0% found this document useful (0 votes)
298 views2 pages

Sup St500 Temperature Transmitter User Manual

The intelligent temperature transmitter (round card) is used as a signal input for RTDs and thermocouples, outputting a 4-20mA analog signal. It has a round casing that is installed inside a sensor housing. It can be configured via a serial port connection and has inputs for RTDs, thermocouples, and linear resistance with outputs between -200-850°C and -180-1372°C depending on sensor type. It is powered by 12-40VDC and has IP66 protection for outdoor installation.

Uploaded by

Wolf Black
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
298 views2 pages

Sup St500 Temperature Transmitter User Manual

The intelligent temperature transmitter (round card) is used as a signal input for RTDs and thermocouples, outputting a 4-20mA analog signal. It has a round casing that is installed inside a sensor housing. It can be configured via a serial port connection and has inputs for RTDs, thermocouples, and linear resistance with outputs between -200-850°C and -180-1372°C depending on sensor type. It is powered by 12-40VDC and has IP66 protection for outdoor installation.

Uploaded by

Wolf Black
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Intelligent Temperature Transmitter (Round Card)

Operating Manual
Product Introduction
Intelligent temperature transmitter (round card) is used for the signal input of resistance temperature detector
(RTD) and thermocouple (TC),and 4 - 20mA analog output of the two-wire system. It is installed inside the
sensor(Form B).
1. Appearance Structure Diagram 2. Installation Schematic Diagram

Lid

Installed screw
and spring

Sensor wiring

Unit:mm(in)

Power line Head installed


3. Configuration Schematic Diagram

V1 or V8 serial port programming line

24V Power supply-


Note: no 24V power supply is required when using the V8 serial port programming line

4. Wiring Diagram

12~40VDC
4~20mA

Note: RTD (platinum resistance) input: three wire resistance value must be equal, each wire resistance is no more than 10 Ω

-1-
5. Technical Parameters

Input
Input signal Resistance temperature detector (RTD),
thermocouple (TC), and linear resistance.
Cold-junction compensation temperature scope -20~60℃
Compensation precision ±1℃
Output
Output signal 4-20mA
Load resistance RL≤(Ue-12)/0.021
Output current of upper and lower limit overflow alarm IH=21mA、IL=3.8mA
Output current of input disconnection alarm 21mA
Power supply
Supply voltage DC12-40V
Other parameters
Transmission precision (20℃) 0.1%FS
Temperature drift 0.01%FS/℃
Response time Reach to 90% of the final value for 1s
Used environmental temperature -40~80℃
Storage temperature -40~100℃
Condensation Allowable
Protection level IP00; IP66 (installation)
Aseismicity 4g/2~150Hz
Installation angle Unlimited
Installation area B-type top cassette installation
Electromagnetic compatibility Conform to GB/T18268 industrial equipment
application requirements (IEC 61326-1)

Input Type Table:

Model Type Measurement scope Minimum measurement scope


Resistance temperature Pt100 -200~850℃ 10℃
detector (RTD) Cu50 -50~150℃ 10℃
B 400~1820℃ 500℃
E -100~1000℃ 50℃
J -100~1200℃ 50℃
K -180~1372℃ 50℃
Thermocouple N -180~1300℃ 50℃
(TC) R -50~1760℃ 500℃
S -50~1760℃ 500℃
T -200~400℃ 50℃
Wre3-25 0~2315℃ 500℃
Wre5-26 0~2310℃ 500℃

-2-