Week 2 Part 1 Microscope
Week 2 Part 1 Microscope
PRINCIPLES OF MICROSCOPY
• In the compound microscope, a magnified
intermediate image of the illuminated
specimen is formed in the optical tube by
each objective lens.
• This image is then magnified again and
viewed through the eyepiece as an enlarged
virtual image that appears to be located
about 10 inches from the eye.
• Microscopists must focus their eyes in that Compound microscope
more distant plane, rather than trying to • Two separate lens systems are used
focus at the distance of the microscope (objective and eyepiece).
stage.
• Each objective lens forms a magnified image 4. The neck, or arm— provides a structural
of the illuminated specimen in the optical site of attachment for the revolving
tube. nosepiece.
• Eyepiece lenses further magnify this image 5. The stand— is the main vertical support of
so that the microscopist sees an enlarged the microscope. The stage assembly,
virtual image that appears to be together with the condenser and base, is
approximately 10 inches from the eye. supported by the stand.
• It employs two separate lens systems; 6. The revolving nosepiece holds the
objective and eyepiece, the product of objectives and allows for easy rotation from
which produces the final magnification. one objective lens to another.
• Standard microscopes use bright-field - The working distance (WD) between the
illumination in which light passes through objectives and the slide varies with the
the thin specimen make and model of the microscope.
• There are usually three or four objective
COMPONENT PARTS AND THEIR FUNCTIONS lenses , each with a specific power of
1. The eyepieces or oculars - usually are magnification.
equipped with 103 lenses (degree of • Engraved on the barrel of each objective
magnification is 103). lens is the power of magnification and the
- The lenses magnify the intermediate numerical aperture (NA).
image formed by the objective lenses in • The NA is related to the angle of light
the optical tube; they also limit the area collected by the objective; in essence, it
of visibility. indicates the light gathering ability of the
- may have either one or two adjustable objective lens.
eyepieces. ➢ Functionally, the larger the NA, the
- All eyepieces should be used correctly greater the resolution or the ability
for optimal focus. to distinguish between fine details
- Eyepieces should not be interchanged of two closely situated objects.
with the eyepieces of the same model • Four standard powers of magnification and
or other models of microscopes, NA used in the hematology laboratory are;
because the eyepieces in a pair are Low power 103/0.25
optically matched. High power, dry 403/0.65 or 453/0.66
2. The interpupillary control is used to adjust
OI (oil Immersion) 503/0.90
the lateral separation of the eyepieces for
OI 1003/ 1.25
each individual. When it is properly ▪ The smaller the magnification, the
adjusted, the user should be able to focus larger the viewing field; the larger the
both eyes comfortably on the specimen and magnification, the smaller the viewing
visualize one clear image. field.
3. The optical tube connects the eyepieces ▪ Total magnification= magnification of
with the objective lens. The intermediate the eyepiece X magnification of the
image is formed in this component. The objective lens. (eyepiece X objective
standard length is 160 mm, which,
lens)
functionally, is the distance from the real
image plane (eyepieces) to the objective
lenses.
image contrast. Closing it beyond this Tungsten possesses a high melting point
point leads to a loss of resolution. and gives off bright yellowish light. A blue
- The stage and condenser consist of a (daylight) filter should be used to eliminate
swing-out lens, an aperture diaphragm, the yellow color produced by tungsten. The
a control for vertical adjustment of the rheostat or light control knob or lever turns
condenser, and two centering screws on the light and should be used to regulate
for adjustment of the condenser. the brightness of the light needed to
- The condenser top lens can swing out of visualize the specimen. The aperture
position. diaphragm control lever should never be
- The stage controls located under the used for this purpose, because closing it
stage move it along an x- or a y-axis. reduces resolving ability.
10. The field diaphragm is located below the
condenser within the base. OPERATING PROCEDURE WITH KOEHLER
- When it is open, it allows a maximally ILLUMINATION
sized circle of light to illuminate the The procedure outlined here applies to
slide. microscopes with a nonfixed condenser. The
- Almost closing the diaphragm, when following steps should be performed at the start
low power is used, assists in centering of each laboratory session in which the oil
the condenser apparatus by the use of objectives will be used:
two centering screws. 1. Connect the microscope to the power
- Some microscopes have permanently supply.
centered condensers, whereas in others 2. Turn on the light source with the power
the screws are used for this function. switch.
- The glass on top of the field diaphragm 3. Open the condenser aperture and field
protects the diaphragm from dust and diaphragms.
mechanical damage. 4. Revolve the nosepiece until the 103
• Microscopes depend on electricity as the objective lens is directly above the stage.
primary source for illumination power. 5. Place a stained blood film on the stage and
There are two types of brightfield focus on it, using the fixed eyepiece, while
illumination: (1) critical illumination, in covering the other eye. (Do not simply close
which the light source is focused at the the other eye, because this would
specimen, which results in increased but necessitate adjustment of the pupil when
uneven brightness; and (2) the Koehler (or you focus with the other eyepiece.)
Köhler) system, in which the light source 6. Adjust the interpupillary control so that
and the condenser are properly aligned. The looking through both eyepieces yields one
end result of Koehler illumination is a field clear image.
of evenly distributed brightness across the 7. Using the adjustable eyepiece and covering
specimen. This is especially important when the opposite eye, focus on the specimen.
using the oil objectives or when taking Start with the eyepiece all the way out, and
photomicrographs. Tungsten-halogen light adjust inward. If using two adjustable
bulbs are used most frequently as the eyepieces, focus each individually.
illumination source. They consist of a 8. Raise the condenser to its upper limit.
tungsten filament enclosed in a small quartz
bulb that is filled with a halogen gas.
9. Focus the field so that the cells become and its chromatin pattern; the cytoplasm
sharp and clear. Concentrate on one cell and its color and texture. The objective lens
and place it in the center of the field. should dip into the oil slightly.
10. Close the field (lower) diaphragm. Look
through the eyepieces. A small circle of light Considerations
should be seen. If the light is not in the 1. When revolving the nosepiece from one
center of the field, center it by using the power to another, rotate it in such a
two centering screws located on the direction that the 103 and 403 objective
condenser. This step is essential, because an lenses never come into contact with the oil
off-center condenser will result in uneven on a slide. If oil inadvertently gets onto the
distribution of light. Adjust the vertical high dry objective, clean the objective
height of the substage condenser so that immediately.
you see a sharp image of the field 2. Parcentric refers to the ability to center a
diaphragm, ringed by a magenta halo. If the cell in question in the microscopic field and
condenser is raised too much, the halo is rotate from one magnification power to
orange; if it is lowered too far, the halo is another while retaining the cell close to the
blue. center of the viewing field. Recentering of
11. 11. Reopen the field diaphragm until it is the cell at each step is minimal. Most
nearly at the edge of the field, and fine-tune laboratory microscopes have this feature.
the centering process. 3. In general, when the 103 and 403 objective
12. Open the field diaphragm slightly until it lenses are used, the light intensity should
just disappears from view. be low. When the 503 and 1003 objective
13. Remove one eyepiece and, while looking lenses are used, increase the intensity of
through the microscope (without the the light by adjusting only the rheostat (light
eyepiece), close the condenser aperture control knob or lever) or by varying neutral
diaphragm completely. Reopen the density filters. Neutral density filters are
condenser aperture diaphragm until the used to reduce the amplitude of light and
leaves just disappear from view. Replace the are available in a variety of densities.
eyepiece. 4. Do not change the position of the
14. Rotate the nosepiece until the 403 objective condenser or the aperture diaphragm
lens is above the slide. Adjust the focus (the control lever to regulate light intensity when
correction should be minimal) and find the viewing specimens with the oil immersion
cell that you had centered. If it is slightly off objectives. The condenser should always be
center, center it again with the stage x-y in its upward position as set during the
control. Note the greater amount of detail Koehler illumination adjustment. The
that you can see. aperture diaphragm may be adjusted to
15. Move the 403 objective out of place. Place a achieve proper contrast of the features of
drop of immersion oil on top of the slide. the specimen being viewed.
Rotate the nosepiece until the 1003 5. After setting the Koehler illumination, when
objective lens is directly above the slide. a new slide is to be examined, always bring
Avoid moving a non–oil immersion objective the specimen into focus with the 103
through the drop of oil. Adjust the focus objective first, and then move to the higher
(the correction should be minimal) and magnifications.
observe the detail of the cell: the nucleus
IMMERSION OIL AND TYPES removing the dust may scratch the lens.
• Immersion oil is required to increase the Do not use laboratory wipes or facial
refractive index when either the 503 or the tissue to clean the lenses.
1003 oil immersion objective lens is used. 3. Avoid placing fingers on the lens
The refractive index is the speed at which surface. Fingerprints affect the contrast
light travels in air divided by the speed at and resolution of the image.
which light travels through a substance. This 4. Use solvent sparingly. The use of xylene
oil, which has the same properties as glass, is discouraged, because it contains a
allows the objective lens to collect light carcinogenic component (benzene).
from a wide NA, which provides high Xylene is also a poor cleaning agent,
resolution of detail. leaving an oily film on the lens. Lens
• Three types of immersion oil, differing in cleaner or 70% isopropyl alcohol
viscosity, are employed in the clinical employed sparingly on a cotton
laboratory: applicator stick can be used to clean the
1. Type A has very low viscosity and is objective lenses. Alcohol should be kept
used in fluorescence and darkfield away from the periphery of the lenses,
studies. because alcohol can dissolve the
2. Type B has high viscosity and is used cement and seep into the back side of
in brightfield and standard clinical the lens.
microscopy. In hematology, this oil 5. When fresh oil is added to residual oil
is routinely used. on the 1003 objective lens, there may
3. Type C has very high viscosity and is be loss of contrast. Clean off all residual
used with inclined microscopes with oil first.
long-focus objective lenses and 6. Do not use water to clean lenses. If no
wide condenser gaps. lens cleaner is available, use a clean
• Bubbles in the oil tend to act as prisms and microfiber cloth.
consequently reduce resolution. Bubbles 7. When transporting the microscope,
may be created when oil is applied to the place one hand under the base as
slide. They are caused by lowering the support and one hand firmly around the
objective immediately into the oil. Sweeping arm.
the objective from right to left in the oil • In addition to daily care of the microscope,
eliminates bubbles. semiannual or annual maintenance with
thorough cleaning should be done by a
CARE OF THE MICROSCOPE professional. Microscope professionals may
Care of the microscope involves the following recognize and correct problems with
details: mechanics or optics before they are
1. When not in use for an extended period detected by the microscope user. They can
of time, always cover the microscope to correct problems such as sticking of stage
protect it from dust. controls or incorrect optical alignment that
2. Before use, inspect the component can lead to physical problems like carpal
parts. If dust is found, use an air syringe, tunnel syndrome and headaches.
a camel hair brush, or a soft lint-free
cloth to remove it. Using lens paper BASIC TROUBLESHOOTING
directly on a dirty lens without first
Most common problems are related to inability installation of an annular diaphragm in the
to focus. Once the operator has ensured that he condenser, together with a phase-shifting
or she is not trying to obtain a “flat field” using element, creates excellent contrast of a cell
an objective lens that is not plan achromatic, against its surrounding background.
the following checklist can aid in identifying the • The principle of phase contrast is related to
problem: the index of refraction and the thickness of
• Eyepieces Clean? a specimen, which produce differences in
• Securely assembled? the optical path. Light passing through a
• Objective lens transparent specimen travels slightly slower
• Screwed in tightly? than light that is unobstructed. The
• Dry objective free of oil? difference is so small that it is not
• Condenser noticeable to the viewer. When a
• Adjusted to proper height? transparent phase plate is placed into the
• Free of oil? microscope, however, the change in phase
• Slide can be increased to half a wavelength,
• Correct side up? which makes the otherwise transparent
objective visible
• Coverslip
• Correct side of blood film?
• Only one coverslip on slide?
• Free of mounting media?
• Light source • This phase difference produces variation in
• Fingerprints on bulb? light intensity from bright to dark, creating
• Bulb in need of changing? contrast in the image. Often the objects
• Light source aligned correctly? appear to have “haloes” surrounding them.
• In hematology, phase-contrast microscopy is
OTHER MICROSCOPES USED IN THE CLINICAL employed in counting platelets in a
LABORATORY hemacytometer, since they are difficult to
Phase-Contrast Microscope visualize and count using brightfield
• The ability to view a stained specimen by microscopy. It also can be used to view
the use of brightfield microscopy is affected formed elements in unstained urine
by two features: sediments.
1. the ability of the specimen to absorb Polarized Light Microscope
the light hitting it, and • Polarized light microscopy is another
2. the degree to which light waves contrast-enhancing technique used to
traveling through the specimen remain identify substances such as crystals in urine
in phase and other body fluids. With brightfield
microscopy, light vibrates in all directions. If
a polarizer (filter) is placed in the light path,
the light vibrates in only one direction or
• Specimens that are transparent or colorless, plane, which creates polarized light. To
such as unstained cells, are not clearly convert a brightfield microscope to a
visualized with brightfield microscopy. polarizing one, two filters are needed. One
Phase-contrast microscopy, through the filter (the polarizer) is placed below the
condenser and allows only light vibrating in This diffracted light is picked up by the
the east-west direction perpendicular to the objective lens and appears as bright detail
light path to pass through the specimen. on a black background. Darkfield
The second filter (the analyzer) is placed microscopy is helpful in microbiology in the
between the objective and the eyepiece identification of spirochetes.
and allows only light vibrating in a north-
south direction to pass to the eyepiece.
When the transmission axes of these two
filters are oriented at right angles, no light Specimen collection and
can pass through the pair to the eyepieces. Bacterial Identification
When polarized light (vibrating in an east-
west direction) passes through an optically
active substance such as a monosodium
urate crystal, however, the light is refracted Basic Principles of Specimen Collection
into two beams, one vibrating in the original 1. Collect the specimen in the acute phase
direction (east-west) and one vibrating in a of the infection and before antibacterial
plane 90 degrees to it (i.e., north-south). antibiotics are administered
The refracted light vibrating in the north- 2. Select the correct anatomic site for
south direction can pass through the second collection of the specimen
filter (the analyzer) and is visible at the 3. Collect the specimen using the proper
eyepiece. The magnified crystal appears technique and supplies with minimal
white against a black background. If a first- contamination from normal biota
order red compensator filter also is placed (normal flora)
in the light path below the stage, the 4. Collect the appropriate quantity of
background becomes pink-red, and the specimen
crystal appears yellow or blue, depending 5. Package the specimen in a container or
on its physical orientation relative to the transport medium designed to maintain
incident light path (east-west). Some the viability of the organisms and avoid
crystals can be specifically identified based the hazards that may result from
on their unique birefringent (doubly leakage
refractive) characteristics when polarizing 6. Label the specimen accurately with the
microscopy is used. specific anatomic site and the patient
Darkfield Microscope information – patient’s name and
• Darkfield microscopy is a contrast- unique identification number
enhancing technique that employs a special 7. Transport the specimen to the
condenser. The condenser sends light up laboratory promptly or store the
toward the specimen in a hollow cone. specimen in a transport medium
Because of the high angle of this cone, none 8. Notify the laboratory in advance if
of the illuminating rays enters the objective unsual pathogens or agents of
lens. Without the specimen in place, the bioterrorism is suspected
field would appear black because of the
absence of light. When the specimen is in Collection Procedure
place, and if fine detail exists in the • Clinical specimens should be collected
specimen, light is diffracted in all directions. in sterile containers except for stool
Microscopic Observation
1. Cocci are spherical cells, bacilli are rod
shaped. Bacteria that have thick cell
walls are termed as Gram positive
because of the way they take up the
Gram stain. Those with thin walls are
termed as Gram negative
2. The Gram positive bacteria are bacteria
with thick cell walls containing the
teichoic acid retain the crystal violet –
iodine complex dye after decolorization
and appear as deep blue
3. The Gram negative bacteria are bacteria
with thinner cell walls containing
lipopolysaccharides (LPS) that do not
retain the dye complex
Stains
Stains can be categorized as: GRAM STAIN
Kinyoun Stain
1. Cover the smear with carbolfuchsin and
Material for Fluorochrome Stain stain for 5 minutes
▪ TB auramine-rhodamine stain 2. Wash slides with running tap water
▪ Carbofuchsin stain (either commercially 3. Move slides to staining rack on acid
prepared or laboratory prepared) alcohol collection container
▪ 0.5% acid alcohol (0.5% HCl in 70% ethanol)
4. Decolorize with acid alcohol until no GREEN; if counter stain is Methyline Blue-
more color appears in the washings background is BLUE).
5. Wash slides with running tap water and
move them to original staining rack General Rules for Gram Staining
6. Flood slides with methylene blue • All cocci are gram positive except Neisseria,
counterstain for 1 minute Branhamella, Moraxella, and Velionella.
7. Wash with running water, drain and air • All bacilli are gram negative except
dry Mycobacteria, Clostridium,
8. Examine the smears with oil immersion Corynebacterium, Bacillus, Erysipelothrix,
objective x100 Listeria, Lactobacillus.
Modified Kinyoun Stain (Partial Acid-Fast) • Higher forms of organisms (Actinomyces,
1. Flood the slides with carbofuchsin stain Streptomyces, yeasts, and molds) are gram
for 5 minutes positive.
2. Rinse with running tap water • All spiral organisms are reported as gram
3. Flood slides with 70% ethanol and rinse
with tap water. Repeat until excess red Reason why Gram (+) becomes (-)
dye is removed • Overdecolorization
4. Move slides to rack on acid collection • Smear too thick
container • Improper washing between steps
5. Continuously drop 1% sulfuric acid on • Too old Culture
the smear • Impure or mixed culture
6. Rinse with running tap water Reason why Gram (-) becomes (+)
7. Move slides to original staining rack • Underdecolorization
8. Counterstain with methylene blue for • Smear too thick
30 seconds
Acid-fast Stain
• All bacteria are non-acid fast except
Zeihl–Neelsen Stain (HOT METHOD) Mycobacteria group
1. Cover the smear with a piece of filter • One does not belong to the
paper cut slightly smaller than the slide Mycobacteria group but is slighty acid
2. Layer filter paper with carbofuchsin fast: Nocardia asteroides
stain, with a bunsen burner, heat the
• Acid Fast organism (AFO) are very hard
smears gently until steaming occurs.
to stain due to the presence of
Stain for 5 minutes without additional
unsaponifiable wax called mycolic acid
heating
or hydroxymethoxy acid.
3. Proceed as for Kinyoun stain, beginning
• In staining, the mycolic acid is
with Step 2
temporarily removed through the
steaming process.
Result for Kinyoun stain and Ziehl-Neelsen stain
▪ Mycobacteria stain red
▪ Background material and non-acid-fast
bacteria stain either blue or green.
(Depending on the counterstain used; if
uses Malachite Green- background is
Biochemical Testing
Biochemical Tests
• Are never done in mixed plates
• Pure cultures are used
Carbohydrate Fermentation
• Fermentation – utilization of CHO by
bacteria
• In bacteriology, this process is detected
by observing changes in pH indicator as
acid products are formed
Gas Production
• (+) bubble formation or splitting of the INDOLE TEST
media or complete displacement of the • Based on the ability of organism to produce
media from the bottom of the tube indole from tryptophan
• Media: Tryptophan broth/SIM broth
Beta-galactosidase and ONPG test • Reagent: Ehrlich’s
(Orthonitrophenyl beta-D-galactopyranoside) Reagent/KOVAC’s/Paradimethylbenzaldehyd
• Is a compound structurally similar to lactose e
test that detects the enzyme beta- • Result: (+) pink to wine colored ring at the
galactosidase junction, (-) no color development
• Used to distinguish enteric bacteria • Rapid Spot Indole test (blue) – filter paper
(Salmonella (-), Citrobacter (+)) and to strips impregnated with p-
identify Pseudomonas amminocinamaldehyde reagent, screening
• Orthonitrophenol is a chromophore that is for indole production
released into the medium and detected by a Interpretation Developmen of cherry red color
pale yellow color at the surface of the reagent and broth within
• (+) yellow, (-) no color change seconds after adding KOVAC’s reagent indicates
the presence of indole and the test is positive. If
SIM TEST no color change is observed, then the test is
• Indole production – ability of organism to negative
split tryptophan to form compound indole
• Result (+) pink to wine red colored ring
• Sulfide = (+) blackening of agar = Motility
VOGES-PROSKAUER TEST
• Principle: This test determines the
capability of some organisms to produce
non-acidic or neutral end products, such as
acetyl methyl corbinol (acetoin), from the
organic acid that results from glucose
metabolism. This test identifies bacteria
that ferments glucose and leading to 2,3-
butanediol accumulation in the medium.
OXIDASE TEST (CYTOCHROME
• Reagent: alpha napthol and KOH (Barritt’s OXIDASE/INDOPHENOL BLUE)
Medthod) – (+) pink to red color (E. cloacae)
• Principle: This test uses certain reagent dyes
/ (-) no color change (E. coli)
such as p-phenylenediamine
• Reagent: alpha napthol and 40% KOH in dihydrochloride that substitute for oxygen
creatine (Coblentz) – (+) red (S. mutans) / (-) as artificial electron acceptors. It is colorless
yellow (S. minis) in reduced state. In the presence of
cytochrome oxidase and atmospheric
oxygen , p-phenylenediamine is oxidized
forming indophenol blue.
o (+) bluish purple – P. aeruginosa
o (-) no color change – E. coli
H2S Production
• Differentiating test for Salmonella and
Shigella
• Systems for H2S detection:
o Lead acetate paper
o SIM tube
o Hektoen and SS agar Mug Test
o XLD agar • Uses 4-methylumbelliferyl-beta-D-
o TSI glucoronide
• (+) – Salmonella, Edwardsiella, Citrobacter, • (+) blue fluorescence – E. coli
Proteus • (-) no fluorescence – P. Aeruginosa
• Result: (+) production of black color
Motility Test
• Ability of an organism to produce proteases
that hydrolyzes gelatin and liquefy solid
gelatin medium.
• Used in the identification of Clostridium, KCN Broth Test
Serratia, Flavobacterium, and Pseudomonas • This test determines whether the microbe
• (+) result gel liquifies / (-) gel solidifies can grow in a medium where potassium
cyanide is present as a carbon and nitrogen
Nitrate Reduction Test source.
• Principle: This test determines • Result (+) tubid; (-) clear
microorganisms that can utilize nitrate as a
terminal electron acceptor during anaerobic
respiration. Nitrate is reduced to nitrite by
nitrate reductase.
the production of alkaline metabolites that will bind to the surface of the
induce a color change organisms
• Result (+) blue; (-) green • Used to identify bacteria that are difficult to
cultivate in a culture medium
• +Result: Presence of visible clumpls or
agglutination
• Interpretation: Specific antibodies bind to
the bacterial surface antigens, causing the
organism to clump
• S. aureus organism that contains protein A • Commonly used for the rapid identification
in its cell walls is utilized in this procedure of bacterial and viral antigens in body fluids,
as part of the reagent swabs, and infected cells.
• Monoclonal or polyclonal antibodies, which
are attached to fluorescent dyes, are
Complement Fixation
applied to an antigen that is previously
• Composed of two parts: a test system which treated with either formalin, acetone, or
includes the antigen that causes the disease alcohol, ten the reaction is visualized under
in the patient serum and an indicator the fluorescent microscope where final
system which consists of the sheep's red results can be released within an hour.
blood cells, complement-fixing antibody like • Borrelia, Legionella, Mycoplasma
the immunoglobulin G(IgG), and an pneumonia, Rickettsia, and TORCH
exogenous complement. • Example of fluorescent dyes: Auramine,
• +Result: Absence of lysis with the red blood rhodamine, and fluorescein
cells which means presence of antibody in isothiocyanate(FITC)
the patient serum • +Result: Occurrence of fluorescence as seen
• •-Result: Occurrence of lysis with the red under the microscope
blood cells and the absence of antibodies in • Types: Direct Fluorescent Antibody (DFA)
the test serum Test and Indirect Fluorescent Antibody (IFA)
Test
Flocculation Test
• uses a soluble antigen that reacts with the
antibody in a serum sample. Types of Western Blot
• +Result: Occurrence of macroscopic or Pulsed-field Gel Electrophoresis
microscopic flocculation (precipitation) • Enzyme-digested chromosomal fragments
• Some examples of this test are venereal of bacteria are isolated electrophoretically
disease research laboratory (VDRL) test and in this approach.
rapid plasma reagin (RPR) test. • Can be used in an outbreak investigation.
Bacteriophage Typing
Enzyme-linked Immunosorbent Assay (ELISA) • Based on the specificity of phage surface
• A sensitive and specific method for the receptors for cell surface receptors
detection of antibodies against certain • A bacteriophage is a virus that attacks
pathogens. bacteria.
• This method consist of enzyme-bound Chromatographic Method(Gas Chromatography
antibodies with the antibody-binding sites and High Performance Liquid Chromatography)
free to react with their specific antigen. • Analysis of microbial metabolites, cellular
• Utilized for the diagnosis of infectious fatty acids, and products of pyrolysis of the
diseases such as those caused by Legionella whole bacterial cells.
• Alkaline phosphatase or horseradish • Gas-liquid chromatography is used to detect
peroxidase is used as an enzyme-conjugate the cellular fatty acids of anaerobes.
antibody reagent. • HPLC is utilized for the fatty acid analysis of
• +Result: Colored end product the Nocardia species.
Immunofluorescent Assays