0% found this document useful (0 votes)
34 views5 pages

Mamo Digitek

This document provides instructions for calibrating the compression force display of a Lorad MIV x-ray machine. The calibration process involves: 1) Configuring the machine with an 18x24 bucky, compression paddle, and scale centered on the bucky; 2) Opening the right side gantry cover; and 3) Setting switches on the Host CPU to calibration mode. The calibration ensures a compression force of at least 25 pounds is provided as required by safety standards.

Uploaded by

franklin fermin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
34 views5 pages

Mamo Digitek

This document provides instructions for calibrating the compression force display of a Lorad MIV x-ray machine. The calibration process involves: 1) Configuring the machine with an 18x24 bucky, compression paddle, and scale centered on the bucky; 2) Opening the right side gantry cover; and 3) Setting switches on the Host CPU to calibration mode. The calibration ensures a compression force of at least 25 pounds is provided as required by safety standards.

Uploaded by

franklin fermin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

DigiTecNotes

Number: 4001
Model: Lorad MIV

Skip intro

800.344.4832 • digitecmedical.com
DigiTecNotes
Number: 4001 Model: Lorad MIV

COMPRESSION FORCE CALIBRATION


PREREQUISITES: None
OBJECTIVE: To calibrate the compression force display
a) Compression scale
Preferred: RMI Model 163
Alternate: Standard Bathroom Scale
b) Basic service hand tools
ESTIMATED TIME: 15 minutes

PERFORMANCE
STANDARD: MQSA / ACR: Compression force of at least 25 pounds shall be provided. The
initial power drive must be between 25 pounds (111 newtons) and 45
pounds (200 newtons).
SAFETY: The production of x-rays is not required to complete this adjustment.

Start Calibration Process

800.344.4832 • digitecmedical.com
DigiTecNotes
Number: 4001 Model: Lorad MIV

CALIBRATION PROCESS (continued)


1) Configure MIV:
18 x 24 bucky
18 x 24 compression paddle
Compression scale tool centered laterally on bucky along chestwall

View video

800.344.4832 • digitecmedical.com
DigiTecNotes
Number: 4001 Model: Lorad MIV

CALIBRATION PROCESS (continued)


2) Open right side gantry cover
To release lock
Turn hex head lock bolt counter clockwise
OR Push down recessed philips head screw

View video

800.344.4832 • digitecmedical.com
DigiTecNotes
Number: 4001 Model: Lorad MIV

CALIBRATION PROCESS (continued)


3) Set switch positions on Host CPU
S6, DIP 4 = on = calibration mode
S6, DIP 8 = off
S4, rotary = 5
S2, pushbutton Display = C20f xxx

View video