Lecture 1 EC410 - V3
Lecture 1 EC410 - V3
Lecture (1):
Introduction to Measurement Systems
2
Grading System
3
References
4
Course Objective
• This course aims at bridging traditional concepts with
modern technologies of electrical and electronics
measurements and instrumentation, as well as to be
introduced to several applications.
How?!
5
Analytical Thinking: Problem solving
6
Course Outline
7
Outline
• Concept of Measurement Systems
• Methods of Measurements
8
Concept of Measurement Systems
• Definition of Measurement
Process:
It is the act, or the result, of a
quantitative comparison between a
given quantity and a quantity of
the same kind chosen as a unit.
Direct Indirect
• The value of the unknown quantity is
• The unknown quantity is measured determined by measuring the functionally
directly instead of comparing it with a related quantity and calculating the
standard. desired quantity rather than measuring it
directly
• Ex: current by ammeter, voltage by
voltmeter..etc • Ex: R= V/I
10
Measurement Standard
• Any time you measure anything, you are comparing it to something
whose value you think you know. You assume your ruler is 1 ft long.
But who says what a foot is?
12
13
What is the meaning of
instrument calibration?
The calibration of a measuring instrument lets us know
how far off the measurement on the instrument is from
the real value of what we are measuring.
14
What is the adjustment of a measuring system?
Calibration meaning is often confused with the adjustment of a
measuring system. The adjustment is a set of operations that are
done on a measuring system so that it provides the correct
measurements for the quantity you want to measure. After an
adjustment of a measuring system, the measuring system must
usually be recalibrated.
15
Why do calibrations need to be done?
• The four most common reasons why calibrations are necessary are:
16
What is a calibrator?
A calibrator is a measurement standard used in
calibration. A calibrator is often used as a
reference to measure other quantities of the
same kind. This helps to establish values and
measurement uncertainties.
17
1-Pressure Calibration
• Pressure calibration is a key function across multiple industries where measurement equipment is used to
monitor process performance and safety, typically measuring gas and hydraulic pressure.
• Nowadays, many companies are certified to quality standards such as the ISO9000. There are
numerous quality procedures in order to maintain quality standards, and because many
industrial processes rely on the measurement of pressure, pressure calibration plays an
important part in company’s quality assurance.
➢ Digital Indicators
➢ Transducers
➢ Transmitters
18
2. Temperature Calibration
❑ In all processed where temperature readings have a critical role, temperature calibration are
conducted and are carried out in a controlled environment.
In the process of temperature calibration, thermistors, thermocouples or Platinum
resistance thermometers (PRTs), also called resistance temperature devices (RTDs) are
usually used.
❑ A temperature calibration can only be performed by comparing a known standard, in a stable
temperature environment to the probe being tested.
A few examples of equipment that require temperature calibration on a periodic basis are:
Thermometers/Thermocouples
Dial Thermometers
Chambers/Furnaces
19
Basic Requirements of a Measurement System / Instrument
The following are the basic requirements of a good quality measurement system /
instrument:
a) Ruggedness
b) Linearity
d) Repeatability
20
Measurement Systems and Its Elements
Gnereally…
21
Measurement Systems and Its Elements
The main functional elements:
Primary sensing element Signal conditioning element Data transmission element
22
Primary sensing element
• A transducer is defined as a device which converts a physical quantity into an electrical quantity. The
output of the sensor and detector element employed for measuring a quantity could be in different
analogous form.
• This output is then converted into an electrical signal by a transducer. This is true of most of the cases
but is not true for all. In many cases, the physical quantity is directly converted into an electrical
quantity by a detector transducer. The first stage of a measurement system is known as a detector
transducer stage.
23
Signal conditioning element
• The output signal of transducers contains information which is further processed by the system.
• Many transducers develop usually a voltage or some other kind of electrical signal and quite often the signal developed is
of very low voltages, may be of the order of mV.
Problem
• This signal could be contaminated by unwanted signals like noise due to an extraneous source which may interfere with
the original output signal.
• Another problem is that the signal could also be distorted by processing equipment itself. If the signal after being sensed
contains unwanted contamination or distortion, there is a need to remove the interfering noise / sources before its
transmission to next stage. Otherwise we may get highly distorted results which are far from its true value.
Solution
• The solution to these problems is to prevent or remove the signal contamination or distortion.
• The operations performed on the signal, to remove the signal contamination or distortion, is called Signal Conditioning.
• The term signal conditioning includes many other functions in addition to variable conversion and variable manipulation.
• Many signal conditioning processes may be linear, such as, amplification, attenuation, integration, differentiation,
addition and subtraction.
• Some may be non-linear processes, such as, modulation, filtering, clipping, etc.
• The signal conditioning processes are performed on the signal … why??to bring it to the desired form for further
transmission to next stage in the system. 24
Variable conversion element
• The output signal of the variable sensing element may be any
kind. It could be a mechanical or electrical signal. It may be a
deflection of elastic member or some electrical parameter,
such as, voltage, frequency etc.
• Also if the output at one stage is analogue form and the next
stage of the system accepts input signal only in digital form.
In such cases, we will have to use as Analogue /Digital
25
converter.
Data transmission element
• There are several situations where the elements
of an instrument are actually physically
separated.
• In such situations it becomes necessary to
transmit data from one element to another.
• The element that performs this function is called
a Data Transmission Element.
• For example satellites or the air planes are
physically separated from the control stations at
earth. For guiding the movements of satellites or
the air planes control stations send the radio by a
complicated telemetry systems.
• The signal conditioning and transmission stage is
commonly known as Intermediate Stage. The telemetry consists of
sensors for pressure,
A telemeter is a physical device used in telemetry. It
temperature,
consists of a sensor, a transmission path, and a
display, recording, or control device. Electronic and humidity and a wireless
devices are widely used in telemetry and can be transmitter to return the
wireless or hard-wired, analog or digital. captured
data to an aircraft. 26
Data presentation element
• The information conveyed must be in a convenient form. In
case data is to be monitored, visual display devices are
needed.
• In case the data is to be recorded, recorders like magnetic
tapes, high speed camera and T.V. equipment; storage type
C.R.T., printers, analogue and digital computers may be
used
• For control and analysis purpose computers and the control
elements are used.
• The final stage in a measurement system is known as
terminating stage.
27
..Let’s get into the blocks of a measurement process..
28
Example of a complete measurement system:
Functional Elements of a Clinical Thermometer
let us consider the simple clinical thermometer shown in Figure
• The thermometer bulb containing mercury acts as the primary sensing
element as well as a variable conversion element.
• It senses the input quantity, the temperature.
• On account of the increase in temperature the mercury in bulb expands
and its volume is increased.
• The temperature signal is converted into volume displacement.
• As the mercury expands it move through the capillary tube in the
thermometer stem, integrated to the bulb.
• The cross section area of the capillary being constant, the volume signal is
thus converted into linear distance signal.
• The capillary thus has the role of signal manipulation and data
transportation elements.
• The final data presentation stage consists of the scale on the thermometer
stem, which is calibrated to give the indication of the temperature signal
applied to the thermometer bulb.
• A restriction bend is provided in the clinical thermometers at the junction
of the bulb and the capillary, which does not allow the back flow of
mercury to the bulb once it has expanded to the capillary. Thus the
restriction in the capillary acts as the data storage function of the
29
instrument.
How to evaluate a measurand?
• Data taken in experiments are not perfect because the accuracy of the data depends on
the accuracy of the test equipment and the conditions under which the measurement was
made.
• In order to properly report measured data, the error associated with the measurement
should be taken into account.
• The difference between the true or best-accepted value of some quantity and the
measured value is the error
33
34
Errors in Measurement - Absolute Error and Relative Error
❑Definition:
• The errors in measurement are a mathematical way to show the uncertainty in the measurement.
• It is the difference between the result of the measurement and the true value of what you were
measuring.
• Any measurement that you make is just an approximation, 100% accuracy is not possible.
• If you measure the same object two different times, the two measurements may not be exactly the
same.
• The difference between the two measurements is called a variation in the measurements.
• This variation introduces an unwanted but unavoidable uncertainty in our measurement.
• This uncertainty is called the Errors in measurement. This ‘error’ should not be confused with a
‘mistake’.
• Error, unlike mistake, does not mean that you got the wrong answer. It just means you didn’t get as close
to the true value as possible.
35
How to find errors in measurement
Tolerance/ Margin of Error
• Error in measurement may be represented by a tolerance (margin of error).
• Machines used in manufacturing often set tolerance intervals or ranges in
which product measurements will be tolerated or accepted before they are
considered flawed.
❑Tolerance refers to the greatest range of variation that can be allowed or is
acceptable and it is one half of the precision of an instrument. So to obtain the
range, we add and subtract one half of the precision of measuring instruments.
• The absolute error of the measurement shows how large the error actually is.
• Absolute error is inadequate due to the fact that it does not give any details
regarding the importance or the graveness of the error.
37
Relative Error
❑ Relative error represents the ratio of the absolute error of the measurement to the accepted
measurement.
• This way we can determine the magnitude of the absolute error in terms of the actual size
of the measurement.
• If the true measurement of the object is not known then the relative error is found using the
measured value.
• The expression for relative error is:
38
Significant Digits
39
40
How to determine the number of significant figures
Example:
the number 12,100 Ω can have three, four, or five significant digits. To clarify the
significant digits, scientific notation (or a metric prefix) should be used. For example,
12.10 kΩ has four significant digits
41
Example:
42
Example:
43