BMID2 Lec.9 - Electromagnetic Module
BMID2 Lec.9 - Electromagnetic Module
Electromagnetic
Module
Magnetic Resonance Imaging
No Title
1 Introduction.
2 MRI Magnet.
3 Gradients.
4 Spatial Encoding.
5 Image Formation.
6 Sequences.
7 Block Diagram.
1. Introduction
- The magnetic field strength resulted from the rotation and this
field with specific direction depends upon:
1. Location of the Proton inside the tissue of the body.
2. Total number of the Hydrogen proton per tissue.
- Within a large external magnetic field (called Bo), slightly
more than half align with the field.
- This imbalance creates Mo. Some of the spins align with the
field (enphase) and some align against field (dephase).
Mo B o
Why is Hydrogen used in MRI?
377
Magnetization. Z axis = ?
Relaxation.
Return to equilibrium of net magnetization is called
Relaxation
378
T1 Relaxation
Signal
With no other signal considerations this
would mean that the protons with short White Matter
Fat
T1 time would contribute the highest
signal on the image.
Hydrogen in differing
Pixel containing protons tissues has a different
with short T1 time. (e.g. Fat) T1 Time.
TR
T2 Decay
With no other signal considerations Signal
this would mean that the protons with a
long T2 time would contribute the
highest signal on the image.
Water
Notes:
1. The sensitive volume of surface coils,
being placed in direct contact with the
zone of interest, has less depth and is
more heterogeneous.
2. The homogeneity and sensitive volume
of surface coils can be improved by
combining them into a phased array.
They still have the advantage of a
better SNR, but at the cost of more
complex signal processing.
4. Spatial Encoding.
Gz
1.5 Bo
As this gradient is
applied simultaneously on
receiving the signal, the
frequency data is included.
5. Image Formation.
- This signal is digitized and raw data are written into a data
matrix called K-space.
- K-space data are equivalent to a Fourier plane.
- To go from a k-space data to an image requires using a 2D
inverse Fourier Transform.
Single frequency
To describe a sine wave, we need its
amplitude and frequency, but also its
phase. The following animation illustrates
the consequence of a phase change on the
sine wave (shift in time domain).
Multiple frequencies
- The Fourier transform is a mathematical
procedure that decomposes a signal into a sum
of sine waves of different frequencies, phases
and amplitude.
- The human ear does the same processing with
sounds, which are analyzed as a spectrum of
elementary frequencies.
- Knowing frequency, amplitude and phase of
each sine wave, it is possible to reconstruct the
signal (inverse Fourier transform).
Spatial Frequency
STIR TSE Turbo STIR Fast STIR Fast STIR Fast STIR
FLAIR FLAIR FLAIR FLAIR FLAIR FLAIR
FLAIR TSE Turbo FLAIR Fast FLAIR Fast FLAIR Fast FLAIR
7. MRI Block Diagram
- Functions such as gates and envelopes for the NMR pulses, blanking for the pre-amplifier and RF power
amplifier and voltage waveforms for the gradient magnetic fields are all under software control.
- The computer performs the various data processing tasks including the FT, image reconstruction, data filtering,
image display and storage.
- Therefore, the computer must have sufficient memory and speed to handle large image arrays and data
processing, in addition to interfacing facilities.
The Magnet
- There is an increasing tendency to use higher magnetic Fields (3.0T) which stems from the fact that
SNR increases with the field strength, allowing greater sensitivity to contrasts of interest, including
functional blood oxygen level dependent (BOLD) contrasts, and higher spatial resolution.
- However, as more experience is gained in functional MR imaging at 3.0 T and higher, it turns out that
there are also rather complex physiological dependencies that are much less obvious and cannot be
assessed easily by physical laws.
RF Receiver:
- Spatial distribution information can be obtained by using the fact that the resonance frequency depends on
the Bo.
- By varying the field in a known manner through the specimen volume, it is possible to select the region of
the specimen from which the information is derived on the basis of the frequency of the signal (Slice
Selection).
- The strength of the signal at each frequency can be interpreted as the density of the hydrogen nuclei in the
plane within the object where the magnetic field corresponds to that frequency.
- NMR imaging methods exploit this property by way of carefully controlled, well-defined gradients to
modulate the NMR signal in a known manner such that the spatial information can later be decoded and
plotted as an image.
- Typically, the gradients are chosen with linear spatial dependence so that the NMR frequency spectrum
directly corresponds to the position or even one or more spatial co-ordinate axes.
- The imaging methods differ mainly in the nature of the gradient time dependence (static, continuously time-
depended or pulsed), and in the type of NMR pulse sequence employed.
A block diagram of the system used is shown in Fig. The hardware can be broken down into four sub-systems.
First sub-system:
- Includes the interface between the computer and the gradient control system.
- Its primary function is to allow the independent positioning of the three time invariant planes (X, Y and Z).
Third sub-system:
- The output of the attenuators is then voltage-amplified by two op amps prior to the driven circuits.
- Current control requires through the shim coil so that the control used to adjust the static field gradients be
available for setting the DC levels upon which the alternating gradients are superimposed.
- An op amp serves the differential voltage drop across a dummy load (having the same resistance as the
shim coil) and produces an output which is then DC coupled to the drivers.
- The high current drivers use a conventional design with a single op amp providing the input to a driver and
a complimentary pair of power transistors to provide a sufficient current to the gradient coil.
Fourth sub-system:
- In typical scanners, gradient coils have an electric resistance of about 1 W and an inductance of 1 mH.
- The gradient fields are required to be switched from 0 to 10 mT/m in about 0.5 ms.
- The current switches from 0 to about 100 A in this interval.
- The power dissipation during the switching interval is about 20 kW.
- This places very strong demands on the power supply and it is often necessary to use water cooling to prevent
overheating of the gradient coils.
Imager System: