X-QUANG DI DỘNG POLYMI PLUS ANH+VIỆT

The document provides service instructions for a medical imaging device. It details checking assemblies, troubleshooting errors, performing service programs and measurements, replacing important components, and changes from previous versions. The document is technical in nature and aimed at servicing the described medical device.

Uploaded by

Linh Nguyen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
22 views54 pages

X-QUANG DI DỘNG POLYMI PLUS ANH+VIỆT

The document provides service instructions for a medical imaging device. It details checking assemblies, troubleshooting errors, performing service programs and measurements, replacing important components, and changes from previous versions. The document is technical in nature and aimed at servicing the described medical device.

Uploaded by

Linh Nguyen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 54

POLYMOBIL Plus

SP

Service Instructions

© Siemens AG 1998
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model or design, are reserved.

English
Print No.: SPR8-125.061.01.02.02 Release Date: 01.99
Replaces: SPR8-125.061.01.01.02
0- Revisio
Chapter Page Rev.
0 all 02
1 all 02
2 all 02
3 all 02
4 all 02
5 all 02
6 all 02
7 all 02

POLYMOBIL SPR8- Page 2 Siemens


125.061.01 of 4 AG Medical
Content 0-
Page
1 Prerequisites 1-1

Text emphasis...................................................................................................................1-1
Required documents.........................................................................................................1-1
Tools, measurement and auxiliary devices required.........................................................1-1
Safety information and protective measures.....................................................................1-2
Replacing damaged or missing screws.............................................................................1-4
Cleaning............................................................................................................................1-4

2 Checking the assemblies 2-1

Preparation........................................................................................................................2-1
Position and function of the LEDs, fuses, test points and potentiometers........................2-4
D915 CPU, control and measurement value acquisition.............................................2-4
D925 Rotating anode starter, filament inverter, ON/OFF switching circuit..................2-5
U1 power supply unit 5 V ± 5% / 15 V ± 10 %.............................................................2-6
U2 power supply unit 12.5 V ± 10 %...........................................................................2-6
D950 capacitor charging circuit...................................................................................2-7
D970 Capacitor bank...................................................................................................2-7
Status of the LED’s (optimum condition).....................................................................2-8

3 Troubleshooting 3-1

Error messages.................................................................................................................3-1
Troubleshooting after error messages..............................................................................3-2

4 Service programs 4-1

Service programs available...............................................................................................4-2

5 Measurements 5-1

Line voltage.......................................................................................................................5-1
Measuring KVact and Itube...............................................................................................5-1
Capacitor charging circuit..................................................................................................5-2
Charging frequency.....................................................................................................5-2
Capacitor charging current..........................................................................................5-3
Capacitor voltage (Uc) and charging time...................................................................5-4
Rotating anode starter.......................................................................................................5-5
Acceleration.................................................................................................................5-5
Brakes..........................................................................................................................5-6
kV Control..........................................................................................................................5-7
Checking the maximum main inverter frequency........................................................5-7
Measuring the oscillating current.................................................................................5-8
Measuring the kVact and kVnom.................................................................................5-9
Error in the kVact acquisition.....................................................................................5-10
Testing the filament circuit...............................................................................................5-11
Filament circuit in standby.........................................................................................5-11
Setting the maximum filament inverter frequency.....................................................5-12

0-4 Contents
Siemens AG SPR8- Page 3 of POLYMOBIL
Medical Rev. 01. 4
0- Revisio
Page
Measuring the filament nominal value.......................................................................5-13
Setting the tube current for the exposure..................................................................5-13
Testing and setting the mAs counter...............................................................................5-15
Coincidence of light and radiation fields..........................................................................5-16
Testing and readjusting the counterbalance....................................................................5-19
Readjusting the counterbalance with accessories attached...........................................5-19

6 Replacing important components 6-1

Replacing the single tank generator..................................................................................6-1


Replacing the collimator....................................................................................................6-3
Replacing the lamp for the localizer..................................................................................6-4
Replacing the steering caster............................................................................................6-5
Replacing the capacitor bank............................................................................................6-6

7 Changes to previous version 7-1

POLYMOBIL SPR8- Page 4 Siemens


125.061.01 of 4 AG Medical
Prerequisite 1-
Text emphasis
WARNING "Warnings" are information provided with special emphasis
when there is the potential for personal injury to the operator or
patient.

CAUTION "Cautions" are information provided with special emphasis


when there is the potential for damage to the equipment.

NOTICE "Notices" are information provided with special emphasis to


facil- itate proper use of the equipment or proper execution of a
proce- dure.

Required documents
 Block circuit diagram X 069I
 Function description SPR8-125.041.01...

A detailed circuit diagram, DIN A 3, can be ordered under item


NOTE number 18 20 377 X069I.

Tools, measurement and auxiliary devices required


The tools, measurement and auxiliary devices required, with the exception of the standard
service equipment and the current transformer, are listed in part 3 of the ARTD.

 Standard service kit


 Digital multimeter
 2-channel memory oscilloscope
 mAs meter
 Protective conductor meter
 Equivalent leakage current meter
 Torque wrench 20 Nm - 100 Nm
 Loctite 242
 Current transformer (50 A : 50 mA) (31 51 289 B2716)
with 10 Ohm; 0.5 W; 1% metal film (10 14 984 B0512)
resistor Refer also to Speed Info 06/91

CAUTION The oscilloscope has to be connected to ground to perform mea-


surements. A TEK video isolator and the trigger set have to be
used when ground loops may distort the measurement results.

Siemens AG SPR8- Page 3 of POLYMOBIL


Medical Rev. 01. 4
1- Prerequisite
Safety information and protective measures
 When performing service work and tests, please note:
CAUTION
- the product-specific safety information in the document,
- the safety information in RA0-000.012.. .in the Register of the
POLYMOBIL Plus binder, as well as
- the safety information contained in the ARTD Part 2.
 Tests and adjustments that must be performed with
radiation ON, are identified by the radiation warning symbol
. Radiation protection must be worn during these types
of adjustments.

POLYMOBIL SPR8- Page 2 Siemens


Rev. 01. of 4 AG Medical
Prerequisite 1-

Fig. 1

WARNING When working on the open system, there is the danger of


Electric shock !
 The capacitor bank can still be charged.
Do not attempt to work on the system while this condition
exists.
 After switching off the system, approximately 350 V may still
be present in the system ; even after disconnecting the line
voltage plug. Within 10 minutes this voltage will drop to
approximately 10 V.
 Always measure the actual voltage present with the DVM at
test points -VCC and + VCC on the D 960 inverter board or
(more accessible) on the D 970 capacitor board at the +
connection point of capacitor C3 and on the right side of fuse
F3 (refer to Fig. 1).
 LED’s V1 ... V10 on D 970 go out at a significantly higher
voltage level and therefore they are not reliable safety
indicators.
 If a fuse on the D 970 has responded, high voltage may still
be present at the affected capacitor even after a prolonged
period of time.
 The capacitor discharging circuit utilizes the D925 board,
CS and LS relays. If connectors X3 or X9 on D 925 or D 950
are not inserted or if there is a defect in the circuit, the C-
bank will not discharge.
This can cause life-threatening voltage to be present in the
system even after a prolonged period of time.
 Refer also to chapter "Replacing the capacitor bank".

Siemens AG SPR8- Page 3 POLYMOBIL


Medical Rev. 01. of 4
1- Prerequisite

 Connect the POLYMOBIL Plus only to a line voltage supply (line voltage receptacle)
that complies with the requirements of VDE 0107, or corresponds to the local national
regulations.
 Disconnect the POLYMOBIL Plus at the line voltage OFF switch on the
operating console and disconnect the line voltage plug prior to any service
work.
 Remove or insert boards with the generator switched OFF only;
observe ESD guidelines when handling boards.

Replacing damaged or missing screws


 Damaged or missing screws may only be replaced by steel screws as specified in the
installation drawings conforming to DIN267 and must have the specified tensile
strength.
CAUTION
All Allen screws must have a tensile strength rating of 8.8.

Cleaning
 The unit must always be switched off or disconnected before cleaning.
 Use only water to clean the unit or a lukewarm, dilute solution consisting of water
and a household cleaning agent.
 Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol
or spot remover. Do not spray with water.
 For additional information, refer to the Operating Instructions "Cleaning/Disinfection".

POLYMOBIL SPR8- Page 4 Siemens


Rev. 01. of 4 AG Medical
Checking the 2-
Preparation

Fig. 1

WARNING When working on the open system, there is the danger of


Electric shock !
 The capacitor bank can still be charged.
Do not attempt to work on the system while this condition
exists.
 For details see chapter "Prerequisites".

Siemens AG SPR8- Page 1 POLYMOBIL


Medical Rev. 01. of 8
2- Checking the

Fig. 2 Fig. 3

To check the assemblies, remove the upper covers, the handles and the lower front cover
of the switchbox.

Proceed as follows:
 Switch POLYMOBIL Plus OFF and disconnect the line voltage plug.
 Remove the 4 Allen screws (1/Fig.2) in the handle and remove the
handle (refer to Fig.2 and Fig.3).

POLYMOBIL SPR8- Page 2 Siemens


Rev. 01. of 8 AG Medical
Checking the 2-

Fig. 4 Fig. 5

 Remove the four Allen screws at the bottom of the switchbox cover (2/Fig.4) and
remove the cover.
 After removing the four panel screws (3/Fig.4), remove the bottom cover at the front
with the cassette tray.

WARNING Always measure the actual voltage present with the DVM at test
points -VCC and + VCC on the D 960 inverter board or (more
accessible) on the D 970 capacitor board at the + connection
point of capacitor C3 and on the right side of fuse F3.

 Remove the four screws of the top housing (4/Fig.5)


 To stay clear of the exposure guide holder, tilt the housing forward and lift it up
to remove it.
Siemens AG SPR8- Page 3 POLYMOBIL
Medical Rev. 01. of 8
2- Checking the
Position and function of the LEDs, fuses, test points and
potentiometers

NOTICE Location of components and boards, see block diagram.

D915 CPU, control and measurement value acquisition


LED:
V22 Preparation ON
V23 Exposure ON
V24 Error occurred
V25 not used
V34 Radiation ON
Test points:
X15 Pin Color
1 black Ground (+ 5 V)
3 black Ground (+ 5 V)
4 red + 5 V  5%
5 brown +15 V  10 %
6 blue Ground (+/- 15 V)
7 yellow -15 V  10 %
Potentiometers:
P1 Setting of the mAs counter
P2 OFFSET mAS - converter
Set at the factory - Do not change!
P3 Max. filament inverter
frequency P4 Max. main inverter
frequency
P5 Adjustment/setting of filament heating circuit
P6 Nominal filament current (standby and ZB), nominal tube current
(exposure)
Set at the factory - Do not change!
P7 Nominal kV value

POLYMOBIL SPR8- Page 4 Siemens


Rev. 01. of 8 AG Medical
Checking the 2-
Reset - switch
SW2

SW2

X12
D 915 P6 P7
V22 V23 V24 V25

X15
P5
ST
P4

P3 P2
P1 X11 V34 X20 Service

X10

1 1 1

GNDJR REG CAL KVP KV KVN IH JRS KVS F1 II1 I2 TRG JIM SCHGND VC BAT KVM SCM VR

Fig. 6

D925 Rotating anode starter, filament inverter, ON/OFF switching circuit


LED:

VP +15 V for inverter frequency ok


+12V +12.5 V auxiliary voltage ok
I/O +12.5 V auxiliary voltage for I/O ok
AR Line voltage for anode starter ok
+DC, -DC 300 V voltage for filament ok

Fuses:

F1 10 AT + 12.5V for collimator lamp


F3 0.6 AT Auxiliary voltage +12.5 V for I/O
F4 0.6 AT Auxiliary voltage +12.5 V
F5 3.2 AF Line voltage for anode starter
F6 1.0 AT Line voltage for filament circuit
F7 10 AT Power
F8 10 AT Power

Siemens AG SPR8- Page 5 POLYMOBIL


Medical Rev. 01. of 8
2- Checking the
Measurement points:

VP +15 V  10 % for inverter frequency


P12 + 12.5 V  10 % auxiliary voltage
GNDL Ground (+ 12.5 V)

X4 Pin Color
1 red +12.5 V  10 %
2 red +12.5 V  10 %
3 red +12.5 V  10 %
4 black Ground (+12.5 V)
5 black Ground (+12.5 V)
6 black Ground (+12.5 V)

U1 power supply unit 5 V  5% / 15 V  10 %


Fuse: 2 A T

U2 power supply unit 12.5 V  10 %


Fuse: 5 A T

POLYMOBIL SPR8- Page 6 Siemens


Rev. 01. of 8 AG Medical
Checking the 2-
D950 capacitor charging circuit
LED:

LINE (V25) Line voltage ok


350 Vcc (V30) Capacitor voltage present (the voltage can also be considerably lower
than 350 V)

Fuse:

F1 10 AT Charging circuit

Test points:

ERR Error output charging error


VC Capacitor voltage 1 V 100 V
IC Capacitor charging current
FR1 Charging circuit frequency (dependent on the line voltage and VC)
GND Ground

D970 Capacitor bank


LED:

V1 - V10 Capacitor voltage present at C1 - C10

Fuse:

F1 - F10 20AF Capacitor charging voltage for C1 - C10


F11 80AF Capacitor voltage for inverter
F12 20AF Charging voltage for capacitor bank

Test points:

-VCC, +VCC Capacitor voltage

Siemens AG SPR8- Page 7 POLYMOBIL


Medical Rev. 01. of 8
2- Checking the
Status of the LED’s (optimum condition)
a) POLYMOBIL Plus OFF, line voltage plug connected:

D925: LED +12 V ON


LED I/O ON

D950: LED 350 Vcc (V30) glows for 10 minutes following shutdown

D970: all LED’s glows for 10 minutes following shutdown

b) POLYMOBIL Plus ON, standby

D915: The exposure values are displayed

D925: LED VP ON
LED +12 V ON
LED I/O ON
LED AR ON
LED +DC ON
LED -DC ON

D950: LED LINE (V25) ON


LED 350 Vcc (V30) ON

D970: all LED’s ON

c) Preparation
All LED’s as described under b) and also V22 on D915.

d) Exposure
All LED’s as described under b) and also V22 and V23 as well as radiation indicator V34
on D915.
In case of an error, V24 on D915 glow (red).

POLYMOBIL SPR8- Page 8 Siemens


Rev. 01. of 8 AG Medical
Troubleshootin 3-
Error messages
NOTICE Error messages Err 1, and initialization errors Err 90, Err 96 and
Err 97 block the generator so that the system must be switched
off and switched back on again.
"StandBy" error messages disappear when the error is cleared.
Error messages which occur during an exposure will continue
flashing until another key is activated.

a) During initialization

Err 90 EPROM Checksum


Err 96 KVnom error
Err 97 mAnom error

b)During stand by

Err 1 Capacity charge error Charging circuit error


Err 2 +15 VError
Err 3 Vfil < Vstby
Err 4 Vfil > Vstby
Err 5 KVact  0
Err 6 mAact  0
Err 7 Rot  0
Err 8 Braking failure Error in the rotating anode circuit
Err 9 Pressure switch Oil pressure switch in the single tank
is active
Err 30 Fast capacitor charge Capacitor bank is charging too fast
Err 31 Slow capacitor charge Capacitor bank is charging too slow
Err 33 Inverter Error
Err 34 Filament circuit Inverter Short-Cir-
cuit

c)During exposure

Err 10 Rot  0 No rotating anode acceleration


Err 11 Short circuit in main inverter Short circuit in the main inverter
Err 12 KVmax
Err 13 Imax
Err 14 KVact < KVnom
Err 15 mAact < mAnom
Err 17 Maximum exposure time Max. exposure reached
Err 18 Exposure not completed Exposure time not completed(expo-
sure release switch released)
Err 20 Capacitor not ready Capacitor bank not completely
charged
Err 21 Vfil > Vmax during prep Maximum filament heating exceeded
during preparation
Err 22 Maximum prep time Max. preparation time exceeded

Siemens AG SPR8- Page 1 POLYMOBIL


Medical Rev. 01. of 4
3- Troubleshootin
Troubleshooting after error messages
Err 1 Capacitor charge error
- Has the system not been used for a longer period of
time (higher capacitor leakage current)?
If the error occurs after switching the system off and
back on again, configure the capacitor bank
(refer to the "Service program" chapter, pages 4-2).
- Are the discharge resistors permanently connected?
- Has the temperature sensor on D950
responded (D950 TP ’TP’)?
- Has overvoltage in the C-bank been detected? (D950 TP ’VC’)
- Is there a short on D950? The error appears only when the short circuit
is continuous.
Replace D950.
Err 2 + 15 V - Error
- Check the fuse on power supply U1.
- Is power supply U1 defective?
Err 3 Vfil < Vstby (none or insufficient filament current)
- Do ’V1’ (-DC) and ’V2’ (+DC) flash on D925? Otherwise check fuse F6.
- No filament frequency? Check D915 .
- Connection D925 =>X-ray tube unit (X5).
- ConnectionD925 => D915 (X11).
- Check X-ray tube assembly.
- D925 defective.
Err 4 Vfil > Vstby (filament current is too high)
- Do ’V1’ (-DC) and ’V2’ (+DC) flash on D925? Otherwise check fuse F6.
- Filament frequency too high?
- A/D converter on D915 defective? => Replace D915.
- Short-circuit at the output of the filament inverter =>check connection
to/in the X-ray tube assembly.

Err 5 KVact  0
- No connection to the single tank generator (D915 X8)?
- Plug-in connection on D900 in the single tank generator ok?
Err 6 mAact  0
- Error in the measurement acquisition on D915? => Replace D915 .
- No connection to the single tank generator (D915 X8)?
- Plug-in connection on D900 in the single tank generator ok?
Err 7 Rot  0 (rotating anode speed  0)
- Error on D915 or D925.

POLYMOBIL SPR8- Page 2 Siemens


Rev. 01. of 4 AG Medical
Troubleshootin 3-

Err 8 Braking failure (error when braking the rotating anode - a current was
measured in the auxiliary capacitor C2 during the braking.)
- D915TP ’I1’. No current must be measured here when braking.
- D915 TP ’I2’. Braking current must be measured here.
- TRIAC ’V7’ or ’V8’ defective? => Replace D925.
Err 9 Pressure switch (oil pressure switch in the single tank generator is
active)
- Tube is overheated?
- Measure D915 plug X8 at pin 8. In case of malfunction, 0V are measured.
Err 10 Rot = 0 (rotating anode speed is 0 on exposure)
The error is only displayed as long preparation is depressed.
- LED ’AR’ on D925 flashes ; is fuse F5 ok?
- Measure at TP ’I1’ and ’I2’. Current has to be present during acceleration.
- Check D925.
- Is the power factor correction capacitor C2 defective?
- Connection to/in the single tank generator ok?
Err 11 Short circuit in main inverter
- Check D960 .
- Check D915 .
- Main inverter frequency too high?
Err 12 KVmax
- Interruption of KVact measurement. Measure TP ’KVN’ and ’KVP’ on
D915. Both signals must have the same amplitude.
- ’KVN’ or ’KVP’ > 70 kV.
- Error in kV measurement circuit?
- Error in kV control circuit?
Err 13 Imax
The tube or filament current is recorded.
- Check the filament circuit.
- Exposure filament basic value incorrectly set (P5 on D915)?
- Tube arcing?
If kVact < kVnom, Err 14 or Err 13 is detected and displayed. Refer also to the
troubleshooting information under Err 14.
Err 14 KVact < KVnom
- Voltage in the capacitor bank too low (D950 TP ’VC’)?
- Defective capacitors /fuses on the capacitor bank (D970)?
- No KVact?
- Error in the measurement circuit on D915?
- Main inverter frequency too low?
- Defect in the main inverter D960?
Err 15 mAACT  mANOM
- Check the filament circuit.

Siemens AG SPR8- Page 3 POLYMOBIL


Medical Rev. 01. of 4
3- Troubleshootin

Err 17 Maximum exposure time (Max. exposure time reached)


- Filament current too low?
- mAs counter is set incorrectly (P1 on D915)?
Err 18 Exposure not completed (exposure not completed by the system - expo-
sure switch was released)
- Operator error?
- Intermittent contact in the release switch?
- Error in the exposure release circuit?
Err 20 Capacitor bank not ready
Device was not ready yet / had not released yet.
- F1 on D950 defect?
- Error on D950?
Err 21 Vfil > Vmax (Max. Filament heating exceeded during preparation)
- Error during the acquisition of the filament heating?
- Basic value of the exposure filament incorrectly set (P5 on D915)?
- Error on D915?
Err 22 Maximum preparation (preparation time too long)
- Prep level pressed too long.
Err 30 Fast capacitor charge (capacitor bank charges too fast)
- Are all fuses on the capacitor bank (D970) ok?
- Are there defective capacitors?
- Is the line voltage too high (> 250 V)?
- Is the charging frequency too high?
- Configure the capacitor bank.
- D950 defect?
Err 31 Slow capacitor charge (capacitor bank charges too slow)
- ’V25’ (line) on D950 flashing? Check fuse F1
- None or insufficient charging frequency?
- Check D950 (’V22 - V24’ defective?)
- Poor connections to / in the capacitor bank?
Err 33 Inverter Error
- Check ribbon cable X20.
- FMAX not adusted?
- IGBT short - circuited?
Err 34 Filament Inverter short - circuit
- FMAX not adjusted?
- IGBT short - circuited?
Err 90 EPROM checksum
- EPROM defective.
Err 96 kVnom failure (hardware error , refer also to Err 5)
Err 97 mAnom failure (hardware error, refer also to Err 6)

POLYMOBIL SPR8- Page 4 Siemens


Rev. 01. of 4 AG Medical
Service 4-

2
SW2

X12
D 915 P6 P7
V22 V23 V24 V25

X15
P5
P4 ST
"OPEN"
P3 P2
P1 Service

X11 X10 X20


1

1 1

GNDJR REG CAL KVP KV KVN IH JRS KVS F1 II1 I2 TRG JIM SCHGND VC BAT KVM SCM VR

Fig. 1

To bypass the service mode, the ST slot (1/Fig.1, below the display on D915) has to be
jumpered before switching on the POLYMOBIL Plus.
As soon as "Pr. 1" appears in the display, bring the jumper back to the "OPEN" position.

KV +/- keys toggle between the programs. When you are


already in a program, the next or previous program
step is selected.
Setting of the kV values in "Pr. 5" and "Pr. 6".

mAs +/- Setting of the mAs values in "Pr. 5" and "Pr. 6".

Lamp switch Activating or quitting the current program

To quit the service mode, switch the POLYMOBIL Plus off/on or press the Reset button
SW 2 (2/Fig.1).

Siemens AG SPR8- Page 1 of POLYMOBIL


Medical Rev. 01. 2
4-2 Service programs

Service programs available


Pr. 1 Configures the capacitor bank.
After starting the service program,"CAP 150" appears on the display.
The voltage of the capacitor increases in increments from 150 V to 250
V, 300 V, 325 V, 337 V up to 345 V.
The program remains for about 6 minutes at the 150 V to 337 V level.
The system remains at the 345 V level for one hour, after which
the program automatically ends; "CAP END" appears on the display.
Pr. 2 Reads out the exposure counter.
Pr. 3 Error memory with up to 20 errors.The left display indicates how many
errors there are.The error message is displayed in the right display.
The- latest error is displayed first.
Pr. 4 Deletes error memory.
Press mAs + key for approximately 4 seconds.
Pr. 5 Sets the default exposure parameters (according to customer request).
Pr. 6 Sets the maximum possible KV and mAs values(according to customer
request.
Pr. 7 Adjusts FMAX on the main inverter.
"ADJ F" appears on the display after starting the service program.
Press S27.
"F on" appears on the display.
FMAX can be measured at TP "REG" on D 915, and can be set with the
potentiometer P4.
Pr. 8 Tests the anode starter unit.
Press S 27.
Pr. 9 Tests the filament unit.
"FIL 3" appears on the display after starting the service program.
Press S27.
Filament boosts; "FIL 5" appears on the display.

There is no radiation present during these service programs.


NOTICE

POLYMOBIL SPR8- Page 2 of Siemens


Rev. 01. 2 AG Medical
Measurement 5-
Line voltage
The line voltage must be set between 110 V ( + 10%) or 230 V ( + 10%).
The line voltage is adjusted automatically.

Measuring KVact and Itube


 Connect the oscilloscope to TP "JR", "KV" and "GND" on D915.

CH1 TP "JR" 1.0 V/Div


CH2 TP "KV" 1.0 V/Div
Ground at "GND"
10 ms/Div
Trigger external at TP "TRG"

 Set 90 kV and 10 mAs at the POLYMOBIL Plus and release an


exposure. The oscillogram should correspond to the one shown in Fig.
1.

Fig. 1

Siemens AG SPR8- Page 1 of POLYMOBIL


Medical Rev. 01. 20
5- Measurement
Capacitor charging circuit
Signals are only present during charging of the capacitors.
NOTICE

Charging frequency
 Connect the oscilloscope to D 950 at TP "FR1" and "GND".

CH1 TP "FR1" 5.0 V/Div


Ground at "GND"
20 µs/Div

 After switch on the following oscillogram can be measured:

Fig. 2

The capacitor charging frequency is 25 kHz and cannot be adjusted.

POLYMOBIL SPR8- Page 2 of Siemens


Rev. 01. 20 AG Medical
Measurement 5-
Capacitor charging current
 Connect the oscilloscope to D 950 at TP "IC" and "GND".

CH1 TP "IC" 500 mV/Div


Ground at "GND"
20 µs/Div

 After switch on the following oscillogram can be measured:

Fig. 3

The charging current is dependent on the line voltage present and the charge sta-
tus(capacitor voltage Uc) of the capacitor bank.
Fig. 3 shows the charging current in standby, i. e. when the C-bank is charged.

Siemens AG SPR8- Page 3 of POLYMOBIL


Medical Rev. 01. 20
5- Measurement
Capacitor voltage (Uc) and charging time
In exposure-ready status, capacitor voltage is approximately 350 V and can be measured
on D 950 TP "VC".
If the device is switched off because of overvoltage, the signal "ERR" is output on D950.

You can measure the capacitor voltage at TP "VC" only if


CAUTION
POLYMOBIL Plus is switched on.

The charging time depends upon the residual charge of the capacitor bank and on the
available line voltage, e.g. with 230 V the max. charging time is about 15 s.
In the worst case, ( line voltage <100 V and residual charge = 0V), charging time
can take 40 seconds.
For the following measurement, the line voltage was 230 V and the residual charge 30 V.

D915 test point "VC" (1 V 100 V)

Fig. 4

POLYMOBIL SPR8- Page 4 of Siemens


Rev. 01. 20 AG Medical
Measurement 5-
Rotating anode starter
Acceleration
Fig. 5 shows the stator currents during rotating anode acceleration (230 V).

CH2 D915 test point "I1" (1 V 0.25 A)


CH1 D915 test point "I2" (1 V 0.25 A)
10 ms /Div

Fig. 5

The currents are correspondingly lower at 100V line voltage.


NOTICE

Siemens AG SPR8- Page 5 of POLYMOBIL


Medical Rev. 01. 20
5- Measurement
Brakes
Fig. 6 shows the stator currents during braking (230 V)

CH2 D915 test point "I1" (1 V 0.25 A)


CH1 D915 test point "I2" (1 V 0.25 A)
10 ms /Div

Fig. 6

The currents are correspondingly lower at 100V line voltage.


NOTICE

POLYMOBIL SPR8- Page 6 of Siemens


Rev. 01. 20 AG Medical
Measurement 5-
kV Control
Checking the maximum main inverter frequency

NOTICE It is generally not necessary to set the main inverter frequency in


the field. When ordered as a replacement, D915 is preset at the
factory.

 Connect the oscilloscope to D915 at TP "REG" and "GND".


 Measure FMAX in the service program Pr. 7
(Refer to "Service program", section on "Service programs available")

5 V/Div.
20 s/Div.
0

FMAX = 15.5 kHz


TMIN = 64.5 s Tolerance: -0 s /+2s

Fig. 7

 Setting the maximum main inverter frequency with potentiometer


P4 on D 915.

Siemens AG SPR8- Page 7 of POLYMOBIL


Medical Rev. 01. 20
5- Measurement
Measuring the oscillating current
WARNING Prior to beginning the following instructions, you must discharge
the capacitor bank, because life-threatening DC voltage is
present. Check the actual charge with the DVM.

 Switch OFF POLYMOBIL Plus and wait until the capacitor voltage has dropped
to a value < 10 V.
 Disconnect the cable from the inverter to capacitor C1 and push the cable through
the current transformer 50 A : 50 mA (with a parallel resistor of 10 ohms) and
connect the cable again.
 Connect the oscilloscope to the current transformer.
 Switch ON POLYMOBIL Plus .

 Set 60 KV, 10 mAs on POLYMOBIL Plus and release exposure.


The diagram must correspond to Fig 8.

CAUTION The maximum main inverter frequency can only be measured


in the ascending part of the high voltage, i.e. in the first
microsec- onds, because afterwards the frequency is lower.

CH2: 1 V/Div, Trigger: Test point "TRG" external on D 915


20 µs /Div

Fig. 8

POLYMOBIL SPR8- Page 8 of Siemens


Rev. 01. 20 AG Medical
Measurement 5-
Measuring the kVact and kVnom
 Connect the oscilloscope to D915 at TP "KV", "KVS" and "GND".
 Set 60 KV, 10 mAs on POLYMOBIL Plus and release exposure.
 The oscillogram must appear like the one in Fig. 9.
 The KVnom value can be set with P7 on D915.

Fig. 9

kVact: CH1 D915 test point "KV" 1V 20 KV


kVnom: CH2 D915 test point "KVS" 1V 30 KV
20 ms /Div

Siemens AG SPR8- Page 9 of POLYMOBIL


Medical Rev. 01. 20
5 - 10 Measurement
Error in the kVact acquisition
Fig. 10 shows that one kVact cable from the single tank generator to the D915 is
damaged or disconnected.
The monitor (KVM) responds (curve 3) because KVN and KVP have different amplitudes;
the exposure is terminated with ERR 12.
Refer to function description "KV monitoring".

Fig. 10

Exposure data: 102 KV 10 mAs


KVact: CH1: D915 test point "KV"
Itube: CH2: D915 test point "JR"
CH3: D915 test point "KVM"
1 ms /Div

POLYMOBIL SPR8- Page 10 of Siemens


Rev. 01. 20 AG Medical
Measurement 5 - 11
Testing the filament circuit
Filament circuit in standby
 Connect the oscilloscope to D915 at TP "CAL", "I" and "GND".

Fig. 11

Filament inverter frequency CH1: D915 test point "CAL"


Filament current CH2: D915 test point "I"

Siemens AG SPR8- Page 11 of POLYMOBIL


Medical Rev. 01. 20
5 - 12 Measurement
Setting the maximum filament inverter frequency
The maximum filament inverter freqency is 25 kHz and can be set with P3 on D915.
Proceed as follows:
 Switch POLYMOBIL Plus OFF.
 Remove fuse F6 on D925.
 Connect the oscilloscope to D915 at TP "CAL" and "GND".
 Switch POLYMOBIL Plus ON (Err 03 is blinking).
The oscillogram must appear like the one in Fig. 12.

CH1: D915 test point "CAL"


20 µs /Div

Fig. 12

 Setting the maximum filament inverter frequency with potentiometer P3 on D915.

POLYMOBIL SPR8- Page 12 of Siemens


Rev. 01. 20 AG Medical
Measurement 5 - 13
Measuring the filament nominal value
The filament nominal value is set at the factory and should not be changed.
The filament nominal value can be measured on D915 at TP "JRS" and "GND" with the
DVM and is 1.5 V  4% in standby.

NOTICE If this value is changed, the nominal values for preparation and
exposure are also affected.

Setting the tube current for the exposure


The tube current must be reset for each tube.
Proceed as follows:
 Connect the oscilloscope to D915 at TP "JR", "IH" and "GND".

 Set 60 kV, 10 mAs on POLYMOBIL Plus and release an


exposure. The oscillogram must appear like the one in Fig. 13.

Fig. 13

Actual tube current valueCH1: D915 test point "JR"


Actual fil. current value CH3: D915 test point "IH"

If this is not the case, adjust P5 on D915 until there is minimal current overshoot at the
beginning of the exposure.
The standby heating of the exposure is set via P5 IH , IH ).
( Fig. 13 shows the correct setting.
In Fig. 14 the standby heating is set too high.
In Fig. 15 the standby heating is set too low.

Siemens AG SPR8- Page 13 of POLYMOBIL


Medical Rev. 01. 20
5 - 14 Measurement

Tek Stop: Single Seq 2.50kS/s

3
Ch1 1.0 V M20.0ms Ch1 1.56V
Ch3 500mV
Standby heating too high

Fig. 14

Standby heating too low


Fig. 15

POLYMOBIL SPR8- Page 14 of Siemens


Rev. 01. 20 AG Medical
Measurement 5 - 15
Testing and setting the mAs counter
 Unscrew the lid of the single tank cover.
 Pull out the mAs jumper on D900 and connect the mAs meter.

 Set 70 kV, 80 mAs on POLYMOBIL Plus and release exposure.


If the value displayed on the mAs meter does not coincide with the value set on
POLYMOBIL Plus, adjust the mAs counter with P1 on D915.

Release another exposure and compare the display.

WARNING Observe the cooling intervals for the tube.

Siemens AG SPR8- Page 15 of POLYMOBIL


Medical Rev. 01. 20
5 - 16 Measurement

Length

Marker

Side marker

Fig. 16

Coincidence of light and radiation fields


Procedure:
 Insert film into the 24 cm x 30 cm or 10” x 12” cassette and place it on a table or
similar surface.
 Adjust the vertical SID of 100 cm or 40” to the top edge of the cassette
using a measuring tape.
 Use the knobs to set a format of 18 cm x 24 cm or 8” x 10”.
 Switch on the light localizer and align the cassette.
 Apply radiopaque markers (e.g. washers, coins) as shown in Fig.
16 . Use one washer as a side marker.
 Release an exposure (60 kV, 10 mAs) and develop the film.
 Log the following data on the developed film using a water-resistant felt-tip pen.
- SID setting
- Film size
- Radiation field size

POLYMOBIL SPR8- Page 16 of Siemens


Rev. 01. 20 AG Medical
Measurement 5 - 17
Evaluation:
 Measure the deviations (X1, X2, Y1, Y2) between the edges of the light field and
the radiation field on all four sides, corresponding to Fig. 17.
 Calculate the total deviations in the X and Y directions (disregarding the algebraic sign).

Y1
Film

Light field

X1
X2

Radiation field

Y2

Fig. 17

 The deviations in length (total Y) as well as the deviations in width (total X) must
be smaller than 1.6 cm respectively.

Siemens AG SPR8- Page 17 of POLYMOBIL


Medical Rev. 01. 20
5 - 18 Measurement

Fig. 18 Fig. 19

 If the deviation is larger, loosen the 4 Allen screws a little ( arrow /Fig.18/19) and shift
the collimator accordingly.
After adjusting the collimator, tighten the screws again (arrow / Fig.18/19).
 Repeat the test and readjust the collimator until the light and radiation fields are
within the acceptable tolerance (< 1.6 cm).

POLYMOBIL SPR8- Page 18 of Siemens


Rev. 01. 20 AG Medical
Measurement 5 - 19

1
2

Fig. 20 Fig. 21

Testing and readjusting the counterbalance


 Loosen the rotating knob (2/Fig.20) and unlock the C-arm (1/Fig.20).
It should be possible to move the C-arm without any attached accessories easily across
the whole range and the arm should come to a stop in any desired position.

Readjusting the counterbalance with accessories attached


If the C-arm with accessories attached does not come to a stop in any desired position,
readjust the spring tension for the counterbalance.
 Remove the cover (3/Fig.20) at the stand.
 Release the supplemental brake at the C-arm (2/Fig.20) completely.
 Move the C-arm into a vertical position.
Siemens AG SPR8- Page 19 of POLYMOBIL
Medical Rev. 01. 20
5 - 20 Measurement
 Tighten the screw using a 17 mm Allen key (4/Fig.21).
Adjust the spring tension in such a way, that there is a balance when the C-arm is in a
vertical position, i.e. that equal force is required to lift and lower the tube assembly.

POLYMOBIL SPR8- Page 20 of Siemens


Rev. 01. 20 AG Medical
Measurement 5 - 21
This page intentionally left blank.

Siemens AG SPR8- Page 21 of POLYMOBIL


Medical Rev. 01. 20
Replacing important 6-

NOTICE To repair paint chips, the POLYMOBIL Plus colors


can be ordered in spray cans:

"white" spray paint 84 27 734 RE999


"ergo gray" spray paint 84 27 742 RE999
"medical blue" spray paint (in preperation)

Fig. 1 Fig. 2

Replacing the single tank generator


To replace the single tank generator, remove it from the stand as described in the follow-
ing:
 Remove the collimator (refer to “Replacing the collimator”).
 Move the arm into the lowest position and check whether the safety mechanism (bolt
1 / Fig. 1) is locked into position.
Now the stand cannot be moved upwards.
 Remove the four Allen screws of the single tank generator cover (2/Fig.2), remove
the cover, disconnect the connections and the strain relief of the corrugated tubing
(3/Fig.2) and remove it.

Siemens AG SPR8- Page 1 POLYMOBIL


Medical Rev. 01. of 8
6- Replacing important

Fig. 3

 Take out one of the four mounting screws and insert it into the threaded hole and
tighten it (arrow/Fig.3).
WARNING This is a safety device to prevent the arm of the unit from flying
upward if the lever for the transport lock is inadvertantly
activated.
 Place a pad under the single tank generator.
 Disconnect the single tank generator together with the bracket from the stand
by removing the three remaining screws (4/Fig.3).
 Lower the single tank generator to the pad.
 The single tank generator is installed in reverse sequence.
- Secure the three mounting screws (4/Fig.3) with Loctite 242 and tighten them.
- Remove the fourth mounting screw from the threaded hole (arrow/Fig.3) and
reinstall it in its former position. Secure the screw with Loctite 242.
- Test the torque of the mounting screws
; Nom: 25 Nm, Tolerance  10%.
 Reinstall the collimator in reverse sequence.
 Set the tube current (refer to "Setting the tube current for the exposure", pp 5-12).
 Check the coincidence of the light and radiation fields and adjust, if
necessary (refer to "Coincidence of light and radiation fields", pp 5-15)

POLYMOBIL SPR8- Page 2 Siemens


Rev. 01. of 8 AG Medical
Replacing important 6-

Fig. 4 Fig. 5

Replacing the collimator


In case of damage, the collimator has to be completely replaced.
Proceed as follows:
 Remove the two screws ( 5/Fig.5) and remove the back panel.
 Mark the 5 connecting cables of the collimator and disconnect them.
 Remove two screws on each side (6/Fig.5) of the upper cover.
 Turn the two knobs to the end position and remove them.
 Remove the front panel after you have removed the two screws (7/Fig. 6).
Caution : The front panel is connected to the collimator via a cable (9/Fig. 7).
 Remove the collimator after removing the four Allen screws (arrow /Fig.4/5).

Siemens AG SPR8- Page 3 POLYMOBIL


CAUTION
Medical Rev. 01. of 8
6- Replacing important
Hold the collimator securely.
Lift the upper cover and pull the cable through the housing.
 Attach the new collimator with the four screws (arrow /Fig.4/5) and center it.
 Reconnect the cables and the back panel.
 Check the coincidence of the light and radiation fields and adjust, if
required. (Refer to “Coincidence of light and radiation field”, page 5-15.)

POLYMOBIL SPR8- Page 4 Siemens


Rev. 01. of 8 AG Medical
Replacing important 6-

9 10
7

11

Fig. 6 Fig. 7

Replacing the lamp for the localizer


 The replacement light localizer lamp is located below the cassette module.
 Turn both knobs (8/Fig.6) to the end position and remove them.
 Remove the front panel after removing both screws (7/Fig.6) .
Caution: The front panel is connected to the collimator via a cable (9/Fig.7).
 Remove both screws (10/Fig.7) and take off the metal bracket for the
temperature switch).
 Take out the defective lamp (11/Fig. 7) and replace it with a new lamp.

CAUTION Do not touch the glass envelope with your fingers.

 Attach the bracket and the front panel.


 Adjust both knobs and attach them.
 Check the function of the light localizer lamp.
 Check the considence of light and radiation fields (see "Considence of light
and radiation fields" in chapter 5).

Siemens AG SPR8- Page 5 POLYMOBIL


Medical Rev. 01. of 8
6- Replacing important

Fig. 8

Replacing the steering caster


Step:
1 Loosen only Allen screws on the left and right (2.5 mm wrench.
2 Pull back hexagonal shaft
3 Remove pedal
4 Remove Allen screws (wrench size 6 mm).
5 Pull out the caster.
6 Align the threaded holes on the new caster.

 Install in the reverse sequence.


 Test the pedal positions (Refer to Operating Instructions).

POLYMOBIL SPR8- Page 6 Siemens


Rev. 01. of 8 AG Medical
Replacing important 6-

Fig. 9

Replacing the capacitor bank


WARNING
When working on the open system, there is the danger of
Electric shock !
 The capacitor bank can still be charged.
Do not attempt to work on the system while this condition exists.
 For details see chapter "Prerequisites".

Siemens AG SPR8- Page 7 POLYMOBIL


Medical Rev. 01. of 8
6- Replacing important

1
2

Fig. 10 Fig. 11

 Switch the unit off; disconnect the line voltage plug;


 Remove the 2 cover screws on each side (1/Fig.10) and remove the cover(with
cassette drawer).
 Measure the actual voltage present in the capacitor bank with the DVM at test point -
VCC and +VCC on the inverter board D960 or (more accessible) on the capacitor
board D970 at the + connecting point of the C3 capacitor and on the right-hand side
of fuse F3 (see Fig. 11).

WARNING Continue work only if the capacitor unit is discharged (U  10 V)

 Remove one cover screw on each side (2/Fig.10).


 Unscrew the three Allen screws (2/Fig.11).
 Pull out the connectors X3 and X9 (3/Fig.11) on the charging board D950.
 Remove flat ribbon cable X20 (4/Fig.11) from the inverter board D960 and from
the charging board D950.
 Disconnect and remove the cables of oscillating circuit capacitor C1 and
oscillating circuit coil L1 to inverter board D960.
 Pull the entire capacitor unit out towards the front.

WARNING If a fuse on D970 is defective, high voltage may still be present at


the affected capacitor.
In this case, discharge the affected capacitor with the
discharging resistor R5 (5/Fig. 11) .

POLYMOBIL SPR8- Page 8 Siemens


Rev. 01. of 8 AG Medical
Replacing important 6-

 Assembly is performed in the reverse sequence.


 Check the capacitor voltage (Uc) and the charging time (refer to "Capacitor voltage
(Uc) and charging time ", page 5-3)

Siemens AG SPR8- Page 9 POLYMOBIL


Medical Rev. 01. of 8
6Changes
- to previous version
Replacing important 7-1
Page 5-1, 5-4, 5-5 and 5-6:
"240 V" changed into "230 V".

SMS Iselin / O’Donnell


TD SD 24 / Am-Ende
TD SD 24 / Groß

POLYMOBIL SPR8- Page 2 of Siemens


Rev. 01. 2 AG Medical
7-2 Changes to previous version
This page intentionally left blank.

Siemens AG SPR8- Page 1 of POLYMOBIL


Medical Rev. 01. 2