IQ Muffle Furnace
IQ Muffle Furnace
S. No Contents Page No
2 Object 2
3 Scope 2
4 Responsibilities 2
6 Instrument specification 4
8 Packing List 6
10 Instrument safety 8
12 Summary report 10
Page 1 of 10
Installation Qualification
Objective:
To qualify the Vulcan A 130 Muffle furnace, for use in residue on ignition testing for quality control analytical
weighing.
Scope:
The scope of this report is limited to Vulcan A 13 Muffle furnace,
Responsibilities:
Vendor
Purchaser/ Buyer
Page 2 of 10
Installation Qualification
Instrument Vulcan A 130 Muffle furnace
Manufacturer Vulcan
Tel: +92-42-35846571
Service Contact Tel/Fax: +92-42-35864077
E-Mail: [email protected]
Serial Number
New
Condition when received Used ✓
Date Received
Instrument Specification
S. No User Requirements Desired features Functions Decision
Page 3 of 10
Installation Qualification
Pass Fail
The temperature limit for
1 Temperature range 25ºC to 1100 ºC ✓
sulphated ash measurement
To check the equipment
2 Temperature resolution 50ºC performance is robust, accurate ✓
and precise during testing
Common options available on
3 Switch ON and OFF Buttons Should be available equipments to switch on and ✓
switch off whenever required
Comments: the equipment received as per URS and USP Specified standards
Page 4 of 10
Installation Qualification
Requirements Pass Fail
Ambient ✓
1 (22 to 26) ºC Our lab common temperature range
temperature of Lab
Max relative ✓
2 NMT 56% The humidity threshold of lab set
humidity
Supply voltage and 220V -240V 50/60HZ Electric supply requirement of muffle ✓
3
frequency 1060W furnace
No direct sun light Temperature fluctuations, including those
4 Sun light ✓
exposure caused by direct sunlight
The equipment should be designed to
Should be inert to
5 Reactive gases
reactive gases
avoid any corrosive reaction with reactive ✓
gases
Packing List
Page 5 of 10
Installation Qualification
S. No Description Quantity
1 Muffle furnace 1
2 Power Cord 1
3 Instruction Manual 1
Page 6 of 10
Installation Qualification
Unpack the equipment carefully, Are any items missing against packing list?
Yes No ✓
Yes No ✓
Instrument Safety
Page 7 of 10
Installation Qualification
The humidity of the location and the risk of pronounced changes in the relative
✓
humidity
✓
Dust and other foreign bodies
Magnetic forces ✓
✓
Position on level, noncombustible surface indoor
✓
Area free from aggressive/ explosive chemical mixture
✓
Ensure correct power supply and all power switches easily accessible
✓
Do not use to heat material that could cause fire or hazard
✓
Read operator manual before use
Page 8 of 10
Installation Qualification
Assembly and User Specialized Engineer ✓
installed by
✓
Unpack and retain packaging
✓
Assembly
✓
Site and content to power supply
✓
As mentioned on instruction manual
Page 9 of 10
Installation Qualification
Model No: Vulcan A 130
Summary Report:
Remarks: successful completion of the proceeding activities and checks indicate that this instrument has been
successfully delivered and installed. The instrument has passed installation qualification procedure and may
now be submitted for operation qualification