SOP-SRC-QC-02 - Calibration of Equipments and Measuring Instruments

This document provides a standard operating procedure for calibrating equipment and measuring instruments at Dubai Ship Building & Engineering. It outlines responsibilities for ensuring equipment is calibrated as required, including documenting calibrations. Departments must visually check equipment before use and coordinate calibration internally or with an external provider. The QA/QC department maintains a calibration log and receives updated certificates. Calibration reveals if readings match a master, and equipment may be adjusted, repaired or scrapped depending on results.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as RTF, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
44 views5 pages

SOP-SRC-QC-02 - Calibration of Equipments and Measuring Instruments

This document provides a standard operating procedure for calibrating equipment and measuring instruments at Dubai Ship Building & Engineering. It outlines responsibilities for ensuring equipment is calibrated as required, including documenting calibrations. Departments must visually check equipment before use and coordinate calibration internally or with an external provider. The QA/QC department maintains a calibration log and receives updated certificates. Calibration reveals if readings match a master, and equipment may be adjusted, repaired or scrapped depending on results.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as RTF, PDF, TXT or read online on Scribd
You are on page 1/ 5

Document Control

HSEQ
Number: SOP-SRC-QC-
STANDARD OPERATING PROCEDURES
02_Rev. 1
Calibration of Equipment & Measuring Instrument

Quality
Management
System

Dubai Ship Building & Engineering Dubai,


United Arab Emirates.
P.O Box 35515
Phone: +971 4 586 5260
Fax: +971 4 568 5298
info.dsbe.ae

Division Ship Repair & Conversion

Department QA/QC

Type of Document Procedure

Title of Document Punch List Management


Calibration of Equipment & Measuring Instrument

Document Number _Rev. No. SOP-SRC-QC-02_Rev. 1

Document Owner QHSE Manager

Next Review Latest by August-2026

Approval
Rev.
Date Prepared Reviewed Approved
No.

Dhaneesh Chandrasekaran
1 11-Dec-2023 Nedal Ali Abwd
Vellilapulli Balamurugan

REVISION NO: 1
DATE: 11/12/2023
DUBAI SHIPBUILDING & ENGINEERING LLC. Page 1 of 5
Document Control
HSEQ
Number: SOP-SRC-QC-
STANDARD OPERATING PROCEDURES
02_Rev. 1
Calibration of Equipment & Measuring Instrument

Record of Amendments

1 Revision made with updating of calibration log (Form #115) in the procedure

0 New procedure implemented to control the calibration of equipment and instruments

REVISION NO: 1
DATE: 11/12/2023
DUBAI SHIPBUILDING & ENGINEERING LLC. Page 2 of 5
Document Control
HSEQ
Number: SOP-SRC-QC-
STANDARD OPERATING PROCEDURES
02_Rev. 1
Calibration of Equipment & Measuring Instrument

1. Purpose
1.1. To define the process of controlling the calibration of Equipment and measuring instruments.

2. Scope
2.1. Applicable to all equipment and instruments and that requires calibration by DSBE Departments
or by External Provider.

3. References
3.1 ISO 9001: 2015 Clause 7.1.5, Monitoring & Measuring Resources.

4. Definitions / Abbreviations
4.1. DSBE: Dubai Ship Building & Engineering

4.2. QMS: Quality Management System

4.3. QA/QC: Quality Assurance / Quality Control

4.4. RFI: Request for inspection

4.5. External Provider: An approved supplier of DSBE qualified for the calibration

5. Responsibility
5.1. It is the responsibility of the concerned Department Supervisors to ensure that all relevant
testing, measuring and monitoring equipment are calibrated as often as required.

5.2. It is the responsibility of the Department Manager to ensure that the calibration takes place as
often as required.

5.3. It is the responsibility of the QA/QC Department to verify whether the equipment and instrument
that are in use in the yard is calibrated properly.

5.4. QA/QC Department will keep the calibration record of instruments and monitor the due date of
calibration.

5.5. QA/QC Department shall be kept all updated calibration certificates for reference.

5.6. QA/QC Department or Trade supervisors may be coordinate with External provider for
calibration.

5.7. Procurement Department is responsible for collecting the calibration certificates for hired
Equipments.

6. Procedure

6.1. Monitoring of measuring equipment must be visually checked by the User before use. The
visual check may include to identify:

6.1.1. Physical damage (wear and tear).

REVISION NO: 1
DATE: 11/12/2023
DUBAI SHIPBUILDING & ENGINEERING LLC. Page 3 of 5
Document Control
HSEQ
Number: SOP-SRC-QC-
STANDARD OPERATING PROCEDURES
02_Rev. 1
Calibration of Equipment & Measuring Instrument

6.1.2. Equipment calibration due date label


6.1.3. Any damage to calibration sticker.

6.2. The equipment shall not be used, if any of the above checks are not adequately met.

6.3. Some equipment such as measuring tapes, rulers, and vernier callipers up to 200mm are not
calibrated but visually checked for wear and tear before each use. There are no records kept for
such visual checks of equipment.
6.4. The calibration of testing, measuring and monitoring equipment will be carried out in one of the
following ways:

6.4.1. By the User department.


6.4.2. By the Electrical Department.
6.4.3. By an External Provider

6.5. QA/QC Department shall maintain a calibration Log of equipment contains Device name,
Identification Number, Manufacturer, Range / Capacity, Certificate Number, Calibration Interval,
Calibration date, calibration Due Date.

6.6. For the calibration of Welding machines and Ovens, Welding supervisor will notify the Electrical
department by email with calibration log.

6.7. Calibration sticker should be there with Identification Number, Calibration date & Calibration Due
date.

6.8. If any equipment or instrument is scrapped, Lost and permanently damaged, Particular
department will notify to QA/QC Department to amend the calibration Log.

6.9. Calibration will reveal the correctness / deviations of readings with respect to a Master
Equipment. During calibration, if the equipment denotes readings which are outside the
acceptable range, then it is adjusted to ensure that readings are within the acceptable range. If
the adjustment is not feasible, It may recommend the equipment for repair or scrap

7. Documented Information

7.1. Calibration Log (Form #115) – Attachment 1

7.2. Calibration certificates.

REVISION NO: 1
DATE: 11/12/2023
DUBAI SHIPBUILDING & ENGINEERING LLC. Page 4 of 5
Document Control
HSEQ
Number: SOP-SRC-QC-
STANDARD OPERATING PROCEDURES
02_Rev. 1
Calibration of Equipment & Measuring Instrument

ATTACHMENT #1

REVISION NO: 1
DATE: 11/12/2023
DUBAI SHIPBUILDING & ENGINEERING LLC. Page 5 of 5