Uncertainty. of Mean Rib Area
Uncertainty. of Mean Rib Area
L/FF/22
Uncertainty measurement of test parameters (Mechanical tests) as per NABL 141 guidelines
CALCULATIONS
Uncert. Source of Type Observation Unit Deviation Unit (Ai-Ᾱ)² UnitStd. deviation σ Unit
U uncertainty (Ai-Ᾱ) = √∑(Ai-Ᾱ)²
U1 Repeatability A1 = 2.40 mm2/mm 0.1 0.01 √ n-1
A2 = 2.28 -0.02 0.0004 = 0.083 mm2/mm
A3 = 2.25 -0.05 0.0025 Std Uncertainty u1
A4 = 2.37 0.07 0.0049 = σ/ √n
A A5 = 2.20 -0.1 0.01 = 0.0371 mm2/mm
2.3 mm2/ ∑ = 0.0278 mm2/ Relative Uncertainty
Ᾱ= mm mm
n= 5 = u1/Ᾱ
= 0.0161
U2 Vernier B Length Standard Uncertainty u2 = Uc/2 Relative U2
Caliper ± Uc from Cal Cert = u2
M 300 mm = 0.0092 mm = 0.0046 mm M
at k = 2 = 0
Uncertainty Budget
Uncert. Source of Estimate Limits Type F Std. Sensi-tivity Uncert. Divisor Relative DOF
Uncertainty Value Uncertainty co- Contribu- (M) Uncertainty
efficient tion Contribu-tion
Combined Rel. Std. Uc = √ {(U1)² + (U2)² + (U3)² + (U4)² + (U5)² + (U6)²+ (U7)²}
Uc = 0.016