KEMA99ATEX1157
KEMA99ATEX1157
TI328F/00/en
Liquiphant M FTL50(H), FTL51(H)
Table of contents
2 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Medium conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Thermal shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Medium pressure pe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Test pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
State of aggregation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Solids content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 25
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Dimensions (in mm; 1 mm = 3.94 in) . . . . . . . . . . . . . . . . . . . . 26
Weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Electronic inserts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Compact housings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . 37
Product structure Liquiphant M
FTL50, FTL51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Product structure Liquiphant M
FTL50H, FTL51H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Welding neck G ¾ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Welding neck G 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Welding neck G 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Welding neck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
DRD welding flange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Lap joint flange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Lap joint flanges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
3 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Application
Level limit detection Maximum or minimum detection in tanks or pipes containing all kinds of liquids, including use in hazardous
areas, food and pharmaceuticals
L00-FTL5xxxx-11-05-xx-xx-000
L00-FTL5xxxx-15-05-xx-xx-000
Level sensor
FEL55/56/57/58/50A
Liquiphant M FTL with electronic versions
FEL55, FEL56, FEL57, FEL58
Ex i
for connecting to a separate switching unit or
or an isolating amplifier FEL50A Switching unit
for connecting to PROFIBUS PA segment PLC
Isolating
amplifier
Segment
coupler
…
EX EX
L00-FTL5xxxx-15-05-xx-en-000
4 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Design FTL50:
Compact
FTL51:
With extension pipe
FTL50H:
Compact, with polished tuning fork and hygienic process connections
FTL51H:
With extension pipe, polished tuning fork and hygienic process connections
Input
Measured variable Level (limit value)
Product density Adjustment on the electronic insert > 0.5 g/cm3 or > 0.7 g/cm3 (other on request)
Endress+Hauser 5
Liquiphant M FTL50(H), FTL51(H)
IL
1 2
IL = load current Max.
(switched through)
< 3.8 mA
1 2
< 3.8 mA = residual current
(blocked)
IL
1 2
= lit
Min.
Signal on alarm Output signal on power failure or in the event of damaged sensor: < 3.8 mA
Connectable load • For relays with a minimum holding power/rated power > 2.5 VA at 253 V AC (10 mA) or
> 0.5 VA at 24 V AC (20 mA)
• Relays with a lower holding power/rated power can be operated by means of an RC module connected
inparallel
• For relays with a maximum holding power/rated power < 89 VA at 253 V AC or < 8.4 VA at 24 V AC
• Voltage drop across FEL51 max. 12 V
• Residual current with blocked thyristor max. 3.8 mA
• Load switched directly into the power supply circuit via the thyristor.
Transient (40 ms) max. 1.5 A, max. 375 VA at 253 V or max. 36 VA at 24 V (not short-circuit proof)
6 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
IL
1 3
IL = load current Max.
(switched through)
< 3.8 mA
1 3
< 3.8 mA = residual current
(blocked)
IL
1 2
= lit
Min.
Signal on alarm Output signal on power failure or in the event of damaged sensor: < 3.8 mA
Connectable load • For relays with a minimum holding power/rated power > 2.5 VA at 253 V AC (10 mA) or
> 0.5 VA at 24 V AC (20 mA)
• Relays with a lower holding power/rated power can be operated by means of an RC module connected
inparallel
• For relays with a maximum holding power/rated power < 89 VA at 253 V AC or < 8.4 VA at 24 V AC
• Voltage drop across FEL51 max. 12 V
• Residual current with blocked thyristor max. 3.8 mA
• Load switched directly into the power supply circuit via the thyristor.
Transient (40 ms) max. 1.5 A, max. 375 VA at 253 V or max. 36 VA at 24 V (not short-circuit proof)
Endress+Hauser 7
Liquiphant M FTL50(H), FTL51(H)
0.5 A
e.g.
Relay
PLC
–
L+ L–
U–
... 10…55 V DC
L00-FTL5xxxx-04-05-xx-en-001
L+ IL +
1 3
IL = load current Max.
(switched through)
< 100 µA
1 3
< 100 µA = residual current
(blocked)
L+ IL +
1 3
= lit Min.
Signal on alarm Output signal on power failure or in the event of damaged sensor: < 100 µA
Connectable load • Load switched via the transistor and separate PNP connection, max. 55 V DC
• Load current max. 350 mA (pulsed overload and short-circuit protection)
• Residual current < 100 µA (with transistor blocked)
• Capacitive load max. 0.5 µF at 55 V, max. 1.0 µF at 24 V
• Residual voltage < 3 V (with transistor switched through)
8 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
R R
FTL5#(H)- #######N3#
(M12x1)
52010285 / 52024216 0.5 A 0.5 A
L– L+ L– L+
FTL5#(H)- #######D3#
(Pg11) 1 1
or
2 2
FTL5#(H)- #######E3#
(NPT ½") 3 + 3
+
R
– R
–
0.5 A 0.5 A
PE L– L+ PE L– L+
(Ground) (Ground)
3 + 3
(BN) (BN) +
R
– R
–
0.5 A 0.5 A
PE L– L+ PE L– L+
(Ground) (Ground)
L00-FTL5xxxx-04-05-xx-en-010
L+ IL +
3 2
IL = load current Max.
(switched through)
L+ < 100 µA +
3 2
< 100 µA = residual current
(blocked) L+ +
IL
2 3
= lit
Min.
Endress+Hauser 9
Liquiphant M FTL50(H), FTL51(H)
L00-FTL5xxxx-16-05-
xx-xx-002
L+ IL –
1 2
Max.
IL = load current
L+ < 100 µA –
(switched through) 1 2
L00-FTL2xxxx-07-05-
xx-xx-000
ye 2 gn
Safety mode Level Output signal LEDs
L00-FTL5xxxx-16-05-
xx-xx-001
L+ IL –
1 2
Max.
IL = load current
L+ < 100 µA –
(switched through) 1 2
L00-FTL2xxxx-07-05-
xx-xx-000
Signal on alarm Output signal on power failure or in the event of damaged sensor: < 100 µA
Connectable load • Load switched via the transistor and separate PNP connection, max. 55 V DC
• Load current max. 350 mA (pulsed overload and short-circuit protection)
• Residual current < 100 µA (with transistor blocked)
• Capacitive load max. 0.5 µF at 55 V, max. 1.0 µF at 24 V
• Residual voltage < 3 V (with transistor switched through)
10 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Output:
When connecting an instrument with 1 2 3 4 56 7 8
high inductance, provide a spark arrester
to protect the relay contact.
A fine-wire fuse (depending on the
load connected) protects the relay
contact on short-circuiting.
Both relay contacts switch simultaneously.
*
0.5 A
* When jumpered, the relay
output works with NPN logic.
NO C NC NO C NC
a u r a u r
L1 N PE
(Ground)
** **
L+ L–
U~ 19…253 V AC, 50/60 Hz
U–
... 19… 55 V DC
L00-FTL5xxxx-04-05-xx-xx-002
3 4 5 6 7 8
Max.
= relay energised 3 4 5 6 7 8
= relay de-energised
3 4 5 6 7 8
= lit
Min.
= unlit
3 4 5 6 7 8
L00-FTL2xxxx-07-05-
xx-xx-001 L00-FTL5xxxx-04-05-xx-xx-005
Signal on alarm Output signal on power failure or in the event of damaged sensor: relay de-energised
Endress+Hauser 11
Liquiphant M FTL50(H), FTL51(H)
EEx ia
EX
EX
– +
U–
... 11…36
11…36 V DC
e.g. PLC
L00-FTL5xxxx-04-05-xx-en-000
+ ~16 mA
2 1
Max.
+ ~8 mA
~ 16 mA = 16 mA ± 5 % 2 1
~ 8 mA = 8 mA ± 6 %
+ ~16 mA
2 1
= lit
Min.
= unlit + ~8 mA
2 1
L00-FTL2xxxx-07-05-
L00-FTL5xxxx-04-05-xx-xx-006
xx-xx-000
Signal on alarm Output signal on power failure or in the event of damaged sensor: < 3.6 mA
12 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
I
EX
L
– +
Isolating amplifier
to IEC 60947-5-6
(NAMUR)
L00-FTL5xxxx-04-05-xx-en-004
0.6 …
+ 1.0 mA
2 1
Max.
2.2 …
+ 2.8 mA
2 1
= lit 0.6 …
+ 1.0 mA
2 1
= flashes
Min.
2.2 …
= unlit + 2.8 mA
2 1
L00-FTL5xxxx-07-05-
L00-FTL5xxxx-04-05-xx-xx-003
xx-xx-002
Signal on alarm Output signal in the event of damaged sensor: > 2.2 mA
Connectable load • See Technical Data of isolating amplifier connected according to IEC 60947-5-6 (NAMUR)
Endress+Hauser 13
Liquiphant M FTL50(H), FTL51(H)
! Note! I
EX
For Ex-d applications, the additional function L
can only be used if the housing is not exposed
to an explosive atmosphere.
Connecting to multiplexer: – +
Set clock time to min. 2 s. Isolating amplifier
to IEC 60947-5-6
(NAMUR)
L00-FTL5xxxx-04-05-xx-en-002
= lit 2.2 …
+ 3.5 mA
2 1
= flashes
Min.
0.6 …
= unlit + 1.0 mA
2 1
L00-FTL5xxxx-07-05- L00-FTL5xxxx-04-05-xx-xx-007
xx-xx-002
Signal on alarm Output signal in the event of damaged sensor: < 1.0 mA
Connectable load • See Technical Data of isolating amplifier connected according to IEC 60947-5-6 (NAMUR)
• Connection also to isolating amplifiers which have special safety circuits (I > 3.0 mA)
14 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
(H-L edge) – + – +
Connecting to multiplexer:
Set clock time to min. 3 s. + – + –
FTL5#(H)- #######C3#
The NAMUR interface has 1 1 2
(BU) (BU) (BK)
a defined power consumption.
Thus, it is not possible to use
the M12 connector with 3
integrated LED (52018763). (BN)
+ – + –
L00-FTL5xxxx-04-05-xx-en-009
= lit 2.2…
+ 3.5 mA –
1 2
= flashes Min.
0.6…
= unlit + 1.0 mA –
1 2
L00-FTL5xxxx-07-05-
L00-FTL5xxxx-04-05-xx-xx-007a
xx-xx-002
Signal on alarm Output signal in the event of damaged sensor: < 1.0 mA
Connectable load • See Technical Data of isolating amplifier connected according to IEC 60947-5-6 (NAMUR)
• Connection also to isolating amplifiers which have special safety circuits (I > 3.0 mA)
Endress+Hauser 15
Liquiphant M FTL50(H), FTL51(H)
16 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
L00-FTL5xxxx-05-05-xx-en-000
Please note this switching response and function of the plant, especially when replacing a Liquiphant with an
EL17Z or FEL37 electronic insert by a Liquiphant M with the FEL57 electronic insert.
150 Hz
= lit
= unlit 50 Hz
L00-FTL2xxxx-07-05-
xx-xx-000
L00-FTL5xxxx-04-05-xx-xx-008
Signal on alarm Output signal on power failure or in the event of damaged sensor: 0 Hz
Connectable load • Floating relay contacts in the connected switching unit Nivotester FTL320, FTL325P, FTL370, FTL372,
FTL375P or Commutec SIF101, SIF111
• For contact load see the Technical Data of the switching unit
Endress+Hauser 17
Liquiphant M FTL50(H), FTL51(H)
Additional functions:
– Digital communication enables the
representation, reading and editing
1 2
of the following parameters:
Fork frequency, switch-on frequency,
switch-off frequency, switch-on time and
switch-off time, status, measured value,
density switch.
– Matrix locking possible
– Switch to WHG mode possible EX
(WHG approval).
– For a detailed description see BA198F
– You can also visit
www.profibus.com for more information EX
PA– PA+
–
–
PROFIBUS PA
U–
... 9…32 V DC
–
Segment coupler
–
L00-FTL5xxxx-04-05-xx-en-005
PC with Commuwin II
Proficard or Profiboard
PLC
Segment coupler
PROFIBUS DP
EX
PROFIBUS PA
EX
L00-FTL5xxxx-04-05-xx-en-006
18 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
OUT_D = 1
PA bus signal
= lit inverted
OUT_D = 0
= unlit PA bus signal
L00-FTL2xxxx-07-05-
xx-xx-000 L00-FTL5xxxx-04-05-xx-xx-009
Signal on alarm • Failure information can be opened using the following interfaces:
Yellow LED flashing, status code, diagnostic code; see BA198F
Endress+Hauser 19
Liquiphant M FTL50(H), FTL51(H)
Switch-on behaviour When switching on the power supply, the output assumes the alarm signal.
After max. 3 s it assumes the correct switching mode (Exception: FEL57)
Performance characteristics
Reference operating Ambient temperature: 23 °C
conditions
Medium temperature: 23 °C
Product density: 1 g/cm3 (water) Switchpoint
13 mm
2 for
Viscosity: 1 mm /s reference conditions
Repeatability 0.1 mm
Hysteresis Approx. 2 mm
20 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Operating conditions
Installation Installation instructions
Switch points on the sensor depend on the mounting position, with reference to water,
density 1 g/cm3, 23 °C, pe 0 bar.
4 mm
4 mm
13 mm
L00-FTL5xxxx-06-05-xx-xx-001
Mounting from above Mounting from below Mounting from the side
! Note!
The switchpoints of the Liquiphant M are at other positions to those of the previous version Liquiphant II.
Examples of mounting Examples of mounting with regard to the viscosity ν of the liquid and the amount of build-up
L00-FTL5xxxx-11-05-xx-xx-001
Endress+Hauser 21
Liquiphant M FTL50(H), FTL51(H)
*
*
L00-FTL5xxxx-11-05-xx-xx-002
Vertical from above Protruding into the tank from the side
*
* min. 25
(min. 1 in)
L00-FTL5xxxx-11-05-xx-en-003
L00-FTL5xxxx-11-05-xx-xx-004
22 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
L00-FTL5xxxx-11-05-xx-xx-005
.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..
Ensure adequate space outside the tank .................................................................................................................................
.. .. .. .. .. .. .. .. .. .. .. .. .
L00-FTL5xxxx-11-05-xx-xx-006
Orientation FTL50(H) and FTL51(H) with short pipe (up to approx. 500 mm) – any position,
FTL51(H) with long pipe – vertical
Environment
Ambient temperature range Permitted ambient temperature Ta at the housing depending on the medium temperature Tp in the tank:
Ta
Ta *
70 °C
50 °C
0 °C
Tp
–50 °C 0 °C 50 °C 100 °C 150 °C Tp
90 °C
–50 °C
L00-FTL5xxxx-05-05-xx-xx-001
* Additional temperature range for devices with a temperature spacer or flameproof bushing
Endress+Hauser 23
Liquiphant M FTL50(H), FTL51(H)
Vibration resistance To IEC 68, Part 2-6 (10…55 Hz, 0.15 mm, 100 cycles)
Medium conditions
Medium temperature range –50…+150 °C; for exceptions, see "Process connections"
Medium pressure pe
pe
bar
(psi)
(psi)
*
100
(1450)
64
(928)
Tp
–1 °C
–50 (–14.5) 0 (°F)
150
(–58) (32) (300)
L00-FTL5xxxx-05-05-xx-xx-003
* Allowed pressure rating by selecting the "100 bar" option (see "Product structure" code 060)
Exceptions see "Process connections"
24 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Mechanical construction
Design Summary of all electrical and mechanical versions
Housing
Process connections
G ¾, DIN ISO 228/I G 1, DIN ISO 228/I Diverse Flanges to DIN, ANSI, JIS
R ¾, DIN 2999 R 1, DIN 2999 hygienic and aseptic from DN 25 / 1"
NPT ¾, ANSI B 1.20.1 NPT 1, ANSI B 1.20.1 connections
(AF 32) (AF 41)
Endress+Hauser 25
Liquiphant M FTL50(H), FTL51(H)
L00-FTL5xxxx-03-05-xx-xx-018
Bushings
Temperature spacer and flameproof bushing
L00-FTL5xxxx-03-05-xx-xx-005
141
133
206
199
32 AF 32 AF 32 AF
1. 5 m cable
*
2. M12 connector
3. Pg11/NPT ½ connector
L00-FTL5xxxx-06-05-xx-en-008
max. 40.7
*
17.5
L00-FTL5xxxx-06-05-xx-xx-004
26 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
max. 150
~25
*
L00-FTL5xxxx-06-05-xx-xx-005
max. 173
ø21.5
*
10
L00-FTL5xxxx-06-05-xx-xx-006
max. 190
*
L00-FTL5xxxx-06-05-xx-xx-007
! Note!
The switchpoints of the Liquiphant M are at other positions to those of the previous version Liquiphant II.
Flameproof bushing
Protects the housing from pressures
up to 100 bar if the sensor is damaged. Vessel
insulation
Provides sealed insulation
for the vessel and normal ambient additional length
140 mm
temperatures for the housing.
L00-FTL5xxxx-11-05-xx-en-000
Endress+Hauser 27
Liquiphant M FTL50(H), FTL51(H)
G¾ GQ2 66.5
max. 100 bar
DIN ISO 228/l GQ5 max. 150 °C
with defined 32 AF 50.5
thread start
With elastomer
flat seal to DIN 7603:
supplied L00-FTL5xxxx-06-05-xx-en-001
G¾ GQ2 66.5
Welding neck max. 25 bar
DIN ISO 228/l GQ5 (with defined max. 150 °C
with defined 32 AF 50.5 thread start)
thread start with silicone O-ring max. 40 bar
Endress+Hauser max. 100 °C
For flush-mounted 52001052
installation in
L00-FTL5xxxx-06-05-xx-en-001 In conformity with
welding neck
FDA*
See "Accessories"
With elastomer
flat seal to DIN 7603:
supplied
L00-FTL5xxxx-06-05-xx-en-002
or
R¾ GE2
DIN 2999 GE5
L00-FTL5xxxx-06-05-xx-en-004
or
R1 GF2
DIN 2999 GF5
L00-FTL5xxxx-06-05-xx-en-005
L00-FTL5xxxx-06-05-xx-xx-008
28 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Tri-Clamp 66.5
Clamping ring and max. 16 bar
1 ½" = ø50.5 mm TC2 front seal max. 120 °C
2" = ø64.0 mm TE2 installed on site
ISO 2852 max. 2 bar
In conformity with
max. 150 °C
FDA*
L00-FTL5xxxx-06-05-xx-xx-009
See "Accessories"
L00-FTL5xxxx-06-05-xx-xx-012
L00-FTL5xxxx-06-05-xx-xx-014
FDA*
max. 150 °C
L00-FTL5xxxx-06-05-xx-xx-015
Endress+Hauser 29
Liquiphant M FTL50(H), FTL51(H)
L
L
ø21.5
L00-FTL5xxxx-06-05-xx-xx-017 L00-FTL5xxxx-06-05-xx-xx-018
L00-FTL5xxxx-06-05-xx-xx-016
Any length L:
148...3000 mm (6...115 in); special version (TSP) on request up to 6000 mm (235 in)
! Note!
The switchpoints of the Liquiphant M are at other positions to those of the previous version Liquiphant II.
Special length "L II":
With vertical mounting from above the same switchpoint as for the Liquiphant II
FTL360, FTL365, FDL30, FDL35
"L II" depends on process connection:
115 mm for flanges and flange-like process connections
99 mm for threads NPT and R (BSPT)
118 mm for threads G 1 (BSP 1)
115 mm for threads G ¾ (BSP ¾)
104 mm for flush-mounted 1" (Endress+Hauser)
30 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Process connections – Parallel thread G ¾, G 1 to DIN ISO 228/I with flat seal to DIN 7603
– Tapered thread R ¾, R 1 to DIN 2999 Part 1
– Tapered thread ¾ -14 NPT, 1 - 11½ NPT to ANSI B 1.20.1
– Flush-mounted with welding neck to factory standard Endress+Hauser (G ¾, G 1)
– Flush-mounted with welding neck to factory standard Endress+Hauser (1"),
Sensor can be positioned
– Tri-Clamp 1½", 2" to ISO 2852
– Threaded pipe joint DN 32, 40, 50 to DIN 11851
– Aseptic connection DN 50 to DIN 11864-1 Form A for pipe DIN 11850
– SMS connection 2" (DN 51)
– DRD flange
– Varivent® DN 50 (50/40) to factory standard Tuchenhagen
– Flanges to EN/DIN from DN 25, for standards see "Product structure", to ANSI B 16.5 from 1",
to JIS B2220 (RF)
Human interface
Electronic inserts With FEL51, FEL52, FEL54, FEL55:
• 2 switches for safety mode and density change,
FEL51
• green LED to indicate operational status, L1 N U~
• red LED to indicate the switching status, 19...253V AC
1 Max >0,7
50/60Hz
flashes in the event of corrosion damage 2
I max : 350mA
1 2
With FEL56:
• 2 switches for safety mode and density change, Connecting terminal LEDs Switches
• green LED flashes to indicate operational status,
• red LED to indicate the switching status, L00-FTL5xxxx-03-05-xx-en-001
With FEL57:
• 2 switches for density change and cyclical checking,
• green LED to indicate operational status,
• yellow LED to indicate the covered status,
flashes in the event of corrosion damage
on sensor or if the electronics are defect
With FEL58:
• 2 switches for safety mode and density change,
• green LED
– flashes quickly to indicate operational status,
– flashes slowly to indicate corrosion damage
at the sensor or if the electronics are defect, L00-FTL5xxxx-03-05-xx-xx-013
With FEL50A:
FEL50A
• 8 switches for configuring the device address, Address
L00-FTL5xxxx-03-05-xx-en-002
Endress+Hauser 31
Liquiphant M FTL50(H), FTL51(H)
L00-FTL5xxxx-19-05-xx-xx-001
Light signals
AC/DC NAMUR
rd ye
gn gn
L00-FTL5xxxx-07-05-xx-xx-005
32 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
ye rd
gn
L00-FTL5xxxx-07-05-xx-xx-003
Endress+Hauser 33
Liquiphant M FTL50(H), FTL51(H)
ye 1
ye 2 gn
L00-FTL5xxxx-07-05-xx-xx-004
Green light (gn) lights up, both yellow lights (ye 1+2) do not light up:
Liquiphant M has detected a fault.
34 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
+
Process connections Order Code
L00-FTL5xxxx-01-05-xx-xx-001
Aseptic HE2 X
DRD PE2 X
SMS UE2 X
Varivent® WE2 X
! Note!
For CIP (Clean in Place) and SIP (Sterilize in Place) processes the pressure and temperture specifications of the
process connections must be observed.
# Warning!
To avoid risk of contamination, install according to the "Hygienic Equipment Design Criteria (HDC)" as stated
in the Subgroup Design Principles of the EHEDG, Doc. 8, July 1993. The flow of liquid during cleaning is
important and should be in compliance with HDC.
Combination Based on the various certificates, permissible combinations of housings *) and electronic inserts are given in
of housings and the following table.
electronic inserts *) Abbreviations: Polyester = PBT, Steel 316L = St., Aluminium = Alu
Aluminium housing with separate connection compartment = Alu/sep
B ATEX II 3G EEx nC IIC T6, WHG PBT, St., Alu, Alu/sep. FEL54
ATEX II 1/2 G, EEx ia IIC T6, WHG St., Alu, Alu/sep. FEL55/56/57/58/50A
ATEX II 1/2 D, T80°C
Endress+Hauser 35
Liquiphant M FTL50(H), FTL51(H)
Q FM, XP, Cl. I, II, III, Div. 1, Gr. A-G Alu FEL51/52/54,
with NPT cable entry FEL55/56/57/58
R FM, NI, Cl. I, Div. 2, Gr. A-D St., Alu, Alu/sep. FEL51/52/54,
with NPT cable entry FEL55/56/57/58
PBT FEL55/56/57/58
with NPT cable entry
PBT FEL51/52,
with NPT cable entry FEL55/56/57/58
S CSA, IS, Cl. I, II, III, Div. 1, Gr. A-G PBT, St., Alu, Alu/sep. FEL55/56/57/58
with NPT cable entry
T CSA, XP, Cl. I, II, III, Div. 1, Gr. A-G Alu FEL51/52/54,
with NPT cable entry FEL55/56/57/58
V TIIS Ex ia IIC T3 PBT, St., Alu FEL57
! Note!
on polyester housing (PBT), electric connecting cables run in pipes:
Do not screw cable entries firmly to the piping. Use flexible connections (e.g. with armoured hose).
If the piping is used for earthing then ensure that there is a continuous electrical connection.
! Note!
Despite the additional Dust-Ex-certificate, the FTL5x(H) is to be used as a liquid limit switch only.
36 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Ordering information
! Note!
This overview does not mark options wich are mutually exclusive.
Endress+Hauser 37
Liquiphant M FTL50(H), FTL51(H)
38 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
FTL51
BB ....... mm; 316L** Ra <3.2 µm/80 grit
BE ....... mm; AlloyC4** Ra <3.2 µm/80 grit
CB ....... inch; 316L** Ra <3.2 µm/80 grit
CE ....... inch; AlloyC4** Ra <3.2 µm/80 grit 2.3 kg/100 in
DB Length: type II*; 316L Ra <3.2 µm/80 grit 0.1 kg
DE Length: type II*; AlloyC4 Ra <3.2 µm/80 grit 0.1 kg
JB ....... mm; 316L** + temperature spacer 0.9 kg/m
+0.6 kg
JE ....... mm; AlloyC4** + temperature spacer 0.9 kg/m
+0.6 kg
KB ....... inch; 316L** + temperature spacer 2.3 kg/100 in
+0.6 kg
KE ....... inch; AlloyC4** + temperature spacer 2.3 kg/100 in
+0.6 kg
LB Length: type II*; 316L + temperature spacer 0.1 kg
+0.6 kg
LE Length: type II*; AlloyC4 + temperature spacer 0.1 kg
+0.6 kg
RB ....... mm; 316L** + pressure tight feed through 0.9 kg/m
+0.7 kg
RE ....... mm; AlloyC4** + pressure tight feed through 0.9 kg/m
+0.7 kg
SB ....... inch; 316L** + pressure tight feed through 2.3 kg/100 in
+0.7 kg
SE ....... inch; AlloyC4** + pressure tight feed through 2.3 kg/100 in
+0.7 kg
TB Length: type II*; 316L + pressure tight feed through 0.1 kg
+0.7 kg
TE Length: type II*; AlloyC4 + pressure tight feed through 0.1 kg
+0.7 kg
YY Special version
40 Electronics; Output:
A FEL50A PROFIBUS PA
1 FEL51* 2-wire, 19...253 V AC
2 FEL52* 3-wire PNP, 10... 55 V DC
4 FEL54 relay DPDT, 19...253 V AC, 19...55 V DC
5 FEL55 8/16 mA, 11... 36 V DC
6 FEL56 NAMUR (L-H signal)
7 FEL57 2-wire PFM
8 FEL58* NAMUR + test button (H-L signal)
9 Special version
Endress+Hauser 39
Liquiphant M FTL50(H), FTL51(H)
60 Additional Option:
A Basic version
C EN 10204 - 3.1 material (316L wetted parts)
Inspection certificate
N EN 10204 - 3.1 material, NACE MR0175 (316L wetted parts)
Inspection certificate
P 100 bar process pressure (FTL51)
R 100 bar process pressure, EN 10204 - 3.1 material,
NACE MR0175 (316L wetted parts) (FTL51)
Inspection certificate
S GL/ABS marine certificate (FTL51: max. 1600 mm)
Y Special version
! Note!
The basic weight includes the compact sensor, thread adapter G ¾, electronic insert, polyester housing
40 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Endress+Hauser 41
Liquiphant M FTL50(H), FTL51(H)
FTL51H
BC ....... mm; Ra <1.5 µm/120 grit 0.9 kg/m
BD ....... mm; Ra <0.3 µm/320 grit / A3 0.9 kg/m
CC ....... inch; Ra <1.5 µm/120 grit 2.3 kg/100 in
CD ....... inch; Ra <0.3 µm/320 grit / A3 2.3 kg/100 in
DC Length: type II*; Ra <1.5 µm/120 grit 0.1 kg
DD Length: type II*; Ra <0.3 µm/320 grit / A3 0.1 kg
JC ....... mm; Ra <1.5 µm/120 grit 0.9 kg/m
+ temperature spacer +0.6 kg
JD ....... mm; Ra <0.3 µm/320 grit 0.9 kg/m
+ temperature spacer +0.6 kg
KC ....... inch; Ra <1.5 µm/120 grit 2.3 kg/100 in
+ temperature spacer +0.6 kg
KD ....... inch; Ra <0.3 µm/320 grit 2.3 kg/100 in
+ temperature spacer +0.6 kg
LC Length: type II*; Ra <1.5 µm/120 grit 0.1 kg
+ temperature spacer +0.6 kg
LD Length: type II*; Ra <0.3 µm/320 grit 0.1 kg
+ temperature spacer +0.6 kg
RC ....... mm; Ra <1.5 µm/120 grit 0.9 kg/m
+ pressure tight feed through +0.7 kg
RD ....... mm; Ra <0.3 µm/320 grit 0.9 kg/m
+ pressure tight feed through +0.7 kg
SC ....... inch; Ra <1.5 µm/120 grit 2.3 kg/100 in
+ pressure tight feed through +0.7 kg
SD ....... inch; Ra <0.3 µm/320 grit 2.3 kg/100 in
+ pressure tight feed through +0.7 kg
42 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
YY Special version
40 Electronics; Output:
A FEL50A PROFIBUS PA
1 FEL51* 2-wire, 19...253 V AC
2 FEL52* 3-wire PNP, 10... 55 V DC
4 FEL54 relay DPDT, 19...253 V AC, 19...55 V DC
5 FEL55 8/16 mA, 11... 36 V DC
6 FEL56 NAMUR (L-H signal)
7 FEL57 2-wire PFM
8 FEL58* NAMUR + test button (H-L signal)
9 Special version
60 Additional Option:
A Basic version
B CoC, EN 10204 - 3.1 material (316L wetted parts)
Inspection certificate
C EN 10204 - 3.1 material (316L wetted parts)
Inspection certificate
S GL/ABS marine certificate (FTL51H: max. 1600 mm)
Y Special version
! Note!
The basic weight includes the compact sensor, thread adapter G ¾, electronic insert, steel housing
Endress+Hauser 43
Liquiphant M FTL50(H), FTL51(H)
Accessories
Welding neck G ¾ Order number: 52001052 max. 25 bar / max. 150 °C
with 3.1.B material certificate: 52011897
max. 40 bar / max. 100 °C
for flush-mounting
a Liquiphant FTL50 or FTL50H 26
with process connection GQ2
21
• With defined thread start *
• Sensor cannot be positioned
Material: corrosion-resistant steel
ø55 –0.4
ISO 228
G¾
ø32
1.4435 (AISI 316L)
Weight: 0.13 kg
Seal: silicone O-ring
Order number: 52001387 L00-FTL5xxxx-06-05-xx-xx-019
! Note!
Use only for FTL50/51 and FTL50H/51H!
ø60 –0.4
ISO 228
Weight: 0.19 kg
Seal: silicone O-ring
Order number: 52001386
L00-FTL5xxxx-06-05-xx-xx-020
FDA approved materials according to
21 CFR Part 177.1550/2600
* The tolerance of the defined thread beginnings between welding boss and sensor amounts to ± 15°.
44 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
ø65 –0.3
Material (wetted):
corrosion-resistant steel
1.4435 (AISI 316L)
Weight: 0.43 kg
5
Seal: silicone moulded seal L00-FTL5xxxx-06-05-xx-xx-021
Order number: 942816-0000
FDA approved materials according to
21 CFR Part 177.1550/2600
RD 52 x 1 6
ø42.3
ø65
Weight: 0.15 kg
Seal on Liquiphant: silicone moulded seal
Order number: 942816-0000
FDA approved materials according to
21 CFR Part 177.1550/2600 ~26
L00-FTL5xxxx-06-05-xx-xx-022
ø125
ø66
ø50
ø84
13
16
L00-FTL5xxxx-06-05-xx-xx-023
Endress+Hauser 45
Liquiphant M FTL50(H), FTL51(H)
14
Pressure up to 40 bar
92
Material: corrosion-resistant steel
92
1.4301 (AISI 304)
Weight: 0.54 kg
G1 12
L00-FTL5xxxx-06-05-xx-xx-024
M6 (3x)
Weight for G 1, NPT 1: 0.21 kg
Weight for G 1½, NPT 1½: 0.54 kg G1A
(1 NPT)
55 AF
19
pe =
0 bar
M6 (3x)
22
G 1½ A
(1½ NPT)
L00-FTL5xxxx-06-05-xx-en-006
G1 DIN ISO 228/I 1.4435 (AISI 316L) 52011888 3.1 material certificate
G 1½ DIN ISO 228/I 1.4435 (AISI 316L) 52011890 3.1 material certificate
NPT 1½ ANSI B 1.20.1 1.4435 (AISI 316L) 52011891 3.1 material certificate
46 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
High pressure sliding sleeves for continuous adjustment of the switch point
of a Liquiphant M FTL51
Material: corrosion-resistant steel
1.4435 (AISI 316L) or AlloyC4 ø60 FTL51
~70
Weight for G 1, NPT 1: 1.13 kg 50 AF
Weight for G 1½, NPT 1½: 1.32 kg
Seal package made of graphite
18
2
G1A
(1 NPT)
ø60
~72
50 AF
22
G 1½ A
(1½ NPT)
L00-FTL5xxxx-06-05-xx-en-007
G1 DIN ISO 228/1 1.4435 (AISI 316L) 52011880 3.1 material certificate
NPT 1 ANSI B 1.20.1 1.4435 (AISI 316L) 52011881 3.1 material certificate
G 1½ DIN ISO 228/1 1.4435 (AISI 316L) 52011882 3.1 material certificate
G 1½ DIN ISO 228/1 AlloyC4 52003666
NPT 1½ ANSI B 1.20.1 1.4435 (AISI 316L) 52011883 3.1 material certificate
NPT 1½ ANSI B 1.20.1 AlloyC4 52003670
pe
bar
(psi)
(psi)
Liquiphant S
100
(1450)
80
(1160)
(1160)
64
(928)
Liquiphant M
Tp
–1 °C
–60 –50 (–14.5) 0
150 260 280 (°F)
(–76)(–58) (32) (300) (500) (540)
L00-FTL5xxxx-05-05-xx-xx-002
Endress+Hauser 47
Liquiphant M FTL50(H), FTL51(H)
L00-FTL5xxxx-03-05-xx-xx-016
27.5
Body: PUR (blue)
Coupling nut: Cu Sn/Ni min. 40
Protection: IP67
L00-FTL20Hxx-07-05-xx-xx-004
Temperature range: –25...+70 °C
48 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Supplementary Documentation
! Note!
This supplementary documentation can be found on our product pages on www.endress.com
Endress+Hauser 49
Liquiphant M FTL50(H), FTL51(H)
50 Endress+Hauser
Liquiphant M FTL50(H), FTL51(H)
Endress+Hauser 51
Liquiphant M FTL50(H), FTL51(H)
Endress+Hauser
GmbH+Co. KG
Instruments International
Colmarer Str. 6
79576 Weil am Rhein
Deutschland
Tel. +49 76 21 9 75 02
Fax +49 76 21 9 75 34 5
www.endress.com
[email protected]
TI328F/00/en/03.06
SL/FM+SGML6.0 ProMoDo