COULOMETRIC Methods of Analysis
COULOMETRIC Methods of Analysis
Methods of
Analysis
Electrolysis
Fundamentals of Electrolysis
Anode: Cathode:
H2O = ½ O2 (g) + 2H+ + 2e- Cu2+ + 2e- = Cu (s)
Fundamentals of Electrolysis
OP = overpotential
IR = ohmic potential
CP = concentration polarization
Fundamentals of Electrolysis
Ecell = Ecathode – Eanode – OP – IR – CP
1) Overpotential (OP)
• Eohmic = V = IR
Fundamentals of Electrolysis
Ecell = Ecathode – Eanode – OP – IR – CP
Q=Ixt
• analyte’s concentration
and current (I) decrease
over time (t)
Fundamental Equations
• Faraday’s law
Q = (nA) (nF) ; Q = It
moles reacted, nA = I x t
nF
, n = number of electrons per molecule
, F = Faraday constant, 96,485 coulombs (1 faraday)
is equal to one mole of electron
Example
• A sodium sulfate solution was electrolyzed using inert Pt
electrodes. The cathode reaction was 2H2O = H2 + 2OH-.
If a current of 3.0 amp was used for 30 min, what weight
(g) of H2O (MM = 18.0) gas would be produced?
moles reacted, n = Ixt
Q = (nA) (nF) ; Q = It ; A
nF
moles reacted, nA = I x t
nF
• is conducted in a three-
electrode cell
Q = amperes x seconds
Principle ● Coulometry
ü Michael Faraday reported that the
mass, m, of elements deposited
at an electrode in g is directly
proportional to the charge, Q, in
coulombs
maQ
Q = It
• Exhaustive means:
1) the analyte is completely oxidized or reduced at
the working electrode, or
2) that it reacts completely with a reagent
generated at the working electrode
Coulometry
• Example:
H+ and OH- can be produced electrochemically
2 H2O + 2e- = H2 + 2 OH-
2 H2O = O2 + 4 H+ + 4 e-
Coulometry
• Based on the stoichiometry associated with the conversion of
species from one oxidation state to another:
1 CONTROLLED-POTENTIAL COULOMETRY
• Is more selective than constant-current
coulometry, but slower
•The quantity of
electricity is measured
with an electronic
integrator
• Integrating the area under the curve
from t = 0 to t = te gives us the total
charge
(te is electrolysis time)
Selecting a Constant Potential
Cu2+ (aq) + 2e- = Cu (s) E0 = +0.324 V
• The potential is selected so that the desired
oxidation or reduction reaction goes to
completion without interference from redox
reactions involving other components or the
sample matrix
• The reaction is favored when the working
electrode’s potential is more negative than
+0.324 V
• To maintain a 100% current efficiency, the
potential must be selected so that the
reduction of H+ to H2 does not contribute
significantly to the total charge passed at the
electrode
1 Potentiostatic Coulometry: Instrumentation
• consists of potentiostat, electrolysis cell, chemical
coulometer for determining the number of coulombs and
placed in series with the working electrode
1 Potentiostatic Coulometry: Instrumentation
(1) CELL ASSEMBLY
• Ag/AgCl electrode
1 Potentiostatic Coulometry: Instrumentation
(2) POTENTIOSTAT
• maintains the potential if the working electrode at a constant
level relative to a reference electrode
• Amperostatic coulometry or
Coulometric titration
• controlled-current
coulometry normally is
carried our using a two-
electrode galvanostat
• Consisting of a
- Usually Pt wire
Coulometric
Titration
5 Sources of Errors: Coulometric Titration
§ REDOX Titrations
e.g. coulometric titration of ascorbic acid using
electrogenerated iodine: 2I- ® I2 + 2e-
I2 + ascorbic acid ® dehydroascorbic acid + HI
Summary of Coulometric Titrations (Neutralization,
Precipitation and Complex formation Reactions
Ca2+, Cu2+, Zn2+, HgNH3Y2- + NH4+ 2e- = HY3- + Ca2+ = CaY2- + H+,
Pb2+ Hg(l) + 2NH3 + HY3- etc.
Summary of Coulometric Titrations Involving REDOX Reactions
Q = It ; nA = Q / nF ; (nF) (nA) = I te
, nA = number of moles of analyte
, n = number of moles of electrons
transferred in the half-cell reaction
, F = 96,487 C mol-
` , te = electrolysis time
nF (g Na2S2O3 / MM Na2S2O3 ) = I te
Sample Calculation
1) The purity of a sample of Na2S2O3 (158.1 g/mol) was determined by a coulometric redox
titration using I- as a mediator, and I3- as the "titrant“. A sample weighing 0.1342 g is
transferred to a 100-mL volumetric flask and diluted to volume with distilled water. A 10.00-
mL portion is transferred to an electrochemical cell along with 25 mL of 1 M KI, 75 mL of a
pH 7.0 phosphate buffer, and several drops of a starch indicator solution. Electrolysis at a
constant current of 36.45 mA required 221.8 s to reach the starch indicator end point.
Determine the purity of the sample.
Ixt
Q = (nA) (nF) ; Q = It ; moles reacted, nA =
nF
2S2O32- (aq) + I3- (aq) = S4O62- (aq) + 3I- (aq)
• oxidizing S2O32- to S4O62- requires one electron per S2O32- (n=1)
grams Na2S2O3
1 1 mol x 158.1 g
= (0.03645 A) (221.8 s) x x
1 e- 96,500 A s mol
weight Na2S2O3 (g) = 0.01324 g
Sample Calculation
1) The purity of a sample of Na2S2O3 (158.1 g/mol) was determined by a coulometric
redox titration using I- as a mediator, and I3- as the "titrant“. A sample weighing 0.1342 g
is transferred to a 100-mL volumetric flask and diluted to volume with distilled water. A
10.00-mL portion is transferred to an electrochemical cell along with 25 mL of 1 M KI, 75
mL of a pH 7.0 phosphate buffer, and several drops of a starch indicator solution.
Electrolysis at a constant current of 36.45 mA required 221.8 s to reach the starch
indicator end point. Determine the purity of the sample.
n = 4.01