Medical Oxy Life Oxygen Concentrator 10l
Medical Oxy Life Oxygen Concentrator 10l
Contents
1 Foreword…………………………………..........……………2
2 Symbols…..........……………………………….……………2
3 Safety notice……...............…………………..……………3
4 Product introduction…....................................…..……4
5 Operation condition………….....……………………..……5
6 Scope of application…………..........………………………5
7 Technical parameter………........………..…………………5
8 Structures and Functions……………….....………….……6
9 Operation instructions…………................................…8
10 Alarms and safety devices.........………....………….…11
11 Maintenance…...........……………………..…………..…12
12 Troubleshooting……................................……………14
13 EMC Declaration..…....................................……..…16
14 Accessories……………………………………...…………19
15 Condition for transportation and storage....................20
16 Quality warranty......................................................20
Date of manufacture Alternating current :Always place the concentrator on a hard surface. Never place the concentrator on a
surface such as bed or couch, where the concentrator may tip or fall.
T6.3AL/
Manufacturer Type and rating of fuse :NEVER leave the concentrator unattended when plugged in.
250V
:Ensure the bottom smooth exhaustion during operating, otherwise the oxygen
Fragile, handle with care Keep dry
concentrator will be over-heated.
This way up Stacking limit by number :5 minutes are needed from oxygen concentrator from warming up to reach regular
function and normal performance.
NOTE: If oxygen does not seem to flow, first verify that the flowmeter ball is registering a 5 Operation conditions and Environment
flow. Then, place the tip of the cannula into a glass of water; if bubbles come out of the
Ambient temperature: 10℃-40℃
cannula, oxygen is flowing. If bubbles do not appear, turn off the oxygen concentrator
Relative humidity: 30%-85%
immediately and refer to Troubleshooting.
Air pressure: 700 hPa-1060 hPa
NOTE: There is never a danger of depleting the oxygen in a room when you use your Altitude: Up to 2286m without degradation; Consult your equipment provider for further
oxygen concentrator. information regarding to 2286m to 4000m
No corrosive gas and strong magnetic field around.
Radio Frequency Interference
Most electronic equipment is influenced by Radio Frequency Interference (RFI). When
6 Scope of application:
there is strong electromagnetic interference, maybe the LCD will be slightly affected, but
the oxygen concentrator is still running. ALWAYS exercise CAUTION with regard to the For Medical Use:
use of portable communications equipment in the area around such equipment. Medical oxygen supplied by the concentrator is beneficial to cure the disease or heart and
blood vessel system, chronic pulmonary system, the brain and blood vessel system,
chronic pulmonary tuberculosis, and other oxygen lacking symptoms, etc.
Requirement of environment protection For Health care:
Medical oxygen can be used for athletics and intellectuals and brainworkers, etc. to
The materials used in the system won’t create environment hazard. The packing materials
eliminate fatigue and also suit for the departments of health care, sanatorium, healthy,
of the system are recyclable, and they must be collected and disposed according to the
plateau military camps and hotels and other places where need oxygen.
related regulation in the country or region where the package of the system or its
accessories is opened. The nasal oxygen tube is made of medical PVC, and if it is thrown
7 Technical Parameters
away, it could not be bio-degradable, so it will cause the pollution. Any material of the
system, that may cause pollution in the environment, must be collected disposed strictly
Serial 3L 4L 5L 6L 8L 10L
complied with the local rules and requirements.
Rated power (VA) 300 300 400 300 400 550
Operation voltage
4 Product introduction AC230/50
(V/Hz)
Oxygen flow (L/mln) 0.5-3 0.5-4 0.5-5 0.5-6 0.5-8 0.5-10
The oxygen concentrator is a device that extracts oxygen from atmospheric air. It will
Oxygen
typically be an electrically-powered molecular sieve (artificial zeolite) used to separate 93%±3%
concentration (%)
nitrogen from ambient air. It could be applied widely in the hospitals at all different level, Outlet pressure
0.04—0.07
clinics, health centers and family nursing, health care for the old person, mental workers (Mpa)
and students, etc.. The oxygen concentrator can supply 1-2 patients simultaneously, with Power failure; low&high pressure ;
Alarm Optional:temperature; low purity; maintenance reminding after
steady oxygen flowing out, safe and reliable, low cost, adjustable flow. The key parts of
3000 hours use
the concentrator adopt anti-tiring and anti-aging design, and the planned life of the whole
Sound level (dB(A)) ≤43 ≤45 ≤50 ≤45 ≤50 ≤55
concentrator reaches up to 20,000 hours or 5 years, whichever comes first.
LCD display accumulating timing; present timing; timing;
8.9 Timing buttons 9.2. Connect the nasal oxygen canula to the humidifier outlet nozzle or to the
The two buttons are used for timing adjustment, and each press of the left button(△) concentrator outlet if a humidifier has not been prescribed. Then set the nasal oxygen
will increase timing by 10min, the max timing is 40 hours. And each press of the right cannula over patient’s ears, insert the nasal oxygen cannula into patient’s nostrils to
button(▽) will decrease timing by 10min. When the right button(▽) is pressed to absorb oxygen; The nasal oxygen canula should be limited to 20 meters long, in order
reduce timing till “0”, the oxygen concentrator will turn off automatically. to ensure that the oxygen flow rate remains
10L: 10position flow value from 0.5 ~ 10L/min on flowmeter. The maximum recommended atomization treatment.
flow: 10L/min. d. Do clean the atomization devices followed by the instructions of the atomization
In compliance with the ISO8359 standard, the flow supplied is equal to the flow set on the devices.
flowmeter, accurate to within 10% or 200ml/min, whichever is the larger of the two USE OF PULSE OXIMETER (for optional)
The variation of the maximum recommended flow does not exceed 10% of the indicated 1. Insert the pulse oximeter probe joint to the pulse oximeter connector on the oxygen
value when a back pressure of 7kPa is applied to the output of the device. The maximum concentrator.(see Figure 11), and the LCD will show the Figure 12.
outlet pressure is 70kPa 2. Put the finger completely into the pulse oximeter
Unplug the power plug
Oxygen Concentration: 3. The SPO2(measuring range: 35%~100%; accuracy: 70% to 100% ±2 digits) and Pulse
-at 2L/min: >90% Rate(measuring range: 30bpm~240bpm; accuracy:±2bpm or±2%) reading will display
-at 3L/min: 93%(±3%) on the screen in few seconds.(see Figure 13)
-at 4L/min: 93%(±3%)
-at 5L/min: 93%(±3%)
-at 6L/min: 93%(±3%)
-at 8L/min: 93%(±3%)
-at 10L/min: 93%(±3%) Figure 10
:It is very important to select only the prescribed level of oxygen. Change the flow
selection only under the guidance of your physician.
Pulse oximeter probe
9.5 . When finished the absorbing, set the I/O power switch to the “O” position to turn off
the unit, if there is discontinuous use, please unplug the power plug, as shown in
Figure 11
Figure 10
9.6. If the patient needs timing oxygen absorbing, please refer to 8.9 on page 8
connected and select the proper angle to gain a better atomization. a. Power failure alarm: In case of a loss of mains power or when the power cord is not
a. Open the cover of medical cup, and add atomization remedy that needed, then close plugged into the wall outlet, an audible alarm is activated with red indicator on(8.1.1b
the cover. or 8.1.2b on page 7). The troubleshooting is referred to 12 on page 14.
b. Connect the joint of atomization nozzle (or mask) with the cover of medical cup, and b. low&high pressure alarm: There is a pressure sensor on the main board to check
then connect the other end of atomizer connection tube with the atomization outlet and the system pressure, when the pressure is lower than 0.1Mpa,there is an audible
turn on atomizer. alarm with yellow indicator on(8.1.1d or 8.1.2c on page 7) and the oxygen
c. Turn on the power of oxygen concentrator, and shut up flow meter, then it is ready for concentrator is still working; When the pressure is higher than 0.23Mpa,there is an
audible alarm with red indicator on(8.1.1c or 8.1.2d on page 7) and the oxygen 11.2. Clean air filter: It is a critical step for daily maintenance to clean intake air filter, at
concentrator is stopped. The troubleshooting is referred to 12 on page 14. least twice a month.
c. Low oxygen concentration alarm(OCSI): The oxygen concentration will rise to the Detail steps: remove the two intake air filters on both sides of the body, clean them
normal level in five minutes of operation. When oxygen purity is≥85%, the green with mild household cleaner and clean it with clean water completely, get ride of
lamp(8.1.1f or 8.1.2e on page 7) is on, when oxygen purity is <85%, red lamp is on extra water and dry up naturally, finally set back after dry up, as shown in Figure 15.
for audible alarm, indicating low purity(8.1.1g or 8.1.2f on page 7). Refer to the
troubleshooting on page 14. Or call your supplier to service the device
d. Temperature alarm: There is a temperature sensor on the main board to check the
internal temperature, when the temperature is higher than 50°C in the oxygen
concentrator, there is an audible alarm with red indicator on(see H.T. on the lamp)
and the oxygen concentrator will be stopped. The troubleshooting is referred to 12 on
page 14. Air filter
10.2 Safety devices
a. Compressor motor:
Thermal safety is ensured by a thermal switch situated in the motor winding(145±5℃).
b. Safety valve:
This is fitted on the compressor outlet and is calibrated to 2.5 bar(250kPa).
11 Maintenance Figure 15
11.3. Clean secondary filter
:Disconnect the power cord from the electrical outlet before you clean the cabinet. Method I: Clean secondary filter an interval about 800 hours, open the top cover and
:Do not operate the concentrator without the filters installed, or while filters are wet. take out storage case,Rotate the air filter with counterclockwise direction as Figure
These actions could permanently damage the concentrator. 16,take down the filter cloth, then clean it with detergent, and then clean it out with
NOTE: If legally binding regulations govern the installation, service and/or the operation of clean water completely, get rid of the extra water, and dry it naturally, finally set back
the product, it is the responsibility for the operator to observe and follow these regulations. after dry up.
NOTE: Modifying the product is not permitted.
11.1. Clean the whole body: In the condition of power off, make a clean for the outside
body by soft towel with little mild household cleaner , and then wipe it up with dry
towel, once or twice per month.
Figure 16
Method II: First, remove the silicone tube connected with oxygen output, take down
the humidification bottle, Rotate the air filter with counterclockwise direction as
Figure 17,take down the filter cloth, clean with detergent, wash by clean water, wipe
Figure 14 off the water. and make it to be dry in air naturally ,then install back to the Air filter.
Guidance and manufacturer’s declaration – electromagnetic immunity Recommended separation distances between portable and mobile RF
The 5L machine is intended for use in the electromagnetic environment specified below. communications equipment and the 5L machine
The customer or the user of the 5L machine should assure that it is used in such an The 5L machine is intended for use in an electromagnetic environment in which radiated
electromagnetic environment. RF disturbances are controlled. The customer or the user of the 5L machine can help
IMMUNITY IEC 60601 Compliance Electromagnetic environment – prevent electromagnetic interference by maintaining a minimum distance between
test test level level guidance portable and mobile RF communications equipment (transmitters) and the 5L machine as
Portable and mobile RF recommended below, according to the maximum output power of the communications
communications equipment should be equipment.
used no closer to any part of the 5L Separation distance according to frequency of transmitter
Rated maximum
machine, including cables, than the m
output power
recommended separation distance 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
calculated from the equation applicable of transmitter
to the frequency of the transmitter. W
Recommended separation distance
0.01 0.12 0.12 0.07
16 Quality Warranty
Warranty for whole unit:15 months
LCD
Warranty for magnetic valve:24 months
Warranty for compressor:24 months
Control valve
M M
Fan Compressor
PCB
Oxy. sensor
DISTRIBUTOR:
ADDRESS:
TELEPHONE:
P.F.
FAX:
Transformer
Sensor