0% found this document useful (0 votes)
69 views

8625 - EN-torque Sensor

This high precision torque sensor measures torque from 0.01 to 10 Nm with high accuracy. It has standardized output signals of ±10V or USB and can be configured using software to set averaging, filters, and tare values. Accessories include mounting brackets and flanges to integrate the sensor into test and measurement setups.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
69 views

8625 - EN-torque Sensor

This high precision torque sensor measures torque from 0.01 to 10 Nm with high accuracy. It has standardized output signals of ±10V or USB and can be configured using software to set averaging, filters, and tare values. Accessories include mounting brackets and flanges to integrate the sensor into test and measurement setups.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

High Precision Torque Sensor

Code: 8625 EN
For non-rotating applications Delivery: ex stock / 2 weeks

Warranty: 24 months
Model 8625

NEW

n Measurement ranges from


0 ... 0.01 Nm to 0 ... 10 Nm
n Linearity error from ≤ 0.05 % F.S.
n Standardized output signal
n Output signal ± 10 V / USB (optional)

8625 EN
n Tare function, filter and average values configurable
n Optional with burster TEDS

Application Description
This high precision torque sensor is designed for both static The strain-gauge based sensor’s modular design allows pre-
and dynamic measurements on non-rotating applications. cise configuration for the desired application:
It is particularly suitable for torque measurements on, for ► mV/V with standardized output signal
instance, extremely small electrical actuating drives and
micro-mechanical actuator elements, or for measuring reaction ► ± 10 V output signal, configuration via USB
torques e.g. on micro-motors. ► ± 10 V output signal, configuration and measurement
The high accuracy of measurement also makes this sensor ideal via USB
for use as a reference in many fields of industrial manufacture With the integrated amplifier option, the sensor directly
as well as laboratory research and development projects. Not supplies a voltage signal of 0 ... ± 10 V that is proportional to
containing any rotating parts, it requires no maintenance if the torque. The sensor can be configured via the micro-USB
properly used. interface, providing access to, for example, a filter frequency
Available accessories include mounting brackets and flange setting, averaging, and a tare function. With the USB option,
adapters, which enable quick, easy and practical integration in addition to the voltage output, the measurement function
of the sensor into existing or newly developed setups and is available via USB as well. The supplied DigiVision software
test benches. can be used for measuring and storing data, or additionally
drivers for e.g. LabVIEW are available. Integration into custom
Other possible applications: software is possible via DLL.

► Test setup for precision mechanics The burster TEDS option (electronic data sheet, memory
chip with sensor-specific data) allows rapid configuration of
► Measuring the frictional torque of bearings compatible evaluation units (instrumentation amplifier, indi-
► Measuring the torques applied to vehicle control cator, ...).
elements and knobs
► Acquisition of breakage moments on screw caps

Technical changes reserved. All data sheets at www.burster.com burster praezisionsmesstechnik gmbh & co kg Germany www.burster.com
Talstr. 1-5 . Gernsbach 76593 . Phone +49-7224-6450 [email protected]
8625 - 2
Technical Data
Order code Measurement Dimensions [mm]
Range
L A B C D E
8625-4010-VXXXXX 0 ... ± 0.01 Nm 59 5,5 5,5 5 4 8
8625-4020-VXXXXX 0 ... ± 0.02 Nm 59 5,5 5,5 5 4 8
8625-4050-VXXXXX 0 ... ± 0.05 Nm 65 8 9 7 6 8
8625-4100-VXXXXX 0 ... ± 0.1 Nm 85 18 19 17 8 10
8625-4200-VXXXXX 0 ... ± 0.2 Nm 85 18 19 17 8 10
8625-4500-VXXXXX 0 ... ± 0.5 Nm 85 18 19 17 8 10
8625-5001-VXXXXX 0 ... ± 1 Nm 85 18 19 17 8 10
8625-5002-VXXXXX 0 ... ± 2 Nm 85 18 19 17 8 10
8625-5005-VXXXXX 0 ... ± 5 Nm 85 18 19 17 8 10
8625-5010-VXXXXX 0 ... ± 10 Nm 85 18 19 17 10 12
higher measurement ranges on request

Dimensional drawing standard sensor

Dimensional drawing bracket-mounted model

Dimensional drawing flange-mounted model


2487-008625EN-5672-111527

Technical changes reserved. All data sheets at www.burster.com burster praezisionsmesstechnik gmbh & co kg Germany www.burster.com
Talstr. 1-5 . Gernsbach 76593 . Phone +49-7224-6450 [email protected]
8625 - 3

Flange-mounted model Bracket-mounted model

The flange adapter allows easy integration of the sensor The bracket provides a quick-to-fit and stable fixture
in existing equipment with a flange connection. When for the sensor. When ordered with the sensor, the
ordered with the sensor, the flange adapter comes pre- bracket comes pre-fitted; please refer to order code.
fitted; please refer to order code.
Alternatively it can be ordered separately as an accessory.
Alternatively it can be ordered separately as an accessory. Model 8625-Z002
Model 8625-Z001 Please refer to the accessories data sheet for further
Please refer to the accessories data sheet for further technical details.
technical details.

Torque sensor with built-in


USB port (option) Metal-bellows coupling, 8691 series, accessory

This sensor model comes with a USB port in addition to Metal-bellows couplings provide optimum misalignment
the 0 ... ± 10 V output. correction. For the best possible misalignment
Two versions are available: correction, we recommend torsionally rigid metal-
bellows couplings. These couplings feature extremely
n ± 10 V output signal, USB used solely for configuration high torsional stiffness under applied torque and
n ± 10 V output signal, USB used for both configuration extremely low restoring forces. The clamp fasteners
and measurement come in two parts for easy and reliable fitting/removal.
When a USB-based measurement is launched, the analog Please refer to the accessories data sheet for further
output signal is disabled because it is not possible to use technical details.
both forms of output simultaneously.

8625
DigiVision configuration and analysis software
DigiVision Features
►► Can be used to actuate tare function,
with value stored in sensor
►► Configuration options for averaging
and filters; value stored in sensor
►► Intuitive user interface
►► Automatic sensor identification
►► Sensor calibration data readout

DigiVision Light PC software USB measurement option


DigiVision configuration and analysis software ►► Numerical & graphical display and measurement of the
(supplied with sensor) Model 8625-P001 physical torque value
DigiVision Standard PC software ►► Practical start and stop trigger functions
DigiVision configuration and analysis software; up to ►► 4 limits can be configured for each measurement channel
400 measurements/s for up to 16 channels Model 8625-P100
►► MIN/MAX value acquisition
DigiVision Professional PC software
►► Automatic scaling
DigiVision configuration and analysis software
with additional configurable maths channel; ►► Measurement reports can be saved as Excel or PDF file
up to 32 channels Model 8625-P200
►► Archive viewer for displaying sets of curves
►► Full version allows multichannel measurements, even with
different sensors (e.g. 9206, 8661)

Technical changes reserved. All data sheets at www.burster.com burster praezisionsmesstechnik gmbh & co kg Germany www.burster.com
Talstr. 1-5 . Gernsbach 76593 . Phone +49-7224-6450 [email protected]
8625 EN - 2
Technical Data
Order Code End of Rel. Non- Rel. Tolerance of Sensi- Maximum Maximum Spring Mass Moment of Weight
Measuring linearity Hysteresis Sensitivity: tivity Axial Load Radial Constant Inertia Measuring
Range [% F.S.] [% F.S.] [% F.S.] [mV/V] [N] Load [N] [Nm/rad] side [10-6 kg*m2] [g]
8625-4010-VXXXXX ± 0.01 Nm 0.15 0.15 0.2 0.25 50 1 5 0.022 150
8625-4020-VXXXXX ± 0.02 Nm 0.1 0.1 0.1 0.25 50 1 8 0.026 150
8625-4050-VXXXXX ± 0.05 Nm 0.1 0.1 0.1 0.5 50 1 10 0.059 150
8625-4100-VXXXXX ± 0.1 Nm 0.05 0.1 0.1 0.5 50 1 18 0.749 180
8625-4200-VXXXXX ± 0.2 Nm 0.05 0.1 0.1 0.5 50 1.5 41 0.812 180
8625-4500-VXXXXX ± 0.5 Nm 0.05 0.1 0.1 0.5 50 2 115 0.886 180
8625-5001-VXXXXX ±1 Nm 0.05 0.1 0.1 0.5 50 3 261 1.15 190
8625-5002-VXXXXX ±2 Nm 0.05 0.1 0.1 1.0 50 6 304 1.17 190
8625-5005-VXXXXX ±5 Nm 0.05 0.1 0.1 1.0 200 15 1242 1.44 190
8625-5010-VXXXXX ± 10 Nm 0.05 0.1 0.1 1.0 200 30 2604 2.2 210
Higher measuring ranges on request.
Technical Data without amplifier Mounting instructions
Electrical values ► make sure the connecting shaft is exactly aligned
Bridge resistance (full bridge): 1000 Ω ► suitable couplings must be used to prevent any forces arising
from a parallel or angular offset of the shafts
Excitation voltage: 5V
► do not exceed the permitted axial and radial forces during fitting
Max. excitation voltage: 10 V
and operation
Environmental conditions ► follow the mounting instructions when fitting the flange adapter
Range of operating and nominal temperature: - 20 °C ... + 80 °C or bracket to an existing sensor
Sensitivity of temperature effects: ► please refer to our operating instructions for detailed information
at zero: ≤ 0.05 Nm 0.020 % F.S./K (www.burster.com)
≥ 0.1 Nm 0.015 % F.S./K
on final value: ≤ 0.05 Nm 0.015 % F.S./K Accessories
≥ 0.1 Nm 0.010 % F.S./K Mating connection Model 9900-V594
Electrical connection Mating connection 90°-angle Model 9900-V596
7 pins plug connection (mating connector included on sensor delivery) Connecting cable, length 3 m,
other end free Model 99594-000A-0150030
Technical Data with amplifier/USB Connecting cable for burster desktop instruments
Electrical values with 12 pin socket, 3 m Model 99141-594A-0150030
Rated supply voltage range: 5 ... 30 VDC for model 9235 and model 9311 Model 99209-594A-0150030
(or 5 V via USB)
DC power consumption: approx. 1 W Order code
Output voltage at ± rated torque: ± 10 V Torque sensor Model 8625-XXXX-V00 □□0
Output resistance: < 500 Ohm
Output voltage 10 V incl. configuring USB 0
Insulation resistance: zero (binding capability) Output voltage 10 V incl. configuring and measuring USB 1
-3 dB cut-off frequency: 5000 Hz Output signal standardized, mV/V 3
Ripple: < 50 mVss Output signal standardized, mV/V with TEDS 4
Calibration signal: 10.00 VDC
both round shaft ends 0
Environmental conditions flange-mounted 4
Range of operating and nominal temperature: - 20 °C ... + 60 °C bracket-mounted 7
Sensitivity of temperature effects:
at zero: ≤ 0.05 Nm 0.020 % F.S./K Order information
≥ 0.1 Nm 0.015 % F.S./K
8625 with 10 Nm measurement range, ± 10 V output signal, USB
on final value: ≤ 0.05 Nm 0,015 % F.S./K port, measurement via USB, flange-mounted, including 8661-P001
≥ 0.1 Nm 0.010 % F.S./K measurement and analysis software 8625-5010-V00140
Electrical connection
7-pin miniature connector, additionally micro-USB interface for Manufacturer Calibration Certificate (WKS)
configuration/measurement (mating connector and USB cable Calibration of a sensor or a measuring chain, clockwise and/or
supplied) counterclockwise torque in 20 % steps, increasing and decreasing.

Mechanical values German-accredited DAkkS calibration


Linearity error and hysteresis: refer to tables The DAkkS calibration certificate (in accordance with German
Dynamic overload safe: up to 70 % from nominal value Calibration Service DKD-R 6-1 guidelines, clockwise and/or
Protection class: acc. EN 60529 IP40 anticlockwise torque) includes at least three measurement cycles in
steps of 10% of the measurement range, rising and falling.
Max. operation torque ≤ 0.1 Nm: 200 % of nominal torque
Max. operation torque ≥ 0.2 Nm: 150 % of nominal torque Please ask for out new torque brochure
Breakaway torque: 300 % of nominal torque or take a look at www.burster.com
Alternating load: 70 % of nominal torque
2487-008625EN-5672-111527

Material:
housing: made of anodized aluminium
shaft ≤ 0.05 Nm: high-strength aluminium 3.1354
shaft ≥ 0.1 Nm: steel shell 1.4542
Weight: refer to tables
Mounting: refer to dimensional drawing

Technical changes reserved. All data sheets at www.burster.com burster praezisionsmesstechnik gmbh & co kg Germany www.burster.com
Talstr. 1-5 . Gernsbach 76593 . Phone +49-7224-6450 [email protected]