Manual PXR-501T
Manual PXR-501T
PXR-501T
CONTENTS
-1-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
CONTENTS
WIRE CONNECTION
-2-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
-3-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
[Fig. 1]
[Fig. 2]
-4-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
-5-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
1. BOARD CALIBRATION
VR3 Small Scale (10mA ~ 80mA) mA FEEDBACK to mAs adjustment with VF converter
Measuring Point:TP5
Adjust the VR3 until the TP5 set to 25kHz± 0.5kHz.
Simple Adjustment: Connect mAs tester to the connector 8,9(mA TEST)of HV tank
and shoot x-ray and adjust mAs.
Conditions: 40kV 50mA / 10mAs
Measuring: 10 ± 0.2mAs
-6-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
TP15
GND
U25
VR6
TP8
kV FB
VR2
TP4
mA FB
-7-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
VR1
TP8
TP9
-8-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
1.5 AGING
-9-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
1.6 kV CALIBRATION
- 10 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
1.7 mA CALIBRATION
Calibration Method: Execute in panel calibration mode.( from C1-21 39 : mA, from C2-21
39 : Pre_mA )
Setting Range:
mA Sec kV Remark
10mA 0.1sec 40kV, 80kV, 125kV
100mA 0.1sec 40kV, 80kV, 125kV
200mA 0.1sec 40kV, 80kV, 125kV
320mA 0.1sec 40kV, 80kV, 125kV
- 11 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 12 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Select the bucky type by choosing no. DIP SW 1.on the HT controller board.
DIP SW 1 Bucky Type Remark
1PIN 2PIN 1BUCKY 2BUCKY
OFF OFF MOTOR Type MOTOR Type
ON OFF Solenoid Type Solenoid Type
OFF ON MOTOR Type Solenoid Type
ON ON Solenoid Type MOTOR Type
- 13 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 14 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
22
10
12
16 5
9 3 4 11 1 2 6
23 24
13 14 15 20 17 21 7 8
18 19
- 15 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
[1]Power on Switch
Switch on the power supply
[9]OVER Indication(Red)
The LED is switched on when operation exceeds radiography conditions. The exposure
is prohibited while [the over LED] is lighted on. In case the LED is turned on due to
changed exposure conditions in the APR mode, the memory of condition should not be
saved. Under the AEC mode selection, if radiography time exceeds backup time, it
blinks. Backup release switch cancels over display at the time of AEC mode.
[10]ERROR Indicator(Red)
The LED is turned on when an error occurs.
- 16 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
[16]Setting switch of a Patient Size and Display and Radiography memory switch
This switch is used for selecting Exposure size. And selected size is displayed. And in
that case of APR Exposure condition revision, store it.
* It is prohibited under DENSITY≠0, and OVER Display on to store APR Exposure
condition.
- 17 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
2. DIP SW FUNCTION
Adjust DIP Switch SW32 (8P) located on the back front of control panel and operate the each
calibration.
DIP SW (SW 32: 8P) Function Display
DIPSW-1 ON (DIPSW-2,3,4 OFF):Calibration Mode
DIPSW-2 ON (DIPSW-1,3,4 OFF):Not in use
DIPSW-3 ON (DIPSW-1,2,4 OFF):Not in use
DIPSW-4 ON (DIPSW-1,2,3 OFF):Not in use
DIPSW-5 ON :AEC selection admission
DIPSW-6 ON :Not in use
DIPSW-7 ON :Not in use
DIPSW-8 ON :Select Exposure End Buzzer
DIPSW-1,2,3,4 ON:Counter Mode
- 18 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 19 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
*
5) Setup of Exposure mode (mAs /sec radiography)
6) Execute Exposure Mode with mAs/sec selection switch
mAs setting of Exposure accumulation time.
Select the exposure accumulation time with mAs/sec setting switch
- Setting range:0.1mAs ~ 999mAs
- Step: 0.1、0.2、0.3、0.4、0.5、0.6、0.8、1.0、1.3、1.6、2.0、2.5、3.2、4.0、
5.0、6.3、8.0、10.0、12.5、16、20、25、32、40、50、63、80、100、125、
160、200、250、320、400、500、630 mAs
Exposure time『sec』setting
Select the exposure time with mAs/sec 5 setting switch.
- Setting range:0.001sec ~ 10sec
- Step: 0.001、0.002、0.003、0.004、0.005、0.006、0.008、0.010、0.013、
0.016、0.020、0.025、0.032、0.040、0.050、0.064、0.080、0.100、
0.125、0.160、0.200、0.250、0.320、0.400、0.500、0.640、0.800、
1.00、1.25、1.60、2.00、2.50、3.20、4 .00、5.00、6.3、8.0 、10sec
- 20 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
* It could display “OVER” when the exposure condition exceeds the item’s permitted
capacity. During that period, Exposure is prohibited. Please reset the exposure
condition.
- 21 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 22 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
2) Selection of Field
Select the target region with setting switch. You can choose the field by pushing switches.
The selected field of Ion Chamber will be displayed on the LED panel.
You cannot make radiography without selecting the field.
- 23 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
4.RADIOGRAPHY
The selection of exposure on and off buzzer is operated by the DIP SW-8.
(SW-OFF/ select exposure on buzzer, SW-ON/ select exposure(during 1sec) off buzzer)
- 24 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 25 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 26 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 27 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 28 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 29 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 30 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
5. DISPLAY OF PANEL
- 31 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.
Push the power on [1] switch and turn the power on.
Control panel is displayed as a calibration mode.
C-1:KVREF&MAREF SETTING
Select the numbers in the calibration mode with [4] mA setting switch.
The selected number in calibration mode is appeared on mA display on the panel.
The calibration data is shown on the [5] mAs/sec display after the calibration mode and
numbers are selected.
-20∼ 00 ∼+20
- 32 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
* Execute the x-ray exposure with following methods after the data is selected in
KVREF/MAREF setting mode.
[18] Execute the conversion of monitor(the monitor after calibration)with backup release
switch. During the monitor recording, it is possible to make x-ray exposures. The setting
of mA during calibration is fixed for the time. APR•AEC Mode is not in option this time.
- 33 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Shift the KVREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(25 mA fixed).
Shift the KVREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(32 mA fixed).
- 34 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Shift the KVREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(40 mA fixed).
Shift the KVREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(50 mA fixed).
Shift the KVREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W
It is possible to change on the screen with [18] backup switch(63 mA fixed).
- 35 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 36 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 37 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Shift the MAREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W
It is possible to change on the screen with [18] backup switch(40 mA fixed).
- 38 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 39 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 40 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 41 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
After selecting setting mode, no., you can see setting data on mAs/sec display[5]
You can change calibrate data with mAs/sec setting switch
Calibration data range is different from each other.
- 42 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
C-2 __0 : [10mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 10.
Data range is from 000 to 004 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __1 : [12.5mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 12.5.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __2 : [16mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 16.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __3 : [20mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 20.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __4 : [25mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 25.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __5 : [32mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 32.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __6 : [40mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 40.
- 43 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
C-2 __7 : [50mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 50.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __8 : [63mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 63.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __9 : [80mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 80.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __10 : [100mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 100.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __11 : [125mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 125.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __13 : [200mA : large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 200.
Data range is from 000 to 009 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
- 44 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
C-2 __14 : [250mA :large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 250.
C-2 __15 : [320mA : large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 320.
Data range is from 000 to 014 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
- 45 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 46 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 47 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 48 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Use in case C-2 __44 Bucky 1Table selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
Operation in Bucky 1Table selection
Magnet for Bucky 1Table is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
Use in case C-2 __45 Bucky 2stand selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
Operation in Bucky 1Table selection
Magnet for Bucky 2table is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
- 49 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 50 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
C-2 _66: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
- 51 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
C-2 _67: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
C-2 _68: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
C-2 _69: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
- 53 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Push the [2]power off switch and turn the power off.
Press 1st switch of DIP SW(SW32 : 8P)and turn it on.
Push the [1] power on switch and turn the power on.
Control panel is shown as a calibration mode.
Execute the initialization of calibration data after selecting calibration mode and numbers.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.
- 54 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.
- 55 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 56 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
The operation of x-ray Exposure is prohibited under the APR Calibration Mode.
Only the changes of APR radiography conditions are possible.
Push the [2]power off switch and turn the power off.
Press only 1st switch of DIPSW(SW 32 : 8P)and turn it on.
Push the power on [1] switch and turn the power on.
Control panel is displayed as Calibration Mode.
- 57 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 58 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
-3 70 -3 70 -3 85
-2 80 -2 90 -2 90
-1 90 -1 80 -1 95
0 0 0
+1 110 +1 130 +1 105
+2 120 +2 120 +2 110
- 59 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
5
3 4 1 2
17
- 60 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 61 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 62 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 63 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
When problems occurred during the use of equipment, they should be marked on the [3]kV,
[4]mA setting section. As well, the [10] error LED will be lighted on with the occurrence of
error.
Error Code
The error code is displayed when problems, such as abnormalities, occur to equipment.
While the error code is displayed, radiography operation is forbidden for safety.
A power supply is re-switched off and on, then an error code is cleared. If the problem of
equipment is not solved at the time of a power-supply re-switched on, the display is lighted
on again.
Check code
Check codes are warning displays when an error is on operation of equipment unlike an error
code. While the check code is displayed, radiography operation is forbidden for safety. Check
code is cleared when the backup release switch is switched on. Please refer to the check code
list for the case the clearance method is different.
- 64 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 65 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 66 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 67 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
CODE Description
Ch_01 TUBE THERMOSTAT OPERATION
The temperature of the slim and long shaped light bulb is higher than 80C.
Thermo switch is in open position.
Release the check code at Thermo switch off condition.
Ch_02 When another switch is pushed during exposure
Release the check code with back up reset switch.
Ch_03 When PREP. Switch on OP PANEL is detached during exposure
Release the check code with back up reset switch.
Ch_04 When X-RAY switch on OP PANEL is detached during exposure
Release the check code with back up reset switch.
Ch_05 When a hand switch (PREP) is detached during exposure
Release the check code with back up reset switch.
Ch_06 When a hand switch (X-RAY) is detached during exposure
Release the check code with back up reset switch.
Ch_07 Shortage of power-supply capacity for exposure
The power voltage drop is below AC160V and DC 150V.
Release the check code with input of power.
Ch_08 Simultaneous exposure prohibition
Other x-ray system is executed simultaneously when the exposure ready switch
was pushed.
Release the check code by off the power of other x-ray system.
Ch_09 Warning to X-rays room door opening
The door of x-ray room was opened when the exposure ready switch was pushed.
Stop the exposure of x-ray in this case and the check code is released when the
door is closed.
Ch_10 AEC
X-ray exposure does not satisfy AEC Reference Level(Voltage).
Expose again after pushing back up switch and resetting the condition .
- 68 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
[C-3]It implys the APR Initial Data transmitted by the initialization of calibration data.
*If possible, initial data is recommended to be 1 type, and in case you need several kinds of
initial data, different APR conditions about ROM should be prepared and correspond.
- 69 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 70 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 71 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
2. Check Point
Check Code No. Parts Check Point
Check 01 I/F Connection
Tube Tube Temperature
Check 02 OP Panel S/W Input
Check 03 OP Panel S/W Input
Check 04 OP Panel S/W Input
Check 05 Hand S/W Hand S/W Input
Check 06 Hand S/W Hand S/W Input
Check 07 MAIN AC Input
High-Voltage Capacitor Voltage Level
(DC 300~310V)
Check 08 I/F I/F Connection
Check 09 I/F I/F Connection
Check 10 Ion Chamber Ion Chamber Input Level
- 72 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 73 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 74 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 75 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 76 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 77 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
WIRE CONNECTION
- 78 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
1. HV TANK CONNECTION
T2 T1 OIL INJECTION
CATHOD ANODE
GND
CORE
* CORE
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)
- 79 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
TB502
CN503
CN502
TB501
TB503
TB504
- 80 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
- 81 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Bucky Motion
(RY3)
BUCKY1 (RY1)
BUCKY2 (RY2)
0.1sec
간찰기 (RY7)
LIGHT IN USE
(RY4)
TB 504 (BR-944C)
PIN I/O Signal
1 O BUCKY1 SPLY
2 O BUCKY1 DR BUCKY1 POWER
3 O BUCKY1 EXP1 (AC220 or AC110)
4 O BUCKY1 EXP2
5 O SPARE OUT1-1
6 I SPARE OUT1-2
7 I SPARE OUT2-1
8 O SPARE OUT2-2
9 O BUCKY2 SPLY
10 O BUCKY2 DR BUCKY1 POWER
11 O BUCKY2 EXP1 (AC220 or AC110)
12 O BUCKY2 EXP2
13 I BUCYK2 OK
14 I BUCYK2 OK (GND)
15 O BUCYK1 OK
16 O BUCYK1 OK (GND)
17 O +24V
18 O 24V GND
- 82 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM CO., LTD.
Ion Chamber
AEC BOARD
AEC2
9pin D-sub (BUCKY2)
CN602 Cable
CN603
Ion Chamber
CN601 CN604
9pin D-sub
Cable
AEC1
(BUCKY1)
Remark
shield cable
1
1
1 1
2 2
Ion - CN1
3 3
AEC CN60
4 4
5
5
6 6
7 7
8 8
9 9
9 9
외함 GND
15m
Connection Table
NO Wire Color Start Point End Point Length Remark Function
1 BLACK Ion - CN1 1 AEC CN60 5 15m N/C
2 UL2464 BROWN Ion - CN1 2 AEC CN60 4 15m FIELD 2
3 AWG24 RED Ion - CN1 3 AEC CN60 3 15m FIELD 1
4 9P (SHIELD CABLE) ORANGE Ion - CN1 4 AEC CN60 2 15m RESET
5 YELLOW Ion - CN1 5 AEC CN60 1 15m OUTPUT
6 GREEN Ion - CN1 6 AEC CN60 9 15m FIELD 3
7 BLUE Ion - CN1 7 AEC CN60 8 15m -15V
8 VIOLET Ion - CN1 8 AEC CN60 7 15m +15V
9 WHITE/SHIELD Ion - CN1 9 AEC CN60 6 15m GND
- 83 -