z50 Mindray
z50 Mindray
Operator’s Manual
[Basic Volume]
Contents
Intellectual Property Statement ...................................................................................................... I
Responsibility on the Manufacturer Party....................................................................................... I
Warranty ...................................................................................................................................... II
Exemptions ............................................................................................................................. II
Customer Service Department ................................................................................................ II
Important Information .................................................................................................................. III
About This Manual ...................................................................................................................... III
Notation Conventions .................................................................................................................. III
Operator’s Manuals..................................................................................................................... IV
Manuals on Paper ....................................................................................................................... IV
Software Interfaces in this Manual ................................................................................................ V
Conventions ................................................................................................................................. V
Product Differences...................................................................................................................... V
1 Safety Precautions.................................................................................................... 1-1
1.1 Safety Classification ........................................................................................................ 1-1
1.2 Meaning of Signal Words................................................................................................. 1-2
1.3 Meaning of Safety Symbols ............................................................................................. 1-2
1.4 Safety Precautions .......................................................................................................... 1-3
1.5 Latex Alert ..................................................................................................................... 1-10
1.6 Warning Labels .............................................................................................................. 1-11
2 System Overview ...................................................................................................... 2-1
2.1 Intended Use ................................................................................................................... 2-1
2.2 Contraindication .............................................................................................................. 2-1
2.3 Product and Model Code ................................................................................................. 2-1
2.4 Product Specifications ..................................................................................................... 2-1
2.4.1 Imaging Mode .......................................................................................................... 2-1
2.4.2 Power supply ........................................................................................................... 2-2
2.4.3 Environmental Conditions......................................................................................... 2-2
2.4.4 Size and weights ...................................................................................................... 2-2
2.5 System Configuration ...................................................................................................... 2-2
2.5.1 Standard Configuration............................................................................................. 2-2
2.5.2 Probes Available ...................................................................................................... 2-3
2.5.3 Options .................................................................................................................... 2-4
2.5.4 Peripherals Supported ............................................................................................. 2-5
2.6 Introduction of Each Unit ................................................................................................. 2-5
2.6.1 I/O Panel.................................................................................................................. 2-7
2.6.2 Power Supply Panel ................................................................................................. 2-8
2.6.3 Control Panel ........................................................................................................... 2-8
2.7 Symbols ........................................................................................................................ 2-12
3 System Preparation .................................................................................................. 3-1
3.1 Move/Position the System ............................................................................................... 3-1
3.2 Power Supply .................................................................................................................. 3-1
3.2.1 Connecting External Power Supply .......................................................................... 3-1
3.2.2 Powered by Battery .................................................................................................. 3-2
3.3 Powering ON/ OFF .......................................................................................................... 3-2
3.3.1 Powering ON ........................................................................................................... 3-2
3.3.2 Powering OFF.......................................................................................................... 3-3
i
3.4 Connecting / Disconnecting a Probe ................................................................................ 3-4
3.4.1 Connecting a Probe ................................................................................................. 3-4
3.4.2 Disconnecting a Probe ............................................................................................. 3-4
3.5 Connecting the Footswitch............................................................................................... 3-5
3.6 Connecting/ Removing a USB Storage Device ................................................................. 3-5
3.7 Graph / Text Printer ......................................................................................................... 3-5
3.8 Digital Video Printer ......................................................................................................... 3-9
3.9 Analog Video Printer........................................................................................................ 3-9
3.10 External DVD ................................................................................................................ 3-10
3.11 Basic Screen and Operation .......................................................................................... 3-10
3.11.1 Basic Screen.......................................................................................................... 3-10
3.11.2 Basic Operations of Screens .................................................................................. 3-12
4 Exam Preparation ..................................................................................................... 4-1
4.1 Start an Exam.................................................................................................................. 4-1
4.2 Patient Information .......................................................................................................... 4-1
4.2.1 New Patient Information ........................................................................................... 4-1
4.2.2 Retrieve Patient Information ..................................................................................... 4-5
4.3 Select Exam Mode and Probe ......................................................................................... 4-7
4.3.1 Selecting Exam Mode and Probe ............................................................................. 4-7
4.4 Selecting Imaging Mode .................................................................................................. 4-8
4.5 Activate& Continue an Exam ........................................................................................... 4-8
4.5.1 Activate an Exam ..................................................................................................... 4-8
4.5.2 Continue an Exam ................................................................................................... 4-8
4.6 Pause & End an Exam .................................................................................................... 4-8
4.6.1 Pause and Exam ...................................................................................................... 4-8
4.6.2 End an Exam ........................................................................................................... 4-9
5 Image Optimization ................................................................................................... 5-1
5.1 Imaging Mode ................................................................................................................. 5-1
5.2 Basic Operations ............................................................................................................. 5-1
5.3 Quickly Saving Image Setting (QSave) ............................................................................ 5-2
5.4 B Mode ........................................................................................................................... 5-2
5.4.1 B Mode Exam Protocol............................................................................................. 5-2
5.4.2 B Mode Parameters ................................................................................................. 5-2
5.4.3 B Mode Image Optimization ..................................................................................... 5-3
5.5 M Mode ........................................................................................................................... 5-7
5.5.1 M Mode Exam Protocol ............................................................................................ 5-7
5.5.2 M Mode Parameters................................................................................................. 5-8
5.5.3 M Mode Image Optimization..................................................................................... 5-8
5.6 Color Mode Image Optimization..................................................................................... 5-10
5.6.1 Color Mode Exam Protocol..................................................................................... 5-10
5.6.2 Color Mode Image Optimization ............................................................................. 5-10
5.6.3 Color Mode Image Optimization ............................................................................. 5-10
5.7 Power Mode Image Optimization ................................................................................... 5-13
5.7.1 Basic Procedures for Power Mode Imaging ............................................................ 5-13
5.7.2 Power Mode Image Parameters ............................................................................. 5-14
5.7.3 Power Mode Image Optimization............................................................................ 5-14
5.8 PW Doppler Mode ......................................................................................................... 5-15
5.8.1 Basic Procedures for PW Mode Exam .................................................................... 5-15
5.8.2 PW Mode Image Parameters ................................................................................. 5-16
5.8.3 PW Mode Image Optimization ................................................................................ 5-16
ii
5.9 Anatomical M Mode (Free Xros M) ................................................................................ 5-20
5.10 TDI ................................................................................................................................ 5-22
5.10.1 Basic Procedures for TDI Imaging .......................................................................... 5-22
5.10.2 TDI Image Parameters ........................................................................................... 5-22
5.10.3 TDI Image Optimization.......................................................................................... 5-22
5.11 iScape ........................................................................................................................... 5-23
5.11.1 Basic Procedures for iScape Imaging ..................................................................... 5-23
5.11.2 Image Acquisition ................................................................................................... 5-24
5.11.3 iScape Viewing ...................................................................................................... 5-24
5.11.4 Cine Review........................................................................................................... 5-26
5.12 3D/4D............................................................................................................................ 5-27
5.12.1 Note before Use ..................................................................................................... 5-27
5.12.2 Overview................................................................................................................ 5-28
5.12.3 Static 3D ................................................................................................................ 5-31
5.12.4 Smart 3D ............................................................................................................... 5-40
5.12.5 4D.......................................................................................................................... 5-42
5.12.6 iLive ....................................................................................................................... 5-43
5.12.7 iPage ..................................................................................................................... 5-44
5.12.8 Smart Face ............................................................................................................ 5-46
6 Display & Cine Review ............................................................................................. 6-1
6.1 Image Display ................................................................................................................. 6-1
6.1.1 Splitting Display........................................................................................................ 6-1
6.1.2 Image Magnification ................................................................................................. 6-1
6.1.3 Freeze/ Unfreeze the Image ..................................................................................... 6-2
6.2 Cine Review .................................................................................................................... 6-3
6.2.1 Entering/ Exiting Cine Review .................................................................................. 6-3
6.2.2 Cine Review in 2D Mode .......................................................................................... 6-3
6.2.3 Cine Review in M/ PW /TVD Mode ........................................................................... 6-4
6.2.4 Linked Cine Review ................................................................................................. 6-5
6.3 Image Compare............................................................................................................... 6-5
6.3.1 Image Compare in Review Mode.............................................................................. 6-5
6.3.2 Frame Compare ....................................................................................................... 6-6
6.4 Cine Memory................................................................................................................... 6-6
6.4.1 Cine Memory Setting................................................................................................ 6-6
6.4.2 Cine Memory Clear .................................................................................................. 6-7
6.5 Preset ............................................................................................................................. 6-7
7 Measurement............................................................................................................. 7-1
7.1 Basic Operations ............................................................................................................. 7-1
7.2 General Measurements ................................................................................................... 7-1
7.2.1 2D General Measurements ...................................................................................... 7-1
7.2.2 M General Measurements ........................................................................................ 7-2
7.2.3 Doppler General Measurements ............................................................................... 7-3
7.3 Application Measurement ................................................................................................ 7-3
7.4 Measurement Accuracy ................................................................................................... 7-4
8 Comments and Body Marks ..................................................................................... 8-1
8.1 Comments....................................................................................................................... 8-1
8.1.1 Comment Basic Procedures ..................................................................................... 8-1
8.1.2 Comment Menu ....................................................................................................... 8-1
8.1.3 Adding Comments.................................................................................................... 8-2
8.1.4 Moving Comments ................................................................................................... 8-3
iii
8.1.5 Editing Comments .................................................................................................... 8-3
8.1.6 Deleting Comments.................................................................................................. 8-3
8.2 Body Mark ....................................................................................................................... 8-4
8.2.1 Body Mark Operation Procedures............................................................................. 8-4
8.2.2 Menu ....................................................................................................................... 8-4
8.2.3 Adding Body Marks .................................................................................................. 8-4
8.2.4 Moving Body Marks.................................................................................................. 8-4
8.2.5 Deleting Body Marks ................................................................................................ 8-5
9 Patient Data Management......................................................................................... 9-1
9.1 Patient Information Management ..................................................................................... 9-1
9.1.1 Enter Patient Information.......................................................................................... 9-1
9.2 Image File Management .................................................................................................. 9-1
9.2.1 Storage Media.......................................................................................................... 9-1
9.2.2 Image File Formats .................................................................................................. 9-1
9.2.3 Image Storage Preset .............................................................................................. 9-2
9.2.4 Saving Images to the System ................................................................................... 9-2
9.2.5 Quickly Saving Images to USB Flash Drive .............................................................. 9-3
9.2.6 Quickly Saving Full Screen Image to the System ...................................................... 9-3
9.2.7 Thumbnails .............................................................................................................. 9-3
9.2.8 Image Review and Analysis...................................................................................... 9-4
9.2.9 iVision ...................................................................................................................... 9-6
9.2.10 Sending Image File .................................................................................................. 9-7
9.3 Report Management ........................................................................................................ 9-7
9.4 Patient Data Management (iStation) ................................................................................ 9-9
9.4.1 Searching a Patient .................................................................................................. 9-9
9.4.2 Patient Data View & Management .......................................................................... 9-10
9.5 Backing Up and Erasing Files through DVD Drive.......................................................... 9-12
9.6 Patient Task Manager .................................................................................................... 9-12
9.7 Access Control .............................................................................................................. 9-14
9.7.1 Access Setting ....................................................................................................... 9-14
9.7.2 Setting Access Control ........................................................................................... 9-14
9.7.3 System Login ......................................................................................................... 9-14
9.7.4 Add/ Delete a User ................................................................................................. 9-15
9.7.5 Modify Password .................................................................................................... 9-17
10 DICOM...................................................................................................................... 10-1
10.1 DICOM Preset ............................................................................................................... 10-1
10.1.1 Network Preset ...................................................................................................... 10-1
10.1.2 DICOM Preset ....................................................................................................... 10-2
10.1.3 DICOM Service ...................................................................................................... 10-4
10.2 Verify Connectivity ......................................................................................................... 10-9
10.3 DICOM Service ........................................................................................................... 10-10
10.3.1 DICOM Storage.................................................................................................... 10-10
10.3.2 DICOM Print ......................................................................................................... 10-11
10.3.3 DICOM Worklist .................................................................................................... 10-11
10.3.4 MPPS .................................................................................................................. 10-13
10.3.5 Storage Commitment ........................................................................................... 10-13
10.3.6 Query/Retrieve..................................................................................................... 10-14
10.4 DICOM Media Storage ................................................................................................ 10-15
10.5 Structured Report ........................................................................................................ 10-16
10.6 DICOM Task Manager ................................................................................................. 10-16
iv
11 Setup........................................................................................................................ 11-1
11.1 System Preset ................................................................................................................ 11-1
11.1.1 Region .................................................................................................................... 11-1
11.1.2 General ................................................................................................................... 11-2
11.1.3 Image Preset .......................................................................................................... 11-3
11.1.4 Application .............................................................................................................. 11-4
11.1.5 OB .......................................................................................................................... 11-5
11.1.6 Key Config .............................................................................................................. 11-5
11.1.7 Admin ..................................................................................................................... 11-6
11.2 Exam Preset................................................................................................................... 11-6
11.3 Measure Preset .............................................................................................................. 11-7
11.4 Comment Preset ............................................................................................................ 11-7
11.5 Bodymark Preset ............................................................................................................ 11-8
11.6 Print Preset .................................................................................................................... 11-9
11.7 Network Preset ............................................................................................................. 11-10
11.8 Maintenance ................................................................................................................ 11-11
11.8.1 Option ................................................................................................................... 11-11
11.8.2 Other Settings ....................................................................................................... 11-12
11.9 System Information ...................................................................................................... 11-12
12 Probes and Biopsy ................................................................................................. 12-1
12.1 Probe ............................................................................................................................ 12-1
12.1.1 Name and Function of Each Part of the Transducer ................................................ 12-3
12.1.2 Orientation of the Ultrasound Image and the Transducer Head ............................... 12-3
12.1.3 Operating Procedures ............................................................................................ 12-4
12.1.4 Wearing the Transducer Sheath ............................................................................. 12-7
12.1.5 Probes Cleaning and Disinfection........................................................................... 12-8
12.1.6 Storage and Transportation .................................................................................. 12-10
12.2 Biopsy Guide ................................................................................................................ 12-11
12.2.1 Basic Procedures for Biopsy Guiding.................................................................... 12-13
12.2.2 Needle-guided Brackets ....................................................................................... 12-13
12.2.3 Needle-guided Bracket Inspection and Installation................................................ 12-16
12.2.4 Biopsy Menu ........................................................................................................ 12-19
12.2.5 iNeedle (Needle Visualization Enhancement ) ...................................................... 12-19
12.2.6 Verify Biopsy Guide Line ...................................................................................... 12-21
12.2.7 Removing the Needle-guided Bracket .................................................................. 12-22
12.2.8 Clean and Sterilizethe Needle-guided Bracket ...................................................... 12-23
12.2.9 Storage and Transportation .................................................................................. 12-25
12.2.10 Disposal ............................................................................................................... 12-25
12.3 Lithotrity ...................................................................................................................... 12-25
13 Battery ..................................................................................................................... 13-1
13.1 Overview ....................................................................................................................... 13-1
13.2 Precautions ................................................................................................................... 13-2
13.3 Installing and Removing the Batteries ............................................................................ 13-2
13.4 Battery Status Indicator.................................................................................................. 13-2
13.5 One Full Discharge / Charge Cycle................................................................................ 13-3
13.6 Checking Battery Performance ...................................................................................... 13-3
13.7 Battery Disposal ............................................................................................................ 13-3
14 Acoustic Output ...................................................................................................... 14-1
14.1 Concerns with Bioeffects ............................................................................................... 14-1
14.2 Prudent Use Statement.................................................................................................. 14-1
v
14.3 ALARA Principle (As Low As Reasonably Achievable) ................................................... 14-1
14.4 MI/TI Explanation .......................................................................................................... 14-2
14.4.1 Basic Knowledge of MI and TI ................................................................................ 14-2
14.4.2 MI/TI Display .......................................................................................................... 14-2
14.5 Acoustic Power Setting.................................................................................................. 14-3
14.6 Acoustic Power Control ................................................................................................. 14-3
14.7 Acoustic Output ............................................................................................................. 14-4
14.7.1 Derated Ultrasonic Output Parameters ................................................................... 14-4
14.7.2 Limits of Acoustic Output ........................................................................................ 14-4
14.7.3 Differences between Actual and Displayed MI and TI.............................................. 14-5
14.8 Measurement Uncertainty.............................................................................................. 14-5
14.9 References for Acoustic Power and Safety .................................................................... 14-6
15 EMC Guidance and Manufacturer's Declaration ................................................... 15-1
16 System Maintenance .............................................................................................. 16-1
16.1 Daily Maintenance ......................................................................................................... 16-1
16.1.1 Cleaning the System .............................................................................................. 16-1
16.1.2 Checking Transducer ............................................................................................. 16-3
16.1.3 Backup of the System Hard Drive........................................................................... 16-3
16.2 Maintenance Checks by Service Engineer ..................................................................... 16-3
16.3 Consumables and Periodic Part Replacement ............................................................... 16-3
16.4 Troubleshooting............................................................................................................. 16-3
Appendix A Electrical Safety Inspection .................................................................A-1
Appendix B Barcode Reader ....................................................................................B-1
Appendix C iWorks (Auto Workflow Protocol) ........................................................C-1
Appendix D Printer Adapter ......................................................................................D-1
vi
©2019 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issue date is 2019-07.
, , , , , BeneView, WATO,
BeneHeart, are the trademarks, registered or otherwise, of Mindray in China and other
countries. All other trademarks that appear in this manual are used only for informational or
editorial purposes. They are the property of their respective owners.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national
and local requirements; and
the product is used in accordance with the instructions for use.
Note
This equipment must be operated by skilled/trained clinical professionals.
I
Warning
It is important for the hospital or organization that employs this equipment to carry out a
reasonable service/maintenance plan. Neglect of this may result in machine breakdown or
personal injury.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
II
Important Information
1. It is the customer’s responsibility to maintain and manage the system after delivery.
2. The warranty does not cover the following items, even during the warranty period:
(1) Damage or loss due to misuse or abuse.
(2) Damage or loss caused by Acts of God such as fires, earthquakes, floods, lightning,
etc.
(3) Damage or loss caused by failure to meet the specified conditions for this system,
such as inadequate power supply, improper installation or environmental conditions.
(4) Damage or loss due to use of the system outside the region where the system was
originally sold.
(5) Damage or loss involving the system purchased from a source other than Mindray or
its authorized agents.
3. This system shall not be used by persons other than fully qualified and certified medical
personnel.
4. Do not make changes or modifications to the software or hardware of this system.
5. In no event shall Mindray be liable for problems, damage, or loss caused by relocation,
modification, or repair performed by personnel other than those designated by Mindray.
6. The purpose of this system is to provide physicians with data for clinical diagnosis. It is
the physician’s responsibility for diagnostic procedures. Mindray shall not be liable for the
results of diagnostic procedures.
7. Important data must be backed up on external memory media.
8. Mindray shall not be liable for loss of data stored in the memory of this system caused by
operator error or accidents.
9. This manual contains warnings regarding foreseeable potential dangers, but you shall
always be alert to dangers other than those indicated as well. Mindray shall not be liable
for damage or loss that results from negligence or from ignoring the precautions and
operating instructions described in this operator’s manual.
10. If the manager for this system is changed, be sure to hand over this operator’s manual to
the new manager.
Notation Conventions
In this operator’s manual, the following words are used besides the safety precautions (refer
to "Safety Precautions"). Please read this operator’s manual before using the system.
III
Operator’s Manuals
You may receive multi-language manuals in compact disc or paper. Please refer to English
manual for latest information and register information.
The content of the operator manual, such as screens, menus or descriptions, may be different
from what you see in your system. The content varies depending upon the software version,
options and configuration of the system.
Manuals on Paper
Operator’s Manual [Basic Volume]: Describes the basic functions and operations of
the system, safety precautions, exam modes, imaging modes, preset, maintenance
and acoustic output, etc.
Operator’s Manual [Advanced Volume]: Describes measurement preset,
measurements and calculations, etc.
Operator’s Manual [Acoustic Power Data and Surface Temperature Data]: Contains
data tables of acoustic output for transducers.
Operation Note: Contains quick guide for basic operations of the system.
NOTE: 1. The manuals in CD are the manuals translated into languages other than
English according to English manuals.
2. When you find that the contents of the manuals in CD are NOT consistent with
the system or English manuals, please ONLY refer to the corresponding English
manuals.
3. The accompanying manuals may vary depending upon the specific system you
purchased. Please refer to the packing list.
IV
Software Interfaces in this Manual
Depending on the software version, preset settings and optional configuration, the actual
interfaces may be different from those in this manual.
Conventions
In this manual, these conventions are used to describe the buttons on the control panel, the
items in menu, buttons in dialog box and some basic operations:
<Buttons>: The angular bracket indicates buttons, knobs and other controls on
control panel.
[Items in menu and buttons in dialog box]: The square bracket indicates items in
menu or buttons in dialog box.
Click [Items or Button]: Move the cursor to the item or button and press <Set>, or
click it on the menu.
[Items in Menu][Items in Submenu]: Selects a submenu item following the path.
[Dyn Rng (Value)]: Indicates menu items with parameter, (value) shows the current
value of the item.
Product Differences
B-Hist B-Profile Double Dist Trace Length Keyboard
Protective Film
Z5 × √ √ √ ×
Z5BW × √ × √ ×
Z5T × × √ √ ×
Z50 √ √ √ √ ×
Z50BW √ √ × √ ×
Z50T √ × √ √ ×
V
1 Safety Precautions
Caution.
NOTE: 1. DO NOT use the system in the vicinity of strong electromagnetic field (such
as a transformer), which may affect the performance of the system.
2. DO NOT use the system in the vicinity of high-frequency radiation source,
which may affect the performance of the system or even lead to failure.
3. When using or placing the system, keep the system horizontal to avoid
disbalance.
Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the probes.
The ultrasonic probe is only for use with the specified
WARNING: 1. ultrasonic diagnostic system. Please refer to the “2.5.2
Probes Available” to select the proper probe.
NOTE: 1. Read the following precautions to prevent the probe from malfunction:
Clean and disinfect the transducer before and after each examination.
After the examination, wipe off the ultrasound gel thoroughly.
Otherwise, the ultrasound gel may solidify and the image quality
would be degraded.
2. Ambient conditions:
To prevent the transducer from being damaged, do not use it where it will be
exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
Use the probes under the following ambient conditions:
Ambient temperature: 0°C ~ 40°C
Relative humidity: 30% to 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Use the probe 10L24EA under the following ambient conditions:
Ambient temperature: 0°C ~ 40°C
Relative humidity: 30% to 90% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Use the probe D6-2EA under the following ambient conditions:
Ambient temperature: 18°C ~ 30°C
Relative humidity: 20% to 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
3. Repeated disinfection will eventually damage the probe, please check the
probe's performance periodically.
b c
2.2 Contraindication
This system is not intended for ophthalmic use.
Model code
Product code
NOTE: The functions described in the operator’s manual may vary depending upon the
specific system you purchased.
NGB-004
65EC10EA
metal/needle un- / 16G, 17G, 18G
65EC10EB
detachable
2.5.3 Options
No. Item
1 iScape View
2 Color/Power module
3 PW module
4 Smart 3D
5 HPRF (PW should be configured first)
6 IMT
7 Free Xros M
8 iLive
9 Tissue Doppler Imaging
10 iNeedle
11 iWorks
12 Smart Face
13 Smart OB
DICOM Basic (including: task management, DICOM storage, DICOM print,
14 DICOM storage commitment, DICOM media storage (including DICOM DIR)
etc.)
15 DICOM worklist DICOM
Basic should
16 DICOM MPPS
be
17 DICOM DICOM Query/Retrieve configured
first
DICOM OB/GYN Structured
18
Report
19 DICOM Vascular Structured Report
20 DICOM Cardiac Structured Report
21 Footswitch
22 Battery Pack
23 Mobile trolley (UMT-150 or UMT-160)
24 Needle-guide bracket
25 Pack
NOTE: If the ultrasound system can not recognize the SONY UP-X898MD printer
automatically, you may need to change the settings on the printer: push <PUSH
ENTER> to enter the main menu and select [DIGITAL]->[DRIVER], and select
[897].
Rear View
Bottom View
<1>
<2> <3> <4>
12V
5V
A
<8>
Caution
Dangerous voltage
Equipotentiality
Power button
Network port
USB ports
Video output
AC (Alternating current)
Standby indicator
Probe port A
Probe port B
Probe port C
Manufacture date
NOTE: When you start the system or switch between probes, you will hear clicking
sounds – this is expected behavior.
To remove the USB storage device: Click to open the [Remove USB Device] screen.
Select the device to be removed. Click [OK].
Data cord
USB port
1. Connect the data cable with the USB port on the ultrasound system.
2. Plug the other end power plug into an appropriate outlet.
3. Power on the system and the printer.
5. Return to “Print Service” page, and select “Report Print” in the list, and set attributes in
the Property box under the screen:
Select the printer model from the drop-down list right to the “Printer”;
Set Paper Size.
Model of the connected printer is not displayed in the “Printer Driver” list;
Click “Printer Driver”->[Add Local Printer], the system will prompts you to update the
ppd file.
At this time, you need to install the printer driver manually as described in the following:
a) Download the ppd file from the printer manufacturer’s official website (contact
service engineer if necessary), and copy the ppd file to the storage device (USB
disk as an example).
b) Connect the USB disk to the ultrasound system, click [Setup]->[Print Preset]-
>”Printer Driver”->[Add Local Printer] to run the ppd file and finish the installation.
NOTE: All printer drivers integrated in the ultrasound system come from official website of
each printer manufacturer (you can check models in the [Setup]->[Print Preset]-
>”Printer Driver”->[Supported Printer List]).Printer drivers may not be updated in
time due to software version and region limitation, consult printer manufacturer for
details.
NOTE: 1. Before connecting the shared printer, make sure the ultrasound system and
the computer or server (connected with shared printer) are in the same
network domain, and the network is working normally.
2. The IP address and the server name should be valid, e.g. \\10.2.40.123 or
\\5-HP, otherwise, the system will fail to connect.
3. If the server has set accessing limitation, the system will prompt a dialogue
box to identify the user. Enter the correct user name, domain name and
password, and then click [OK].
(3) Select the service type as "Digital Image Print" and enter the service name manually.
(4) Click OK to return to the Printer Service page.
(5) Set the items in the Property box and click [Save] to save the settings.
Image print
For DICOM image printing, refer to “10 DICOM”.
Modify print service:
a) Select an existed printer service in the list.
b) Select the printer type in the Property box.
c) Set the attribute: paper size, orientation, etc.
d) Click [OK] to confirm.
Image print
Select the image to be printed on the iStation or Review screen, and click [Send
To] to select the printer to print.
Please refer to the accompanying manuals of the printers for more details.
Information Area
The information area displays manufacturer logo, hospital name, exam date & time,
acoustic power & MI/TI, freeze icon, patient information, probe model, current exam mode,
and accession #, etc. It can be preset whether to display the operator, patient gender, age,
ID, name etc. via [Setup]-> [System Preset]-> [General]. See “11.1.2 General” for detailed
preset procedures.
Contents
Control
button
Composition Description
The title bar is used to give a description for the content and
Title Bar
function of the screen.
Use the selection pointer and <Set> key to open / close the
Page Tab
available pages.
Radio button: click to select the item.
Check box: click to check or uncheck the item.
Content Entry box: enter characters manually via the keyboard. .
Drop-down list box: click [▼] to show the list and select
an item.
When the operation of a screen is complete, press the [OK]
[OK] and [Cancel] or [Cancel] button to save or cancel the operation, and close
the screen.
To reposition a dialogue box:
1. Roll the trackball to move the cursor onto the title bar of the dialogue box. At this time the
cursor becomes a ; press the [Set] key.
2. Roll the trackball and reposition the rectangular graphic to the new desired location.
3. Press the <Set> key, and the dialogue box is moved to the desired position.
2
3
4
Information includes:
1. General information
Name
Enter patient name through the keyboard. Characters of A through Z and 0 through 9 and
“.” are allowed. “\”, “^” and “=” are not permitted.
Patient ID
Patient ID is generated automatically by the system after starting a new patient, and can
be modified manually. Once you enter the ID and confirm it, you are not allowed to
change it.
Other ID
The second ID of the patient used for other information like insurance ID.
ABD
Weight /
(Abdomen) BSA (body After the height and weight are inputted, the system will
surface automatically calculate the BSA based on the formula which
area) is set via [Setup]-> [System Preset]-> [General].
Calculate gestation age (GA) and estimated delivery date
(EDD) based on last menstrual period (LMP), in vitro
fertilization (IVF), basic body temperature (BBT), previous
exam date (PRV). Select LMP, IVF, PRV, BBT, or EDD from
the drop-down list; or, calculates GA and LMP according to
the EDD and entered date.
LMP: After you enter LMP, the system will calculate
Calculation and display GA and EDD.
index IVF: After you enter IVF, the system will calculate GA
OB and EDD.
(Obstetric) PRV: input the date and GA of the last exam, the
system will calculate a new GA and EDD.
BBT: input BBT, the system will calculate the GA and
EDD.
EDD: after you enter EDD, the system will calculate
and display GA and LMP.
Gravida Times of pregnancy.
Ectopic Times of abnormal pregnancy. e.g. extrauterine pregnancy
Gestations Number of embryos (default value is 1)
BREAST Height /
(Breast) Weight /
3. Operating Information
Ref. Physician: the people who requires the operator to do the ultrasound operation.
“\”, “^” and “=” are not permitted.
Diagnostician: people who is responsible for exam. “\”, “^”, “=” and “,” are not
permitted.
Operator: people who is responsible for images collection and scanning. “\”, “^” and
“=” are not permitted.
Accession #: exam number used in DICOM, “\” is not permitted.
Comment: exam-specific explanation or remarks.
4. Functional key
[New Patient]: click to clear the current patient information in the patient information
screen in order to input new patient information.
[New Exam]: click to clear the current exam information in order to create a new
exam for the current patient.
[Pause Exam]: to pause the current exam.
[Cancel Exam]: to cancel the current exam.
(2) Roll the trackball and press <Set> to select the exam mode, and use the direction
keys to turn pages of the exam modes.
5.4 B Mode
B mode is the basic imaging mode that displays real-time images of anatomical tissues and
organs.
Depth
Description This function is used to adjust the display depth of sampling, the real-time
value of which is displayed on the image parameter area in the upper left
corner of the screen.
Operation Use the <Depth/Zoom> knob to adjust the depth;
The adjustable depth values vary depending upon the probe types.
Effects Increase the depth to see tissue in deeper locations, while decrease the depth
to see tissue in shallower locations.
Impacts Depth increase will cause a decrease in the frame rate.
TGC
Description The system compensates the signals from deeper tissue by segments to
optimize the image.
There are 8-segment TGC sliders on the control panel corresponding to the
areas in the image.
Operation To increase the gain compensation at an area of interest, move the TGC slider
to the right. To decrease the gain compensation at the corresponding area of
interest, move the TGC slider to the left.
About 1.5s after the adjustment is finished, the TGC curve disappears.
Effects Adjust the signal gain for the certain image area to get a balanced image.
Frequency
Description This function is used to select the operating frequency of the current probe,
the real-time value of which is displayed in the image parameter area in the
upper left corner of the screen, where “F” represents B mode frequency, and
“FH” represents harmonic frequency.
A. power
Description Refers to the power of ultrasonic wave transmitted by the probe, the real-time
value of which is displayed in the image parameter area in the upper left corner
of the screen.
Operation Adjust through the [A.Power] item in the image menu;
The adjusting range is 7%-100% in increments of 3%.
Effects Generally, increasing the acoustic power will increase the brightness and
contrast of the image as well as the force of penetration.
Impacts You should perform exams according to actual situation and follow the ALARA
Principle.
Focus
Description Refers to adjustment of focus of the ultrasonic beams, symbols as " " of
which will be displayed on the right of the image.
Operation Adjust the focus number through the [Focus Number] in the menu;
Adjust the focus position through the [Focus Position] in the menu;
In B Mode, focus number can be switched among 1-4.
Effects The area that is focused will be of a higher contrast and resolution to provide
a much clearer image.
Impacts The greater the number of focus, the slower the frame rate to image.
Line Density
Description The function determines the quality and information of the image.
Operation Adjust through the [Line Density] item in the menu;
Levels: UH/ H/ M/ L.
Effects The higher the line density, the higher the resolution, and the lower the frame
rate.
Dynamic Range
Description This function is used to adjust the B image resolution to compress or expand
the gray display range.
The real-time dynamic range value is displayed on the image parameter area
in the upper left corner of the screen.
Operation Adjust through the [Dyn Ra.] item in the menu;
iClear
Description This function is used to increase image profile, so as to distinguish the image
boundary.
Operation Adjust through the [iClear] item in the menu;
The system provides 5 levels of iClear effects adjustment, off represents no
iClear is turned on, and the bigger the value the stronger the effect.
Effects The bigger the value the clearer the profile of the image.
Persistence
Description This function is used to superimpose and average adjacent B images, so as to
optimize the image and remove noises.
Operation Adjust through the [Persistence] item in the menu;
The system provides 8 level of frame average adjustment, the bigger the
value the stronger the effect.
Effects Persistence can remove image noise to make details to be clearer.
Impacts Persistence increase may lead to signal missing.
Rotation/ Invert
Description This function provides a better observation for image display.
iBeam
Description This function is used to superimpose and average images of different steer
angles to obtain image optimization.
Operation Adjust through the [iBeam] item in the menu;
Off: no iBeam
On: maximum iBeam optimization
Effects Images after iBeam processing can be optimized with less spot noise and
higher resolution, so that more details for the structure are revealed.
Impacts iBeam is not available when ExFov function is turned on.
Auto Merge
Description In the Dual-split mode, when the images of the two windows use the same
probe type, depth, invert status, rotation status and magnification factor, the
system will merge the two images so as to extend the field of vision.
Operation Turn on/ off the function through the [Auto Merge] item in the menu;
Impacts Available only for linear probes.
The function is available in real-time imaging, freeze or cine review status.
Gray Map
Description Adjusting grayscale contras to optimize the image.
Operation Select among the maps through the [Gray Map] item in the menu.
The system provides 1-25 gray maps to be selected among.
Impacts The function is available in real-time imaging, freeze or cine review status.
iTouch
Description To optimize image parameters as per the current tissue characteristics for a
better image effect.
Operation Press <Gain/ iTouch> on the control panel, the iTouch symbol will display in
the image parameter area.
Click [iTouch] on the image menu to adjust gain in iTouch status among -12
through 12dB.
Gray Invert
Description Reverse the polarity of the image.
Operation Click [Gray Invert] to turn on or off the image inversion.
Impacts The gray invert functions is available in real-time imaging, freeze or cine review
status.
Its post process adjustments will not influence the cine review.
HScale
Description Display or hide the width scale (horizontal scale).
The scale of the horizontal scale is the same as that of vertical scale (depth),
they change together in zoom mode, or when the number of the image
window changes. The HScale will be inverted when image is turned up/down.
Operation Click [HScale] on the menu to display or hide the scale.
LGC
Description Adjust the gain of scan lines to increase the image lateral resolution.
Operation Click [LGC] to access the adjusting dialog box.
The 8 LGC items displayed on the screen indicate the corresponding image
areas on the image screen.
Click corresponding [LGC1-8] to adjust the gain. The higher the value, the
higher the gain.
The system also provides several preset parameters for imaging.
5.5 M Mode
5.5.1 M Mode Exam Protocol
1. Select a high-quality image during B mode scanning, and adjust to place the area of
interest in the center of the B mode image.
2. Press <M> on the control panel, and roll the trackball to adjust the sampling line.
Focus Position
Description To change the focus position in M mode, symbols as " " of which are displayed
on the right of the image.
Operation Adjust the focus position through the [Focus Position] item in the menu.
Display Format
Description To set the display format of M mode image with B mode image.
Operation Adjust through the [Display Format] item in the menu;
There are 4 formats available for the images display: V1:1, V1:2, V2:1, Full.
Effects Adjust according to the actual situation and obtain a desired analysis through
comparison.
There are 6 levels of scan speed available, the smaller the value the faster the
speed.
Effects Speed changing makes it easier to identify disorders in cardiac cycles.
Gray Map
Description Adjusting grayscale contrast to optimize the image.
Operation Select among the maps through the [Gray Map] item in the menu.
The system provides 1-25 gray maps to be selected among.
Impacts The function is available in real-time imaging, freeze or cine review status.
Edge Enhance
Description This function is used to increase image profile, so as to distinguish the image
boundary.
Operation Adjust through the [Edge Enhance] item in the menu;
The system provides 14 levels of edge enhance effects, off represents no edge
enhance is turned on, and the bigger the value the stronger the effect.
Impacts Larger edge enhance may lead to noise increasing.
Dynamic Range
Description Adjusts contrast resolution of an image, compresses or expands gray display
range. The real-time dynamic range value is displayed on the image parameter
area in the upper left corner of the screen.
Operation Adjust through the [Dyn Ra.] item in the menu;
Steer
Description The feature is used to adjust the ROI of color flow of different angles with
immobility of the linear probe.
Operations Adjust through the <Steer> key on control panel or [Steer] item on image
menu.
Effects This function is used to adjust the scan angle of linear probes, so as to change
the angle between the transmitting beam and flow direction.
Impacts Steer is valid only for linear probes.
Scale
Description This function is used to adjust the speed range of color flow, which is adjusted
through PRF in the system. The real-time PRF value is displayed in the image
parameter area in the upper right corner of the screen.
Operations Press the <Scale> key on the control panel and rotate the Multifunctional Knob
on the right to adjust.
The adjusting range varies by frequency, probe and depth; adjust according to
the actual situation.
Effects To provide a much clearer color flow image.
Use low PRF to observe low-velocity flows, and high PRF to observe high-
velocity flows.
Impacts Aliasing may occur if low velocity scale is used and high velocities are
encountered.
Low velocities may not be identified when a high velocity scale is used.
Baseline
Description Refers to the area where the velocity is zero in the scale. Adjust according to
the actual situation so as to get an optimum flow display.
Operations Press the <Baseline> key on the control panel and rotate the Multifunctional
Knob on the right to adjust.
Positive value means to increase the signals above the baseline, and negative
value means to increase the signals below the baseline.
Invert
Description To set the display mode of color flow and the color scale will be inverted when
the function is turned on.
Priority
Description This function is used to set levels of the flow display, to display the grayscale
signal or color signal.
Operations Click the [Priority] item on the image menu to select the value.
The adjusting range of the priority is 0-100%.
The higher the value, color signals are prior to be displayed; while the lower
the value, grayscale signals are prior to be displayed.
Packet Size
Description This function is an indication of the ability to detect flow, which is used to
adjust the accuracy of color flow.
Operations Click the [Packet Size] item on the image menu to adjust the value.
There are 4 levels of packet size available, 0 represents no packet size control
and the bigger the value the higher the sensitivity.
Effects The higher the packet size, the more sensitive indication for low-velocity flow.
Impacts Adjusting the packet size may lead to frame rate changing.
Persistence
Description This function is to adjust the temporal smooth in Color mode to optimize the
image.
Operations Click the [Persistence] item on the image menu.
The system provides 5 levels of persistence adjustment, 0 represents no
persistence and the bigger the value the stronger the effect.
WF (Wall Filter)
Description It filters out low-velocity signals to provide effective information, and this
function is used to adjust the filtered frequency. The real-time value is
displayed in the image parameter area in the upper right corner of the screen.
Operations Click the [WF] item on the image menu.
There are 8 levels of wall filter function available, and adjust according to the
actual situation.
Impacts Flow signals may be missing.
Color Map
Description This function is a combination of several image parameters, which indicates
the display effect of color image.
Operations Click the [Color Map] item on the image menu to select among the maps.
The system provides 21 different maps to be selected among, where the V
group provides 11 ordinary maps and the VV group provides 10 2-D maps.
Dual Live
Description This function is used to display B image and Color image synchronously.
Operations Turn on or off the function through the [Dual Live] item on the image menu.
When the function is turned on, the window will be automatically switched to
the dual windows (one for B image, and the other for Color image).
Line Density
Description Line density determines the quality and information of the image.
Operations Adjust through the [Line Density] item on the image menu.
There are 4 levels of line density provided: H, L, UH, M.
Effects The higher the line density, the higher the resolution.
Impacts The higher the line density, the lower the frame rate.
ROI Adjustment
Description This function is to adjust the width and position of ROI in Color mode.
Operations When the ROI box is dotted line, roll the trackball to change the size.
When the ROI box is solid line, roll the trackball to change the position.
Press <Set> to switch between the solid line and the dotted line.
Impacts The larger the ROI box, the lower the frame rate, and the lower the resolution
and color sensitivity.
Smooth
Description This feature is used to reject noise and smooth the image.
Operations Adjust through the [Smooth] item on the image menu.
The system provides 5 levels of smooth function, the bigger the value the
higher the smooth.
Power Map
Description This feature indicates the display effect of Power image.
The maps in the Power mode image are grouped into two categories: Power
maps and Directional Power maps.
Operations To select among the maps, turn the knob under [Map] on the image menu.
There are 8 kinds of maps provided: P0-3 belong to Power mode maps, while
Dp0-3 belong to Directional Power mode maps.
The Power maps provide information of blood flow, which are highly sensitive
to the low-velocity flows.
The Directional Power maps provide information of flow direction.
Dynamic Range
Description This function is to adjust the transformation of echo intensity into color signal.
Operations Click the [Dynamic Range] item on the image menu to adjust the dynamic
range.
The adjusting range is 10-70dB in increments of 5dB.
PW Sampling Line SV
Adjustment
Angle
SVD
3. Set the position of the sample line by moving the trackball left and right, and set the SVD
by moving the trackball up and down, adjust the angle and SV size according to the
actual situation.
4. Press <PW> or <Update> to enter PW mode again and perform the examination. You can
also adjust the SV size, angle and depth in real-time scanning.
5. Adjust the image parameters during PW mode scanning to obtain optimized image.
6. Perform other operations (e.g. measurement and calculation) if necessary.
Frequency
Description Refers to the operating frequency in PW mode of the probe, the real-time value
of which is displayed in the image parameter area in the upper left corner of the
screen.
Operation Select the frequency value through the [Frequency] item in the image menu or
rotate the <Focus/Freq./THI> knob on the control panel.
Values of frequency vary depending upon the probe types.
Select the frequency according to the detection depth and current tissue
characteristics.
Effects The higher the frequency, the better the resolution and sensitivity, and the
worse the force of penetration.
Baseline
Description Refers to the area where the velocity is zero in the spectrum.
Operations Press the <Baseline> key on the control panel and rotate the Multifunctional
Knob on the right to adjust.
Effects To change the range of flow velocity to optimize the image.
PW Steer
Description This function provides adjustment on angles for sampling line.
Operations Adjust through the <Steer> key on control panel or [Steer] item on image
menu.
Effects This feature is used to steer the direction of the beam so as to change the
angle between the beam and flow direction with immobility of the linear probe.
Values of steer angles vary by probe.
Steer is available only for linear probes.
Scale
Description This function is used to adjust the speed range of flow, which is adjusted
through PRF in the system.
The real-time PRF value is displayed in the image parameter area in the
upper right corner of the screen.
Operations Press the <Scale> key on the control panel and rotate the Multifunctional
Knob on the right to adjust.
Effects To provide a much clearer color flow image.
Use low PRF to observe low-velocity flows, and use high PRF to observe
high-velocity flows.
Impacts Aliasing may occur if low velocity scale is used and high velocities are
encountered.
Low velocities may not be identified when a high velocity scale is used.
Wall Filter
Description It filters out low-velocity signals to provide effective information, and this
function is used to adjust the filtered frequency. The real-time value is
displayed in the image parameter area in the upper right corner of the screen.
Operations Select through the [WF] item on the image menu.
There are 7 levels of wall filter function provided.
Impacts Signals of low-velocity flow may be missing.
T/F Res
Description This function is used to adjust for a balance between time resolution and
spatial resolution.
Operations Adjust through the [T/F Res] item on the image menu.
There are 5 levels of T/F Res values available.
Dynamic Range
Description The dynamic range conveys the information that being transformed from echo
intensity to gray scale.
Operations Adjust through the [Dynamic Range] item on the image menu.
The adjusting range is 24-72dB in increments of 2dB.
Effects The more the dynamic range, the more specific the information, and the lower
the contrast with more noise.
Speed
Description This function is used to set the scanning speed of PW mode imaging.
Operations Click the [Speed] item on the image menu.
There are 6 levels of scan speed available, the smaller the value the faster the
speed.
Effects Changing the speed makes it easier to identify the cardiac cycles and to
detect more details.
Display Format
Description To set the display format of PW mode image with B mode image.
Operations Click the [Display format] item on the image menu to adjust.
Options: V2:1, V1:2, V1:1, Full.
Audio
Description This function is used to adjust the output audio in spectrum map.
Operations Click the <Audio> item on the image menu to adjust the volume.
The adjusting range of the audio is 0-100%.
Effects Utilizing the output audio helps to identify the feature and status of flow.
Gray Map
Description Adjusting grayscale contrast to optimize the image.
Operation Select among the maps through the [Gray Map] item in the menu.
The system provides 1-25 gray maps to be selected among.
Impacts The function is available in real-time imaging, freeze or cine review status.
Duplex/ Triplex
Description This function is used to set if B image (B + Color image) and PW image are
displayed synchronously.
Auto Calculation
Description This function is used to trace the spectrum and calculate parameters of PW
mode image, and the results of which are displayed in the result window.
Auto Turn on or off the auto calculation function through the [Auto Calc] item on the
Calculation image menu.
Auto Select parameters in the dialogue box prompted by clicking [Auto Calc Param]
Calculation on the image menu.
Parameter
Auto Calc To set the number of heart cycle for auto calculation.
Cycle Click [Auto Calc Cycle] on the image menu to select the cycle number.
Operations In real-time scanning, the results displayed are derived from the calculation of
latest cardiac cycle;
In the freeze and cine status, the results displayed are calculated from the
current selected area.
Trace
Trace Area To set the trace area of the Doppler wave in the spectrum map, applicable for
auto calculation.
Change the trace area through the [Trace Area] item on the image menu.
The available selections are: Above, Below, All.
SV
Description To adjust the SV position and size of sampling in PW mode, the real-time
value of SV and SVD are displayed in the image parameter area in the upper
right corner of the screen.
SV size Click the [SV] on the image menu to adjust the SV size.
Value: 0.5-20mm.
SVD Roll the trackball to select the SV depth.
Effects The smaller the SV size, the more accurate the result; and more details are
obtained when selected large SV size.
iTouch
Description To optimize image parameters as per the current tissue characteristics for a
better image effect.
Operations Press <Gain/ iTouch> on the control panel to turn on the function.
Angle
Description This function is used to adjust the angle between Doppler vector and flow to
make the velocity more accurate.
The real-time adjusting angle value is displayed on the right part of the
spectrum map.
Operations Click the <Angle> item on the image menu to adjust.
The adjustable angle range is -89~89°, in increments of 1°.
For an image in the traditional M mode, the M-mark line goes along the beams transmitted
from the probe. Thus it is difficult to obtain a good plane for difficult-to-image patients that
cannot be moved easily. However, in the Anatomical M mode, you can manipulate the M-mark
line to move to any position at desired angles. The system supports anatomical M scanning
(including Free Xros M mode) in 2D imaging modes (B, Color, Power and TVI mode).
Shortcut key setting
You can preset a user-defined key for entering Free Xros M mode: [Setup] -> [System
Preset] -> [Key Config].
Free Xros M Imaging Procedures
1. In real-time 2D mode or M mode, adjust the probe and image to obtain the desired plane.
Or select the desired 2D cine files.
2. Click [Free Xros M] on the image menu of the B mode to enter Free Xros M mode, or
press the user-defined key.
3. Adjust the M-mark line to obtain optimized images and necessary information.
In Dual-split or Quad-split mode, it is the current activated window that enters the Free
Xros M mode.
Free Xros M imaging is supported on frozen B image, B+M image and B+Power/Color/TVI
image.
TDI (Tissue Doppler imaging) mode is intended to provide information of low-velocity and
high-amplitude tissue motion, specifically for cardiac movement.
There are 3 types of TDI mode available:
Tissue Velocity Imaging (TVI): This imaging mode is used to detect tissue movement
with direction and speed information. Generally the warm color indicates the
movement towards the probe, while the cool color indicates the movement away from
the probe.
Tissue Energy Imaging (TEI): This imaging mode reflects the status of cardiac
movement by displaying the intensity of tissue, the brighter the color the less the
intensity.
Tissue Velocity Doppler Imaging(TVD): This imaging mode provides direction and
velocity information of the tissue quantificationally.
Only probe 35C20EA can be applied for TDI function.
NOTE: 1. The measurement accuracy for the spliced image may be degraded, exercise
caution when measurements are performed on an iScape image.
2. If there is a trace during the retracing, please don’t perform measurement
across the trace.
a b c
5.12.2 Overview
Ultrasound data based on three-dimensional imaging methods can be used to image any
structure where a view can’t be achieved by standard 2D-mode to improve understanding of
complex structures.
Terms
Volume: a three-dimensional content.
Volume data: the image data set of a 3D object rendered from 2D image sequence.
3D image (VR): the image displayed to represent the volume data.
View point: a position for viewing volume data/3D image.
Section image (or MultiPlaner Rendering as MPR): tangent planes of the 3D image
that obtained by algorithm. As shown in the figure below, XY-paralleled plane is C-
section, XZ-paralleled plane is B-section, and YZ-paralleled plane is A-section. The
probe is moved along the X-axis.
ROI (Region of Interest): a volume box used to determine the height and width of
scanning volume.
VOI (Volume of Interest): a volume box used to determine the area of a sectional
plane for 3D imaging.
View point
2 3
1 4 X
6
7
5 8
Z
ROI and VOI
Before image acquisition after the system enters Smart 3D imaging, a B image with ROI
displays on the screen. A line (shown in the following figure) that shows the upper edge
position of VOI is inside ROI.
ROI
Cut plane
c d
e f
View Direction
a. Up/Down b. Down/Up
c. Left/Right d. Right/Left
e. Front/Back f. Back/Front
Wire cage
When you view a Smart 3D image on the display monitor, it’s sometimes difficult to
recognize the orientation. To help, the system displays a three-dimensional drawing to
illustrate the orientation. Of which, the blue plane presents the image acquisition where
started, while the red plane presents the image acquisition where ended, besides, a
yellow plane in the wire cage presents the position of the sectional plane. See the graphic
below:
The ultrasound images are provided for reference only, not for
CAUTION: confirming a diagnosis. Please use caution to avoid misdiagnosis.
5.12.3 Static 3D
Static 3D provides single frame image acquisition of 3D images. During scanning, the probe
performs the scanning automatically.
The probe D6-2EA supports the static 3D.
Draw a circle
here
Cross cursor on
the VOI curve
ROI
Activate MPR
Click [VR/MPR] to activate MPR or 3D image (VR).
MPR Viewing
In the actual display, different colors for the window box and the section line are used to
identify the MPR A, B and C.
View Direction
The Region of Interest (ROI), also referred to as the Render Box in rendering, contains the
section of the volume you want to render. Therefore, objects that are not inside the box are
not included in the render process and are cut out (this is important in surface mode to allow a
free line of sight). This may or may not be the entire VOI.
You can adjust the view direction of the ROI.
a b
e f
View Direction
a. Up/Down b. Down/Up
c. Left/Right d. Right/Left
e. Front/Back f. Back/Front
Adjust VOI
VOI On
The VR image displays VOI information.
1. In image viewing status, click [VOI] to turn it to “On.”
2. Roll the trackball to adjust the VOI position, size and curved VOI, and press <Set> to
switch between the adjusting status.Or use [Slice] item to adjust the relative position of
MPRs so as to slice through the VR.
VOI Off
The VR image displays ROI information.
Click [VOI] to turn it to “Off.” The ROI image is displayed on the screen. Roll the trackball to
observe the MPR.
Accept VOI
1. This function is usually used for MPR observation and to determine the relative position
of the MPR to the VR.
2. Click [Accept VOI].
3. Select the desired MPR by clicking [A], [B] or [C] or [VR].
4. Roll the trackball to view the current active MPR. The other two MPRs change
correspondingly.Or use [Slice] item to adjust the relative position of MPRs so as to slice
through the VR.
In Accept VOI status, when the VR is active or the MPR which is perpendicular to the view
direction is active, the center point of the VR displays. Adjust the position using the trackball.
Image Zooming
To adjust the zoom factor of VR. The sectional images will be zoomed in/out accordingly.
Operation
Set the VR window as the current window.
Press <Zoom> to increase or decrease the magnification factor.
5.12.4 Smart 3D
The operator manually moves the probe to change its position/angle when performing the
scan. After scanning, the system carries out image rendering automatically, then displays a
frame of the 3D image.
Smart 3D is an option. All probes for the Ultrasound System support the Smart 3D except for
D6-2EA.
5.12.5 4D
4D provides continuous, high-volume acquisition of 3D images. During scanning, the probe
performs the scanning automatically.
4D image acquisition operations are similar to those of Static 3D. The only difference is that in
static 3D mode, only a single frame 3D image is captured, while in 4D mode, continuous,
high-volume acquisition of 3D images is provided.
4D imaging is an option. The probe D6-2EA supports the 4D mode.
5.12.6 iLive
iLive brings you a better imaging experience by adding a light rendering effect to the
traditional method. It supports the global lighting mode as well as the partial scattering mode,
allowing human tissue texture to be revealed more clearly.
iLive is an option, and is available under Smart 3D, Static 3D and 4D modes. To use the iLive
function, you must configure the Smart 3D module or the 4D module.
To Activate iLive
1. Enter 3D/4D image viewing status, or double-click the saved 3D/4D cine file in the
iStation or Review screen.
2. Select the render mode to be iLive on the screen, and adjust the parameters.
Imaging using iLive
1. Select the imaging mode:
Use the ordinary probe and press <3D/4D> to enter Smart 3D.
Or, use 4D probe and press <3D/4D> twice to enter 4D mode or Static 3D mode.
2. Select the render mode to be iLive on the screen, and set the related parameters (quality
and angle, etc.).
3. Press <Update> to begin acquisition.
4. The system finishes acquisition and enters the image viewing screen.
In image review status, you can perform operations such as VOI setting, image editing,
comment adding, body mark adding, etc.
5. Or press <B> or <3D/4D> to exit the mode.
1
9
7
2
5
6
8
3
10
1
1
Layout
The system supports several types of display layout: 2*2, 3*3, 4*4 and 5*5. Click the
corresponding icon on the screen to select, and the selection [Slice Number] changes
accordingly.
Reference image
Click [A Plane], [B Plane] or [C Plane] to select the reference image.
Slice and slice line
Central slice: the central plane which corresponds to the central slice line is the
central slice, marked with a green “*” in the top-left corner of the image.
View vertical or horizontal slices.
Active slice: the central plane which corresponds to the green slice line is the active
slice, marked with a green box. The default active slice is the central slice.
Slice order number: indicates the order of the slices. The position of the central slice
is “0”, the slices before the central slice are marked with negative integral numbers,
and the slices after the central slice are marked with positive integral numbers.
Basic Procedures
1. Acquire Static 3D image or frozen 4D single-frame image of fetal face.
Parameter adjusting
Parameters under Smart Face are similar to those under Static 3D mode.
FaceContact
Description The higher the value, the more adjacent the VR image is to the fetal
face, and the more obstacles cleared.
The lower the value, the further the VR image is to the fetal face, and
the less obstacles cleared.
Operation Rotate [FaceContact] to adjust the parameter.
The adjusting range is: -15-15 in increment of 1.
Quick Rot.
Description Rotate the image quickly.
Operation Adjust in VR Orientation: 0°, 90°, 180°, 270°.
Total frames
End mark
Auto Review
Reviewing all
a) In the manual cine review status, click [Auto Play] on the image menu to activate
auto cine review.
b) Reviewing speed: In the auto cine review status, use [Auto Play] item in the
image menu to adjust the review speed.
c) In auto play status, set [Auto Play] item to Stop to exit auto play.
Setting Region of Auto Review
a) Set start frame: move the playback mark to the desired frame, and click [Set First
Frame] in the image menu to set the frame as start point.
b) Set end frame: move the playback mark to the desired frame, and click [Set End
Frame] in the image menu to set the frame as end point.
c) Click [Auto Play] in the image menu; the system plays the auto review region
automatically.
d) Rotate the knob on the left of the control panel to increase/decrease the auto
play speed.
e) In the auto cine review, setting [Auto Play] to Stop or rolling the trackball will stop
the auto cine review and enter the manual cine review.
f) Click [Skip to First]/ [Skip to Last] to review the first or last frame.
Tips: you can perform cine review on each image window in the dual/ quad splitting
mode, and set auto review region for each window.
Start mark
Playback mark
End mark
The frame mark on the time mark of M/PW/TDI image indicates the corresponding 2D
image. In the statuses rather than the dual live, you can only review images in the currently active
window.
6.5 Preset
Open [Setup] →[System Preset]→[General] to preset the cine storage length.
Measurement 7-1
Measurement Tools Function
Volume The volume of a target.
The length of two line segments, which are perpendicular to each
Cross Line
other.
Parallel Line The distance between each pair of parallel lines in a sequence.
Trace Length(Trace) Measures the length of a curve on the image.
Trace Len (Spline) Measures the length of a curve on the image.
Measures the length of two line segments, which are perpendicular to
Double Dist
each other.
Measures the lengths of any two line segments and the calculated
Distance Ratio
ratio.
Area Ratio The areas of any two regions and the calculated ratio.
The grayscale distribution of ultrasonic echo signals in a closed
B histogram
region.
B profile The grayscale distribution of ultrasonic echo signals across a line.
Color Velocity Color flow velocity (only valid for Color mode).
Volume Flow Blood flow through some vascular cross section per unit time.
Measures the distance between LI (Lumen-Intima) and MA (Media-
Adventia)
IMT
NOTE: The IMT function is unavailable on the product not configured
with IMT.
7-2 Measurement
7.2.3 Doppler General Measurements
Doppler general measurements refer to general measurements on PW-mode images. The
measurements listed below can be performed:
Measurement
Tools Function
Measurement 7-3
7.4 Measurement Accuracy
Table 1 Error of 2D Images
Parameter Value Range Error
Distance Full screen Within ±4%
Area Full screen Within ±8%
Circumference / Within ±20%
Angle Full screen Within ±3%.
Linear and convex-wide probe: within ±10%
Distance (iScape) Full screen
Micr-convex probe: within ±15%
Table 2 Volume Measurements
Parameter Value Range Error
Volume Full screen Within ±12%.
Table 3 Time/Motion Measurements
Parameter Value Range Error
Distance Full screen Within ±4%
Time Timeline Display Within ±2%.
Heart rate Timeline Display Within ±5%.
PW velocity 10cm/s~200cm/s ≤10% (correction angle ≤60º)
NOTE: Within the selected field range, the measurement accuracy is ensured within the
range mentioned above. The accuracy specifications are performance in the worst
conditions, or based on the real test for the system, regardless of acoustic speed
error.
7-4 Measurement
8 Comments and Body Marks
8.1 Comments
Comments can be added to an ultrasound image to bring attention, notate or communicate
information observed during the examination. You can add comments to: zoomed image, cine
review image, real-time image, frozen image. You can type the character as comments; insert
the pre-defined comments from the comment library; or insert arrow markers.
You must ensure that the entered comments are correct.
WARNING: Incorrect comments may cause misdiagnosis!
Double press the <Set> key to enter the edit status, use the or key to
move the cursor to a location where needs to insert characters, and type characters.
2. Press the <Del> key to delete the comment character or text on the right side of the
cursor; Press the <Backspace> key to delete the comment character or text on the
left side of the cursor.
3. Roll the trackball or press the <Set> key or the multifunctional knob to confirm the
modification and to exit the edit status, and the color of the comments turns yellow.
Modifying (Editing) Arrows
1. Move the cursor on the arrow that needs to be modified. After the cursor turns to ,
press the <Set> key. The color of the current arrow turns green, and there is green
frame around the arrow, indicating the arrow can be edited. Move the cursor to
change the arrow position.
2. Rotate the multifunctional knob to modify the arrow direction.
3. Press the <Set> or <Enter> key to complete the modifying operation.
8.2.2 Menu
In the body mark menu. You can:
Select Bodymark library
Press the [update] key to select the desired library.
Custom
Click [Custom] to load the body mark pictures.
You can import a user-defined body mark of PNG and BMP (24/ 32/ 256 bit) image of
75*75 pixels.
NOTE: In the Body Mark mode, if no object is selected, pressing the [Clear] key will
clear all comments, body marks and general measurements from the screen.
Tips:
Powering off, preset returning, switching the exam/ patient/ mode/ probe will clear the
body marks.
9.2.7 Thumbnails
The stored images or cineloops are displayed in the form of thumbnails on the screen:
In the iStation screen, the thumbnails refer to the images stored for the selected
exam or the selected patient.
In the scanning or freeze mode, the thumbnails refer to the images stored in the
current exam.
In the Review screen, the thumbnails refer to the images stored in the same exam.
In the Review screen, open an image to enter the image analyzing status, all the
thumbnails belong to the exam are displayed.
When you move the cursor onto a thumbnail, its name and format will be displayed.
Select an exam of a patient in the iStation screen, and click to enter the Review
screen to review the images of the patient. You can also select more than one exam,
and the system will display images of the lately exam images in Review status.
The Review screen is shown as follows:
To exit Review:
Click [Exit] on the Review screen; or,
Press <ESC> or <Review> again.
Basic operations
Move the cursor onto an exam item in the Exam History area and press <Set>. The
selected item is highlighted. Click [Info] or [Report] to view patient information or report.
Double-click a thumbnail to view and analyze an image. Rotating the Multifunctional knob
will navigate through thumbnails.
The function buttons are described as follows:
Exam History:
You can select one certain exam from the exam directory to review the images.
Demonstration item
The demonstration items are the image files in the formats that the system supports. You
can add the exam data in patient database or system supported image files and folders to
demonstration list. For files and folders in demonstration list, the images in the directory
and subdirectory are played one by one, and the system will automatically jump over the
files that can’t be opened.
Demonstration item
There are two kinds of catalogs: Demo Catalog and Customize Catalog.
Demo Catalog: demo catalog is the folder in hard disk(E disc), where the factory
DEMO is stored. The system plays the images in this folder when performs
demonstration.
The system supports import, delete or clear the data in demo catalog.
Click [Demo Manager] to operate:
[>]: to import data to demo catalog.
In the iStation screen, select patient data, click (Restore) or (Backup) to import or
export patient information, images and reports from or to an external memory device. See
the following figure:
Review
Select an exam of a patient, click on the right side to display the patient information
of this exam.
Report
After you select an exam of a patient, click to view the report of this exam for this
patient. If no report is generated in the exam, the system prompts that “There is no report
belongs to the exam”.
Delete
Select an exam or a patient, click to delete. However, you cannot delete patient data
being printed, exported or sent, or delete the current exam.
To delete an image, select the image and click the Delete icon on the right side.
Backup/ Restore
You can back up the selected patient data to the system-supported media in order to view
it on PC, or restore the patient data to the system from an external media.
: Backup. Click to export the selected patient data to the system-supported media.
: Restore. Click to import the patient data from an external media. If no external data
source is connected, then the button is unavailable.
Select the patient record, click in the menu to send exam data or images of the
selected record.
Select the image, click the Send To Arrow to send the selected image.
Send patient exam data to USB devices, DVD drive and iStorage.
Send images to USB devices, DVD drive, DICOM storage server, DICOM printer,
video printer, text/ graph printer and iStorage.
Send images with report to USB devices, DVD drive and iStorage.
Format transfer is available when sending images to USB devices, DVD or
iStorage. See "9.2.10 Sending Image File" for details.
Press <Shift> on the control panel to select more than one exam or image at one
time.
Activate exam
After you select an exam, which has been performed within 24 hours, click to
activate the exam and load the basic patient information and measurement data to
continue the exam.
If you want to select a patient data in an external memory database to start a new exam
or recover the exam, you have to first allow the system to load the patient data to the
system’s patient database.
Continue Exam
Select an exam that is paused within 24 hours, click to activate the exam and load
the basic patient information and measurement data to continue the exam.
If you want to continue an exam which data lies in an external memory database, you
have to first allow the system to load the patient data to the system’s patient database.
Recycle bin
The recycle bin is used to store the deleted patient data, exam data and images (time
lasts from deleting to system is powered off). The system supports recovery of those data
from the recycle bin. NOTE: The recycle bin will be cleared after the system is powered
off.
To recover the deleted patient data, click at the lower right corner of the screen
(when the button is gray, the operation is unavailable) to enter the Patient Recycle Bin
screen.
(1) Select items to be recovered in the list.
(2) Select operations:
Click [Restore Items] to restore the item back to iStation;
Click [Delete] to delete the item permanently, and the item can never be restored
again;
Click [Restore All Items] to restore all the items back to iStation;
Click [Empty Recycle Bin] to empty the recycle bin and all items can never be
restored again.
Click [Exit] to exit Recycle Bin screen and return to iStation.
(2) Select the data to be backed up, click or in the screen (in iStation or
Review). Select the target drive in the Send To or Back Up Patient Record dialogue
box.
(4) After the writing process is completed, click to pop up the Disc Option dialogue
box, and select [Eject] to eject the CD/DVD.
(2) Double-click the symbol to pop up the [Disc Option] screen, as shown in the
figure below.
(3) Click [Erase] to erase data from a CD/DVD;
(4) After the erasing process is completed, click [Eject] in the Disc Option dialogue box.
Note:
Writing data using “Send To” supports PC format transfer function, while CD/DVD
written by “Backup” supports only system-relevant formats.
The symbol indicates that the input CD/DVD is damaged or with data of wrong
format.
During the backup process, if a CD/DVD is forcibly taken out or
CAUTION: you perform other operations, the backup process will fail or
the system may malfunction.
When there is/are task(s) undergoing, the task management icon displays as , you
can click the icon to check the process.
When there is/are task(s) failed, the task management icon displays as , you can
click the icon to check the failure reason.
Emergency operators are general ones, they can enter the system without entering
password. But they can’t modify or delete the password.
(2) Select the user name in the drop-down list of User Name.
(3) Enter password and click [Login].
3. Enter the user name(you are not allowed to enter the same name or modify the name
already exist).
4. Enter user name and the password.
5. Set the user role in the drop-down list: administrator or operator.
6. Click [OK] to confirm the setting and exit the dialogue box, then the new user will appear
on the User List.
(3) Enter the previous and the new password in the dialogue box.
(4) Click [OK] to exit.
Abbr. Descriptions
Digital Imaging and Communications in
DICOM
Medicine
AE Application Entity
PDU Protocol Data Unit
SCU Service Class User (DICOM client)
SCP Service Class Provider (DICOM server)
SOP Service-Object Pair
DICOM 10-1
1. Press <Setup> to enter the [Setup] menu.
2. Select [Network Preset].
10-2 DICOM
3. Preset local DICOM properties and DICOM server.
Localhost DICOM Service Property
Name Description
Application entity title of the ultrasound system.
AE Title The AE title here should be the same with the one of the acceptable
SCU set in the server.
DICOM communication port, which should be the same with the one
Port
in the server.
Maximum PDU data package size ranging from 16384 to 65536.
PDU
The default value is 32768.
Enter AE Title, Port, PDU according to the actual situation.
Note:
AE Title should be the same with the SCU AE Title preset in the server (PACS/ RIS/
HIS), for example, if the AE Title of the server preset in the print server is AAA, and
the AE Title of the accepted SCU is preset as MMM, then in the figure above, the AE
Title of Local should be MMM, and the AE Title of print server should be AAA.
DICOM Server Setting
Name Description
Device Name of the device supporting DICOM services.
IP Address IP address of the server.
DICOM 10-3
Name Description
You can ping other machines to verify connection after entering the
Ping correct IP address. Also you can check the connection of the
already added server in the list.
[Add] Click to add servers to the device list.
[Set DICOM Service] Click to enter DICOM service preset, see “10.1.3 DICOM Service”.
[Delete] Click to delete the selected server (s) in the device list.
Note:
If the currently entered name has already existed, the system will pop up: “The server
name exists!” Click [OK] to enter another name.
Enter [Setup]-> [DICOM Preset], Click [Set DICOM Service], you can add, delete and set the
property of the service on DICOM server.
10.1.3.1 Storage
1. Enter [Setup]-> [DICOM Preset]-> [Set DICOM Service]-> [Storage].
2. Select device, enter the information. For device setting, please refer to “10.1.2 DICOM
Preset.”
Click [Add] to add the service to the Service list;
Select an item in the service list, change the parameters in the above area, and click
[Update] to update the item in the service list; click [Cancel] to cancel changing.
Select an item in the service list, and click [Delete] to delete the service.
Select an item in the service list, and click [Default] to set the server to be the default
service.
3. Select an item in the service list, and click [Verify] to verify the connection.
4. Click [Exit] to save and exit the preset.
10-4 DICOM
DICOM storage setting items are described as follows:
Name Description
After you set the server (s) in DICOM Server Setting, the name
Device (s) will appear in the drop-down list, select the name of the
storage server.
Service Name Default is xxx-Storage, and it can be modified.
Application Entity title, Here, it should be consistent with that of
AE Title
the storage server.
DICOM communication port, 104 by default. Here, the port should
Port
be consistent with that of the storage server port.
Maximum Retries: Range: 0-9.
Interval Time(s): Reserved feature.
Advanced Preset Timeout: Refers to time after which the system will stop trying
to establish a connection to the service. Value: 5-60s, in
increments of 5s, and 15 by default.
Color Mode Gray/ Mixed/ Color
Select the compression mode: uncompressed, RLE, JPEG and
Compression Mode
JPEG2000.
Compression Ratio Select the compression ratio: lossless, low, medium and high.
Allow Multiframe If SCP supports this function, then select it.
SR Storage Option To enable or disenable structured reporting sending.
[Add] Add the DICOM service to the service list.
[Cancel] Click to cancel the parameter setting.
DICOM 10-5
Name Description
Select an item in the service list, change the parameters in the
[Update] above area, and click [Update] to update the item in the service
list.
[Delete] Click to delete the selected service in the service list
Select an item in the service list, click [Default] and you can see
[Default]
“Y” in the Default column.
Click to verify if the two DICOM application entities are normally
[Verify]
connected.
[Exit] Click to exit the screen.
Tips: RLE, JPEG and JPEG2000 are not supported by all SCPs. Please refer to the electronic
file DICOM CONFORMANCE STATEMENT of the SCP to check whether SCP supports or not.
Do not select these compression modes if the storage server doesn’t support them.
10-6 DICOM
DICOM print setting items are described as follows:
Name Description
After you set the server (s) in DICOM Server Setting, the
Device name (s) will appear in the drop-down list, select the name
of the print server.
Service Name Default is xxx-Print, and it can be modified.
Application Entity title. Here, it should be consistent with that
AE Title
of the print server.
DICOM communication port, 104 by default. Here, the port
Port
should be consistent with that of the print server port.
Maximum Retries: Range: 0-9.
Interval Time(s): Reserved feature.
Advanced Preset Timeout: Refers to time after which the system will stop
trying to establish a connection to the service. Value: 5-
60s, in increments of 5s, and 15 by default.
Refer to copies of printed files. You can select among 1
Copies
through 5, or directly enter the number.
The system supports RGB(color printing) and
Settings MONOCHROME2 (black and white printing). Please select
the type the printer supports.
Film Orientation Select between LANDSCAPE and PORTRAIT.
Priority Specify printing task priority among HIGH, MED and LOW.
Select film size among the selections listed in the drop-down
Film Size
list.
DICOM 10-7
Name Description
Specify quantity of printed files,e.g. STANDARD\2, 3
Display Format
indicates 6 images are printed for each page.
Medium Type Specify print medium: Paper, Clear Film, Blue Film.
Specify whether you want a trim box to be printed around
Trim
each image on the film: Yes or No.
Min Density Enter the minimum density of the film
Max Density Enter the maximum density of the film.
Configuration Info Enter configuration information in the field.
Specify where the file is exposed:
Destination MAGAZINE (stored in the magazine), or
PROCESSOR (exposed in the processor)
Replicate: interpolated pixels belong to duplicate of adjacent
pixels;
Bilinear: interpolated pixels are generated from bilinear
interpolations between adjacent pixels;
Magnification Type
Cubic: interpolated pixels are generated from cubic
interpolations between adjacent pixels; None: without
interpolation.
Select how the printer magnifies an image to fit the film.
[Add] Add the DICOM service to the service list.
[Cancel] Click to cancel the parameter setting.
Select an item in the service list, change the parameters in
[Update] the above area, and click [Update] to update the item in the
service list.
[Delete] Click to delete the selected service in the service list.
Select an item in the service list, click [Default] and you can
[Default]
see “Y” in the Default column.
Click to verify if the two DICOM application entities are
[Verify]
normally connected.
[Exit] Click to exit the screen.
Parameter setting should be consistent with the printer performance:
For instance, if the printer cannot support film size to be “8IN*10IN”, please select the
printer supported size.
Settings: RGB is color printing and MONOCHROME2 is black and white printing.
Medium Type: for black and white printing, usually Blue Film or Clear Film is used,
and for color printing, Paper is usually used. For details, please refer to printer
conformance statement.
Change the setting if necessary.
10-8 DICOM
DICOM Worklist service parameters are similar to those described in DICOM Storage Preset,
please refer to “10.1.3.1 Storage” for details.
Special setting item for DICOM Storage Commitment service is Associated Storage Service,
as described in the following; other parameters are similar to those described in DICOM
Storage Preset, please refer to “10.1.3.1 Storage” for details.
Name Description
The associated storage server should be preset before storage
Associated Storage
commitment, only after the exam is sent out, can storage
Service
commitment be created.
DICOM 10-9
The server supports the verification, but this function is not activated. Please check if
the verification function is activated.
Note:
Not all the SCPs can support verification; please consult SCP belongings to confirm whether
SCP can support this service. If not, the verification won’t pass.
(3) Select DICOM in the “Target” list, select a server in the “Storage Server” list.
(4) Click [OK] to start the sending.
10-10 DICOM
To send image for storage after an exam ends
(1) Open “[Setup] → [System Preset] → [General]”, and then check
.
(2) Set a default storage server.
a) Enter the DICOM Service Preset screen via “[Setup] → [DICOM Preset] → [Set
DICOM Service]”.
b) Select a storage server in the Service List and click [Default]; you can see “Y” in
the Default column.
c) Click [Exit] to exit the page and return to Setup menu, then click [Save] on the
Setup menu to make the preset to take effect.
(3) After finishing resets, you can perform image scanning, each time you press <End
Exam> on the control panel, the system will send the image to the default DICOM
storage server for storage.
DICOM 10-11
To query patient information via Worklist server:
(1) Press <Patient> to enter Patient Info screen.
(2) Click [WorkList] to enter the WorkList page.
10-12 DICOM
(4) Press <Patient> to enter Patient Info screen.
(5) Click [WorkList] to enter the WorkList page.
(6) The system queries intraday patients via Worklist server automatically and the patient
records will appear in the list.
In the off-line status, you can:
Perform a second query; or,
Click [Show Detail] to view the patient information in details:
10.3.4 MPPS
MPPS is used to send exam state information to the configured server. This will facilitate the
other systems to obtain the exam progress in time.
The status information is described as below:
When you begin an exam or send image(s) during the exam, the system sends
status information “Active” to MPPS server.
When the exam is completed, the system sends status information “End” to MPPS
server.
When a paused exam is continued, the system sends status information “Active” to
MPPS server.
When an exam is cancelled, the system sends status information “Cancelled” to
MPPS server.
(2) Select an exam record (image (s) is/are stored in the exam record), click to
open the Send To dialogue box.
(3) Click to select “DICOM” in the “Target” box on the left side, then select the DICOM
storage server in the “Storage Server” box on the right side.
(4) Click [OK] to start sending. The system will send all the images stored in the exam
record to the storage server, meanwhile, it will send storage commitment to the
storage commitment server.
To send storage commitment after an exam ends
(1) Open [Setup] -> [System Preset]-> [General], and then select [Sending/ Printing after
End Exam] in the Patient Info area.
(2) Set the default storage server and storage commitment server.
a) Enter the DICOM Service Preset screen via “[Setup] -> [DICOM Preset] -> [Set
DICOM Service]”.
b) Select a server in the Service List of Storage page and Storage Commitment
page and click [Default]; you can see “Y” is marked in the Default column.
c) Click [Exit] to exit the page and return to Setup menu, then click [Save] on the
Setup menu to make the preset to take effect.
(3) When finished the presets, you can perform image scanning, each time you press
<End Exam> on the control panel, the system will send the image to the default
DICOM storage server for storage and send storage commitment to storage
commitment server.
DICOM 10-13
If images are successfully sent to the storage server, the storage commitment server will
return information about the successful image storage. In the iStation screen, you will see
there is a tick “√” marked in the list below .
Tips:
Storage commitment is confined to the whole exam; not each image sending can be indicated.
NOTE: Multi-frame storage is not allowed if “Allow Multiframe” is not selected ([Setup]->
[DICOM Preset] -> [Set DICOM Service] -> [Storage]). For example, if there is
multi-frame file in the exam to be sent, then only single-frame image storage will be
performed, and after the storage is completed, there is no “√” marked in the list
below in the iStation screen.
10.3.6 Query/Retrieve
The query/retrieve function is employed to query and retrieve the patient exam records in a
designated server.
After setting the DICOM query/retrieve server, you can perform the query/retrieve function in
iStation screen.
1. Open iStation screen: press <iStation> on the control panel; or, press <Patient> on the
control panel, and then click [iStation] on the Patient Info screen.
2. Click [Query/Retrieve] to open the screen.
3. Select the server in “Server and Service” area (both the source and the destination).
4. Enter the query information, such as Patient ID, Patient Name, Accession #, Exam Date,
or key words.
Click [Clear] to empty the entered query information.
5. Click [Query], the system performs the query and lists out the results into the patient
(source) list.
10-14 DICOM
You can perform further query basing on the results by entering new query information.
6. Select one or more patient records according to the actual situation.
Click [Select All] to select all the patient records in the list.
Click [Deselect All] to deselect all the patient records in the list.
7. Click [Retrieve] to retrieve the patient records in the DICOM query/retrieve server into the
local machine.
8. Click [Exit], you can see the retrieved patient records are listed in the iStation screen.
DICOM 10-15
Data Restore:
1. If the DICOM format data is backed up to external media, you can restore the data to the
system from the media.
2. Review the data stored in the external media.
3. Select the data to be restored in iStation.
10-16 DICOM
11 Setup
The Setup function is designed to set the configuration parameters of operating the system
and maintaining user workflow setup data. The setup data of the user and system are stored
to the hard drive, and should be backed up to CD/DVD or USB memory devices.
When the setup data is changed, be sure to save the preferences
CAUTION: according to the methods described in this chapter. Mindray is
not responsible for the loss of the setup data.
To enter Setup:
Press the <Setup> to enter Setup.
To exit Setup:
Click [Save] on the Setup page to close the Setup menu, and the parameter settings is
saved.
Click [Cancel] or press <Esc> on the control panel to exit the setup.
11.1.1 Region
Open the Region page via [Setup]-> [System Preset]-> [Region].
Item Description
To set the hospital relevant information like name, address, telephone
Hospital Information
and so on.
Setup 11-1
Item Description
To select a language for the system, the available languages are
Chinese, English, French, German, Italian, Portuguese, Russian,
Spanish, Polish, Czech, Turkish, Finnish, Danish, Icelandic,
Language Norwegian, and Swedish.
The system will restart automatically after you change the language
and return from the Setup menu.
Time Zone To select the time zone.
To set the date format among DD/MM/YYYY, MM/DD/YYYY and
Date Format
YYYY/MM/DD.
Time Format To select the time format between 12Hour and 24 Hour.
To set the system date.
Posit the cursor into the corresponding field, and enter the data
System Date
through the keyboard; or, click the calendar icon , and then select
the date.
Enter time to the text box; or modify the time via spinner buttons on
System Time
the right side.
11.1.2 General
Open the General page via [Setup]-> [System Preset]-> [General].
11-2 Setup
Type Item Description
To select if to display the following patient information
Info displays in
on the image banner: Gender, Age, Operator, ID,
image banner
Name, Hospital
Patient Info
H&W Unit To set the unit for patient height and weight.
Surface Formula To set the surface formula.
To set the size for a stored image:
Image Area
Image Output Size
Standard Area
Other Full Screen
Clip length/ Time
Times: 1~60s
(s)
Output Mode PAL/ NTSC
Patient Sending/printing Select if to automatically archive the exam data to
Management after End Exam DICOM server for storage/print.
Enable To enable screen saver function, and select the waiting
ScreenSaver time until screen saver.
Screen Enable Standby Available after “Enable ScreenSaver” is selected.
Saver
After screen saver function is enabled, click [Browse] to
Select Picture select the figure used for screen saver, and click
[Preview] to see the effect.
Status after exam Select to enter the imaging scanning, patient info or the
Exam Setup
ends Worklist after ending an exam.
Color temperature Cold/ Warm
Display Brightness/Contrast Load the factory data of the display brightness and
Load Factory contrast.
Setup 11-3
Type Item Description
Reset Config Probe To set the default probe model for the system.
To set MI TI indexes displayed for current probe/exam
/ MITI
mode.
Freeze
Status after Freeze To set the system status after image is frozen.
Config
To set the steer mode in B+ Color +PW imaging mode.
C&PW: select to adjust the sample volume in color
Steer mode and sample line in PW mode together.
C/PW: select to adjust the sample volume in color
mode and sample line in PW mode individually.
Image The spectrum can automatically invert when the color
Auto Invert flow is steered to a certain angle, thus accommodating
operator’s habit of distinguishing flow direction.
Cine Memory To set the cine memory splitting type.
iScape Ruler To set if to display the iScape ruler on iScape imaging
Display mode.
11.1.4 Application
Open the Meas page via [Setup]-> [System Preset]-> [Application]. Through the page, you
can set the measure ruler, follicle and relevant information. For details, please refer to the
Operator’s Manual [Advanced Volume]
11-4 Setup
11.1.5 OB
Open the OB page via [Setup]-> [System Preset]-> [OB]. Through the page, you can set the
gestational age formula, fetal growth formula, fetal weight formula and the relevant
information. For details, please refer to the Operator’s Manual [Advanced Volume].
The following introduces an example as setting <F1> as “Send Cine to USB disk”:
(1) Select F1 in the “Key Function” list at the left side of the page.
(2) Select “Send Cine to USB disk” in “Function” at the right side of the page.
(3) Click [Save] to complete the setting.
Setup 11-5
Other Settings
Item Description
Key Brightness To set the lightness for the keys.
Key Volume To set the key volum. Zero means no sound.
Trackball Speed To set the trackball speed when move the trackball.
Trackball Brightness To set the color for the trackball.
11.1.7 Admin
Open the Admin page via [Setup]-> [System Preset]-> [Admin].
For details of access control, please refer to “9.7 Access Control”.
Procedures:
1. To select a probe:
Move the cursor to [Probe], and select a probe from the drop-down list.
2. Selecting/delete exam modes
On the right side of the screen, you can view the exam types supported by the current
probe. On the left side, you can view all the exam modes supported by the system, i.e.,
Exam Library.
11-6 Setup
[>]: add a selected exam mode in the [Exam Library] to the [Exam Selected] list.
[>>]: add all exam modes in the library to the [Exam Selected] list.
[<]: remove an exam mode selected from the [Exam Selected] list.
Click [Up] and [Down] to adjust the sequence of the items in Selected Items.
Click [Default] to set a selected exam mode as the default exam mode.
3. Click [Save] to confirm the modified setup; or click [Cancel] to cancel the modified setting.
Procedures:
1. Select an exam mode.
2. Enter Library Name: you can enter characters for the library name, or accept the default
name (the same as that for the exam mode).
3. Select an available item.
4. Add user-defined comments: directly enter comment texts, or select comment texts for
the comment library.
Directly enter comment texts: position the cursor in the field box above [Add
Comment], enter the text comment through the keyboard, and then click [Add
Comment]. Then the directly-entered comment will be added to the Available Items
and Selected Items.
Setup 11-7
Select Available Items: First select a comment library in the drop-down list beside
“Available Items”, and then click [Set] on one item displayed below “Available Items”.
Click [>] to add the item in Available Items on the left into Selected Items on the
right.
Click [>>] to add all items in Available Items on the left into Selected Items on the
right.
5. Change position of the selected items: select an item on the right side box and click [Up],
[Down], [Left] or [Right] button to change the position of the item.
6. Delete comments:
Remove item (from the library or user-defined) in the Selected Items list:
Select an item in Selected Items list, and click [<] to remove it to the Available Items
list.
Click [<<] to remove all items in Selected Items.
Delete a user-defined item in the Available Items box. You can only delete the user-
defined items rather than the items in the system library. After a user-defined item is
deleted, it will not be available.
Select a user-defined item in the Available Items box, and click [Delete]. After the
user-defined item in the Available Items box is deleted, the same item in the Selected
Items box on the right will disappear as well.
7. After you customize comments, click [Save] to confirm and exit the [Comment Preset]
screen.
11-8 Setup
To preset body mark for exam mode:
1. Select an exam mode in the drop-down list, the current exam mode is set by default.
2. Enter the library name of custom body marks.
3. Select a package from the drop-down list besides Available Item.
4. Add or delete body mark(s)
To add the item selected from the [Available Items] into the [Selected
[>]
Items].
[>>] To add all items from the [Available Items] into the [Selected Items].
To remove the selected item in [Selected Items] to the [Available
[<]
Items].
[<<] To remove all items from the [Selected Items].
5. Click [Up] and [Down]: to adjust the sequence of the items in Selected Items.
6. Preset the probe mark of the selected body mark in Selected Items
Move the cursor onto the probe mark of the body mark in Probe area and press <Set> to
make the cursor turn green. Roll the trackball to change the position of the probe, rotate
the multifunctional knob to change the angle, and press <Set> to confirm the orientation
of the probe.
7. Select the body mark in Selected Items and click [Default], and then the body mark will be
marked with a tick ”√”.
8. Click [Save] to complete the setting.
Setup 11-9
Print Service Setting
Add Service: click to begin print service adding.
Remove Service: click to delete the selected print service.
Rename Service: click to rename the selected print service.
Property: to preset the property of print services.
Printer Driver Setting
Printers listed in the operator’s manual are all supported by the system, no drive is
required.
Click [Print Driver] page to enter printer driver setting screen:
Display the printer name as well as print status.
You can add printer including network printer.
Check the printer attribute.
For details about DICOM print, please refer to “10 DICOM”.
Printer Adapter
For details, please refer to the “Appendix D Printer Adapter”.
Name Description
Service Name Name of the device, cannot be empty
Port Port for transmitting.
11-10 Setup
Name Description
IP address of the PC installed with iStorage software, cannot be
IP Address
empty
Press to verify connection with the PC server. On PC server, if the
storage path has not been confirmed, a dialog box will pop up and
Connect guide the user to set it. If the storage path on PC has already been
set, here it displays connection successful after clicking this button.
Default Click to set an selected item in the list as the default iStorage server.
11.8 Maintenance
In the Setup menu, select [Maintenance] enter the screen.
11.8.1 Option
The system enters Option page after you enter the Maintenance screen. In Option list, the
system lists out all the system supported options and the installation status (Not installed or
Installed).
Install and uninstall
Click [Install] to begin the installation of a disabled option.
Click [Uninstall] to begin the un-installation of an already installed option.
Trial
Select the option item in the list to see the information on the Comment region on the
right. For functions that are not configured, you can use the function for free for 3 months.
Click to select the function(s) and click [Trial].
Tips: Every trial option can be used only once. Please refer to the dependence
relationship between options in chapter “2.5.3 Options”.
Please contact Mindray Customer Service Department or sales representative for details.
Setup 11-11
11.8.2 Other Settings
Type Item Description
Export Log Export the operation log.
Log
Upload Log Upload the operation log.
Load Factory Local factory default settings.
Preset Manager Export Export preset data to disk.
Import Import the preset data into the system.
11-12 Setup
12 Probes and Biopsy
12.1 Probe
The system supports the following probes:
No. Probe model Illustration
1. 35C50EA
2. 65C15EA
3. 65EC10EA
4. 75L38EA
5. 10L24EA
6. 35C20EA
7. 35C50EB
8. 75L38EB
9. 65EC10EB
10. D6-2EA
Note: For details of storage time and condition for disinfected probes or sterilized probes
and brackets, please refer to Technical standard for Disinfection of Medical and Health
Structures
<1>
<3>
<4>
<5>
Mark
Examinations
Biopsy procedure
Draining/drying
Draining/drying
Storage
Examinations
Washing the
transducer with water
Draining/drying
Disinfection
Draining/drying
Storage
3. Secure the sheath with enclosed elastic 4. Inspect the sheath to ensure there are no
bands. holes or tears.
NOTE: 1. After the examination, wipe off the ultrasound gel thoroughly. Otherwise, the
ultrasound gel may solidify and degrade the image quality of the transducer.
2. DO NOT make the probe to become overheated (more than 55°C) during
cleaning and disinfections. High temperature may cause the probe to become
deformed or damaged.
Cleaning
Please refer to the instructions in the manual and follow your hospital policy and procedures
for cleaning.
1. Disconnect the probe from the system.
2. Wear sterile gloves to prevent infection.
3. Wash the transducer with clean water or soapy water to remove all the foreign matters, or,
wipe the transducer with a soft ethyl carbamate sponge. Avoid using a brush, because it
may damage the transducer.
4. Dry the transducer using a sterile cloth or gauze after rinsing. Do not dry the transducer
by heating it.
Strain relief
NOTE: Observe the graph here carefully to perform disinfection. Do not spray the strain
relief on the connector end or the connector.
Disinfecting by Immersion
1. Wear sterile gloves to prevent infection.
2. Clean the transducer before disinfecting it.MINDRAY recommends the following solutions
to disinfect the transducer.
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the disinfectant solution, method of disinfection and dilution and
cautions during use.Do not soak the transducer connector or the cable near it into
water or any solution.
Soak the transducer into the disinfectant solution for the shortest time the
manufacturer recommends (for example, the shortest time recommended by the
manufacturer for soaking Cidex OPA is 12 minutes).
Follow local regulations when selecting and using the disinfectant.
3. Rinse the transducer with plenty of sterile water (about 2 gallons) for at least 1 minute to
remove all chemical residues on it.Or, follow the rinsing method recommended by the
disinfectant manufacturer to rinse the transducer.
4. Wipe off the water on the transducer with sterile cloth or gauze after rinsing it.Do not dry
the transducer by heating.
NOTE: 1. Observe the graph here carefully to immerse the transducer. Only soak parts
of the transducer below the strain relief.
2. Repeated disinfection will eventually damage the probe, please check the
probe performance periodically.
Compatible Disinfectants
Please see Mindray Transducer Disinfectant Recommendation for the disinfectant use in
details.
Biopsy
Target
Probe
Needle
Target
Ultrasound beam
Name of Parts
This section describes the parts and corresponding functions of each needle-guided bracket.
Needle guide
hole
Clamp
Locating pit
Needle guide rack
Locating
groove
Grip knob
Needle-guided
bracket
Transducer
Needle guide
Retaining clamp
Locating bulge
Transducer
Locating groove
NGB-016
No Name Description
Clamp of needle-guided Used for installing the needle-guided bracket on the
<1>
bracket transducer.
Groove of the needle-
<2> Matches with the tab of the transducer.
guided bracket
<3> Angle adjusting base There are 3 types of angles available to be adjusted.
Angle shift sign (30°, 40°, Matches with the biopsy angle
<4>
50°) (30°, 40°, 50°).
NGB-001
1. Put on the sterile transducer sheath.
2. Hold the transducer by one hand, select the proper needle-guided bracket, and hold
it with the other hand. Match the groove of the bracket with the tab of the transducer.
Amount the bracket onto the transducer.
3. Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the transducer.
(3) Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the transducer.
(4) Select a proper guiding block and push it into the groove above the angle block.
NGB-004
1. Put on the sterile probe sheath.
3. When the retaining clamp is turned to the right position, the locking nut will lock the
retaining clamp and the needle-guided bracket is then mounted to the right position.
NGB-016
1. Put on the sterile transducer sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the
transducer. Mount the bracket onto the transducer.
3. Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the transducer.
4. Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
Ensure that all guide parts are seated properly prior to performing
CAUTION: a biopsy.
Figure 1 Figure 2
To enter/exit iNeedle
To enter iNeedle
Click [iNeedle] item in B page on the screen.
Or, you can assign a user-defined key for entering iNeedle.
Open iNeedle in Biopsy status
1. Perform scanning and locate the target, press <Biopsy> to enter the screen.
2. Click [iNeedle] to enter the status; available adjusting parameters are displayed on the
menu.
To exit iNeedle
Press the user-defined key or click [iNeedle] to exit the status and enter B mode.
Needle Steer
Description This function adjusts the biopsy needle angle via changing the steer angle of
the scan line. The iNeedle affecting region changes correspondingly.
Operation Click [Needle Steer] item on the screen.
B/iNeedle
Description This function is used to display B image and iNeedle image synchronously.
Operation To turn on or off the function, click [B/iNeedle] on the screen.
Tips: iZoom (full-screen magnification) is available on iNeedle status.
NOTE: 1. You can perform guide line verification on a single live B image only, and
all biopsy-irrelevant operations are forbidden.
2. For bi-planar probe applied biopsy, the verification is performed on the first
guide line, the other guide lines can move together with the first one in
parallel.
1. Confirm that the needle-guided bracket has been installed securely in the correct position.
2. Prepare a container filled with sterile water.
3. Place the head of the transducer in the sterile water, and get a biopsy needle into the
needle guide.
4. When the biopsy needle appears on the image, please adjust the parameters of the
menu on the screen to make the biopsy needle is displayed at almost the same position
as the selected needle mark.
(2) Separate the residual part of the needle-guide bracket and the transducer from the
needle.
NGB-004
Hold the transducer in the left hand, unscrew the locking nut with the right hand to open
the retaining clamp, and then raise the needle-guided bracket to separate the locating
bulge from the locating grooves.
NGB-016
1. Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
2. Separate the residual part of the needle-guide bracket and the transducer from the
needle.
3. Screw the pinch nut of the bracket, and remove the needle-guided bracket from the
transducer.
Sterilization
1. Wear sterile gloves to prevent infection.
2. Clean the needle-guided bracket before sterilizing it. MINDRAY recommends the
following solution or sterilizing system to sterilize the needle-guided bracket.
3. Follow local regulations when selecting and using the sterilant.
12.2.10 Disposal
Be sure to dispose the needle-guided bracket only after sterilizing it.
Contact your MINDRAY representative when disposing of this device.
12.3 Lithotrity
“Lithotrity” helps to locate the focus point of lithotrity wave during lithotrity treatment. By
watching the procedure of lithotrity in real-time and adjusting the intension and frequency of
the lithotrity wave, the harm to the patients can be reduced to the least.
To enter lithotrity mode: Click [Lithotrity] in the image menuto turn on the lithotrity.
In lithotrity mode:
The lithotrity line is a vertical dotted line located in the middle of the screen, the
position and direction of which cannot be changed.
The lithotrity mark is a “×” mark located on the lithotrity line which can be moved up
and down along the lithotrity line by rolling the track ball.
The depth of the mark is displayed in the image parameter area of the screen.
13.1 Overview
The battery charges when the system is connected to the AC power supply.
Generally, if the system is turned off, a completely discharged battery will fully charge
in less than 3 hours.
NOTE: 1. It is recommended to charge the batteries when the system is turned off,
thus you can quickly charge the battery and save time.
2. Power off the system if you will not use the system for a long period of time
(including storage/ transportation condition), and you should not allow the
system in standby status, otherwise the batteries will be out of power and
permanently damaged.
When the external power supply is not connected, a lithium-ion battery will provide the power.
The model of the chargeable lithium-ion battery (hereinafter called battery for short) is
LI34I002A.
Battery 13-1
13.2 Precautions
1. Before using the battery, carefully read the description in the label on the surface of the
battery.
2. When you use the battery at the first time and find that it is dirty or emit an odor, do not
use it.
3. Do not expose the battery to heat sources or high pressure.
4. Do not leave the battery in direct sunlight.
5. Store the battery out of the reach of children. Store the battery out of the reach of children.
6. The battery is designed to be charged only in this system; charge the battery only when
the ambient temperature is between 0 and 40 ℃.
7. When the battery is out of power, please charge it immediately.
8. When the battery will not be used for 10 days or over, take it out of the system and store it
with caution. DO NOT wrap it with conductive material; store it in shady, cool and dry
place.
9. Do not knock the battery.
13-2 Battery
Warning! Battery is out of power! Please connect to power supply or system will be shut down
in one minute. Connect the power supply to afford normal work.
Battery 13-3
14 Acoustic Output
This section of the operator’s manual applies to the overall system including the main unit,
probes, accessories and peripherals. This section contains important safety information for
operators of the device, pertaining to acoustic output and how to control patient exposure
through use of the ALARA (as low as reasonably achievable) principle. Also this section
contains information regarding the acoustic output testing and the real-time output display.
Read this information carefully before using the system.
fawf CMI
NOTE: 1 The system needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
2 Other devices may interfere with this system even though they meet the
requirements of CISPR.
3 Preventing conducted RF immunity. Due to technological limitations, the
conducted RF immunity level are limited to 3Vrms level, conducted RF
interference above 3Vrms may cause wrong diagnosis and measurements. We
suggest that you position system further from sources of conducted RF noise.
4 Portable and mobile RF communications equipment can affects system. See
tables 1, 2, 3, and 4 below.
TABLE 1
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
ELECTROMAGNETIC ENVIROMENT-
EMISSIONS TEST COMPLIANCE
GUIDANCE
The system uses RF energy only for its
internal function. Therefore, its RF
RF emissions emissions are very low and are not likely
Group 1
CISPR 11 to cause any interference in nearby
electronic equipment.
RF emissions
Class B
CISPR 11 The system is suitable for use in all
establishments including domestic
Harmonic Emissions
Class A establishments and those directly
IEC 61000-3-2 connected to the public low-voltage
Voltage Fluctuations/ Flicker power supply network that supplies
Emissions Compliance buildings used for domestic purposes
IEC 61000-3-3
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular
/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the
location in which system is used exceeds the applicable RF compliance level above,
system should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
system.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Cable sample
No. Name Cable length (m) Shield or not Remarks
1 Power input 2.5m Not shielded /
2 SIP/SOP <3.0m Shielding /
3 ECG cable 4.0m Shielding /
4 Probe Cable <3.0m Shielding /
5 Footswitch Cable 2.9m Shielding /
Clamp ring
Track ball
Rotate clamp ring 35 degrees anticlockwise Remove clamp ring Draw out the ball
b) Cleaning
Clean the lens slightly using the tissue until there is no foreign matter. Then clean the other
parts and the dust excluder. Make sure to exert force properly when clean the small ball,
otherwise, the small ball may drop down. See the figure below. No system shutdown is
required during the whole clean. Reinstall the ball and clamp ring after the clean is finished.
Lens
Small ball (x3)
Waterspout
When liquid is spray onto the trackball area, most of the liquid can be discharged through the
waterspout, besides, you can dry the leaving water using the tissue or cloth.
c) Installing the trackball
Put back the ball, align the buckle with gap of the front cover, press the clamp ring with both
hands and rotate it 35 degrees clockwise to click the buckle. At this moment, the clamp ring
cannot be moved any further, it indicates that the clamp ring clicks in position. See the figure
below. See figure below.
Before installing Turn 35 degrees Installed
clockwise
Buckle (x2)
16.4 Troubleshooting
To ensure proper system operation and function, it is recommended that a maintenance and
inspection plan be established to periodically check the safety of the system. If any system
malfunction is experienced, contact Mindray Customer Service Department or sales
representative.
If any persistent system malfunction is experienced, e.g. an onscreen error message, blank
imaging screen, absent menus, please refer to the following table below. If the failure cannot
be eliminated, please contact Mindray Customer Service Department or sales representative.
All tests can be performed using commercially available safety analyzer test equipment. These
procedures assume the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer.
Other popular testers complying with IEC 60601-1 used in Europe such as Fluke, Metron, or Gerb may
require modifications to the procedure. Follow the instructions of the analyzer manufacturer.
The electrical safety inspection should be periodically performed every two years The safety
analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of line voltage and
grounding, as well as total current loads.
The following outlet conditions apply when performing the Earth Leakage test.
normal polarity( Normal Condition),
reverse polarity( Normal Condition),
LIMITS
For UL60601-1,
300 μA in Normal Condition
1000 μA in Single Fault Condition
For IEC60601-1,
500 μA in Normal Condition
1000 μA in Single Fault Condition
LIMITS
For UL60601-1,
100μA in Normal Condition
300 μA in Single Fault Condition
For IEC60601-1:
100μA in Normal Condition
500 μA in Single Fault Condition
The following outlet conditions apply when performing the Patient Leakage Current test.
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity
LIMITS
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).
LIMITS
Note: Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
Follow the instructions of the analyzer manufacturer.
B.1.3 Setting
The reader has factory settings; please refer to A.4 for details.
The reader supports some user-defined functions as introduced below.
For more details, please contact the SYMBOL reader agents or Mindray Customer Service Department.
Volume setting:
Scan the following barcode to set the volume parameter.
I 2 of 5 symbols setting:
Select this option to decode only I 2 of 5 symbols containing a selected length. Select the length using
the numeric barcodes below. For example, to decode only I 2 of 5 symbols with 8 characters, scan I 2
of 5 - One Discrete Length, then scan 0 followed by 8.
3. Upon successful decode, the reader beeps and the LED turns green.
Tips: Do not hold the reader directly over the barcode. Laser light reflecting directly back into the reader
from the barcode is known as specular reflection. This specular reflection can make decoding difficult.
Screw Mount
1. Position the assembled base on a flat surface.
2. Screw one #10 wood screw into each screw-mount hole until the base of the stand is secure
Tape Mount
1. Peel the paper liner off one side of each piece of tape and place the sticky surface over each of the
three rectangular tape holders.
2. Peel the paper liner off the exposed sides of each piece of tape and press the stand on a flat
surface until it is secure.
Scanning in Hands-Free Mode
When the reader is seated in the stand’s “cup”, the reader’s built-in sensor places the reader in hands-
free mode. When you remove the reader from the stand it operates in its normal hand-held mode.
Interleaved 2 of 5 (ITF)
Interleaved 2 of 5 (ITF) Enable Enable
Set Lengths for I 2 of 5 14
I 2 of 5 Check Digit Verification Disable
B.3 Maintenance
Cleaning the exit window is the only maintenance required. A dirty window can affect scanning accuracy.
Do not allow any abrasive material to touch the window
Remove any dirt particles with a damp cloth
Wipe the window using a tissue moistened with ammonia/water
Do not spray water or other cleaning liquids directly into the window.
C.1 Overview
The major objective of ultrasound workflow automation (iWorks) is to speed up exam time and reduce
the excessive number of user interface manual key strokes that can lead to repetitive motion injuries
over time. It automates a clinical work flow in common exam protocols in a logical, “step by step”
manner. It also prevents missing an important part of an examination as well as decreases exam time.
A Protocol Event contains series workflow events (annotation comments, body mark and
measurements) and image modal commands to assist the user in routine ultrasound examinations.
The system provides different protocol events based on the different application regions.
iWorks is an option.
1 2 3
Name Description
2 The current active view, with green solid frame around the image.