Guide to Compressed or Liquefied Gases in Laboratories
Table of Contents
1.0 Introduction
2.0 General Precautions
2.1 Types of Compressed Gas Cylinders
2.2 Securing Cylinders
2.3 Lecture Bottles
2.4 Storage
2.5 Return and Transport
2.6 Connections
3.0 Piping and Manifolds
4.0 Toxic Gases
5.0 Flammable Gases
5.1 Acetylene
5.2 Hydrogen
6.0 Cryogenic Liquids and Liquefied Gases
6.1 Dewars and Transfer Equipment
6.2 Monitoring for Oxygen Deficiency
6.3 Minimum Ventilation Rate
7.0 Quantity Limits on Campus
1.0 Introduction
Hazards can result from improper handling of gas cylinders and high pressure equipment. A leaking
cylinder could produce an atmosphere that is toxic, anesthetic, asphyxiating, or explosive; and in the
event of a rapid escape, the cylinder becomes a randomly directed missile. The main purpose of
properly handling compressed gases is, therefore, to prevent uncontrolled escape of the gas.
A compressed gas is defined by the Department of Transportation (DOT) as “any material or mixture
which exerts in the packaging an absolute pressure of 280 kPa (40.6psia) or greater at 20C (68F).”
2.0 General Precautions
• Never drop cylinders or permit them to strike each other violently.
• Do not expose cylinders to temperatures higher than 50° C. Some rupture devices on cylinders
release at about 65° C.
• Never tamper with pressure relief devices in valves or cylinders.
2
• Before using cylinders, read all label information and Safety Data Sheet associated with the gas
being used.
• Always use a cylinder dolly when moving cylinders.
2.1 Types of Compressed Gas Cylinders
There are three major groups of compressed gases stored in cylinders.
• Liquefied gases are partially liquid at normal temperature and charge pressure. Examples:
chlorine, propane, nitrous oxide.
• Non-liquefied gases are entirely gaseous at normal temperatures regardless of charge pressure.
Examples: argon, oxygen, nitrogen. The standard 5 foot gas cylinders supplied by gas vendors at
a pressure of 2,200 – 2,400psi, contain an average of 250 cu. ft. of gas at normal temperature.
• Dissolved gases are dissolved in a liquid phase solvent. Dissolved gas cylinders are packed with
an inert, porous filter saturated with the solvent which stabilizes the volatile gas. Acetylene is
the only common dissolved gas.
Some gases, such as carbon dioxide, are commonly used in both a liquid and gas form. Cylinders
designed for liquid phase dispensing have a siphon, or "dip", tube.
2.2 Securing Cylinders
Keep cylinders secured to the bench or wall and to keep the caps on when they are not in use. Cylinders
may be chained to the wall, secured by bench straps, or held by commercial floor stands. Bungee cords,
zip ties, ropes or wires are not acceptable means of securing cylinders.
2.3 Lecture Bottles
In addition to standard precautions, the following special rules apply to work with lecture bottles in the
laboratory:
• Lecture bottles do not have pressure-relief devices to prevent rupturing or a transport cap.
3
• Unlike larger cylinders, lecture bottles all have identical valve threads, irrespective of the gas
contained within.
• If labels and valve tags do not agree or if there is any question as to the contents of a lecture
bottle, return the unused bottle to the supplier. Whenever possible, purchase lecture bottles
from suppliers who will accept the return of empty or partially empty bottles.
• When transporting lecture bottles, use a cart and block the bottles to prevent rolling and falling.
2.4 Storage
Maximum allowable storage quantities vary depending on building, floor, control area, fire rated design
and type of gas. Contact the College of Engineering Director for Safety for assistance with determining
allowable storage in your lab. When a cylinder is not in use, close the main cylinder valve tightly and add
the protective cap.
Promptly remove the regulator from an empty cylinder, replace the protective cap, and label the
cylinder by using an "empty" tag or writing on the side of the cylinder with chalk. Never bleed cylinders
completely empty; leave a slight pressure to keep contaminants out. Empty cylinders shall be promptly
removed.
2.5 Return and Transport
When returning empty cylinders leave some positive pressure in the cylinder. Transport using a wheeled
cylinder cart or cylinder dolly with the capped cylinder strapped securely to the cart/dolly. When
returning empty cylinders to the loading dock, always chain the cylinders in the marked locations.
Ensure the return tag clearly lists the lab specific contact.
4
2.6 Connections
Threads on cylinder-valve outlet connections have been standardized by the Compressed Gas
Association and are not the same on all cylinders.
This prevents accidental mixing of incompatible gases from an interchange of connections. Left hand
fittings for fuel gases have a cut mark through the nut.
Never lubricate, modify, force, or tamper with cylinder valves. Especially do not put oil or grease on the
high-pressure side of a cylinder containing oxygen, chlorine, or another oxidizing agent. An auto ignition
or explosion could result. Unlike larger cylinders, lecture bottles all have identical valve threads,
irrespective of the gas contained within.
5
Regulators are gas-specific which limits interchange and adds safety. Special installation processes, not
mentioned here, are used for toxic or high purity gases. Always make sure that the regulator and valve
fittings are compatible.
To select the appropriate regulator:
1. Determine the gas pressure needed
2. Determine the maximum pressure the system might require
3. Select a delivery pressure range so the required pressures are in the 25%-90% range of the
regulator delivery pressure
4. Check with the gas supplier about compatible connections and regulators
Check cylinder outlet and regulator inlet connections for debris or contamination before connecting.
Some gases, such as carbon dioxide, require a gasket. Insure that a required gasket is in place before
assembling the regulator onto the cylinder.
Tighten connecting nut with a smooth jaw wrench. Back out the adjusting knob or key on the regulator.
Open the cylinder valve just enough to indicate pressure on the regulator gauge (no more than one full
turn). Check connections checked with a soap solution for leaks. Never use oil or grease on the regulator
of a cylinder valve.
All compressed gas regulators should, at a minimum, be checked for external leakage and internal
leakage (creep or crawl) regularly. Regulators should be removed from service at least every five years
(more frequent in some cases) and returned to the manufacturer, or a competent agent to be inspected
and/or refurbished as necessary. Regulators should also be tagged or labeled to identify the last date of
inspection. Users should consult the manufacturer for specific procedures on how to check for external
6
and internal leakage as well as the recommended frequency of the tests. Regulators are continuously
exposed to high stresses due to cylinder pressures. In addition to that, the materials of construction are
attacked internally by both mildly and severely corrosive gases. External corrosive environments can
cause gauges and springs to corrode. Argon, helium and nitrogen regulators (CGA 580) will, under a
given set of conditions, have a longer service life than regulators used for hydrogen chloride and
hydrogen sulfide (CGA 330) simply because the gas service is more severe (corrosive).
The most common type of regulator failure is the internal leak, sometimes called creep or crawl. This
can occur when the seat becomes damaged or displaced due to a foreign particle such as a metal chip or
other material. When the seat cannot close completely, delivery pressure will not be maintained and
regulator pressure cannot reach a state of equilibrium. Downstream or delivery pressure will continue to
climb until the safety relief mechanism on the regulator is activated (usually a relief valve or a
diaphragm burst hole). Checking for this type of failure is relatively easy if the device has a gauge that
reads regulated pressure. The gauge pressure will start to rise above the set point and continue upward.
This creates a potentially hazardous condition where any downstream equipment would be subjected to
pressures beyond the rated limit. Regulators should be visually checked for this type of failure. Excessive
flexing of metal regulator diaphragms can cause a radial crack, which allows gas to escape to the
atmosphere through the vent hole in the bonnet.
3.0 Piping and Manifolds
Piping, tubing, valves and fittings conveying hazardous materials shall be designed and installed in
accordance with ASME B31 (Reference IFC 2012 – 5003.2.2). The system’s weakest component
determines the overall pressure limit.
4.0 Toxic Gases
Flow-restricting orifices are recommended on cylinders of toxic gases. All portable tanks and cylinders
must be marked to indicate the orifice (inches) on the certification tags and the vessel themselves.
Toxic-gas cylinders shall be stored in continuously mechanically ventilated enclosures with an
extinguishing system (IFC 2012 – Section 6004.1.2). If the net toxic gas content exceeds one pound per
cylinder no more than three cylinders of toxic gas are allowed per enclosure (gas cabinet). Any new
laboratory construction shall require vented gas cabinets for storage of highly toxic gases. Gas cylinder
cabinets for toxic gases must have a fire extinguishing system (IFC 2012 – 6004.1.2). Purchase of diluted
toxic gas, if feasible below a concentration known to be dangerous, will serve to reduce exposure risk.
Waste toxic gases shall be treated by absorption, wet or dry scrubbing, combustion, or condensation via
refrigeration, before being vented to chemical fume hoods or other local exhaust arrangements. The
safe venting of pressure-relief devices should be considered. (IFC 2012 – Section 6003.1.3 Treatment
Systems)
If the physiological warning properties for the toxic or highly toxic gas(es) are above the PEL an
emergency alarm system is required (IFC 2012 – 6004.2.2.10); consult with the UW Environmental
Health and Safety Office regarding this determination. Consult the emergency plan for the given lab area
to determine the action expected during a leak situation. The Office of Environmental and Occupational
Health (EOH) and Occupational Medicine shall be contacted for information on selection, fit testing, and
training if respirators or Self Contained Breathing Apparatus (SCBA) have been provided. No one may
7
use respirators on the job without prior medical approval, fit testing, and training. Any required leak
detection system shall send all alarm signals to University Police as part of a mutually agreed upon
emergency response plan.
5.0 Flammable Gases
Purchase of diluted flammable gas, if feasible below the explosive range, will serve to reduce the
explosion risk.
Where gases or liquids having a NFPA 704 hazard ranking of Flammability Class 4 are carried in
pressurized piping above 15 pounds per square inch gauge (psig), an approved means of leak detection
and emergency shutoff or excess flow control shall be provided. (Reference: IFC 2012 - 5003.2.2.1).
Any required continuous atmosphere detection system shall send all alarm signals to University Police as
part of a mutually agreed upon emergency response plan.
Oxygen/fuel gas system for welding, brazing or glass blowing shall follow OSHA CFR 1910.253. This
requires the use of Grade T welding hose color coded red/green for all oxygas welding applications and
the installation of flash arresters on hydrogen and acetylene cylinders.
Flammable gas cylinders must be stored 20ft away from oxidizers and oxygen gas cylinders or separated
by a fire rated wall.
5.1 Acetylene
The in house transfer, handling, storage, and utilization of acetylene in cylinders shall be in accordance
with Compressed Gas Association Pamphlet G-1-2015. Acetylene cylinders have a porous filler material
filled with acetone and dissolved acetylene. The cylinder must only be used in the upright position. If a
cylinder has been handled in a non-upright position, do not use it until it has sat upright for at least 30
minutes.
Some tubing materials, such as copper and lead solder, form explosive acetylides.
Never exceed the delivery pressure limit of 15psig indicated by the warning red line of an acetylene
pressure gauge. The use of an excess flow control valve is not recommended. Install a flash
arrestor downstream from the regulator and check valves wherever backflow needs to be prevented.
5.2 Hydrogen
Individual hydrogen gas cylinders should contain less than 400scf. Larger hydrogen gas vessels may
require a laboratory designed with explosion control and other safety measures. The installation of a
flash arrestor is required and the installation of an excess flow control valve is recommended. When
piping hydrogen, use stainless steel tubing to prevent the buildup and potential spark from static
electricity.
8
6.0 Cryogenic Liquids and Liquefied Gases
The hazards of cryogenic liquids include fire or explosion, pressure buildup, embrittlement of structural
materials, asphyxiation, and destruction of living tissue on contact. Liquid helium, argon or nitrogen may
displace air and create an atmosphere without sufficient oxygen. Portable cylinders of cryogens must
only be stored in well ventilated areas. Storage of cryogenic liquids (i.e. liquid nitrogen) or liquefied
gases (i.e. carbon dioxide) in cold rooms or other rooms without external ventilation is prohibited. Fire
or explosion may occur when the liquid form of flammable gases, such as hydrogen, is used without
proper management of the gaseous phase. Liquid oxygen may produce an enriched oxygen atmosphere,
which increases the flammability of ordinary combustible materials. Enriched oxygen levels may also
cause some nonflammable materials, such as carbon steel, to burn readily.
Contact with cryogenic liquids generally causes tissue freezing and frostbite. Even brief contacts may be
intense and painful. Prolonged contact may result in blood clots. Appropriate protective clothing, gloves,
and eye protection — preferably a face shield — shall be worn when cryogenic liquids are handled.
Choice of personal protective equipment (PPE) should be carefully evaluated, as gloves not designed for
use with cryogenic liquids can saturate and cause more extensive cold damage to the skin.
6.1 Dewars and Transfer Equipment
Use a phase separator or special filling funnel to prevent splashing and spilling when transferring liquid
nitrogen into or from a dewar. The top of the funnel should be partly covered to reduce splashing. Use
only small, easily handled dewars for pouring liquid. For the larger, heavier containers, use a cryogenic
liquid withdrawal device to transfer liquid from one container to another. Be sure to follow instructions
supplied with the withdrawal device. The receiving vessel must be raised so the delivery tube is
immediately above the mouth of the vessel (i.e., the cryogenic liquid should never be allowed to fall
through air to reach the receiving vessel). When a warm tube is inserted into liquid nitrogen, liquid will
spout from the bottom of the tube due to gasification and rapid expansion of liquid inside the tube.
Wooden or solid metal dipsticks are recommended; avoid using plastics that may become very brittle at
cryogenic temperatures.
6.2 Monitoring for Oxygen Deficiency
Indoor areas where bulk inert gas systems are newly installed shall be continuously monitored with an
atmosphere monitoring system. The system shall provide an audible and visual alarm (red light) when
the oxygen level drops to 19.5%. The audible and visual alarm shall be located inside the area and
immediately outside of all entrances to the indoor area. Any required atmosphere monitoring system
shall send all alarm signals to University Police.
9
6.3 Minimum Ventilation Rate
Natural or mechanical ventilation shall be provided when bulk inert gas systems are installed in
buildings, rooms, or any indoor confined area. Ventilation shall be provided throughout the space at the
rate of not less than 1.0 cu. foot per minute per square foot of floor area determined by the area
enclosed. (Reference: IFC2012 – 5004.3.1)
7.0 Quantity Limits on Campus
The scope of this section generally applies to areas constructed for business occupancy (B). For quantity
limits for all other occupancies, especially High Hazard Group (H) contact the UW Environmental Health
and Safety Office.
Building users on the UW Campus generally have to comply with the limits according to the
International Building Code (IBC) and the International Fire Code (IFC). Groups of research laboratories
within a building may be subdivided into fire control areas. For more specific information regarding
laboratory design and fire control specification contact the UW Environmental Health and Safety Office.
The total quantities of flammable or combustible liquids allowed in a fire control area (laboratory or
suite of laboratories) are limited by the location in the building and the construction specifications.
7.1 IFC Maximum Allowable Quantities in Storage per Building - Sprinklered Fire Control Area
Hazardous Material Below grade/basement 1stfloor/grade 2nd floor 3rd floor 4th to 6th floor
Cryogenic flammable 67 gal 90 gal 67 gal 45 gal 11 gal
Cryogenic Oxidizing 67 gal 90 gal 67 gal 45 gal 11 gal
Flammable gas (gaseous) 1,500 cu. ft.a 2,000 cu. ft.a 1,500 cu. ft.a 1,000 cu. ft.a 250 cu. ft.a
Oxidizing gas (gaseous) 2,250 cu. ft.a 3,000 cu. ft.a 2,250 cu. ft.a 1,500 cu. ft.a 375 cu. ft.a
Oxidizing gas (liquefied) 112 poundsa 150 poundsa 112 poundsa 75 poundsa 18.7 poundsa
Pyrophoric (needs EHS 37 cu. ft.a 50 cu. ft.a 37 cu. ft.a 25 cu. ft.a 6 cu. ft.a
approval)
Corrosive 1,215 cu. ft.a 1,620 cu. ft.a 1,215 cu. ft.a 810 cu. ft.a 202 cu. ft.a
Highly toxic (needs EHS 15 cu. ft.b 20 cu. ft.b 15 cu. ft.b 10 cu. ft.b 2 cu. ft.b
approval)
Toxic 1,215 cu. ft.a 1,620 cu. ft.a 1,215 cu. ft.a 810 cu. ft.a 202 cu. ft.a
a. Quantities shall be increased 100% when stored in approved cabinets, gas cabinets, or exhausted enclosures as specified by
the International Fire Code. Maximum quantities shall be increased 100% for buildings equipped throughout with an automatic
sprinkler system. Where both are present, the increase shall be applied accumulatively.
b. Allowed only when stored in approved exhausted gas cabinets or exhausted enclosures.
10