Osmometro 3d3
Osmometro 3d3
Service Manual
7. Schematics
Hitachi Processor Board
PCB600 Assembly Drawing
PCB600 Main Board Schematic
Interconnection Schematic - Hitachi
Intel Processor Board
PCB605 Assembly Drawing
PCB605 Application Board Schematic
4D3605 Assembly Drawing
PCB620 Assembly Drawing
PCB620 Processor Board Schematic
Interconnection Schematic - Intel
APPENDICES
A: Symbol Definitions
B: Product Disposal and Recycling
C: Service Log
Index
Fuses: 2-Amp time delay for 100-130 VAC; Start-up Time: From stand-by: immediately
1-Amp time delay for 200-250 VAC From power on: under 5 minutes
Operating Temperature
(ambient): 18°C to 35°C
(64°F to 95°F)
Resolution: ±1 m°C or m°H Refer to Regulatory Notices (see page 11) for appli-
cable standards.
Repeatability: ±2 m°C or m°H (1 S.D.)
Communications:
DTE RS-232C, parallel printer, barcode
Operating Temperature
(ambient): 18°C to 35°C
(64°F to 95°F)
This symbol indicates the presence • Route power cords so that they will not be
of alternating current (AC). walked on, tripped on, or pinched by items
placed upon or against them. Pay particu-
• WARNING: Handle all biohazardous mate- U.S. Safety This product has been listed by
rials according to established good laborato- ETL testing laboratories as being
ry practices and follow your institution’s in compliance with the require-
exposure control plan. Persons handling ments of UL 3101-1, 1st Edition,
human blood and body fluid samples must "Electrical Equipment for
Laboratory Use; Part 1: General
be trained in blood-borne hazards and
Requirements". The "US" in the
observe universal precautions. Universal pre- lower right of the ETL mark
cautions is an approach to infection control, demonstrates this listing.
where all human blood and body fluids are
treated as if known to be infectious. Use per- Canadian Safety This product has been listed by
sonal protective equipment such as gloves, ETL testing laboratories as being
in compliance with the require-
gowns, etc., to prevent exposure. Store bio-
ments of CAN/CSA C22.2
hazardous materials in regulated waste con- No.1010.1-92, "Safety
tainers and dispose of these materials in a Requirements for Electrical
safe and acceptable manner that is in com- Equipment for Measurement,
pliance with all country, state and local Control and Laboratory Use;
requirements. General Requirements";
Including Amendment Two. The
• If a biohazardous material is spilled on or "C" in the lower left of the ETL
inside the equipment, decontaminate the mark demonstrates this listing.
equipment using a 1% bleach solution, or as
EC Declaration of This product meets the intent of
outlined by those policies and procedures Conformity - EMC Directive 89/336/EEC for
established within your institution. Electromagnetic Compatibility.
Compliance was demonstrated
• To avoid injury or fire hazard, do not operate using the following standards, as
this product in an explosive atmosphere. listed in the Official Journal of
the European Communities:
Consult the Declaration of Con-
Regulatory Notices formance certificate shipped with
the product for the latest update.
• This product has been designed and manu- • EN 61326: 1998, Group 1,
Class B, "Electrical Equipment
factured in accordance with U.S., Canadian,
for Measurement, Control, and
and European regulatory requirements as Laboratory Use".
outlined below. Modifications made to this
product that are not expressly approved in EC Declaration of: This product meets the intent of
writing by the manufacturer will void the Conformity - Low Directive 73/23/EEC, the Low
user’s authority to operate this product, pre- Voltage Voltage Directive. Compliance
was demonstrated using the fol-
viously issued factory approvals, and the
lowing standards as listed in the
user’s rights under the warranty. Official Journal of the European
Communities: Consult the
• The distributor or dealer may have applied
Declaration of Conformance cer-
additional local, national, or international tificate shipped with the product
approvals to this product. Consult the dis- (if required) for the latest update.
tributor or dealer for more information and · EN 61010-1 (1993), "Safety
documentation. Requirements for Electrical
Equipment for Measurement,
• Connections to this product must be made Control and Laboratory Use -
with shielded cables. Use of non-shielded General Requirements";
cables may violate RFI/EMI limits. Including Amendment Two
(1995).
Symbol Conventions
Over the years, Advanced Instruments, Inc. The larger board is called the application
has made several improvements to the board, and the smaller board is called the
3D3/4D3 system. It is important to identify processor board. The processor board
some of these changes to properly service contains an Intel 80C186 microprocessor
your instrument. Although every attempt has and two field-programmable Flash
been made to cover all the configurations, EPROM's. All references to this configu-
some descriptions may vary from what is ration will use the terms control board
present in your particular situation. If you are set, application board, or processor
having difficulty comparing your system to board.
this manual, consult the factory for assistance.
Please refer to the applicable system photo- Motor control relays
graphs and circuit descriptions once you have
1. The original configuration used two chas-
identified your configuration. You may also
sis-mounted relays located on the center
refer to suffix history chart located in the
rear interior wall.
Design Changes section of this manual, but
keep in mind that many of these changes were 2. The second configuration used two PCB-
backward-compatible, so use the chart only as mounted relays located on the center rear
a reference. interior wall. This version may be used
as a replacement for the first configura-
tion.
System control electronics
1. The original main control board was a
Operator/Supervisor keyswitch
single board located on the right side,
containing a Hitachi 63A03 microproces- 1. The first configuration had no keyswitch.
sor and a 256K Byte EPROM. This All references to keyswitch position in
board can be identified by the heat sink this manual should be ignored.
mounted component in the upper left cor-
ner of the main control board. This board 2. The second configuration used a black
and firmware are no longer supported, as plastic keyswitch located on the rear of
of October 1, 2001. the instrument. This keyswitch, when in
the operator position, locked the user out
2. The second main control board was a sin- of these features: setup, diagnostic test
gle board located on the right side, also utilities, and calibration.
containing a Hitachi 63A03 microproces-
sor, but had a 512K Byte EPROM. This 3. The third and current configuration uses a
board can be identified by the lack of a metal keyswitch located on the rear of the
heat sink mounted component in the instrument. This keyswitch, when in the
upper left corner of the main control operator position, locks the user out of
board. This board and firmware are no these features: setup and calibration.
longer supported, but can be replaced Note: If you have configuration number 2
with the Intel configuration. All refer- (above) and you replace the main con-
ences to this configuration will use the trol board with the control board set,
term main control board. the operator and supervisor positions
3. The third and current control board, locat- will be reversed, and your keyswitch
ed on the right side, is a two-board set. will now only lock out the setup and
calibration features.
L M N O
K P Q R
I
H
G
F
S
D T
C
U
B A
Processor
Board
Barcode Power
Port Supply
Applications Cooling
Serial Port
Board System
Memory Map: The 1 mega-byte of memory Memory: Two types of memory are avail-
address space contains read-only memory able: read-only and read/write. The chips
(Flash EPROM), read/write memory (static used here have 8-bit wide data buses. Used in
RAM), and the real time clock. About half of pairs, the memory becomes compatible with
the address space is unused, allowing for the processor’s 16-bit wide data bus. The
future expansion. The read-only memory is read-only memory is implemented using two
divided into four sections: reset vector, boot 128 kilo-bytes x 8 bits Flash EPROMs. Flash
code, parameter blocks (unused), and appli- memory is only programmable when a pro-
caiton code. The reset vector tells the proces- gramming voltage (in this case, 12 volts DC)
sor where to first start executing code; in this is applied to the memory; otherwise, it
case, the boot code is executed first after behaves like read-only memory. The boot
reset. The boot code determines if new soft- code controls the programming voltage, since
ware will be downloaded by checking if the this is the only place where new software can
dip switch is in the “PROGRAM” position. be downloaded. The read/write memory is
If it is, the boot code waits for new software implemented using two 128 kilo-bytes x 8
to be downloaded through the serial port. If bits static RAMs.
not, the application code starts.
Watchdog Timer: The watchdog timer pro-
The real time clock maintains the current date vides a mechanism to reset the processor
and time, and contains nonvolatile memory when the software is not behaving normally.
where the instrument’s operation parameters The software “pets” the watchdog every 100 -
are stored. A internal lithium battery pre- 200 milliseconds. As long as the watchdog is
serves the contents of the nonvolatile memory “petted”, nothing happens. However, if the
when the instrument is turned off. Finally, the sotware stops “petting” the watchdog, the
read-write memory is used for stacks, vari- watchdog “bites” the processor by activating
ables, etc., except for the small section allo- the processor’s RESET pin, causing the
cated for the interrupt vectors. The interrupt processor to reset. The watchdog will also
vectors tell the processor what code to run reset the processor if the power supply falls
when an interrupt occurs, and functions simi- below 4.75 VDC. These two features allow
larly to the reset vector. the instrument to recover from abnormal soft-
ware and hardware conditions.
Please note that all memory addresses and
sizes are always defined in terms of bytes, Real Time Clock (RTC): The real time
even though the processor handles words (1 clock maintains the current date and time.
word = 2 bytes, 1 byte = 8 bits; therefore, 16 The RTC uses an internal clock circuit with
bits = 1 word). an internal, 10-year life, lithium battery to
Voltages: Three voltages are present on the Drive Circuitry: Drive circuitry is provided
CPU board: +5 VDC, +12 VDC, and VPP. to turn on and off the four high current loads
+5 VDC powers all the logic. +12 VDC is such as the stir wire, the head motor, the ther-
switched on and off to create the flash memo- moelectric cooler, and the fan.
ry’s programming voltage, VPP.
The stir wire coil is controlled by the rear
Glue Logic: The glue logic performs the panel-mounted darlington transistor (Q2B).
functions of creating the RAM memory chip The drive circuit consists of a programmable
selects, the Flash memory write signals, the timer, D/A converter, and interface op-amps.
VPP control signal, the watchdog control sig- The timer provides a square wave of approxi-
nal, and accessing the dip switch. The CPU mately 71Hz, while the D/A converter con-
board has two software accessible hardware trols the output amplitude to the darlington
registers to read the dip switch and to control transistor.
watchdog and VPP.
The head motor is controlled by two relays
Connectors: A 64-pin connector provides located at the rear of the chassis that interface
address, data and interrupt signals. A 16-pin the 120 VAC motor to the DC logic. The
connector provides general chip selects and upper relay raises the head, while the lower
serial receive/transmit signals. The applica- relay lowers the head. The relays get their
tion PCB uses a subset of the available sig- commands from the application board via two
nals. FETs and the application logic. LEDs are
provided on the application board for moni-
toring the drive signals. LEDs are also pro-
PCB605 Application Board vided to monitor the signals from the head
sensor board, described below.
The application PCB contains all circuitry
specific to the 3D3/4D3 instruments. The The thermoelectric cooler is controlled by the
board includes voltage supplies, indicators, FET driver transistor mounted on the rear
analog-to-digital converters, drive circuitry, panel (Q1B). The FET is, in turn, controlled
parallel ports, serial ports, and various switch- by the microprocessor through the PLD appli-
es. cation logic. The processor varies the duty
cycle square wave in response to software
Analog-to-Digital Converters: Two analog- commands and block probe resistance.
to-digital converters are provided; one for the
sample thermistor probe, and one for the The fan is controlled by a FET transistor on
block thermistor probe. The thermistors vary the application board, that is interfaced to the
in resistance from approximately 2 kilo-ohms processor through the application logic.
at room temperature, to approximately 10
Power Supply
Electronics Cooling
Serial Port
Package System
Note that your instrument is only designed to * 4D3 cryoscopes that have a suffix before
support unidirectional communication with an "E" and 3D3 osmometers that have a suf-
external device. At this time, there is no pro- fix before "C" do not have the head tran-
tocol for bidirectional communication. sition board.
** 4D3 cryoscopes without a suffix do not
For a sample RS-232 Port Setup, please see have the head sensor board or the keypad
the RS-232 Supplemental Information in the transition board.
Appendix at the end of this user’s guide.
Serial Suffix ‘A’ CSA certification changes. Serial Suffix ‘O’ ---
July 1993 omitted
Serial Suffix ‘B’ Changes to accommodate sharing Serial Suffix ‘P’ Correct reboot problem.
October 1993 parts with Mark 2. January 1998
Serial Suffix ‘C’ • Software for multi-language Serial Suffix ‘Q’ 4-digit year display, for year
March 1994 capability. June 1998 2000.
• Plug-in probe.
Serial Suffix ‘R’ Sample well changed.
Serial Suffix ‘D’ 512K EPROM. June 1998
May 1994
Serial Suffix ‘S’ New IEEE display and new
Serial Suffix ‘E’ Software upgrade to add adjust- October 1998 cable.
April 1995 able buzz point.
No suffix Original model released. Serial Suffix ‘N’ Power supply changed.
August 1991 June 1997
Serial Suffix ‘A’ • Case manufactured in one Serial Suffix ‘O’ ---
August 1992 piece, instead of two. omitted
• Front panel (keypad) mounted
to the base assembly. Serial Suffix ‘P’ Reduction of reset problem.
• Keypad pulls down. November 1997
• Printer power output added to
the rear panel. Serial Suffix ‘Q’ Correct reboot problem.
• Head position sensors mounted January 1998
to a printed circuit board.
• Heat tansfer fluid door pulls Serial Suffix ‘R’ 4-digit year display, for year
down. June 1998 2000.
Serial Suffix ‘B’ One-way valves redesigned. Serial Suffix ‘S’ Sample well changed.
July 1993 June 1998
Serial Suffix ‘C’ • Supervisor/Operator keyswitch Serial Suffix ‘T’ New IEEE display and new
August 1993 added. October 1998 cable.
• Software change.
Serial Suffix ‘U’ • Intel Processor board set,
Serial Suffix ‘D’ • Transistors moved to the back December 2001 4D3605.
October 1993 wall. • Added Barcode port.
• New power entry module. • New power entry.
• FLASH memory upgrade.
Serial Suffix ‘E’ • Sample probe designed to plug
March 1994 into head unit. Serial Suffix ‘V’ • Added ISO 5764 seeking
• Software incorporated several April 2002 method.
languages. • New Delrin probe.
3D3 4D3
100 mOsm/kg Calibration Standard -422 m°H Calibration Standard
(10 5-mL ampules) . . . . . . . . . . . .3LA011 (-408 m°C) . . . . . . . . . . . . . . . . . .3LA023
500 mOsm/kg Calibration Standard -621 m°H Calibration Standard
(10 5-mL ampules) . . . . . . . . . . . .3LA051 (-600 m°C) . . . . . . . . . . . . . . . . . .3LA033
900 mOsm/kg Calibration Standard -530 m°H Lactrol® 530 Reference
(10 5-mL ampules) . . . . . . . . . . . .3LA091 Solution (-512 m°C) . . . . . . . . . .3LA030
The 3D3/4D3 cooling assembly should be 7. Insert an empty sample tube into the
cleaned once every two months to insure opti- freezing chamber and press START.
mal performance.
8. Run this test for 10-15 minutes and then
To flush the system: press STOP.
1. Open the door and remove the heat trans- 9. Remove the bottle with the diluted bleach
fer fluid bottle. and install a new bottle of heat transfer
fluid. Insert only the tube with the filter
2. Empty the contents and fill the bottle (A).
halfway with a solution of 20% bleach
and 1:5 distilled water. 10. Press START twice. Run this test until
the colored fluid begins draining out of
3. Replace the bottle and insert only the tube the return tube, and then press STOP.
with the filter (A). Completely immerse
the filter in the fluid. 11. Remove the return tube from the contain-
er and insert it into the heat transfer fluid
4. Insert the return tube (B) into another bottle. Do not immerse the return tube in
container. the heat transfer fluid (B).
Problems with the performance of the Before you repair or replace any
3D3/4D3 are usually identified by the soft- parts, turn the instrument off and
ware and displayed as error messages. The then back on after a short delay.
troubleshooting table on the following pages This allows the software to reinitial-
lists all error messages, indicates why the ize, and may correct what appears to
software generated the message, and lists be a serious hardware problem.
what may be causing the problem.
To check and reset the freeze amplitude set- 5. Check the position of the yoke (B) and
ting: clapper. The yoke must be 90º and flush
against the back wall of the head. Bend
1. Turn on the instrument. If there is a the yoke if the clapper appears too close
Supervisor/Operator keyswitch, turn it to or too far away from the coil frame. The
the Supervisor position (if required). clapper does not have to be directly paral-
lel to the back wall of the unit.
2. Press TEST. Press < or > to select “Stir/
Freeze Test”, and then press START.
D F G
1. Turn on the instrument. If there is a • The sample well fluid passage (D) is not
Supervisor/Operator keyswitch, turn it to obstructed by debris, and the pump is mov-
the Supervisor position (if required). ing the heat transfer fluid into the freezing
chamber.
2. Press TEST. Press < or > to select
“Head Up/ Down Test”, and then press Note: If the sample well fluid passage is
START. obstructed, disassemble and clean the
sample well following the instructions
3. Place an empty sample tube into the found earlier in this section. Also,
freezing chamber. examine the one-way check valves (E)
to make sure they are not stuck open or
4. At “[START] Test [STOP]”, press closed.
START.
Note: If your instrument is equipped with an
The head should move continually up and Intel control board set, a series of
down. diagnostic LEDs have been provided to
allow visual identification of drive sig-
Use this test to ensure the following compo- nals and sensor activation. The yellow
nents are working properly: LEDs marked “Drive Up” and “Drive
Dn” indicate that the control electron-
• The motor (A) is running.
ics have issued a command to the head
• The gears (B) are turning properly. motor relays. The red LEDs marked
“Head Up” and “Head Dn” indicate
• The sensors (C) are detecting the head-up that the corresponding optical sensor
and head-down conditions. has been tripped.
B C D
To set the correct sample bin number, fol- 4. At “Select Setup Item”, press < or > to
low steps 6-8. select “Select Block Bin #”, and then
press START.
If the current sample bin number matches
the number just recorded, press STOP 5. Enter the correct number and then press
twice to exit the menu. ENTER.
2. Replace the cover. Plug in the instrument 4. Check the setting by placing a sample
and turn it on. If there is a Supervisor/ (water or standard solution) in position on
Operator keyswitch, turn it to the the probe. If bubbles form in the sample,
Supervisor position (if required). or if the stir wire hits the tube, the current
setting is too high and must be adjusted
3. At “Press START to Continue”, press downward.
SETUP.
5. When the setting is correct, record the
new stir amplutide number for your
records, and press ENTER.
3. Press STOP twice to end the test and exit Barcode Test
the Test menu. (Only Intel instruments with a barcode con-
nector.)
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
1. Turn off the power and unplug the instru- Option Switch Settings:
ment. Processor board 3D3620/4D3620 is equipped
with a three-position dip switch. Each switch
2. Remove the instrument cover (see section position and combination of positions can be
2 for instructions on removing the instru- used to configure different programming
ment cover). options, such as the programing option listed
above for instruments without the three-posi-
3. Locate the 3D3620/4D3620 processor tion external keyswitch. If your settings
board (smaller of two). should accidentally get changed, or your
replacement processor board is shipped from
4. Locate SW1 dip switch on the processor the factory with the incorrect settings for your
board. Locate switch position 3 (lower product, the following table details the appro-
switch) and select the CLOSED, ON, or 1 priate switch positions for restoring your
position. Do not change any other switch instrument to proper operation.
setting.
Option Switch Setup Table
5. Attach the Advanced Instruments serial
cable between your PC and the instru-
ment. SW1
1 2 3
6. Plug in and turn on the instrument. The 3D3 ON OFF OFF
display should read “Flash BOOT MK3 ON ON OFF
Program”. 4D3 OFF OFF OFF
MK2 OFF ON OFF
7. Perform the download, using the instruc- Program N/A N/A ON
tions provided with the update software. NOTE:
ON = 1 = CLOSED
OFF = 0 = OPEN
After download is complete, the instru-
ment will reboot and report the new soft-
ware version.
2. “A/D Init Failure” One or both A/D converters lost initial- 1. Turn power off and on.
“A/D High Filter Error” ization. 2. If problem persists, replace application
“A/D Low Filter Error” board.
“A/D Cal Mode Error”
3. “Baudrate Error” The baudrate setting is out of range. Reset baudrate in SETUP menu.
4. “Block Probe Failure” The block probe has failed (usually Check the block probe resistance (2,300
shorted). ohms ±300 at room temp.)
5. "Block Probe Open?" The block probe resistance is greater 1. Is the probe plugged in?
than A/D converter can measure. 2. Check the block probe resistance (2,300
ohms ±300 at room temp.)
6. “Calibration Cancelled” The STOP button was pressed during Calibration cancelled.
calibration.
7. “Calibration Out of The range of the samples tested did not 1. Recalibrate the instrument.
Range; Please Repeat” repeat well enough for calibration. 2. Check freeze pulse.
“Calibration Out of 3. Check sample size and quality.
Range; Repeat Calib” 4. Check probe/size alignment.
“Calibration Out of
Range; Repeat”
8. "Check Calibration" Buzz point setting has been changed. 1. Spot-check calibration.
(3D3 only) 2. Recalibrate, if needed.
9. “Cooling System The cooling assembly may not be func- 1. Check the cooling assembly.
Error” tioning properly. 2. Make sure there is heat transfer fluid in
the cooling well.
10. “Count Error; A memory check error occurred in the 1. Turn the power off and on.
Parameter RAM” RAM. 2. If problem persists, replace processor
board.
11. "Error Reading The data received from the barcode 1. Turn the power off and on.
Barcode” reader was not formatted correctly. 2. Check barcode scanner settings.
3. If problem persists, replace application
board.
12. “Event Record Lost” The instrument has lost its record from 1. Turn the power off and on.
memory. 2. If problem persists, replace processor
board.
13. “Fan Driver Failure” The fan is not functioning properly. 1. Check fan driver/control circuit.
“Fan System Failure” 2. Replace the fan or application board.
14. “Flash Boot Program” The instrument is configured to receive 1. If intentional, perform download process
Flash EPROM updated code. as described in this manual.
2. If not intentional, turn off power and set
three-position Operator/Supervisor key-
switch to Operator or Supervisor position.
3. Check processor board option switch
SW1 position 3 and set to OPEN/OFF/0
(see Option Switch Setup Table).
15. “Head Not Up” The head up/down mechanism may not 1. Check the motor, clutch assembly and
“Head Not Down” be functioning properly. relays by running the head up/down test.
2. Check the Head Up or Head Down posi-
tion sensor.
3. Check LED indicators on application
board.
16. “New Software EPROM has been changed, or new soft- For Hitachi systems:
Version” ware has been downloaded. 1. Reset the sample and block bin numbers.
2. Reset the stir amplitude.
3. Recalibrate the instrument.
4. Reset plateau mode/range.
5. Reset instrument serial number.
6. If problem persists, replace processor
board.
17. “No Parameters in The battery-backed RAM has lost its 1. Reset sample and block probe bin num-
RAM” data or has been cleared. bers. Recalibrate the instrument.
“Parameter RAM 2. Reset the stir amplitude.
Failed” The instrument has lost the calibration, 3. Reset the buzz point (3D3 only).
data, and probe bin numbers from mem- 4. Reset plateau mode/range.
ory. 5. Reset instrument serial number.
6. If problem persists, replace processor
board.
18. “No Plateau Detected, The sample did not reach a valid 1. Make sure the sample is not too viscous.
Please Repeat” plateau. 2. Check the sample probe.
“No Plateau, Repeat 3. Make sure there is heat transfer fluid in
Test” the cooling well.
19. “Overrange, Use High The sample tested in the low range set- Retest the sample in the high range setting.
Range” ting was over the low range limit.
“Result >2000, Use
High Range”
“Result >2500, Use
High Range”
(3D3 only)
20. “RAM Failure” The main board has failed. Replace the processor board.
“EPROM Failure”
21. “Recalibration The calibration data has been lost from- 1. Recalibrate the instrument.
Needed” memory. This is caused by: 2. If error message persists after successful
a. battery back-up failure, calibration, replace the processor board.
b. installation of a new processor
board,
c. corruption of system memory, or
d. new software/firmware.
22. “Reset Probe The probe bin data has been lost from 1. Reset sample and block probe bin num-
Configuration” memory. This is caused by: bers.
a. battery back-up failure, 2. If problem persists, replace processor
b. installation of a new processor board.
board,
c. corruption of system memory, or
d. new software/firmware.
23. “Result = 0, Bad A. The sample tested was de-ionized 1. Recalibrate the instrument.
Calibration?” H2O (this is not an error). 2. Check the sample/block probe bin set-
“Result = 0, tings.
Calibration Okay?” B. The instrument was not recalibrated
(Hitachi processor only) after the sample probe was replaced.
24. “Result >4000, Out of The sample tested was beyond the range Dilute the sample and retest.
Range” (3D3 only) of the instrument.
25. “Sample Did Not The sample did not freeze (after 1. Make sure the sample tube is not empty,
Freeze, Please Repeat” “buzzing”). and make sure the sample is not too vis-
“Sample Did Not cous.
Freeze, Repeat Test” 2. Check the freeze amplitude.
3. Make sure the instrument is in the proper
range (3D3 ony).
26. “Sample Freeze Error, The sample did not freeze properly. 1. The sample may have been contaminated.
Repeat” Retest with a fresh sample.
“Sample Freeze Error, 2. Check the freeze amplitude.
Repeat Test” 3. Check the sample probe.
27. “Sample Pre-Freeze, The sample has frozen prematurely 1. The sample may have been contaminated.
Please Repeat” (before the “buzz”). Retest with a fresh sample.
“Sample Pre-Freeze, 2. Check sample/block probe bin settings.
Repeat Test” 3. Check the stir amplitude.
4. Check for stir driver noise.
28. “Sample Probe The sample probe has failed (usually Check the sample probe resistance (2,300
Failure” shorted). ohms ±300 at room temp).
29. “Sample Probe Open?” The sample probe resistance is greater 1. Is the probe plugged in?
than 10558 ohms. 2. Check the sample probe resistance (2,300
ohms ±300 at room temp).
30. “Serial Loop Failed” RS-232 communications test failed. 1. Was loop back connector attached?
(Hitachi processor only) 2. Try lower baud rate.
3. If problem persists, consult factory serv-
ice.
31. “Standards Reversed? The instrument was expecting “high” Recalibrate the instrument with the calibrator
Please Repeat” calibrator standard and detected a value standards in the correct order.
“Standards Reversed? below, equal or close to the “low” cali-
Repeat Calib” brator standard.
32. “Stir/Freeze Failure” The stir/freeze mechanism may not be 1. Check the stir and freeze ampligudes.
“Stir System Failure” functioning properly. 2. Check stir driver/control circuits.
3. If problem persists, replace the processor
board.
33. “Supervisor Key Operator/Supervisor keyswitch is in the 1. Press the STOP key.
Needed, Press Stop” Operator position. 2. Set the keyswitch to the Supervisor posi-
tion.
34. “System Error: Trap” The software did not execute properly. 1. Turn the power off and on.
“System Error: NMI” 2. If problem persists, replace the processor
“System Error: ID” board.
“System Error: Unknown”
“System Error: Requests”
“System Error: Primitive”
“System Error: Interrupt”
“System Error:
Communications”
“Illegal Error Code:
“Unknown Error”
“Memory Allocation
Error”
“ROM Serial Number
Error”
35. “TE Driver Failure” The thermoelectric driver may not be 1. Use the A/D test to check the cooling
“Thermoelectrics functioning properly. assembly.
Driver Failure” 2. Check Q1B driver.
“Thermoelectrics 3. Check application board.
System Failure”
36. “Test Cancelled” The STOP button was pressed during Retest with a fresh sample.
the test.
37. “Test Time-out Error” The test has taken too long to complete. 1. Make sure the sample tube is not empty,
and make sure the sample is not too vis-
cous.
2. Make sure there is heat transfer fluid in
the cooling well.
3. Check the stir amplitude.
4. Check the sample/probe block bin setting.
5. Check the cooling assembly well.
6. Use A/D test to verify cooling system
function.
38. “Underrange, Use Low The sample tested in the high range set- Retest the sample in the low range setting.
Range” ting was under 1400 mOsm.
“Result <1400, Use
Low Range”
(3D3 only)
39. Wrong instrument Processor board 3D3620/4D3620 option Reset switch SW1 (see Option Switch Setup
model displayed switch SW1. Table).
Instruction:
C
4SP340 Rev3
(4D35 Service Manual)
Page 2 of 2
Reference: Use this instruction with replace- 6. Replace the old clapper with the new one
ment part 3C2241R. and reassemble by reversing steps 2-5.
Tools Needed: Phillips screwdriver, flat- 7. Use the probe alignment tool and instruc-
bladed screwdriver, 5/64” tions to properly position the probe and
Allen wrench, probe align- stir/freeze wire.
ment tool and instructions.
Instruction:
D
B
Reference: Use this instruction with replace- The following instructions have been pre-
ment part 3D3605R, 4D3605R, pared to cover several different replacement
PCB605R, 3D3620R, or scenarios. Please read carefully and select the
4D3620R. appropriate set of instructions for your situa-
tion.
Tools Needed: Phillips screwdriver, flat-
bladed screwdriver, static GENERAL INSTRUCTIONS:
grounding (earthing) wrist
strap (included). 1. Record current sample and block probe
bin numbers, sample stir amplitude, and
any other customized settings that may
Warning-Hazardous Voltage need to be reset after this procedure.
Warning-Internal components may 2. Turn off the power and unplug the instru-
be damaged by static electricity. ment.
B C
Two-Board Set
4D3P620 Rev3
(4D35 Service Manual)
Page 2 of 4
6. Use the SETUP menu to reset the sample 3. Cycle through the menu options by press-
and block probe bin numbers and the ing < or > until you reach “Serial
sample stir amplitude, and any other cus- Number”, and then press START. The
tomized settings. display will read “Serial#:********”.
Note: If values are unavailable, refer to the 4. Press the following keys consecutively:
User’s Guide for setup procedures. 2, 4, CLEAR. The displayed serial num-
ber will disappear.
7. Reset the serial number (see the following
procedure). 5. Enter each digit of the serial number as it
appears on the label on the back of the
8. Press STOP to exit the Setup menu. The instrument (Note: the letter suffix at the
instrument will run the self-diagnostics end of the serial number will not be
program. entered). When the number is correct,
press ENTER.
9. Recalibrate the instrument. For further
information, refer to your User’s Guide. The serial number will be recorded in
memory and should appear the next time
10. If you have upgraded from a single con- you power up your instrument.
trol board to the two-board set, see the
Supplemental Information document.
11. If you have upgraded from a single con- XXX620 Option Switch Setup
trol board to the two-board set and your
instrument has the black plastic keyswitch SW1
located on the rear of the instrument, your 1 2 3
Operator and Supervisor positions will 3D3 ON OFF OFF
have changed. Replacement labels have MK3 ON ON OFF
been provided along with this upgrade. 4D3 OFF OFF OFF
MK2 OFF ON OFF
Place the “OPERATOR” label along the Program N/A N/A ON
top of the keyswitch, and place the NOTE:
ON = 1 = CLOSED
“SUPERVISOR” label along the right OFF = 0 = OPEN
side.
4D3P620 Rev3
(4D35 Service Manual)
Page 4 of 4
Reference: Use this instruction with replace- b. Remove the two screws from the rear
ment part 3D3300 or 4D3300. panel of the back cover and lift off the
back cover.
Tools Needed: Phillips screwdriver, flat-
c. Remove the four screws from under-
bladed screwdriver.
neath the front cover, lift the front
cover slightly up and then slide it for-
ward and off.
Warning-Hazardous Voltage
3. Drain the system:
Warning-Internal components may
be damaged by static electricity. a. Remove the heat transfer fluid bottle
and empty the contents.
CAUTION: Improper connections
may cause damage to the instrument. b. Replace the empty bottle and insert
the tubes.
c. Press gently on the pump at the bot-
Instruction: tom of the head. Slide until all liquid
has been expelled (the unit may need
1. Turn off the power and unplug the instru- to be tipped on its side to remove the
ment. liquid).
The following instructions refer to 4D3 The following instructions refer to 4D3
cryoscopes with suffix "A" or later, and cryoscopes with suffix "A" or later, and
to all 3D3 osmometers: to all 3D3 osmometers:
a. Remove the two sets of screws from a. Remove the four screws from the
the side panels and the two screws deck (A), and remove the deck.
from the rear panel. b. Remove the two screws that secure
b. Slide the instrument cover up and off. the bottle bracket to the cooling
assembly (B).
The following instructions refer to 4D3 c. Remove the two screws that secure
cryoscopes without a suffix: the pump bracket to the cooling
a. Remove the heat transfer fluid from assembly (C).
the front cover. d. Disconnect the hose from the pump
mechanism (D).
C B
4D3P300 Rev1
(4D35 Service Manual)
Page 2 of 2
The following instructions refer to 4D3 The following instructions refer to 4D3
cryoscopes with suffix "A" or later, and cryoscopes with suffix "A" or later, and
to all 3D3 osmometers: to all 3D3 osmometers:
a. Remove the two sets of screws from a. Remove the four screws from the
the side panels and the two screws deck (A), and remove the deck.
from the rear panel. b. Remove the two screws that secure
b. Slide the instrument cover up and off. the bottle bracket to the cooling
assembly (B).
The following instructions refer to 4D3 c. Remove the two screws that secure
cryoscopes without a suffix: the pump bracket to the cooling
a. Remove the heat transfer fluid from assembly (C).
the front cover. d. Disconnect the hose from the pump
mechanism (D).
C B
4DP360 Rev0
(4D35 Service Manual)
Page 2 of 2
A
B B
4DP851R Rev1
(4D35 Service Manual)
Page 2 of 2
Reference: Use this instruction with replace- To determine if your instrument contains a
ment part 716811R. 256K EPROM main board, remove the cover
and look at the upper left corner of the main
Tools Needed: Phillips screwdriver, flat- board. If your board contains component Q9,
bladed screwdriver, static you still have the 256K EPROM main board
grounding (earthing) wrist and can use this upgrade.
strap (included), 28-pin IC
removal tool (alternatively, a Upon completion of this replacement, the
small screwdriver may be software version reported at power-up will
used with caution so as not to match the version number printed on the new
damage the IC or the socket). EPROM.
Warning-Hazardous Voltage
Instruction:
Warning-Internal components may
1. Turn off the power and unplug the instru-
be damaged by static electricity.
ment.
CAUTIONS:
2. Attach the static grounding strap to your
• Instrument calibration and all cus- wrist and stick the adhesive end to a suit-
tomized settings will be lost when able bare-metal ground.
the EPROM is replaced. Therefore,
make sure the current sample and 3. Remove the screws securing the instru-
block probe bin numbers and the ment cover, and remove the cover:
sample stir amplitude are recorded
in the service log so that you can a. Remove the two sets of screws from
reset these vital numbers into the the side panels and the two screws
replacement EPROM. from the rear panel.
• Never unpack, touch or handle an b. Slide the instrument cover up and off.
EPROM or any other integrated
circuit without wearing a ground- 4. Locate EPROM IC U8 on the circuit
ing (earthing) strap to minimize board. (U8 bears a white label indicating
your static charge. the software version number.) Note the
orientation of the small locating notch in
one corner, and/or locating mark on one
*EPROM Replacement 716811R may be used edge of the IC.
to update the operating software in all
Advanced Model 4D3 cryoscopes with serial 5. Make sure your static draining connection
numbers prior to suffix F that contain a 256K is intact, and then remove IC U8 from the
EPROM main board. circuit board.
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
71P811R Rev1
(4D35 Service Manual)
Page 2 of 2
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
71P824R Rev1
(4D35 Service Manual)
Page 2 of 2
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
71P825R Rev1
(4D35 Service Manual)
Page 2 of 2
1. Turn off the instrument power and discon- This option is the standard factory installation.
nect the power cord. These photographs have been provided as ref-
erence only, for determination of the installed
2. Remove the two screws holding the cover upgrade. Field upgrades should use option A,
on the sample head assembly. as described above.
Ferrite Beads
NOTE: For installation Option A, locate two ferrite beads on the white wire exiting the head shaft and
attached to TR2 on the head transitionn board.
Ferrite Bead
Ferrite Bead
4SP475 Rev0
(4D35 Service Manual)
Page 2 of 2
Reference: Use this instruction with replace- The following instructions refer to 4D3
ment part 70011 (1.0-Amp) or cryoscopes with suffix before "D", and to
70022 (2.0-Amp). 3D3 osmometers with suffix before “B”:
700P11 Rev0
(4D35 Service Manual)
Page 2 of 3
Reference: Use this instruction with replace- 2. Use a flat-bladed screwdriver to pry out
ment part 70011 (1.0-Amp) or the fuse holder door and then the fuse
70022 (2.0-Amp). holder module.
Tools Needed: Flat-bladed screwdriver. 4. Place the fuse holder back into the instru-
ment, with the correct voltage on the fuse
holder showing in the window (A) on the
Warning-Hazardous Voltage fuse holder socket.
700P11 Rev0
(4D35 Service Manual)
Page 3 of 3
Instruction:
Head Up Relay
1. Turn off the power and unplug the instru- A
ment.
Reference: Use this instruction with replace- cryoscopes use a closed-loop circulation sys-
ment parts 4D3690, 4D3710, tem for the heat transfer fluid in the cooling
4D3705, or 99133R. system. Various design modifications to the
Pump Assembly affect the part replacement
Tools Needed: Phillips screwdriver, flat- instructions. Please turn to the appropriate
bladed screwdriver. section listed in the table of contents, based
upon the part being replaced and the model
instrument that is being serviced.
Warning-Hazardous Voltage
Instruments shipped prior to 1996 have a
Warning-Internal components may cooling assembly inlet tube with a much
be damaged by static electricity. smaller diameter. A tubing adapter has been
provided in the 4D3690 kit to allow easy
CAUTION: Improper tubing connec- connection to this smaller tube.
tions may result in fluid leakage and
may cause damage to the instrument. Instruments shipped prior to 2004 use the
same type of one-way check valve (4D3705)
in two places. One is located next to the
General Information: Fluid Filter on the pump intake in the Heat
Transfer Fluid bottle. The other is located at
Advanced Instruments, Inc. Models 3D3, the outlet end of the Pump Assembly
4D3, 3250, and 4250, and Fiske Associates attached to the inlet of the Cooling Assembly.
Models Mark 2 and Mark 3 osmometers and Effective in 2004, the one-way check valve
Table of Contents
Replacement Part Model Page
Heat Transfer Fluid Pump Assembly (4D3690) 3D3, Mark 3, 4D3, Mark 2 2
Heat Transfer Fluid Pump Assembly (4D3690) 3250, 4250 4
Fluid Filter (4D3710) All 5
Bottle One-way Check Valve (4D3705) All 5
Cooling System Inlet One-way Check Valve All 6
(99133R)
Figure 1
attached to the Cooling Assembly inlet has 2. Remove the screws securing the instru-
been changed to a different, larger check ment cover, and remove the cover (see
valve (99133R) to improve instrument per- below):
formance and reliability.
The following instructions refer to 4D3
Figure 1 shows the current 4D3690 Pump cryoscopes without a suffix:
Assembly with the various components a. Remove the heat transfer fluid from
labeled. the front cover.
CAUTION: The heat transfer fluid b. Remove the two screws from the rear
contains hazardous chemicals. panel of the back cover and lift off the
Consult the material safety data sheet back cover.
(MSDS) and use appropriate personal c. Remove the four screws from under-
protective equipment. neath the front cover, lift the front
cover slightly up, and then slide it for-
ward and off.
4D3690 Heat Transfer Fluid Pump
Assembly Replacement The following instructions refer to 4D3
Models 3D3 and Mark 3 Osmometers cryoscopes with suffix "A" or later,
Models 4D3 and Mark 2 Cryoscopes and to all 3D3 osmometers:
a. Remove the two sets of screws from
Instruction: the side panels and the two screws
from the rear panel.
1. Turn off the power and unplug the instru-
ment. b. Slide the instrument cover up and off.
4DP690 Rev2
(4D35/3255/MK05 Service Manual)
Page 2 of 7
Figure 2 Figure 3
4DP690 Rev2
(4D35/3255/MK05 Service Manual)
Page 3 of 7
Instruction:
4DP690 Rev2
(4D35/3255/MK05 Service Manual)
Page 4 of 7
Instruction:
4. Replace the old filter with the new one, 3. Holding the tube in an elevated position
and then replace the tube into the heat to minimize leakage, firmly grip the tube
transfer fluid bottle. just above the old check valve, then pull
off the old check valve and filter.
Note: The instrument cover does not need to 5. Remove the old check valve from the
be removed to replace this part. short piece of tube that connects it to the
filter, then reattach the tube and filter to
Instruction: the new check valve. Replace the tube
with filter into the heat transfer fluid
1. Open the heat transfer fluid compartment bottle.
door.
4DP690 Rev2
(4D35/3255/MK05 Service Manual)
Page 5 of 7
Instruction:
1. Turn off the power and unplug the instru-
ment.
The following instructions refer to 5. Firmly grip the cooling assembly inlet
4250 cryoscopes and 3250 osmometers: tube and remove the check valve.
Discard the old check valve and cable
a. Remove the two sets of screws from clamp.
the side panels and the four screws
from the rear panel. 6. Install the new check valve, making sure
b. Lift and tilt the cover so that the rec- to orient the check valve properly (see
tangular opening clears the front of Figure 4). Connect the new check valve
the head cover first, then move the to the cooling system inlet tube, then fas-
instrument cover back, up and off. ten the new cable clamp using the hard-
ware saved in step 3.
3. If present, remove and save the screw On instruments shipped prior to 1996, the
holding the cable clamp (B or F) to the cooling assembly inlet tube was a much
4DP690 Rev2
(4D35/3255/MK05 Service Manual)
Page 6 of 7
4DP690 Rev2
(4D35/3255/MK05 Service Manual)
Page 7 of 7
Contents: Instruction:
1 4D3102 Non-Metallic Sample Probe Kit
1 716824R Model 4D3 Software Upgrade 1. Replace the sample probe according to
Kit the instructions contained in the 4D3102
Sample Probe Replacement kit.
Contents: Instruction:
1 4D3102 Non-Metallic Sample Probe Kit
1 4D3990 Model 4D3 Software Upgrade 1. Replace the sample probe according to
Kit: Disk and Cable the instructions contained in the 4D3102
Sample Probe Replacement kit.
Reference: Use this instruction to install part 2. Remove the instrument cover.
425150R.
3. Pull and drop the keypad and display for-
ward.
For Use On Models:
4D3 with Serial Suffix N or later 4. Locate two #2 holes with threaded inserts
3D3 with Serial Suffix L or later just to the left of the power supply termi-
MK2 with Serial Suffix G or later nal block and install the 425150 load
MK3 with Serial Suffix E or later resistor using provided #2x3/8 screws (A),
3250 -- All as shown in photo.
4250 -- All
5. Disconnect cooling assembly red and
Tools Needed: Small Phillips screwdriver. black wired connector from the M20150
harness.
Warning-Hazardous Voltage 6. Plug both connectors into the mating con-
nector of the load resistor assembly.
Warning-Internal components may
be damaged by static electricity. 7. Replace the instrument cover and restore
power.
8. Verify correct operation using the “A/D
Instruction: Test” in TEST utilities.
5. Carefully, push the new mandrel up past 11. Plug in the power cord and return the
the probe and into position on the flange. instrument to operation.
The following instructions refer to 4D3 6. Remove the two screws (B) that secure
cryoscopes without a suffix: the bracket to the deck.
a. Remove the heat transfer fluid from 7. Remove the two Allen (C) screws that
the front cover. secure the clutch to the motor shaft. If
4DP350 Rev0
(4D35 Service Manual)
Page 2 of 2
CAUTION: Unplug the power cord • If using PCB620, remove the processor
prior to opening or removing covers, board from the application board and
or else you may be exposed to electric locate U15 on the back side.
shock, excessive temperatures, or • If using 200016PC, locate U11.
mechanical hazards.
• If using 210102PC, locate U11.
Performing this service or maintenance
should only be done by a qualified 5. Once you have located the NVRAM chip,
service technician. locate the small notch and insert the small
screwdriver, as described in figure 1.
A discharge of static electricity from
contact with the human body or other
conductor may damage system boards
or static sensitive devices. Never per-
form internal maintenance without fol-
lowing recommended static protection
procedures.
Figure 1
71P27R Rev3
For additional information or technical assistance, please
(2025/2105/3255/3305/3325/
contact Advanced Instruments Hot-Line® Service Center.
(U.S. 1-800-225-4034, outside North America +US 1-781- 4D35/MK05 Service Manuals)
320-9000. After normal business hours, dial extension 2191.) Page 1 of 2
7. Remove the replacement powercap from 11. Turn off the power and wait ten minutes.
the anti-static bag and install it on the Turn the instrument back on and verify
component base, as described in figure 2. that the correct time has been maintained.
8. Reinstall the processor board, if removed. 12. Turn off the power and restore the instru-
ment cover.
9. Restore power to the instrument and veri-
fy correct power-up sequence without any 13. Reset all stored parameters (e.g., block
LOW BATTERY warnings. and sample bin numbers, plateau modes,
etc.) and recalibrate the instrument.
Figure 2
71P27R Rev3
(2025/2105/3255/3305/3325/
4D35/MK05 Service Manuals)
Page 2 of 2
*Not for use on Models 3250/4250 with 5. Remove the screws securing the protec-
serial suffix D and higher. tive black plastic cap on the outside rear
of the instrument.
Reference: Use this instruction to install part
6. Remove the old transistor and insulating
PCB520R, which is intended as a
materials by gently prying with a flat-
direct replacement for part 35010
bladed screwdriver.
metal can FET transistor used on
some of the above models. 7. Remove the transistor socket Q1B from
the rear panel and clear the mounting
Materials Included: holes of any remaining hardware.
1 - PCB520 assembly
2 - H50606, #6x3/8 Sems 8. Cut transistor socket Q1B from harness
2 - H50608, #6x1/2 Sems M20150.
2 - H63093, #6 large flat washers 9. Strip wire insulation ¼” on four (4) wires
2 - 90049, Hex swage standoffs formerly soldered to the transistor socket.
2 - 91038, ¼” plastic hole plugs Twist the stripped ends to prevent fraying.
Tools Needed: Wire cutters, wire strippers, 10. Install the hex standoffs to PCB520 using
small flat-bladed screwdriver, #6x3/8 screws, with the flat end of the
medium Phillips screwdriver. standoff against the board.
11. Connect white/red wire to terminal 1 (A).
Warning-Hazardous Voltage
12. Connect white/brown wire to terminal 2
Warning-Internal components may be (B).
damaged by static electricity. 13. Connect brown wire to terminal 3 (C).
14. Connect black wire to terminal 4 (D).
4D3, 3D3, 3250, and 4250 Instruction:
1. Turn off the instrument power and discon-
nect the power cord.
2. Remove the intrument cover. A
3. Locate transistor Q1B on left-rear of the
B
chassis. Q1B is the transistor nearest to
the top of the rear panel.
C
4. If your instrument already has PCB520 in
this location, remove the old PCB and D
skip to step 11.
PCP520R Rev4
(4D35/4C35/3255/3905 Service Manual)
Page 2 of 3
G
H
I
J
PCP520R Rev4
(4D35/4C35/3255/3905 Service Manual)
Page 3 of 3
*Not for use on Models 3250/4250 with 5. Remove the screws securing the protec-
serial suffix D and higher. tive black plastic cap on the outside rear
of the instrument.
Reference: Use this instruction to install
part PCB522R, which is intend- 6. Remove the old transistor and insulating
ed as a direct replacement for materials by gently prying with a flat-
part 35050 metal can Darlington bladed screwdriver.
transistor used on some of the 7. Remove the transistor socket Q2B from
above models. the rear panel and clear the mounting
holes of any remaining hardware.
Materials Included:
1 - PCB520 assembly 8. Cut transistor socket Q2B from harness
2 - H50606, #6x3/8 Sems M20150.
2 - H50608, #6x1/2 Sems 9. Strip wire insulation ¼” on four (4) wires
2 - H63093, #6 large flat washers formerly soldered to the transistor socket.
2 - 90049, Hex swage standoffs Twist the stripped ends to prevent fraying.
2 - 91038, ¼” plastic hole plugs
10. Install the hex standoffs to PCB522 using
Tools Needed: Wire cutters, wire strippers, #6x3/8 screws, with the flat end of the
small flat-bladed screwdriver, standoff against the board.
medium Phillips screwdriver.
11. Connect white/blue wire to terminal 1
(A).
Warning-Hazardous Voltage
12. Connect violet wire to terminal 2 (B).
Warning-Internal components may
13. Terminal 3 (C) is unused.
be damaged by static electricity.
19. Turn on the instrument and verify correct 15. Connect blue/red wire to terminal 4 (J).
stir/freeze operation using the test utility 16. Connect blue wire to terminal 4 (J).
“Stir/Freeze Test” (see user’s guide).
17. Install completed assembly on the back
4C3 and 3900 Instruction: side of the heat sink where the previous
transistor socket was mounted. Use
1. Turn off the instrument power and dis- #6x1/2 screws and large flat washers to
connect the power cord. secure swaged end of standoffs to the
2. Remove the intrument rear cover. heat sink, using the old transistor mount-
ing holes. The ¼” plastic hole plugs are
3. Locate transistor Q2 on the transistor heat not used in this installation and may be
sink mounted to the bottom of the chas- discarded.
sis. Q2 is the transistor 2nd from the
right. 18. Replace the instrument cover and connect
the power cord.
4. If your instrument already has PCB522 in
this location, remove the old PCB and 19. Turn on the instrument and verify correct
skip to step 12. stir/freeze operation using the test utility
“Stir/Freeze Test” (see user’s guide).
5. Remove the screws securing the heat sink
to the chassis and gently lift out the heat
sink.
6. Remove the screws holding the old tran-
sistor and insulating materials.
PCP522R Rev3
(4D35/4C35/3255/3905 Service Manual)
Page 2 of 3
G
H
I
J
PCP522R Rev3
(4D35/4C35/3255/3905 Service Manual)
Page 3 of 3
red
Warning-Hazardous Voltage
black
Warning-Internal components may black
be damaged by static electricity.
violet
CAUTION: Improper connections yellow
may cause damage to the instrument.
orange
green/yellow
4D3950 Instruction:
blue/white
1. Turn off the power and unplug the instru- brown/white
ment.
A B E
325P950 Rev2
(4D35/3255/MK05 Service Manual)
Page 2 of 3
325P950 Rev2
(4D35/3255/MK05 Service Manual)
Page 3 of 3
1. Turn off the power and unplug the instru- c. Remove the four screws from under-
ment. neath the front cover, lift the front
cover slightly up and then slide it for-
2. Remove the screws securing the instru- ward and off.
ment cover, and remove the cover (see
below):
red
black
black
violet
yellow
orange
green/yellow
blue/white
brown/white
4DP950 Rev2
(4D35 Service Manual)
Page 2 of 2
Head
Mandrel Setscrews
Mandrel
Sample Probe
Stir Wire
Yoke
3LP700 Rev10
(4C35/4D35/3255/MK05 Service Manual)
Page 2 of 5
Raised V
Centerline This portion of the stir wire should
be parallel to sample probe, and
the space should be about .04
Alignment inches (1 mm).
Tool
Stir Wire
Sample Probe
Bottom of sample probe End of stir wire should
should rest on this level. rest on this small step.
gently retighten the probe setscrew 7. If necessary, make stir wire adjustments,
(overtightening will crack or distort as follows:
the thin plastic wall of the mandrel).
a. To realign the stir wire in the slot in
the mandrel, note the position of the
5. After completing any adjustments,
top of the yoke relative to the top of
recheck the probe alignment and level.
the head, then partially loosen the coil
core/yoke screw (see Figure 4) from
6. Check the stir wire alignment and level
the back of the head. Move the yoke/
(see Figure 3 notes). In the front-to-back
clapper/stir wire assembly side-to-side
direction, the stir wire should be even
until the stir wire is in the correct
with the sample probe.
position relative to the slot. Make
sure to maintain the previously noted
2. Check the probe alignment by making 6. Check the stir wire alignment and level
sure that the probe shaft is lined up along (see Figure 5 notes). In the front-to-back
the raised V centerline of the alignment direction, the stir wire should be even
tool and the probe level by making sure with the sample probe.
that the end of the probe rests on the hori-
zontal surface below the raised V center- 7. If necessary, make stir wire adjustments:
line, as shown in Figure 5. Note that the
a. To realign the stir wire in the slot in
orientation of the alignment tool is 90º
the mandrel, note the position of the
from the way the tool is used with an
top of the yoke relative to the top of
osmometer.
the head, then partially loosen the coil
core/yoke screw (see Figure 4) from
3. If necessary, make probe centering adjust-
the back of the head. Move the
ments:
yoke/clapper/stir wire assembly side-
a. If aligning a metal probe, bend the to-side until the stir wire is in the cor-
metal probe stem gently by hand, as rect position relative to the slot.
necessary, to align the probe along the Make sure to maintain the previously
raised V centerline of the probe align- noted yoke vertical position relative
ment tool. to the head. Retighten the screw,
making sure that the yoke/clapper/stir
b. If aligning a non-metallic probe, DO
wire assembly does not rotate in
NOT ATTEMPT TO BEND THE
place.
PROBE; instead, loosen the probe
setscrew (see Figure 1) and then b. To adjust the vertical position of the
rotate the probe so that any curvature stir wire, loosen the two stir wire
3LP700 Rev10
(4C35/4D35/3255/MK05 Service Manual)
Page 4 of 5
Raised V
Stir wire should be parallel to
Centerline sample probe, and the space
should be about .06 inches
(1.5 mm).
Alignment
Tool
Stir Wire
Sample Probe
Bottom of sample
probe and end of
This level is not used with stir wire should rest
cryoscopes. on this level.
3LP700 Rev10
(4C35/4D35/3255/MK05 Service Manual)
Page 5 of 5
Reference: Use this instruction with replace- leads as close to the terminal strip as pos-
ment part 4D3700. sible.
Tools Needed: Phillips screwdriver; flat- 5. Loosen the probe setscrew (C) and gently
bladed screwdriver; soldering pull the probe down and out of the man-
iron and SN62, 2% silver, drel, guiding the leads through the center
eutectic, rosin-core solder; hole so that they do not catch.
probe alignment tool and
instructions (included). B
Instruction:
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
4DP700 Rev2
(4D35 Service Manual)
Page 2 of 2
Reference: Use this instruction with replace- 2. Turn off the power and unplug the instru-
ment part 3D3700 (stainless ment.
steel) or 4D3102 (non-metallic),
and any of the following models 3. Remove the two screws from the head
of Advanced Instruments, Inc. or cover (A) and lift off the head cover.
Fiske Associates® osmometer or
cryoscope: 4. Unplug the sample probe connector from
Advanced Instruments Inc. the head transition board (B).
4250 Cryoscopes (Serial Suffix E - W)
4D3 Cryoscopes (Serial Suffix E - W) 5. Loosen the two mandrel setscrews (C).
3250 Osmometers (Serial Suffix C - T)
3D3 Osmometers (Serial Suffix C - T) B
Fiske Associates
Mark 2 Cryoscopes C
Mark 3 Osmometers
Instruction:
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
3DP700 Rev2
(3255/4D35/MK05 Service Manual)
Page 2 of 2
Instruction:
12. Arrange the probe leads so that they will 7. At “Select Setup Item”, press < or > to
not touch the clapper or be pinched by the select “Select Sample Bin #”, and then
head cover, and replace the cover. press START.
13. Plug in the power cord and turn on the 8. If the current sample bin number is not
instrument. the same as the one recorded in step 5,
enter the correct number, press ENTER,
14. Set the sample bin number (see the fol- and then press STOP twice to exit the
lowing procedure). menu.
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
4D3P02 Rev1
replaced by 3DP700
(4D35/3255 Service Manual)
Page 2 of 2
A new, composite, stainless steel and plastic C) Why has the design been changed?
body cryoscope sample probe has been
Answer: The stainless steel probe, Type
released for use in the Advanced Instruments,
A, exceeded the thermal conductance lim-
Inc. single and multi-sample cryoscopes
its specified in ISO 5764, therefore, the
(Models 4D3, 4C3 and 4250).
plastic probe, Type B, was developed to
meet these new requirements. Type B,
A) What is the difference between the three
although compliant with the new ISO
versions of sample probe?
5764 standard, proved to be difficult to
Answer: Only the body material is differ- manufacture, which has led to the com-
ent. Type A used a brazed stainless steel posite design probe, Type C. This design
body; Type B used a one-piece plastic uses the strength of stainless steel up in
body; and Type C uses a composite stain- the mandrel, yet has the lower thermal
less steel and plastic body. All three use conductance of plastic where the probe is
the same high accuracy glass thermistor immersed in the sample.
as the sensing element.
D) My cryoscope has been operating without
B) Is this new probe used in the osmome- any problems. Do I need to change the
ters? probe? Will I see any performance
change?
Answer: No, this change is only for
cryoscopes and does not affect the all- Answer: If your cryoscope needs to be
stainless steel body sample probe used in compliant to the ISO 5764 standard and
the osmometers. you are currently using a Type A probe,
4DP002 Rev1
(4D3/4C3/3255 Service Manual)
Page 2 of 2
3DP404 Rev0
(4D35 Service Manual)
Page 2 of 2
Reference: Use this instruction with replace- 6. Use the probe alignment tool and instruc-
ment part 3LH243 or 4LH243. tions to properly position the stir/freeze
wire.
Tools Needed: Phillips screwdriver, flat-
bladed screwdriver, 5/64”
Allen wrench, probe align-
ment tool and instructions.
B
CAUTION: Improper connections may
cause damage to the instrument.
Instruction:
Reference: Use this instruction to install part 6. Remove the transistor socket Q1B from
PCB520, which is intended as a the rear panel and clear the mounting
direct replacement for part 35010 holes of any remaining hardware.
metal can FET transistor used on
7. Cut transistor socket Q1B from harness
the above models.
M20150.
Materials Included: 8. Strip wire insulation ¼” on four (4) wires
1 - PCB520 assembly formerly soldered to the transistor socket.
2 - H50606, #6x3/8 Sems Twist the stripped ends to prevent fraying.
2 - H50608, #6x1/2 Sems
2 - H63093, #6 large flat washers 9. Install the hex standoffs to PCB520 using
2 - 90049, 5/16 Hex swage standoffs #6x3/8 screws, with the flat end of the
2 - 91038, ¼” plastic hole plugs standoff against the board.
10. Connect white/red wire to terminal 1 (A).
Tools Needed: Wire cutters, wire strippers,
small flat-bladed screwdriver, 11. Connect white/brown wire to terminal 2
medium Phillips screwdriver. (B).
12. Connect brown wire to terminal 3 (C).
Warning-Hazardous Voltage
13. Connect black wire to terminal 4 (D).
Warning-Internal components may 14. Install completed assembly inside the unit,
be damaged by static electricity. against the rear wall with the terminal
PCP520R Rev0
(4D35/4C35/3255/3905 Service Manual)
Page 2 of 3
PCP520R Rev0
(4D35/4C35/3255/3905 Service Manual)
Page 3 of 3
Osmometer Instruction:
3. Using a pair of Tru-arc pliers, remove the Remove the spring (D) from inside the
Tru-arc retainer ring (B) from inside the well and replace it with the new spring,
well. wide-end down.
Instruction:
C D
4MP540 Rev3
(4D35SM)
Page 2 of 7
4MP540 Rev3
(4D35SM)
Page 3 of 7
The printer can be tested and setup using the Follow these steps to access the Setup Menu:
steps described in this section. However,
please note that the printer has been setup to 1. Unplug the printer at the outlet or at the
work with your instrument, as shipped. No back of the printer.
setup changes are necessary for correct opera-
tion. 2. Press and hold the left side of the rocker
switch as you plug in the printer. The
Before attempting to change your printer set- printer will advance the paper.
tings, be sure to disconnect the data cable
from the instrument. 3. After the paper advance has stopped,
count 3-5 seconds and release the switch.
The following is printed:
Printer Test
Follow these steps to test the printer: *** SETUP MENU ***
CONFIGURE... [NEXT/OK]
4MP540 Rev3
(4D35SM)
Page 4 of 7
Invert
Use this parameter to change the orientation
of text in your printouts. Choose YES if you
want inverted (upside down) text, or NO if
you want non-inverted (right-side up) text.
Font
Choose from a 5 x 5 or a 5 x 8 dot matrix
print pattern. The 5 x 5 dot matrix pattern
produces only uppercase (capital) letters.
4MP540 Rev3
(4D35SM)
Page 5 of 7
_Underscore
Choose which symbol the same ASCII code
will print -- an _underscore, or a left arrow. RESET SEQ#...
4MP540 Rev3
(4D35SM)
Page 6 of 7
4MP540 Rev3
(4D35SM)
Page 7 of 7
Test Calibrator
Start Content
Stop Control
Calibration Flammable
RS232 Sterile
D R
Replacement Parts . . . . . . . . . . . . . . . . . . .32
Design Changes . . . . . . . . . . . . . . . . . . . . .30
Display/Printer Test . . . . . . . . . . . . . . . . . .53
S
F Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Sample Bin Setting . . . . . . . . . . . . . . . . . . .51
Freeze Amplitude . . . . . . . . . . . . . . . . . . . .46
Serial Number . . . . . . . . . . . . . . . . . . . . . .54
Service Assistance . . . . . . . . . . . . . . . . . . . .9
G Software Update . . . . . . . . . . . . . . . . . . . . .55
General Description and Purpose . . . . . . . . .9 Specifications . . . . . . . . . . . . . . . . . . . . . . . .7
General Overview . . . . . . . . . . . . . . . . . . .18 Stir Amplitude . . . . . . . . . . . . . . . . . . . . . .52
Stir Driver Noise . . . . . . . . . . . . . . . . . . . .47
I Suffix History . . . . . . . . . . . . . . . . . . . . . . .30
Instrument Software Update . . . . . . . . . . . .55 Supplies & Accessories . . . . . . . . . . . . . . .32
K T
Keypad Test . . . . . . . . . . . . . . . . . . . . . . . .53 Troubleshooting Table . . . . . . . . . . . . . . . .29
Key/Beeper Test . . . . . . . . . . . . . . . . . . . . .27