The document is a service report for preventative maintenance performed on a Philips Allura FC OPTIQ angiography system located in Banda Aceh, Indonesia, which was experiencing a blue screen issue; upon inspection, the technician found the PC console was stuck with a blue screen due to memory and component board issues, which were resolved by reinserting the memory and PPC board. The report provides equipment details, describes the activities performed and resolutions, and makes recommendations for continued maintenance and calibration.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views

Service Report

The document is a service report for preventative maintenance performed on a Philips Allura FC OPTIQ angiography system located in Banda Aceh, Indonesia, which was experiencing a blue screen issue; upon inspection, the technician found the PC console was stuck with a blue screen due to memory and component board issues, which were resolved by reinserting the memory and PPC board. The report provides equipment details, describes the activities performed and resolutions, and makes recommendations for continued maintenance and calibration.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Service Request Type: Service Request Number:

Service Report Corrective Repair 2022-10-22/120

Customer Details

Customer Name: RS.Kesdam Iskandar Muda Aceh Location: Banda Aceh - Indonesia
PO #: Contact Phone: +6287796032272
Contact Nome: Mrs.Esty Email Address:
Initial Problem Description: Preventive Maintenance

Equipment Details
System ID: 723003 Initial Equipment Status: Ready to use
Serial Number: 196 Entitlement: --
Product Description: Philips Allura FC OPTIQ

Activity Description

Activity #: Activity Close Dote:


Activity Type: Field Support Engineer Name: sukrisdianto
Tube Slice Reading: 700mA
Problem Description: Blue Screen and system down issue

Couse/ Diagnosis: Okt 22/2022


- Blue Screen from PC Console C-Arm
- X-ray Not available

Resolution/ Service Okt,22/2022


Performentd - Check and inspect supply electricity from hospital
- Turn on Unit OK
- Check and inspect indicator LED from cabinet generator OK
- Check and inspect mainboard from cabinet generator OK
- Check and inspect monitor Exam OK
- Functional Test Table movement OK
- Functional Test C-arm LAO,RAO,CAUD and CRAN OK
- Functional Test Flouro and cine OK
- PC Console stuck and blue screen
- Turn off unit
- All check fuse from cabinet generator OK
- All check component board from cabinet generator OK
- Reinsert memory from PC console C-Arm OK
- Reinsert PPC bord from PC console C-Arm OK
- Check and inspect connection cable from console to cabinet generator OK
- Check and inspect installation cable LAN from console to cabinet generator OK
- Turn on Unit and All check OK
- Functional ALL test OK
- Flouro and cine OK
- Turn off unit and ALL general cleaning cover from angiography OK
Page No. 1 of 1
Recommendation ;
1. Maintenance on schedule
2. Calibration on schedule
3. Back up supply UPS Min 10KVA
4. Performent check and test

Page No. 2 of 2
Service Request Type:
Service Report Request Number:
Corrective Maintenance
2320045-44

Date Time Hours Labour Type


Okt 22,2022 11:00-16:37 Zona travel
And
Maintenance

Ports Used Details


Port Number Description
0tY Asset Serial No

Expense Tracker
Expense Dote Type Comments
0tY

Engineer Signoture:
22-Okt-2022 11:00 Customer Signoture:
SUKRISDIANTO

Page No.3 of 3