Pagewriter Xli Users Guide
Pagewriter Xli Users Guide
PageWriter XLi
M1700A
ii
Safety Summary
Safety Symbols Marked on the Cardiograph
Alternating current
Keep dry
Fragile
iii
Conventions Used in This Manual
WARNING Warning statements describe conditions or actions that can result in personal injury
or loss of life.
CAUTION Caution statements describe conditions or actions that can result in damage to
the equipment or software.
Softkey represents the temporary key labels that appear on the keyboard
display.
iv
Documentation Map
Documentation Map
Enter patient ID
Store ECGs
Transmit or receive ECGs
Troubleshoot problems
Load paper
Change applications
v
Documentation Map
Order supplies
Use filters
vi
Contents
1 Introduction
2 Acquiring an ECG
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Understanding TPI Variables . . . . . . . . . . . . . . . . . . . . 3-2
Understanding ACI TIPI Variables . . . . . . . . . . . . . . . 3-3
Using the TPI and ACI-TIPI Applications . . . . . . . . . 3-3
Analyzing an ECG with the Predictive Instruments . . 3-4
Generating Reports with the Special Applications Off 3-5
Generating Reports with the Special Applications On 3-6
Auto Analysis and the Default Choice . . . . . . . . . . . . 3-7
Generating a STAT ECG Report . . . . . . . . . . . . . . . . . 3-8
Contents-1
4 Choosing Report Features
5 ECG Storage
Contents-2
Printing the Configuration . . . . . . . . . . . . . . . 6-16
Contents-3
Receiving ECG Reports from a 5600C
ECG Management System . . . . . . . . . . . . . 7-22
Receiving ECGs from Philips M1730A/
M3700A TraceMaster ECG Management
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
8 Troubleshooting
9 Supplies
Availability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
A Lead Systems
Glossary
Contents-4
1 Introduction
1
2
About This Manual
This guide contains reference information and configuration instructions for
experienced PageWriter XLi cardiograph users. For additional help on using
your cardiograph refer to the Philips PageWriter XLi Operating Guide.
1-1
2 Acquiring an ECG
ECG Technique
ECG technique is very important, both to avoid difficulty when taking
the ECG and to achieve the best quality result. There are three key
aspects of good ECG technique:
For best results, perform the following steps in the order given. More
details on good technique follow this list.
2-1
Acquiring an ECG
The more the patient relaxes, the less the ECG will be affected by
noise. Your good technique helps the patient relax. You can help the
patient to relax by the following:
l Your calm, relaxed attitude will help put the patient at ease.
2-2
Acquiring an ECG
The tip of each lead wire is lettered and color coded for easy lead
identification. For example, make sure that the RA lead wire and
electrode connect to the right arm and the RL lead wire and electrode
connect to the right leg.
RL On the right leg (inside calf, midway between knee and ankle)
LL On the left leg (inside calf, midway between knee and ankle)
2-3
Acquiring an ECG
l Making sure that all electrodes are clean and bright. (Dirty or
corroded electrodes prevent a good electrical connection.)
l Avoiding bony areas. Select flat, fleshy sites. You don’t have
to shave hair from the skin unless the hair is very thick.
2-4
Acquiring an ECG
A good test for firm electrode contact is to grasp the electrode and try
2
to move it. If it moves easily, the electrode connection is too loose. If
it digs into the flesh, the electrode is too tight. Do not allow the chest
electrodes to move in any way. Check the patient module display for
indications that the connections are good. Ideally, the noise bar should
remain in the green zone.
2-5
Acquiring an ECG
2-6
Acquiring an ECG
You can stop the recording before or during printing if you see artifact
or other ECG waveform problems on the screen. Modify lead
placement or improve patient preparation and resume recording the
ECG.
For further instructions on using the patient module display and the
preview screen, refer to the PageWriter XLi Operating Guide. 2
Quality Messages on the Cardiograph’s Display
After you start an Auto ECG recording, the cardiograph’s display will
show messages which indicate the quality of the recording. If there is
minimal noise and all electrodes are securely attached, the message
ECG ok will appear.
Some messages indicate problems with the leads. They list the
condition and the action to take. There are four conditions which
affect the ECG:
l Leads off
l AC noise
l Artifact
l Baseline wander
If any of these conditions are severe, the message directs you to retry
the recording. Correct the problem and then resume recording the
ECG.
2-7
3 Understanding the PageWriter XLi Special
Applications
Overview
The ACI TIPI (Acute Cardiac Ischemia - Time Insensitive Predictive
Instrument) and the TPI (Thrombolytic Predictive Instrument) are
software products that enhance the computer-assisted ECG analysis 3
capabilities of the PageWriter XLi Cardiograph. These "Predictive
Instruments" generate 0-100% Predicted Probability scores of ACI
(Acute Cardiac Ischemia) and patient outcome with and without
thrombolytic therapy for acute myocardial infarction (AMI). These
predicted probabilities are based on ECG features, patient age,
gender, blood pressure, chest pain status and time since ischemic
symptom onset. The cardiograph can be configured to automatically
print these probabilities on the Auto ECG report.
3-1
Understanding the PageWriter XLi Special Applications
acquired, can then be used along with actual patient outcome to help
improve patient management practices retrospectively.
l patient age
l patient gender
3-2
Understanding the PageWriter XLi Special Applications
NOTE For each of the clinical factors listed above, patient data must be
entered in order to produce a TPI report.
l patient age
l patient gender
NOTE For each of the clinical factors listed above, patient data must be
entered in order to produce the ACI-TIPI report.
3-3
Understanding the PageWriter XLi Special Applications
The exclusionary cases for both the TPI and ACI-TIPI applications
are listed in the Predictive Instrument Physician’s Guide. Please refer
to this document for information.
To use the TPI and ACI-TIPI applications, you must configure the
cardiograph and enable the applications. There are several types of
reports that are produced by the cardiograph. These reports are
summarized in Table 3-1.
3-4
Understanding the PageWriter XLi Special Applications
3-5
Understanding the PageWriter XLi Special Applications
When first turned on the PageWriter XLi cardiograph will have the
Special Applications turned off. The Special Application choices are
part of the Global Configuration Menu and enable access to the
following settings.
3-6
Understanding the PageWriter XLi Special Applications
For each of these report options, here are the resulting reports given
when Special Applications are Off and the Auto button has been
pressed:
3
l Adult: the XLi will do an 09 report (regardless of patient age)
l TPI: the XLi will do a TPI report. TPI screening will not
occur.
1. From the main display, press the F1 key until Auto Analysis
appears.
3-7
Understanding the PageWriter XLi Special Applications
l TPI: the XLi will do a TPI report. TPI Screening will not
occur.
l TPI
3-8
Understanding the PageWriter XLi Special Applications
l ACI-TIPI
l Standard ECG
l Vectorcardiography
3-9
Understanding the PageWriter XLi Special Applications
3-10
4 Choosing Report Features
This chapter describes the various ECG reports and how to print the
Extended Measurements report.
ECG Formats
The Auto Report
l Auto 3 x 5
The Auto 3 x 5 format displays consecutive 2 second
segments of 12 leads. The 5th lead column shows the
extended pediatric leads: V3R, V4R, and V7. One or three
leads can be displayed as rhythm strips at the bottom of the
Auto 3 x 5 report. The rhythm strips show the same 10 second
segments as in the Auto 3 x 5 section of the report.
l Auto 4 x 4
The Auto 4 x 4 format displays consecutive 2.5 second
segments of 12 leads. The 4th row consists of the extended
research leads: VX1-VX4 or V3R, V4R, V7, and V8. The
Auto 4 x 4 report can show 1 rhythm strip.
4-1
Choosing Report Features
l Auto 6 x 2
The Auto 6 x 2 format displays consecutive 5 second
segments of 12 leads, six leads at a time.
PR PR interval milliseconds
QT QT interval milliseconds
4-2
Choosing Report Features
l Normal
l Otherwise Normal
l Borderline
l Abnormal
l Defective Data
4-3
Choosing Report Features
l Or, type the patient ID number and press Enter . The most
recent ECG with that patient ID number appears. If you want
a different ECG for that patient, scroll through the list by
pressing Previous . Then press Select .
4-4
Choosing Report Features
You can also configure the cardiograph for Choice to print this report
when copying an ECG.
Manual Formats
Manual ECGs print continuously after you press Manual until you
press Stop . Manual ECGs are printed as the ECG waveform occurs.
Besides the conventional 12 leads, you may use one of the following
sets of supplemental leads:
Frank leads X, Y, Z
4-5
Choosing Report Features
12 I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
12 |I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
I, II, aVF, V4R, V3R, V1, V2, V3, V4, V5, V6, V7
4-6
Choosing Report Features
12 I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
I, II, aVF, V1, V2, V3, V4, V5, V6, X, Y, Z 4
Table 3-5 Manual Standard plus Generic Research Leads
4-7
Choosing Report Features
12 I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
I, II, V4R, V3R, V1, V2, V3, V4, V5, V6, V7, V8
12 aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6
4-8
Choosing Report Features
12 aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6
I, II, aVF, V4R, V3R, V1, V2, V3, V4, V5, V6, V7 4
Table 3-9 Manual Cabrera plus Frank Leads
4-9
Choosing Report Features
12 aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6
I, II, V1, V2, V3, V4, V5, V6, VX1, VX2, VX3, VX4
4-10
5 ECG Storage
Standard mode 5
l stored at 250 samples per second
l contain only the waveform segments printed on the Auto
report
l more than 100 ECGs can be stored on disk in Standard mode
Special mode
l stored at 500 samples per second
l contain a full ten seconds of information for all leads
l approximately 35 ECGs can be stored on a disk in Special
mode
l only ECGs stored in Special mode can be re-analyzed or re-
formatted
5-1
ECG Storage
Press Store to archive the last ECG report generated. The archived
report will be the last one generated.
For the report choices of: Adult, Pediatric, TIPI, and TPI, pressing
Store will archive the last report generated. Risk Management
Reports are never stored.
Pressing Store when Auto Analysis is set as the default setting will
generate the Default Store report type previously defined in the
Global Menu settings.
5-2
ECG Storage
l Do not eject the disk, move the cardiograph or turn off the
cardiograph when the disk is active in the drive. The disk is
active when the drive access light is on.
l Clean the disk drive every six months unless the working
environment is particularly dusty or dirty. Clean the drive
more often in dusty or dirty environments.
5-3
ECG Storage
The Store-Log lists all Auto ECGs stored on the flexible disk which
is in the cardiograph’s disk drive. The Store-Log is updated
automatically when you store an ECG and when you delete a stored
ECG from the flexible disk.
5-4
ECG Storage
A B C
D E F G H I J K
5-5
ECG Storage
A B C D
E F G H I J K LMNO
Description
A Information about ECGs stored on the flexible disk in the cardiograph
B Number of ECGs stored on the flexible disk
C Percentage of space available on the disk
D Date and time of the report
E Numerical order in which the ECG was stored
F Date and time when the ECG was recorded
G Patient name and identification number
H ECG format
I Initials of the person who recorded the ECG
J Department in which the ECG was recorded
K Institution, department and cardiograph identification number
L S-ECG storage mode (Standard is 250, Special is 500; 3x5 and 4x4 Reports are 500; SAECG is
SAE)
M T - Indication that ECG was transmitted. For 3x5 and 4x4 Reports a (-) will appear
N E - Patient ID information edited, Yes or No
O A - Criteria used to interpret the ECG
5-6
ECG Storage
5-7
6 Configuring Your Cardiograph
6-1
Configuring Your Cardiograph
NOTE If this is not the initial configuration, you may need to type the
configured password before you can make any changes.
Here are the basic descriptions for each softkey in this menu:
6-2
Configuring Your Cardiograph
The initial display is similar to the first parameter in the Global list.
The message line on top displays the parameter with its current
setting. The softkey labels are displayed on the bottom line.
6-3
Configuring Your Cardiograph
Configuration
GlobalID Transmit Files Exit
Table 6-1 lists each Global parameter. Factory default settings are
shown in boldface type. Additional information on parameter settings
can be found after Table 6-1.
Setting Choices
Parameter Comments
(default in bold)
6-4
Configuring Your Cardiograph
Setting Choices
Parameter Comments
(default in bold)
6-5
Configuring Your Cardiograph
Setting Choices
Parameter Comments
(default in bold)
6-6
Configuring Your Cardiograph
Setting Choices
Parameter Comments
(default in bold)
6-7
Configuring Your Cardiograph
Setting Choices
Parameter Comments
(default in bold)
Confirmation? PRELIMINARY -
MD MUST REVIEW.
Type a label,
30 characters maximum
6-8
Configuring Your Cardiograph
Setting Choices
Parameter Comments
(default in bold)
6-9
Configuring Your Cardiograph
Setting Choices
Parameter Comments
(default in bold)
ACI-TIPI? Off | On
TPI? On | Off
6-10
Configuring Your Cardiograph
Interpretation Parameters
6-11
Configuring Your Cardiograph
Line Frequency
This parameter selects the frequency of the filter enabled by the line
power filter. The user usually selects the frequency during initial
installation in response to a prompt. However, the user will not be
prompted when a custom configuration is installed. Including this
parameter in Global Configuration allows the user to change the
frequency selection after initial software installation.
Besides the pre-defined lead groups, you can configure one group of
three leads and one group of six leads that may be selected for Auto 3
x 4, 3R, Manual 3 or Manual 6 reports. These lead groups will appear
and be recorded when you select Custom 3-lead group and Custom
Manual 6-lead group reports from the main display.
1. Configure custom lead groups from Global Configuration when
the following display appears.
3. Use Tab to move to the next lead field. In this case, the next lead
field is aVF.
Selecting AutoCopy
6-12
Configuring Your Cardiograph
To conserve battery power, the cardiograph turns off after the period
of time specified by this parameter. The time is measured from the
completion of the last operation, the last key press, or disconnection
of all patient leads. The timeout function is disabled when the
cardiograph is plugged into AC power and the AC switch is on. If the
cardiograph turns off, current ID information and Copy memory are
lost.
6-13
Configuring Your Cardiograph
Setting a Password
CAUTION If the password is set and forgotten, you will be unable to change the
cardiograph configuration without re-installing the supplied software.
Configuration
GlobalID Transmit Files Exit
6-14
Configuring Your Cardiograph
l Allow BP entry*?
l Allow DX entry*?
l Allow RX entry*?
6-15
Configuring Your Cardiograph
6-16
Configuring Your Cardiograph
NOTE If an unformatted flexible disk is inserted into the disk drive, press
Yes to format the disk. Formatting a disk takes about 90 seconds.
You may also insert a formatted flexible disk at this time.
6-17
Configuring Your Cardiograph
6-18
7 Setting Up Your Cardiograph for Transmitting
ECGs
NOTE You will only be able to select from transmittable ECGs when you
press Transmit . You cannot transmit Manual ECGs, ECGs recorded
with research leads, or ECGs interpreted with a different criteria
program version than on the transmitting cardiograph. The criteria
program is the software used to interpret the ECG. The receiving
7-1
Setting Up Your Cardiograph for Transmitting ECGs
7-2
Setting Up Your Cardiograph for Transmitting ECGs
7-3
Setting Up Your Cardiograph for Transmitting ECGs
Configuration
GlobalID Transmit Files Exit
Configuration
Transmit Receive AutoDial Exit
Configuration Site
#1 #2 #3 #4 Exit
Connection? None
Global Choose Exit
6. Press Choose until you see the choice Direct Connection, then
press Enter . The next display will allow you to select the baud
rate. When you choose direct connection, the default baud rate
will be 9600.
7-4
Setting Up Your Cardiograph for Transmitting ECGs
Before using the M1706B modem you must connect the cables.
Figure 7-2 shows how to connect the cables for transmitting ECGs by
modem.
7-5
Setting Up Your Cardiograph for Transmitting ECGs
Refer to Figure 7-2 and perform the following steps to install the
modem.
1. Turn the cardiograph to Standby.
2. Turn the modem power switch off.
NOTE Make sure that the battery level is sufficient to prevent loss of
configuration settings and software in the cardiograph when AC
power is removed. Check the Battery Level indicator lights on the
cardiograph’s key panel to ensure that they are all on.
7-6
Setting Up Your Cardiograph for Transmitting ECGs
NOTE Connect the modem to the cardiograph and to the telephone line in the
following order. This will avoid the possible electrical hazards from
the outlet and telephone line.
6. Plug the modem power cable into the plug on the modem marked
9.0V ~AC.
7. Plug the other end of the modem power cable into the cardio-
graph.
8. Insert either end of the telephone cable into the connector on the
back panel of the modem marked To Line.
9. Insert the free end of the telephone cable into the telephone wall
jack.
10. Plug the cardiograph power cable into the cardiograph and an AC
outlet if they are not already plugged in.
11. Turn on the cardiograph’s AC power switch.
12. Turn on the modem.
13. Turn on the cardiograph.
The modem is now installed and ready for use. You must configure
the cardiograph for modem usage before transmitting ECGs.
7-7
Setting Up Your Cardiograph for Transmitting ECGs
Configuration
Global ID Transmit Files Exit
Configuration Site
Transmit Receive AutoDial Exit
Configuration Site
#1 #2 #3 #4 Exit
7-8
Setting Up Your Cardiograph for Transmitting ECGs
Parameter Selection
7-9
Setting Up Your Cardiograph for Transmitting ECGs
2. Press Choose to select the site (#1, #2, #3, or #4) to which the
cardiograph will automatically transmit when the operator presses
Transmit .
3. Press Enter and the selected site will be set. The display will
return to the Transmission Configuration menu. If you have not
configured the transmission site, you can configure it from this
menu.
7-10
Setting Up Your Cardiograph for Transmitting ECGs
7-11
Setting Up Your Cardiograph for Transmitting ECGs
WARNING Some Group III fax machines distort the received fax transmission by com-
pressing the image to fit the viewing or printing device. This distortion can
affect the amplitude and time scale of an ECG transmitted by fax.
For example, PC based fax/modems and fax cards will normally compress
the received image when displaying the fax transmitted ECG on the PC
monitor and can compress the image when the fax is printed.
To determine whether a fax machine distorts the received ECG fax, you will
need to measure the ECG report’s grid spacing or the calibration pulse and
check these measurements against the printed scale factors shown at the
bottom of a printed ECG report. The grid is nominally printed at 1 mm inter-
vals in both the time and amplitude directions. The width of the calibration
pulse is 200 msec and its height represents 1 mV. Therefore an ECG report
with a 10 mm/mV amplitude scale and 25 mm/sec time scale should have a
calibration pulse that is 10 mm tall and 5 mm wide.
7-12
Setting Up Your Cardiograph for Transmitting ECGs
Before using the FAX/modem you must connect the cables as shown
in Figure 7-4.
Refer to Figure 7-4 and perform the following steps to install the
modem.
7-13
Setting Up Your Cardiograph for Transmitting ECGs
NOTE Make sure that the battery level is sufficient to prevent any loss of
configuration settings and software in the cardiograph when AC
power is removed. Check the Battery Level indicator lights on the
cardiograph’s key panel to ensure that they are all on.
CAUTION Connect the modem to the cardiograph and to the telephone line in the
following order. This will avoid the possible electrical hazards from
the outlet and telephone line.
6. Plug the modem power cable into the plug on the modem marked
9.0 V ~ AC.
7. Plug the other end of the modem power cable into the cardio-
graph.
8. Insert either end of the telephone cable into the connector on the
back panel of the modem marked To Line.
9. Insert the free end of the telephone cable into the telephone wall
jack.
10. Plug the cardiograph power cable into the cardiograph and an AC
outlet if they are not already plugged in.
11. Turn on the cardiograph’s AC power switch.
12. Turn on the modem.
7-14
Setting Up Your Cardiograph for Transmitting ECGs
The FAX/modem is now installed and ready for use. You must
configure the cardiograph for FAX/modem usage transmitting ECGs
by FAX.
Configuration
Global ID Transmit Files Exit
7
3. Press Transmit . The Transmit Configuration menu appears.
Configuration Site
Transmit Receive AutoDial Exit
Configuration Site
#1 #2 #3 #4 Exit
7-15
Setting Up Your Cardiograph for Transmitting ECGs
Connection? None
Enter Choose Exit
Selection
Parameter
(default setting in bold)
7-16
Setting Up Your Cardiograph for Transmitting ECGs
7-17
Setting Up Your Cardiograph for Transmitting ECGs
7-18
Setting Up Your Cardiograph for Transmitting ECGs
WARNING The PageWriter XLi A.05.00 software allows you to print ECGs on a Laser-
Jet. The LaserJet printer is NOT isolated to UL544/IEC-601-1 leakage cur-
rent specifications. Be sure to disconnect the LaserJet from the
cardiograph when you want to record ECGs.
1. With the printer on, press the Online key until the light turns
off.
7-19
Setting Up Your Cardiograph for Transmitting ECGs
After connecting the cables, you must configure the cardiograph for
printing on the HP LaserJet.
Configuration
Global ID Transmit Files Exit
Configuration Site
Transmit Receive AutoDial Exit
Configuration Site
#1 #2 #3 #4 Exit
Connection? None
Enter Choose Exit
6. Press Choose until you see the choice LaserJet. Press Enter .
The baud rate display appears.
7-20
Setting Up Your Cardiograph for Transmitting ECGs
7. Press Choose until you see the baud rate 19200. Press Enter .
8. Select whether to print a grid. If you have unlined paper in the
printer, this selection will print the grid with the ECG.
Receiving ECGs
When configured and connected for receiving, the cardiograph may
receive ECGs whenever it is turned on. ECG reports are printed
immediately when they are received. These reports cannot be copied
or stored on the flexible disk.
Configuration Site
Transmit Receive AutoDial Exit
7-21
Setting Up Your Cardiograph for Transmitting ECGs
Connection? None
Enter Choose Exit
Configuration
Global ID Transmit Files Exit
Configuration Site
Transmit Receive AutoDial Exit
Connection? None
Enter Choose Exit
7-22
Setting Up Your Cardiograph for Transmitting ECGs
The cardiograph may receive ECGs via the FAX/modem any time the
cardiograph is configured for FAX/modem usage and is idle.
Configuration
Global ID Transmit Files Exit
Configuration Site
Transmit Receive AutoDial Exit
Connection? None
Enter Choose Exit
7
Choose the FAX programmable modem selection for connection type.
The default baud rate for the FAX programmable modem is 2400.
After you have configured a site number for the 5600C ECG
Management System with the appropriate connection parameters, you
can receive reports from that system.
7-23
Setting Up Your Cardiograph for Transmitting ECGs
3. Press Query . You may then select the site number which is
assigned to the 5600C ECG Management System.
4. After you press the site number, the cardiograph will call the sys-
tem, ECGs will be transmitted from the ECG Management Sys-
tem to the cardiograph, and the cardiograph will print the reports
as they are received.
There are two types of Query, Standard query and Interactive query.
7-24
Setting Up Your Cardiograph for Transmitting ECGs
NOTE Query will timeout after an idle period. The length of this period is
determined by how the ECG Management system is configured.
NOTE Press No to start a batch query. All reports queued for your location
will now be printed.
7
4. Press Yes to select Interactive query. You will be asked for your
ID number. Your ID number is the same as your password. See
your system administrator if you do not know your ID number.
5. Type a valid ID number and press Enter . You will be asked for
an institution number:
7-25
Setting Up Your Cardiograph for Transmitting ECGs
6. Press .
7. After you enter a valid institution number, you will be asked for
patient ID. Type either a patient ID or a patient name no more
than 16 characters long.
7-26
Setting Up Your Cardiograph for Transmitting ECGs
From this display you can scroll through a list of selected ECGs.
The top line contains:
13. Press More and the softkey selections will change as shown
below.
III PATIENTID 01/26/01 06:21:52
Name NewInst New ID More
7-27
Setting Up Your Cardiograph for Transmitting ECGs
You may end the interactive query session any time by pressing the
Stop key on the cardiograph. The transmission link will end
immediately.
7-28
8 Troubleshooting
RL Dotted line on all leads X on all leads Flat line on all leads
RA, Dotted line on all leads X on lead which is off and Flat line on all leads
LA, X on chest leads
and/or
LL
Chest Dotted line on lead(s) X on lead(s) which is off Flat line on lead(s) which
lead(s) which is off is off
8
Troubleshooting ECG Noise
Use the electrode status indicators (noise bars) on the patient module
to check the quality of the ECG signal. The preview screen is another
way to check the ECG signal quality before recording the ECG. Noise
on the ECG recording is often due to poor electrode contact or to
8-1
Troubleshooting
Configurable Settings
Filter Type Interference Sources In the Filter Box
(default in bold)
8-2
Troubleshooting
If an error number appears, note the number and then perform the
following steps.
Latin America
8
Tel: 305-269-7500
8-3
Troubleshooting
Canada
Australia France
Germany Italy
Belgium Finland
Spain
Tel: 34 91 631 31 00
8-4
Troubleshooting
Symptom What to Do
Message: No battery. Install the battery, then try turning on the cardiograph.
Load battery. Cardiograph won’t turn
on.
Battery charge level indicator never 1. With the cardiograph connected to AC power, turn on the
indicates full capacity. cardiograph. If only a single battery level indicator is lit
(other than the Battery Low Warning), reload the
operating system, insuring the paper compartment door
is closed. If the operating system is loaded with this door
open, only a single indicator will light.
2. If after completing the step above, all battery level
Battery charge level indicator always indicators are not lit, the battery requires further
indicates full capacity on AC power charging. When checking the battery level, always turn
with battery. the cardiograph on while connected to AC power.
3. Replace battery.
4. Call your Philips service representative.
8
8-5
Troubleshooting
NOTE When dealing with noise problems, remember that most noise is the
result of technique or poor electrode contact. If noise persists after
checking the electrode placement and other elements of technique, the
noise may be the result of faulty electrode wires. Only when you
eliminate technique and electrode wires as the causes of the noise
should you consider calling your Philips service representative.
Symptom What to Do
Leads off not indicated when l Remove electrode wire from patient module.
wire is off.
l Ensure the contacts on the patient module are
clean and dry.
l If Leads off is not indicated, call your
Philips service representative.
Leads off indicated when wire is l Remove electrode wire and install shorting plug.
not off. If Leads off is indicated, call your Philips service
representative.
l If Leads off is not indicated, replace electrode
wire.
8-6
Troubleshooting
Symptom What to Do
Electrode off on ECG when no l Be sure STAT ECG is not taken within 15 sec-
electrode is off. onds of cardiograph power-up.
Noisy lead(s): AC (regular l See Table 6-1 in the PageWriter XLi Cardiograph
pattern) Operating Guide.
For 60 Hz, 12 peaks/5 mm
l Reposition patient module cable away from AC
at 25 mm/ sec; for 50 Hz, 10
power.
peaks/5 mm
l Try plugging cardiograph into AC outlet.
Noisy lead(s): Muscle artifact. l See Table 6-1 in the PageWriter XLi Cardiograph
Operating Guide.
l Replace disposable electrodes or lead adapters. 8
l Call your Philips service representative.
8-7
Troubleshooting
Symptom What to Do
Noise in only one V lead. Noise l See Table 6-1 in the PageWriter XLi Cardiograph
in general. Operating Guide.
l Isolate defective wires and replace.
Symptom What to Do
Paper doesn’t move; printer error message l Replace paper or clear paper jam.
displayed after checking paper supply.
l Call your Philips service
representative.
Paper moves then stops and displays an l Make sure cardiograph is loaded properly with
error message. Check paper supply. approved thermal paper.
l Clean paper sensor lens. See PageWriter Cardio-
graph XLi Operating Guide for instructions.
l Call your Philips service representative.
Paper moves but nothing prints. l Make sure cardiograph is loaded with approved
Paper moves but printing is faint. thermal paper.
l Call your Philips service representative.
8-8
Troubleshooting
Symptom What to Do
Paper stops in the wrong place after l Make sure cardiograph is loaded with approved
pressing Page Advance or Auto . No thermal paper.
error message. l Clean sensor lens.
l Call your Philips service representative.
Poor print quality or some dots not l Clean printhead. See the PageWriter XLi Cardio-
printing. graph Operating Guide for instructions.
l Call your Philips service representative.
Printing is dark on one side of page but l Call your Philips service representative.
faint on the other side.
Symptom What to Do
Error message Try new disk appears. l Try known good disk.
l Call your Philips service representative.
Error message Unrecognized file l Corrupt file, try known good file.
type appears.
8
8-9
Troubleshooting
Symptom What to Do
Washed out display. l Adjust brightness and contrast controls for best dis-
play appearance.
8-10
Troubleshooting
Symptom What to Do
Keys won’t work. Unable to enter l Call your Philips service representative.
data or operate controls.
Symptom What to Do
Message ERROR 153X: Remote site l This is a message from the remote site informing
failure. the user that the communication link has been ter-
minated.
Unable to transmit or receive. l Check to see that modem is turned on and plugged
in, if using a modem.
l Call your Philips service representative.
Message No answer at site. Retry? l Type an X before typing the phone number. The
when there is a dial tone. The modem will not wait for a dial tone.
cardiograph will not be able to start the
transmission because the modem doesn’t
recognize the dial tone.
3 x 5, 4 x 4 untransmittable, FAX only l This is a message to indicate that the user has
attempted to transmit reports that are unrecognized
by the remote site.
8
8-11
9 Supplies
Availability
Pricing and availability of the following and other supplies are
available from Philips Medical Supplies Centers.
l U.S.: 1-800-225-0230
Paper
Battery
9-1
Supplies
Electrodes
40418A Welsh electrode; 15mm base 5cc bulb; screw connect; 6/box
40421A Welsh electrode; 15mm base 5cc bulb; push-in connect; 6/box
(AHA Leads)
40490E Welsh electrode; 4mm conn.(banana); 15mm cup; 2cc bulb (IEC
Leads)
9301-0119 Large welsh electrode; 30mm dia cup; screw connect; 1/box
40419A Replacement bulb; 5cc for all 15mm dia. cups; 12/pk
40422A Small replacement bulb; 2cc for all 15mm dia. cups; 12/pk
Lead Adapters
9-2
Supplies
Lead Adapters
40431B Lead adapter from 1/8" post to alligator clip; 10/pk. (AHA Leads)
40432B Lead adapter from 4mm post to alligator clip; 10/pk. (IEC Leads)
Patient Cables
9-3
Supplies
Replacement Leads
M1713B AHA Complete Lead Set; Chest leads, 24", limb leads 39"
M1716B IEC Complete Lead Set; chest leads 60 cm; limb leads 100 cm
M1717B Extra Leads Set, AHA or IEC (Pediatric, Frank, & Research); 24"/60 cm
(qty 4)
M1718A Lead Support Kit; 4 shorting plugs, 10 banana adapters, 2 lead separators
Other
M1700-89004 Box of 10 High Density Disks (available from your Philips service
office)
9-4
A Lead Systems
A
Frank Leads
Table A-1 Frank Lead Electrode Positions
A-1
Lead Systems
Additional leads can be placed into the patient module. When the
power to the cardiograph is turned on, the additional leads will be
recognized by the cardiograph. The cardiograph will identify the
additional leads as V3R, V4R, V7 and V8. The V8 lead position on
the module is labeled H.
A-2
Lead Systems
Pediatric 12 Leads + 3 x 4, 6 x 2,
V4R, V3R, V7 3x5
Standard 12 Leads 3 x 4, 6 x 2
A-3
Lead Systems
Auto
Patient Module /
Global Configuration Lead Type Lead Set Report
Shorting Plug
Format
A-4
B Glossary
B
AC filter
The configurable filter which screens out ECG artifact caused by power
line interference.
adult criteria
AHA leads
alphanumeric
analysis criteria
Rules used to interpret ECGs. (See adult criteria, pediatric criteria, and
ECG analysis)
Glossary-1
applications
Software used for a specific task. The PageWriter XLi contains the
SAECG application and the Preview Plus application.
artifact
ECG waveform distortion that may diminish ECG quality. ECG artifact
(or noise) may be caused by electrical interference, poor electrode
connections, or patient movement.
artifact filter
Philips term for filter which screens out noise on the ECG caused by
muscle tremor. Operators turn on the artifact filter by pressing the Filter
key on the key panel.
Auto ECG
Twelve-lead ECG recorded and analyzed over a ten second period and
printed in a preselected format.
AutoCopy
AutoDial
AutoStore
Glossary-2
baseline wander
Philips term for the process cardiograph uses to turn off automatically
after a preset time period to conserve power. The number of minutes
before battery timeout can be set in configuration mode.
baud rate
The speed at which data (ECGs) can be transmitted from one instrument
to another.
Cabrera
calibration pulse
A 200 ms, 1 mV square wave pulse which appears on the printed record.
Calibration pulse shows the sensitivity at which the ECG was recorded
and may show the effect of the filters.
CheckDisk
Philips term for softkey function which checks the percentage of flexible
disk storage available for storing Auto ECGs.
Glossary-3
configuration
continuous preview
The Preview Plus application that allows you to view the ECG waveform
on the preview screen without pauses when you print an ECG report.
cycle power
Data Commport
The cardiograph connector into which the modem data cable or direct
connection cable is inserted for ECG transmission.
delayed Manual
The Preview Plus application that allows you to capture and print ECG
information that you see on the preview screen. Ordinarily the ECG
information you see printed is what occurred after you pressed Manual .
ECG analysis
ECG report
Glossary-4
ECG-Log
Philips term for the softkey function which accesses the list of the last 60
ECGs recorded on a cardiograph.
FAX
A device that allows the cardiograph to send and receive ECGs and
FAXes (Group III only).
file
flexible disk
A disk which can be inserted in the disk drive to store data such as Auto
ECGs or the system configuration. Flexible disks can also be used to load
software.
format
The manner in which ECG waveforms are presented on the printed ECG
report. ECG format is selected by the operator.
Frank leads
frequency response
front panel
Cardiograph area that includes the front panel display and the keyboard.
Glossary-5
global configuration
Hertz (Hz)
ID fields
Philips term for the areas where variable patient information can be
entered. Using the ID fields, the operator can key in information such as
patient identification number, name, and age.
IEC leads
jittery waveform
Irregular up and down movement on the baseline of the ECG often caused
by patient movement or muscle tremor.
Manual ECG
ECG report format which runs continuously until the operator stops the
recording. The ECG may show three, six, or twelve lead waveforms.
Many institutions and physicians may identify this format as a rhythm
strip.
measurements
Glossary-6
Menu key
modem
operator
overread
password
patient module
Philips term for the remote unit that contains all of the cardiograph’s ECG
data acquisition electronics, the display where electrode status appears,
and a remote start/stop key. The patient module connects to the patient
data cable and to the leads attached to the patient.
Philips page description language for printers. Used for defining a page,
selecting fonts, and other formatting elements.
Glossary-7
pediatric criteria
preliminary report
An ECG report that has not been reviewed by a qualified physician. (See
overread)
Preview Plus
preview screen
Philips term for the optional screen which, when installed on the
cardiograph, shows the ECG traces as they will appear on the printed
ECG report.
power on application
query
reformatted report
A printed report that is printed with a different lead presentation than the
original ECG report. For example, you can reformat a 3 x 4 ECG and
print it as a 6 x 2 ECG.
Glossary-8
research lead set
rhythm strip
Philips term for ten second recording of a particular lead that is printed at
B
the bottom of an Auto ECG report. (See Manual and Auto ECG)
shorting plug
Small spacer plug inserted in patient module lead slot when lead is not in
use.
softkeys
standard leads
The conventional twelve lead order is I, II, III, aVR, aVL, aVF, V1 - V6.
(See Cabrera)
Store-Log
Philips term for function which accesses a list of all ECGS stored on the
flexible disk.
Glossary-9
transmission site
Welsh cups
Glossary-10
Index
i
Index
ii
Index
iii
Index
iv
Index
v
Index
vi
Index
vii
Index
viii
3DUW1XPEHU0
3ULQWHGLQ86$-DQXDU\
(GLWLRQ
'$(&&&'