175 0010 13 DECADE Elite Lite User Manual
175 0010 13 DECADE Elite Lite User Manual
Hoorn 131
2404 HH Alphen a/d Rijn
The Netherlands
DECADE Elite/Lite
User Manual
Copyright ©2022, Antec, The Netherlands. Contents of this publication may not be reproduced in any form or by any means
(including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent
from the copyright of the owner. Antec reserves the right to make changes to the design and specifications of the instrument
and this manual without prior notice.
DECADE Elite, DECADE Lite, ROXY, ALEXYS, ROXY Exceed, INTRO, FlexCell, ReactorCell, µ-PrepCell, ISAAC, HyREF,
SenCell and SynthesisCell are trademarks of Antec Scientific. Whatman™ (word and device) and Whatman™ (word only) are
trademarks of Whatman international Ltd. SOLVENT IFD™ and AQUEOUS IFD™ are trademarks of Arbor Technologies, Inc.
Clarity®, DataApex® are trademarks of DataApex Ltd. Microsoft® and Windows™ are trademarks of Microsoft Corporation. Ex-
cel is a registered trademark of the Microsoft Corporation. Thermo Scientific ChromeleonTM is a trademark of ThermoFisher
Scientific. EmpowerTM is a trademark of Waters Corporation. Agilent OpenLabTM is a trademark of Agilent Technologies Inc.
All other trademarks are the property of their respective owners.
The software and the information provided herein is believed to be reliable. Antec shall not be liable for errors contained
herein or for incidental or consequential damages in connection with the furnishing, performance, or use of software or this
manual. All use of the software shall be entirely at the user’s own risk.
Declaration of conformity
We Antec Scientific, Alphen a/d Rijn, The Netherlands, declare that the product:
EN 61010-1:2010
Safety requirements for electrical equipment for measurement, control, and laboratory use; part 1:
General Requirements.
EN 61010-2-010:2014
Safety requirements for electrical equipment for measurement, control, and laboratory use; part 2-
010: Particular requirements for laboratory equipment for the heating of materials.
EN 61326-1:2013
Electrical equipment for measurement, control and laboratory use - EMC requirements; part 1: Gen-
eral Requirements (Class B equipment).
EN 55032:2015
Electromagnetic compatibility of multimedia equipment
Emission requirements (Class B equipment).
Only use manufacturer-supplied cable(s) to connect with other devices. Thoroughly connect
shielding to common. Manufacturer will not accept any liability for damage, direct or indirect,
caused by connecting this instrument to devices and with cables which do not meet relevant
safety standards. All cables used should not exceed a length of 3 meters.
This DoC applies to above-listed products (2020 model). A 2020 model can be identified by the letter
B in the part number. For older models see previous revision of this manual.
Table of contents
Symbols 7
Intended use 9
WEEE 10
ROHS 10
ISO 10
Safety instructions 11
CHAPTER 1 - Introduction 15
Instrument description 16
DECADE Elite – Front side 16
DECADE Lite – Front side 17
DECADE Elite – Back side 18
DECADE Elite – Oven compartment 19
CHAPTER 2 - Installation 21
Storage requirements 21
Site Preparation Requirements 21
Personal Computer 21
Laboratory requirements 22
Electrical and power requirements 24
Chemicals 25
Unpacking 26
Mains connection 28
PC connection 28
Software 32
Dialogue Elite 33
Clarity Chromatography software 34
HPLC fluid connections 37
Tubing connectors 37
Mobile Phase 39
Installation & start-up 41
DIAG menu 54
Parameters 55
Dialogue Elite 60
Clarity CDS 62
Chromeleon CDS 63
OpenLab CDS 64
Empower CDS 65
INDEX 155
CHAPTER 1 Table of contents 7
Symbols
The following symbols might be used in this guide or may be found on the
instrument:
The toxic hazard sign draws attention to the fact that use
of toxic solvents or samples may carry a significant
health risk.
8 175_0010 - DECADE Elite / Lite user manual - rev 13
Intended use
The DECADE Elite /Lite Electrochemical Detector is used in combination
with (Ultra) High Performance Liquid Chromatography for the electrochem-
ical detection of suitable analytes in liquid samples. With this technique the
amount of electroactive substances in mobile phase can be quantified. The
instrument can be used for the chromatographic analysis of a wide range
of electroactive analytes in the fields of for example:
• Bioanalytical analyses
• Food analyses
• Environmental analyses
WEEE
All equipment of Antec Scientific which are subjected to the WEEE di-
rective shipped after August 13, 2005 are compliant with the WEEE mark-
ing requirements. Such products are labelled with the “crossed out
wheelie”, depicted on the left site.
The symbol on the product indicates that the product must not be
disposed as unsorted municipality waste.
Antec Scientific
Hoorn 131
2404 HH Alphen a/d Rijn
The Netherlands
ROHS
ISO
Safety instructions
Adhere to the following standard quality control procedures and the follow-
ing equipment guidelines when using the DECADE Elite / Lite detector.
The following safety practices are intended to ensure safe operation of the
instrument.
System Operation
To assure optimal performance keep of the detector we recommend that
the instrument is checked regularly and maintenance procedures are car-
ried out. Preventive maintenance contracts are available for that purpose.
Please contact your local dealer or the nearest sales office for more infor-
mation.
Electrical safety
The removal of protective panels on the instrument can result in exposure
to potentially dangerous voltages. Therefore, disconnect the instrument
from all power sources before disassembly.
Untrained personnel should not open the instrument, this may only be
done by authorized service engineers. Replace or repair faulty insula-
tion on power cords immediately after discovery of the fault. Check that the
actual power voltage is the same as the voltage for which the instruments
are wired. Make sure power cords are connected to correct voltage
sources: grounded AC power source, line voltage 100 – 240 VAC. The in-
strument should be connected to a protective earth via a ground socket.
12 175_0010 - DECADE Elite / Lite user manual - rev 13
The DECADE Elite must only be used with appliances and power sources
with proper protective grounding to prevent damage through build-up of
static electricity. The power source should exhibit minimal power transients
and fluctuations. If necessary connect to a filtered mains socket.
EMPLOYER UNIQUEMENT
AVEC DES FUSIBLES DE 250V
Replace blown fuses with fuses of proper type and rating as indicated on
the rear panel and as listed in the list of accessories and spares (appendix
D). The fuse holder is integrated in the mains connector. Ensure that the
instrument is never put in operation with fuses of a different type. This
could cause fire.
Solvents
The solvents used may be flammable, toxic or corrosive. The room in
which the system is installed should be well ventilated to prevent that sol-
vent vapors cause poisoning or ignite and cause a fire. Use of open fire in
the vicinity of this system must be strictly prohibited. Do not install the sys-
tem in the same room with any other equipment that emits or could poten-
tially emit sparks. Provide protective equipment near the instrument, when
solvent gets into the eyes or on the skin, it must be flushed away immedi-
ately. Provide equipment, such eye wash stations and safety showers, as
close to system as possible. Use proper eye and skin protection when
CHAPTER 1 Table of contents 13
Biological Hazard
When you analyze biological fluids you need possible precautions and
treat all specimens as potentially infectious. Always wear protective
And gloves when handling toxic or biologically infectious samples to pre-
vent bio hazards or hazards while working with the DECADE Elite .If nec-
essary the instrument must be decontaminated before decommis-
sioning or shipment of the instrument for repair to Antec or its repre-
sentatives. When shipped to Antec every instrument must be accompa-
nied with a decontamination form which should be completely filled in and
signed by the customer. Without this decontamination form the instrument
will not be processed by Antec (either repaired or disposed).
Waste disposal
Perform periodic leak checks on LC tubing and connections. Do not close
or block the drain in the oven compartment. Do not allow flammable and/or
toxic solvents to accumulate. Follow a regulated, approved waste
Disposal program. Never dispose of flammable and/ toxic solvents through
the municipal sewage system.
Using the DECADE Elite / Lite in other ways than indicated in the
manual or defined by good laboratory practice may result in erratic or
unsafe operation.
14 175_0010 - DECADE Elite / Lite user manual - rev 13
CHAPTER 1 Introduction 15
C H A P T E R 1
Introduction
Congratulations on your purchase of the DECADE Elite or DECADE Lite.
This detector enables you to perform all (U)HPLC applications using elec-
trochemical detection. The DECADE Elite includes a highly stable Fara-
day-shielded oven compartment accommodating column and flow cell.
This flow cell has surprised researchers for its unsurpassed S/N ratio and
therefore you now possess the best possible combination for extremely
sensitive EC analyses. The DECADE Elite has 3 operational measurement
modes: DC, SCAN and PULSE mode. Furthermore, a Diag(nostics) and
Config(uration) mode are available. In addition, crucial parameters can be
controlled by either relays or TTL. The DECADE Elite can support up to 3
flow cells (optional), which makes it possible to perform 3 independent
measurements with one detector. The DECADE Lite (right side) is a “light”
version of the Elite with a front panel without keyboard and display, it is for
single flow cell operations in DC mode only. For the DECADE Lite PC soft-
ware for parametric control and data-acquisition is required.
Instrument description
DECADE Elite – Front side
# Description # Description
1 Instrument housing 7 '+' and '-' value keys
2 LC tubing inlet/outlet 8 Cursor keys
3 Instrument door panel 9 Door handle (for opening door)
4 4 x 40 Ch LCD display
5 Function keys
6 <Enter> key
CHAPTER 1 Introduction 17
# Description # Description
1 Instrument housing
2 LC tubing inlet/outlet
3 Instrument door panel
4 Door handle (for opening door)
18 175_0010 - DECADE Elite / Lite user manual - rev 13
# Description # Description
1 Instrument rear panel 7 USB connector (USB B)
2 Type label (PN, SN etc.) 8 Fuse & power rating
3 Digital I/O connector (25-p sub-D fem) 9 Mains switch/inlet
4 Analog data (9-pins sub-D fem) 10 Grounding stud
5 Valve connector (9-pins sub-D male) 11 Ventilation holes
6 LAN connector (RJ45 jack) 12 Fuse compartment
CHAPTER 1 Introduction 19
# Description # Description
1 Cell cabinet 8 Door lock
2* Cell connector (9-pin sub-D) 9 Mounting hole M4 (cell clamp)
3 Top fan heater (intake) 10 Bottom fan heater (exhaust)
4 Door sensor 11 Mounting hole M3 (column clamp)
5 Mounting plate 12 Drain
6 Column clamp 13 Door panel, rear
7 Flow cell clamp (SenCell) 14 Type label
*) The Lite version has 1 flow cell (1 cell connector) only.
20 175_0010 - DECADE Elite / Lite user manual - rev 13
CHAPTER 2 Installation 21
C H A P T E R 2
Installation
Storage requirements
The DECADE Elite is shipped in one box to your facility with the following
dimensions:
Make sure to have sufficient space to store the packed instrument under
the following storage conditions:
Parameter Requirement
Storage temperature -25 – 50 °C (-14 – 122 °F)
Storage humidity 20 – 80%, non-condensing
Personal Computer
In case the instrument is used via remote control by PC software (Dialogue
Elite, DataApex Clarity, Waters Empower, Agilent OpenLab CDS, Agilent
OpenLab Chemstation or Thermo Chromeleon CDS) or a firmware (FW)
update needs to be performed a desktop computer is required with the fol-
lowing requirements:
22 175_0010 - DECADE Elite / Lite user manual - rev 13
- Screensavers
- (USB, LAN) hibernate mode
- Auto hard disk shut down (energy saving)
- Automatic Windows updates
- Avoid exhaustive scanning by virus scanners*
Laboratory requirements
Your instrument is intended for indoor use only in an industrial or commer-
cial environment (EN55011 group 1 class B ISM equipment). It is suitable
for the following categories: Installation category II, Pollution degree 2,
equipment class I.
CHAPTER 2 Installation 23
Parameter Requirement
Operating temperature 10 – 35 °C (50 – 104°F)
Maximum Altitude < 2000 meter (< 7500 ft)
Operating humidity 20 – 80%, non-condensing
Do not place the system next to heating or cooling pipes or expose the
instrument to direct sun light or expose it to air drafts (AC system /
open windows).
Requirements for the laboratory bench on which the instrument will be in-
stalled:
For regions with deviating mains plugs/sockets like (for example UK,
Switzerland, Brazil etc.) make sure to have the appropriate power
cords available on-site at the date of installation. Make sure these
power cables are properly grounded and meet the relevant safety
standards which apply in your country. In case of questions contact
your local distributor.
Chemicals
Mobile phase and flush/storage solutions must be clean as it is in direct
contact with the working electrode of the electrochemical reactor cell. High
purity chemicals including water is a prerequisite. So all chemicals should
be electrochemically clean, HPLC grade or better. For water used for the
preparation of mobile phases a water purification apparatus is advised
which is able to supply high purity deionized water with resistivity of >18
MOhm.cm and low TOC level (<10 ppb).
26 175_0010 - DECADE Elite / Lite user manual - rev 13
Unpacking
Inspect the transport box for possible damage as it arrives. Immediately in-
form the transport company in case of damage, otherwise she may not ac-
cept any responsibility. Keep the transport box as it is designed for opti-
mum protection during transport and it may be needed again. Carefully un-
pack the system and inspect it for completeness and for possible damage.
Contact your supplier in case of damage or if not all marked items on the
checklist are included. Prior to shipment, your detector has been thor-
oughly inspected and tested to meet the highest possible demands. The
results of all tests are included.
*) Note that flow cells are not part of the DECADE Elite detector and have
to be ordered separately.
To unpack the detector, lift it from its box by both hands (Fig. 1). Never lift
the detector at its front door, but at its sides.
With both hands under the instrument lift the detector and bring it to its op-
eration location. Install the detector in an area which meets the environ-
mental conditions.
Fig. 2. Location of ventilation holes in the DECADE Elite (bottom & rear).
Remove the protective tape from the detector LCD screen. Leave the in-
strument to adopt ambient temperature for at least half an hour in the
place of installation.
In case the instrument was stored cold (< 10°C) and switched on immedi-
ately, the oven might not work. This is by design (at temperatures below
10°C the heater is switched off in the embedded software). Remedy: just
allow the detector (electronics) to warm up a little longer to adopt ambient
before switching on the oven.
Use the detector indoors only. Place the potentiostat upright (on its
instrument feet) on a stable, flat and smooth surface. Do not place the
instrument in an area subject to excessive dust or shocks.
Do not place it near a source of heat or in direct sun light, as this may
influence the heating capabilities of the instrument. Make sure the
potentiostat is placed in such a way that the mains power connection
can be reached easily to disconnect it from the mains power by
removing the Mains power cable. Do not block the ventilation holes at
the back and bottom of the instrument (Fig. 2.). Blocking might impair
the cooling capability of the power supply.
objects are placed at only one side of the instrument. and 10 cm from the
instrument, if objects are placed on both sides of the instrument.
Mains connection
Check that fuses and voltage range on the rear side of the instrument
match that for the power outlet to be used.
Leave the instrument powered off until specifically mentioned in the proce-
dure below.
PC connection
Follow the instructions in this paragraph when the instrument is used with
PC control over LAN using the Dialogue Elite, DataApex Clarity, Agilent
OpenLab CDS, Agilent OpenLab Chemstation or Thermo Chromeleon
CDS. This section can be skipped if the instrument is used stand-alone
(DECADE Elite only). To be able to communicate over LAN a computer is
required with a free (PCI, PCI Express or PCI-X) LAN port.
The DECADE Elite has a fixed IP address: 192.168.5.1, with subnet mask:
255.255.255.0. Gateway and DNS are not filled in.
1. Right click on the Network icon in the bottom right of the Windows
taskbar and open the menu 'Network & Internet settings'. Alterna-
tive: open Windows start menu, open Setting menu and open Net-
work & Internet menu.
2. Open the menu 'Change adapter settings' on the right panel of the
'Network & Internet window.
30 175_0010 - DECADE Elite / Lite user manual - rev 13
3. Right click on the Local Area Connection icon of the LAN card in
your PC and click on properties to open the Network card setting.
Software
The DECADE Elite can be used remote in combination with PC control
software. The following software packages are available for control of the
DECADE Elite electrochemical detector:
1) Besides a valid CDS license an additional (paid) license is required to enable the detector
control driver in the CDS. Contact Antec Scientific for detailed info and pricing.
2) The Chromeleon CDS license should include an Instrument class 3 license to be able to
control the DECADE Elite.
3) The Empower 3 license should also include a Agilent LC control license to be able to
control the DECADE Elite. Empower ICF Support Layer V3.5 or higher is required.
†) Version compatibility: compatible with the listed version and up.
In the chromatography CDS drivers only DC and Pulse mode are sup-
ported (no SCAN mode support), and up to a maximum of two flow cells
can be controlled (exception: in Clarity up to 3 flow cells). Only DECADE
Elites with FW version 1.09 or higher can be controlled with the CDS soft-
ware.
The Antec Dialogue Elite software is used for instrument control and Quali-
fications and is also required for the upload of new firmware (= embedded
software controlling the DECADE Elite electronics). In this section the in-
stallation of Dialogue Elite and configuration is shortly described. Note: this
is by no means a replacement of the installation documentation available
for the software packages. Please refer to the software documentation for
details.
Dialogue Elite
To install the software:
• Download the latest version of the Dialogue Elite from the Antec
website www.antecscientific.com (register to get access).
• Double click on the setup.exe file to start the installation wizard
• Follow the instructions of the installation wizard for successful
installation of the software.
• Insert the Dialogue Elite license dongle to get full access to the
software (without dongle it will operate in demo mode).
• Make sure that the LAN connection is configured and the LAN
cable is connected.
• Power up the detector by means of the mains switch on the rear
panel.
• Start the program Dialogue Elite from the Windows start menu.
• During start-up the ‘Select devices’ menu will pop-up as shown
below.
• The pull down field shows all responding devices with their IP and
MAC address. In case of doubt check the device unique MAC
address on the rear panel IO connector.
• Type OK. The instrument will connect and is ready for use.
Use proper eye and skin protection when working with solvents. The
solvents used may be flammable, toxic or corrosive. Organic sol-
vents are toxic above a certain concentration. Ensure that work areas
are always well-ventilated! Use of open fire in the vicinity of this sys-
tem must be strictly prohibited. Do not install the system in the same
room with any other equipment that emits or could potentially emit
sparks. Wear protective gloves, safety glasses and other relevant
protective clothing when working on the device!
The manufacturer will not accept any liability for damage, direct or
indirect, caused by connecting this instrument to devices that do not
meet the relevant safety standards.
Tubing connectors
For optimal operation it is of the utmost importance that all tubing connec-
tions on the injection valve, columns and flow cells of the injection valves
are made without introducing dead volumes to minimize peak broadening,
carry-over etc.
Use only the original polymeric fingertights supplied by Antec with the
purchased flow cell to make LC connections on the flow cells inlet and
outlet. Do not use metal tubing on the flow cell because it may lead to
damage or incorrect operation of the flow cell. Use PEEK, PEEKsil or
Fused Silica tubing (with FEP sleeves).
For columns and injection valves etc. use only nuts, ferrules or finger tights
recommend by the respective manufacturer of the parts. So for Vici Valco
38 175_0010 - DECADE Elite / Lite user manual - rev 13
valves use Valco nuts and ferrules, for Rheodyne valves use Rheodyne
valves etc. The use of unsuitable connectors may lead to damage of the
parts or the introduction of dead volumes. In case the DECADE Elite was
purchased as a part of an ALEXYS HPLC-ECD analyzer, the instrument is
shipped with a complete set of dedicated tubing assemblies (LC connec-
tion kits) tailor-made for the application the analyzer was specifically de-
fined for.
Please note that the tubing length (length L below) required to make a
good connection differs for each brand of connection. If length L is not cor-
rect, this will result in faulty peaks and carry-over. Essentially, when you
create a connection the ferrule on the tubing is compressed into the valve
to make sure that the connection is leak-tight. Take the following into ac-
count when creating the connection:
If L is too long, the ferrule cannot form a seal in the connection. This may
cause
Irreparable damage port of the valve, column or other part, as:
• Leakage
• Dead volume at the end of the ferrule (a ‘mixing chamber’).
Mobile Phase
Electrochemical detection is a sensitive detection technique characterized
by extremely low detection limits. A typical detection limit of 100 pmol/L or
lower for catecholamines is no exception. Improving detection limits will al-
ways be limited by the weakest link in an LC-EC system. In daily practice a
couple of ‘rules’ must be obeyed to fully exploit the incredible linear dy-
namic range and low detection limits of an EC detector. These are not only
hardware related, but also refer to mobile phase composition, degassing,
temperature and pH stability and several other issues.
Mobile phase must be clean as it is in direct contact with the working elec-
trode in EC detection. High purity chemicals including water is a pre requi-
site. In some applications EDTA is added to the mobile phase to traps
electrochemically metals such as Fe2+ by forming an inactive complex.
40 175_0010 - DECADE Elite / Lite user manual - rev 13
However at higher working potentials (typically > 1.2 V vs. salt bridge AgCl
ref) also EDTA can become electrochemically active and is not recom-
mended. In that case only a passivation step with 15% HNO3 is recom-
mended (see installation guide).
Electrolytes assure contact between 3 electrodes in an electrochemical
flow cell. Low ion strength destabilizes an EC system and noise will in-
crease. Extremely high buffer concentrations cause problems of salt for-
mation. Therefore concentrations between 20 and 200 mmol/L are recom-
mended.
20 min. The acid is flushed through the pump, the pump tubing,
the dampener, the injector (in load and inject position) and to
waste
Make sure that all parts that are not acid-resistant such as: nylon inlet
filters, column and flow cell are not connected during this step.
8. Consult your flow cell manual for detailed information about the
installation of the flow cell. Although the oven of the detector can
be operated in a temperature range from 7 °C above ambient to
60°C, do not operate the flow cells above 50°C.
9. See figure 5 for reference how to make the electrical and fluid
connections. Connect the flow cell to the corresponding cell
connector in the oven compartment. All cell connectors are
marked with a label for identification. In case of a DECADE Elite
SCC connect the flow cell to the cell connector on the left side
marked “Cell 1”.
The cell connector inside the oven compartment is ESD sensitive. Make
sure that the flow cell is OFF when removing or connecting the cell
cable.
Fig. 5. Left: SenCell with ISAAC reference mounted under an angle of ap-
proximately 45° in the detector. [1] Cell clamp, [2] Cell Inlet (tubing connec-
tion from column–to–cell), [3] Cell outlet (tubing connection from cell-to-
waste) make sure that the outlet is positioned on the top side to prevent
entrapped air bubbles, [4] WE contact (red), [5] AUX contact (blue), [6]
REF contact (black). Top-right: electrical connections of WE (red con-
nector) and AUX electrode (blue connector). Bottom-right: SenCell with
Salt Bridge reference electrode.
44 175_0010 - DECADE Elite / Lite user manual - rev 13
10. Before switching ON the flow cell, make sure that the mobile
phase contains sufficient electrolyte (buffer ions). A stable
baseline will never be obtained if the cell is switched ON with only
water or another non-conducting mobile phase. Also be sure that
no air bubbles are trapped in the flow cell.
11. The outlet tubing from the flow cell should lead to a reservoir that
is at a higher level than the flow cell. This ensures a small back
pressure which prevents air-bubble entrapment. The outlet tubing
should be under the liquid level, to avoid electrical noise induced
by ‘dripping’ of mobile phase.
12. Set the cell potential (see page 99 for optimization of the
potential), switch ON the flow cell and allow the system to stabilize
for approximately 30 min. A ‘good’ stabilization curve shows a
mono-exponential decline without jumps and/or spikes.
Your system is now ready for use. The DECADE Elite has been developed
for continuous operation. For maximum stability it is advised to leave the
system ON continuously. If preferred, the flow cell may be switched OFF at
night.
CHAPTER 3 Maintenance & Shutdown 45
C H A P T E R 3
Maintenance
In this paragraph all maintenance is described which can be performed by
the end-user, all other maintenance & service procedures may only be per-
formed by authorized service engineers only.
In case the actual temperature exceeds 70°C switch off the potentiostat
immediately do not touch the metal parts inside the oven compartment
because they could be hot, and contact the manufacturer or its
representatives for service. Do not use/switch on the instrument before
the instrument is serviced at the factory.
In the event of a heater defect and the temperature exceeds 70°C the in-
strument will come up with an error message (Err 23).
46 175_0010 - DECADE Elite / Lite user manual - rev 13
Flow cell
Check the performance of the detector & flow cell on a daily basis by eval-
uating background current, noise and signal. An increase in background
current, noise and/or loss of sensitivity may be a sign of contamination of
the working electrode (WE) and/or a sign that maintenance is required on
the reference electrode (REF) of the flow cell. If necessary perform mainte-
nance on the flow cell. Detailed information about flow cell maintenance in-
structions are described in the user manual supplied with the specific flow
cell (SenCell, VT-03 or FlexCell) purchased in combination with your de-
tector.
Cleaning
In general, the DECADE Elite needs very little maintenance. The outside
of the detector may be cleaned with a non-aggressive cleaning liquid.
Do not use any organic solvents to clean the exterior of the detector,
because this may lead to damage of the paint layer.
In case of leakage in the cell cabinet (tubing, connectors, cell, column etc.)
remove the spilled mobile phase or other solutions as soon as possible be-
cause this may damage the paint layer, or result in the deposition of salt
crusts (in case of buffered mobile phases), which could block the drain in
the bottom of the cell cabinet. Remove any dust on the protective screens
that cover the fans in the oven compartment.
Replacement of fuses
Replace blown fuses with fuses of proper type and rating as stipulated on
the rear panel and specified in the installation section of this manual. The
fuse holder is integrated in the mains connector. Ensure that the instru-
ment is never put in operation with fuses of a different type. This could
cause fire.
In case the fuses blow out repetitively contact Antec or its representatives
for instructions and/or service of the instrument.
CHAPTER 3 Maintenance & Shutdown 47
There are a couple of steps to take to switch off an LC system with electro-
chemical detector for a longer period of time. Shutting down is not different
from most other HPLC systems. Perform the following procedure:
• Switch off the flow cell using the keyboard (standalone) or via the
software (Clarity or Dialogue Elite).
• Check the column(s) documentation for the appropriate storage
liquid, apply this and make sure the column is properly flushed. A
reversed phase C18 column is usually stored with 50%
Acetonitrile/water.
• Take out the column, mount the corresponding end-caps and store
the column in an appropriate place.
C H A P T E R 4
Introduction
The DECADE Elite has been designed for maximum functionality and ease
of use. The control of ECD parameters via the keyboard & LCD display
(DECADE Elite only) is such that without reading this chapter, it should be
possible to operate the detector. This chapter is intended as a reference
guide in case questions arise during operation.
The information shown in the numerous screens is presented in alphabeti-
cal order. For each item an explanation is given, together with the item’s
nature and the screen(s) of appearance. The nature of an item can be:
1. Control: parameters with a cursor box (‘’) can be attained via
cursor buttons and changed by the ‘value’ button.
2. Status: without a cursor box a parameter reflects the current
status.
3. Functions: parameters in CAPITALS are commands accessible via
function buttons F1 - F5.
4. The 'Enter' button is only used to accept changes in cell potential
and range.
In the top right corner of each screen the name of the present screen is
displayed. If available, the bottom left function button displays a previous
screen, and the bottom right one the next screen.
DC mode
Chapter 4 DECADE Elite controller 51
Pulse mode
52 175_0010 - DECADE Elite / Lite user manual - rev 13
SCAN mode
Chapter 4 DECADE Elite controller 53
CONFIG menu
54 175_0010 - DECADE Elite / Lite user manual - rev 13
DIAG menu
Chapter 4 DECADE Elite controller 55
Parameters
Explanation: Type S is status, F is function and C is control.
Dialogue Elite
Dialogue Elite for Windows is a multi-functional program to control the
DECADE Elite and DECADE Lite electrochemical detector.
There are no tools for further data analysis, therefore Dialogue cannot be
considered as fully featured chromatography software package. The free
demo version of this program has already all functionality implemented,
functionality is unlocked by a license dongle.
The free version is primarily used for controlling all operational parame-
ters of the detector. The free version cannot be used for starting a run, run-
ning a calibration or OQ. The DECADE Lite does not have a keyboard and
display, and are fully controlled by the free Dialogue software.
For details about the use and functionality of the Dialogue Elite software
please refer to document 175.0015 Dialogue Elite user manual which can
be downloaded from the Antec website (www.antecscientific.com).
62 175_0010 - DECADE Elite / Lite user manual - rev 13
Fig. 10. DECADE Elite method window in Clarity. On the main tab the
main measurement conditions can be set/controlled (measurement mode
& potential settings).
Chromeleon CDS
Full control and data acquisition of the DC and Pulse measurement mode
for the DECADE Elite is supported in Chromeleon CDS version 7.2 SR5
and higher (ThermoFisher Scientific, USA).
Fig. 11. Chromeleon console showing a ECD e-panel with method settings
and acquisition window.
For details about the requirements, use and functionality of the DECADE
Elite control driver in Chromeleon please refer to document 176.0108
DECADE Elite driver for Chromeleon CDS user manual, which can be
downloaded from the Antec website (www.antecscientific.com).
64 175_0010 - DECADE Elite / Lite user manual - rev 13
OpenLab CDS
Full control and data acquisition of the DC and Pulse measurement mode
for the DECADE Elite is supported in OpenLab CDS version 2.3 and
higher (Agilent, USA). The DECADE Elite is supported in OpenLab Chem-
station version C.01.09 and higher.
Fig. 12. Agilent Status window showing the Instrument status and acquisi-
tion window showing the acquired data.
For details about the requirements, use and functionality of the DECADE
Elite control driver in OpenLab and OpenLab Chemstation edition please
refer to document 176.0102 and 176.0102C DECADE Elite driver for
OpenLab CDS user manual, respectively. The documents can be down-
loaded from the Antec website (www.antecscientific.com).
Empower CDS
Empower™ CDS is a fully featured data acquisition software from Waters
Corporation. Full control and data acquisition of the DC and Pulse meas-
urement mode for the DECADE Elite is supported in Empower 3.0 (with
Empower ICF Support Layer v3.5) and higher.
The DECADE Elite driver adds full instrument control for Electrochemical
Detection (ECD) up to two flow cells (DCC). Supported is the DC mode
with constant potential measurement for most applications, and the
PULSE mode for special applications such as carbohydrates and amino-
glycosides. A time table is supported for time based control of several op-
erational parameters and triggers for external devices by switching relays.
Also, an external electric valve can be controlled with the DECADE Elite
case of special applications such as column switching during an analysis.
For details about the requirements, use and functionality of the DECADE
Elite control driver in Empower OpenLab please refer to document
176.0112 Waters ECD driver for Empower 3 user manual. The documents
can be downloaded from the Antec website (www.antecscientific.com).
C H A P T E R 5
Introduction
One of the characteristics of electrochemical detection is its tremendous
dynamic range. In amperometric detection peak heights may vary from mi-
cro-amperes down to the pico-ampere range. The DECADE Elite covers
such a wide range from 200 µA down to 10 pA full scale, without being lim-
ited by electronic noise. For this reason the DECADE Elite is equipped
with a 24-bit ADC and 16-bit DAC for analogue data output. One of the key
features is that data can be sampled with data collection rates up to 100
Hz (100 pts/sec) in DC mode which assures that fast peak responses typi-
cal for UHPLC can be detected with sufficient resolution.
Three-electrode configuration
The circuitry of the DECADE Elite detector is designed for operation with
electrochemical flow cells with a three-electrode configuration (Fig. 14).
The working potential is set between the working electrode (WE) and the
auxiliary electrode (AUX). The AUX is kept at a precisely defined reference
electrode (REF) potential by means of the so-called voltage clamp. This is
an electronic feedback circuit that compensates for polarization effects at
the electrodes.
Internal organization
At the working electrode (WE) in the electrochemical flow cell the electron
transfer takes place due to an oxidation or reduction reaction. The result-
ing electrical current is amplified by the current-potential (I/E) converter
(Fig. 15).
Fig. 15. DECADE Elite signal processing from electrochemical flow cell to
output.
The signal from the I/E converter can be compensated with auto zero or
offset, and is digitized using a 24-bits ADC. In the CPU the signal is pro-
cessed, for example noise filtering, or more complex data processing in
PAD. Finally after the 16-bits DAC the signal is set to a 1 V full scale ana-
log data output (by default Output = ADC). Also the true analog signal from
the I/E converter (before AD conversion) is available via the ‘Analog data
output’ connector. This output can be selected in the CONFIG menu by
setting the parameter Output = I/E.
Fig. 16. Typical configuration for serial mode detection. Cell 1 is a FlexCell,
cell 2 is a SenCell for detection. Channel 1 and 2 use the same time base
of system 1.
Fig. 17. Typical configuration for parallel mode detection. Two independent
HPLC systems with dual channel support from OR 110, DECADE Elite and
Clarity software.
70 175_0010 - DECADE Elite / Lite user manual - rev 13
Fig. 18. DECADE Elite with 2 columns and 2 SenCell flow cells for parallel
detection.
new status screen is available in dual cell systems which indicates the sta-
tus of both cells in a single screen. However, for convenience it is advisa-
ble to use PC control from the Clarity Chromatography software or Dia-
logue Elite software when working with 2 flow cells.
M A I N 1 2
E C D e t e c t o r 1 . 1 0
# C O N F I G > D C P U L S E S C A N D I A G
QCC 1 D I c = + 2 3 . 4 5 n A R E M O T E 10A
2 P I c = + 1 2 0 . 6 n A
2 5 > 3 0 ° C
P R E V
Fig. 19. DECADE Elite main menu (top) with active cell indicator in top
right corner. Multi-STAT screen showing cell 1 (DC mode) and cell 2
(PULSE mode).
Parameters
Operational parameters are controlled from the SETUP screens in the
DECADE Elite. Parameters are filter, cell potential and offset. Temperature
is set in CONFIG menu.
T e m p = 3 0 ° C C o n t r a s t = 2 0 C O N F I G
V a l v e = 3 V o u t s o u r c e = D A C
I D 1 m a s t e r = n o
P R E V I O
R a n g e = 5 0 n A M a x C o m p = 2 . 5 µ A D C 1
E c = + 0 . 5 0 V # O f f s = + 1 0 % S E T U P
F i l t = . 0 0 2 H z
P R E V P O L A R = + E V E N T S N E X T
Range
Range selection is done in the ‘SETUP’ or ‘STAT’ screen in DC, PULSE
and SCAN mode. A number of ranges can be selected; the maximum cur-
rent that can be compensated for using auto zero and offset differs. The
high sensitivity ranges (10 pA - 5 nA) have the best noise specifications. In
fact, there is a trade-off between best noise specification at sensitive
ranges, and maximum compensation at the less sensitive ranges. This is
72 175_0010 - DECADE Elite / Lite user manual - rev 13
In the PULSE and SCAN mode (available in the DECADE Elite only), cur-
rent is much higher than in DC mode. Therefore it is not possible to select
pA ranges.
Offset
A maximum offset of +50% and - 50% in 5% steps can be set. For exam-
ple, 20% is a 200 mV offset when the maximum output is 1.0 Volt.
Chapter 5 Detection and Parameters 73
Polarity
The polarity of the output can be inversed. Oxidative and reductive anal-
yses generate opposite currents. For data acquisition, traditionally chroma-
tographic peaks have a positive amplitude. Therefore selection of polarity
is useful.
Filter
High frequency noise is efficiently removed and chromatographic peaks
can be detected with better signal to noise ratio.
Fig. 21. Signal to noise ratio is improved using a filter (A vs. B).
The DECADE Elite is equipped with ADF (Advanced Digital Filter) as a tool
to filter the acquired signal and improve the sensitivity of the analysis (Sig-
nal-to-noise ratio). In the next chapter the filter setting is explained includ-
ing detailed background information about filtering.
DC mode
In the tables below the available filter settings for the DC mode are listed
with the corresponding data rate of the output. Data rate is expressed as
number of data points per second (Hz). In the DC mode the data rate is
not an adjustable parameter but is coupled to the filter setting, except for
RAW. RAW is special, the incoming data are not filtered and a data rate
between 1 and 100 Hz can be selected.
74 175_0010 - DECADE Elite / Lite user manual - rev 13
Filter OFF is also a special case. The data rate is fixed to 10 Hz, and the
data is not filtered. Setting OFF is therefore the same as RAW at 10 Hz.
Pulse mode
In the pulse mode the working electrode is dynamically and continuously
regenerated by a series of potential steps in a cyclic manner. Data is pro-
cessed differently, and the data rate is defined by the total duration of the 5
potential steps in a pulse: t1 + t2 + t3 + t4 + t5. Usually the typical pulse
duration is between 0.5 and 2s (data rate between 2 – 0.5 Hz). Filter set-
tings in the pulse mode are therefore selectable between 0.5 and 0.001
Hz, and OFF.
Autozero
In the STAT screen the Autozero function (F4) is available. For the DC
mode this screen is shown below.
V o u t = + 0 . 5 0 0 V I c = + 2 . 5 0 0 n A D C 1 12
R a n g e = 5 n A E c = + 0 . 8 0 V S T A T
F i l t = o f f C o m p = O F F 3 5 → 3 5 ° C 2 1 1 . 2 0
P R E V C E L L = O N M A R K A Z E R O N E X T
V o u t = + 0 . 0 0 0 V I c = + 2 . 5 0 0 n A D C 1 12
R a n g e = 5 n A E c = + 0 . 8 0 V S T A T
F i l t = o f f C o m p = O N 3 5 → 3 5 ° C 2 1 1 . 2 0
P R E V C E L L = O N M A R K A Z E R O N E X T
Fig. 22. Stat screen in DC mode. Top: compensation off. Bottom: after ex-
ecuting an Autozero (Function button F4).
The Autozero function can be used to set the output voltage of the I/E con-
verter (signal) to 0 Volt. This is done by means of an compensation ampli-
fier in the analog measurement circuitry, see Fig. 15. This compensation
circuitry enables the possibility to measure small current signals (analyte
peaks) in a sensitive range even if the background cell current is high. For
example with a background current of 20 nA one can still do measure-
ments in the 1 nA range, because the maximum compensation for that
range is 25 nA.
Applying an Autozero will set the baseline (close) to 0 nA level. The full
scale is available for data acquisition. The Icell does not change, it is the
uncompensated current.
Remedy: In the event it does happen, the issue can be simply solved by
setting the Compensation to OFF (Comp = OFF) in the STAT screen or
76 175_0010 - DECADE Elite / Lite user manual - rev 13
CDS control software. Subsequently, wait till the background current is sta-
ble/stabilized and execute an Autozero again. The detector should now
display the correct current again.
In case event tables are programmed in LC methods in the CDS
software with an Autozero action, make sure that the autozero is pro-
grammed in a part of the chromatogram were the signal is stable. So at the
beginning of the run, during the hold-up time (so before the unretained/sol-
vent peak elutes). Or at the end of the run.
Chapter 6 Measurement modes 77
C H A P T E R 6
Measurement modes
DC mode
In Direct Current (DC) mode a static potential is constantly applied to the
EC flow cell to establish an electrochemical oxidation or reduction reaction
of an suitable analyte. In DC mode the resulting current signal is continu-
ously measured and sent to the detector output.
Fig. 23. Top: Plot of cell potential versus time. Bottom: DC mode method
settings window in Dialogue Elite. The potential (top right) can be set be-
tween -2.5 and +2.5 Volt.
78 175_0010 - DECADE Elite / Lite user manual - rev 13
The DC mode can be used for detection using relatively inert working elec-
trode (WE) materials such as Glassy Carbon (GC) or Boron-Doped Dia-
mond. Suitable analytes in this case are typically aromatic compounds with
functional groups which can be easily oxidized at relatively low potentials.
In addition, one other requirement for successful operation in DC mode is
that the target compounds and reaction products should not easily absorb
or contaminating the electrode surface leading to inactivation of WE elec-
trode activity and subsequent loss of response.
Pulse mode
Besides the DC mode the DECADE Elite can also be operated in Pulse
mode. The pulse mode is quite different from the DC mode. Instead of a
constant potential, a series of potential steps is applied in a cyclic manner.
The signal is sampled during a fraction of the total pulse cycle. During the
sampling time (ts) the signal is collected and this value is sent to the detec-
tor output. The frequency of data output is determined by the pulse dura-
tion: t1 + t2 + t3 + t4 + t5. Typically the duration is between 0.5 and 2s
(data rate between 2 – 0.5 Hz).
This mode is particularly useful for certain applications where the working
electrode is rapidly fouled due to adsorption of reactant or reaction prod-
ucts or/and when using metal electrodes like gold, platinum, titanium elec-
trodes etc. Metal electrodes are the electrodes of choice for oxidation/re-
duction of aliphatic compounds with functional groups. Such compounds
are hard to oxidize on inert electrode materials. Due to surface adsorption
of the target compound on metal electrodes the activation energy barrier
Chapter 6 Measurement modes 79
Fig. 25. Pulse mode method settings window in Dialogue Elite. The poten-
tial waveform settings are shown on the right side.
Fig. 26. Potential steps in pulse mode. During t1 the measurement poten-
tial is applied, the actual current measurement occurs during ts. Steps t2,
t3 and t4 are for regenerating the electrode. This process repeats itself
continuously as soon as the cell is on.
80 175_0010 - DECADE Elite / Lite user manual - rev 13
In pulse mode up to five potential steps can be defined. The sampling in-
terval (period at which the current is measured) is applied at the end of E1.
More detailed information about pulsed amperometric detection is availa-
ble in chapter 8 of this user manual.
MPAD mode
In the DECADE Elite (FW 1.09 and up) an Multistep Pulsed Amperometric
Detection (MPAD) mode is available also known as pulse mode 2. In
MPAD mode it is possible to program a multi-step waveform, with up to 30
time, potential (t,E) coordinates with a maximum pulse duration of 4 sec-
onds, see figure below.
Fig. 27. The new pulse mode 2 with freely programmable t,E table. In red
the sampling of data for acquisition.
Chapter 6 Measurement modes 81
Scan mode
For method development the scan mode is available. In the scan mode the
cell potential is swept between two pre-set values (E1 and E2) using a cer-
tain scan speed (in mV/s) and the current is measured during the sweep.
This method can be used to study the REDOX behavior and to determine
the optimum detection potential of a pure analyte dissolved in mobile
phase.
Fig. 28. Scan mode example: plot of cell potential versus time in scan
mode (two full scans).
82 175_0010 - DECADE Elite / Lite user manual - rev 13
Fig. 29. Scan mode method settings window in Dialogue Elite. Continuous
scan (10 scans) between 0.20 and 1.20V with a scan rate of 50 mV/s.
The scan mode can be used in method development but is not a measure-
ment mode used in HPLC-ECD analysis itself. Therefore it is only available
in Dialogue Elite and in stand-alone operation. In Clarity, Chromeleon,
OpenLab and OpenLab Station Ed. CDS the scan mode is not available.
See chapter 9 describing the details how to optimize the working potential
for detection using scanning voltammetry.
Chapter 7 Measurement modes 83
C H A P T E R 7
Introduction
Besides for its tremendous linear dynamic range and selectivity, electro-
chemical detection is well-known for its very low limits of detection. To fur-
ther improve these detection limits the Antec engineers have developed
ADF (Advanced Digital Filter) and the DECADE Elite has been equipped
with it as a standard. The improvement factor in signal-to-noise (S/N) ratio
depends on the frequency relation of signal and baseline noise. S/N im-
provements from a factor 5 up to more than 100 have been obtained.
To understand how a digital filter works, first the importance of frequencies
in chromatographic analysis will be explained. Then we will look at peak
width, filter settings, cut off frequency, amplitude response plots and finally
at a few chromatograms before and after applying ADF.
Frequency
A scientific definition of frequency is “the number of completed alterations
per unit time”. It has two dimensions: count and time. Frequency is usually
expressed in Hz, which is counts per second.
The counts themselves can run in a regular, evenly spaced manner, as
with sine waves whose curve shapes do not change. Alternatively, the
counts can run in an irregular manner within the specified unit of time. If
the latter happens, frequencies would vary if broken down into t smaller
units of time.
In the example of Fig. 30 a signal is shown with a frequency of 12 altera-
tions in 5 minutes. To express its frequency in a more scientific way a full
period is precisely determined and expressed in Hertz (or s-1). It is a sine
wave with a frequency of 0.04 Hz (Fig. 31).
84 175_0010 - DECADE Elite / Lite user manual - rev 13
400000
300000
200000
100000
0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
-100000
-200000
-300000
-400000
Fig. 30. Example of a signal with regular evenly spaced alterations: a sine.
400000
300000
200000
100000
0
4.0 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0
-100000
t (min)
-200000
-400000
Fig. 31. Sine of Fig. 30. A full period is 0.41 min (25 s) which corresponds
to a frequency of 1/25 = 0.04 Hz.
540000
3.8 4.05
440000
340000
240000
140000
40000
3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0
Fig. 32. Frequency tells how often something happens: 1 peak in about
0.25 min (15 s), f = 1/15 = 0.07 Hz.
Chapter 7 Measurement modes 85
540000
440000
340000
240000
140000
40000
3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0
698000 ADF
598000 10 s 40 s
0.10 Hz 0.025 Hz
498000
398000
h
298000
198000
98000
-2000
1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0
min
20
sine f = 10 Hz
15
10 sine f = 0.4 Hz
I (pA)
-5
30 31 32 33 34 35 36 37 38 39 40
t (s)
Fig. 35. Typical random noise in chromatography (lower trace). Both fre-
quencies (0.4 and 10 Hz) can be recognized amongst others.
Looking closely to the lower noise trace both frequencies (and others) can
be recognized. This is typical for noise in chromatography: a collection of
more or less random frequencies.
form a running average filter takes the average of n data points to create a
new data point. For example in a 5-points running average filter output
data point y[80] is calculated from measured data points x[80] – x[84] as:
Each input data point has the same weighting factor of 1/5. In more ad-
vanced digital signal processing a more complicated equation is used to
calculate the output data point y[n]:
1.2
pass
1
Amplitude
0.8
0.6
0.4
0.2
filte r block
0
0 1 2 3 4 5
Freq uency (Hz)
Fig. 36. Amplitude response plot of an ideal low pass filter with a cut-off
frequency of 1 Hz.
There are several ways to describe the filter characteristics. An amplitude
response plot gives important information on filter behavior. Suppose our
signal of interest has a frequency between 0 - 1 Hz, and all higher frequen-
cies are noise. An ideal filter is shown in Fig. 36 where signal frequencies
88 175_0010 - DECADE Elite / Lite user manual - rev 13
In practice filters behave a bit different from the ideal situation. Amplitude
response plot shows a more gradual attenuation profile at higher fre-
quency. This cut off frequency is where the output signal amplitude is 70%
of the input signal, also known as 3 dB point.
1.2
1
Amplitude
0.8
0.6
0.4
B A
0.2
C
0
0 1 2 3 4 5
Freq uency (Hz)
Fig. 37. An amplitude response plot of a low pass filter with a cut-off fre-
quency of 1 Hz. It is a 2 (A), 4 (B) and 8 (C) pole Bessel filter.
In Fig. 37 it is shown that the number of poles is important, a filter behaves
more ideal with increasing number of poles. In a hardware filter the num-
ber of poles is the number of filter circuits that are placed in series.
A digital filter does not have poles, but it is characterized by the number of
input data points used to calculate a new output data point. For example a
9-point digital filter (Savitzky-Golay) is given as:
Note that the sum of coefficients is exactly 1. Y[n] is the output data point,
X[n] are input data points. Generally spoken, the performance of a digital
filter improves with more input data points, but also more processor capac-
ity is required for the large number of calculations.
90 175_0010 - DECADE Elite / Lite user manual - rev 13
400000
300000
200000
100000
0 0
0 5 10 15 5 6 7 8 9 10
400000
300000
200000
100000
0 0
0 5 10 15 5 6 7 8 9 10
400000
300000
200000
100000
0 0
0 5 10 15 5 6 7 8 9 10
Fig. 39. From top to bottom filter setting of 0.5, 0.02 and 0.002 Hz. Narrow
peaks in front of the chromatogram are deformed at 0.005 Hz, whereas
wider peaks show hardly any deformation (see peak at t~ 13 min). Attenu-
ation of noise is shown in the close up on the right.
Prerequisite for a ‘good’ noise filter for data acquisition in liquid chromatog-
raphy is that it improves the S/N ratio without significant distortion of the
Chapter 7 Measurement modes 91
C H A P T E R 8
Introduction
The DECADE Elite can operate in PAD mode (Not available in the
DECADE Lite). In pulsed amperometric detection (PAD) the working elec-
trode (WE) is regenerated at a frequency of 0.5 - 3 Hz by the application of
a series of potential changes. This is particularly useful for certain applica-
tions where the working electrode is rapidly fouled due to adsorption of in-
soluble reaction products. A well-known application area of PAD is the
analysis of carbohydrates (Fig. 41) [1].
Pulse settings
In PAD of carbohydrates a series of potentials is applied in a continuous
cyclic manner. During time interval t1 the detection potential is applied.
The data collection occurs within t1, during time interval ts (sampling time).
The time difference t1 - ts is the stabilization time.
During the next time intervals (t2..t4) the electrode is ‘cleaned’ by reductive
and oxidative potential steps.
Output frequency
An important difference between the DC and the pulse mode is the fre-
quency of the output signal. In the DC mode the signal has a 1-100 Hz fre-
quency, in the pulse mode the frequency is determined by the duration of
the pulse. Once every cycle, the ts signal is sent to the output.
If a stepwise pattern in the chromatogram is seen when zooming in, it
means that the data acquisition has an unnecessarily high sampling fre-
quency. This leads to large data files, but certainly not to a better chroma-
togram. Usually, data acquisition at 1 Hz is sufficient.
References
1. D.C. Johnson, D. Dobberpuhl, R. Roberts and P. Vandenberg, Review.
Pulsed amperometric detection of carbohydrates, amines and sulphur spe-
cies in ion chromatography - the current state of research, J. Chromatogr.
640 (1993) 79-96
2. D.C. Johnson, W.R. LaCourse, LC with pulsed ECD at gold and plati-
num electrodes, Anal. Chem., 62 (1990) 589A-597A
3. W.R. LaCourse, D.C. Johnson, Optimization of waveforms for pulsed
amperometric detection of carbohydrates following separation by LC, Car-
bohydrate Research, 215 (1991) 159-178
4. W.R. LaCourse, D.C. Johnson, Optimization of waveforms for pulsed
amperometric detection of carbohydrates based on pulsed voltammetry,
Anal. Chem. 65 (1993) 50-55
Chapter 9 Measurement modes 99
C H A P T E R 9
Introduction
A current - voltage (I/E) relationship, or voltammogram, characterizes an
analyte. It gives information on the optimum working potential, which can
be used to improve detection sensitivity and selectivity.
There are several ways to obtain a voltammogram:
Electrochemical reactions
In electrochemical detection (ECD) a reaction of the analyte at an elec-
trode surface is monitored. This distinguishes ECD from most other detec-
tion techniques where detection is based on the physical properties of an
analyte (i.e. mass spectrometry: molecular mass, absorbance detection:
molar absorptivity). For electrochemically active compounds, the potential
between reference electrode (REF) and working electrode (WE) deter-
mines the reactivity of the analyte at the WE. The potential difference sup-
100 175_0010 - DECADE Elite / Lite user manual - rev 13
plies the energy level needed to initiate or enhance the electrochemical re-
action. Different analytes may have different oxidation or reduction poten-
tials, which determines the selectivity of ECD.
10 A
Icell (nA)
0
0.0 0.5 E1 1.0 1.5
Ecell (V)
Fig. 45. Hydrodynamic voltammogram of norepinephrine (A) at a glassy
carbon working electrode, and the current of the baseline (B). At E1 the
electrochemical signal becomes diffusion limited.
102 175_0010 - DECADE Elite / Lite user manual - rev 13
Scanning voltammogram
An alternative for the chromatographic construction of an I/E relationship is
the application of scanning voltammetry (DECADE Elite only). In a scan-
ning voltammetry experiment the working electrode potential is ramped up
and down between two preset potentials (E1 and E2) and the current is
measured while the analyte is continuously flushed through the flow cell.
This is repeated as many times as desired. The rate of voltage change
over time is defined as the scan rate (mV/s).
E2
)
Working Potential (V)
(V/s
nrate
Sca
E1
t0 t1 t2
Time (Seconds)
As can be seen in both Fig. 45 and Fig. 47, when the working potential is
increased the electrochemical reaction is enhanced hence the signal in-
creases. At a certain potential the I/E curve flattens. All analyte molecules
that reach the working electrode are converted at such a high rate that the
analyte supply becomes the limiting factor. At the working electrode sur-
face a stagnant double layer exists, where molecular transport takes place
by diffusion only. Therefore, the current at (and beyond) this potential is
called the diffusion limited current.
A
10
B
Icell (nA)
0
0.0 0.4 0.8 1.2
Ecell (V)
Fig. 47. Scanning voltammetry of 1.0 µmol/L norepinephrine (A) at a
glassy carbon working electrode, at a scan speed of 10 mV/s. Scan (B) is
the blank solvent.
Fig. 49. At a higher pH the I/E curve of norepinephrine is shifted to the left.
R a n g e = 2 µ A S p d = 2 0 m V / s S E T U P
O f f s = + 0 % C y c = h a l f
P R E V C E L L = O F F N E X T
Fig. 51. Programming the scan mode in the 'SCAN SETUP' screen.
Fig. 52. Programming a scan in the Elite Dialogue software, half sweep.
In the above example a ‘half’ scan is programmed, sweeping the potential
from 0 V to 1.0 V. A full scan would include the reverse scan, i.e. from 0 V
to 1.0 V and back to 0 V. In the continuous mode the voltage is swept up
and down between both potentials for a pre-defined run time.
108 175_0010 - DECADE Elite / Lite user manual - rev 13
Fig. 53. DECADE Elite scanning voltammetry set-up with a syringe pump.
The Antec dual syringe infusion pumps (pn 188.0035, 36 and 37) which
can be obtained as accessory has as an advantage that it can be con-
trolled in Dialogue Elite software as well.
When scanning with the Elite Dialogue software (‘continuous’) all scans
are displayed and can be selected individually.
C H A P T E R 1 0
General
DC mode
C H A P T E R 1 1
In this chapter all rear panel functionality is described. The DECADE Elite
has besides the mains inlet in total 5 connectors on the rear panel for com-
munication, data output and I/O. A photo of the rear panel connectors is
shown below for reference.
USB B connector
USB type B connector for serial instrument control over USB, for service
use only:
LAN connector
RJ-45 bus for serial instrument control over LAN:
Please consult the installation section for details about configuration and
set-up of communication over LAN.
VALVE connector
Serial D9 (9-pins subD male) valve control connector for electrical (Vici)
and manual valves:
T e m p = 3 5 ° C C o n t r a s t = 2 0 C O N F I G
V a l v e = P r e s e n t V o u t s o u r c e = D A C
I D 1 m a s t e r = n o
P R E V I O
118 175_0010 - DECADE Elite / Lite user manual - rev 13
Not ‘hot swappable’: Always restart the detector (by re-powering via the
mains switch on the rear panel) when connecting the valve for correct
initialization of an electrically-actuated Vici valve.
In the case an electrical Vici valve is detected, the Dialogue Elite software
automatically shows the presence of the electrical valve in the device set-
tings window (EC device) and valve control buttons will appear in the de-
tector window.
Dialogue Elite: Go to the device settings and the EC device tab and check
the manual valve check box.
Fig. 60. Dialogue Elite EC device settings, manual valve check box.
In stand-alone mode the valve position can be controlled via the F5 func-
tion button in the STAT screen. In the case of a manual valve the infor-
mation above the F5 button (INJ=L) in the STAT screen shows the status
(position) of the valve only, no valve control is possible with this type of
valve.
Make sure that the valve flow path is connected correctly and LOAD
corresponds with position A and INJECT with position B. See example
in the figure below for a 6-port valve.
In Dialogue Elite the valve status/position is shown under the monitor tab.
To make it visible in the left table, check the valve position checkbox in the
item selector (by default the valve position is not show).
CHAPTER 11 Rear panel I/O 121
Fig. 62. Dialogue Elite Monitor tab with valve status (position) displayed.
Fig. 63. Dialogue Elite Events tab with an example of two programmed
events switching the valve to Inject (t = 0.01 min) and back to Load (t =
0.05 min).
T e m p = 3 5 ° C C o n t r a s t = 2 0 C O N F I G
V a l v e = P r e s e n t V o u t s o u r c e = D A C
I D 1 m a s t e r = n o
P R E V I O
Fig. 64. Top: CONFIG screen. Bottom: DECADE Elite signal processing
from electrochemical flow cell to output. R is a selectable I/E resistor of
100M, 10M, 1M, 100K, 10K or 1K in the I/E converter circuit.
The output parameter Vout source can also be selected/set using Dialogue
Elite. Go to the device settings menu and Acquisition tab and set the ana-
logue output to DAC or I/E converter. The instrument is by default set to
DAC. For regular measurements it is advised to use the DAC output.
DAC output
The DAC output is the processed signal by the DECADE Elite’s CPU and
is identical to that of the digital cell current signal obtain via data-acquisi-
tion over LAN using the Elite Dialogue or the Clarity Chromatography soft-
ware.
The signal in mV from this output is directly related to the range setting of
the DECADE Elite. So for example in the case the 200 µA measurement
range is selected ‘Range = 200 µA’ the DAC signal on the analog data
connector corresponds with +1000 mV = +200 µA and -1000 mV = -200
µA (so +/- 1V full scale). To convert the signal in mV to the actual cell cur-
rent in nA use the following calculation:
So for example a signal on the output of 250 mV in the 200 nA range cor-
responds with an actual cell current of (250/1000)*200 nA = 50 nA.
I/E output
The I/E output is the unprocessed analog signal from the I/E converter cir-
cuit. This signal is a true analog signal which is as close as practically pos-
sible to the working electrode (WE). The maximum output voltage of the
I/E output is +/- 2.5V under all conditions and is independent of the detec-
tor range setting. The signal in mV is related to the selectable I/E resistor
of 100M, 10M, 1M, 100K or 1K in the I/E converter circuit. The actual cell
current can be calculated from the analog I/E output in Volt using the fol-
lowing formula:
Table VI. DECADE Elite I/E converter resistors (valid for FW1.09 or higher)
Current Ranges I/E Resistor (Ohm)
10pA, 20pA, 50pA, 100pA, 200pA, 500pA and 1nA 100M
2nA, 5nA 10M
10nA, 20nA, 50nA 1M
100nA, 200nA, 500nA, 1µA, 2µA 100K
5µA, 10µA, 20µA 10K
50µA, 100µA, 200µA 1k
For reference, the layout of the analog data out connector is shown in the
table below.
To measure for example the analog signal of cell 1 with an external A/D
converter:
• Connect the signal wire, lead of pin 1 (Vout 1), to the analog
measurement channel of the A/D converter.
• Connect the GND wire, lead of pin 6 (or 7-9V), to the
corresponding analog ground connection of the A/D converter.
Relays
The DECADE Elite has 2 free programmable contact closure outputs:
The maximum rating for these contact closure outputs are 24 VDC (switch-
ing voltage) and 0.25 A. The relays can be controlled in the Elite Dialogue
software and Clarity Chromatography software.
AUX
The DECADE Elite has 4 free programmable TTL outputs AUX1 – AUX4
(pin 7 – 10). These contacts are default ‘high’ 3.3V (inactive), when active
the status is ‘low’ 0V.
Overload
The overload output (pin 11) can be used to monitor if the cell current goes
out of range during a chromatographic run. An ‘Out of range’ error appears
when the cell current Icell exceeds the limit of the current range at which the
measurement is performed. See figure on the next page.
CHAPTER 11 Rear panel I/O 127
Fig. 66. Example of a chromatogram were the cell current exceeds the
maximum current level and the signal is ‘out of range’.
If for instance the cell current goes in out of range during the recording of
an analyte peak, it can be (in most cases) easily recognized by a flat top of
the peak and a very abrupt transition to the flat top at the edges.
By default the status of the overload output is ‘high’ 3.3V. When the cell
current has the status ‘out of range’ the overload output will change status
to ‘low’ 0 Volt, until the cell current returns to a value within the measure-
ment range again. The overload output (pin 11) is one of the configurable
I/O’s.
P 1 1 ( O V L D ) = 1 P 1 8 ( A Z E R O ) = 1 I O
P 1 2 ( C - O N ) = A L L P 2 1 ( S T A R T ) = A L L
P 1 5 ( C - O F F ) = 2
P R E V
The following options can be selected for the configuration of pin 11 (in
case of a DECADE Elite TCC) :
*) When this option is selected, the overload output will be active for all
cells present in the DECADE Elite. If the cell current of either one of those
cells will go ‘Out of range’ the overload input pin 11 will become active.
Autozero
The DECADE Elite has 3 TTL inputs (pin 18 – 20) available to autozero
the cell current of the cell(s). Triggering these inputs enables external acti-
vation of the auto zero command. This function is active only when the ‘I-
cell’ is displayed. One autozero input is configurable (pin 18) in the IO
menu (see previous chapter about the overload output). The configuration
settings of this input are: 1, 2, 3, 4, 5,’ ‘, and all. In case ‘all’ is selected the
cell current of all cells present in the DECADE Elite will be zeroed when
the input is triggered.
Start
The DECADE Elite has 4 TTL inputs (pin 21 – 24) available to start data-
acquisition and/or start a scan. One start input is configurable (pin 21) in
the IO menu and can be used for example to start the data-acquisition of
all cells synchronously using only one trigger input when the setting ‘All’ is
selected.
CHAPTER 11 Rear panel I/O 129
The manufacturer will not accept any liability for damage, direct or
indirect, caused by connecting this instrument to devices that do not
meet the relevant safety standards.
Programming outputs
In the Dialogue Elite software the DECADE Elite/Lite outputs (Relays and
AUX) can be controlled and programmed. This can be done via the Detec-
tor tab on the main window and via timed event tables. By default the I/O is
hidden and not shown in the monitor tab. To be able to see the status and
control buttons for the I/O go to the Device settings window (settings tab)
and check the ‘Show status outputs’ checkbox under the controls section.
Fig. 67. Dialogue Elite Detector tab with I/O status and control buttons.
The status of the Relays and TTL inputs can be set to active () or inac-
tive (). The selected state of the outputs will be activated after clicking
the ‘Send to device button’.
The outputs can also be switched via timed events under the events tab.
By opening a ‘New event’ under the Events pull-down menu, a new Event
table/file can be created. Subsequently, the Event editor will open. Select
130 175_0010 - DECADE Elite / Lite user manual - rev 13
the event ‘Output A’ and assign a time at which the output should be
switched and select which output should be activated (Relay, Aux) under
the value field. See example below in the figure below. Make sure to pro-
gram a second step to inactivate the Output again and put it back in its ini-
tial inactive state again.
Fig. 69. Dialogue Elite Event table window with an example of program-
ming timed events for output A (in this case Relay 1).
CHAPTER 11 Rear panel I/O 131
In the detector accessory kit (pn 175.0200) a dedicated I/O cable is sup-
plied with an I/O connector board with screw terminal connections for easy
connection of open-ended electrical wiring for I/O:
For use, connect the I/O cable to the 25-pins I/O connector on the detector
rear panel, subsequently connect the I/O connector board to the end of the
I/O cable.
On the rear side of the I/O connector board a label is adhered for
reference with the lay-out of the digital I/O connector.
CHAPTER 11 Rear panel I/O 133
Fig. 70. Top: DECADE Elite with I/O screw terminal board installed. Bot-
tom-left: back side of screw terminal board.
134 175_0010 - DECADE Elite / Lite user manual - rev 13
Fig. 71. Left: DECADE Elite rear panel grounding stud. Right: example of
shielding the flow cell by grounding the solvent outlet tubing.
Use the chassis grounding stud for shielding only, not for safety
grounding .
CHAPTER 12 Troubleshooting 135
C H A P T E R 1 2
Troubleshooting
Even though great care was taken in the design of the DECADE Elite,
problems may occur during operation of the instrument. The information in
this chapter may help you to identify and solve the source of the problems.
Errors can be categorized in two types:
• Instrument errors
• Analytical problems
In the next sections both types of errors are described. In the event that
the problems cannot be solved after following the instructions in this trou-
bleshooting section, contact your local supplier for further assistance.
Instrument errors
Incidental fault conditions may occur in any instrument. The DECADE Elite
will generate an error message containing an error number with a short de-
scription for several hardware fault conditions, which will be shown in the
instruments LCD display.
Error 23: In case the actual temperature exceeds 70°C switch off the
potentiostat immediately do not touch the metal parts inside the oven
compartment because they could be hot, and contact the manufacturer
or its representatives for service. Do not use/switch on the instrument
before the instrument is serviced at the factory.
136 175_0010 - DECADE Elite / Lite user manual - rev 13
No detector response
Possible cause Remedy
No power Check line voltage setting, plug in power cord
Power switch off Turn this switch ON (at the rear panel)
Faulty fuse Replace fuse
Divergent mains voltage Check line voltage
Message Advice
01 Out of range* Output is either above +1.0V or below –1.0V.
Pressing AZERO may give an adequate read-
out again. If the message remains after
pressing AZERO, the autozero function is
unable to compensate the background cell
current. Advice: use a less sensitive range in
the SETUP menu.
02 PAD overload Charging current in pulse mode out of range.
Pressing AZERO may give an adequate read-
out again. If not, it is advisable to change the
pulse settings (increase t1) or use a less
sensitive range.
*) An ‘Out of range’ error appears when the cell current Icell exceeds the
limit of the current range at which the measurement is performed. See fig-
ure on the next page.
CHAPTER 12 Troubleshooting 137
Fig. 72. Example of a chromatogram were the cell current exceeds the
maximum current level and the signal is ‘out of range’.
Remedy: In the event it does happen, the issue can be simply solved by
setting the Compensation to OFF (Comp = OFF) in the STAT screen or
CDS control software. Subsequently, wait till the background current is sta-
ble/stabilized and execute an Autozero again.
In the case event tables are programmed in LC methods in the CDS soft-
ware with an Autozero action, make sure that the autozero is programmed
in a part of the chromatogram were the signal is stable. So at the begin-
ning of the run, during the hold-up time (so before the unretained/solvent
peak elutes). Or at the end of the run.
138 175_0010 - DECADE Elite / Lite user manual - rev 13
Analytical troubleshooting
Analytical problems like loss of signal, increase in noise level, high cell cur-
rent, loss in sensitivity etc. may occur in any (U)HPLC-ECD system. It may
be hard to find the cause and several checks need to be performed to
eliminate the source of the problem. The first step is to determine whether
the problem is caused by the DECADE Elite electrochemical detector or
the rest of the (U)HPLC system. For that purpose two basic checks should
be performed by the end-user:
Dummy cell test: The outcome of the dummy cell test, which is described
in the next paragraph will give an indication if the problems are caused by
the detector hardware (electronics).
Stop flow test: The stop flow test will eliminate if the problems are caused
by the electrochemical flow cell, or is originating from the rest of the
(U)HPLC system (pump, autosampler, pulse damper, column, mobile
phase etc.)
CHAPTER 12 Troubleshooting 139
Fig. 73. Left: cell cable with external dummy flow cell (pn 250.0040). Right:
detector with external dummy flow cell.
140 175_0010 - DECADE Elite / Lite user manual - rev 13
The dummy flow cell consists of a resistor of 300 MOhm and a capacitor of
0.47 µF in parallel. The current is measured over the resistor according to
Ohm’s law (V = I * R), hence with a working potential of 0.80 V the current
will be about 2.67 nA.
Small differences are due to the tolerance of the resistor. The capacitor
functions as a ‘noise generator’ and resembles the capacitance of a flow
cell in an (U)HPLC set-up. The noise should be less than 2 pA, provided
that the dummy cell is within the fully closed Faraday shielded instrument
at the same position as the flow cell. The cell and cable must be positioned
in such way that it is not in the air flow of an oven fan.
Parameter Setting
Mode DC
Cell potential 0.80 V
Oven 35 °C, door closed, stable*
Filter Off
Range 1 nA/V
Data rate noise measurement ≤ 10 Hz
Cell and cable position Not in full air flow of an oven fan
* It is allowed to run a dummy cell test at a lower temperature (incl oven off) or with shorter
stabilization time if the instrument appears stabilized. However, in case ‘failed’, repeat the test
under above conditions to rule out short stabilization or temperature issue.
Noise measurement
A baseline trace is obtained for 15 minutes to measure the noise. Evalua-
tion of the noise level is done by averaging 30 peak to peak values from
segments of 30 s (total of 15 min) using appropriate software.
n1 + n2 + n3 ................. + n30
noise =
30
However, in case ‘failed’ or near 2 pA value, repeat the test with 30 seg-
ments to rule out an error in the estimated value.
Test criteria
The results (cell current and noise) of the dummy cell test should be
within the above mentioned test criteria. If the current value Icell and
the noise are not within the criteria it is an indication that something
could be wrong with the detector hardware. Please consult your local
representative.
CHAPTER 12 Troubleshooting 143
The Vout of 0.538V reflects the analog output signal of the ECD. In case
an AD converter is used for data acquisition this value should correspond
with what is acquired in the chromatogram. If applied, any conversion or
multiplier factor in the CDS should be considered as well.
The range setting of the internal dummy cell test is 5 nA/V, to facilitate this
Output test. At 1 nA/V the signal would run of scale completely. The visu-
ally checked noise spec at 5 nA/V range is 4 pA.
No display
For a DECADE Lite (ECD without keyboard and display) the internal
dummy cell test can be performed with the Dialogue Elite software. See
figure below. First, start “Options/Dummy cell noise test”. When the script
has started, in ‘monitor’ tab check () the Internal dummy cell checkbox.
The cell on/off will automatically uncheck, which is correct because the ex-
ternal cell circuit must be off.
Evaluate the cell current in the monitor window. The internal dummy cell
current and noise level should fall within the criteria specified on the previ-
ous page. Make sure to uncheck the Internal dummy cell option again from
the device settings menu after the test.
144 175_0010 - DECADE Elite / Lite user manual - rev 13
Fig. 75. After starting the “Dummy cell noise test”, the internal dummy cell
must be checked. Note that the (external) cell is off.
Results:
Please bear in mind that analytical problems may also be caused by exter-
nal Influences like temperature, unstable samples. Make sure that the ap-
plication was running trouble-free before and that no changes have been
made to the system. A number of causes and possible solutions for analyt-
ical problems is listed below. Contact your local representative if you need
further help.
No detector response
Possible cause Remedy
No power Check line voltage setting, plug in power cord
Power switch off Turn this switch ON (at the rear panel)
Faulty fuse Replace fuse
Divergent mains voltage Check line voltage
Cell disconnected, or Check connection
switched off
Output disconnected Check connection
Fouled WE Clean WE
CHAPTER 12 Troubleshooting 147
Noisy baseline
Saturation of output
Solved issues
In FW version 1.11.16 the known issues listed below are solved. For us-
ers with a DECADE Elite running on FW version 1.09.12 we advise to up-
date the FW to version 1.11.16 or us the work arounds advised in the table
below. In the second column (instr.) the type of instrument to which the is-
sue applies is indicated. SCC = Single Cell control (standard product),
DCC = Dual Cell control (option).
C H A P T E R 1 3
Detector accessories
For these and other DECADE Elite parts or flow cells contact your local
supplier.
CHAPTER 13 Detector accessories 151
Fig. 76. Left: pn 250.128A DECADE Elite SCC output cable D9 - Shi-
madzu ADC for connection to for example (right) an Shimadzu CBM-20A
equipped with a PC-55N AD board.
For Agilent LC systems also some dedicated trigger and autozero cables
are available which can be connected to free remote ports on any of the
Agilent LC modules
The parts 250.0129B and 250.0032G are for Agilent Infinity II LC systems
with 15-pins ERI remote interface. Part 250.0129A and 32H for Infinity I or
older LC systems with 9-pins AGP remote interface. For more information
contact your local Antec Scientific distributor.
CHAPTER 13 Detector accessories 153
This part consist of an metal union with two 10-32 fingertights connectors.
The union has an electrical ground wire which can be connected to the
grounding stud on the rear panel of a DECADE II / SDC and DECADE
Elite / Lite. The grounding stud is connected to instrument ground and in
such a manner the mobile phase passing through the union is grounded.
Cut the waste line (outlet tubing from the cell going to the waste bottle)
with a PEEK tubing cutter and fix the union in between as depicted below.
154 175_0010 - DECADE Elite / Lite user manual - rev 13
This part consist of an grounding lead with an alligator clip on one side and
a faston connector on the other end. This grounding lead is used inside the
detector oven compartment. The alligator clip can be clipped onto a col-
umn with a metal housing (stainless steel) or a metal T-splitter. See pic-
tures below. T-splitters are used in HPLC-ECD applications based on post-
column addition of sodium hydroxide (the splitter is than a part of the post-
column mixing assembly). The faston connector can be connected to the
grounding point in the oven compartment (faston tab) located at the top
side (centered) of the oven compartment.
CHAPTER 13 Detector accessories 155
156 175_0010 - DECADE Elite / Lite user manual - rev 13
Index
Index 157
ADC, 66
ADF, 83
Amplitude response, 87
ANALOG DATA, 122
Applications: quality control, 13
auxiliary electrode, 66
Biological Hazard, 13
Buffer pH, 39
buttons DECADE Elite keyboard, 49
Carbohydrate analysis, 93
cleaning of detector, 46
column, 42
compensation, 71
cut off frequency, 88
DAC output, 124
degassing mobile phase, 42
digital filter, 86
digital I/O connector, 125
DSP, 87
dual flow cell control, 68
dual flow cell navigation, 70
dummy cell
external, 139
internal, 140
dummy cell test, 138
EDTA, 39
Electrical safety, 11
electrochemical reaction, 100
Environmental specifications, 23
filter, 73, 86
frequency, 83
fuses
replacement of, 12, 46
Helium degassing, 40
I/E converter, 66, 68
I/E output, 124
I/E relationship, 99
inject marker, trigger, 117
installation
DECADE Elite, 21, 148
Instrument description, 16
integrator
158 175_0010 - DECADE Elite / Lite user manual - rev 13
example, 107
scanning voltammogram, 102
Serial mode detection, 69
Solvents, 12
specifications DECADE Elite, 111, 115
stop flow test, 138
Symbols, 7
System Operation, 11
Tefzel rotor, 95
three-electrode configuration, 66
Time table, 62
transport box, 26
ts, 94
USB, 116
Vacuum degassing, 40
valve configuration, 117
Vespel rotor, 95
Vici Valco, 41
voltage clamp, 66
voltammogram
construction, 101, 106
introduction, 99
warning sign, 7
Warning Symbols, 7
Waste disposal, 13
WEEE directive, 10
working electrode, 66
material, 97
Working environment & safety, 11