Sample Adapter LMX05 LAVM10

This document provides instructions for using sample adapters with the SPECTROMAXx LMX05 and SPECTROLAB LAVM10 instruments. It describes how to use the small sample adapter, foil adapter, and wire adapter and includes pictures to illustrate the steps. Safety precautions are also outlined.

Uploaded by

prasath
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

SPECTROMAXx LMX05 / SPECTROLAB LAVM10

1 Use of the SPECTROMAXx/SPECTROLAB sample


adapters

1.1 Preparation .....................................................................1


1.2 Small sample adapter .....................................................2
1.3 Foil adapter.....................................................................3
1.4 Wire adapter ...................................................................4

Adapter kit (PN


75260125)
Electrode spacer
Special spark stand plate
for small parts
Foil adapter
Plastic holder for small
parts
Protection cover
Wire adapter

1.1 Preparation

Caution!
When measuring small samples, wires, foils, etc. it is important
to choose the correct method, which can be identified by the
extension “–SP” after the name.

SPECTRO Analytical Instruments GmbH


Document 80111203 1 of 4
Revision 1; 02.12.2009
SPECTROMAXx LMX05 / SPECTROLAB LAVM10

Remove the previous spark


stand plate
Loosen the electrode. It will
be pushed up by a spring.
Fasten the electrode. It is
now used to center the
spark stand plate for small
samples correctly.
Mount the new spark stand
plate; do not forget the O-
ring

Adjust the electrode gap


with the electrode spacer
(1) for small parts.

1.2 Small sample adapter

Put the sample onto the


spark stand and fix it with
the plastic holder for small
parts.
1
The plastic holder is
transparent to make sure
that the sample is still
centered.
Fix the sample and holder
with the special clamp (1)

SPECTRO Analytical Instruments GmbH


Document 80111203 2 of 4
Revision 1; 02.12.2009
SPECTROMAXx LMX05 / SPECTROLAB LAVM10

Slide the protection cover


on
The spark generates UV
radiation. Therefore do not
measure without the
protection cover when the
spark stand opening is not
fully covered.
Lower the sample clamp to
close the safety circuit.

1.3 Foil adapter

Put the sample onto


the spark stand.
Lay the foil adapter on
top. It will dissipate the
heat.
1 Fix the sample and
adapter with the special
clamp (1)
Lower the sample
clamp to close the
safety circuit.
.

SPECTRO Analytical Instruments GmbH


Document 80111203 3 of 4
Revision 1; 02.12.2009
SPECTROMAXx LMX05 / SPECTROLAB LAVM10

1.4 Wire adapter

The wire adapter has


two grooves for wires
with different
diameters.

Fix the sample and


holder with the special
clamp (1)
Lower the sample
clamp to close the
safety circuit.

SPECTRO Analytical Instruments GmbH


Document 80111203 4 of 4
Revision 1; 02.12.2009