainless steel or coper pipes.
Adapter fittings for other pipe diameters are available upon request. Working
space for maintenance and operation is recommended in front of and above the
instrument.
Care must be taken at mounting of the pipes connections to avoid any torsion of
the housing and damage of the instrument.
7 Piping
The instrument has to be supplied with a representative sample of the separator
effluent water stream. The sample stream through the instrument is
recommended to be approx. 2 Liters per Minute.
Connect the OMD-2008 EV-FC Monitor to the sample point of the oily-water
separator outlet and to a source of oil free water. Fittings on the Instrument are
for 6 mm OD copper or stainless steel pipe. Adapter fittings for other pipe
diameters are available upon request.
The sample point should be located on a vertical section of the separator outflow
piping to minimize the effects of any entrained air. The tapping point should be at
a level above the outlet of the monitor to ensure the sample cell is flooded at all
times.
Fig. 3
8 Wiring
See Section 2 for important notes concerning wiring.
This unit must be connected to the mains supply via a suitable rated and
approved fused isolator unless such fusing / isolation is provided by associated
equipment. When fitted, the isolator should be close, readily accessible and
marked as to function.
Electrical connections are made through the metric cable gland openings
prepared underneath the instrument. Not all openings may be free.
Fig. 4
Precise wiring details will vary dependent upon the control system to be
employed but the most frequently used systems employ alarm relay 1 for alarm
only and alarm relay 2 for control purposes.
Electrical connections are made to the terminal blocks inside the computer
housing. Wires are connected to the terminals by pushing a suitable screwdriver
into the clamp holes to release the internal spring loaded clamps. After the wire is
inserted to the terminal and the screwdriver is removed, the wire is fixed.
The instrument provides a pilot voltage output at Terminals 3&4. This is internally
connected to the power supply input (Terminals 1&2 via Fuse T2A). The pilot
Please note: any device connected to the pilot voltage output must be rated for
the voltage the instrument is supplied with. Do not use the pilot voltage for driving
motors, heaters or other high load devices. The pilot voltage is intended for alarm
purposes only.
The separation process will continue until such time as the pollution level falls
below the alarm set point at which time the discharge will be directed overboard.
9 Power Supply
See Section 2 for important notes.
10 Commissioning
See Section 2 for important notes.
On completion of the installation, wiring, and piping carry out the following
checks:
c) Check that the grounding (PE) connections have been made according to
the relevant regulations.
b) On OMD-2008 EV-FC instruments adjust the flow rate through the unit by
O-Ring 4.5x2 inside O-Ring 8.5x2 outside Main O-Ring 11.5x3 always present
using the small O-Rings in the cell cap.
The flow rate adjustment is important for the time delay between the sample point
and the monitor. If the sample flow rate is too small, the instrument will be in
Alarm and display “Status: Flow”.
MCU
(Manual Cleaning Unit)
b) Verify that the water is flowing with the proper flow rate trough the
instrument.
c) Switch on the instrument and make sure, that the Power LED is
illuminated and
the display is
showing the
initializing display
for about 15 sec.
After that time it
will change to the standard display, showing the actual measurement.
d) Switch the Clean Water Valve to clean water. While oil free water is
running through the monitor check the Zero adjustment. The display
should be "0" to “2” and the status will show “FW”. If the display varies by
greater amounts, it may be that air entrainment is present. If this is the
case, the cause must be located and rectified.
OMD-2008
The clock is factory set for GMT,
Greenwich Mean Time, and cannot be
changed.
AL1 AL2
SYS
Exit from
Pressing the SYS Select if you want
SYSTEM-info
button directly leads information about the
menu by pressing
into SYSTEM menu. instrument or
the ESC button.
information about the
measuring cell.
Exit from
MEASURING
CELL menu by
pressing the ESC
ON
TEST
Exit from
information
Desiccator status
LOG
11 Operating instructions
The OMD-2008 has a manual clean water switchover valve. Whenever ths valve
is not in normal operation position, the instrument will be in alarm condition.
The OMD-2008 EV has an electric switchover valve to switch over from sample
water to clean water. Whenever clean water is used, the instrument will be in
Alarm condition.
The OMD-2008 EV-FC has an additional Flow Sensor to detect the flow rate of
the water through the instrument. If the sample does not flow, the instrument will
be in alarm condition.
The OMD-2008 EV-FC MCU has the Flow Sensor, and additionally it has the
Manual Cleaning Unit.
c) Switch the Clean Water Valve to clean water stream and let oil free water
flow through the system for a few minutes. Check that the display shows 0
to 2 ppm. If not, clean Measuring Cell sample Glass Tube properly.
12 Operator Maintenance
d) Unscrew and remove the cell cap. For instruments equipped with Manual
Cleaning Unit (MCU), unscrew the MCU instead.
e) Insert a suitable Cell Cleaning brush (Art. No. 77555) into the cell and
clean it with upwards and downwards motion through the entire length of
the cell several times.
f) Remove the Cell Cleaning brush and replace the cell cap.
g) Open clean water valve and allow oil free water to flow through the
instrument for a few minutes.
h) Observe that the display is showing "0" to “2”. If not, clean again.
The Manual Cleaning Unit (MCU) facilitates cleaning of the cell without the need
of removing the cell cap. Regular use of this device should prevent malfunction of
the monitor due to common fouling mechanisms of the sample tube, and may
allow longer maintenance intervals. The MCU (Order No. 77780) can be installed
at any time.
Vergleichsbild: normale
Kopfschraube
Operating Instructions
b) Activate the manual cell clean unit by pressing the handle several times.
In some cases of high reading with clean water the measuring cell has a problem
with internal coating of the glass tube. Just cleaning with brush and clean water
will not help in this case.
Please carry out the following instructions to make sure, that the glass tube is
really clean. Then the unit will show 0 to 2 ppm with clean water.
Clean the glass tube by using the cleaning brush under assistance from some
cleaner.
Iron Oxide can be converted into a water soluble compound using Citric Acid.
Citric Acid is commonly used to remove calcite from Coffee Machines. Fill some
Citric Acid into the drained Sample Glass Tube, add water, and let it soak
overnight (the reaction takes some time). Flush with clean water.
Additional use of tooth paste may also assist in cleaning as a last resort. Do not
use abrasive cleaners containing hard particles. Hard abrasive particles may
scratch the glass surface, permanently damaging the instrument.
13 Fault finding
See Section 2 for important notes.
The OMD-2008 EV-FC will indicate several malfunctions in the status line of the
display. Pressing the “OK” button will lead into an information window, similar to
the items listed in the table below.
S R Syst Alarm- Reason Servicing
t ea em- circuit
a di Alar 1,2
t ng m-
u LED
s
O 0.. Gree Norma Normal -
K 49 n/ l operation
Blinki operati
ng on
O E Gree Alarm Sample Wait until oil
K E n/ reading is content is
Blinki out of within the
ng range: range, clean
Oil content sample tube
too high,
dirty
sample
tube
14 Calibration
15 ppm Bilge Alarms built according MEPC.107(49) have to be protected against
access beyond the checks of instrument drift, repeatability of the instrument
reading and zero adjustment. For this reason the instrument is electronically
sealed, so that only the manufacturer or his authorized persons, equipped with
the related tools, are able to get access for changing the calibration.
To provide a simple procedure to check the instrument aboard ship, the OMD-
2008 EV-FC is constructed in that way, that the zero check also confirms the
instrument drift within the specifications.
d) If it is sure, that non aerated, clean water is in the instrument, the reading
should be 0 ppm ± 2 ppm.
To provide a simple procedure for check the instrument aboard ship, the OMD-
2008 EV-FC is constructed in that way, that the zero check also confirms the
instrument drift within the specifications. The Test button starts a self test routine
and allows to put both alarms contacts into alarm condition. The instrument will
count down from an assumed high reading (30ppm) downwards until the
assumed value is equal to the actual measured ppm value. Note that this test will
only switch the alarm contacts to non-alarm condition, if the sample contains less
than 15ppm oil content and all other conditions for proper measurement are OK.
15 Spare Parts
When ordering spares, it is important to supply details of the type of monitor, part
number of each spare required, its description and any relevant serial number.
DESCRIPTION ART-NUMBER
Service Kit 18570
Desiccator 79550
Cell Cleaning Brush 77555
O-Ring Set 77775
Memory Card 79186
Optional Items
Manual Cell Clean Unit 77780
Wiper Seal for Cell Clean Unit 77606
OMD-CR Memory Card Reader 75185
Function Test Kit 18510
Flow Sensor 16700
DESCRIPTION ART-NUMBER
1 off Service Kit 18570
16 REMARKS
All the modifications and deviations from the standard form, which have to be
carried out in the supply, should be attached at this paragraph.
Remarks: