MegaSig AM 581 DataSheet - EN-20221013

The document describes the Artificial Mouth 581, which is designed for accurate and repeatable electroacoustic measurements. It has very low distortion, typically less than 1% THD. When used with a PM 0083, it provides a classical solution for quality control testing of frequency response and distortion of devices like mobile phones and microphones. Key specifications include low distortion, impedance of 4 ohms, and diameter and height dimensions.

Uploaded by

hungdv
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
123 views

MegaSig AM 581 DataSheet - EN-20221013

The document describes the Artificial Mouth 581, which is designed for accurate and repeatable electroacoustic measurements. It has very low distortion, typically less than 1% THD. When used with a PM 0083, it provides a classical solution for quality control testing of frequency response and distortion of devices like mobile phones and microphones. Key specifications include low distortion, impedance of 4 ohms, and diameter and height dimensions.

Uploaded by

hungdv
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Equips engineers to create better products

AM 581 Version:1.0_221013_EN

Artificial Mouth 581 is designed for accurate and repeatable electroacoustic


measurements. It has extraordinary distortion performance, typical value is: after
equalization to 94dBSPL, 25mm distance, the THD can be lower than 1%.
PM 0083 + AM 581, is a classical solution for quality control testing of frequency
response and distortion of mobile phone, telephone transmitters and close-talking
microphones.

Features
 Meet the requirements of IEEE269, 661 international standard and "mouth reference point" recommended by ITU-T
P51 for acoustic signal transmission
 Acoustic measurement standard sound source
 Low distortion, high stability

Specification

THD 200Hz – 10kHz:criteria 1%(94dB SPL, 25mmMRP)

Frequency Response After equalization 94dB ± 0.1dB(100Hz - 10kHz)

Anechoic box, smaller 20dB


Compensation value (EQ Range)
Anechoic room, smaller 15dB

Impedance 4 ohm

Continuous output power 10W

Sound ring diameter Ø20mm

Out:Ø51mm
Lip ring
Height: 25mm

Diameter Ø102mm

Height 105mm

Weight 1160g

Interface BNC

Shenzhen MegaSig Co.,Ltd www.megasig.com Page 1 of 3


Equips engineers to create better products

Curve

After EQ, 94dBSPL(20mm MRP)

After EQ, 94dBSPL THD(20mm MRP)

Shenzhen MegaSig Co.,Ltd www.megasig.com Page 2 of 3


Equips engineers to create better products

Dimension

MegaSig reserves the right to change Shenzhen MegaSig Measurement & Control Technology Co.,Ltd
specifications and accessories without
notice.
Tel:0755-8950.839 Fax:0755-8950.8392
Sales email:[email protected]
Technical support email:[email protected]
www.megasig.com Address:Room 1002, Unit 2, Building 1, Hongxing Chuangzhi Plaza,
Guangming District, Shenzhen,China

Shenzhen MegaSig Co.,Ltd www.megasig.com Page 3 of 3