Um TQ GC Ms 7000e 7010c Series Concepts Guide g7006 90008 en Agilent
Um TQ GC Ms 7000e 7010c Series Concepts Guide g7006 90008 en Agilent
Concepts Guide
Notices
© Agilent Technologies, Inc. 2022 Warranty Safety Notices
No part of this manual may be repro- The material contained in this docu-
duced in any form or by any means ment is provided “as is,” and is subject
(including electronic storage and to being changed, without notice, in C AU T I O N
retrieval or translation into a foreign lan-
future editions. Further, to the maxi-
guage) without prior agreement and
mum extent permitted by applicable A CAUTION notice denotes a
written consent from Agilent Technolo-
law, Agilent disclaims all warranties,
gies, Inc. as governed by United States hazard. It calls attention to an
either express or implied, with regard
and international copyright laws. operating procedure, practice,
to this manual and any information
contained herein, including but not lim- or the like that, if not correctly
Manual Part Number
ited to the implied warranties of mer- performed or adhered to, could
G7006-90008 chantability and fitness for a particular result in damage to the product
purpose. Agilent shall not be liable for
Edition or loss of important data. Do
errors or for incidental or consequen-
tial damages in connection with the not proceed beyond a CAUTION
First Edition, July 2022
furnishing, use, or performance of this notice until the indicated
Printed in USA
document or of any information con- conditions are fully understood
Agilent Technologies, Inc. tained herein. Should Agilent and the and met.
5301 Stevens Creek Blvd. user have a separate written agree-
Santa Clara, CA 95051 ment with warranty terms covering the
material in this document that conflict
Microsoft® and Windows® are U.S. reg- with these terms, the warranty terms in
WAR N IN G
istered trademarks of Microsoft Corpo- the separate agreement shall control.
ration. A WARNING notice denotes a
Technology Licenses hazard. It calls attention to an
Software Revision operating procedure, practice,
The hardware and/or software described
This guide applies to the Agilent in this document are furnished under a or the like that, if not correctly
Mass-Hunter Workstation Software -- license and may be used or copied only performed or adhered to, could
Data Acquisition for 7000/7010 Series in accordance with the terms of such
Triple Quadrupole program version
result in personal injury or
license.
B.03.00 or higher until superseded. death. Do not proceed beyond a
Restricted Rights Legend WARNING notice until the
indicated conditions are fully
U.S. Government Restricted Rights. Soft-
ware and technical data rights granted to understood and met.
the federal government include only
those rights customarily provided to end
user customers. Agilent provides this
customary commercial license in Soft-
ware and technical data pursuant to FAR
12.211 (Technical Data) and 12.212
(Computer Software) and, for the Depart-
ment of Defense, DFARS 252.227-7015
(Technical Data - Commercial Items) and
DFARS 227.7202-3 (Rights in Commer-
cial Computer Software or Computer
Software Documentation).
In This Guide...
The Concepts Guide presents “The Big Picture” behind the operation of the
Agilent 7000/7010 Series Triple Quadrupole GC/MS Systems by helping you
understand how the hardware and software work.
1 “Overview”
Learn how 7000/7010 Series Triple Quads help you do your job.
1 Overview
System Description 8
Help for applications 8
Help for acquisition 9
Help for data analysis 11
System Description
An Agilent 7000/7010 Series Triple Quadrupole MS is a standalone triple
quadrupole mass spectrometer for use with the Agilent 8890 and 9000 gas
chromatographs. The 7000/7010 Series Triple Quads feature:
• One split flow turbomolecular vacuum pump
• Rotary vane or optional dry scroll foreline pump
• Independently MS-heated electron-ionization ion source
• Two independently MS-heated hyperbolic quadrupole mass filters
• Single hexapole collision cell
• High-energy dynode (HED) electron multiplier detector
• Independently GC-heated GC/MS interface
This configuration has advantages for many applications. The data is interpreted
through the use of the MassHunter Workstation software, which provides
quantitative and qualitative analyses of the data obtained.
The 7000/7010 Series Triple Quads are the only triple quadrupole GC/MS
combinations that incorporates a hexapole collision cell, blanketed with a
combination of nitrogen and helium gas, to improve the ion fragmentation prior
to final filtration, detection, and quantification.
Paired with Agilent 8890, 9000, or 7890 GCs, the 7000/7010 Series Triple Quads
deliver sensitive, reproducible analyses of target compounds in complex
matrices. This provides the following:
• Femtogram-level limits of detection and quantification
• Selective quantification of target compounds in high chemical background
samples
• Improved signal-to-noise ratios (S/N) in complex matrices
• Ability to meet stricter regulations regarding sample analytical limits for
certain applications
• Simplified operation with Agilent’s instrument control and data analysis
software
The 7000/7010 Series Triple Quads offer the high sensitivity in GC/MS/MS
analyses that is required by many commercial and regulatory applications.
Acquire data
• Enter sample information and pre- or post-programs (scripts) and run single
samples interactively
• Enter and automatically run both individual samples and samples organized
in a sequence of samples
• Set up pre- and post-scripts to run between samples in a sequence
• Set up and run a sequence to optimize MS acquisition parameters
• Print a sequence report
• View system events, including start and stop times, run events, and errors
• Print an event log report
To learn how to get started with the Agilent Triple Quadrupole GC/MS, see the
7000E/7010C Series Triple Quad GC/MS Quick Start Guide.
To learn more about how to use the Agilent Triple Quadrupole GC/MS with real
samples and data, see the 7000E/7010C Series Triple Quad GC/MS
Familiarization Guide.
To learn how to do individual tasks with the GC/MS, see the online help.
To learn more about your Agilent GC, see the Agilent user documentation for
your specific GC model.
This chapter explains the inner workings of the 7000/7010 Series Triple Quads.
The foundation for understanding the operation of a triple quadrupole mass
spectrometer is the operation of a single quadrupole mass spectrometer.
Therefore, an explanation of the workings of a single quadrupole mass
spectrometer is presented first.
The quadrupole mass analyzers consist of four parallel rods to which specific DC
(direct current) and RF (radio frequency) voltages are applied. These rods filter
out all ions except those of one or more particular mass-to-charge (m/z).
The RF is applied to all four rods, but the negative (–) rods are 180 degrees out of
phase with the positive (+) rods. The rods are labeled + and – in reference to the
DC voltages applied to them.
All ions that comprise the sample are generated at the source. However, when a
specific set of voltages is applied, only ions of the corresponding m/z value may
pass through the quadrupole to reach the detector. As the voltages are altered,
ions with other m/z values are allowed to pass through. A full MS scan is
obtained by increasing the DC and RF voltages applied to the four rods over an
expanded range of values.
Detector
In the model:
• All of the ions contained in a sample are formed in the external ionization
source and collected in a funnel. The balls of different colors and sizes
represent different ions having different m/z values.
• The quadrupole mass analyzer is represented by a moving belt that serves to
filter the ions as they pass through openings of various sizes. The ions pass
from the funnel, through the filter, to the detector.
• The detector is represented by the collecting funnel below the filtering belt.
As the belt (the analyzer) moves, or the voltages on the rods are changed, ions
with different m/z values are filtered through the mass spectrometer.
As the analyzer moves from a small m/z value to increasingly larger values, a full
MS scan is created.
If the belt does not move, the detector continues to monitor the same single m/z
value over the entire scan period. This type of analysis is known as selected ion
monitoring, or SIM. It is the most sensitive operating mode for a single
quadrupole mass spectrometer.
The scan period is selected (fixed) by the user. The user may set the dwell time
to scan a specific mass range (for example, m/z 50 to 1,000) or to remain on one
selected ion (SIM) or to move to several selected ions during the scan period.
The quadrupole mass filter is not scanned over a range in the SIM mode. The
required RF and DC voltages are set to filter a single mass for a specified time
before moving to the setting for the next SIM.
Detector
In this example:
1 All of the ions (+, –, and neutrals) are formed in the ionization source. The
Agilent ion sources consist of a series of lenses and a repeller assembly that
directs the ion beam into the analyzer.
2 Ion optics guide the ions to the quadrupole mass analyzer.
3 In the analyzer, only ions of a particular m/z value, represented in Figure 3 by
blue balls, are allowed to pass through to the detector.
Detector
The full scan MS mode is less sensitive because the duty cycle for each m/z
value is considerably less than 100 percent. The quadrupole mass analyzer
scans sequentially, passing each m/z value in the selected mass range to the
detector.
A full scan MS is still a useful mode of operation because it shows all of the ions
that are being formed in the ion source. This alerts analysts to other compounds
co-eluting with compounds of interest, and is helpful information for developing
SIM acquisitions.
Detector
The diagram shows that fragment ions, also known as product ions, are formed
by fragmenting or breaking apart precursor ions. Precursor ions formed in the
ion source travel through the mass analyzer without change, unless extra energy
is applied to their motion in a region where fragmentation can occur. If the ions
run into gas molecules and the translational energy is high enough, these
collisions convert the translational energy into molecular vibrations that can
cause the ions to fragment. This is called collision-induced dissociation (CID).
Fragmentation, or CID, can be carried out in a low-pressure region between the
ion source and the mass analyzer. The outlet of the ion source is under vacuum,
which is created by a two-stage vacuum pump. On the Agilent single quadrupole
mass spectrometer, the region between the ion source and the quadrupole
exhibits a gas pressure of about 10 to 20 × 10-5 Torr, which is well below
atmosphere pressure (760 Torr). Under normal operation, a voltage is applied
across this region to keep the ions passing through to continue on to the mass
analyzer. Even if these ions collide with the gas molecules in this region, they
usually do not have enough energy to fragment.
However, as the voltage is increased, the ions have more translational energy.
Then, if the ions run into gas molecules (Figure 6), CID can occur. Even though
this fragmentation does not occur where the ions are formed, this type of
fragmentation is still referenced as “in-source CID.”
As in the single quadrupole MS, the analyzers consist of four parallel hyperbolic
rods through which selected ions are filtered. After passing through the first
quad, the filtered ions then enter a collision cell where they are fragmented. The
collision cell is typically called the second quadrupole, but in this case,
geometrically it is actually a hexapole filled with a combination of nitrogen and
helium.
The product ions formed in the collision cell are then sent to the third quadrupole
for a second filtering stage to enable isolation and examination of multiple
precursor to product ion transitions.
The split flow high-performance turbo pump provides vacuum and efficiently
eliminates carrier gas and un-ionized material to allow for accurate product
quantification.
Once through the HED, the ions are quantitated by an electron multiplier detector.
Gain normalized tuning of this detector provides consistent sensitivity over the
life of the electron multiplier.
External
Ionization
Collision
Cell
Precursor
Quad
Filter Q1
Product
Quad Detector
Filter Q3
Representing the quadrupole mass analyzers as moving belts, a collision cell can
be placed between the belts to fragment the ions. The first belt can be fixed to
select which precursor ion travels to the collision cell. Different types of collision
cells can be used.
The cell can be another quadrupole, a hexapole (six rods like the one used in the
7000/7010 Series Triple Quads), an octopole (eight rods), or even a transverse
wave guide.
Whichever geometry is used, an inert, nonreactive gas is required for use as the
collision gas. In addition, the voltages applied to the collision cell must be
different from those applied to the quadrupoles to accelerate the movement of
all of the ions so that their collision energy is sufficient to cause fragmentation.
In the example shown in Figure 8 on page 22, a precursor ion is selected using
the first quadrupole and is sent to the collision cell for fragmentation. The
fragments are then scanned through the third quadrupole, resulting in a
product-ion scan MS/MS. Since the fragment ions are pieces of the precursor,
they represent portions of the overall structure of the precursor molecule. They
are selected by the third quadrupole according to their m/z ratio in the same way
the precursor ions are chosen by the first quadrupole. In this manner, a triple
quad provides enhanced selectivity for accurate target compound analysis.
Full scan MS/MS (first belt remains steady, second belt moves) using a triple
quadrupole MS is not the most sensitive mode for the same reason that a full
scan MS using a single quadrupole is not the most sensitive mode. The most
sensitive mode of operation for the triple quadrupole MS instrument is to fix both
belts and only monitor a specific precursor ion and a specific product ion. This
mode is called selected reaction monitoring or SRM.
In normal operation, a triple quadrupole MS instrument involves running multiple
SRMs for the same precursor ions. This is called multiple reaction monitoring or
MRM.
This chapter shows how the 7000/7010 Series Triple Quad reduces chemical and
electronic noise and how each component contributes to enhanced instrument
sensitivity.
Sensitivity
In the design of the 7000/7010 Series Triple Quads, sensitivity was addressed
within all stages of instrumentation, from the ion source to the detector. See
Figure 9.
3
1 2 4 5 4 6 7
8
Figure 9. Design elements that improve sensitivity
The 7000/7010 Series Triple Quads pass through four transitional steps in
translating a signal in the MRM process (Figure 11).
Step 1 The spectrum at the far left represents everything that is being ionized
at the ion source. A triple quad GC/MS reduces chemical noise for low-level
quantitation in a dirty matrix more than a single quad GC/MS does.
Step 3 After Q1 (MS1), fragment ions are generated in the collision cell. The
corresponding MS/MS spectrum is shown below the collision cell.
Step 4 Particular fragment ions can be selected to pass through the Q3 (MS2)
quadrupole. These are selected for quantitation and confirmation.
The second stage of selectivity using the Q3 (MS2) quadrupole removes much of
the chemical background. Typically, the chance of an isobaric interference at the
same exact mass as the fragmentation ion is remote.
Single MS:
SIM 283.8
MS/MS:
283.8—>213.9
Figure 13. HCB analysis by 7000/7010 Series Triple Quad in MRM mode
The 7000/7010 Series Triple Quads produce clear data to help evaluate complex
samples at lower levels. Figure 14 illustrates a chromatogram of multiple PCB
congeners in mussel extract measured at 2 pg on-column. The resulting data
provide strong signals on a flat baseline for clear data interpretation.
Reproducibility is an expression of how well the data agrees from test to test.
Figure 15 illustrates the GC/MS results when the organophosphorous
insecticide cyanophos, spiked into garlic at 0.5 ppb, is analyzed in MRM mode.
An overlay of five injections showing the quantitation transition of
243109 m/z demonstrates strong reproducibility.
Ion sources
The 7000/7010 Series Triple Quads offer the same ionization methods as the
single quadrupole mass spectrometers: electron impact ionization (EI) and
chemical ionization (CI).
vacuum system to pump away ions of carrier gas and un-ionized material as the
sample ions enter the quad, thereby reducing neutral noise and improving
sensitivity.
Filament assembly
Repeller
Source body
Lenses
Lenses
Source body
Repeller
Filament
Filament
Lenses
Drawout plate
Source body
Repeller
Filament
Dummy filament
Collision cell
What is the collision cell?
The collision cell collects percursor ions selected by the first quadrupole mass
filter and fragments them. The resulting product ions are transmitted to the
second quadrupole mass filter. In the 7000/7010 Series Triple Quads, the
collision cell is a high-pressure hexapole assembly with its linear acceleration
adjusted to optimize MS/MS fragmentation while eliminating crosstalk even at
very low dwell times (Figure 20).
Front analyzer
Post-filter
Figure 20. Collision cell technology produces higher sensitivity and faster responses without memory or
cross-talk effects
The components that contribute to this higher sensitivity and faster response
are:
• Small diameter hexapole collision cell
• High frequency hexapole collision cell
• Linear axial acceleration
• High pressure collision cell
• High speed digital electronics
The collision cell contains nitrogen and helium. Helium has been shown by
Agilent to provide more control over the fragmentation process, especially with
higher mass ions. It was added to the nitrogen stream to reduce neutral noise by
thermalizing metastables without letting them hit the detector. The helium is
then eliminated by the vacuum pump along with carrier gas and unfragmented
sample ions. The small diameter of the hexapole assembly assists in capturing
fragmented ions.
Why a hexapole?
The geometry of a hexapole provides advantages in two domains: ion focusing
and ion transmission (Figure 21).
• The first advantage is in ion focusing. Studies have shown that a quadrupole
provides better ion focusing than a hexapole and a hexapole provides better
ion focusing than an octopole. Therefore, ion focusing improves with a lower
number of poles in the filter.
• The second advantage involves ion transmission across a wide mass range,
or m/z bandwidth. In this case, the octopole is better than the hexapole,
which is better than the quadrupole.
The hexapole was chosen after extensive modeling, simulation, and
experimentation because it offered the best compromise between the focusing
of a quadrupole and the ion transmission of an octopole.
Figure 21. Broad mass range transmission and improved transmission efficiency using a hexapole
A potential difference is always present. This ensures that the precursor ions
coming from Q1 (MS1), or fragment ions generated in the collision cell, are
transmitted and not allowed to drift around at random.
Sweeping out the ions in this manner avoids the issue of crosstalk, where
residual product ions from a previous MRM experiment can interfere with the
product ion spectrum of a subsequent MRM experiment (see Figure 23). A
collision energy voltage is applied over the accelerating linear voltage to generate
fragments or product ions.
Figure 23. Collision cell clearing profile (500 pg Alprazolam, 20 ms dwell time)
Detector
The detector assembly is unique to Agilent (Figure 24). It is a high-energy dynode
coupled with an electron multiplier.
The high-energy dynode is located off-axis from the center of the rear quad
analyzer. This orientation reduces the possibility of neutral molecules impacting
the detector while at the same time attracting the ions with high voltages. When
the ion beam hits the dynode, the ions are converted to electrons before they
impact the multiplier. These electrons are attracted to the more positively
charged electron multiplier horn. The off-axis design of the detector allows
neutrals to pass through and be eliminated by the vacuum system without hitting
the detector.
The multiplier has a long lifetime since only electrons are allowed to impact it.
Ions never impact its surface. Gain normalized tuning provides consistent
sensitivity over the life of the electron multiplier. This also translates into
consistency from MS to MS, and lab to lab.
Pumping system
A single split flow turbomolecular pump is used for the entire vacuum system.
Sufficient vacuum for the entire process is achieved by partitioning the turbo
pump to create multiple vacuum stages. The vacuum system removes
molecules of carrier gas and any un-ionized or unfragmented sample molecules
from the ion source outlet, the collision cell, and both analyzers. This pump is
backed by a single mechanical foreline (roughing) pump (Figure 25).
This section will help you understand the design and operation of the Agilent
MassHunter Workstation Software for GC/MS Instrument Control for the
7000/7010 Series Triple Quads.
Description
The Instrument Control program (Figure 26) has the following features:
• Showing the instrument at work through real-time plots
• Running multiple samples through the sequence table, a spreadsheet-like
interface
• Controlling and monitoring instrument settings
• Tuning the instrument
• Setting up acquisition parameters for the GC and the Triple Quad
• Monitoring the chromatogram and mass spectra as samples are analyzed
• Setting up sequences of samples
Tuning
Autotune
An autotune taking about 8 minutes can be used when an extensive tune is
recommended (Figure 27). In this mode, everything is automatic. The tuning mix
is delivered by the calibrant delivery system (CDS), which is switched on
automatically during the tune.
Manual tune
A manual tune of user-defined ion masses with six corresponding profile masses
is available.
Tune reports
Tune reports are also available.
Acquisition
The 7000/7010 Series Triple Quad GC/MS systems can be controlled and
monitored from the Instrument Control panel, which is the window used for
accessing acquisition settings and sequence lists (Figure 28, Figure 29 on
page 49, and Figure 30 on page 50).
The real-time plot pane also can show the MS and GC results in real time.
*G7006-90008*
G7006-90008