Promag 200 Operation Instruction
Promag 200 Operation Instruction
18
71423454
Products Solutions Services
2018-12-03
Serial number
2. www.endress.com/deviceviewer Endress+Hauser
Operations App
3.
A0023555
2 Endress+Hauser
Flowmeter Proline 200 Brief Operating Instructions for the flowmeter
Endress+Hauser 3
Table of contents Flowmeter Proline 200
Table of contents
1 Document information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Mounting the pressure measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Mounting the transmitter of the remote version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.3 Turning the transmitter housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.4 Turning the display module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.5 Transmitter post-installation check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.1 Connection conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.2 Connecting the measuring device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Ensuring the degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.4 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6 Operation options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.1 Overview of operation options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.2 Structure and function of the operating menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.3 Access to the operating menu via the local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.4 Access to the operating menu via the operating tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7 System integration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.2 Switching on the measuring device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.3 Setting the operating language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.4 Configuring the measuring device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.5 Defining the tag name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.6 Protecting settings from unauthorized access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8.7 Application-specific commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9 Diagnostic information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4 Endress+Hauser
Flowmeter Proline 200 Document information
1 Document information
1.1 Symbols used
1.1.1 Safety symbols
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Permitted Preferred
Procedures, processes or actions that Procedures, processes or actions that
are permitted. are preferred.
Forbidden Tip
Procedures, processes or actions that Indicates additional information.
are forbidden.
Endress+Hauser 5
Document information Flowmeter Proline 200
Symbol Meaning
LED LED
Light emitting diode is on. Light emitting diode is flashing.
Open-ended wrench
6 Endress+Hauser
Flowmeter Proline 200 Basic safety instructions
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
Measuring devices for use in hazardous areas, in hygienic applications or where there is an
increased risk due to process pressure, are labeled accordingly on the nameplate.
To ensure that the measuring device remains in proper condition for the operation time:
‣ Keep within the specified pressure and temperature range.
‣ Only use the measuring device in full compliance with the data on the nameplate and the
general conditions listed in the Operating Instructions and supplementary documentation.
‣ Based on the nameplate, check whether the ordered device is permitted for the intended
use in the hazardous area (e.g. explosion protection, pressure vessel safety).
‣ Use the measuring device only for media to which the process-wetted materials are
sufficiently resistant.
‣ If the measuring device is not operated at atmospheric temperature, compliance with the
relevant basic conditions specified in the associated device documentation is absolutely
essential: "Documentation" section.
‣ Protect the measuring device permanently against corrosion from environmental
influences.
Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage caused
by improper or non-designated use.
LWARNING
Danger of breakage due to corrosive or abrasive fluids!
‣ Verify the compatibility of the process fluid with the sensor material.
‣ Ensure the resistance of all fluid-wetted materials in the process.
‣ Keep within the specified pressure and temperature range.
Endress+Hauser 7
Basic safety instructions Flowmeter Proline 200
NOTICE
Verification for borderline cases:
‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in
verifying the corrosion resistance of fluid-wetted materials, but does not accept any
warranty or liability as minute changes in the temperature, concentration or level of
contamination in the process can alter the corrosion resistance properties.
Residual risks
LWARNING
The electronics and the medium may cause the surfaces to heat up. This presents a burn
hazard!
‣ For elevated fluid temperatures, ensure protection against contact to prevent burns.
2.6 IT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any
inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
8 Endress+Hauser
Flowmeter Proline 200 Product description
3 Product description
The device consists of a transmitter and a sensor.
Two device versions are available:
• Compact version – transmitter and sensor form a mechanical unit.
• Remote version - transmitter and sensor are mounted in separate locations.
For detailed information on the product description, see the Operating Instructions for
the device
4 Installation
For detailed information about mounting the sensor, see the Sensor Brief Operating
Instructions → 3
LCAUTION
Excessive force can damage the housing!
‣ Avoid excessive mechanical stress.
The transmitter of the remote version can be mounted in the following ways:
• Wall mounting
• Pipe mounting
Endress+Hauser 9
Installation Flowmeter Proline 200
! 8.6 (0.39)
80 (3.15)
M8
19 (0.6) 80 (3.15)
A0033484
1 mm (in)
! 20…70
(! 0.79 to 2.75)
4x SW 13
A0033486
2 mm (in)
10 Endress+Hauser
Flowmeter Proline 200 Installation
2.
max. 350°
1. 3.
8 mm 8 mm
A0032242
4. 1.
1
3.
2.
–
+
E
3 mm
A0032238
1. Loosen the securing clamp of the electronics compartment cover using an Allen key.
2. Unscrew cover of the electronics compartment from the transmitter housing.
3. Optional: pull out the display module with a gentle rotational movement.
4. Turn the display module to the desired position: max. 8 × 45° in every direction.
5. Without display module pulled out:
Allow display module to engage at desired position.
Endress+Hauser 11
Installation Flowmeter Proline 200
12 Endress+Hauser
Flowmeter Proline 200 Electrical connection
5 Electrical connection
Electrical safety
In accordance with applicable federal/national regulations.
Signal cable
Current output 4 to 20 mA HART
A shielded cable is recommended. Observe grounding concept of the plant.
Current output 4 to 20 mA
Standard installation cable is sufficient.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Current input
Standard installation cable is sufficient.
Cable diameter
• Cable glands supplied:
M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Plug-in spring terminals for device version without integrated overvoltage protection: wire
cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Screw terminals for device version with integrated overvoltage protection: wire cross-
sections 0.2 to 2.5 mm2 (24 to 14 AWG)
Endress+Hauser 13
Electrical connection Flowmeter Proline 200
Standard cable 2 × 2 × 0.5 mm2 (22 AWG) PVC cable with common shield (2 pairs, pair-stranded) 1)
Cable length 5 m (16 ft), 10 m (32 ft), 20 m (65 ft), 30 m (98 ft)
Operating temperature When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable can
move freely: –25 to +105 °C (–13 to +221 °F)
1) UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to sun as much
as possible.
Cable, reinforced 2 × 2 × 0.34 mm2 (22 AWG) PVC cable with common shield (2 pairs, pair-stranded)
and additional steel-wire braided sheath 1)
Cable length 5 m (16 ft), 10 m (32 ft), 20 m (65 ft), 30 m (98 ft)
Operating temperature When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable can
move freely: –25 to +105 °C (–13 to +221 °F)
1) UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to sun as much
as possible.
Standard cable [(3 × 2) + 1] × 0.34 mm2 (22 AWG)PVC cable with common shield (3 pairs, pair-
stranded) 1)
14 Endress+Hauser
Flowmeter Proline 200 Electrical connection
Operating temperature When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable can
move freely: –25 to +105 °C (–13 to +221 °F)
1) UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to sun as much
as possible.
Endress+Hauser 15
Electrical connection Flowmeter Proline 200
Transmitter
4-20 mA HART connection version with additional inputs and outputs
3 2 1 4 3 2 1 4
5 6 3 4 1 2 5 6 3 4 1 2
+ – + – + – + – + – + –
A0033475 A0033475
Maximum number of terminals Maximum number of terminals for order code for
Terminals 1 to 6: "Accessory mounted", option NA "Overvoltage
Without integrated overvoltage protection protection"
• Terminals 1 to 4:
With integrated overvoltage protection
• Terminals 5 to 6:
Without integrated overvoltage protection
Pulse/frequency/switch
Option B 1) 4-20 mA HART (passive) -
output (passive)
16 Endress+Hauser
Flowmeter Proline 200 Electrical connection
+ –
1 1 2 3 4
BN WH YE GN
1 1 2 3 4
+ –
A0033476
3 Terminals for connection compartment in the transmitter wall holder and the sensor connection
housing
1 Terminals for connecting cable
2 Grounding via the cable strain relief
Endress+Hauser 17
Electrical connection Flowmeter Proline 200
2 Grounding White
BN WH GN RD BK YE BU
1 1 2 3 4 5 6 7
– + RES GND VCC + –
A0034571
4 Terminals for connection compartment in the transmitter wall holder and the sensor connection
housing
1 Terminals for connecting cable
2 Grounding via the cable strain relief
3 Reset Green
5 Grounding Black
18 Endress+Hauser
Flowmeter Proline 200 Electrical connection
Supply voltage
Transmitter
An external power supply is required for each output.
1) In event of external supply voltage of the power supply unit with load
2) The minimum terminal voltage increases if local operation is used: see the following table
3) Voltage drop 2.2 to 3 V for 3.59 to 22 mA
Order code for "Sensor version; DSC sensor; measuring tube", option DA, DB, DC,
DD: + DC 1 V
Mass (pressure-/temperature-compensated)
Load
Load for current output: 0 to 500 Ω, depending on the external supply voltage of the power
supply unit
Calculation of the maximum load
Depending on the supply voltage of the power supply unit (US), the maximum load (RB)
including line resistance must be observed to ensure adequate terminal voltage at the device.
In doing so, observe the minimum terminal voltage
Endress+Hauser 19
Electrical connection Flowmeter Proline 200
500
400
300
200
100
0 Us [V]
12 14 16 18 20 22 24 26 28 30 32 34 36
A0033472
Sample calculation
Supply voltage of power supply unit:
– US = 19 V
– Uterm. min = 12 V (measuring device) + 1 V (local operation without lighting) = 13 V
Maximum load: RB ≤ (19 V - 13 V): 0.022 A = 273 Ω
The minimum terminal voltage (UKl min) increases if local operation is used.→ 19.
NOTICE
Insufficient sealing of the housing!
Operational reliability of the measuring device could be compromised.
‣ Use suitable cable glands corresponding to the degree of protection.
20 Endress+Hauser
Flowmeter Proline 200 Electrical connection
2. 1. 5. 6. 3.
3 mm 20 mm
10 (0.4)
4. mm (in)
A0032239
Endress+Hauser 21
Electrical connection Flowmeter Proline 200
3. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
4. Strip the cable and cable ends. In the case of stranded cables, also fit ferrules.
5. Connect cable in accordance with terminal assignment→ 16.. For HART
communication: when connecting the cable shielding to the ground clamp, observe the
grounding concept of the facility.
6. LWARNING
Housing degree of protection may be voided due to insufficient sealing of the
housing.
‣ Screw in the screw without using any lubricant. The threads on the cover are coated
with a dry lubricant.
Removing a cable
3 (0.12)
3 4 1 2
A0032240
‣ To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between
the two terminal holes while simultaneously pulling the cable end out of the terminal.
22 Endress+Hauser
Flowmeter Proline 200 Electrical connection
The following procedure (in the action sequence given) is recommended for the remote
version:
1. Mount the sensor and transmitter.
2. Connect the connecting cable for the remote version.
3. Connect the transmitter.
How the connecting cable is connected in the transmitter housing depends on the
measuring device approval and the version of the connecting cable used.
In the following versions, only terminals can be used for connection in the transmitter
housing:
• Certain approvals: Ex nA, Ex ec, Ex tb and Division 1
• Use of reinforced connecting cable
• Order code for "Sensor version; DSC sensor; measuring tube", option DA, DB, DC, DD
In the following versions, an M12 device connector is used for connection in the
transmitter housing:
• All other approvals
• Use of connecting cable (standard)
Terminals are always used to connect the connecting cable in the sensor connection
housing (tightening torques for screws for cable strain relief: 1.2 to 1.7 Nm).
2.
1.
3 mm
A0034167
Endress+Hauser 23
Electrical connection Flowmeter Proline 200
4.
3.
A0034171
6 Sample graphic
24 Endress+Hauser
Flowmeter Proline 200 Electrical connection
A0034172
2. 3. 1.
–
+
E
3 mm
A0034173
Endress+Hauser 25
Electrical connection Flowmeter Proline 200
4.
TX 10
5.
8 mm
A0034174
6.
~15°
7.
A0034175
7 Sample graphic
6. Turn the transmitter housing to the right until it reaches the marking.
7. NOTICE
The connection board of the wall housing is connected to the electronics board of
the transmitter via a signal cable!
‣ Pay attention to the signal cable when lifting the transmitter housing!
Lift the transmitter housing.
26 Endress+Hauser
Flowmeter Proline 200 Electrical connection
8.
A0034176
8 Sample graphic
10.
9.
A0034177
9 Sample graphic
Endress+Hauser 27
Electrical connection Flowmeter Proline 200
11. Connect the cable shield via the cable strain relief.
12. Tighten the screws for the cable strain relief using a torque in the range of
1.2 to 1.7 Nm.
13. Reverse the removal procedure to reassemble the transmitter housing.
5.2.3 Connecting the connecting cable for the pressure measuring cell
When delivered to the customer, the connecting cable is connected as follows:
• Compact version: to transmitter housing
• Remote version: to sensor connection housing
For connection to sensor and pressure measuring cell:
‣ Insert M12 plug of connecting cable into pressure measuring cell and screw into place.
5.2.4 Ensuring potential equalization
Requirements
Please consider the following to ensure correct measurement:
• Same electrical potential for the medium and sensor
• Remote version: same electrical potential for the sensor and transmitter
• Company-internal grounding concepts
• Pipe material and grounding
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following steps
after the electrical connection:
1. Check that the housing seals are clean and fitted correctly.
2. Dry, clean or replace the seals if necessary.
3. Tighten all housing screws and screw covers.
4. Firmly tighten the cable glands.
5. To ensure that moisture does not enter the cable entry:
Route the cable so that it loops down before the cable entry ("water trap").
A0029278
Endress+Hauser 29
Operation options Flowmeter Proline 200
6 Operation options
6.1 Overview of operation options
ESC
- + E
1 2 3 4 5
A0032226
30 Endress+Hauser
Flowmeter Proline 200 Operation options
Operator
Language !
Maintenance
Operation
task-oriented
Setup
Diagnostics
Expert function-oriented
A0014058-EN
Endress+Hauser 31
Operation options Flowmeter Proline 200
1 2 XXXXXXX 2.1
2.3 19.184 mA
2.2
12.5 2.4
2.6 l/s
2.5
E 4 Xxxxxxx
A B C D E F G H I J K
L M N O P Q R S T U V W
X Y Z
1.7 5 +0.000 Xx
0 1 2 3 4
5 6 7 8 9
+ –
A0014013
1 Operational display with measured value shown as "1 value, max." (example)
1.1 Device tag
1.2 Display area for measured values (4-line)
1.3 Explanatory symbols for measured value: Measured value type, measuring channel number, symbol
for diagnostic behavior
1.4 Status area
1.5 Measured value
1.6 Unit for the measured value
1.7 Operating elements
2 Operational display with measured value shown as "1 bar graph + 1 value" (example)
2.1 Bar graph display for measured value 1
2.2 Measured value 1 with unit
2.3 Explanatory symbols for measured value 1: measured value type, measuring channel number
2.4 Measured value 2
2.5 Unit for measured value 2
2.6 Explanatory symbols for measured value 2: measured value type, measuring channel number
3 Navigation view: picklist of a parameter
3.1 Navigation path and status area
3.2 Display area for navigation: designates the current parameter value
4 Editing view: text editor with input mask
5 Editing view: numeric editor with input mask
32 Endress+Hauser
Flowmeter Proline 200 Operation options
• Depends on the device version, e.g.: The following symbols appear in the status area of the
– : Volume flow operational display at the top right:
– : Mass flow • Status signals
– : Density – F: Failure
– G: Conductivity – C: Function check
– : Temperature – S: Out of specification
• : Totalizer – M: Maintenance required
• : Output • Diagnostic behavior
• : Input – : Alarm
• : Measurement channel number 1) – : Warning
• Diagnostic behavior 2) • : Locking (locked via hardware))
– : Alarm • : Communication via remote operation is active.
– : Warning
1) If there is more than one channel for the same measured variable type (totalizer, output etc.).
2) For a diagnostic event that concerns the displayed measured variable.
The following appears in the status area of the navigation • Icons for menus
view in the top right corner: – : Operation
• In the submenu – : Setup
– The direct access code for the parameter you are – : Diagnostics
navigating to (e.g. 0022-1) – : Expert
– If a diagnostic event is present, the diagnostic • ▸: Submenus
behavior and status signal • : Wizards
• In the wizard • : Parameters within a wizard
If a diagnostic event is present, the diagnostic behavior • : Parameter locked
and status signal
Exits the input without applying the changes. Moves the input position one position to the
right.
Clears all entered characters. Moves the input position one position to the left.
Switches to the selection of the correction tools. Deletes one character immediately to the left of
the input position.
Aa1 Toggle
• Between upper-case and lower-case letters
• For entering numbers
• For entering special characters
Endress+Hauser 33
Operation options Flowmeter Proline 200
Numeric editor
Confirms selection. Moves the input position one position to the left.
Exits the input without applying the changes. . Inserts decimal separator at the input position.
– Inserts minus sign at the input position. Clears all entered characters.
Enter key
Minus key
Plus key
In a menu, submenu
• Pressing the key briefly
– Exits the current menu level and takes you to the next higher level.
– If help text is open, closes the help text of the parameter.
• Pressing the key for 2 s for the parameter: Returns you to the operational display ("home position").
With a wizard: Exits the wizard and takes you to the next higher level.
With a text and numeric editor: Closes the text or numeric editor without applying changes.
34 Endress+Hauser
Flowmeter Proline 200 System integration
+ Plus/Enter key combination (press and hold down the keys simultaneously)
7 System integration
For detailed information on system integration, see the Operating Instructions for the
device.
• Overview of device description files:
– Current version data for the device
– Operating tools
• Measured variables via HART protocol
• Burst mode functionality in accordance with HART 7 Specification
8 Commissioning
8.1 Function check
Before commissioning the measuring device:
‣ Make sure that the post-installation and post-connection checks have been performed.
• "Post-installation check" checklist→ 12
• "Post-connection check" checklist → 29
Endress+Hauser 35
Commissioning Flowmeter Proline 200
If nothing appears on the local display or a diagnostic message is displayed, refer to the
Operating Instructions for the device → 2
XXXXXXXXX
20.50
XX XX
Main menu 0104-1
1. Display language
English
Operation
Setup
Display language 0104-1
2. Ã English
Deutsch
Español
Français
Display language 0104-1
3. Ã English
Deutsch
Español
Français
Hauptmenü 0104-1
4. Sprache
Deutsch
Betrieb
Setup
A0029420
36 Endress+Hauser
Flowmeter Proline 200 Commissioning
The desired units can be selected in the System units submenu. The wizards systematically
guide the user through all the parameters required for configuration, such as parameters for
measurement or outputs.
The wizards available in the particular device can vary on account of the device version
(e.g. sensor).
Wizard Meaning
Low flow cut off Set the low flow cut off
1 XXXXXXXXX
A0029422
Navigation
"Setup" menu → Device tag
Endress+Hauser 37
Commissioning Flowmeter Proline 200
Device tag Enter the name for the Max. 32 characters, such as Prowirl
measuring point. letters, numbers or special
characters (e.g. @, %, /).
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Steam option.
3. When pressure measured value is read in 1):
In the Steam calculation mode parameter, select the Automatic (p-/T-compensated)
option.
4. If pressure measured value is not read in:
In the Steam calculation mode parameter, select the Saturated steam (T-
compensated) option.
5. In the Steam quality value parameter, enter the steam quality present in the pipe.
Without Wet Steam Detection/Measurement application package: Measuring
device uses this value to calculate the mass flow of the steam.
With Wet Steam Detection/Measurement application package: Measuring device
uses this value if the steam quality cannot be calculated (steam quality is not
compliant with basic conditions).
1) Sensor version option "mass (integrated pressure and temperature measurement)", Pressure read in via current
input/HART/
38 Endress+Hauser
Flowmeter Proline 200 Commissioning
For detailed information on the basic conditions for wet steam applications, see the
Special Documentation.
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Liquid option.
3. In the Select liquid type parameter, select the User-specific liquid option.
4. In the Enthalpy type parameter, select the Heat option.
Heat option: Non-flammable liquid that serves as a heat carrier.
Calorific value option: Flammable liquid whose combustion energy is calculated.
Endress+Hauser 39
Commissioning Flowmeter Proline 200
Single gas
Combustion gas, e. g. methane CH4
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Gas option.
3. In the Select gas type parameter, select the Single gas option.
4. In the Gas type parameter, select the Methane CH4 option.
40 Endress+Hauser
Flowmeter Proline 200 Commissioning
Gas mixture
Forming gas for steel mills and rolling mills, e. g. N2/H2
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Gas option.
3. In the Select gas type parameter, select the Gas mixture option.
Air
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Gas option.
3. In the Select gas type parameter, select the Air option.
The density is determined according to NEL 40.
Endress+Hauser 41
Commissioning Flowmeter Proline 200
Endress+Hauser recommends the use of active pressure compensation. This fully rules
out the risk of measured errors due to pressure variations and incorrect entries .
Natural gas
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Gas option.
3. In the Select gas type parameter, select the Natural gas option.
4. In the Fixed process pressure parameter, enter the value of the process pressure
present.
5. In the Enthalpy calculation parameter, select one of the following options:
AGA5
ISO 6976 option (contains GPA 2172)
6. In the Density calculation parameter, select one of the following options.
AGA Nx19
ISO 12213- 2 option (contains AGA8-DC92)
ISO 12213- 3 option (contains SGERG-88, AGA8 Gross Method 1)
42 Endress+Hauser
Flowmeter Proline 200 Commissioning
Endress+Hauser recommends the use of active pressure compensation. This fully rules
out the risk of measured errors due to pressure variations and incorrect entries .
Ideal gas
The unit "corrected volume flow" is often used to measure industrial gas mixtures, in particular
natural gas. To do so, the calculated mass flow is divided by a reference density. To calculate
the mass flow, knowledge of the exact composition of the gas is essential. In practice,
however, this information is often not available (e. g. as it varies over time). In this case, it can
be useful to regard the gas as an ideal gas. This means that only the operating temperature
and operating pressure variables as well as the reference temperature and reference pressure
variables are needed to calculate the corrected volume flow. The error resulting from this
assumption (typically 1 to 5 %) is often considerably smaller than the error caused by
inaccurate composition data. This method should not be used for condensing gases (e. g.
saturated steam).
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Gas option.
3. In the Select gas type parameter, select the User-specific gas option.
4. For non-flammable gas:
In the Enthalpy type parameter, select the Heat option.
Endress+Hauser 43
Diagnostic information Flowmeter Proline 200
6. In the Reference density parameter, enter the reference density of the fluid.
7. In the Reference pressure parameter, enter the reference pressure of the fluid.
8. In the Reference temperature parameter, enter the fluid temperature associated with
the reference density.
9. In the Reference Z-factor parameter, enter the value 1.
10. If specific heat capacity is to be measured:
In the Specific heat capacity parameter, enter the heat capacity of the fluid.
11. In the Z-factor parameter, enter the value 1.
12. In the Dynamic viscosity parameter, enter the viscosity of the fluid under operating
conditions.
9 Diagnostic information
Faults detected by the self-monitoring system of the measuring device are displayed as a
diagnostic message in alternation with the operational display. The message on remedial
measures can be called up from the diagnostic message, and contains important information
on the fault.
44 Endress+Hauser
Flowmeter Proline 200 Diagnostic information
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
X XX i Menu
1.
Diagnostic list S
Diagnostics 1
1 S801 Supply voltage
Diagnostics 2
Diagnostics 3
2.
3.
A0029431-EN
Endress+Hauser 45
www.addresses.endress.com