Ad_Multi_HardwareOutline, 6 September 2004

Hardware Outline
Installation Notes
© Toshiba Medical Systems Corporation 2003 1
* Except A3522033 TMSE
A3522076 TAP
A3522093 TAMS
In General A3522096
A3522134
TAMS
TMSE

Somewhat different
Gantry - Reinforced and balanced for 0.4 sec rotation*
DD Motor Servo - Prepared for 0.4 sec rotation
From January 2004: new DD motor and servo for 0.4 sec rotation
Tube - Reinforced for 0.4 sec rotation - CXB-750D
Optics assembly - DR wedge instead of flat wedge
Software - Dedicated for 16 slice hardware
KGTSM - Adapted for 16 slice system
KGTSP - Adapted for new couch

Quite different
Couch - new design - Foot pedals -Accessories - etc.
MUDAT - ten channels
Detector - sixteen rows
DAS - processes sixteen rows of data
Console - two units - new reconstruction unit - new computer

© Toshiba Medical Systems Corporation 2003 2


New or Ref. Detector SRU Gantry
big Change X-Ray tube
(Trans, Inverter) Rotation
Detector DAS
Minor
Change Slit

Same Wedge
as 4 slice Multi X-Ray Control DAS Rotation
OPCONTA (XC) P.S.U. elec. part
Laser projector

MR DD motor LCSR
sensor
LPBA RPDM
Monitor Hybrid Mouse
Keyboard Speaker

Monitor Speaker SPDA LPBM


MIC

Power
SS/ Amp. Control
ADI2 GCIFA

GCIFA D Operation panel R

O2 R12K Hybrid Mouse


Keyboard Operation panel L
FC I/F Power sequence

KGTSM
RTM-D Tilt encoder
position sensor
RTM
H Laser projector
U Image
B HDD MHR FC Look guide
RDD Fiber Channel KGTSP
Gantry
OCTANE2 MHR FC
Fiber Channel
Tilt motor Stationary
MIC

Image Couch
Recon. MHR DP &
MO MHR BP SI Pump motor
drive Elec valves
Image Back plane
Slide servo AQUILION 16
Power Sensors Slide motor
HDD Power
Control Console
In more Detail
• Low noise fans - in the roof -
Speed depends on gantry temperature
• Access to R-base power supply units from the rear
Also for replacement of power supplies
• Three wedges
Small, Large and Dose Reduction (DR)
• Upper cover service position
• Easy adjustment of MR sensor
• Firmware for gantry control sequence has been changed
• Additional 3.3V power supplies - stationary and rotational
• Improved Visibility of slice counter - fitted under an angle -
• Improved accessibility of switches on the various boards

© Toshiba Medical Systems Corporation 2003 4


In more Detail - MUDAT
• 10 channels
• Data transfer speed from rotation part to stationary part is
doubled - 16Mhz to 32Mhz
• Optical fiber cable between Gantry and Console
• Dedicated interfaces - OPCONTA - GCIFA+GCIFA-D
With Retransmit function
• Access hatch for MR sensor position check and adjustment

© Toshiba Medical Systems Corporation 2003 5


In more Detail - Detector & DAS
16 row Detector
16 rows of 0.5mm detectors in the center and 12 rows of 1mm
detectors on either side

16 row DAS
32x Converter-16
1x Control-16

DAS power supply


2x 5.7V - PS1 and 3
1x 3.3V - PS2
Detector and DAS are one unit
Weight is about 76 kg
Mounting jig is standard - STAY -
© Toshiba Medical Systems Corporation 2003 6
Console - CKCN-011A
• CPU Box and Reconstruction Box
• Power controller in CPU Box
• CPU Box - Scan console O2 R12K - Display console Octane2
• 46 cm (18”) Colour flat-panel monitors
• Reconstruction Box - STAR - MHR
• MOD connected to Display console - no CD-ROM player -
• RDD - 5 x 36 GB - RAID level 3 - byte level striping with parity

© Toshiba Medical Systems Corporation 2003 7


Console - CPU Box

Item Scan Console Display Console


CPU O2 R12K Octane2
Memory 512 MB 1 GB
System Disk 9 or 18 GB 9 GB
Image Disk 36 GB 36 GB
Media CD-ROM
For service purposes
5 inch MOD

© Toshiba Medical Systems Corporation 2003 8


Console - CPU Box

Octane2 R12K

Handle With Care!


© Toshiba Medical Systems Corporation 2003 9
Reconstruction Box
RTM + RTM-D
Hardware is the same but firmware is different

MHR-FC board x 2 - Fiber Channel -


For preprocessing and RDD I/F

MHR-DP board x 3 - Dual Processor -


For reconstruction

MHR-BP board x 3 - Back Projector -


For reconstruction

MHR - Multi-functional High-speed Reconstruction Unit -


STAR - Sixteen-slice Advanced Reconstruction Unit -

© Toshiba Medical Systems Corporation 2003 10


Reconstruction Box

RDD

cPCI

© Toshiba Medical Systems Corporation 2003 11


Site Planning
• New Couch is longer than existing model -
Long version is standard
• Two Console Units - Boxes -
• Reinforced gantry base - New bringing-in jigs - FSM3123

Refer to the relevant Product Data


and Site Planning Manual

© Toshiba Medical Systems Corporation 2003 12


Site Planning - Installation
Two console units
• CPU Box and Reconstruction Box
• Keyboard and monitor cables - standard length is 1.5m -
Connect to CPU Box
• Maximum cable length for D-console is 6 m
• Casters under chassis not for bringing-in but for service use
after installation - Use transportation tool for bringing-in
• Keep system transformer away from the console

Cables between units


• 6 m between CPU-Box and Rec-Box - also optical cable -
• DAS optical cable from the gantry connects to Rec-Box
Keep the protection caps of the optical cables at a safe place on site

© Toshiba Medical Systems Corporation 2003 13


Installation - Console
• Capacity of AC 100V outlet is larger than previous consoles
Don’t connect other company’s equipment!
• Do not touch the surface of the Octane2 connectors with bare
hands as this causes damage to the connectors which is not
covered by warranty
Two C-POP covers for Octane2 are included with console accessories -
also keep them available at the site -
• Maximum calibration data to be stored is 400 stages
- Make an appropriate selection
- Calibration is time consuming

© Toshiba Medical Systems Corporation 2003 14


Tube Alignment
• X-ray tube alignment should be done with SMALL focus
Most applicable scan conditions use SMALL focus

• Use the F/B and L/R alignment programs

V1.30 - has F/B alignment program available

© Toshiba Medical Systems Corporation 2003 15


Tube F/B Alignment
• Data is registered by Quit
• No scan - no new registration
• Registered data has no meaning in software

This section was once meant for future use

© Toshiba Medical Systems Corporation 2003 16


Limitation of input value for Large Focus

Tube L/R Alignment


• An abnormal image* may occur with 4 row MUSCOT
reconstruction if L/R alignment value for LF out of range
(448.0 =< x < 448.5)
This is caused by a shortcoming of the reconstruction firmware

• Temporary workaround:
Align the tube as usual with Small Focus
Check the LF center channel offset value as usual
If less than 448.0 - Input 448.0
If more than 448.5 - Input 448.49

• LF may be out of range if there is a big difference with SF

• Corrected in V1.40
* FSM CT3189

© Toshiba Medical Systems Corporation 2003 17


Operation
• After switching on the gantry, allow one minute for GCIF-D
initialization before using the system
• 80kV and 100kV are not standard but can be enabled
- Special Configuration -
• Dose Reduction wedge (DR) is enabled - default setting -
- Special configuration - Calibrations!
• Automatic air calibration can take long
- Set only the necessary warm up scan conditions -
• Calibration window opens with “air” calibration menu
- Select NRA to take NRA data -
• Wait for at least 30 seconds after switching off the gantry
before switching on again - to avoid damage to the DD
motor controller -
© Toshiba Medical Systems Corporation 2003 18
As from April 2004 production

Aquilion16 PC based
• New computers
Design change only - NOT marketed as a new product
Two PC’s in CPU box - for Aquilion16 and Aquilion8

• Improved 3D
Competitive specs
• 95% re-usage of existing technology
as developed for Super4
• Gantry and couch unchanged

© Toshiba Medical Systems Corporation 2003 19


Compared with Aquilion16

Aquilion16 PC based - Software


Current system PC based system
Scan DSS 3D Scan DSS 3D
Recon. Display /MPR Recon. Display sub /MPR
sub system system
A part of
correction Diversion Diversion
Driver Common Windows IAP
middleware 3D engine
Driver Common Octane/O2 A part of
middleware IAP correction Diversion Diversion
3D engine
Collection Windows manager Purchase
recon.,
Collection peripheral Exceed Purchase
SGI;O2/IRIX
recon., equipment MKS(Nutcracker) Purchase
peripheral Purchase
equipment
No Windows
change PC(Intel family CPU)

© Toshiba Medical Systems Corporation 2003 20