Agilent 7697A Headspace Sampler: Installation and First Startup
Agilent 7697A Headspace Sampler: Installation and First Startup
Headspace Sampler
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2014 Warranty Safety Notices
No part of this manual may be reproduced The material contained in this document
in any form or by any means (including is provided “as is,” and is subject to being
electronic storage and retrieval or
CAUTION
changed, without notice, in future
translation into a foreign language) without editions. Further, to the maximum extent A CAUTION notice denotes a hazard. It
prior agreement and written consent from permitted by applicable law, Agilent calls attention to an operating
Agilent Technologies, Inc. as governed by disclaims all warranties, either express
United States and international copyright procedure, practice, or the like that, if
or implied, with regard to this manual and
laws. not correctly performed or adhered to,
any information contained herein,
could result in damage to the product
including but not limited to the implied
Manual Part Number warranties of merchantability and fitness or loss of important data. Do not
for a particular purpose. Agilent shall not proceed beyond a CAUTION notice
G4556-90013 be liable for errors or for incidental or until the indicated conditions are fully
consequential damages in connection understood and met.
Edition with the furnishing, use, or performance
of this document or of any information
Fourth edition, April 2014
contained herein. Should Agilent and the WA R N I N G
Third edition, September 2012
user have a separate written agreement
Second edition, November 2011
with warranty terms covering the A WARNING notice denotes a hazard.
First edition, January 2011 It calls attention to an operating
material in this document that conflict
Printed in USA with these terms, the warranty terms in procedure, practice, or the like that, if
the separate agreement shall control. not correctly performed or adhered to,
Agilent Technologies, Inc. could result in personal injury or
2850 Centerville Road
death. Do not proceed beyond a
Wilmington, DE 19808-1610 USA
WARNING notice until the indicated
安捷伦科技 (上海)有限公司 conditions are fully understood and
上海市浦东新区外高桥保税区 met.
英伦路 412 号
联系电话:(800)820 3278
Contents
1 Installation
Before You Begin 6
Tools required for installation 6
Personnel needed for installation 6
Parts supplied with the instrument 7
Verify site preparation 8
Prepare the GC and data system 8
Determine how you will connect the headspace sampler to the GC 9
Install the GC consumables needed for HS checkout 13
Inspecting the Headspace Sampler 14
Place the Headspace Sampler on the Bench 15
Install the Configuration Plug 17
Verify the available power configuration 17
Install the configuration plug 18
Install the Vial Tray (111 Vial Model Only) 21
Connect Cabling 26
Cabling diagrams for Agilent GCs 27
Other systems 28
Connect Gas Supplies 30
Connect the vial pressurization gas 31
Connect the carrier gas (using GC inlet carrier control) 32
Connect the Transfer Line to the GC 35
Install the transfer line to the Headspace 36
Attach the transfer line to a GC inlet 45
Install the transfer line support bracket 47
Turn On the Headspace Sampler 49
Set the IP Address 50
To set the LAN address at the keyboard 50
To use a DHCP server 50
Configure the Headspace Sampler 51
2 First Startup
Tools and Materials Required 54
Load the Checkout Method in the HS 55
Make the First Connection 57
Set the Checkout Method Parameters 58
1
Installation
Before You Begin 6
Inspecting the Headspace Sampler 14
Place the Headspace Sampler on the Bench 15
Install the Configuration Plug 17
Install the Vial Tray (111 Vial Model Only) 21
Connect Cabling 26
Connect Gas Supplies 30
Connect the Transfer Line to the GC 35
Turn On the Headspace Sampler 49
Set the IP Address 50
Configure the Headspace Sampler 51
Agilent Technologies 5
1 Installation
The other necessary parts and tools are supplied with the
new instrument.
Nut and ferrule set, 1/8-in. Swagelok, 5181-7482 1 Connect the 1/8-in. end of the tubing to the
stainless steel gas supply.
Transfer Line Connection Kit
Fused silica, deactivated, .530 mm id x 5 m 160-2535-5 1 Transfer line.
long
Fitting, internal reducer, 1/16-in. to 1/32-in. 0100-2594 1 Transfer line connection.
Ferrule, polyamide graphite 1/32-in. 0100-2595 1 Transfer line connection.
Transfer line support assembly G3504-60620 1 Support transfer line at GC inlet. Consists of
bracket, clamp, and knurled thumbscrew.
Tubing, post-drawn, 1650 mm 1530-2239 1 Connect HS Carrier fitting to the gas supply
from the GC inlet flow module.
Union SS 1/16-inch tubing 0100-0124 1 Connect the 1/16-in. end of the post-drawn
tubing to the open end of the GC inlet carrier
line.
Column cutting wafer 5181-7487 1 Trim fused silica transfer line.
CAUTION The tray gantry in secured only by the foam packing material. Be
careful when unpacking the tray. The gantry can move when you lift
the tray.
Product Line voltage (VAC) Frequency (Hz) Maximum Current rating Power outlet
continuous (amps) current rating
power
consumption
(VA)
111 Vial model 100*/120 50/60 850 6.2 15 Amp
single phase
(–10% / +10%)
111 Vial model 200/220/230/240 50/60 850 3.8/3.4/3.3/3.1 10 Amp
single/split phase
(–10% / +10%)
12 Vial model 100*/120 50/60 850 6.2 15 Amp
single phase
(–10% / +10%)
12 Vial model 200/220/230/240 50/60 850 3.8/3.4/3.3/3.1 10 Amp
single/split phase
(–10% / +10%)
* Users of 100 V power (Japan) receive the 120 V US power option with a 100 V/120 V step up transformer.
1 Locate the cover plate for the AC board on the lower back
right corner of the HS.
Screw
Hole for
configuration
plug label
2 Roll the tray back until the rollers engage the clips in the
chassis. See Figure 3.
Cable in clip
Cable connections
4 Slowly lower the front of the tray. Align the vial oven
access through- hole over the raised rim of the vial oven
access port. The tray front aligns with the front of the
instrument. The tray should be level and flat against the
instrument chassis.
5 If needed, gently slide the gantry all the way to the back
of the tray.
Connect Cabling
1 Connect the power cable. See Figure 5.
G1530-60930
8121-0940
G1530-60930*
LAN cable LAN cable 8121-0940 APG Remote
8121-0940 cable
GC:
7890A, 7820A, 6890
with LAN, 6850, HS
PC LAN switch 5975T LTM
or hub GC/MSD
G1530-60930*
LAN cable APG Remote
8121-0940 cable
GC HS
LAN switch 6890A GPIB
PC
or hub
G1530-61200
APG Remote Y-cable
LAN switch
or hub GC
MS or 7890A, 7820A,
MSD 6890 with LAN, HS
PC 6850
Other systems
Use the Agilent general purpose remote start/stop cable,
35900- 60670, to synchronize an Agilent product and a
non- Agilent product. See Figure 9 and Figure 10.
35900-60670
LAN cable General purpose
8121-0940 remote cable
non-Agilent HS
PC LAN switch
or hub
Agilent recommends installing shutoff valves (not supplied) near the vial
NOTE
pressurization and carrier gas fittings. Using the valves provides a fast way
to isolate the HS during troubleshooting and periods of inactivity.
1/8-in. flare
Carrier 1/8-in.
This procedure assumes that you will manually splice the HS into the
NOTE
carrier gas stream. If you have purchased accessory G3521A, 7890A S/SL
MMI Weldment for Headspace, you will skip some steps. See below. This
procedure does not discuss installation using other accessories or inlet
types.
If you purchased the G3521A 7890A S/SL MMI Weldment for Headspace, install it
NOTE now. Skip step 4.
CAUTION Cut the tubing as square as possible. Be careful to avoid kinks in the
tubing. Avoid deforming the tubing. The cut should not be beveled or
angled.
c Plug the inlet carrier line at the septum head using the
1/16- inch Swagelok nut, ferrules, and cap assembly
provided in the ship kit.
5 Install the male half of a 1/16- inch union onto the open
end of the carrier gas tubing coming from the inlet EPC
module.
Valve cover
Sample probe
10 O’Clock
position Vial pressurization gas
line
Carrier gas line
Sample loop
6 Slide the stainless steel ferrule over the 1/16- inch tube
end of the reducer, then install into the open valve port
(at 10 O’clock). Finger- tighten, then tighten 1/4- turn
more.
Polyimide ferrule
For detailed instructions on how to trim a fused silica column, refer to your
NOTE
HS maintenance manual or the GC maintenance manual.
Volatiles Interface
1 Slide the transfer line nut (G2319- 20210) onto the tubing.
2 Install a SilTite ferrule and make a SilTite connection as
described below. Refer to Table 9 and use the SilTite
ferrule appropriate for your transfer line.
CAUTION The SilTite ferrules are delicate. Follow the instructions in the next
steps very carefully to avoid overtightening.
Nut
Pre-swaging tool
Back mounting
post hole
Back mounting
detail
Front mounting
post hole
Front mounting
detail
Transfer line
Clamp
Support bracket
Knurled thumbscrew
Agilent 7697A
A.01.01
Power on successful
Now that the gases and HS are turned on, the vial
pressurization gas will begin to purge the HS sampling
system.
2
First Startup
Load the Checkout Method in the HS 55
Make the First Connection 57
Set the Checkout Method Parameters 58
If using an Agilent data system 58
Checkout method GC parameters 58
Prepare the Checkout Sample 61
Run the Checkout Sample 62
Compare Results 64
Prepare for Your Method (Optional) 67
Agilent Technologies 53
2 First Startup
Configuration
Vial Gas type He
Loop Volume (mL) 1
Advanced Functions
Extraction mode Single
Purge flow 100.0
Purge time 1.000
Temperature
Oven 100.0 °C
Loop/Valve 110 °C
Transfer line 115 °C
Time
GC cycle time 4 min. (FID, NPD)
5 min. (µECD)
6.5 min. (MS, MSD)
Vial equib time 7 min.
Pres equib time 0.1 min.
Inject time 0.5 min
Vial
The table lists the settings for a 7890A GC. Other GC types use similar
NOTE
settings.
Oven
Initial temperature 100 °C
Initial time 4.00 min
Rate 0 (Off)
Equilibration time 1.00 min
Run time 4.00 min
Inlet (multimode, volatiles interface, or split/splitless)
Mode Split
Initial temperature 250 °C (On)
Pressure 134.4 kPa (19.5 psi)
Split ratio 20:1
Split flow 80 mL/min
5
1
4 1
3
Compare Results
When the run completes, compare the output chromatogram
against the examples below. Is correctly installed and
operating properly, the output chromatogram should be
similar to the example here.
2.037
FID1 A, (HS_ECD\HSVI0004.D)
0.880
pA
2.476
160
Ethanol Nitrobenzene
140 (40,000 counts
2.825
typical)
120
80
60
40
0.858
20
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 min
2.384
NPD1 B, (HS_NPD\HSNPD002.D)
pA
Nitrobenzene
250
200
150
100
Ethanol
50
ECD1 A, (HS_ECD\HSEC0000.D)
Hz
Dichlorobenzene
120000
100000
2.461
80000
60000
Ethanol
2.810
40000
0.850
20000
0 1 2 3 4 min
Dichlorobenzene
1000000
TIC: 0201002.D
800000
Abundance
600000
t-Butyl disulfide
400000 Nitrobenzene
200000
0
3.20 3.60 4.00 4.40 4.80 5.20 5.60 6.00 6.40
Time
A
Making Swagelok Connections
Making Swagelok Connections 70
Using a Swagelok Tee 73
Agilent Technologies 69
A Making Swagelok Connections
Front ferrule
Important!
Back ferrule
The narrow end of the
Nut back ferrule fits into
the rear of the front
ferrule.
Tubing
CAUTION Use a separate stainless steel fitting in a vise for initial tightening of
the nut. Do not use an inlet or detector fitting. Strong forces are
Do not
Plug or
fitting held Tubing
in vise
Back
ferrule
Withdraw tubing
1-2 mm
Tighten nut
Pencil line
Do not combine valve actuator air with flame ionization air. The valve
NOTE
action will cause major upsets in the detector signal.
Materials needed:
• 1/8- inch preconditioned copper tubing
• Tubing cutter
• 1/8- inch Swagelok nuts and front and back ferrules
• 1/8- inch Swagelok Tee
• Two 7/16- inch wrenches
• 1/8- inch Swagelok cap (optional)
1 Cut the tubing where you want to install the Tee. Connect
the tubing and Tee with a Swagelok fitting. See Figure 32.
Cap
Tee fitting
Nut and
ferrules