Unity 2
Unity 2
User manual
UNITY 2™
www.markes.com
Contents
1. Preface ................................................................................................. 5
1.1 Notices ............................................................................................ 5
1.2 Warranty.......................................................................................... 5
1.3 Regulatory compliance .................................................................. 5
1.4 Important safety warnings ............................................................. 5
1.4.1 Labels/symbols................................................................. 6
1.4.2 Mains voltages .................................................................. 6
1.4.3 High temperatures ............................................................ 6
1.4.4 Hydrogen gas safety ......................................................... 7
1.4.5 Cleaning and decontamination........................................ 7
1.5 Environment operating conditions ................................................ 7
1.5.1 Temperature...................................................................... 7
1.5.2 Humidity ............................................................................ 7
1.5.3 Altitude .............................................................................. 8
1.6 Technical Specifications ................................................................ 8
1.6.1 Physical properties ........................................................... 8
1.6.2 Electrical properties.......................................................... 8
1.7 Technical support contact details ................................................. 8
2. Instrument familiarisation .................................................................. 9
2.1 Hardware ........................................................................................ 10
2.1.1 Sample tubes .................................................................... 10
2.1.2 Tube desorption oven ....................................................... 10
2.1.3 Tube filters and seals ....................................................... 10
2.1.4 The cold trap ..................................................................... 10
2.1.5 Cold trap cooling and heating .......................................... 10
2.1.6 Trap filters and seals ........................................................ 11
2.1.7 Split filters ......................................................................... 11
2.2 Sequence of operation................................................................... 12
2.2.1 Standby.............................................................................. 12
2.2.2 Leak test ............................................................................ 13
2.2.3 Prepurge ............................................................................ 14
2.2.4 Primary (tube) desorption ................................................ 15
2.2.5 Pre trap fire purge............................................................. 16
2.2.6 Secondary (trap) desorption ............................................ 17
2.3 Insertion and removal of a sample tube in UNITY 2 .................... 18
2.3.1 Insertion............................................................................. 18
2.3.2 Removal............................................................................. 18
2.4 Desorb and split flows.................................................................... 18
2.4.1 Gas flow through the cold trap ........................................ 18
www.markes.com
www.markes.com
www.markes.com
1. Preface
This manual provides detailed instructions on the use of the UNITY 2. It is suitable for users with little or
no prior experience of the system and details the software/hardware interface and procedure for first
sample runs.
1.1 Notices
No part of this manual may be reproduced in any form or by any means (including electronic storage and
retrieval or translation into a foreign language) without prior agreement and written consent from Markes
International.
Please note that the material contained in this document is subject to being changed, without notice, in
future editions. Markes International shall not be liable for errors or for incidental or consequential
damages in connection with the supply, use or performance of this document or of any information
contained herein, unless a separate agreement between Markes International and the user should take
precedence.
1.2 Warranty
UNITY 2 is designed for laboratory use only. It is not intended for use in domestic establishments or
establishments directly connected to a low-voltage power supply network that supplies buildings used for
domestic purposes. Where equipment is used in a field placement environment, care must be taken to
ensure that the instrument is not exposed to detrimental conditions, i.e. rain, wind, or sun. Exposure may
diminish the performance, cause damage to the instrument and/or cause the equipment to become
unsafe to the user.
If the equipment is not used in a way specified by Markes International, the safety protection provided by
the equipment may be reduced. Furthermore, system failures arising from such use may not be covered in
standard warranty and/or service contract agreements.
www.markes.com
1.4.1 Labels/symbols
Throughout this manual symbols will appear where carrying out an operation may involve hazards.
Symbols and warnings also appear on the instrument, and these should be adhered to at all times.
Markes International can accept no liability for failure to do so.
BURN HAZARD
A hot surface may result in a burn injury.
INSTRUMENT/PART DAMAGE
Damage to the instrument or module may occur. This damage may not be covered under
the standard warranties.
LIFTING HAZARD
Injury may occur if proper lifting procedures are not followed.
WARNING
General warning symbol to alert the user that personal injury or instrument damage may
occur if the instrument is improperly used or if instructions
instructions are not followed correctly.
DISPOSAL
This label indicates the instrument must not be disposed in regular waste but in
accordance with the WEEE scheme.
scheme
www.markes.com
Due to the high temperatures involved in the flow path, other zones of the instrument
will be at higher temperatures during operation. These may not on visual inspection
be obvious to the user. These zones are:
o The insulation of the GC transfer line.
o Top and side covers (especially directly above the heated valve).
See Section 2 for the location of these zones.
1.5.1 Temperature
Recommended operating ambient temperature range is 15 to 30°C.
1.5.2 Humidity
Recommended operating humidity range is 5 to 95% non-condensing.
www.markes.com
1.5.3 Altitude
The product has been evaluated for operation up to a maximum of 6500 ft (2000 m), in accordance with
safety standards. Higher altitudes pose no safety risk, but instrument performance may be reduced.
NOTE: For storage or shipping the allowable temperature range is -40 to 70°C and the allowable humidity
range is 5-95% non-condensing. After instrument exposure to extremes of temperature or
humidity, allow 2 hours for return to the recommended ranges before switching on.
www.markes.com
2. Instrument familiarisation
www.markes.com
2.1 Hardware
www.markes.com
capillary chromatography is produced without on-column focusing and with desorption flows as low as 2
ml/min. This facilitates splitless operation with high-resolution capillary GC.
www.markes.com
2.2.1 Standby
STANDBY
GC
Heated Valve
Sample Tube
Split Tube
Cold Trap
KEY
Pressure and flow
Pressure, no flow
www.markes.com
GC
Heated Valve
Sample Tube
Split Tube
Cold Trap
KEY
Pressure and flow
Pressure, no flow
From the Standby position the UNITY 2 Status changes to Leak test. The system allows carrier gas through
the tube and into the cold trap and split filter and the whole sample flow path is pressurised to capillary
column head pressure. Carrier gas continues to be supplied to the capillary analytical column throughout
the leak test (as it is through every other stage of UNITY 2 operation).
After 5 seconds the flow path is isolated from the capillary column head pressure and a pressure
transducer monitors the pressure in the sample flow path. If the measured pressure drops by more than
5% in 30 seconds this is classified as a leak in the sample flow path and UNITY 2 status changes to Tube
Leaked.
If the system maintains pressure, no leak is detected and UNITY 2 passes on to the second part of the
leak test. The flow path is depressurised briefly. The sample flow path pressure is again monitored to see
if it increases by 5% over 30 seconds. This would indicate an internal leak across the heated valve. If
UNITY 2 fails this part of the leak test, the status again changes to Tube Leaked. Failure of this part of the
leak test requires the attention of a trained service engineer.
The main causes of Leak Test Failure are:
Wearing of the O-rings which seal the tube
Interference with the tube seal by fibres and particles
Damaged O-rings
Leaking split filter tube seal
Wearing of the cold trap seals
See the Technical Support Manual for further information.
www.markes.com
2.2.3 Prepurge
GC
Heated Valve
Sample Tube
Split Tube
Cold Trap
KEY
Pressure and flow
Pressure, no flow
Each tube must be purged thoroughly with carrier gas to remove air before heat is applied. Even the
smallest trace of oxygen could result in sorbent and possible analyte oxidation, generating artifacts and
compromising data quality.
There is also the option in this step to perform an elevated temperature purge. See Section 3.6.4.
www.markes.com
GC
Heated Valve
Sample Tube
Split Tube
Cold Trap
KEY
Pressure and flow
Pressure, no flow
At the start of primary (tube) desorption the tube oven begins to heat (either from ambient or from the
temperature of the optional elevated temperature purge if selected). Note that the tube desorption time is
measured from the beginning of tube oven heating and not from the time at which the sample tube
reaches the desorption temperature.
www.markes.com
GC
Heated Valve
Sample Tube
Split Tube
Cold Trap
KEY
Pressure and flow
Pressure, no flow
Following primary desorption, the heated valve is moved and carrier gas is flushed through the split tube
and the trap to remove any residual air and water prior to trap injection. It may also be used to dry purge
the cold trap prior to injection when direct desorbing solid or humid samples.
This pre-trap fire purge time can be adjusted in the TD method.
www.markes.com
GC
Heated Valve
Sample Tube
Split Tube
Cold Trap
KEY
Pressure and flow
Pressure, no flow
The final stage is secondary trap desorption. The cold trap is heated at a rate of 100°C/s during the first
seconds ensuring rapid desorption of analytes into the carrier gas stream. The flow through the cold trap
at this stage is determined by the column flow and any outlet split.
NOTE: the flow path of UNITY 2 remains in the trap desorption configuration until the trap has cooled
below 50°C.
www.markes.com
2.3.1 Insertion
Tubes are inserted using the left-hand lever mechanism. Orientate the tube such that the sampling
(grooved) end is pointing to the rear of the instrument. Before inserting a sample ensure that the tube
oven is relatively cool, note that the oven temperature is displayed on the UNITY 2 software status bar
(see section 3.2). Place the tube horizontally in the tube oven, such that it is clear of the seals at both
ends. Operate the lever towards the rear of the instrument to seal the tube into the UNITY 2 flow path.
The instrument is then ready to run.
2.3.2 Removal
As soon as a tube has been desorbed (i.e. after primary desorption) it can be removed from the desorption
oven. If the oven and tube are still hot (you can check this by looking at the software status bar, it is
strongly recommended that the tube extractor tool, included in the shipping kit, is used to remove the
tube. Use the tool to grasp the tube at the end released from the tube seal and pull steadily.
www.markes.com
If the analysis method is to include a split (single or double) then answer ‘Yes’ to the question in the box.
The UNITY 2 flow path will then alter to allow both the desorb flow and the split flow to be adjusted and
measured at the split and desorb flow vents.
If the analysis method is completely splitless or if only the desorb flow needs to be set, then answer ‘No’ to
the question in the box. The UNITY 2 flow path will then alter so that there is no split flow, only the desorb
flow, to be adjusted and measured.
www.markes.com
The column flow should be measured at the detector end using procedures recommended for the GC in
question. Once adjusted and measured as required, all the flows can be entered into the Confirm/Enter
Flows dialog box on the controlling method, see Section 3.6.2.
www.markes.com
Flame Ionisation Detector (FID): Typical detection limits 10-100 pg for individual hydrocarbons.
Electron Capture Detector (ECD): Highly selective detector. Used for halogenated solvents. Detection
limits well below 1 pg for compounds such as carbon tetrachloride and some heavily halogenated
pesticides.
Thermal Conductivity Detector (TCD): General purpose, low sensitivity detector. Typical detection limits in
the order of 10 ng.
Zero split or Unusually among thermal desorbers, During tube desorption the desorb flow must
splitless UNITY 2 can operate in splitless mode be set to at least 10 ml/min to provide
in conjunction with narrow bore (0.32 enough flow through the sample tube for
mm ID) columns and MS detectors as efficient thermal desorption.
the minimum flow required through
During tube desorption the desorb flow should
the cold trap during trap heat is 2
not far exceed 50 ml/min or there may be a
ml/min.
risk of breakthrough from the cold trap.
Single split During single split operation the split During tube desorption the desorb flow must
may either be open during tube or trap be set to at least 10 ml/min to provide
desorption. The advantages of having enough flow through the sample tube for
the split open on the way into the trap efficient thermal desorption.
(inlet split) are:
During tube desorption the trap flow should
1. That a relatively fast flow can be not exceed 50 ml/min or there may be a risk
used through the tube to of breakthrough from the cold trap.
facilitate desorption while, at the
During tube desorption the trap flow should
same time, the flow passing
not be less than 2 ml/min to ensure efficient
through the cold trap is kept low
sweeping of analytes onto the trap sorbent.
to aid retention and focusing of
analytes.
www.markes.com
www.markes.com
3. Software
UNITY software will run on most 32-bit versions of Windows® and 64-bit versions of Windows 7, however,
use of currently supported versions of Windows is strongly recommended. In general, a PC with sufficient
resources to run 32-bit Windows should have enough performance to control UNITY.
Other Windows software may be active on the PC at the same time as UNITY 2 software, for example data
handling software, word processing and spreadsheet functions.
Elapsed
Status Pressure UNITY zones GC
Time
The level of information presented for each category may be selected by a right click of the mouse on the
appropriate section of the status bar. The parameters shown on the status bar are as follows:
www.markes.com
Status Shows the overall status of UNITY 2 as it progresses through the various stages of
the sequence - Standby, Leak Test, Trap Heat etc. Status will also report tube
failures i.e. Tube Leaked.
Elapsed Time Shows the time elapsed since the start of the function currently operating on UNITY
2 and should be read in conjunction with the Status parameter. For example if the
Status is reading Prepurge and the Elapsed Time is reading 0.3 this tells us that
UNITY 2 is 0.3 minutes (i.e. 18 seconds) into the ambient purge. If the Status is
reading Primary Desorb and the Elapsed Time is 4.9 - then it is 4.9 minutes since
the start of tube desorption etc.
Pressure Shows the pressure of the carrier gas within the UNITY 2 flow path.
UNITY zones UNITY 2 continually monitors the actual temperature of its internal heated/cooled
zones (tube oven, cold trap, transfer line and heated valve) and the actual carrier
gas pressure and compares them to the set values.
When first switched on UNITY 2 status changes to Standby and the display for UNITY
2 temperature will be highlighted blue until all zones reach the relevant set point.
The Purge Gas indicates whether the dry air or nitrogen is at a high enough pressure
for the Peltier coolers to operate. If Purge Gas is displayed as ‘Off’ then this
indicates that the purge gas pressure is not high enough. In this scenario the Peltier
coolers will be switched off.
NOTE: A run may be started when UNITY 2 is Not Ready and the run will then pause,
waiting for the appropriate zone to reach temperature before desorbing the tube.
The status will read Equilibrating at this stage.
NOTE: The actual temperature of the tube oven has no effect on the ready status as
heat is not applied to the oven until tube desorption starts.
GC Indicates whether the GC status is ready () or not ready ().
See Section 3.2.1 below.
www.markes.com
GC is ready, twice during its operation sequence, before starting an analysis and/or before desorbing the
cold trap.
Generating a new New methods are generated by either selecting File > New, by clicking the New
method File Icon ( ) at the top left-hand side of the toolbar, or by pressing CTRL and N.
A new method screen will be brought up containing a set of default parameters
ready for you to modify. Once the method has been generated it can be named
and saved, see below.
Opening a stored Methods are opened by either selecting File > Open or by clicking the open file
method icon ( ) at the top left-hand side of the toolbar. This brings up the directory
(nominated under Options, see Section 3.4.1) containing UNITY 2 method files.
Select the required file name and click OK or double-click on the required file
name to access it.
Saving a method New methods can be saved by
1. Pressing the floppy disk icon ( ) at the left of the tool bar
2. Selecting File > Save
3. Pressing CTRL and S
4. Selecting File > Save As
Using the first three routes brings up the Properties dialogue box. Using route 4
brings up the Method File Directory and allows a file name to be entered for the
method. The Properties dialogue box then automatically appears with the file
name in the appropriate field.
The Method Author and Company names will appear in the Method Properties
dialogue box, as set up under Options, see Section 2.4.1. If nothing has been
entered for these fields under Options the blanks may be filled in at this stage.
The Method Name and Notes sections can be used to provide additional
information on the method if required.
www.markes.com
Method linking Methods can be linked together to perform consecutively on a given tube. This is
most useful where an analytical method is to be followed by a tube conditioning
method or to automate the process of thermal desorption method development.
Select Edit > Copy, or click the copy icon ( ) on the toolbar (the shortcut CTRL
and C can also be used).
Copy will put a "text copy" of the active method onto the Windows Clipboard (in
English Only), thus allowing you to Paste the text string into another application
such as a word processor
Paste Paste is activated after Copy has been used and allows for a copy of the active
method to be pasted to another method (after THAT method has been made
Active).
Note: UNITY 2 does not support pasting from the Windows clipboard i.e. you
cannot paste from the Word processor to UNITY 2
Start run Start run is used to start a desorption sequence.
Start is accessed by clicking the Start Run icon ( ) on the toolbar or selecting
Instrument and then Run.
Set gas flows Used to set the split and desorb flows.
www.markes.com
Exchange split tube Used to divert any split flow in standby mode so that the split/recollection tube
can be removed without loss of carrier gas.
www.markes.com
Method linking sequence which is setup in this example to run two injections with a 2 stage desorption
method followed by a tube conditioning step.
After a method has been run the next method in the sequence is downloaded to UNITY 2 and run. The
procedures to setup and edit linked
ed sequences are:
Open a new sequence Right click and select ‘New Sequence’ to open a new
method linking table.
Open an existing sequence Right click and select ‘Open Sequence’ to enable the user to
open a previously saved method linking table.
Add an existing
xisting method to the link table Right click and select ‘Add Item’ from the list of options.
The appropriate method can then be selected for the
directory.
Insert a previously stored sequence into Right click and select ‘Add Sequence’’ then select the
the current sequence desired sequence file.
Delete a highlighted method from the Right click and select the ‘Delete Item’ option or press the
link table delete key on the keyboard.
Change the number of injections Double click on the existing number and select the
appropriate figure from the drop down method. The
software allows 1 - 100 injections of a particular method.
Recycle the sequence Checking the ‘recycle’ box at the top left corner of the
method linking table. This will recycle the process unt
until
stopped by the user. However, a set number or recycles can
be selected by checking the ‘stop after’ box, and entering
the appropriate number of repeats in the ‘cycles’ box.
www.markes.com
NOTE: Should the stop sequence button be pressed during a linking sequence, a dialogue box containing
a number of options appears. This enables the user to either continue running, or stop the sequence at
one of the stages listed.
NOTE: GC cycle time will operate in the same manner as a single tube desorption.
Once Trap Heat has been selected, UNITY 2 will ask you to confirm that the default method is that which
you wish to use. For example, if the default method was set to Use Controlling Method Parameters then
the box message will read Heat Trap with the Controlling
Controlling Method Parameters? Select Yes, No, or Cancel.
If Yes is selected, UNITY 2 proceeds with trap heat. If No is selected the instrument software immediately
takes you to the relevant section under Options to change your selection.
Also, a ‘Wait for GC Ready, before Heating?’ option is made available. Depending on the response to this
option, UNITY 2 will either wait for an External Ready signal from the GC equipment before proceeding, or
continue the trap heat regardless of the GC status.
Caution: Ensure that the dedicated Trap Heat Method or the Controlling Method, if being used, is
compatible with the maximum temperature of the least stable sorbent in the cold trap.
When using Trap Heat several times to clean a freshly packed trap, it is advisable
advisable to use a dedicated Trap
Heat Method and set the Trap Low Temperature to its maximum +50°C, in order to save the time taken
for the system to equilibrate.
Caution: For
or traps containing carbon molecular sieves Trap Heat Method must not be used until th
the
trap has been extensively purged with helium immediately prior to Trap Heat. Refer to the
cold trap certificate supplied with your trap for further information.
www.markes.com
This is accessed using the status icon ( ) or by selecting View and then Status Show.
The schematic display of UNITY 2 status is linked directly to operation of the instrument. The di
display is
thus not just cosmetic, but is a representation of the actual status of the instrument.
www.markes.com
3.4.1 Configuration
The configuration page contains the current instrument configuration settings.
Instrument Configuration
For UNITY 2 operation the Instrument Configuration will be set to UNITY (manual).
Communications Port
The communications port determines serial port number (e.g. Com1) that is used for communicating with
UNITY.
Sequence Options – Enable Direct Mode
If the Enable Direct Mode option is selected an additional mode of operation will be available. Additional
hardware is required for this mode to be used correctly. Direct mode is only required when the any of the
following hardware accessories
cessories are installed:
U-INLET accessory.
www.markes.com
3.4.2 GC Interface
Used to set the GC Interface logic.
The GC Interface logic is set to reflect whether the GC sense is Open or Closed for both GC Start (Out) and
GC Ready (In) i.e. for the particular GC being used is the contact Open or Closed to reflect a Start signal,
and Open or Closed to reflect a Ready signal. This information is usually provided in the GC Users Manual.
www.markes.com
The Carrier Gas being used can be set using the drop down box.
3.4.4 Methods
This is accessed by selecting ‘View > Options > Methods’. There are two method functions which can be
changed.
www.markes.com
www.markes.com
3.4.5 Reporting
This option influences the information recorded in the ‘sequence reporter’ page of method linking, see
Section 3.3.1. Recording parameters are selected by clicking on the appropriate box(es).
NOTE: the reporting options available will change if an accessory is being used in conjunction with UNITY
2.
Method Name
Displays the name of the method used.
Desorb Start Time
Displays the time at which tube desorption started.
Desorb End Time
Displays the time at which tube desorption was completed.
Peak Desorb Temp
Displays the maximum temperature the tube oven reached during the analysis.
Trap Fire Time
Displays the time at which the cold trap desorption started (and therefore the time at which a GC
connected to the UNITY would have been triggered to start).
Unity Deviation
Displays any error messages encountered during operation (e.g. Tube Leaked)
Injection Count
Displays the number of times a given sample has been analysed.
www.markes.com
Cycle Count
Displays the number of sequence cycles that have been completed.
www.markes.com
www.markes.com
If required, select Trap In Line during the prepurge by clicking on the box to
enter a check. Similarly, click on the Split On box to enter a check if a split is
required during the prepurge. Remember one or other or both of these flows
must be selected during the prepurge.
Tube Desorb Set the Time 1 - desorption time - and the Temp 1 - desorption purge
temperature - by clicking in the relevant boxes and typing in the required values.
www.markes.com
Tube desorption may be carried out split or splitless i.e. all the sample from the
sample tube can be transferred to the trap or part of it sent to the split vent. If a
split is required during tube desorption (a so-called inlet split, see Section 1.4)
check the relevant box. If the split is selected to be off during tube desorption,
all the analytes will be transferred from the tube to the cold trap.
If an elevated temperature purge is required, see Section 2.6.4, then Time 1
and Temp 1 should be set to reflect the required elevated Prepurge time and
temperature. In this case, Time 2 and Temp 2 become the actual analytical tube
desorption time and temperature and time values should be entered
appropriately.
www.markes.com
Most applications work well with a Trap Low of -10°C. Obvious exceptions to
this include the analysis of samples containing high levels of water or standards
containing unwanted solvent. In these cases it may be useful to maintain a Trap
Low temperature at or near ambient (25-30°C). This allows water and volatile
solvents such as methanol and methylene chloride to breakthrough the trap
(depending on which sorbents are in the trap) and be eliminated from the
system before trap desorption and the beginning of the chromatographic
analysis.
Set the Trap High temperature as appropriate for the trap sorbent and target
analytes. The correct Trap Hold time should ensure all analytes are desorbed
from the trap and that the trap is again clean and ready to use for the next
sample. This is typically set to 3 minutes.
Trap desorption can be carried out split or splitless i.e. if a split is required
during trap desorption (a so-called outlet split, see Section 1.4), check the
relevant box. If the split is selected to be off during trap desorption, all the
analytes will be transferred from the trap to the GC analytical column during
desorption.
The flow through the trap when it is hot = the flow through the GC column plus
that through the split vent, if any.
Flow Path Temp
Set the temperature required for the heated valve and sample flow path by
clicking in the box and typing in the required value. The following general
guidelines may be applied:
All analytes, volatile enough to be present in the air in the vapour phase at
ambient temperature, will pass comfortably through the system with the flow
path between 120°C and 150°C.
For ease of use, extended system lifetime and trace level operation a lower
www.markes.com
Once the values have been entered and confirmed by pressing OK, the split
ratios are automatically calculated and displayed.
For measurement of gas flows see Section 1.4.4.
www.markes.com
www.markes.com
In order to access the tube on the split side (charcoal or SecureTD-Q), the exchange split tube icon ( )
must be selected. This will isolate the split tube from the carrier gas supply without disrupting the flow to
the analytical column allowing chromatography to continue uninterrupted as the SecureTD-Q and charcoal
tubes are exchanged.
Once ‘exchange split tube’ has been selected, lift the right-hand levered tube cover to release the charcoal
split tube from one of its seals. Use your fingers or the tube extractor tool to gently pull the tube free from
the other seal. Once the charcoal tube has been removed, the clean (SecureTD-Q) sorbent tube should be
inserted with its sampling (grooved) end pointing to the rear of the instrument. Lower the right-hand
www.markes.com
levered tube cover to seal it into position then press the stop icon ( ) to return the carrier gas flow path
to normal standby status.
If the sample is being re-collected during an analysis, re-activate the ‘exchange split tube’ icon as soon as
the desorber returns to standby mode after trap desorption. Remove the re-collected sample using the
above procedure and cap it immediately. Insert another clean (SecureTD-Q) sorbent tube or the charcoal
split tube as required, and press the stop icon ( ) to return to normal standby mode.
NOTE: SecureTD-Q will not work for methods that are operated entirely splitless. Either an inlet or outlet
split or both are required.
As well as repeat analysis, SecureTD simplifies method development, provides a convenient method
validation tool and allows critical samples to be archived when required.
For further information on split re-collection for repeat analysis see TDTS 24.
www.markes.com
Parameter Range
www.markes.com
3.7.1 Windows 7
1. Click on the start button and select control panel.
www.markes.com
4. Select Uninstall/Change Program and follow the on screen prompts (when asked to remove shared
files click yes). The Markes TD software should now be uninstalled.
3.7.3 Windows XP
1. Click on the Start button and select Control Panel.
2. In the Control Panel select Add/Remove Programs.
3. Scroll down the list presented and select Markes TD 5.1.0
4. Click on Add/Remove and follow the on screen prompts (when asked to remove shared files click
yes). The Markes TD software should now be uninstalled.
www.markes.com
www.markes.com
5. Technical Support
Replace sample tube seals and filters Only if damaged / leak test failing
UltrA ITEMS
www.markes.com
Ensure that empty tubes are fitted with a sorbent retaining gauze in the correct position at the front of the
tube - the position of this gauze is critical for diffusive sampling. (Factory supplied empty tubes arrive with
the gauze pre-fitted). Insert a quantity of quartz wool after the gauss to prevent migration of fine sorbent
particles out of the tube.
Tubes should be gravity filled with sorbent (not filled under pressure). For compliance with International
Standards and accreditation schemes the mass of sorbent packed into the tube should be measured and
recorded.
Once the appropriate mass of sorbent has been poured into the tube a few taps on a hard surface will be
sufficient to settle the packing.
A small amount of quartz wool should be inserted followed by a second sorbent retaining gauze, inserted
using a TubeMate (C-TBMTE) or other suitable tool, until the gauze sits firmly on top of the sorbent
(without compressing the packing).
A gauze retaining spring is then inserted using the TubeMate to hold the gauze in place.
Tubes should be impedence tested prior to use
A range of pre-packed sorbent tubes is available from Markes International (see relevant Accessories for
Thermal Desorption catalogue or contact Markes on [email protected] )
www.markes.com
www.markes.com
It is not necessary to store capped tubes in refrigerated conditions unless they are sampled, multibed
tubes in which case refrigeration may be required to prevent analyte migration to the stronger sorbents.
In fact in most cases differential contraction of the tubes and seals at fridge or freezer temperatures can
cause the caps to come loose. If refrigeration is to be used, caps must be retightened (a quarter turn)
once they have reached their storage temperature.
NOTE: A tube should be allowed to warm up and equilibrate properly at ambient temperature before the
beginning of sampling. If this is not done, a significant mass of water from the atmosphere will be retained
in the tube via simple condensation.
NOTE: If tubes are to be transported in such a way as to be exposed to very cold temperatures, it is
advisable to follow the above retightening procedure by cooling the tubes prior to shipment. When
monitoring trace level atmospheric components, conditioned and sampled tubes can be wrapped in
uncoated aluminum foil and / or placed in a sealed, non-outgassing container, such as an uncoated tin,
during transportation and storage.
www.markes.com
www.markes.com
C-DL1S0 Cap, DiffLok, inert coated, pk 100 - for use on hot end
C-DL010 Cap, DiffLok, st. st. pk 10 – for use on cold end
C-DL100 Cap, DiffLok, st. st. pk 100 – for use on cold end
C-DLP10 Cap, DiffLok, inert coated / st. st, pk 10 pairs
C-DL1P0 Cap, DiffLok, inert coated / st. st, pk 100 pairs
2.7 Changing the nozzle seals (UltrA)
NOTE: Before proceeding ensure that UltrA and UNITY 2 are switched off by closing down the software
and switching off the main power switch at the back of the instruments.
The required parts are:
U-COV07 O-ring, Size 007, pk 10
SERZ-0351 O-ring extraction tool
SERZ-0285 O-ring insertion tool
www.markes.com
Empty cold traps should be packed from the wider (2 mm) bore end using the following procedure.
www.markes.com
First insert a 2-5 mm plug of quartz or glass wool using a suitable flexible tool such as a 15 cm length of
1/16 -inch, narrow bore plastic tubing (PTFE or PEEK tubing is ideal.)
Pour in the required amount of sorbent. Analytes enter and desorb from the trap through the restricted
end. Multibed traps must therefore be packed with the sorbents arranged in order of increasing strength
from the narrow end. 1 to 3 mm plugs of quartz wool must separate different sorbents.
Plug the end of the trap with another glass wool plug, backed with a packing retaining spring.
If unsilanised glass (rather than quartz) wool is used in the cold trap, it may cause degradation and / or
tailing of polar and labile compounds. Silanised glass wool does not suffer from these limitations but
must not be taken above 275 °C or breakdown products from the silylating reagent will coat the sample
flow path of the desorber, reducing recovery of high boiling compounds.
NOTE: A 6 cm length of the wider bore section of the trap tube, measured from the point of the bore
restriction, is subjected to full cooling and heating power. The trap packing, including all but the
back glass / quartz wool plug, should be within this 6 cm length of the trap
NOTE: Do not over-compress the cold trap packing as this will cause high impedance, which may limit
trap desorption flows.
If high trap desorption flows will be required, for example when using a high split ratio, use 40-60 rather
than 60-80 mesh sorbent.
The carrier gas supply and See video tutorials for animated cold trap removal
compressed air to the
instrument and the instrument itself
must be turned off before starting.
www.markes.com
www.markes.com
www.markes.com
www.markes.com
www.markes.com
www.markes.com
2.
www.markes.com
www.markes.com
1.
www.markes.com
www.markes.com
www.markes.com
6. Trademarks
UNITY 2™, ULTRA™, ULTRA 50:50™, Air Server™ and SecureTD-Q™ are all trademarks of Markes
International Ltd., UK
Windows® is a registered trademark of Microsoft Corporation, USA.
www.markes.com