VAL 2006 Validation Plan For Quantitative Method
VAL 2006 Validation Plan For Quantitative Method
It must be revised to reflect your lab’s specific processes and/or specific protocol
requirements.
(Insert Laboratory Header Here)
(Include full laboratory name and address)
Instrument/Method/Reagent
to be validated:
Primary Back-up
(if applicable)
Serial Number(s):
Analyte(s):
Kit Name:
1. Overview
a. This plan was written using “VAL 2001_Quantitative Validation Guidelines” as a
reference, please refer to this document if more details are needed.
b. All raw data reports will be saved in (insert location details)
c. The plan includes the following sections:
Precision
Accuracy
Linearity
Analytic Measurement Range (AMR) and Clinical Reportable Range (CRR)
Analytical Sensitivity and Specificity
Reference Ranges
Method Approval
(Insert/remove additional sections if needed-e.g. Carryover))
2. Precision
a. Precision is reproducibility - the agreement of the measurements of replicate runs of
the same sample. It is the process of determining the range of random error. The
precision is measured in terms of coefficient of variation (CV). Precision will be
tested only on measured analytes, and not calculated analytes.
Short-Term
Analyte Mfg Precision 25% of TEa
ALT (00.00%) (00.00%)
AST (00.00%) (00.00%)
Albumin (00.00%) (00.00%)
Long-Term
Analyte Mfg Precision 33% of TEa
ALT (00.00%) (00.00%)
AST (00.00%) (00.00%)
Albumin (00.00%) (00.00%)
3. Accuracy
a. Accuracy is the true value of a substance being measured. Verification of accuracy is
the process of determining that the test system is producing true, valid results.
Accuracy will be tested on measured analytes only.
b. The ideal number of samples is 40, however a minimum of 20 samples that cover the
reportable range of the method and include points near the medical decision points,
if possible is acceptable.
c. Accuracy will be determined by one of the options below:
i. Option A: A minimum of 20 samples, tested in duplicate. These will be
primarily patient samples but may include commercial proficiency testing or
control samples in order to provide material that covers the reportable range.
The samples will be tested on (insert name of comparison instrument) located
at (insert name and location of laboratory). The samples will be tested in
4. Linearity
a. A quantitative analytical method is said to be linear when measured results from a
series of sample solutions are directly proportional to the concentration or activity in
the test specimens. This means that a straight line can be used to characterize the
relationship between measured results and the concentrations or activity levels of an
analyte for some stated range of analyte values. Linearity will be tested on measured
analytes only.
b. Linearity verification will be determined using the (insert provider and product name)
samples.
i. At a minimum, samples will be run in duplicate.
ii. Known values of the standards will be plotted on the X-axis and the mean of
the measured values will be plotted on the Y-axis.
iii. Slope and intercept will be calculated.
iv. A predicted Y value will be calculated for each X value.
v. Predicted Y values will be plotted versus the corresponding known X values.
A straight line will be drawn to connect the predicted Y points on the graph.
vi. Measured Y values will be subtracted from the associated predicted Y value.
This difference is the systematic error due to non-linearity.
vii. Systematic error will be compared to 50% of the total error.
c. Acceptability criteria:
i. The method is linear if the difference between the predicted Y value and the
measured Y value is less than the allowable error for each specimen point.
ii. The systematic error must be less than 50% of the total error.
other pretreatment not part of the usual assay process. AMR validation is the
process of confirming that the assay system will correctly recover the concentration
or activity of the analyte over the AMR. Reportable range will be verified on
measured analytes only.
b. The Clincal Reportable Range (CRR) is the range of analyte values that a method
can report as a quantitative result, allowing for specimen dilution, concentration or
other pretreatment used to extend the direct AMR. The laboratory should establish a
CRR that covers a range inclusive of Grade 4 Adverse Events on the DAIDS Toxicity
Table
c. Reportable range will be determined using the (insert proficiency provider and survey
name) samples.
i. At a minimum, samples will be run in duplicate.
ii. It may be necessary to dilute the lowest sample to verify the low end of
Analytical Measurement Range (AMR).
iii. The high end of the AMR will only be as high as the highest sample.
iv. The Clinical Reportable Range (CRR) must extend the AMR in order to
include grade 4 events of the Division of AIDS Toxicity Table.
The lab will establish what dilutions are necessary to cover this range,
bearing in mind that a minimum amount of dilution is ideal since
accuracy decreases with increasing dilution.
The laboratory will decide the maximum value of dilution that will be
allowed. Any samples that do not give a numerical value beyond this
allowed dilution should be reported as greater than the upper end of
the CRR.
d. Acceptability criteria:
i. The reportable range must be within the manufacturer’s AMR.
ii. The manufacturer’s upper limit will be accepted if the known sample is within
percent TEa of the AMR upper limit.
iii. The manufacturer’s lower limit will be accepted if the known sample is within
the minimum detectable difference or percent TEa of the lower limit
(whichever is greater).
7. Reference Ranges
b. The reference range studies have been verified (or established) for the following
populations:
i. Adult reference ranges–describe the methods used to establish or verify
reference ranges for each analyte. Include information on how “normal”
subjects were screened, the total number of subjects included, and any other
pertinent information.
ii. Pediatric reference ranges– describe the methods used to establish or verify
reference ranges for each analyte. Include information on how “normal”
subjects were screened, the total number of subjects included, and any other
pertinent information. See example below:
“Reference ranges for pediatrics were adopted from Ugandan
reference ranges in use at the XYZ Research lab in Kampala,
Uganda. The ranges were evaluated by the ABC medical leadership
team in consultation with local pediatricians and determined to be
appropriate for the local Tanzania population. Adoption of these
ranges was approved by the ABC Medical Director. All ranges will be
verified and monitored over time as appropriate data becomes
available.”
c. Acceptability criteria:
i. Establishment: ranges will be determined using a non-parametric statistical
method to determine the 95% reference limits. For most analytes the lower
and upper reference limits are defined as the 2.5th and 97.5th percentiles,
respectively.
ii. Verification: ranges will be considered verified if 90% of values fall within the
proposed range
iii. Verification of pediatric ranges will be dependent on the ability to collect
sufficient pediatric samples in each age category. Additional time may be
required or fewer samples may be acceptable. The Medical Director will
have final approval on the acceptability of pediatric reference range
verification.
8. Method Approval- The final decision on method validation and acceptance is made after a
careful review of all the studies performed as part of the complete method validation
process. The Laboratory Director shall make the ultimate decision on method validation.
Method acceptance is based on the results from the above studies plus an evaluation of the
new method’s cost effectiveness, turn-around-time, laboratory staff training needs, and any
other relevant operational considerations.
Prepared By:
Date: