MDFu 74
MDFu 74
MDF-U74V
MDF-U74VC
SANYO Electric Co., Ltd.
Biomedical Business Division
SM9910225
Effective models
䊶Specifications ------------------------------------------------ 1
- Structural specifications
- Control specifications
- Performance specifications
䊶Dimensions ------------------------------------------------ 4
䊶Cascade refrigeration system --------------------------------------------- 5
䊶Cooling unit parts ------------------------------------------------ 6
䍃Components on PCB ------------------------------------------------ 7
䊶Connection on PCB ------------------------------------------------ 8
䊶Electric parts ------------------------------------------------ 9
ŶStructural specifications
Item MDF-U74V MDF-U74VC
Name Ultra-low Temperature Freezer
External dimensions W1010 × D870 × H2010 (mm)
Internal dimensions W870 × D600 × H1400 (mm)
Effective capacity 728 L
Outer door 1door, painted steel
Inner door 2doors, ABS resin panel with stainless frame
Insulation Vacuum insulation panel + rigid polyurethane foamed-in place
Exterior Painted steel
Interior Painted steel
3shelves, stainless steel
Shelf
Inner dimensions; W848 x D533 (mm) Load; 50kg/shelf
Outer door latch 1pc
Outer door lock 1pc
Caster 4pcs㧔leveling leg㧦2pcs㧕
Monitoring hole 3places(Rear upper, back right bottom, back left bottom)
Inner diameter; ij17mm
Refrigeration circuit Cascade refrigeration system
Compressor High stage side; Hermetic type, Output; 1100W
Low stage side; Hermetic type, Output; 1100W
Evaporator High stage side; Cascade condenser
Low stage side; Tube on sheet㧔Sharing with interior㧕
Condenser High stage side; Fin and Tube type
Low stage side; Cascade condenser
Refrigerant High stage side; R-407D
Low stage side; R-508
Refrigerant oil Ze-NIUSL22SA
Power supply Local voltage
Weight 346 Kg 351 Kg
Accessories 1 set of key, 1 scraper
Optional component Automatic temperature recorder(MTR-G85)
Back-up system (CVK-UB2, CVK-UB2(I)); LCO2, (CVK-UBN2);LN2
Inventory rack (IR-220, IR-224U㧕
Independent inner door (MDF-7ID)
-1-
ŶControl specifications
Item MDF-U74V MDF-U74VC
Temp. controller Micro-processor control system
Temperature setting range: -50͠㨪-90͠ (Unit:1͠)
Non-volatile memory
Thermal sensor Pt.1000ȍ
Temperature display LED display (Unit:1͠)
When chamber temp. reaches to set temp.+5͠~+20͠ (Factory default; +10͠),
high temp. alarm emits.
ALARM lamp blinks, audible alarm sounds intermittently after 15min. passes.
High temp. alarm
Remote alarm contact; Normal Open, Normal Close
Contact activates in reverse after 15min. passes.
Allowable contact capacity; Max. 30VDC, 2A
When chamber temp. reaches to set temp.-5͠~-20͠ (Factory default; -10͠),
low temp. alarm emits.
ALARM lamp blinks, audible alarm sounds intermittently after 15min. passes.
Low temp. alarm
Remote alarm contact; Normal Open, Normal Close
Contact activates in reverse after 15min. passes.
Allowable contact capacity; Max. 30VDC, 2A
Door alarm When a door leaves open, DOOR lamp illuminates after 2min. passes.
When a condenser filter is clogged, FILTER lamp illuminates and audible alarm
Filter alarm
sounds intermittently.
When a power is interrupted, ALARM lamp blinks, audible alarm sounds
Power failure alarm
intermittently and remote alarm contact outputs.
Remote alarm terminal 3P; Max. DC30Vޔ2A N.C.-COM, N.O.-COM
Remote alarm When temp. alarm or power failure alarm emits, or when sensor is failed, remote
alarm contact activates in reverse.
When battery accumulation time reaches to approx. 3years, BATTERY lamp
Notice of battery life
illuminates.
When fan motor accumulation time reaches to approx. 6years, BATTERY lamp
Notice of fan motor life
blinks.
Status-1: When a temperature in AT sensor is lower than 0͠ or higher than
+35͠, it diagnoses that the ambient temperature is abnormal.
Status-2: When a power supply voltage is lower than rated voltage -15%, it
STATUS function
diagnoses that the power supply voltage is abnormal.
Status-3: When a running rate of compressor L is higher than 95%, it diagnoses
that unit is in overloaded operation.
Lamps: ALARM, BATTERY, STATUS, DOOR, FILTER
Lamps and keys on Buzzer stop key: BUZZER Alarm test key: ALARM TEST
Control panel Status key: STATUS Set key: SET
Digit shift key: Numerical value shift key:
Press key for 5 seconds to display Key Lock mode.
Key Lock
L0: Key Lock is OFF L1: Key Lock is ON
When a temperature in cascade sensor is lower than -34͠, compressor L turns
on. When a temperature in cascade sensor is higher than -12͠, compressor L
Compressor protection turns off. When a temperature in filter sensor is higher than +60͠, compressor H
turns off.
Overload relay
If several units are installed on a same site, it is possible to set delay time not to
start them simultaneously after they returns from power failure.
Start delay
Setting range: 3~15 minutes (Unit: 1 minute)
It is impossible to set delay time individually by units.
-2-
ŶPerformance specifications
Model MDF-U74V
Cooling performance -86 C at the center of the chamber (ambient temperature; 30oC, no load)*
o
Model MDF-U74VC
o
Cooling performance -86 C at the center of the chamber
(ambient temperature; 30oC, no load)*
Temperature control range -50oC to -86oC (ambient temperature; 30oC, no load)
Power source 220 V, 60 Hz
Rated power consumption 1230 W
Noise level 51 dB [A] (background noise; 20 dB)
Maximum pressure 2680 kPa
-3-
Dimensions
-4-
<Overlook view>
Power cord
H side Compressor H
L side
Pre-condenser
Dryer H
Frame pipe
Cap. tube H
Pre-condenser
-5-
Cascade condenser
Dryer L
Evaporator
Header
EX tank
<MDF-U74V/U74VC>
Specifications
Item
H side L side
Compressor
220V, 50Hz Type:KS370J1NS-4A Code:7FB-0-M101-001-04
230/240V, 50Hz Type:KS370J1NS-4A1 Code:7FB-0-M101-001-05
220V,60Hz Type:KS370J1NS-7A Code:7FB-0-M101-001-06
*220V, 50Hz Type:KS370J1NS-4AD Code:7FB-0-M101-001-07
Refrigerant oil Ze-NIUSL22SA Charged q’ty: 850cc
Cooling system Forced air cooling㧔partially㧕and oil cooler
Condenser
Type Fin and tube Cascade condenser
Condenser 12 columns x 4 lines P6.35mm
Coil pipe Ǿ6.35mm
Fin 88pcs.
Pre-condenser W 350mm ʊ
Frame pipe Ǿ6.35mm ʊ
Evaporator Cascade condenser Tube on sheet
Type Shell and tube Ǿ80mm Ǿ9.52mm
Capillary tube Ex. capillary
Resistance
78 PSI 0.37Mpa 34 PSI/G
PSIkg/cm2
Length 3000mm 3000mm 500mm
Outer diameter Ǿ2.4mm Ǿ2.0mm Ǿ2.4mm
Inner diameter Ǿ1.2mm Ǿ0.9mm Ǿ1.2mm
Refrigerant R-407D Charged q’ty: 606g R-508 Charged q’ty: 320g
n-Pentane n-Pentane
Oil additive
Charged-q’ty: 39g Charged q’ty: 46g
Dryer 4A-XH-9 Charged q’ty: 18g 4A-XH-6 Charged q’ty: 58g
Condensing fan Ǿ230 mm, 4 blades Material: ABS ʊ
Condensing fan
motor Type:SV4-11AA5P ʊ
Oil separator ʊ Type:SPK-0S02S2
S R
-6-
Components on PCB
CN11 CN8
䋤1-䋤3: To Temp. sensor 䋤1-䋤2: To Battery
䋤3-䋤4: To Transformer
CN1
䋤1-䋤3: To Switching
power supply
CN2
To MTR-480䋨Option䋩
CN7 To MTR-L03䋨Option䋩
䋤1-䋤2: To Door SW
䋤5-䋤6: To AT sensor
䋤7-䋤8: To Filter sensor
䋤 9- 䋤 10: To Cascade
sensor
CN6 CN9
To LCD PCB 䋤1-䋤2: To H comp. relay
CN4
䋤 1- 䋤 2: To Temp. control
relay
䋤3-䋤4: To Heater relay
CN5
䋤1-䋤5: To Switch PCB
䋤6-䋤7: To Buzzer PCB
CN3
To Remote alarm terminal
䋤1:COM.
䋤2:N.O.
䋤3:N.C.
-7-
Connections on PCB
The following shows the connections of connectors on the Temp. controller PCB.
-8-
Electrical Parts
-9-
Specifications of sensor
The following shows the temperature in thermal sensor (502AT-1) and its resistance value.
Temp. Resistance Temp. Resistance Temp. Resistance Temp. Resistance
(C) Value (kǡ) (C) Value (kǡ) (C) Value (kǡ) (C) Value (kǡ)
㧙50 154.5 㧙36 71.80 㧙22 35.65 0 13.29
㧙49 145.9 㧙35 68.15 㧙21 33.99 5 10.80
㧙48 137.8 㧙34 64.71 㧙20 32.43 10 8.84
㧙47 130.2 㧙33 61.48 㧙19 30.92 15 7.20
㧙46 123.1 㧙32 58.43 㧙18 29.50 20 6.01
㧙45 116.5 㧙31 55.55 㧙17 28.14 25 5.00
㧙44 110.2 㧙30 52.84 㧙16 26.87 30 4.17
㧙43 104.4 㧙29 50.23 㧙15 25.65 35 3.50
㧙42 98.87 㧙28 47.77 㧙14 24.51 40 2.96
㧙41 93.70 㧙27 45.45 㧙13 23.42 45 2.51
㧙40 88.85 㧙26 43.26 㧙12 22.39 50 2.13
㧙39 84.18 㧙25 41.19 㧙11 21.41 55 1.82
㧙38 79.80 㧙24 39.24 㧙10 20.48 60 1.56
㧙37 75.67 㧙23 37.39 㧙5 16.43 65 1.35
The following shows the temperature in thermal sensor (PT1000ȍ) and its resistance value.
Temp. Resistance Temp. Resistance Temp. Resistance
(C) Value (ǡ) (C) Value (ǡ) (C) Value (ǡ)
㧙140 450.83 㧙70 729.99 0 1000.0
㧙130 491.47 㧙60 769.02 10 1038.0
㧙120 531.83 㧙50 807.87 20 1076.0
㧙110 571.92 㧙40 846.58 30 1113.8
㧙100 611.76 㧙30 885.13 40 1151.4
㧙90 651.38 㧙20 923.55 50 1189.0
㧙80 690.78 㧙10 961.84 60 1226.4
- 10 -
Wiring diagram
- 11 -
- 12 -
㪚㫀㫉㪺㫌㫀㫋㩷㪻㫀㪸㪾㫉㪸㫄㩷
- 13 -
㩷 Control specifications
1. Key and Switch
BUZZER : In alarm condition, audible alarm silences when this key is pressed.
Remote alarm activates and message is not eliminated.
When a power failure is occurred (battery back-up), press this key to show a
chamber temperature for 5 seconds.
ALARM TEST : When this key is pressed, unit steps into Alarm Test mode with ALARM lamp
blinks, intermittent buzzer beeps, digital LED goes off and remote alarm
activates.
After an elapse time is about 90seconds, unit returns to normal condition. (Auto
Return function)
If Alarm Test is performed when a battery switch is in off position, “E09” blinks
on the display.
nd
SET : Press this key once to activate setting mode with 2 digit in LED blinks.
Press this key again to store a value to be changed.
STATUS : If this key is pressed during STATUS lamp illuminates, status code (‘--- 1’, ‘---2’
and ‘---3’) is displayed.
A blinking digit can change among 1st digit ~ 3rd digit with every time press this
(Digit shift key) key.
If this key is pressed for 5 seconds when a chamber temperature is displayed,
Key Lock activates with “L_0” is displayed.
A blinking digit can increase one by one every time press this key.
(Numerical value If this key is pressed for 5 seconds when a chamber temperature is displayed.
shift key) Function mode activates with “F00” is displayed.
2. Temperature control
Setting range : -50㷄~-90㷄
Display range : -180~50
Setting procedure : Press SET key and set the required value with key and key.
Press SET key to store the value to be changed.
Unacceptable setting If a value which is out of setting range is input and SET key is pressed, it is
range : noticed the value cannot be set with audible alarm sounds for 1second.
3. Temperature alarm
Setting range : High temperature alarm … +5㷄~+20㷄 (Factory default: 10㷄)
Low temperature alarm …. -5㷄~-20㷄 (Factory default: -10㷄)
Setting procedure : Keep pressing key over 5 seconds to step to function mode (F00). Input
“F01” for high temperature alarm or “F02” for low temperature alarm.
st
Press SET key to set the value to be changed with the 1 digit blinks.
Press SET key again to store the value in the non-volatile memory.
Unacceptable setting If a value which is out of setting range is input and SET key is pressed, it is
range : noticed the value cannot be set with audible alarm sounds for 1second.
- 14 -
B) Function mode
Setting range : 00~50
Display range : 00~59
00, 16 and 33~43, 44~49, 51~59 are unused.
Setting procedure : In chamber temperature display, press key for 5 seconds to step㩷 to
function mode (F00 is displayed).
Change the blinking 1st digit to desired function code with key and key.
Press SET key to be function code available.
Unacceptable setting If a value which is out of setting range is input and SET key is pressed, it is
range : noticed the value cannot be set with audible alarm sounds.
5. Error codes
E01: Temp. sensor is open circuited
E02: Temp. sensor is short circuited
E03: Cascade sensor is open circuited
E04: Cascade sensor is short circuited
E05: Filter sensor is open circuited
E06: Filter sensor is short circuited
E07: AT sensor is open circuited
E08: AT sensor is short circuited
E09: Battery switch is in off position
E10: Compressor temperature is abnormal
- 15 -
(4) AT sensor
Open circuit (E07): When a temperature in AT sensor is lower than -60㷄, E07 and chamber
temperature are displayed alternately, audible alarm sounds intermittently
and remote alarm contact activates.
Press BUZZER key to silence audible alarm.
Short circuit(E08): When a temperature in AT sensor is higher than 60㷄, E08 and chamber
temperature are displayed alternately, audible alarm sounds intermittently
and remote alarm contact activates.
Press BUZZER key to silence audible alarm.
(5) Battery SW is in If you press ALARM TEST key when battery switch is in off position or
off position (E09): battery is unconnected, E09 is displayed.
(6)Compressor When a temperature in filter sensor is higher than 60㷄, E10 and chamber
abnormal temperature are displayed alternately and high side compressor is forced to
temperature (E10): turn off to notify compressor temperature is abnormal or fan motor is locked.
Press BUZZER key to silence audible alarm.
When the temperature in filter sensor subtracts ambient temperature is equal
or lower than 10㷄, compressor turns on.
6. Warning function
Door alarm : When an outer door leaves open, DP54 (red lamp) illuminates. Audible alarm
sounds intermittently after 1~15 minutes (Factory default: 2 minutes) elapse.
Audible alarm does not activate simultaneously with remote alarm.
Press BUZZER key to silence audible alarm.(No Ring Back)
High temp. alarm : When chamber temperature is equal or higher than set temperature + high
temp. alarm set temperature +1, ALARM lamp and LED blink, audible alarm
sounds intermittently after 10 minutes delay, and remote alarm activates.
When chamber temperature is equal or lower than set temperature, ALARM
lamp and LED go off, audible alarm silences, and remote alarm turns off.
When BUZZER key is pressed, audible alarm silences, but remote alarm
output does not turn off.
Low temp. alarm : When chamber temperature is equal or lower than set temperature - low
temp. alarm set temperature -1, ALARM lamp and LED blink, audible alarm
sounds intermittently after 10 minutes delay, and remote alarm activates.
When chamber temperature is equal or higher than set temperature, ALARM
lamp and LED go off, audible alarm silences, and remote alarm turns off.
When BUZZER key is pressed, audible alarm silences, but remote alarm
output does not turn off.
- 16 -
Power failure alarm : If a power interrupts for 3 seconds when battery switch is in on-position,
ALARM lamp blinks, audible alarm sounds intermittently and remote alarm
activates.
When a power returns within 3 seconds after the power interrupts,
microprocessor resets and unit will start operation in default settings. At the
time remote alarm will be active.
Press BUZZER key to silence audible alarm, but remote alarm does not turn
off.
Remote alarm should activate until chamber temperature is stable after the
power returns from power failure.
When a power interrupts, press BUZZER key to see chamber temperature
for 5 seconds.
7. Other function
Auto Return : If there are not any key operations for 90 seconds in setting mode, Key Lock
mode and Function mode, unit automatically returns to chamber temperature
display.
Ring Back : To prevent someone except for an operator pressing BUZZER key when a
unit is in alarming condition, audible alarm sounds again after predetermined
setting time elapses even if BUZZER key is pressed to silence audible alarm.
Capillary heater F18: When you input ‘000’ in F18, compressor turns off and capillary heater
forcibly ON/OFF : is forced to turn on.
When you input ‘000’ during capillary heater turns on, it comes to end
to turn capillary heater on.
When you input ‘001’, capillary heater never turns on, but compressor
turns off in every 18 hours.
ROM version : F30: ROM version is displayed (Ex. Ver. 1.00 㸢㩷 Displayed as “1.00”)
- 17 -
8. STATUS
(1) STATUS lamp illuminates in the following conditions;
STATUS 1: When an ambient temperature is higher than 35.0㷄, or lower than 0㷄, ‘----1’
is displayed.
STATUS 2: When a power source voltage becomes low (Lower than 2.01VDC between
TP7 and TP3), ‘--- 2’ is displayed.
STATUS 3: When running rate is higher than 95%, ‘---3’ is displayed.
(Ex.1) Every STATUS codes are displayed in the following order when all of three
STATUS occur simultaneously.
(Ex. 2) Both STATUS codes are displayed in the following order when STATUS 1 and 2
occur simultaneously.
(Note)
When all of STATUS codes are eliminated or 90 seconds elapse, unit returns to chamber
temperature display automatically.
9. Running rate
ON time (Min.) = The time until P3.1 in IC5 first reaches from LOW to HIGH
OFF time (Min.) = The time until P3.1 in IC5 reaches from HIGH to LOW
- 18 -
Chamber
Compressor ON (1st)
temp. Start to calculate Start to diagnose running rate
Start to cycle run
running rate
㧿et temp.
Compressor OFF (1st) 8 hours stabilized Diagnosis is done with every 8 hour
2 hours
Reach to set temp. (Cycle run) until chamber temp is higher than
Set temp + 4.0
- 19 -
Input service code ‘384’ in F06 prior to use function codes which are marked with *.
To cancel service code, input “000” in F06 or turn the power off.
Setting In chamber temperature display, press key for 5 seconds to display “F00”.
procedure: Input Function code by pressing key and key.
Press SET key to be function mode available.
F05: <Purpose> Compressor turns on with delay when a power is supplied or a power
returns from a power failure.
<Operation> Input “F05” and press SET key to display ‘003’ (Factory default).
Setting range is ‘003’~’015’. (Unit: Minute)
Change a value by pressing key and key.
Press SET key to store the value and to return to chamber temperature
display.
- 20 -
<Cancel> Input F06 again and press SET key to display ‘384’.
Change to ‘000’ by pressing key and key.
Press SET key to store the value and to return to chamber temperature
display.
Turn a power off then on to change a value to ‘000’, but it is not stored
in non-volatile memory.
Note) Service code ‘384’ is stored in non-volatile memory during battery
back-up.
F07: <Purpose> To match a temperature in temp. sensor with 1/2H air temperature
<Operation> Input service code in F06 prior to use this mode.
Input F07 and press SET key to display ‘00.0’ (Factory default).
Setting range is ‘-4.9’~’04.9’.
Change a value by pressing key and key.
Press SET key to store the value and to return to chamber temperature
display.
- 21 -
<Initial values in non-volatile memory>
Zero Adjustment value for temperature sensor : 0㷄
Zero Adjustment value for cascade sensor : 0㷄
Capillary heater ON time : 8 minutes
Auto Return : 30 minutes
Chamber set temperature : -80㷄
Door alarm delay time : 2 minutes
High temp. alarm
: +10㷄
Low temp. alarm
Compressor delay time : -10㷄
Communication ID : 3 minutes
Communication mode : 000
Key Lock : 000
Linkage between remote alarm and buzzer : OFF
Buzzer tone for filter alarm : OFF
Diagnosed value for STATUS 3 : OFF
: Fixed
- 22 -
F22: <Purpose> Setting of serial communication mode
<Operation> input F22 and press SET key to display ‘000’ (Factory default)
Change a value by pressing key and key.
Press SET key to store the value and return to chamber temperature
display.
Control mode (the 3rd digit)
0: Local (initial)
1: Remote
Baud rate (the 2nd digit)
0: 2400bps (initial)
1: 4800bps
2: 9600bps
- 23 -
F28: <Purpose> Display of delay time to start measuring running rate
(2hrs timer; 000~120 min)
<Operation> Service code should be input in F06 prior to use this mode.
Input F28 and press SET key to display alternately F28 with ‘xxx’
(present count value for delay time to start measuring running rate).
Press SET key to return to chamber temperature display.
When a delay time expires (a value reaches to ‘120’), unit will start
measuring running rate.
- 24 -
11. Compressor control (differential) value
Compressor H:
Turns on when a chamber temperature is set temperature -0.4㷄.
Compressor L:
1) Turns on when compressor H turns on and a temperature in cascade sensor is -34㷄.
2) Turns on after 3 minutes elapse since compressor H turned on
Compressor H, L:
Turns off when a chamber temperature is set temperature -1.4㷄
Interval:
1 minute *
* ‘Interval’ means that a period one of compressor either H or L turns on after another one was
turned on.
Note) Compressor L does not turn on until 1 minute elapses after compressor H turned on,
if a temperature in cascade sensor is lower than -34㷄.
Set temp.
Chamber Set temp. –0.4㷄
Temp.
Set temp. –1.4㷄
Cascade Cascade
Temp. =-34㷄 Temp. =-34㷄
Cascade Cascade Temp.
Temp. =-34㷄
ON
H comp.. OFF
ON
L comp.. OFF
Compressor H protection:
Compressor H turns off to prevent it from being warmed up that is caused by fan motor locked.
- 25 -
Power failure alarm delay time (3 seconds fixed)
When a power is failed, power failure alarm is triggered with 3 seconds of delay.
Delay time cannot be changed.
CN3
1 – 2 (N.O.) 1 – 3 (N.C.)
Normal Open Close
In alarm Close Open
- 26 -
Settings after unit returns from power failure:
Alarms: OFF
Compressors: OFF
Remote alarm: ON
Door alarm delay time: 2 minutes
Timers: Reset
2H timer, 8H timer: 0 (Reset)
Counting of compressor L OFF period: Reset
Setting data: Read by non-volatile memory
Examples of display:
- 27 -
Buzzer tone:
Alarms (except door alarm) Intermittent tone
Key operation Click tone
Set value memory Click tone
Out of settable range Continuous tone (1 second)
Door alarm Intermittent tone with shorter than other alarms
- 28 -
Parts layout
Control panel
Latch "Close"
Latch "Open"
Handle
Door latch
Fan motor
Boost BOX
(MDF-U74VC only)
Power transformer
(MDF-U74VC only)
Filter sensor
- 29 -
<Grille> Filter
AT sensor
Power switch
<Electrical BOX>
12P terminal
Remote alarm
Main PCB
Switching power supply
Running capacitor H
Starting capacitor H
- 30 -
Test data
*Following data are the reference only.
Pull-down data䇭䇭䇭䇭AT:30㷄
40
20
0
Chamber temp[㷄]
-20
-40
-60
-80
-100
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Time[Minute]
Pull-up data䇭䇭䇭䇭AT:30㷄
0
-10
-20
Chamber temp[㷄]
-30
-40
-50
-60
-70
-80
-90
-100
0 1 2 3 4 5 6 7 8 9
Time[Minute]
- 31 -
Pull-down pressure data䇭AT35㷄, 50Hz
2.5
H䇭Ps
H䇭Pd
2 L䇭Ps
L䇭Pd
Pressure䋨MPa䋩
1.5
0.5
0
0 2 4 6 8 10 12 14 16
Time䋨Hour䋩
3
H䇭Ps
H䇭Pd
2.5 L䇭Ps
L䇭Pd
Pressure䋨MPa䋩
1.5
0.5
0
0 2 4 6 8 10 12 14
Time䋨Hour䋩
- 32 -
Current-Input data AT35㷄 230V,50Hz
10 2
A
9 1.8
kW
8 1.6
7 1.4
6 1.2
Current䋨A䋩
Input䋨kW䋩
5 1
4 0.8
3 0.6
2 0.4
1 0.2
0 0
0 2 4 6 8 10 12 14
Time䋨䌨our䋩
10 1.8
A
9 1.6
kW
8 1.4
7
1.2
Current䋨䌁䋩
6
Input䋨䌫W䋩
1
5
0.8
4
0.6
3
2 0.4
1 0.2
0 0
0 2 4 6 8 10 12 14
Time䋨Hr䋩
- 33 -
H/L EVA OUT temperatures 䇭AT30㷄
40
HEVAout
20 LEVAout
1/2hair
0
Temperatture䋨㷄䋩
-20
-40
-60
-80
-100
0 2 4 6 8 10 12 14
Time䋨Hr䋩
- 34 -
Temperature uniformity - 17 points measured
* Measured points
Back
㽳 Top of interior 㽵
㽶 Upper area measuring points
Front
70mm᳂ 㽲~㽵: 70mm(H) from top of interior
27mm(D), 42.6mm(W) from each
㽲 㽴 corners on Middle shelf.
42.6mmW
27mmD
㽸 㽻 㽺
㾂 10mmH
D
Lower shelf
Bottom area measuring points
᳑ 㽿
㽽 㽼~㽿: 70mm(H) from bottom of interior
㾀 27mm(D), 42.6mm(W) from each
corner on middle shelf
㽼 Bottom of interior 㽾 70mm᳂
㾀 65mm(H) from bottom of interior
Left Right
- 35 -
䋼Conditions䋾
䋼Distribution data䋾
Temperature of the cycle in each area (SV=-80㷄䇮air temperature)
Unit:㷄
Ambient temperature 20㷄
50Hz 60Hz
Middle of Middle of
Maximum Minimum Differential Maximum Minimum Differential
cycle cycle
㽲 Left front -75.9 -79.1 -77.5 ±1.6 -74.3 -80.5 -77.4 ±3.1
Upper area
㽳 Left back -76.6 -79.9 -78.3 ±1.7 -74.3 -81.9 -78.1 ±3.8
㽴 Right front -75.8 -79.5 -77.7 ±1.9 -74.1 -81.6 -77.9 ±3.8
㽵 Right back -76.6 -80.9 -78.8 ±2.2 -75.1 -82.1 -78.6 ±3.5
㽶 Center -77.2 -81.2 -79.2 ±2.0 -75.2 -83.2 -79.2 ±4.0
㽷 Left front -77.1 -79.6 -78.4 ±1.3 -76.3 -80.3 -78.3 ±2.0
Middle area
㽸 Left back -78.2 -81.2 -79.7 ±1.5 -76.0 -83.3 -79.7 ±3.7
㽹 Right front -76.8 -79.2 -78.0 ±1.2 -75.0 -81.0 -78.0 ±3.0
㽺 Right back -79.5 -82.4 -81.0 ±1.5 -78.8 -83.8 -81.3 ±2.5
㽻 Center -77.2 -79.5 -78.4 ±1.2 -75.3 -81.6 -78.5 ±3.2
㽼 Left front -73.2 -75.5 -74.4 ±1.2 -72.3 -77.3 -74.8 ±2.5
Bottom area
㽽 Left back -74.8 -76.7 -75.8 ±1.0 -73.6 -78.6 -76.1 ±2.5
㽾 Right front -72.6 -75.4 -74.0 ±1.4 -71.7 -76.2 -74.0 ±2.3
㽿 Right back -74.2 -76.4 -75.3 ±1.1 -72.8 -77.7 -75.3 ±2.5
㾀 Center -74.1 -75.9 -75.0 ±0.9 -73.2 -77.8 -75.5 ±2.3
㾁 Center of Upper shelf -79.8 -82.3 -81.1 ±1.3 -77.6 -84.0 -80.8 ±3.2
㾂 Center of Lower shelf -75.0 -77.5 -76.3 ±1.3 -73.9 -78.2 -76.1 ±2.2
Average - - -77.6 - - - -77.6 -
Unit:㷄
Ambient temperature 30㷄
50Hz 60Hz
Middle of Middle of
Maximum Minimum Differential Maximum Minimum Differential
cycle cycle
㽲 Left front -74.8 -78.0 -76.4 ±1.6 -75.1 -78.6 -76.9 ±1.8
Upper area
㽳 Left back -75.4 -78.6 -77.0 ±1.6 -75.6 -79.6 -77.6 ±2.0
㽴 Right front -74.6 -78.3 -76.5 ±1.9 -74.8 -78.9 -76.9 ±2.1
㽵 Right back -75.4 -79.6 -77.5 ±2.1 -75.7 -80.6 -78.2 ±2.5
㽶 Center -75.9 -80.0 -78.0 ±2.1 -75.9 -80.5 -78.2 ±2.3
㽷 Left front -76.4 -79.0 -77.7 ±1.3 -76.3 -79.2 -77.8 ±1.5
Middle area
㽸 Left back -77.5 -80.5 -79.0 ±1.5 -77.3 -81.1 -79.2 ±1.9
㽹 Right front -76.0 -78.8 -77.4 ±1.4 -76.3 -79.6 -78.0 ±1.7
㽺 Right back -78.9 -82.1 -80.5 ±1.6 -79.0 -82.5 -80.8 ±1.8
㽻 Center -76.6 -79.0 -77.8 ±1.2 -76.5 -79.2 -77.9 ±1.4
㽼 Left front -73.5 -75.5 -74.5 ±1.0 -72.6 -75.1 -73.9 ±1.3
Bottom area
㽽 Left back -74.4 -76.5 -75.5 ±1.1 -73.1 -76.4 -74.8 ±1.7
㽾 Right front -72.3 -74.5 -73.4 ±1.1 -73.0 -75.1 -74.1 ±1.1
㽿 Right back -74.1 -76.4 -75.3 ±1.2 -74.3 -76.9 -75.6 ±1.3
㾀 Center -73.9 -76.0 -75.0 ±1.1 -73.8 -76.3 -75.1 ±1.3
㾁 Center of Upper shelf -78.9 -81.6 -80.3 ±1.3 -79.0 -82.0 -80.5 ±1.5
㾂 Center of Lower shelf -74.6 -77.3 -76.0 ±1.4 -74.1 -77.4 -75.8 ±1.7
Average - - -76.9 - - - -77.1 -
- 36 -
䋼Distribution data䋾
Temperature of the cycle in each area (SV=-70㷄䇮air temperature
Unit:㷄
Ambient temperature 20㷄
50Hz 60Hz
Middle of Middle of
Maximum Minimum Differential Maximum Minimum Differential
cycle cycle
㽲 Left front -67.9 -72.9 -70.4 ±2.5 -68.1 -73.3 -70.7 ±2.6
Upper area
㽳 Left back -68.5 -74.4 -71.5 ±3.0 -68.8 -74.9 -71.9 ±3.1
㽴 Right front -68.0 -74.1 -71.1 ±3.1 -68.3 -74.5 -71.4 ±3.1
㽵 Right back -68.9 -76.0 -72.5 ±3.6 -69.3 -76.7 -73.0 ±3.7
㽶 Center -69.4 -76.0 -72.7 ±3.3 -69.7 -76.6 -73.2 ±3.5
㽷 Left front -67.7 -71.1 -69.4 ±1.7 -68.0 -71.1 -69.6 ±1.6
Middle area
㽸 Left back -68.8 -74.5 -71.7 ±2.9 -69.8 -74.4 -72.1 ±2.3
㽹 Right front -67.9 -71.5 -69.7 ±1.8 -68.3 -71.6 -70.0 ±1.7
㽺 Right back -70.9 -75.5 -73.2 ±2.3 -71.0 -75.5 -73.3 ±2.3
㽻 Center -68.6 -72.3 -70.5 ±1.9 -68.8 -72.2 -70.5 ±1.7
㽼 Left front -63.2 -65.2 -64.2 ±1.0 -63.2 -65.0 -64.1 ±0.9
Bottom area
㽽 Left back -63.1 -66.8 -65.0 ±1.9 -64.0 -66.8 -65.4 ±1.4
㽾 Right front -63.3 -65.6 -64.5 ±1.2 -63.3 -65.4 -64.4 ±1.1
㽿 Right back -64.8 -67.2 -66.0 ±1.2 -64.9 -67.2 -66.1 ±1.2
㾀 Center -64.0 -66.2 -65.1 ±1.1 -64.2 -66.0 -65.1 ±0.9
㾁 Center of Upper shelf -71.9 -76.0 -74.0 ±2.1 -72.3 -76.2 -74.3 ±2.0
㾂 Center of Lower shelf -64.9 -67.5 -66.2 ±1.3 -65.0 -67.4 -66.2 ±1.2
Average - - -69.1 - - - -69.4 -
Unit:㷄
Ambient temperature 30㷄
50Hz 60Hz
Middle of Middle of
Maximum Minimum Differential Maximum Minimum Differential
cycle cycle
㽲 Left front -67.8 -72.7 -70.3 ±2.5 -67.7 -73.3 -70.5 ±2.8
Upper area
㽳 Left back -68.6 -73.9 -71.3 ±2.7 -68.7 -74.4 -71.6 ±2.9
㽴 Right front -67.6 -73.7 -70.7 ±3.1 -67.5 -74.1 -70.8 ±3.3
㽵 Right back -68.2 -75.4 -71.8 ±3.6 -68.2 -75.7 -72.0 ±3.8
㽶 Center -69.1 -75.5 -72.3 ±3.2 -69.1 -75.9 -72.5 ±3.4
㽷 Left front -67.5 -71.6 -69.6 ±2.1 -67.6 -71.6 -69.6 ±2.0
Middle area
㽸 Left back -69.6 -74.7 -72.2 ±2.6 -69.8 -75.2 -72.5 ±2.7
㽹 Right front -67.1 -71.1 -69.1 ±2.0 -66.9 -71.3 -69.1 ±2.2
㽺 Right back -70.5 -75.9 -73.2 ±2.7 -70.4 -76.0 -73.2 ±2.8
㽻 Center -68.4 -72.4 -70.4 ±2.0 -68.4 -72.8 -70.6 ±2.2
㽼 Left front -62.8 -64.8 -63.8 ±1.0 -62.7 -65.1 -63.9 ±1.2
Bottom area
㽽 Left back -64.1 -66.4 -65.3 ±1.2 -63.8 -66.6 -65.2 ±1.4
㽾 Right front -61.7 -63.6 -62.7 ±0.9 -61.5 -63.6 -62.6 ±1.1
㽿 Right back -63.3 -65.8 -64.6 ±1.3 -63.2 -66.0 -64.6 ±1.4
㾀 Center -63.0 -65.0 -64.0 ±1.0 -62.8 -65.1 -64.0 ±1.2
㾁 Center of Upper shelf -72.0 -76.6 -74.3 ±2.3 -72.1 -76.6 -74.4 ±2.3
㾂 Center of Lower shelf -64.2 -66.8 -65.5 ±1.3 -64.1 -66.9 -65.5 ±1.4
Average - - -68.9 - - - -69.0 -
- 37 -
AT20㷄䇭50䌈䌺䇭SV䋽-80㷄
-70
-72
-74
Chamber temp[㷄]
-76
㾂
-78
㽻
-80 㽶
-82
㾁
-84
0:00 0:30
30 1:00
60 1:30
90 2:00
120 2:30
150
Time[Minute]
AT20㷄䇭50䌈䌺䇭SV䋽-70㷄
-60
-62
-64
㾀
-66
Chamber temp[㷄]
㾂
-68
-70
㽻
-72
-74 㽶
-76 㾁
-78
-80
0:00 0:30
30 1:00
60 1:30
90 2:00
120 2:30
150 3:00
180
Time[Minute]
- 38 -
AT30㷄䇭50䌈䌺䇭SV䋽-80㷄
-70
-72
㾀
-74
Chamber temp[㷄]
-76
㾂
㽻
-78
-80
㽶
-82 㾁
-84
0:00 0:30
30 1:00
60 1:30
90 2:00
120 2:30
150
Time[Minute]
AT30㷄䇭50䌈䌺䇭SV䋽-70㷄
-60
-62
-64 㾀
-66
Chamber temp[㷄]
㾂
-68
-70
㽻
-72
-74
㽶
-76 㾁
-78
-80
0:00 0:30
30 1:00
60 1:30
90 2:00
120 2:30
150 3:00
180
Time[Minute]
- 39 -
AT20㷄䇭60䌈䌺䇭䌓䌖䋽-80㷄
-70
-72
-74 㾀
Chamber temp[㷄]
-76
㾂
-78
㽻
-80 㽶
-82
㾁
-84
0:00 0:30
30 1:00
60 90
1:30 120
2:00 150
2:30 180
3:00
Time[Minute]
AT20㷄䇭60䌈䌺䇭䌓䌖䋽-70㷄
-60
-62
-64
㾀
-66
Chamber temp[㷄]
㾂
-68
-70
㽻
-72
-74 㽶
-76 㾁
-78
-80
0:00 30
0:30 60
1:00 90
1:30 120
2:00 150
2:30
Time[Minute]
- 40 -
AT30㷄䇭60䌈䌺䇭䌓䌖䋽-80㷄
-70
-72
-74
Chamber temp[㷄]
-76
㾂
-78
㽻
-80 㽶
-82 㾁
-84
0:00 30
0:30 60
1:00 90
1:30 120
2:00 150
2:30 3:00
180
Time[Minute]
AT30㷄䇭60䌈䌺䇭䌓䌖䋽-70㷄
-60
-62
㾀
-64
㾂
-66
Chamber temp[㷄]
-68
-70
㽻
-72
-74 㽶
-76
㾁
-78
-80
0:00 30
0:30 60
1:00 90
1:30 120
2:00 150
2:30 180
3:00
Time[Minute]
- 41 -
Sample load test
Note) Following data are the reference only.
<Condition>
500ml water x 240 bottles 䋨Total:120L䋩
Measuring points: a, b, c, d, H EVA OUT, L EVA OUT
Ambient temperature: 30㷄
Front
1st shelf: 500ml×60bottles
20
0
L䌯䌡䌤䇭䌣
Temperature䋨㷄䋩
-40
HEva out
L䌯䌡䌤䇭䌢
-60
L䌯䌡䌤䇭䌡
-80 L䌯䌡䌤䇭䌤
-100
0 10 20 30 40 50 60 70 80 90 100 110
Time䋨Hr䋩
- 42 -
Temperatures after unit reaches to -80㷄
10
H Eva out
0
-10
-20
Temperature䋨㷄䋩
-30
-40
-50
-60
L Eva 䌯䌵䌴
L䌯䌡䌤䇭䌣 L䌯䌡䌤䇭䌤
-70
-80
L䌯䌡䌤䇭䌡 L䌯䌡䌤䇭䌢
-90
0 60 120 180 240 300 360
Time䋨min䋩
-20
-30
Temperature䋨㷄䋩
-50
-60
-70 L䌯䌡䌤䇭䌤
-80
L䌯䌡䌤䇭䌣
L䌯䌡䌤䇭䌡 L䌯䌡䌤䇭䌢
-90
0 60 120 180 240 300 360
Time䋨min䋩
- 43 -
SV-80㷄 Pull-up data
40
30
20
10 H Eva out
0
-10
L Eva 䌯䌵䌴
Temperature䋨㷄䋩
-20
-30
L䌯䌡䌤䇭䌢
-40
L䌯䌡䌤䇭䌡
-50 L䌯䌡䌤䇭䌤
-60
-70 L䌯䌡䌤䇭䌣
-80
-90
-100
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
Time䋨Hr䋩
- 44 -