Can Ozt
Can Ozt
KD-TBED series
Electronic weighing scale instruction manual
a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com
Trang 1
LED display : 7 digits, 3. units to be selected ( g/ct/ozt, 1ct =0.2g, 1ozt=31.1034768g) Be able to count; Options : RS 232 intreface; AC or DC power supply, outfit AC/DC adaptor; Pan site 116mm (round , Max.Cap < 1000g) , 125 mm X 145 mm (Square, Max.Cap >1000g)
II. ATTENTIONS. 1. Prevent it from getting wet. 2. Do not place the balance in an environment with extreme temperature or humidity. 3. Do not shock the balance anh do not exceed the capacity. III. Preparations. 1. Before using the Balance, Please take out the protection screw from the left of the balance, then plug up with the circu plastic cap, Otherwise, the balance with not work. 2. The balance must be in an exactly horizontal postion in order to achieve accurate for high precision balance Re-calibration before each use in prefearable Please dont turn the balane on to work before the above items are well prepared IV. EXPLANATION OF DISPLAY PARTS
: Zero Indicator .
: Tare indicator.
: Power Low indicator. : indicator that the reading is stable. a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com
Trang 2
PCS : indicator that the balance is in counting mode. CT : indicator that the unit is Ct. OZT: indicator that the unit Ozt. g : indicator that the unit g.
V. Keypad functions: Phm chc nng. :On /OFF power swich. Phm M ngun/ Tc ngun : For weigh unit selection (g /ct/ ozt).
Phm chuyn i n v
Sample key, used for samplinh (for calculating the unit weight) in counting : mode. Phm chc nng m s lng PCS : Tare key, used to subtract the weight of an item or container. Press tare key again to exit the tare mode (when empty). Phm tr b : Zero key, press this key to return the display to zero if a small weight is left while unloaded. Phm tr v khng VI. Operations: Press key to turn the balance on ; the default mode is weigh mode.
Weight mode Weight units Press Tare function key to choose the needed unit.
Put a container on the pan and press when the display reading is stanble. The tare weight will be stored into memory and dislay will be brought to zero. Tare indicator in the display will appear, The weight added here aferwards and the objects, the display will show the weight of the container companied by a
a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com
Trang 3
negative sign. To cancel the tare mode, Press the Tare key again to cancel Tare mode. Tare range : Up to balances maximum capacity. Zero fucction key to retur the display to the center of zero if the zero Shifts during Press operation. Counting mode Sampling
1. Press key, the display will show Cnr 10 .Press the key again and again, 10,20,50,100 (pcs) will appear in succession. Stop at the one you want to use. ,the 2. Put the exact quantity of samples as desired on the pan and press determined sample size will be shown.After sampling, its ready to count. 3. If the unit weght is too small ( less than 0.25d) for the counting resolution,
rr will be show.
4. In counting ,ode, press VII. Error messages: When the display shows - -- - - -- and beep out a warning, it means that the balance is overloaded. Please remove the object from the pan immediately so as to avoid damage to the load sensor inside the lalance. VIII. Instructions of RS 232 communication(optional) 1. Model EIA- RS232 Cs UART signal 2. Format (1) Baud rate : 2400 bps (1200 bps, 4800bps,9600bps) (2) Data bits :8 bits (3) Parity bit : none (4) Stop bit : 1 bit (5) Code ASCLL DATA RORMAT: HEAD1, HEAD2 DATA UNIT a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com key will ture to weght mode.
CR
Trang 4
CNG TY CN IN T THNH PHT 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 HEAD1 (2BYTES ) HEAD2 (2BYTES) Us unstable NT net weght mode ST stable GS GROSS WEGHT MODE OL overload DATA (8BYTES) 2D (HEX) =- (negative sign) 20 (HEX) = (Blank) 2E (HEX) =- (Decimal point) UNIT (4 BYTES) G = 20 (HEX) ; 20 (HEX) ; 20(HEX) ; 67 (HEX) CT = 20(HEX) ; 20(HEX) ; 63 (HEX) ; 74 (HEX) OZT = 20 (HEX) ; 6F (HEX) ;7A (HEX) ;74(HEX) Transmission example: Ex: stable net + 0.168 g HEAD1, HEAD2, DATA, ST , NT, +0.168 UNIT g CR 0D0A
IX. Power supply Alternative Power Supplies 1. DC 6V/1.3Ah rechargeable sealed lead- acid battery. 2. 8.5 V / 0.2A AC/DC power adaptor. Low voltage indication When the voltage of the batterries is lower than 5.8 V +(-)0.15V the Power LOW LED will be lighted up. The lead- acid battery should be rechanged. Or elese, the display reading will be unstble and the balance will auto power-off when the voltage of the bstteries lower than 5.1V +(-)0.15V. X. Calibration (1) Press and hold key when press on, it will display CAL
(2) After the stable indicator LED lightd, press key, the screen will show000000 (3) To key in the value of the weght to be placed on the pan, press key to shift the twinkling number; Press fraction is invalid) to increase the value of twinkling number.(decimal
a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com
Trang 5
(4) Place a weight which value equal to the value just entered and after the reading is key to complete external calibration, and the screen show the value stable, press of the weight Note: The weight placed on the pan is not correct if the display shows Err 1 1 .M ngun v nhn gi phm 2. Sau tip tc nhn phm 3. Gii s bng phm cho ti khi mn hnh xut hin ch CAL , mn hnh xut hin 000000
. 4. Tng s bng phm 5. B ti hin c ln mt bn cn , t nht 1/3 ti max ca cn 6. Tip tc nhn phm th mn hnh xut hin trng lng ti ca qu cn. 7. Kt thc qu trnh canh chnh ca cn. XI. Setting filter parameters:. 1. Press and hold the key until display nb0 or nb1 ornb2 ornb3ornb4 ornb5 ornb6 ornb7. 2. Select the applicable parameter by pressing key, and confirm the setting by pressing key. The parameters nb0den nb4 means that response speed is slow den fast. 3. After the setting of filter parameters, the balance will turn to the selection of divisions. First, the current divisions will be show.You can use key to select and key to confirm.Then it will turn to only which is large than the current division.
XII. Setting of Zero parameters and communiccation parameters 1. Press and hold the key to enter the setting of zero tracking range. The display will show 0.0d or 0.5d or1.0d or1.5d or2.0d or3.0d after self-test. 2. Press key to select and press key to confirm and go to the setting of zero display range. The display on the screen is Zer-S (invalid) or Zer-L (display 0 while the weight within the range of +(-)d) 3. Press key to select and press key to confirm and go to the setting of communication mode. The display on the screen is St (output when the reading is stble)or CO (continue output)
a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com
Trang 6
4. Press key to select and press key to confurm and go to the setting of baud rate. The display on the screen is 1200or2400or4800or9600 5. Same as above, Press key to select and press key to confirm. XIII. Single- double range selection Press and hold key when power on, it will display the curret set, such asSIN, press key to select the applicable parameter (SIN single range, dbl- double range), then press key to confirm. MODEL Weighing capacity and readability (g) Standard deviation (e) Linearity (e) Corner deviation (e) Taring range Overload capacity Time of stabilizing (s) Applicable temperature rang Power source Pan size (mm) Housing dimension (mm) Net weight(g) KD TBED 300g 300g0.01g 1 1 1 0-300g KD TBED 600g 600g X 0.01g 1 1 1 KD TBED 1200g 1200g X 0.01g 1 1 1 KD TBED 3000g 3000g X 0.1g 1 1 1 0-3000g
AC adapter(supplied with balance); 116 < 1000G 124 X 144 > 1000g 251 (L) 172 (W) 58 (H) 272 (L) 176 (W) 58 (H) 1050 1000
www.canthuongmai.com
a Ch: 17 C5 ng D1, Phng 25 Qun Bnh Thnh , TP.H Ch Minh Tel : (08) 512.7131, 2437586, 0908.486.774 Fax (08) 512.7966 Email: [email protected] Website : www.canthuongmai.com
Trang 7