FS8233 4.8 Inch Large Screen Multimeter
FS8233 4.8 Inch Large Screen Multimeter
AUTO
6000 t
m COM
Warning
Pay special attention to use of this instrument, for
improper use may cause electric shock or damage to
instrument. During use, observe usual safety regulations
and observe safety measures regulated in use manual.
In order to make full use of instrument functions and
guarantee safe operation, please carefully read and
observe use methods in this manual.
AUTO
6000 t
m COM
3
LCD display
AUTO
6000 t
m COM
AC & DC
Connection/Disconnection indication
Automatic range mode
Automatic shutdown indication
Low battery
Reading hold state
Volt (voltage), ampere (current)
Ohm kilohm and megohm (resistance)
Hertz
DIODE
Microfarad
Centigrade and Fahrenheit
5
3.2 Technical indexes
3.2.1 AC current
Range Resolution Accuracy
m m
± (1.0% reading + 8 digits)
3.2.2 DC current
Range Resolution Accuracy
m m
± (1.0% reading +5 digits)
3.2.3 DC voltage
Range Resolution Accuracy
6V 0.001V
± (0.5% reading + 3 digits)
60V 0.01V
600V 0.1V
- Min. input voltage 0.5 V DC
- Max. input voltage: 600V DC
6
3.2.4 AC voltage
Range Resolution Accuracy
6V 0.001V
60V 0.01V ±( 0.8% reading + 5 digits)
600V 0.1V
- Min. input voltage: 0.5 V AC
- Max. input voltage: 600V AC (valid value)
- Frequency range: 40 ~ 1000Hz
3.2.5 Line on-off test
Range Resolution Function
1Ω If resistance of the line
being measured is less than
50Ω, buzzer in instrument
will make continuous
alarming sounds.
- Overload protection: 600
0V DC or AC (valid value)
3.2.6 Resistance
Range Resolution Accuracy
600Ω 0.1Ω
6kΩ 0.001kΩ
60kΩ 0.01kΩ
± (0.8% reading + 3 digits)
600kΩ 0.1kΩ
6MΩ 0.001MΩ
60MΩ 0.01MΩ
Overload protection: 600V DC or AC (valid value)
7
3.2.8 Capacitor
Range Resolution Accuracy
reading digits
8
Frequency
(viagear A):
Range Resolution Accuracy
reading digits
Via gear V:
Range Resolution Accuracy
reading digits
4. Operation guide
4.1 Reading hold
During measurement, if it is required to hold reading,
touch button, value on display will be locked, touch
button again, to cancel reading hold.
4.2 Backlight/Light
1) During measurement, if environment light is too dark,
causing it difficult to read,you could press key
for more than 2 seconds, to open backlight or light,
which will turn off in about 60 seconds automatically.
9
4.3 Automatic shutdown
1) If there's no operation within 15 minutes after start,
instrument will enter standby state, and shut down to
save energy.
2) After automatic shutdown,
wake the instrument enter work state.
3) Holding SEL/NVC to turn on the power will cancel
the automatic shutdown function.
After shutdown, restart to restore the automatic
shutdown function.
10
4.5 NCV (Non -contact voltage detection)
Switch the rotation switch to any gear, hold down the
"NCV" key to display EF, and bring the NCV sensor close to
the conductor under test, instrument could detect whether
the lead being measured is >90V AC voltage. When
instrument detects AC voltage, instrument buzzer will
alarm and NCV alarm light will flash.
AUTO
6000 t
m COM
Note:
1) Even if there’s no alarm indication, voltage may still
exist. Do not depend on non-contact voltage detector
to judge whether there's voltage in the lead. Detection
operation may beimpacted by factors like different
socket designs and insulation thickness types etc.
11
Connect the measuring pen into the measured signal.
When the measured signal >20mA, the meter will
display the current measured current value.
When the measured ac signal >2000mA,
the instrument pair displays
the ac current frequency,
AUTO
and the measured current
b> 800mA is red backlight
warning.
6000 t
m COM
AUTO
6000 t
m COM
12
AUTO
6000 t
m COM
AUTO
6000 t
m COM
13
Temperature measurement
14
5. Maintenance
5.1 Change battery
Warning
Before opening battery cover of the
instrument, remove the pens from the circuit being
measured, to avoid electric shock.
Warning
When changing the pen, it is required to
replace with an identical pen or a pen of the same
level. The pen must be in good condition, and level
of the pens is: 1000V 10A.