Digital Storage Oscilloscope Quick Guide
Digital Storage Oscilloscope Quick Guide
Quick Guide
Table of Contents
1. General Safety Requirements .......................................................................................... 1
2. Safety Terms and Symbols ............................................................................................... 2
3. Quick Start ....................................................................................................................... 4
Introduction to the Structure of the Oscilloscope..................................................................... 4
Front Panel ............................................................................................................................................. 4
Rear Panel ............................................................................................................................................... 5
Control Area ........................................................................................................................................... 5
i
1.General Safety Requirements
1
2.Safety Terms and Symbols
Safety Symbols
Symbols on the product. The following symbol may appear on the product:
Test Ground
2
2.Safety Terms and Symbols
To avoid body damage and prevent product and connected equipment damage, carefully
read the following safety information before using the test tool. This product can only be
used in the specified applications.
Warning:
The two channels of the oscilloscope are not electrically isolated. The channels should
adopt a common ground during measuring. To prevent short circuits, the 2 probe
grounds must not be connected to 2 different non-isolated DC levels.
The diagram of the oscilloscope ground wire connection:
Signal Input
Power Cord
Ground Clip
Warning:
To avoid fire or electrical shock, when the oscilloscope input signal
connected is more than 42V peak (30Vrms) or on circuits of more than
4800VA, please take note of below items:
Only use accessory insulated voltage probes and test lead.
Check the accessories such as probe before use and replace it if
there are any damages.
Remove probes, test leads and other accessories immediately after
use.
Remove USB cable which connects oscilloscope and computer.
Do not apply input voltages above the rating of the instrument
because the probe tip voltage will directly transmit to the
oscilloscope. Use with caution when the probe is set as 1:1.
Do not use exposed metal BNC or banana plug connectors.
Do not insert metal objects into connectors.
3
3.Quick Start
3. Quick Start
Introduction to the Structure of the Oscilloscope
This chapter makes a simple description of the operation and function of the front panel of
the oscilloscope, enabling you to be familiar with the use of the oscilloscope in the
shortest time.
Front Panel
The front panel has knobs and function buttons. The 5 buttons in the column on the right
side of the display screen are menu selection buttons, through which, you can set the
different options for the current menu. The other buttons are function buttons, through
which, you can enter different function menus or obtain a specific function application
directly.
1 2 3
7 6 5 4
Figure 3-1 Front panel
1. Display area
2. Menu selection buttons: Select the right menu item.
3. Control (button and knob) area
4. Probe Compensation: Measurement signal (5V/1kHz) output.
5. Signal Input Channel
6. USB Host port: It is used to transfer data when external USB equipment connects to
the oscilloscope regarded as "host device". For example: Saving the waveform to USB
flash disk needs to use this port.
7. Power on/off
4
3.Quick Start
Rear Panel
Control Area
6 1
3
5
4
5
3.Quick Start
1. Function button area: Total 6 buttons.
2. Horizontal control area with 1 button and 2 knobs.
"HOR" button refer to horizontal system setting menu, "Horizontal Position" knob
control trigger position, " Horizontal Scale" control time base.
3. Trigger control area with 2 buttons and 1 knob.
The Trigger Level knob is to adjust trigger voltage. Other 2 buttons refer to trigger
system setting.
4. Copy button: This button is the shortcut for Save function in the Utility function
menu. Pressing this button is equal to the Save option in the Save menu. The
waveform, configure or the display screen could be saved according to the chosen type
in the Save menu.
5. Vertical control area with 3 buttons and 4 knobs.
"CH1" and "CH2 " correspond to setting menu in CH1 and CH2, "Math" button refer
to math menu, the math menu consists of six kinds of operations, including CH1-CH2,
CH2-CH1, CH1+CH2, CH1*CH2, CH1/CH2 and FFT. Two "Vertical Position" knob
control the vertical position of CH1/CH2, and two "Scale" knob control voltage scale
of CH1, CH2.
6. M knob(Multipurpose knob): when a symbol appears in the menu, it indicates
you can turn the M knob to select the menu or set the value. You can push it to close
the menu on the left and right.
6
3.Quick Start
1. Waveform Display Area.
2. Run/Stop
3. The state of trigger, including:
Auto: Automatic mode and acquire waveform without triggering.
Trig: Trigger detected and acquire waveform.
Ready: Pre-triggered data captured and ready for a trigger.
Scan: Capture and display the waveform continuously.
Stop: Data acquisition stopped.
4. The two blue dotted lines indicates the vertical position of cursor measurement.
5. The T pointer indicates the horizontal position for the trigger.
6. The pointer indicates the trigger position in the record length.
7. It shows present triggering value and displays the site of present window in
internal memory.
8. It indicates that there is a USB disk connecting with the oscilloscope.
9. Channel identifier of current menu.
10. The waveform of CH1.
11. Right Menu.
12. The waveform of CH2.
13. Current trigger type:
Rising edge triggering
Falling edge triggering
Video line synchronous triggering
Video field synchronous triggering
The reading shows the trigger level value of the corresponding channel.
14. It indicates the measured type and value of the corresponding channel. "T" means
period, "F" means frequency, "V" means the average value, "Vp" the peak-peak
value, "Vr" the root-mean-square value, "Ma" the maximum amplitude value,
"Mi" the minimum amplitude value, "Vt" the Voltage value of the waveform's
flat top value, "Vb" the Voltage value of the waveform's flat base, "Va" the
amplitude value, "Os" the overshoot value, "Ps" the Preshoot value, "RT" the rise
time value, "FT" the fall time value, "PW" the +width value, "NW" the -Width
value, "+D" the +Duty value, "-D" the -Duty value, "PD" the Delay A->B
value, "ND" the Delay A->B value, "TR" the Cycle RMS, "CR" the Cursor
RMS, "WP" the Screen Duty, "RP" the Phase, "+PC" the +Pulse count, "-PC"
the - Pulse count, "+E" the Rise edge count, "-E" the Fall edge count, "AR" the
Area, "CA" the Cycle area.
15. The readings show the record length.
16. The frequency of the trigger signal.
17. The readings show current sample rate.
18. The readings indicate the corresponding Voltage Division and the Zero Point
positions of the channels. "BW" indicates bandwidth limit.
7
3.Quick Start
The icon shows the coupling mode of the channel.
"—" indicates direct current coupling
"~" indicates AC coupling
" " indicates GND coupling
19. The reading shows the setting of main time base.
20. It is cursor measure window, showing the absolute values and the readings of the
cursors.
21. The blue pointer shows the grounding datum point (zero point position) of the
waveform of the CH2 channel. If the pointer is not displayed, it means that this
channel is not opened.
22. The two blue dotted lines indicate the horizontal position of cursor measurement.
23. The yellow pointer indicates the grounding datum point (zero point position) of
the waveform of the CH1 channel. If the pointer is not displayed, it means that
the channel is not opened.
8
3.Quick Start
to the following steps:
1. Connect the power cord to a power source. Press the button on the bottom
left of the instrument.
The instrument carries out all self-check items and shows the Boot Logo. Push the
Utility button, select Function in the right menu. Select Adjust in the left menu,
select Default in the right menu. The default attenuation coefficient set value of the
probe in the menu is 10X.
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope
with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten
the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe
compensator.
3. Push the Autoset Button on the front panel.
The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in
several seconds (see Figure 3-5).
9
3.Quick Start
1. Set the attenuation coefficient of the probe in the menu as 10X and that of the
switch in the probe as 10X (see "How to Set the Probe Attenuation Coefficient"
on P10), and connect the probe with the CH1 channel. If a probe hook tip is used,
ensure that it keeps in close touch with the probe. Connect the probe tip with the
signal connector of the probe compensator and connect the reference wire clamp
with the ground wire connector of the probe connector, and then push the
Autoset button on the front panel.
2. Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see Figure 3-6 and Figure 3-7).
10
3.Quick Start
Caution:
The default attenuation coefficient of the probe on the instrument is preset to
10X.
Make sure that the set value of the attenuation switch in the probe is the
same as the menu selection of the probe attenuation coefficient in the
oscilloscope.
The set values of the probe switch are 1X and 10X (see Figure 3-8).
Caution:
When the attenuation switch is set to 1X, the probe will limit the bandwidth
of the oscilloscope in 5MHz. To use the full bandwidth of the oscilloscope,
the switch must be set to 10X.
Warning:
To avoid electric shock, always keep your finger behind the safety guard
ring of the probe during the operation.
To protect you from suffering from the electric shock, do not touch any
metal part of the probe tip when it is connected to the power supply.
Before making any measurements, always connect the probe to the
instrument and connect the ground terminal to the earth.
11
3.Quick Start
1. Use the Vertical Position knob to show the signal in the center of the waveform
window. The Vertical Position knob functions the regulating of the vertical display
position of the signal. Thus, when the Vertical Position knob is rotated, the pointer of
the earth datum point of the channel is directed to move up and down following the
waveform.
Measuring Skill
If the channel is under the DC coupling mode, you can rapidly measure the DC
component of the signal through the observation of the difference between the wave
form and the signal ground.
If the channel is under the AC mode, the DC component would be filtered out. This
12
3.Quick Start
mode helps you display the AC component of the signal with a higher sensitivity.
Vertical offset back to 0 shortcut key
Turn the Vertical Position knob to change the vertical display position of channel
and push the position knob to set the vertical display position back to 0 as a shortcut
key, this is especially helpful when the trace position is far out of the screen and want
it to get back to the screen center immediately.
2. Change the Vertical Setting and Observe the Consequent State Information Change.
With the information displayed in the status bar at the bottom of the waveform
window, you can determine any changes in the channel vertical scale factor.
Turn the Vertical Scale knob and change the "Vertical Scale Factor (Voltage
Division)", it can be found that the scale factor of the channel corresponding to
the status bar has been changed accordingly.
Push buttons of CH1, CH2 and Math, the operation menu, symbols, waveforms
and scale factor status information of the corresponding channel will be displayed
in the screen.
1. Turn the Horizontal Scale knob to change the horizontal time base setting and
observe the consequent status information change. Turn the Horizontal Scale
knob to change the horizontal time base, and it can be found that the Horizontal
Time Base display in the status bar changes accordingly.
2. Use the Horizontal Position knob to adjust the horizontal position of the signal
in the waveform window. The Horizontal Position knob is used to control the
triggering displacement of the signal or for other special applications. If it is
applied to triggering the displacement, it can be observed that the waveform
moves horizontally with the knob when you rotate the Horizontal Position knob.
Triggering displacement back to 0 shortcut key
Turn the Horizontal Position knob to change the horizontal position of channel
and push the Horizontal Position knob to set the triggering displacement back to 0
as a shortcut key.
13
3.Quick Start
3. Push the Horizontal HOR button to switch between the normal mode and the
wave zoom mode.
14
3.Quick Start
1. Push the Measure button to show the automatic measurement function menu.
2. In the right menu, select Add CH1.
3. In the left Type menu, turn the M knob to select Period.
4. In the right menu, select Add CH1. The period type is added.
5. In the left Type menu, turn the M knob to select Frequency.
6. In the right menu, select Add CH1. The frequency type is added.
The measured value will be displayed at the bottom left of the screen automatically.
15
4.Communication with PC
4. Communication with PC
The oscilloscope supports communications with a PC through USB port. You can use the
Oscilloscope communication software to store, analyze, display the data and remote
control.
To learn about how to operate the software, you can push F1 in the software to open the
help document.
Here is how to connect with PC using USB port.
(1) Install the software: Install the Oscilloscope communication software on the
supplied CD.
(2) Connection: Use a USB data cable to connect the USB Device port in the right panel
of the Oscilloscope to the USB port of a PC.
(3) Install the driver: Run the Oscilloscope communication software on PC, push F1 to
open the help document. Follow the steps of title "I. Device connection" in the
document to install the driver.
(4) Port setting of the software: Run the Oscilloscope software; click
"Communications" in the menu bar, choose "Ports-Settings", in the setting dialog,
choose "Connect using" as "USB". After connect successfully, the connection
information in the bottom right corner of the software will turn green.
16
5.Oscilloscopes Technical Specifications
Display 7" Colored LCD (Liquid Crystal Display),65536 colors, 800 x 480 pixels
Channel 2 channels
Max Record length 10K
Sampling rate / relay time
±100 ppm
accuracy
Input coupling DC, AC , Ground
Input impedance 1 MΩ±2%, in parallel with 20 pF±5 pF
Max. input voltage 400V (DC+AC, PK - PK)
DC Gain Accuracy ±3%
Vertical Sensitivity 5 mV/div – 5 V/div
Trigger type Edge, Video
Trigger mode Auto, Normal, Single
Line/field frequency
Support standard NTSC, PAL and SECAM broadcast systems
(Video)
Period, Frequency, Mean, PK-PK, RMS, Max, Min, Top, Base,
Amplitude, Overshoot, Preshoot, Rise Time, Fall Time, +PulseWidth,
Automatic measurement -PulseWidth, +Duty Cycle, -Duty Cycle, Delay A→B , Delay A→B ,
Cycle RMS, Cursor RMS, Screen Duty, Phase, +PulseCount,
-PulseCount, RiseEdgeCnt, FallEdgeCnt, Area, and Cycle Area.
Waveform Math +, -, ×, ÷, FFT
Waveform storage 16 waveforms
Communication interface USB 2.0, USB flash disk storage
Power supply 100 - 240 VACRMS,50/60 Hz,CATⅡ
Fuse 2 A, T class, 250 V
Mechanical Specifications
Dimension 301 mm× 152 mm×70 mm (L*H*W)
Weight About 1.1 kg
17