0% found this document useful (0 votes)
40 views2 pages

Spesifikasi Teknis CT Revolution CT

The document provides specifications for a dual energy CT 512 slice scanner with spectral imaging. It lists technical details like the number of detector rows, slice thickness, rotation speed, dose reduction features, reconstruction capabilities, and clinical applications.

Uploaded by

Arya Bintang
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
40 views2 pages

Spesifikasi Teknis CT Revolution CT

The document provides specifications for a dual energy CT 512 slice scanner with spectral imaging. It lists technical details like the number of detector rows, slice thickness, rotation speed, dose reduction features, reconstruction capabilities, and clinical applications.

Uploaded by

Arya Bintang
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

NAME OF PRODUCT Dual Energy CT 512 slice with Spectral Imaging

BRAND / MANUFACTURER GE Healthcare


MODEL / TYPE Revolution CT
COUNTRY OF ORIGIN USA
FACTORY OF ORIGIN GE MEDICAL SYSTEM LLC, USA
PRODUCT'S REGISTRATION (AKL/AKD) AKL 21501412197
FDA Clearance Yes
CE Mark (MDD) Yes
WHERE MARKETED World Wide

REFERENCE OF TECHNICAL
TYPE Dual Energy CT 512 slice with Spectral Imaging
Slice per rotation 512 slice
Total detector width, z-axis, mm 160 mm

GANTRY FEATURES
Gantry aperture, cm 80 cm
Scan localizer laser alignment
o
Rotation Time for full 360 0.28s per 360°
Breath Hold Guidance -
Audio/Voice instruction provided
Audio Instruction / Voice
Breath Hold Guidance -
Visual indicator provided
Visual Indicator
Breath Hold Guidance -
Breath hold timer provided
Counting Down Timer
Gantry Display Touch screen Display on Gantry
Patient Information Data on Gantry Name, ID, age
ECG Wave graphics on Gantry Display Provided
Instruction video & Pediatric animation on Gantry LCD display Provided

X-RAY TUBE & GENERATOR


Generator kW output 103 kW
kVp range 70, 80, 100, 120, 140 kV
mA range 10 - 740 mA
Tube focal spots, mm 3 Focal spot

DETECTOR
Field of view 50 cm
Total detector width, z-axis, mm 160 mm
Number of rows 256 rows
Number of detector elements 212,992 cells with individual electronic / DAS channels
Minimum slice width, mm 0.625 mm

PATIENT TABLE
Vertical movement 500– 1030 mm
Horizontal range 2000 mm
Scannable range, cm 2000 mm
Horizontal speed 300 mm/s
Max load capacity, kg (accuracy, mm) 227kg (± 0.25mm)
Footswitch function Up, Down, Inward, Outward

CARDIAC PERFORMANCE & IMAGE QUALITY


High-contrast spatial resolution 0.23mm
Low Contrast Resolution 5 mm @3HU Difference, 8.8mGy CTDIvol
CT Number ±31,743 HU
Spectral Imaging GSI Xtream utilizes ultrafast kVp switching
Metal Artifact Reduction Smart MAR
Low Dose Cardiac Scan 1-Beat cardiac
Fast Acquisition Cardiac Scan 1-Beat cardiac
Prospective ECG cardiac Provided
Effective Cardiac temporal resolution 29 ms
Coronary Motion artifact correction SnapShot Freeze
Equivalent gantry rotation speed 0.058s equivalent gantry rotation speed
Extended coverage perfusion imaging 16 cm coverage perfusion of: Heart, Brain, Liver, Kidneys
Dynamic Acquisition Modes 4D imaging to acquire morphology and perfusion information and vascular flow to the organs

LOW DOSE FEATURES


Lowest mA selection (low dose CT) 70 kV selection.
Highest mA selection (high performance CT) 740 mA
Dose-modulation technique 3D Dose Modulation, SmartmA, Automatic Exposure Control (AEC), ECG Modulated mA, 70 kV Scanning
Pediatric-specific dose control Color Coding for Kids
Iterative image reconstruction ASiR-V
Sliding collimation (overbeaming reduction) Smart Track, Smart Beam
Superficial organ dose reduction Organ Dose Modulation (ODM)
Low-dose cardiac (axial acquisition) 1-Beat cardiac
Max heart rate, bpm. Temporal resolution effective 29 ms
Arrhythmia correction Smart Arrhythmia Management
CTDIvol (CTDI volume), DLP (Dose Length Product), and Dose Efficiency computation and display during scan
Dose check, Dose display, Dose reporting
prescription

IMAGE RECONSTRUCTION
Computer CPU Intel XEON
Monitor displlay Dual Monitors/Screen Size: 24’’/Screen Resolution: 1920x1200
Disk 1 TB
Image capacity (512x512) 700,000 Uncompressed DICOM Images (512x512)
RAM 64 GB
Reconstruction matrices 512 x 512
Display Matrix 1024 x 1024
Reconstruction speed 65 fps

IMAGE PROCESSING
Standard or optional Standard
Recommended postprocessing workstation GE AW advanced workstation provided
Remote access to raw image data Yes
Remote access to clinical applications Yes
DICOM 3D image export Yes

CLINICAL APPLICATIONS AND FUNCTIONALITY


Auto bone removal AutoBone Xpress
Auto vessel mapping AVA Xpress
Vessel analysis (noncardiac) AVA Xpress
Coronary artery calcification scoring SmartScore
Coronary artery vessels measure include stenosis, density of stenosis, and length of stenotic area CardIQ Xpress
2D/3D coronary vasculature tree images with 1-click automatic vessel tracking. CardIQ Xpress
Cardiac quantification of left and right ventricular volumes CardIQ Function Xpress
Ejection fraction and chamber volume analysis on LV and RV automatic ejection fraction CardIQ Function Xpress
Myocardial analysis with wall motion, wall thickness, and mass calculations CardIQ Function Xpress
Lung nodule assisted reading LungVCAR
Lung nodule CAD LungVCAR
Digital Contrast Agent (DCA) LungVCAR
Lung tissue analysis Thoracic VCAR
Quantitative measurements of the lungs/COPD Thoracic VCAR
Parenchyma analysis, automated measurements of wall thickness Thoracic VCAR
Lobe segmentation measurements to the individual lung lobes Thoracic VCAR
Airway analysis, automatic segments and tracks airways Thoracic VCAR
Virtual colonoscopy assisted reading Volume Viewer - Navigator
Virtual colonoscopy CAD Advantage CTC Pro
Brain perfusion post processing: Brain stroke, Brain tumor CT Perfusion 4D - Neuro
Body perfusion post processing: Body tumor, Liver perfusion, Pancreas perfusion, Prostate perfusion, Kidney perfusion, Soft tissue
CT Perfusion 4D - Body
perfusion, Spleen perfusion, Bone perfusion
Perfusion imaging Neuro Perfusion, Organ Perfusion
Liver Software Hepatic VCAR
Spectral Viewer Yes
Stroke Software Stroke VCAR, Fast Stroke

SYSTEM INTEGRATION
DICOM Storage Service Class Standard
Service Class User (SCU) for image send Standard
Service Class Provider (SCP) for image receive Standard
Service Class User (SCU) for storage commitment Standard
DICOM Query/Retrieve Service Class Standard
DICOM Storage Commitment Class Push Standard
DICOM Modality Worklist Standard
DICOM Modality Performed Procedure Step Standard
DICOM Print Standard
DICOM Structured Dose Report Standard

ROOM SIZE
POWER REQUIREMENTS 150 kV
Cooling unit Aircon without water chiller
SHIELDING REQUIREMENTS According to BAPETEN regulation
ECG patient monitor Part of the system
Partial UPS 3 fasa Included
UPS for Operator Console PC Included