DC Programmable Electronic

Loads
Series IT8500plus User’s Manual

Model:IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512A
+/IT8512B+/IT8512C+/IT8512H+/IT8513A+/IT8513C
+/IT8514C+/IT8514B+/IT8516C+
Version:2.0
Notices Safety Notices
Warranty
© Itech Electronic, Co., Ltd. 2015
No part of this manual may be The materials contained in this
reproduced in any form or by any means document are provided “as is”, and
(including electronic storage and is subject to change, without prior
retrieval or translation into a foreign notice, in future editions. Further, to
language) without prior permission and
A CAUTION sign denotes a
the maximum extent permitted by hazard. It calls attention to an
written consent from Itech Electronic,
applicable laws, ITECH disclaims
Co., Ltd. as governed by international operating procedure or practice
copyright laws. all warrants, either express or
implied, with regard to this manual that, if not correctly performed
Manual Part Number and any information contained or adhered to, could result in
IT8500+-402008 herein, including but not limited to damage to the product or loss of
the implied warranties of important data. Do not proceed
Revision merchantability and fitness for a beyond a CAUTION sign until
Second Edition: May 09, particular purpose. ITECH shall the indicated conditions are fully
not be held liable for errors or for understood and met.
2017
incidental or indirect damages in
Itech Electronic, Co., Ltd. connection with the furnishing, use
or application of this document or of
Trademarks
any information contained herein.
Pentium is U.S. registered trademarks Should ITECH and the user enter
of Intel Corporation. into a separate written agreement A WARNING sign denotes a
with warranty terms covering the hazard. It calls attention to an
Microsoft, Visual Studio, Windows and
materials in this document that operating procedure or practice
MS Windows are registered trademarks
of Microsoft Corporation in the United conflict with these terms, the that, if not correctly performed
States and/or other countries and warranty terms in the separate or adhered to, could result in
regions. agreement shall prevail. personal injury or death. Do not
Technology Licenses proceed beyond a WARNING
The hardware and/or software sign until the indicated
described herein are furnished under a conditions are fully understood
license and may be used or copied only and met.
in accordance with the terms of such
license.

Restricted Rights Legend


NOTE
Restricted permissions of the U.S. A NOTE sign denotes
government. Permissions for software
and technical data which are authorized important hint. It calls attention
to the U.S. Government only include to tips or supplementary
those for custom provision to end users. information that is essential for
ITECH follows FAR 12.211 (technical users to refer to.
data), 12.212 (computer software).
DFARS 252.227-7015 (technical
data--commercial products) for national
defense and DFARS 227.7202-3
(permissions for commercial computer
software or computer software
documents) while providing the
customized business licenses of
software and technical data.
IT8500+ User’s Manual

Quality Certification and Assurance


We certify that series IT8500+ electronic load meets all the published
specifications at time of shipment from the factory.

Warranty
ITECH warrants that the product will be free from defects in material and
workmanship under normal use for a period of one (1) year from the date of
delivery (except those described in the Limitation of Warranty below).
For warranty service or repair, the product must be returned to a service center
designated by ITECH.
 The product returned to ITECH for warranty service must be shipped
PREPAID. And ITECH will pay for return of the product to customer.
 If the product is returned to ITECH for warranty service from overseas, all
the freights, duties and other taxes shall be on the account of customer.

Limitation of Warranty
This Warranty will be rendered invalid in case of the following:
 Damage caused by circuit installed by customer or using customer own
products or accessories;
 Modified or repaired by customer without authorization;
 Damage caused by circuit installed by customer or not operating our
products under designated environment;
 The product model or serial number is altered, deleted, removed or made
illegible by customer;
 Damaged as a result of accidents, including but not limited to lightning,
moisture, fire, improper use or negligence.

Safety Symbols
Direct current ON (power on)

Alternating current OFF (power off)

Both direct and alternating Power-on state


current

Protective conductor terminal Power-off state

Earth (ground) terminal Reference


terminal

Caution, risk of electric shock Positive terminal

Warning, risk of danger (refer Negative terminal


to this manual for specific
Warning or Caution
information)

Frame or chassis terminal - -

Copyright © Itech Electronics Co., Ltd. i


IT8500+ User’s Manual

Safety Precautions
The following safety precautions must be observed during all phases of
operation of this instrument. Failure to comply with these precautions or specific
warnings elsewhere in this manual will constitute a default under safety
standards of design, manufacture and intended use of the instrument. ITECH
assumes no liability for the customer’s failure to comply with these precautions.

 Series IT8500+ electronic load supports 110V/220VAC input and need to


switch the input voltage before operation.
 Do not use the instrument if it is damaged. Before operation, check the
casing to see whether it cracks. Do not operate the instrument in the
presence of inflammable gasses, vapors or dusts.
 The electronic load is provided with a three-core power line during delivery
and should be connected to a three-core junction box. Before operation, be
sure that the instrument is well grounded.
 Make sure to use the power cord supplied by ITECH.
 Check all marks on the instrument before connecting the instrument to
power supply.
 Use electric wires of appropriate load. All loading wires should be capable
of bearing maximum short-circuit current of electronic load without
overheating. If there are multiple electronic loads, each pair of the power
cord must be capable of bearing the full-loaded rated short-circuit output
current
 Ensure the voltage fluctuation of mains supply is less than 10% of the
working voltage range in order to reduce risks of fire and electric shock.
 Do not install alternative parts on the instrument or perform any
unauthorized modification.
 Do not use the instrument if the detachable cover is removed or loosen.
 To prevent the possibility of accidental injuries, be sure to use the power
adapter supplied by the manufacturer only.
 Never use the instrument with a life-support system or any other equipment
subject to safety requirements.

 Failure to use the instrument as directed by the manufacturer may render


its protective features void.
 Always clean the casing with a dry cloth. Do not clean the internals.
 Make sure the vent hole is always unblocked.

Copyright © Itech Electronics Co., Ltd. ii


IT8500+ User’s Manual

Environmental Conditions
The instrument is designed for indoor use and an area with low condensation.
The table below shows the general environmental requirements for the
instrument. The speed of fan will change intelligently by the temperature of
radiator. When the temperature is up to 40°C, the fan will be on and adjust
intelligently when temperature changes.
Environmental Conditions Requirements
Operating temperature 0°C to 40°C
Operating humidity 20%-80% (non-condensation)
Storage temperature -20°C to 70 °C
Altitude Operating up to 2,000 meters
Pollution degree Pollution degree 2
Installation category II

Note
To make accurate measurements, allow the instrument to warm up for 30 min before
operation.

Regulatory Markings
The CE mark indicates that the product
complies with all the relevant European
legal directives. The specific year (if any)
affixed refers to the year when the design
was approved.

The instrument complies with the WEEE


Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard the
electrical/electronic product in domestic
household waste.

This symbol indicates the time period


during which no hazardous or toxic
substances are expected to leak or
deteriorate during normal use. The
expected service life of the product is 10
years. The product can be used safely
during the 10-year Environment Friendly
Use Period (EFUP). Upon expiration of
the EFUP, the product must be
immediately recycled.

Waste Electrical and Electronic Equiment (WEEE)


Directive
2002/96/EC Waste Electrical and Electronic Equipment
(WEEE) Directive
This product complies with the WEEE Directive (2002/96/EC)
marking requirement. This affix product label indicates that you
must not discard the electrical/electronic product in domestic
household waste.
Product Category

Copyright © Itech Electronics Co., Ltd. iii


IT8500+ User’s Manual

With reference to the equipment classifications described in the


Annex I of the WEEE Directive, this instrument is classified as a
“Monitoring and Control Instrument”.
To return this unwanted instrument, contact your nearest
ITECH office.

Copyright © Itech Electronics Co., Ltd. iv


IT8500+ User’s Manual

Compliance Information
Complies with the essential requirements of the following applicable European
Directives, and carries the CE marking accordingly:
 Electromagnetic Compatibility (EMC) Directive 2014/30/EU
 Low-Voltage Directive (Safety) 2014/35/EU

Conforms with the following product standards:

EMC Standard
IEC 61326-1:2012/ EN 61326-1:2013 ¹²³
Reference Standards
CISPR 11:2009+A1:2010/ EN 55011:2009+A1:2010 (Group 1, Class A)
IEC 61000-4-2:2008/ EN 61000-4-2:2009
IEC 61000-4-3:2006+A1:2007+A2:2010/ EN 61000-4-3:2006+A1:2008+A2:2010
IEC 61000-4-4:2004+A1:2010/ EN 61000-4-4:2004+A1:2010
IEC 61000-4-5:2005/ EN 61000-4-5:2006
IEC 61000-4-6:2008/ EN 61000-4-6:2009
IEC 61000-4-11:2004/ EN 61000-4-11:2004

1. The product is intended for use in non-residential/non-domestic environments. Use of the


product in residential/domestic environments may cause electromagnetic interference.
2. Connection of the instrument to a test object may produce radiations beyond the specified
limit.
3. Use high-performance shielded interface cable to ensure conformity with the EMC standards
listed above.

Safety Standard
IEC 61010-1:2010/ EN 61010-1:2010

Copyright © Itech Electronics Co., Ltd. v


IT8500+ User’s Manual

Content
DC PROGRAMMABLE ELECTRONIC LOADS ................................................................................................ 1
QUALITY CERTIFICATION AND ASSURANCE ....................................................................................................................... 1
WARRANTY .............................................................................................................................................................. 1
LIMITATION OF WARRANTY .......................................................................................................................................... 1
SAFETY SYMBOLS ....................................................................................................................................................... 1
SAFETY PRECAUTIONS ................................................................................................................................................. 2
ENVIRONMENTAL CONDITIONS...................................................................................................................................... 3
REGULATORY MARKINGS ............................................................................................................................................. 3
WASTE ELECTRICAL AND ELECTRONIC EQUIMENT (WEEE) DIRECTIVE .................................................................................... 3
COMPLIANCE INFORMATION ......................................................................................................................................... 5
CHAPTER1 INSPECTION AND INSTALLATION ...................................................................................................... 1
1.1 VERIFYING THE SHIPMENT ...................................................................................................................................... 1
1.2 INSTRUMENT SIZE INTRODUCTION ............................................................................................................................ 1
1.3 ADJUSTMENT OF LOAD HANDLE............................................................................................................................... 3
1.4 DISASSEMBLY OF LOAD HANDLE............................................................................................................................... 4
1.5 RACK MOUNTING ................................................................................................................................................ 5
1.6 CONNECTING THE POWER CORD .............................................................................................................................. 5
CHAPTER2 QUICK START .................................................................................................................................... 6
2.1 BRIEF INTRODUCTION............................................................................................................................................ 6
2.2 FRONT PANNEL INTRODUCTION ............................................................................................................................... 7
2.3 VFD DISPLAY ANNUNCIATORS ................................................................................................................................. 8
2.4 FRONT PANEL KEYS............................................................................................................................................... 8
2.5 COMBINATION KEYS.............................................................................................................................................. 9
2.6 REAR PANNEL INTRODUCTION ............................................................................................................................... 10
2.7 POWER-ON SELFTEST .......................................................................................................................................... 11
CHAPTER3 FUNCTIONS AND CHARACTERISTICS .............................................................................................. 14
3.1 LOCAL MODE/REMOTE MODE .............................................................................................................................. 14
3.2 OPERATION MODE ............................................................................................................................................. 14
3.2.1 Constant Current Mode (CC) ............................................................................................................ 14
3.2.2 Constant Voltage Mode (CV)............................................................................................................. 15
3.2.3 Constant Resistance Mode (CR) ...................................................................................................... 15
3.2.4 Constant Power Mode (CW).............................................................................................................. 15
3.3 INPUT ON/OFF CONTROL .................................................................................................................................... 16
3.4 SHORT-CIRCUIT ANALOG FUNCTION ....................................................................................................................... 16
3.5 SYSTEM MENU (SYSTEM) ..................................................................................................................................... 16
3.6 CONFIG MENU (CONFIG) ..................................................................................................................................... 17
3.7 TRIGGER FUNCTION ............................................................................................................................................ 19
3.8 TRANSIENT FUNCTION ......................................................................................................................................... 19
3.8.1 Continuous Mode ............................................................................................................................... 19
3.8.2 Pulse Mode ......................................................................................................................................... 19
3.8.3 Toggled Mode ..................................................................................................................................... 20
3.9 LIST OPERATION................................................................................................................................................ 20
3.10 SAVING AND RECALLING SETTINGS ....................................................................................................................... 20
3.11 VON FUNCTION .............................................................................................................................................. 21
3.12 OCP OPERATION ............................................................................................................................................. 22
3.13 OPP OPERATION ............................................................................................................................................. 23
3.14 BATTERY TEST.................................................................................................................................................. 24
3.15 CR-LED TEST FUNCTION ................................................................................................................................... 26
3.16 MEASUREMENT OF VOLTAGE RISE TIME................................................................................................................. 26
3.17 PROTECTION FEATURES ...................................................................................................................................... 27
3.17.1 Over Voltage Protection (OVP) ....................................................................................................... 27
3.17.2 Over Current Protection (OCP) ....................................................................................................... 27
3.17.3 Over Power Protection (OPP) ......................................................................................................... 28
3.17.4 Over Temperature Protection (OTP) .............................................................................................. 28
3.17.5 Reverse Voltage Protection (LRV) .................................................................................................. 28
3.18 KEY LOCK FUNCTION ......................................................................................................................................... 28
Copyright © Itech Electronics Co., Ltd. vi
IT8500+ User’s Manual

3.19 THE TERMINALS ON THE REAR PANEL .................................................................................................................... 28


3.19.1 Remote Sensing ............................................................................................................................... 28
3.19.2 External Triggering ........................................................................................................................... 29
3.19.3 Current Monitoring (I Monitor) ......................................................................................................... 29
3.19.4Ripple Function .................................................................................................................................. 30
CHAPTER4 BASIC OPERATION .......................................................................................................................... 31
4.1 CONSTANT CURRENT OPERATION ........................................................................................................................... 31
4.2 CONSTANT VOLTAGE OPERATION ........................................................................................................................... 31
4.3 CONSTANT POWER OPERATION ............................................................................................................................. 32
4.4 CONSTANT RESISTANCE OPERATION ........................................................................................................................ 32
4.5 TRANSIENT TEST OPERATION ................................................................................................................................ 32
4.5.1 Continuous Transient Operation ....................................................................................................... 33
4.5.2 Pulse Transient Operation ................................................................................................................. 34
4.5.3 Toggle Transient Operation ............................................................................................................... 34
4.6 LIST OPERATION................................................................................................................................................. 35
4.7 TEST FILES ........................................................................................................................................................ 37
CHAPTER5 COMMUNICATION INTERFACES...................................................................................................... 42
5.1COMMUNICATION MODULES INTRUDUCTION ............................................................................................................ 42
5.2 COMMUNICATION WITH PC .................................................................................................................................. 43
5.3USB INTERFACE.................................................................................................................................................. 45
CHAPTER6 SPECIFICATIONS.............................................................................................................................. 46
SPECIFICATIONS ....................................................................................................................................................... 46
SUPPLEMENTARY CHARACTERISTICS.............................................................................................................................. 54
APPENDIX ............................................................................................................................................................... 55
SPECIFICATIONS OF RED AND BLACK TEST LINES .............................................................................................................. 55

Copyright © Itech Electronics Co., Ltd. vii


Inspection and Installation

Chapter1 Inspection and Installation


1.1 Verifying the Shipment
Unpack the box and check the contents before operating the instrument. If
wrong items have been delivered, if items are missing, or if there is a defect
with the appearance of the items, contact the dealer from which you purchased
the instrument immediately. The package contents include:
Checklist of Package Contents
Item Qty. Model Remarks
The IT8500+ series include:
IT8511+/IT8511A+/IT8511B+/IT85
Electronic Loads x1 IT8500+ series 12+/IT8512A+/IT8512B+/IT8512C
+/IT8512H+/IT8513A+/IT8513C+/I
T8514C+/IT8514B+/IT8516C+
User may select an appropriate
power cord that matches the
IT-E171/IT-E172
specifications of power socket
Power cord x1 /IT-E173/IT-E17
used in the area. See the
4
Section Connecting the Power
Cord for details.
It contains IT8500+ electronic load
User’s Manual, Programming
CD x1 -
Guide and other user
documentations.
Ex-factory Test It is the test report of the
x1 -
Report instrument before delivery.

NOTE
Upon verification of the shipment, keep the package and relevant contents thereof in a
safe place. When returning the instrument for warranty service or repair, the specified
packing requirements shall be met.

1.2 Instrument Size Introduction


The instrument should be installed at well-ventilated and rational-sized space.
Please select appropriate space for installation based on the electronic load
size.
IT8500+ series electronic load different models are not the same size, the detail
size of the electronic load are shown as below.

Copyright © Itech Electronics Co., Ltd. 1


Inspection and Installation

IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512B+/IT8512C+/IT8512H+ Model

Dimension:
Width: 214.5mm
Height: 88.2mm
Depth: 354.6mm

Detialed Dimension Drawing

IT8513A+/IT8513C+ Model

Dimension:
Width: 214.5mm
Height: 88.2mm
Depth: 453.5mm

Detailed Demension Drawing

Copyright © Itech Electronics Co., Ltd. 2


Inspection and Installation

IT8514B+/ IT8514C+ Model

Dimension:
Width: 436.5.5mm
Height: 88.2mm
Depth: 463.5mm

Detialed Dimension Drawing

IT8516C+ Model

Dimension:
Width: 482.5mm
Height: 174.5mm
Depth: 531.5mm

Detailed Dimension Drawing

1.3 Adjustment of Load Handle


IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512A+/IT8512B+/IT8512C+/IT8512H
/IT8513A+/IT8513C+ series loads are equipped with a handle for user to easily
carry and place it.

Copyright © Itech Electronics Co., Ltd. 3


Inspection and Installation

The load handle may be adjusted based on three methods (as shown in icons
below). Be sure that appropriate force is applied to adjust the load handle to
appropriate position.

1.4 Disassembly of Load Handle


Please disassemble the load handle before installing equipment on the support.
Disassembly steps:
1. Adjust the handle to the position as shown in the figure below.

NOTE
To easily disassemble handle, align the locking mouth and locking device, which is between
the handle and the instrument.
2. Align the locking mouth, and pull out the handle towards two sides.

NOTE
Do not use too much force and mind your hands during disassembly of load handle.

Copyright © Itech Electronics Co., Ltd. 4


Inspection and Installation

1.5 Rack Mounting


IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512A+/IT8512B+/IT8512C+/IT8512H
+/IT8513A+/IT8513C+ loads can be installed on standard 19-inch rack. ITECH
provides user with IT-E151/IT-E151A rack, as an optional mount kit. The
detailed operation please refer to the User Manual of your mount kit.
IT8514B+/IT8514C+/IT8516C+ need not mount on rack, they can installed on
cabinet directly by screw.

1.6 Connecting the Power Cord


Connect the power cord after checking that the power switch of the instrument
is turned OFF. Only use the power cord supplied as a standard accessory.
Select from the following Schedule of Power Cord Specifications an appropriate
power cord that matches the voltage for the area in which you use the
instrument. If the power cord included in the instrument you purchased does
not match the voltage, contact the dealer or manufacturer for change.

E
E E E
N L N L
L N
L N

China United States & Europe England


Canada
IT-E171 IT-E173 IT-E174
IT-E172

Copyright © Itech Electronics Co., Ltd. 5


Quick Start

Chapter2 Quick Start


This chapter introduces the front panel, the rear panel, key functions and VFD
display function of the electronic load, make sure that you can quickly know the
appearance, instruction and the key function before you operate the load, Help
you make better use of this series of electronic load.

2.1 Brief Introduction


IT8500plus series DC electronic loads are single channel programmable
electronic load which can provide multiple solutions according to the
requirements of your design and test. This series have international advanced
functions and features.
 High-visibility vacuum fluorescent display (VFD)
 Measurement resolution: 0.1mV,0.1mA
 Voltage and current Measurement speed: up to 40KHZ
 Four operation modes:CV(Constant Voltage),CC,CR,CW
 Battery test function
 OCP test , OVP test
 Auto test function:
 Short circuit function
 Remote Sense function
 Memory capacity to save/recall setting parameters: 100 registers
 Intelligent fans
 Build-in Buzzer as alarm signal
 Power off memory function
 Rotary knob, making the operation more convenient
 Measure test function,test the rising/dropping time of the voltage
 List modes
Model Voltage Current Power Communication
Interface
IT8511+ 120V 30A 150W DB9(TTL)
IT8511A+ 150V 30A 150W DB9(TTL)
IT8511B+ 500V 10A 150W DB9(TTL)
IT8512+ 120V 30A 300W DB9(TTL)
IT8512A+ 150V 30A 300W DB9(TTL)
IT8512B+ 500V 15A 300W DB9(TTL)
IT8512C+ 120V 60A 300W DB9(TTL)
IT8512H+ 800V 5A 300W DB9(TTL)
IT8513A+ 150V 60A 400W DB9(TTL)
IT8513C+ 120V 120A 600W DB9(TTL)
IT8514B+ 500V 60A 1500W USB/RS232
IT8514C+ 120V 240A 1500W standard USB/RS232
IT8516C+ 120V 240A 3000W Standard USB/RS232

Copyright © Itech Electronics Co., Ltd. 6


Quick Start

2.2 Front Pannel Introduction


IT8500+ series electronic load different models have different front pannels, the
front panels and keyboards of different models are shown as below.

IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512A+/IT8512B+/IT8512C+/IT8512H+/IT8513A+/IT
8513C+ Model

① Power switch ⑤ Function key:


② vacuum fluorescent display (VFD) Set the operation mode
③ Compound key and the local switch key Control the input state: On/Off
④ Number key: ⑥ Direction function
Set the parameters value, ⑦ Rotary knob
achieve the menu’s function by key ⑧ Input terminal
combination

IT8514B+/IT8514C+ Model

① Power switch ⑤ Function key:


② vacuum fluorescent display (VFD) Set the operation mode
③ Compound key and the local switch key Control the input state: On/Off
④ Number key: ⑥ Direction function
Set the parameters value, ⑦ Rotary knob
achieve the menu’s function by key ⑧ Input terminal
combination

Copyright © Itech Electronics Co., Ltd. 7


Quick Start

IT8516C + Model

① Power switch ⑤ Function key:


② vacuum fluorescent display (VFD) Set the operation mode
③ Com pound key and the local switch key Control the input state: On/Off
④ Num ber key: ⑥ Direction function
Set the parameters value, ⑦ Rotary knob
achieve the menu’s function by key ⑧ Input terminal
combination

2.3 VFD Display Annunciators


OFF The load is off. Error An error has occurred.
CC Constant current Trig Waiting for the trigger
mode signal
CV Constant voltage Sense Remote sensing is on.
mode
CR Constant resistance Prot OCP function is on.
mode
CW Constant power mode Auto Voltage range automatically
seleted function is open
Rmt Instrument is in the Lock The keyboard is locked.
remote state.
Timer LOAN ON is on. Shift Shift button has been
pressed.

2.4 Front Panel Keys

( Blue-green) Shift button is a composite key.


Copyright © Itech Electronics Co., Ltd. 8
Quick Start

(Gray) Local button is used to switch local and remote


mode.
(Gray-white) Power on button
0 ~ 9 Enter the digits 0 to 9.
Decimal point
ESC
The escape key
Choose constant current mode.
Choose constant voltage mode.
Choose constant resistance mode.
Choose constant power mode.
Enter the selected value or setting.
Turns DC Load ON or OFF (OFF is high
impedance state).
Scroll up key
Scroll down key
Scroll left and right key

2.5 Combination Keys


Press [ Shift ] button first and then other keys to achieve all kinds functions in
the following table.
[ Shift ]+ [ 1 ] (Short) Turn short circuit on or off.
[ Shift ]+ [ 2 ] (Tran) Start or stop transient condition.
[ Shift ] + [ 3 ] (List) Set LIST operation parameters.
[ Shift ] + [ 4 ] (Save) Store the DC Load state in non-volatile
memory.
[ Shift ] + [ 5 ] (Battery) Turn on or off battery testing function.
[ Shift ] + [ 6 ] (Prog) Enter auto test function.
[ Shift ] + [ 7 ] (Info) Display product’s Model/SN/Version.
[ Shift ] + [ 8 ] (System) System menu setting
[ Shift ] + [ 9 ] (Config) Configure menu setting
[ Shift ] + [ 0 ] (Pause) Press this button if you need a pause
when runing an auto test file.
[ Shift ] + [ ] (Trigger) Cause an immediate trigger.
[ Shift ] + [ CC ] (OCP) Enter OCP test function.
[ Shift ] + [ CV ] (Setup) Set detailed parameters in
CC/CV/CW/CRmode.
[ Shift ] + [ CW ] (OPP) Enter OPP test function.
[ Shift ] + [ Enter ] (Recall) Recall the DC Load state from
non-volatile memory.
[ Shift ] + [ On/Off ] (Lock) Key lock function

Copyright © Itech Electronics Co., Ltd. 9


Quick Start

2.6 Rear Pannel Introduction


IT8500+ series electronic load different models have different rear pannels, the
rear panels and keyboards of different models are shown as below.
IT8511+/IT8511A+/ IT8511B+/IT8512+/ IT8512B+/IT8512C+/IT8512H/IT8513A+/IT8513C+ Model

①Thermal window ④Current monitoring Terminal


②Line voltage selection ⑤9-Pin serial port interface connector
switch( 110V/220V ) ⑥4 pin trigger and remote sensing connector
③3 pin IEC320 AC input connector

IT8514B+/IT8514C+Model

①Thermal window ⑤RS232 communication cable interface


②Line voltage selection ⑥4 pin trigger and remote sensing connector
switch( 110V/220V ) ⑦USB communication cable interface
③3 pin IEC320 AC input connector
④Current monitoring Terminal

Copyright © Itech Electronics Co., Ltd. 10


Quick Start

IT8516C+Model

①Thermal window ⑤RS232 communication cable interface


②Line voltage selection ⑥4 pin trigger and remote sensing connector
switch( 110V/220V ) ⑦USB communication cable interface
③3 pin IEC320 AC input connector
④Current monitoring Terminal

2.7 Power-on Selftest


A successful test process indicates that the instrucment meets the factory
specifications and can be operated well.
Before operation, please confirm that you have fully understood the safety
instructions.

 To avoid burning out, be sure to confirm that power voltage matches with
supply voltage.
 Be sure to connect the main power socket to the power outlet of protective
grounding. Do not use terminal board without protective grounding. Before
operation, be sure that the power supply is well grounded.
 To avoid burning out, pay attention to marks of positive and negative
polarities before wiring.

Selftest steps
Normal selftest procedures:
1. Correctly connect the power cord. Press [ Power ] key to start up.
2. After selftest, VFD display information below.

0.0000V 0.0000A
0.00W CC=0.000A
OFF CC Auto

Information description:
 The first line display actual voltage and current value.
 The second line display the actual power value and the setting
current/voltage/power/resistance value
Copyright © Itech Electronics Co., Ltd. 11
Quick Start

 The third line display the input state/operation mode.


3. Press [ Shift ] + [ 7 ], VFD display products information. You can press
direction buttons to examine product’s model/SN/software version.

Error Information References


The following error information may occur when an error occurs during Power
On self-test:
 If the EEPROM was damaged, the VFD will display “Eeprom Fail”.
 If the lastest operation data in EEPROM is lost, then VFD will display
“Config Data Lost”.
 If the calibration data in EEPROM is lost, then VFD will display “Cal data
lost”.
 If the system setting data in EEPROM is lost, the VFD will display “Eeprom
data lost”.Please press [ Shift ] + [ 4 ] and [ 0 ] to save after setting
parameters.

Exception handling
If the electronic load cannot start normally, please check and take measures by
reference to steps below.
1. Check whether the power line is correctly connected and confirm whether
the electronic load is powered.
Correct wiring of power line => 2
Incorrect wiring of power line => Re-connect the power line and check
whether the exception is removed.

2. Check whether the power in On.[ Power ] key is under “ ” On


status.
Yes => 3
No => Please check the [ Power ] key to start power and check whether
the exception is removed.
3. Check whether set power voltage of electronic load is larger than the power
supply voltage. If set power voltage is 220 V and the supply voltage is 110V,
the electronic load cannot start.
4. Check whether the fuse of electronic fuse is burned out.
If yes, change fuse. Detailed steps:
 Pull out power line and take out the fuse box at power line jack with a
small screw driver. As shown below.

 If the fuse is fused, please change fuse of same specification based on


machine model. See the table blow for matching information of fuse
and machine model.
Fuse specification Fuse specification
Model
(220VAC) (110VAC)
IT8511+ T0.5A 250V T1.25A 250V
IT8511A+ T0.5A 250V T1.25A 250V
IT8511B+ T0.5A 250V T1.25A 250V

Copyright © Itech Electronics Co., Ltd. 12


Quick Start

Fuse specification Fuse specification


Model
(220VAC) (110VAC)
IT8512A+ T0.5A 250V T1.25A 250V
IT8512A+ T0.5A 250V T1.25A 250V
IT8512B+ T0.5A 250V T1.25A 250V
IT8512C+ T0.5A 250V T1.25A 250V
IT8512H+ T0.5A 250V T1.25A 250V
IT8513A+ T1.25A 250V T2.5A 250V
IT8513C+ T1.25A 250V T2.5A 250V
IT8514B+ T1.25A 250V T2.5A 250V
IT8514C+ T1.25A 250V T2.5A 250V
IT8516C+ T2.5A 250V T5A 250V

 After replacement, install the fuse box back to original position, as


shown below.

Copyright © Itech Electronics Co., Ltd. 13


Functions and Characteristics

Chapter3 Functions and Characteristics


This chapter elaborates on the functions and characteristics of electronic
loads.Contents following sections:
 Switching of local/remote operation modes
 Constant-status operation mode
 Input On/Off function
 Keyboard locking function
 Short -circuit analog function
 System setup function
 Triggering function
 List mode
 Test function
 Save/Recall
 VON function
 Full protection function,OCP,OVP,OTP,OPP,Reverse voltage
 Remote Sense function
 Current monitoring function
 Ripple function

3.1 Local Mode/Remote Mode


There are two types of control modes for IT8500+ series products:Local mode
and Remote mode.
In remote mode, you can operate the electronic loads through PC via
communication cable(optional). While After power on electronic loads, it
defaults in local mode and all buttons are avaiable in this mode.In remote
control mode, the keys on the front pannel can not work except local
key.Customers could through [ Local ] key to switch the control mode.

3.2 Operation Mode


There are four operation modes of IT8500+ series products:
 Constant current mode (CC)
 Constant voltage mode (CV)
 Constant resistance mode (CR)
 Constant power mode(CW)

3.2.1 Constant Current Mode (CC)


In constant current mode, the DC load will comsume a constant current,
regardless of the voltage at its terminals.

Copyright © Itech Electronics Co., Ltd. 14


Functions and Characteristics

I
Load Setting current
current

Input voltage V

CC mode

Diagram 3-1 I-V curve in CC mode

3.2.2 Constant Voltage Mode (CV)


In constant voltage mode, the DC load will cause a constant voltage to appear
at its terminals.

V
Setting voltage
Input voltage

Input current I

CV mode
Diagram 3-2 I-V curve in CV mode

3.2.3 Constant Resistance Mode (CR)


In constant resistance mode, the DC load will behave as a fixed resistance
value.As shown below,the load linearly changes the current value with the
rising of input voltage.
I Slope is equal
Load To the setting
current resistance

Input voltage V

CR mode

Diagram 3-3 I-V curve in CR mode

3.2.4 Constant Power Mode (CW)


In constant power mode,the DC load will cause a constant power to be
dissipated in the load.As shown below,the load current is decreasing with the
rising of input voltage,while power always maintain the setting value.

Copyright © Itech Electronics Co., Ltd. 15


Functions and Characteristics

Diagram 3-4 I-V curve in CP mode

3.3 Input On/Off Control


[ On/Off ] button on the front panel is used to manually toggle the instrument
between its set mode and an infinite impedance state, i.e control input on and
off. [ On/Off ] button lighted indicates the load input is on, meanwhile the [ OFF ]
indicator will disappear.

3.4 Short-circuit Analog Function


Short circuit simulation and short circuit current measurement:you may press
[ Shift ] + [ 1 ] botton to emulate a short state.It can be used to check whether
the tested instrument’s short protection is avaiable.
In short mode,the DC load will draw maximum current from the DC supply in
any of the four operation modes(CC,CV,CW or CR).In CC,CV,or CR mode,you
may press [ Shift ] + [ 1 ] to stop short.The DC load will return to its previous
operation.However,in CW mode,the short current will continue to be drawn.To
stop the short,you must press the [On/Off] key after you press [ Shift ] + [ 1 ].
When emulating a Short in CC, CW or CR mode, the maximum allowable short
current is equal to the 110% of current range. Under CV mode, short circuit
current is equivalent to that constant voltage value of load is 0 V.

3.5 System Menu (System)


Press [ Shift ] + [ 8 ] (system) to enter the system menu.
POWER-ON Power on state of instrument
RST(default) Do not remember state in SAVE 0.Customer can save
POWER-ON a often used data in SAVE 0 to recall when power on
the DC load next time.
SAV0 Remember state in SAVE 0
BUZZER
BUZZER ON(default) Enable audible beep when key is pressed
OFF No sound when key is pressed
KNOB KNOB

Copyright © Itech Electronics Co., Ltd. 16


Functions and Characteristics

UPDATE(default) The value modified with knob during operation will be


saved after load is off.For example; the DC load is set
to 1A by press [CC] and turned on the input.Then
increase the setting value to 2A with knob.When
customer turn off load and trun on again, the setting
value changes to 2A.

OLD As explained above, after the DC load is turned on


again,the setting value is 1A instead of 2A changed
with knob.
SOURCE Set trigger mode
MANUAL(Def) Triggered from the [ Shift ] + [ ] key
EXTERNAL Triggered from a TTL high signal at the trigger
TRIGGER
connector on rear panel
BUS Triggered from a serial bus command 5AH
HOLD Receving a command 9DH
MEMORY Recall the prestored datas
MEMORY GROUP= 0 0:indicates1-10 group;1:indicates 11-20group, by parity
of reasoning
DISP-TIMER Timer function

DISPLAY ON Enable timer function


OFF(default) Disable timer function
RS-232
4800_8N 1 Baudrate 4800, data bit 8, none parity, stop bit 1
RS-232 9600_8N 1 Baudrate 9600, data bit 8, none parity, stop bit 1
19200_8N 1 Baudrate 19200, data bit 8, none parity, stop bit 1
38400_8N 1 Baudrate 38400, data bit 8, none parity, stop bit 1
SCPI Select SCPI protocol
PROTOCOL
FRAME Select FRAME protocol
ADDRESS ADDRESS= 0 Set the instrument’s address(0~31)
RUN Runing mode at power on
NORMAL Normal mode
BATTERY Default in battery test mode at power on
RUNMODE
PROG_TEST Default in autotest mode at power on
OCP_TEST Default in OCP test mode at power on
OPP_TEST Default in OPP test mode at power on
DEFAULT DEFAULT
NO Do not return instrument to factory default settings.
YES Retrun instrument to factory default settings

3.6 Config Menu (Config)


Press [ Shift ]+ [ 9 ] (Config) to enter the menus.

Max-P Set hardware power protection


PROTECT
MAX POWER=150.00W Set hardware OPP value

Copyright © Itech Electronics Co., Ltd. 17


Functions and Characteristics

A-LIMIT Set software current protecting state


A-LIMIT
ON Enable software over current protection function
A-LIM POIN=30.000A Set the software OCP level
A-LIM DELAY=3S Set the OCP delay time
OFF Disable the software OCP funtion
P- LIMIT Set software power protecting state.

P-LIM POIN=150.00W Set the software OPP level.

P-LIM DELAY=3S Set the OPP delay time.

TIMER Set load on timer


LOAD-TIMER
ON Enable load-on timer
LOAD-TIMER=10.0S Set the load on time(0.1S~9999.9S)
OFF Disable load on timer
V-RANGE Voltage auto-rangefuntion
V-RANGE
ON Enable voltage auto range function
OFF Disable voltage auto range function

MEASURE FILTER Set the filter parameter


FILTER COUNT = 2^14 Filter count set, range 2~16
TIME-V1
TIME-VOLT1=0.000V Set the start time, to measure the voltage rise/fall time.
TIME-V2
TIME-VOLT2=120.00V Set the end time, to measure the voltage rise/fall time

CR-LED Imitate LED (in CR mode)

CR-LED Open the function(in CR mode,press [ Shift ]+ [ CV ] to


ON
set Vd value)
OFF Disenable the function
REM- SENSE Remote sense function
SENSE ON Enable remote sense function
OFF Disable remote sense function
VON Set the load’s VON point
LIVING VON point living state
VON VON POINT = 0.10V Set the VON value
LATCH VON point latch state, ON /OFF
VON POINT = 0.10V Set the VON value
RESET Reset the config menu
RESET NO Do not reset
YES Reset

Copyright © Itech Electronics Co., Ltd. 18


Functions and Characteristics

3.7 Trigger Function


Triggering is used with the transient operation; list operation and test
function.There are four types of triggers you can use for IT8500+ products.
Manual: An immediate trigger is created by pressing [ Shift ]+ [ ] (Trigger)
on the front panel.
External(TTL signal): An external trigger is a TTL low signal applied to the
trigger connection on the rear panel.This TTl signal must last for more than
10us.
Bus: The instrument will be triggered if command 5AH is sent via the
communication interface.
Hold: The instrument will be triggered if command 9DH is sent via the
communication interface.

3.8 Transient Function


The transient test allows switching between two different load values.A
common application is to test the dynamic characteristics of DC source.
There are three different types of transient operation: continuous, pulse,
toggled.

3.8.1 Continuous Mode


In continuous transient operation, the load is continuously switched between
two load values.An example is shown in the following figure:

3.8.2 Pulse Mode


In pulse operatrion,the load operates at the A value that has been entered until
a trigger is received.At the trigger,the load switches to the B value and stays at
that level for the B timing value.Then the load switches back to the A value and
stays there until another trigger is received.

Copyright © Itech Electronics Co., Ltd. 19


Functions and Characteristics

3.8.3 Toggled Mode


In toggled transient operation,the load starts at the stored parameters for the
mode.When a trigger is received,the load switches to B value.When another
trigger is received,the load switches to the A level.It stays at the A value until
another trigger is received,at which point it switches to the B value.Here’s an
example:

10A

5A

TRG TRG

Toggled Transient Operation

3.9 LIST Operation


List mode allows you to generate a complex current sequence. Moreover, the
mode.
change can be synchronized with an internal or external signal, to accomplish
dynamic and precise test A list file includes following parameters: file name,
step counts (range 2-84), time width of single step(0.00005s~3600s),step
value and slope.The edited list file can be saved in nonvolatile memory, can be
recalled easily.The DC load provides 7 nonvolatile registers to save list files for
recall later.
In list mode,the DC load start to run the list file once receiving a trigger signal,
continue to run once receiving another trigger signal.To illustrate the use of a
list,we’ll create a list that runs the following constant current profile on a power
supply:

Trigger

0 1 2 3 4 5
List count=1 List count=2

List sequence

3.10 Saving and Recalling Settings


We can save some often-used parameters in the non volatile memory, including
working mode, voltage/current value and so on.IT8500plus series provide 100
non-volatile registers.
They are divided into 10 Memory groups: Group0-9.You can set it in the system
menu.Group0 means you can save and recall parameters in 0-10 registers.

Copyright © Itech Electronics Co., Ltd. 20


Functions and Characteristics

Group1 means you can save and recall parameters in 11-20 registers.
Group2-Group9 can be concluded in the same manner.

Save and Recall operation


For example: the instrucment works in CC mode, setting value is 1A,
Memory Group is 6. Save “CC 1A” in the 61th register and then recall.
1. Set the parameters ok. To save the instrument's settings to a register,
press [ Shift ]+ [4] (Save). Enter number [6] and [1].
2. Then press [Enter], The setting is saved.
3. To recall the instrument's settings from a register, press [ Shift ]+ [Enter]
(Recall).
4. Enter [1].Then the setting is recalled.

NOTE
Saving operation will overwrite any values previously saved in that register.
Recalling operation will light the Enter . You should press [ESC] to escape the recalling state
before setting other parameters.

3.11 VON Function


The DC load can be set to only turn on if the voltage is above a set value(VON
set) under configure menu by pressing [ Shift ]+ [9].There are two types of
VON function:Living and Latch.The following will have detailed description for
the two types.

NOTE
VON set is used to ensure an electronic system under test will not have power applied
unless the supply voltage is above a certain value.If you have no such testing request, do
not set this value arbitrarily.If your instrument can not work normally, for example, set CC=1A,
after turn on the input while the current is still 0A instead of setting value 1A, then you should
check VON set firstly.If VON set is not 0V, then please modify to 0V.

VON LIVING MODE


In Living mode, when power is applied to the DC load, the voltage must rise
above VON setting before the load draws current from the source.If the voltge
below VON setting on the load’s terminals, the load will turn off input.
Copyright © Itech Electronics Co., Ltd. 21
Functions and Characteristics

VON LATCH MODE


In Latch mode,as before,the load will turn on only when the voltage exceed
VON setting,but once on,it will now stay on,even if the voltage drops to zero.

3.12 OCP Operation


OCP test process: After input voltage reaches VON point,the DC load start to
draw a current from the source after a delay time.The current value will
increase by a certain step size at regular intervals.Simultaneously,the DC load
will judge whether the input voltage is lower than OCP voltage you’ve set.If it
is,then the present current value will be compared to see if it is in the current
range you’ve set,in this range,the OCP test will Pass or fail.On the contrary,the
DC load will continue to increase drawing current and compare the voltage.
To start an OCP test,press [ Shift ]+ [ ] (Trigger).
Press [ Shift ]+ [ CC ] (OCP) to enter OCP operation.
OCP TEST
1.VON LEVEL=0.000V Set Voltage threshold
2.VON DELAY=0.00S After delay certain time,the DC
load starts to draw current.
3.RANGE=3.000A Set current range
4.START=0.1000A Set start current
EDIT 5.STEP=0.1000A Set step current
6.STEP DEL=0.20S Set delay time of each step
7.END=2.0000A Set end current
8.OCP VOLT=2.000V Set OVP value
9.MAX TRIP =1.5000A Upper limit of OCP value
10.MIN TRIP=0.9000A Lower limit of OCP value
SAVE OCP FILE=1 Save OCP test file (1-10)

Set the power on mode to be OCP test mode:


Operation Display on front pannel

1.Press [ Shift ]+ [ 8 ] (system) enter into sysmtem 0.0000V 0.000A


menu POWER-ON BUZZER

Copyright © Itech Electronics Co., Ltd. 22


Functions and Characteristics

2.Press right key,select RUNMODE and confirm 0.0000V 0.000A


with [ Enter ] button RUN <NORMAL
3.Press direction key to select OCP_TEST,Press 0.0000V 0.000A
[ Enter ] to confirm. RUN <OCP_TEST
4,Press [ Esc ] button to quit the set. 0.0000V 0.000A
STOP 0.000A

After above steps,press [ ] (Trigger) button to run ocp test file.

Recall OCP File:


1. Press [ Shift ]+ [ Enter ] button to select programe file,the panel displays
“CALL OCP FILE= 1.Enter the file name(1-10),press [ Enter ] button to
confirm.
2. According to the following steps to escape OCP mode:press [ Shift ]+ [ 8 ]
(system)------“RUNMODE”-----[Enter]-------select
“NORMAL”mode----[ Enter ].

3.13 OPP Operation


OPP test process:When the input voltage has reached VON point, power will
begin to work after a delay time.The power value will increase by a step size at
regular intervals.Simultaneously,the DC load will judge wether the input voltage
is lower than OPP voltage(you need to set).If it is,then the present current value
will be compared to see if it is in the current range you’ve set,in this range,the
OPP test will Pass or fail.On the contrary,the power will continue to increase
within the cut-off current range.And then compare OPP voltage with input
voltage too.
To start a OPP test, press [ Shift ]+ [ ] (Trigger).
Press [ Shift ]+ [CW] (OPP) to enter OPP test operation.
OPP TEST
RUN
STOP Run OPP test file
OPP TEST
CALL
Recall OPP File=1 Recall OPP test file(range file1-file10)
OPP TEST
1.VON LEVEL=0.000V Set Voltage on value
2.VON DELAY=0.01S Set Voltage on delay time
3.RANGE=5A Set working current range
OPP TEST
4.START =0.1W Set start power value
5.STEP =1W Set step power value
EDIT
6.STEP DEL=1S Set step delay time
7.END =12W Set cut-off power value
8.OPP VOLT=7V Set OPP value
9.MAX TRIP =6.5W Upper limit of OPP value
10.MIN TRIP =5.6W Lower limit of OPP value
SAVE OPP FILE=1 Save OPP test file

Set the power on mode to be OPP test mode


Operation Display on front pannel

1.Press [ Shift ]+ [ 8 ] (system) enter into 0.0000V 0.000A


sysmtem menu POWER-ON BUZZER
2.Press right key,select RUNMODE and 0.0000V 0.000A
confirm with [ Enter ] button RUN <NORMAL

Copyright © Itech Electronics Co., Ltd. 23


Functions and Characteristics

3.Press direction key to select 0.0000V 0.000A


OCP_TEST,Press [ Enter ] to confirm. RUN <OPP_TEST
4,Press [ Esc ] button to quit the set. 0.0000V 0.000A
STOP 0.000A

After above steps,press [ ] (Trigger) button to run OPP test file.

Recall OPP File


1. Press [ Shift ]+ [ Enter ] button to select programe file,the panel displays
“CALL OPP FILE= 1.Enter the file name(1-10),press [ Enter ] button to
confirm.
2. According to the following steps to escape OPP mode:press [ Shift ]+ [8]
(system)------”RUNMODE”----[ Enter ]------select“NORMAL”mode----[ Ente
r ].

3.14 Battery Test


IT8500plus series products test the battery capability in CC/CW/CR mode.
The test mode should be set first, and then the discharge stop conditions.
There are three discharge stop conditions to be set for IT8500 plus series
products. When user only need to do battery testing in one or two stop
conditions, the other conditions should be set to the specified value (STOP
VOLT:0V;STOP CAP:999.999AH;STOP TIMER:99999S). When the system
checks the discharging time or battery voltage or capacity is equal to the setting
stop value or under an insecurity state, the battery test will stop, and the E-Load
will turn off. The battery voltage, discharge current, discharge time and
discharged capability are displayed on the VFD while testing.
Take CC mode for example, the operations are as below:
( 1 )Voltage-threshold Cut Off
Step Operation Display
1 Press [ Shift ]+ [ 5 ] (Battery),set current range 0.0000V 0.000A
RANGE = 0.00A
2 Set discharge current,for example 2A CURRENT = 2.000A
3 Set the stop voltage,for example 2V, then press STOP VOLT=2V
[ Enter ] to confirm.
4 Set the stop capability to maximum 999.999AH, STOP CAP=999.999AH
press[ Enter ] to confirm.
5 Set the stop timer to maximum 99999S, press STOP TIMER=99999S
[ Enter ] to confirm.
6 Save the battery test to specified file 0.0000V 0.000A
SAVE BATT FILE 1(1-10)
6 Press [ Enter ] to confirm 0.0000V 0.000A
0.00W I = 2.00A
Off cc

(2)Capacity-threshold Cut Off


Step Operation Display
1 Press [ Shift ]+ [ 5 ] (Battery),set current range 0.0000V 0.000A
RANGE = 0.00A
2 Set discharge current,for example 2A CURRENT = 2.000A
3 Set the stop voltage to 0V, then press [ Enter ] to STOP VOLT=0V
confirm.
4 Set the stop capability, for example,7AH,then STOP CAP=7AH
press [ Enter ] to confirm.
5 Set the stop timer to maximum 99999S, press STOP TIMER=99999S
Copyright © Itech Electronics Co., Ltd. 24
Functions and Characteristics

[ Enter ] to confirm.
6 Save the battery test to specified file 0.0000V 0.000A
SAVE BATT FILE 1(1-10)
7 Press [ Enter ] to confirm 0.0000V 0.000A
0.00W I = 2.00A
Off cc

(3)Time-out Cut Off


Step Operation Display
1 Press [ Shift ]+ [ 5 ] (Battery),set current range 0.0000V 0.000A
RANGE = 0.00A
2 Set discharge current,for example 2A CURRENT = 2.000A
3 Set the stop voltage to 0V, then press [ Enter ] to STOP VOLT=0V
confirm.
4 Set the stop capability to maximum 999.999AH, STOP CAP=999.999AH
press [ Enter ] to confirm.
5 Set the stop timer, for example, 3800S, press STOP TIMER=3800S
[ Enter ] to confirm.
6 Save the battery test to specified file 0.0000V 0.000A
SAVE BATT FILE 1(1-10)
7 Press [ Enter ] to confirm 0.0000V 0.000A
0.00W I = 2.00A
Off cc

(4) Any of the Three Conditions Cut Off


Step Operation Display
1 Press [ Shift ]+ [ 5 ] (Battery),set current range 0.0000V 0.000A
RANGE = 0.00A
2 Set discharge current,for example 2A CURRENT = 2.000A
3 Set the stop voltage as needed,for example,2V,then STOP VOLT=2V
press [ Enter ] to confirm.
4 Set the stop capbility as needed,for example,7AH,then STOP CAP=7AH
press [ Enter ] to confirm.
5 Set the stop timer as needed,for example,3800S,then STOP TIMER=3800S
press [ Enter ] to confirm.
6 Save the battery test to specified file 0.0000V 0.000A
SAVE BATT FILE 1(1-10)
7 Press [ Enter ] to confirm 0.0000V 0.000A
0.00W I = 2.00A
Off cc

(5)Go into Battery Test Mode


Operation Display on front pannel

1.Press [ Shift ]+ [ 8 ] (system) enter into 0.0000V 0.000A


sysmtem menu POWER-ON BUZZER
2.Press right key,select RUNMODE and 0.0000V 0.000A
confirm with [ Enter ] button RUN <NORMAL
3.Press direction key to select 0.0000V 0.000A
OCP_TEST,Press [ Enter ] to confirm. RUN <BATTERY
4,Press [ ESC ] button to quit menu set 0.0000V 0.000A
0S 0.000AH
(6)Start Battery Test
Press [ ] (trigger) to provide a signal to start battery test.The discharing
process will be auto terminated when stop conditions are reached.
(7)Recall Battery File
Press [ Shift ] + [ Enter ] button to select programe file,the panel displays
“RECALL BATTERY 1.Enter the file name(1-10),press [ Enter ] button to
Copyright © Itech Electronics Co., Ltd. 25
Functions and Characteristics

confirm.
(8)Pannel Locked in case of Error Operations
Press [ Shift ]+ [ On/Off ] (Lock) button to lock the panel.In this mode,only [ Shift ] and [ On/Off ]
button is enabled.
According to the following steps to escape OPP mode:
Press [Shift ]+ [ 8 ] (system)---”RUNMODE”---[ Enter ]----select “NORMAL” mode---[ Enter ]

3.15 CR-LED Test Function


With adding of diode break-over voltage setting in the IT8500+ series electronic
load under conventional CR mode, the electronic load only works when
voltage applied at its both ends is higher than the diode break-over voltage to
give a real simulation of diode working principle, i.e., the ripple current at real
LED test.
Detailed steps of LED power test:
1. Start CR-LED Function
Press [Shift] + [9] keys to enter configuration menu. Press Right Key and
select “CR-LED”. Press [Enter] key for entry. Select “on” and press [Enter]
key. Press [ESC] key to exit.
2. Set CR Mode and Resistance Value
Press [CR] key and set corresponding constant resistance (as R calculated
below).
3. Set Vd Value
Press [Shift] + [CV] keys for a series of related setting: range=7500.0,
high=130V, low=0V, which may remain the original values. Vd will be set
based on the calculation below.
Calculation method of Vd and R:
Vd=V*0.8 R=0.2V/I
Where:
 V: constant working voltage of load LED of LED constant current
source;
 I: output current of LED constant current source;
 Vd: break-over voltage of diode (string);
 R: constant resistance;

3.16 Measurement of Voltage Rise Time


The IT8500+ series electronic load is provided with special voltage rise/drop
time measurement function. This function gives a simple analog of voltage
rise/drop speed of oscilloscope test power.
Operation methods:
Set initial Voltage and Final Voltage
1. Press [Shift] + [9] keys to enter configuration menu. Press Right key.
Select “Measure” and press [Enter] key.
2. Press [ ] to select “TimeV1”. Press [Enter] key. Press numeric
keys to set initial voltage value and press [Enter] key.
3. Press [ ] to select “TimeV2”. Press [Enter] key. Press numeric
keys to set final voltage value and press [Enter] key.
4. Press [ESC] to exit setting.
Start timer function

Copyright © Itech Electronics Co., Ltd. 26


Functions and Characteristics

5. Press [Shift] +[8] keys to enter system menu. Press Right key till “Displ”
flicks and press [Enter] key.
6. Press [ ] key to select “On”. Start timer function and press [Enter]
key.
7. Press [ESC] to exit setting.
8. VFD second line will display time 0.0000S between power value and set
value.
OFF CC
0.0001V 0.0002A
0.00W 0.0000S CC=0.000A

Measurement of Rise Time


9. Connect DC power to be tested to the input terminal of the electronic load.
The power is set with a value that is higher than the set final voltage value.
Keep power output in OFF status.
10. Set a constant current value on the load and open the load input.
11. Open power output.
12. The electronic load timer starts timing. After ending, time will keep stable,
which is rise time of voltage.
13. Close the power output. The electronic load VFD will display voltage drop
time.

3.17 Protection Features


DC load protection features include: OVP, OCP, OPP, OTP, reverse
voltageprotection(LRV/RRV).

3.17.1 Over Voltage Protection (OVP)


If input voltage exceeds the voltage limit set by the user, the DC load will turn
the input OFF and the buzzer will sound. The display will show OVP.
Operations to clear the OVP state
Disconnect the instrument under test. Press any key on the front panel, the
OVP on the VFD will disappear, then the DC load exits OVP protection state.

3.17.2 Over Current Protection (OCP)


The DC load includes both hardware and software over current protection
features.
Hardware OCP: maximum input current of the DC load will be limited at about
110% of the current range, once the hardware OCP is activated, the status
register’s OC bit will be set; when the hardware OCP is removed, the status
register’s OC bit will be reset. Hardware over current protection won’t change
the DC load’s On/Off state.
Software OCP: users can set the DC load’s software OCP value, steps:
[Shift] +[9] > Protect > Alimit set ON, Apoint set OCP current value, Adelay
set delay time before alarm. When the software OCP function is actived, the DC
load will automatically turn off, VFD displays OCP.
Operations to clear the OCP state
Disconnect the instrument under test. Press any key on the front panel, the
OCP displayed on the VFD will disappear, the DC load exits OCP protection
state.
Copyright © Itech Electronics Co., Ltd. 27
Functions and Characteristics

3.17.3 Over Power Protection (OPP)


The DC load includes both hardware and software OPP features.
Hardware OPP: the DC load allows user to set a power protection limit in
hardware which will limit the power in the range you set when the OPP occur.
The hardware OPP protection will not change the ON/OFF state of the the DC
load.
Software OPP: users can set the DC load’s software OPP value, steps: [Shift]
+[9] > Protect > P-LIMIT > P-LIM POIN set OPP power value, P-LIM DELAY
set alarm delay. If the the DC load’s power value reach OPP limit and after the
delay time, the DC load will automatically turned off, VFD will display OPP.

Operations to Clear the OPP State


Disconnect the instrument under test. Press any key on the front panel, the
OPP displayed on the VFD will disappear, the DC load exits OPP protection
state.

3.17.4 Over Temperature Protection (OTP)


If internal temperature exceeds safety limits(85℃ ;185 °F), the Over
temperature circuitry will be activated. The DC Load will turn off the input, the
buzzer will sound, and the display will show OTP.

Operations to Clear the OTP State


When the DC load temperature drops to the protecting point, press any key on
the front panel, the OTP) displayed on the front panel will disappear, the DC
load exits OTP protection state.

3.17.5 Reverse Voltage Protection (LRV)


This feature protects the DC load in case the DC input terminals are connected
to a power source with reversed polarity. If a reverse voltage condition is
detected, the buzzer will sound and will be displayed on the VFD.

Operations to Clear the Reverse Voltage State


Check whether the connection is reversed; if so disconnect the power source.

3.18 Key Lock Function


Press [Shift] +[On/Off] (Lock)key to lock the front panel keys,VFD will display
a Lock label.In this state,setting values can not be modified,woriking mode can
not be changed.Press [Shift] +[On/Off] (Lock)again will disable this function.

3.19 The Terminals on the Rear Panel


3.19.1 Remote Sensing
Remote sensing is used to counteract the effect of lead resistance. For
example, if you connect a power supply to the DC Load, the voltage at the
power supply's terminals will not be the same as the voltage at the DC Load's
terminals if there is a current flowing because of the finite resistance from the
wires. Using remote sensing, you can sense the voltage at the power supply's
terminals, effectively removing the effect of the voltage drop in the connection
wire.
When using remote sensing, the power displayed by the instrument includes
both the power dissipated inside the instrument and the power dissipated in the
Copyright © Itech Electronics Co., Ltd. 28
Functions and Characteristics

leads from the power supply to the DC Load's input terminals.


 Steps to enable remote sensing in the menu:
1. Press [Shift] +[9] key into the menu
2. VFD displays > , press [Enter] key to confirm
3. Press [ ] to choose > , press [Enter] key to confirm
4. Press [ ] to choose >, press [Enter] key to confirm, then remote
sense function has been set, and VFD display indicator.
Remote Sensing: SENSE (+) and SENSE (–) are the remote sensing inputs.
By eliminating the effect of the inevitable voltage drop in the DC load leads,
remote sensing provides greater accuracy by allowing the DC load to regulate
directly at the source's output terminals.

Wiring Diagram for Remote Sensing:

3.19.2 External Triggering


EXTERNAL:An external trigger is a TTL low signal applied to the Trigger
connection on the back panel. This TTL signal must last for more than
5 ms. A trigger applied to this input can be used to change settings
(voltage, current , resistance), toggle between settings in transient-
toggle mode, or generate a pulse in pulse mode.

Operation to select the trigger source as external:


[Shift] +[8] (system)to enter the menu,use [ ] to select TRIGGER,press
[Enter],and then select EXTERNAL.Press [ESC] to exit the menu.

3.19.3 Current Monitoring (I Monitor)


Current monitoring terminal will output 0-10V analog signal to corresponding to
0 to full range of input current. You can connect an external voltmeter or an
oscilloscope to display the input current’s changing

Copyright © Itech Electronics Co., Ltd. 29


Functions and Characteristics

3.19.4Ripple Function
IT8500plus series DC electronic loads have test ripple function.You can read
ripple voltage and ripple current by sending instructions.See in IT8500+
programming guide.

Copyright © Itech Electronics Co., Ltd. 30


Basic Operation

Chapter4 Basic Operation


4.1 Constant Current Operation
(Set the current from 0 to the current limit)
There are three ways to set the current value:
1. In CC mode, rotate Rotary knob.
2. In CC mode, input value through number keys directly , press [Enter] to
confirm.
3. In CC mode, move the cursor to change the step value by pressing the
[ ], and then adjust the current by pressing the [ ].
To set current range, follow the steps:
Step Operation VFD Display
1 Press [CC], and then [Shift] RANGE=30.000A
+[CV] (Setup).
2 Set the current range,press RANGE =10.000A
[Enter] to confirm
3 Press [Esc] to escape. HIGH=120.00V

Note: when you set the current range to low range(within 3A),the resolution of current will rise.

4.2 Constant Voltage Operation


(Set the voltage from 0.1V to the setting voltage limit)
There are three ways to change the voltage:
1. In CV mode, rotate Rotary knob.
2. In CV mode,input value through number key boards directly,press [Enter]
to confirm
3. In CV mode, move the stepping cursor by pressing [ ], and then
adjust the voltage by pressing [ ].
To set voltage range, follow the steps:
Steps Operation VFD Display
1 Press [CV], then [Shift] +[CV]. RANGE=120.00V

2 Set the voltage range, press RANGE=10.00V


[Enter] to confirm
3 Press [Esc] to escape. HIGH=30.000A

Note: When you set the voltage range to low range,the resolution of voltage will rise.

Copyright © Itech Electronics Co., Ltd. 31


Basic Operation

4.3 Constant Power Operation


(Set a value from 0 to upper limit of power)
There are three ways to set the power value:
1. In CW mode, rotate Rotary konb.
2. In CW mode, input value through number key boards directly,press [Enter]
to confirm
3. In CW mode, move the stepping cursor by pressing [ ], and then
adjust the power by pressing [ ].
To set power range, follow the steps:
Steps Operation VFD Display
1 Press [CW], then [Shift] +[CV]. RANGE=150.00W

2 Set the power range,press [Enter] to RANGE =100.00W


confirm
3 Press [Esc] to escape. HIGH=120.00V

4.4 Constant Resistance Operation


(Allowed setting range is 0.05Ω to 7500Ω)
There are three ways to set the resistance value:
1. In CR mode, rotate Rotary knob.
2. In CR mode, input value through number key boards directly,press [Enter]
to confirm
3. In CR mode, move the stepping cursor by pressing [ ], and then
adjust the resistance by pressing [ ]
To set resistance range, follow the steps:
Steps Operation VFD Display
1 Press [CR], then [Shift] +[CV] RANGE=7500.0Ω
(Setup).
2 Set the resistance range,press [Enter] RANGE =2000Ω
to confirm
3 Press [Esc] to escape. HIGH=120.0V

4.5 Transient Test Operation


Transient operation enables the DC load to periodically switch between two
levels.
To edit a transient test file, related parameters need to be set: A level, B level,
time width (only in pulse mode), frequency, duty and running mode
(Continuous/Pulse/Toggled). If in CC dynamic mode, user can set current rising
and falling slope additionally.
Following is an example to illustrate the three transient operations.

Copyright © Itech Electronics Co., Ltd. 32


Basic Operation

4.5.1 Continuous Transient Operation

Press [Shift] +[2] (Tran) to enter the transient operation setup menu:
Steps Operation VFD Display
Press [Shift]+[2] (Tran), move TRAN ON OFF
1 [ ] key to select ON., press
[Enter] to confirm.
Preess [ ] to select MODE CONTINUOUS >
2 transient operation mode as
CONTINUOUS (the indicator lamp
Trig will be lighted)
Set the rising slope,press [Enter] to UP=1A/uS
3
confirm
Set the descending slope,press DOWN=2/uS
4
[Enter] to confirm
Set level A,press [Enter] to LEVEL A=5A
5
confirm
6 Set level B,press [Enter] to confirm LEVEL B=10A
Set the frequency,press [Enter] to FREQUNCE=50HZ(0.01-10000HZ)
7
confirm
Set the dutyfactor,press [Enter] to DUTY=98%(%0.1-99.9%)
8
confirm
Open the transient test function, TRAN ON OFF
9 mentain on the “on”selection, press
[Enter] to confirm.
Then the VFD will display TRAN A 10.0000V 0.0000A
10 and Trig 0.00W TRAN. 0
Trig
11 Turn on the load, press [Shift]+[ ]
(Trigger) to trigger

Press anyone of [CC/CV/CW/CR] button can quit the transient test,if you
12
want to continue the test again, please repeat 1-11 steps

Copyright © Itech Electronics Co., Ltd. 33


Basic Operation

4.5.2 Pulse Transient Operation

Press [Shift]+[2] (Tran) enter the transient operation setup menu:


Steps Operation VFD Display
1 Press [Shift]+[2] (Tran), TRAN ON OFF
move [ ] key to
select ON., press [Enter]
to confirm.
2 Press [ ] to select MODE CONTINUOUS >
transient operation mode
as PULSE(the indicator
lamp Trig will be lighted)
3 Set the rising slope, press UP=1A/US
[Enter] to confirm
4 Set the descending DOWN=2A/US
slope,press [Enter] to
confirm
5 Set level A,press [Enter] to Level A=5.000A
confirm
6 Set level B,press [Enter] Level B=10.000A
to confirm
7 Set the time width,press WIDTH=5S(0.00005-3600S)
[Enter] to confirm
8 Open the transient test TRAN ON OFF
function, mentain on the
“on”selection , press
[Enter] to confirm
9 Then the VFD will display 10.0000V 0.0000A
TRAN and Trig 0.00W TRAN. 0
Trig
10 Turn on the load, press
[Shift]+[ ] (Trigger) to
trigger
11 Press anyone of [CC/CV/CW/CR] button will quit the
transient test,if you want to continue the test again,please
repeat 1-10 steps.

4.5.3 Toggle Transient Operation

Copyright © Itech Electronics Co., Ltd. 34


Basic Operation

10A

5A

TRG TRG

Toggled Transient Operation

Press [Shift]+[2] (Tran) to enter the transient operation setup menu


Steps Operation VFD Display
1 Press [Shift]+[2] (Tran), move TRAN On Off
[ ] key to select on.,
press [Enter] to confirm.
2 Set transient operation mode as MODE
TOGGLE(the indicator lamp CONTINUOUS >
Trig will be lighted)
3 Set the rising slope, press UP=1A/US
[Enter] to confirm.
4 Set the descending slope,press DOWN=2A/US
[Enter] to confirm
5 Set level A, press [Enter] to LEVEL A=5A
confirm
6 Set level B, press [Enter] to LEVEL B=10A
confirm
7 Open the transient test function, TRAN ON OFF
function,press [Enter] to
confirm
8 Then the VFD will display 10.0000V 0.0000A
TRAN and Trig 0.00W TRAN. 0
Trig
9 Turn on the load, press
[Shift]+[ ] (Trigger) to trigger
10 Press anyone of [CC/CV/CW/CR] button will quit the
transient test,if you want to continue the test again,please
repeat 1-9 steps.

4.6 List Operation


Before run a list file,you should edit the list file firstly and save it in a non-volatile
memory.The following examples will help you understand the function well.In
the example, the output voltage and current are 10V and 3A,and the DC load is
in CC mode.

Copyright © Itech Electronics Co., Ltd. 35


Basic Operation

Trigger

0 1 2 3 4 5
List count=1 List count=2

List sequence

Steps Operation VFD Display


1 Press [Shift]+[3] (List), make sure LIST ON
the ON is flashing,if not,press CALL EDIT
[Enter] to select ON,then press
[ ] key to select EDIT,press
[Enter] to confirm.
2 Set the current range. LIST RANGE=3A
3 Set list step count by number LIST STEP=2(2-84)
keys, and press [Enter] to
confirm.
4 Set the first step’s current, press STEP 01 =1A
[Enter] to confirm.
5 Set the first step’s rise slope, STEP 01 =0.1A/US
press [Enter]to confirm.
6 Set the first step’s time, such as STEP 01 =5S
5S. Press Enter to confirm.
7 Set the second step’s current, STEP 01 =2A
such as 1A, press [Enter] to
confirm.
8 Set the second step’s rise slope, STEP 01 =0.1A/US
such as 1A/uS press [Enter] to
confirm.
9 Set the second step’s time, such STEP 01 =5S
as 5S. Press [Enter] to confirm.
10 Set repeat times, press [Enter] to REAPEAT =3
confirm.
11 Select the position to save file, SAVE LIST =1(1-7)
such as 1, press [Enter] to
confirm.
12 Press [ ] to select ON, LIST ON CALL
press EDIT
[Enter] to confirm (the Trig
indicator will be light now),press
[Esc].
13 Turn on the DC load, press
[Shift]+[ ] (Trigger) to trigger.
14 Press any function ksys if you want to quit list mode

If you want to run a list file you’ve saved, please recall it first.The steps is:
Steps Operation VFD Display
Copyright © Itech Electronics Co., Ltd. 36
Basic Operation

1 Press [Shift]+[3] (List), make sure LIST ON CALL


the ON is flashing,if not,press EDIT
[Enter] to select ON,then press
[ ] to select CALL,press
[Enter] to confirm.
2 Select the list file.Press [Enter] to RECALL LIST =1
confirm.
3 - LIST ON CALL
EDIT

4.7 Test Files


Test files are a generalization of lists—they let you generate a sequence of
tests using different modes, mode parameters, and durations. They are useful
for executing a set of tests on a device, then displaying whether the tests
passed or failed. We will illustrate how to use test files by a short example.
You can edit up to 10 groups of testing files, each file has 10 steps, it can edit
up to 100 steps which can be saved in EEPROM (address).
Suppose we have a small AC to DC power supply (a "wall-wart") and we want
to set up an acceptance test for a number of these devices. Our test will consist
of two steps:
1. Set the DC load to constant current mode to draw the rated current of 1.2A
from the device. The output voltage of the device at the rated current must
be between 4.4V and 4.6V.
2. Set the DC load to constant voltage 3V. The output current of the device is
between 2A and 3A.
3. When the device operates into a short, the supplied current must be larger
than 3.0 A.
Steps Operation VFD Display
1 Press [Shift]+[6] (Prog), ACTIVE =0987654321
2 Press [1] ,[2] and [3], Press ACTIVE =0987654YYY
[Enter] to confirm.
3 Select the step that needs to PAUSE
pause during the test. =NNNNNNN32Y
When it is paused,press [ ]
can continue the test.
4 Step 3 short circuit testing, SHORT
press [3]. And press [Enter] =NNNNNNNY21
to confirm.
5 Set Ton for the first step, if you SEQ01 ON =2S
want to load on 2S, press [2],
and then press [Enter] to
confirm.
Ton range 0~60S
6 Set Toff for the first step, if you SEQ01 OFF =2S
want to load off 2S, press [2],
then press [Enter] to confirm
Toff range 0~60S
7 Set testing delay time, range SEQ01 P/F =1S
0~60S e.g. 1S, press [1].

Copyright © Itech Electronics Co., Ltd. 37


Basic Operation

Steps Operation VFD Display


8 Set Ton for the second step, if SEQ02 ON =2S
you want to load 2S, press
[2], then press [Enter] to
confirm
9 Set Toff of the second step, if SEQ02 OFF =2S
you need 2S, press [2], and
then press [Enter] to confirm.
10 Set testing delay time of the SEQ02 P/F =1S
second step, e.g. 1S, press
[1]
11 Set Ton for the third step, if SEQ03 ON =3S
you want to load on 3S, press
[3], then press [Enter]to
confirm
12 Set Toff of the third step, if you SEQ03 OFF =2S
need 2S, press [2], then press
[Enter] to confirm.
13 Set testing delay time of the SEQ03 P/F =2S
third step, e.g. 2S, press [2].
14 set start voltage.Please refer AUTO START=0.000V
to “function of auto start
voltage”.
15 Set stop condition STOP COMP
FAILURE

Function of Auto Start Voltage:


1. Auto start=0V.

Auto test file start to run when receive a trigger signal by pressing
[shift]+[trigge]r or providing a external trigger signal.
2. Auto start is not equal to 0V(Take 2V as an example)
In this condition,user only need to connect the charger to input termianls of
E-load.The unit can auto start to run test file when detect a rising edge from
0-2V.Auto start voltage is not suggested to be a big value.2V is suitable.

Ton, Toff and Tpf (P/F) Relation:

Copyright © Itech Electronics Co., Ltd. 38


Basic Operation

Tpf is the delay time for a step


15 Set stop conditions: COMP means STOP COMP
stop test when all the steps are FAILURE
completed, FAILURE means stop
test when the testing fails. Press
[Enter] key to confirm.
16 Select the test file to link if you’d like CHAIN PROGRAM
to. The linked file must be saved =0(0-10)
before. 0 stands for not linking to
other files. Press [Enter] key to
confirm.

PROGRAM 1 1 2 3 4 5 6 7 8 9 10
Sequence
Save Group 1 2 3 4 5 6 7 8 9 10
PROGRAM 2 1 2 3 4 5 6 7 8 9 10
Sequence
Save Group 11 12 13 14 15 16 17 18 19 20
PROGRAM 10 1 2 3 4 5 6 7 8 9 10
Sequence
Save Group 91 92 93 94 95 96 97 98 99 100

17 Save the edited files in EEPROM, SAVE PROGRAM


you can save up to 10 groups of =1(1-10)
files, e.g please press [1] to save
the edited file in group 1, and then
press [Enter] to confirm.
18 Select a operation mode and then 10.0000V 0.0000A
press [Shift]+[CV] to set related 0.00W CC=1.000A
parameters
19 Edit the three steps of the test file,
details refer to below procedure.
After all the steps are set ,
Press [ESC] to exit setup, and
then press [Shift]+[4] to save.
You need to recall the test file before runing it

Set the steps of a test file in the example


CC Mode, 1.2A, Voltage Range 4.4V~4.6V
Step Operation VFD Display
1 Press [CC] button,and then RANGE=30.000A
[Shift)+[CV] (Setup) to enter CC
the setting interface
2 Set the current range,press RANGE =1.2A
[Enter] to confirm CC
3 set the upper limit of voltage, HIGH=4.6V
press [Enter] to confirm CC
4 Set the lower limit of voltage, LOW=4.4V
press [Enter] to confirm CC
5 Set the rise speed of current, UP=1A/uS
press [Enter] to confirm CC
6 Set the fall speed of current, DOWN=1A/uS
press [Enter] to confirm CC
7 Finish the setup 10.0000V 0.000A
0.00W CC=0.000A
Copyright © Itech Electronics Co., Ltd. 39
Basic Operation

CV Mode, 3V, Current Range 2A~3A


Steps Operation VFD Display
1 Press [CV] button,press RANGE=120.00V
[Shift]+[CV] to set related
parameters
2 Set the voltage range,press RANGE=3.00V
[Enter] to confirm
3 set the upper limit of HIGH=3A
current,press [Enter] to confirm
4 Set the lower limit of LOW=2A
current,press [Enter] to confirm
5 Finish the setup 10.0000V 0.000A
0.00W CV=10V

The CW and CR is set as the same way:


 CW Mode
Steps Operation VFD Display
1 Press [CW] button,press RANGE=150.00W
[Shift]+[CV] to set related
parameters
2 Set the power range,press RANGE =1.00W
[Enter] to confirm
3 Set the upper limit of HIGH=120.00V
voltage,press [Enter] to confirm
4 Set the lower limit of LOW=0.000V
voltage,press [Enter] to confirm
5 Finish the setup 10.0000V 0.000A
0.00W CW=1.00W

 CR Mode
Steps Operation VFD Display
1 Press [CR] button,press RANGE=7500.0Ω
[Shift]+[CV](Setup) to set
related parameters
2 Set the resistance range,press RANGE =2Ω
[Enter] to confirm
3 set the upper limit of HIGH=120.0V
voltage,press [Enter] to confirm
4 Set the lower limit of LOW=0.000V
voltage,press [Enter] to confirm
5 Finish the setup 10.0000V 0.000A
0.00W CR=2.000Ω

 Go into Autotest Mode


Operation Display on Front Pannel

1.Press [Shift]+[8] (system) enter 0.0000V 0.000A


into sysmtem menu POWER-ON BUZZER
2.Press right key,select 0.0000V 0.000A
RUNMODE and confirm with RUN <NORMAL
[Enter] button

Copyright © Itech Electronics Co., Ltd. 40


Basic Operation

3.Press direction key to select 0.0000V 0.000A


OCP_TEST,Press [Enter] to RUN <PROG_TEST
confirm.
4,Press [Esc] button to quit menu 0.0000V 0.000A
set P01

According to the following steps to escape OPP mode:


Press [Shift]+[8] (system)---”RUNMODE”---[Eneter]----select “NORMAL”
mode---[Enter].
(6)Start Auto Test File
Press [ ](trigger) to provide a signal to start auto test file.The discharing
process will be auto terminated when stop conditions are reached.
(7)Recall Test File
Press [Shift]+[Enter] button to select programe file,the panel displays
“RECALL PROGRAM= 1.Enter the file name(1-10),press [Enter] button to
confirm.

If you need a pause, please press [Shift]+[0] (pause).Press [ ] can continue


the test.

Copyright © Itech Electronics Co., Ltd. 41


Communication Interfaces

Chapter5 Communication Interfaces


DB9 in the rear panel of the DC load could connect with RS-232 through on TTL connector. The
following description may help you to know how to control the output of the DC load through PC.

WARNING: Don’t connect the DC load’s DB9 connector to a standard RS232 instrument; doing
so may damage the instrument.

5.1Communication Modules Intruduction


IT-E121 Communication Module
The DB9 interface connector on the rear panel of the DC load is TTL voltage
level; you can use the communication module IT-E121 and an a standard
RS232 extension cable to connect the DB9 interface connector of the DC load
and the RS-232 interface connector of computer for the communication.

IT-E121A Communication Module


The DB9 interface connector on the rear panel of the DC load is TTL voltage
level; you can use the communication module IT-E121A and an a standard
RS232 extension cable to connect the DB9 interface connector of the DC load
and the RS-232 interface connector of computer for the communication.
IT-E121A is derived on the basis of IT-E121,the main difference between them
is that the DB9 interface connector of the RS232 changes from female to
male ,so that can be directly connected to the standard LAN interface.
IT-E121A communication cable

IT-E122 Communication Module


The DB9 interface connector on the rear panel of the DC load is TTL voltage
level;
Copyright © Itech Electronics Co., Ltd. 42
Communication Interfaces

IT-E122 has a USB interface on one end,you can use IT-E122 and an a
standard USB extension cable to connect the DB9 interface connector of the
DC load and the USB interface connector of computer for the communication.

IT-E123 Communication Module


The DB9 interface connector on the rear panel of the DC load is TTL voltage
level;
The interface on both side port of IT-E123 are DB9 interface and RS485
interface, you
can use the communication module IT-E123 and an a standard RS485-RS232
conversion cable to connect the DB9 interface connector of the DC load and
the RS-232 interface connector of computer for the communication.

RS485 pins

5.2 Communication with PC


Before using the remote operation mode, please make sure that the baud rate
and communication address in the DC load are the same as in the computer
software, otherwise, the communication will fail, you can change the baud rate
and communication address from the front panel or from computer.

Copyright © Itech Electronics Co., Ltd. 43


Communication Interfaces

DB9 Serial Port


In order for the computer to communicate with the DC load, both must be set to
the same RS-232 settings. These communication settings are:
1. Address: the range is from 0 to 31, default setting is 0
2. Baud rate: 4800,9600,19200 and 38400 are selectable, default setting is
9600. Refer to chaptor 1.7.
3. Data bit: 8 bit
4. Stop bit: 1
5. Parity: None,Even, Odd, default is None, refer to chaptor1.7.

Parity=None Start Bit 8 Data Bits Stop Bit

RS-232
IT8514B+/IT8514C+/IT8516C+ series electronic load have a DB9 interface on
rear panel. Connect E-load and computer by cable of COM ends (DB9).
Composite key [Shift]+[8] on front board can be used to enter system menu for
activation.
 RS-232 Interface
In RS-232 interface, all SCPI commands can be used for programming. If
RS-232 interface is selected, in accordance with internal connection of data
terminal equipment (DTE) and data communication equipment (DCE) as
defined in EIA RS-232, the load is connected to another DTE (e.g., PC COM
interface) with direct-connected Modem cable.
Base Pin Description
Number
1 No conjunction
2 TXD, data transmission
RS232 Pins of Plug
3 RXD, data receiving
4 No conjunction
5 GND, grounding
6 No conjunction
7 CTS, clear to send
8 RTS, request to send
9 No conjunction

 Communication Setup
Please ensure the PC and the load have the same configuration in the following
items.
Baudrate: 9600(4800、9600、19200、38400).You could enter the system menu
Copyright © Itech Electronics Co., Ltd. 44
Communication Interfaces

to set the baudrate.


Data bit: 8
Stop bit: 1
Parity bit: (none,even,odd)
EVEN 8 data bits have even parity
ODD 8 data bits have odd parity
NONE 8 data bits have no parity
Native machine address: ( 0 ~31, factory default is 0)

Start Bit 8 Data Bits Parity=None Stop Bit

5.3USB Interface
Connect the load and the computer using a cable with two USB interfaces
(each end). All functions of the load can be programmed via USB.
After connecting the load and computer by USB, you need to install IT-E122
driver or IT-E132 driver ( see in ITECH CD or contact ITECH agent). The device
manager of PC will display ‘Prolific USB-to-Serial COM Port’ after installing.
Notes: Only IT8514B+/8514C+/IT8516C+ models have the USB commucation
interface. Just use USB line to connect in commucation. ( Please don’t connect
DB9 interface at the same time). Don’t need to set the menu.

Copyright © Itech Electronics Co., Ltd. 45


Specifications

Chapter6 Specifications
Specifications
Model IT8511+ IT8512+
input voltage 0~120V 0~120V

Rated input current 0~3A 0~30A 0~3A 0~30A


value input power 150W 300W
( 0~40 ℃) Minimum
operation 0.14V at 3A 1.4V at 30A 0.12V at 3A 1.2V at 30A
value
range 0.1~18V 0.1~120V 0.1~18V 0.1~120V

CV mode resolution 1mV 10mV 1mV


10mV
±(0.05%+0.025%F
accuracy ±(0.05%+0.02%FS) ±(0.05%+0.025%FS) ±(0.05%+0.02%FS)
S)
range 0~3A 0~30A 0~3A 0~30A
CC mode resolution 0.1mA 1mA 0.1mA 1mA

accuracy ±(0.05%+0.05%FS)
range 0.05Ω~10Ω 10Ω~7.5KΩ 0.05Ω~10Ω 10Ω~7.5KΩ
CR mode
resolution 16bit 16bit
*1
accuracy 0.01%+0.08S *2 0.01%+0.0008S 0.01%+0.08S *2 0.01%+0.0008S
range 150W 300W
CP mode
resolution 10mW 10mW
*3
accuracy ±(0.1%+0.1%FS) ±(0.1%+0.1%FS)
Dynamic mode
CC
T1&T2 20uS~3600S /Res:1 uS 20uS~3600S /Res:1 uS
accuracy 2uS±100ppm 2uS±100ppm
Rising/Falling slope *4 0.0001~0.2A/uS 0.001~1.5A/uS 0.0001~0.2A/uS 0.001~1.5A/uS
minimum rise time *5 ≒10uS ≒10uS ≒10uS ≒10uS
Measuring range
range 0~18V 0~120V 0~18V 0~120V
Readback
resolution 0.1 mV 1 mV 0.1 mV 1 mV
voltage
accuracy ±(0.025%+0.025%FS)
range 0~3A 0~30A 0~3A 0~30A
Readback
resolution 0.1mA 1mA 0.1mA 1mA
current
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 150W 300W
Readback
resolution 10mW 10mW
power
accuracy ±(0.1%+0.1%FS) ±(0.1%+0.1%FS)
Protection range
OPP
≒160W ≒320W
Protection
OCP
≒3.3A ≒33A ≒3.3A ≒33A
Protection
OVP
≒125V ≒125V
Protection
OTP
≒85℃ ≒85℃
Protection
Specification
current(CC) ≒3.3/3A ≒33/30A ≒3.3/3A ≒33/30A
Short
voltage(CV) 0V 0V 0V 0V

Copyright © Itech Electronics Co., Ltd. 46


Specifications

resistance(C
≒45mΩ ≒45mΩ ≒40mΩ ≒40mΩ
R)
input
Impendanc 150KΩ 150KΩ
e
Dimension 214.5mm*88.2mm*354.6mm 214.5mm*88.2mm*354.6mm

Model IT8511A+ IT8512A+


input voltage 0~150V 0~150V
Rated
input current 0~3A 0~30A 0~3A 0~30A
value
( 0~40 input power 150W 300W
℃) Minimum
0.25V at 3A 2.5V at 30A 0.14V at 3A 1.4V at 30A
operation value
range 0.1~18V 0.1~150V 0.1~18V 0.1~150V
CV resolution 1mV 10mV 1mV 10mV
mode ±(0.05%+0.025%F
accuracy ±(0.05%+0.02%FS) ±(0.05%+0.025%FS) ±(0.05%+0.02%FS)
S)
range 0~3A 0~30A 0~3A 0~30A
CC resolution 0.1mA 1mA 0.1mA 1mA
mode ±(0.05%+0.05%
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
FS)
CR range 0.05Ω~10Ω 10Ω~7.5KΩ 0.05Ω~10Ω 10Ω~7.5KΩ
mode resolution 16bit 16bit
*1 accuracy 0.01%+0.08S *2 0.01%+0.0008S 0.01%+0.08S *2 0.01%+0.0008S
CP range 150W 300W
mode resolution 10mW 10mW
*3 accuracy ±( 0.1%+0.1%FS ) ±( 0.1%+0.1%FS )
Dynamic mode( CC mode )
T1&T2 20uS~3600S /Res:1 uS 20uS~3600S /Res:1 uS
accuracy 2uS±100ppm 2uS±100ppm
Rising/Falling slope *4 0.0001~0.2A/uS 0.001~1.5A/uS 0.0001~0.2A/uS 0.001~1.5A/uS
minimum rise time *5 ≒10uS ≒10uS ≒10uS ≒10uS
Measuring range
range 0~18V 0~150V 0~18V 0~150V
Readba
ck resolution 0.1 mV 1 mV 0.1 mV 1 mV
voltage ±(0.025%+0.025%
accuracy ±(0.025%+0.025%FS) ±(0.025%+0.025%FS) ±(0.025%+0.025%FS)
FS)
Readba range 0~3A 0~30A 0~3A 0~30A
ck resolution 0.1mA 1mA 0.1mA 1mA
current accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 150W 300W
Readbac
k power
resolution 10mW 10mW
accuracy ±(0.1%+0.1%FS) ±(0.1%+0.1%FS)
Protection range
OPP ≒320W
Protecti ≒160W
on
OCP
Protecti ≒3.3A ≒33A ≒3.3A ≒33A
on
OVP
Protecti ≒160V ≒160V
on
OTP
Protecti ≒85℃ ≒85℃
on

Copyright © Itech Electronics Co., Ltd. 47


Specifications

Specification
current( CC ) ≒3.3/3A ≒33/30A ≒3.3/3A ≒33/30A
Short voltage( CV ) 0V 0V 0V 0V
resistance( CR
≒80mΩ ≒80mΩ ≒40mΩ ≒40mΩ
)
input
Impend 150KΩ 150KΩ
ance
Dimensi
214.5mm*88.2mm*354.6mm 214.5mm*88.2mm*354.6mm
on

Model IT8511B+
input voltage 0~500V
input current 0~3A 0~10A
Rated value input power 150W
( 0~40 ℃) Minimum
operation 1.2V at 3A 4V at 10A
value
range 0.1~50V 0.1~500V
CV mode resolution 1mV 10mV
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 0~3A 0~10A
CC mode resolution 0.1mA 1mA
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 0.5Ω~10Ω 10Ω~7.5KΩ
CR mode*1 resolution 16bit
accuracy 0.01%+0.08S *2 0.01%+0.0008S
range 150W
CP mode
resolution 10mW
*3
accuracy 0.1%+0.2%FS
Dynamic mode
CC mode
T1&T2 20uS~3600S /Res:1 uS
accuracy 2uS±100ppm
Dynamic mode Rising/Falling
0.0001~0.2A/uS 0.001~0.8A/uS
slope*4
minimum rise
time ≒10uS ≒10uS
*5
Measuring range
range 0~50V 0~500V
Readback
resolution 1 mV 10 mV
voltage
accuracy ±(0.025%+0.025%FS) ±(0.025%+0.025%FS)
range 0~3A 0~10A
Readback
resolution 0.1mA 1mA
current
accuracy ±(0.05%+0.05%FS)
range 150W
Readback power resolution 10mW
accuracy ±(0.1%+0.2%FS)
Protection range
OPP Protection ≒160W
OCP Protection ≒3.3A ≒11A
OVP Protection ≒530V
OTP Protection ≒85℃

Copyright © Itech Electronics Co., Ltd. 48


Specifications

Specification
current( CC ) ≒3.3/3A ≒11/10A

Short voltage( CV ) 0V 0V
resistance( C
≒400mΩ ≒400mΩ
R)
input
1MΩ
Impendance
Dimension 214.5mm*88.2mm*354.6mm

Model IT8512B+ IT8512C+


input voltage 0~500V 0~120V
input current 0~3A 0~15A 0~6A 0~60A
Rated value input power 300 W 300 W
( 0~40 ℃) Minimum
operation 0.6V at 3A 3V at 15A 0.25V at 6A 2.5V at 60A
value
range 0.1~50V 0.1~500V 0.1~18V 0.1~120V
CV mode resolution 1mV 10mV 1mV 10mV
±(0.05%+0.025%F
accuracy ±(0.05%+0.02%FS) ±(0.05%+0.025%FS) ±(0.05%+0.02%FS)
S)
range 0~3A 0~15A 0~6A 0~60A
CC mode resolution 0.1mA 1mA 0.1mA 1mA
accuracy ±(0.05%+0.05%FS)
range 0.3Ω~10Ω 10Ω~7.5KΩ 0.3Ω~10Ω 10Ω~7.5KΩ
CR mode
resolution 16bit 16bit
*1
accuracy 0.01%+0.08S *2 0.01%+0.0008S 0.01%+0.08S *2 0.01%+0.0008S
range 300W 300W
CP mode
resolution 10mW 10mW
*3
accuracy ±(0.1%+0.1%FS) ±(0.1%+0.1%FS)
Dynamic mode
CC mode
T1&T2 20uS~3600S /Res:1 uS 20uS~3600S /Res:1 uS
accuracy 2uS±100ppm 2uS±100ppm
Rising/Falling slope *4 0.0001~0.2A/uS 0.001~0.8A/uS 0.0001~0.3A/uS 0.001~3A/uS
minimum rise time *5 ≒10uS ≒10uS ≒10uS ≒10uS
Measuring range
range 0~50V 0~500V 0~18V 0~120V
Readback
resolution 1 mV 10mV 1 mV 10mV
voltage
accuracy ±(0.025%+0.025%FS)
range 0~3A 0~15A 0~6A 0~60A
Readback
resolution 0.1mA 1mA 0.1mA 1mA
current
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 300W 300W
Readback
resolution 10mW 10mW
power
accuracy ±(0.1%+0.1%FS) ±(0.1%+0.1%FS)
Protection range
OPP ≒320W
≒320W
Protection
OCP
≒3.3A ≒16A ≒6.5A ≒65A
Protection
OVP
≒530V ≒125V
Protection
Copyright © Itech Electronics Co., Ltd. 49
Specifications

OTP
≒85℃ ≒85℃
Protection
Specification
current( CC ) ≒3.3/3A ≒16/15A ≒6.5/6A ≒65/60A
Short voltage( CV ) 0V 0V 0V 0V
resistance( C
≒180mΩ ≒180mΩ ≒40mΩ ≒40mΩ
R)
input
Impendanc 1MΩ 150KΩ
e
Dimension 214.5mm*88.2mm*354.6mm 214.5mm*88.2mm*354.6mm

Model IT8512H+
input voltage 0~800V
input current 0~1A 0~5A
Rated value input power 300W
( 0~40 ℃) Minimum
operation 1.4V at 1A 7V at 5A
value
range 0.1~80V 0.1~800V
CV mode resolution 1mV 10mV
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 0~1A 0~5A
CC mode resolution 0.1mA 1mA
accuracy ±(0.05%+0.1%FS) ±(0.05%+0.05%FS)
range 2Ω~10Ω 10Ω~7.5KΩ
CR mode*1 resolution 16bit
accuracy 0.01%+0.08S *2 0.01%+0.0008S
range 300W
CP mode
resolution 10mW
*3
accuracy 0.2%+0.2%FS
Dynamic mode
CC mode
T1&T2 20uS~3600S /Res:1 uS
accuracy 2uS±100ppm
Dynamic mode Rising/Falling
0.0001~0.04A/uS 0.001~0.2A/uS
slope*4
minimum rise
time ≒20uS ≒20uS
*5
Measuring range
range 0~80V 0~800V
Readback
resolution 1 mV 10 mV
voltage
accuracy ±(0.025%+0.025%FS) ±(0.025%+0.025%FS)
range 0~1A 0~5A
Readback
resolution 0.1mA 1mA
current
accuracy ±(0.05%+0.05%FS)
range 300W
Readback power resolution 10mW
accuracy ±(0.2%+0.2%FS)
Protection range
OPP Protection ≒320W
OCP Protection ≒1.1A ≒5.5A

Copyright © Itech Electronics Co., Ltd. 50


Specifications

OVP Protection ≒850V


OTP Protection ≒85℃
Specification
current( CC ) ≒1.1/1A ≒5.5/5A

Short voltage( CV ) 0V 0V
resistance( C
≒1.4Ω
R)
input
2MΩ
Impendance
Dimension 214.5mm*88.2mm*354.6mm

Model IT8513A+
input voltage 0~150V
input current 0~6A 0~60A
Rated value input power 400W
( 0~40 ℃) Minimum
operation 0.25V at 6A 2.5V at 60A
value
range 0.1~18V 0.1~150V
CV mode resolution 1mV 10mV
accuracy ±(0.05%+0.02%FS) ±(0.05%+0.025%FS)
range 0~6A 0~60A
CC mode resolution 0.1mA 1mA
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
range 0.1Ω~10Ω 10Ω~7.5KΩ
CR mode*1 resolution 16bit
accuracy 0.01%+0.08S *2 0.01%+0.0008S
range 400W
CP mode
resolution 10mW
*3
accuracy ±(0.2%+0.2%FS)
Dynamic mode
CC mode
T1&T2 20uS~3600S /Res:1 uS
Dynamic accuracy 2Us+100ppm
mode Rising/Falling
0.001~0.15A/uS 0.01~1 A/uS
slope*4
minimum rise
time ≒50uS ≒60uS
*5
Measuring range
range 0~18V 0~150V
Readback
resolution 0.1 mV 1mV
voltage
accuracy ±(0.025%+0.025%FS) ±(0.025%+0.025%FS)
range 0~6A 0~60A
Readback
resolution 0.1mA 1mA
current
accuracy ±(0.05%+0.05%FS)
range 400W
Readback
power
resolution 10mW
accuracy ±(0.2%+0.2%FS)
Protection range
OPP ≒420W

Copyright © Itech Electronics Co., Ltd. 51


Specifications

Protection
OCP
≒6.6A ≒66A
Protection
OVP
≒165V
Protection
OTP
≒85℃
Protection
Specification
current( CC ) ≒6.6/6A ≒66/60A

Short voltage( CV ) ≒0V


resistance( C
≒30mΩ
R)
input
Impendanc ≒280KΩ
e
Dimension 214.5mm*88.2mm*453.5mm

Model IT8513C+ IT8514C+


input voltage 0~120V 0~120V
input current 0~12A 0~120A 0~24A 0~240A
Rated value input power 600 W 1500W
( 0~40 ℃) Minimum
operation 0.2V at 12A 2V at 120A 0.25V at 24A 2.5V at 240A
value
range 0.1~18V 0.1~120V 0.1~18V 0.1~120V
CV mode resolution 1mV 10mV 1mV 10mV
±(0.05%+0.025%F
accuracy ±(0.05%+0.02%FS) ±(0.05%+0.025%FS) ±(0.05%+0.02%FS)
S)
range 0~12A 0~120A 0~24A 0~240A
CC mode resolution 1mA 10mA 1mA 10mA
accuracy ±(0.05%+0.05%FS) ±(0.1%+0.1%FS)
range 0.05Ω~10Ω 10Ω~7.5KΩ 0.05Ω~10Ω 10Ω~7.5KΩ
CR mode
resolution 16bit 16bit
*1
accuracy 0.01%+0.08S *2 0.01%+0.0008S 0.02%+0.08S *2 0.02%+0.0008S
range 600W 1500W
CP mode
resolution 10mW 10mW
*3
accuracy ±(0.2%+0.2%FS) ±(0.2%+0.2%FS)
Dynamic mode
CC mode
T1&T2 100uS~3600S /Res:1 uS 100uS~3600S /Res:1 uS
accuracy 10uS±100ppm 10uS±100ppm
Rising/Falling slope *4 0.001~0.2A/uS 0.01~1.6A/uS 0.001~0.3A/uS 0.01~3.2A/uS
minimum rise time *5 ≒60uS ≒60uS ≒60uS ≒60uS
Measuring range
range 0~18V 0~120V 0~18V 0~120V
Readback resolution 0.1 mV 1mV 0.1 mV 1mV
voltage
accuracy ±(0.025%+0.025%FS)
range 0~12A 0~120A 0~24A 0~240A
Readback
resolution 1mA 10mA 1mA 10mA
current
accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
Readback range 600W 1500W
power resolution 10mW 10mW

Copyright © Itech Electronics Co., Ltd. 52


Specifications

accuracy ±(0.2%+0.2%FS) ±(0.2%+0.2%FS)


Protection range
OPP ≒1550W
≒620W
Protection
OCP
≒13A ≒130A ≒26.7A ≒267A
Protection
OVP
≒125V ≒125V
Protection
OTP
≒95℃ ≒85℃
Protection
Specification
current( CC ) ≒13/12A ≒130/120A ≒26.7/24A ≒267/240A
Short voltage( CV ) 0V 0V 0V 0V
resistance( C
≒15mΩ ≒15mΩ ≒8mΩ ≒8mΩ
R)
input
Impendanc 150KΩ 150KΩ
e
Dimension 214.5mm*88.2mm*453.5mm 436.5mm*88.2mm*463.5mm

Model IT8514B+ IT8516C+


input voltage 0~500V 0~120V
input current 0~6A 0~60A 0~24A 0~240A
Rated value input power 1500 W 3000W
( 0~40 ℃) Minimum
operation 0.25V at 6A 2.5V at 60A 0.15V at 24A 1.5V at 240A
value
range 0.1~50V 0.1~500V 0.1~18V 0.1~120V
CV mode resolution 1mV 10mV 1mV 10mV
±(0.05%+0.025%F
accuracy ±(0.05%+0.02%FS) ±(0.05%+0.025%FS) ±(0.05%+0.02%FS)
S)
range 0~6A 0~60A 0~24A 0~240A
CC mode resolution 1mA 10mA 1mA 10mA
accuracy ±(0.05%+0.05%FS) ±(0.1%+0.1%FS)
range 0.05Ω~10Ω 10Ω~7.5KΩ 0.05Ω~10Ω 10Ω~7.5KΩ
CR mode
resolution 16bit 16bit
*1
accuracy 0.02%+0.08S *2 0.02%+0.0008S 0.02%+0.08S *2 0.02%+0.0008S
range 1500W 3000W
CP mode
resolution 10mW 10mW
*3
accuracy ±( 0.2%+0.2%FS ) ±( 0.2%+0.2%FS )
Dynamic mode
CC mode
T1&T2 100uS~3600S /Res:1 uS 120uS~3600S /Res:1 uS
accuracy 10uS±100ppm 10uS±100ppm
Rising/Falling slope *4 0.001~0.15A/uS 0.01~0.8A/uS 0.001~0.3A/uS 0.01~2.8A/uS
minimum rise time *5 ≒60uS ≒60uS ≒70uS ≒70uS
Measuring range
range 0~50V 0~500V 0~18V 0~120V
Readback resolution 0.1 mV 1mV 0.1 mV 1mV
voltage
accuracy ±(0.025%+0.025%FS)
range 0~6A 0~60A 0~24A 0~240A
Readback
resolution 1mA 10mA 1mA 10mA
current
accuracy ±(0.05%+0.05%FS) ±(0.1%+0.1%FS)
range 1500W 3000W
Readback
resolution 10mW 10mW
power
accuracy ±(0.2%+0.2%FS) ±(0.2%+0.2%FS)
Copyright © Itech Electronics Co., Ltd. 53
Specifications

Protection range
OPP ≒3050W
≒1550W
Protection
OCP
≒6.7A ≒67A ≒26A ≒260A
Protection
OVP
≒530V ≒125V
Protection
OTP
≒85℃ ≒85℃
Protection
Specification
current( CC ) ≒6.7/6A ≒67/60A ≒26/24A ≒260/240A
Short voltage( CV ) 0V 0V 0V 0V
resistance( C
≒30mΩ ≒30mΩ ≒5mΩ ≒5mΩ
R)
input
Impendanc 150KΩ 150KΩ
e
Dimension 436.5mm*88.2mm*463.5mm 436.5mm*176mm*463.5mm

*1 The voltage/current input is no less than 10% FS


*2 The scope of read-back resistance is:( 1/(1/R+(1/R)*0.01%+0.08),1/(1/R-(1/R)*0.01%-0.08) )
IT8514B+/14C+/16C+: ( 1/(1/R+(1/R)*0.02%+0.08),1/(1/R-(1/R)*0.02%-0.08) )
*3 The voltage/current input is no less than 10% FS
*4 .Ascending/descending slope: 10%-90% current ascending slope from 0 to
maximum current
*5 Minimum rise time: 10%-90% current rise time
* The above specifications may be subject to change without prior notice.

Supplementary Characteristics
 Memory capacity:100 registeres
Suggested calibration frequency:Once a year
 AC input level(A transfer switch is selectable on the rear panel)
Option Opt.1: 220V ±10% 50Hz/60Hz
Option Opt.2: 110V ±10% 50Hz/60Hz
 Cooling type
Intelligent fans
 Fans working principle:
Fans running speed is determined by radiator temperature.When
temperature reaches 40℃,fans start to work and intellig ently adjust its
speed with temperature variation.

Copyright © Itech Electronics Co., Ltd. 54


Appendix

Appendix
Specifications of Red and Black Test Lines
ITECH provides you with optional red and black test lines, which individual
sales and you can select for test. For specifications of ITECH test lines and
maximum current values, refer to the table below.
Model Specification Cross Section Length
IT-E301/10A 10A - 1m
IT-E301/30A 30A 6mm2 1.2m
IT-E301/30A 30A 6mm2 2m
IT-E301/60A 60A 20mm2 1.5m
IT-E301/120A 120A 50mm2 2m
IT-E301/240A 240A 70mm2 1m
IT-E301/240A 240A 70mm2 2m
IT-E301/360A 360A 95mm2 2m

For maximum current of AWG copper wire, refer to table blow.


AWG 10 12 14 16 18 20 22 24 26 28
The 40 25 20 13 10 7 5 3.5 2.5 1.7
Maximum
Current
Value( A)
Note: AWG ( American Wire Gage), it means X wire ( marked on the wire). The
table above lists current capacity of single wire at working temperature of 30°C.
For reference only.

Copyright © Itech Electronics Co., Ltd. 55