Ilovepdf Merged
Ilovepdf Merged
Tables 5.1 and 5.2 give a rough idea of the use of different kinds of
transducers in the measurement of representative non-electrical properties. The
list is not exhaustive but representative. As discussed earlier, we are dealing
with electrical transducers only because of their adaptability to instrumentation.
Copyr.grted materal
Digitallnstruments 149
r
block diagram of Fig. 6.2 (b) can be implemented with the circuit shown in
Fig. 6.2 (c).
:- -------- _____ ..,d ..,c_,V'---.-------,
:
'
''' Attenuator • •
<!C V
- ----- ------ - - -
~. -. -------
S1c
dcV acV
Ohms
de rnA
..,
' Ohms dcmA AID
''' Converter
'• ICompensated I Rectifier ~
•
'
I Attenuator Decade
•' Counter
:' dcV
·--
acV Current
'
~----- - to Digital
dcmA Voltage Readout
Converter
Ohms
1 Constant
Current
Source
I
Ag. G1 (b) Block diagram of a basic digital multimeter
The current to be measured is applied to the summing junction (r. i) at the
input of the opamp. Since the current at the input of the amplifier is close to zero
because of the very high input impedance of the amplifier, the current JR is very
nearly equal to 1;, the current IR causes a voltage drop which is proportional to the
current, to be developed across the resistors. This voltage drop is the input to the
NO converter, thereby providing a reading that is proportional to the unknown
current.
,
-
Unknown r.1 AID Decade Digital
Current Converter Counter Read
Out
6. The function (X - Z)I Y provides off.~etting and scaling with user entered
Z and Y constants (where X is the reading).
7. The function 100 x (X - Y)/Y determines the percentage deviation, and
20 log XI)' displays X in decibels relative to the value of Y. An internal
memory (RAM) can be used to store the results of measurements and
programs for taking the measurements.
The signal may be amplified before being applied to the Schmitt trigger. The
Schmitt trigger converts the input signal into a square wave with fast rise and fall
times, which is then differentiated and clipped. As a result, the output from the
Schmitt trigger is a train of pulses, one pulse for each cycle of the signal.
The output pulses from the Schmitt trigger are fed to a START/STOP gate.
When this gate is enabled, the input pulses pass through this gate and are fed
directly to the electronic counter, which counts the number of pulses.
When this gate is disabled, the counter stops counting the incoming pulses.
The counter displays the number of pulses that have P.i!Ssed through it in the time
interval between start and stop. If this interval is known, the unknown frequency
can be measured.
When the function switch is on the period mode, the control voltage is
connected to proper gates of the logic circuitry, which connects the time base
signals to the counted signal channel of the main gate. At the same time the
logic circuitry connects the input to the gate control for enabling or disabling the
main gate. The other function switches, such as time interval ratio and external
standards perform similar functions. The exact details of switching and control
procedures vary from instrument to instrument.
+ Vee
(
\....r--
--
1~! D ~
Clock s A s 8 f- s
In put
T T 1- T
- R - R R C
A 8 c I o
Fig. G.13 (a) Decade counter
The outputs of the F/F B and D arc high (equal to binary I) after I 0 pulses
have been applied to the counter. Therefore, the output signal of the decade
counter is 1010. This output has to be reset on the very next pulse which is done
by the use of an AND gate that resets all F/F's to 0, when the t'utputs ofB and D
are 1. The wavefom1 shown in Fig. 6.13 (b) shows the pulse train applied to the
trigger i.nput (clock) oft he decade coumer (shown in Fig. 6.13(a)) and the output
wavefonn of each F/F.
At the beginning all the F/Fs are reset to 0000. The clock pulse is applied to
the trigger input T of the F/F. Since this is a negative edged triggered F/ F, at the
negative edge or falling edge oft he trigger input the F/ F A will toggle, and hence
the output of FIF A changes to level l; all other f/fs undergo oo change. The
outputs from the F/Fs will be 00{) l. At the next clock pulse the F/ F A will toggle
back to 0, and the output ofF/FA falls from I to 0 and is applied to the T input
oftbe next Fff B, toggling il The output of the F/F B changes to Eand the output
of the decade counter goes to 00 I 0.
Copynghled mater al
Digirol Instruments 165
Let us consider the case, when the input signal frequency is 48 Hz. The period
i.e. 1/48 s = 20.83 ms. Within this period, the number of pulses wiJI be 5208. (For
20 ms the count is 5000, hence for 20.83 ms the count is
5000 x 20.83 ms _
5208
20ms
Therefore, 20.83 ms the count is 5208). Now, these pulses are counted down
and the display reading is I 0000 - 5208 = 4792; 48Hz is displayed as 47.92 Hz.
and we can read the rpm of the rotating shaft directly. So, the relation between
the gate period and the number of pulses produced by the pickup is G = 60/P. If
we fix the gate period as one second (G = I s), then the revolution pickup must
be capable of producing 60 pulses per revolution.
Figure 6.19 shows a schematic diagram of a digital tachometer.
4 Digit
Display
JlliiiiL.Jlllll
Gate Count Input
Generator Counter/Latch
Decoder
Main/
Gate
Strobe Reset
Reset .......llA'-----4A_
Strobe _ __..,1_---'IL
~
Fig. G.l9 Basic block diagram of a digital tachometer
Copynghted mater ai
Digital Instruments 171
Pre- Zero
Amplifier Crossing
& Detector
Attenuator
As the P0 input signal increases in the positive half cycle, tbe zero crossing
detector changes its state when the input crosses zero (0) giving a high ( I) level
at the output. This causes the J- K F/F- 1 to be set (1), that is, the output (Q) of
F/F- 1 goes high. This high output from the F/F- 1 enables the AND gate, and
pulses from the clock arc fed directly to the counter. The counter starts counting
these pulses. Also this high output level of F/F-1 is applied to the clear input of
F/F-2 which makes the output of the F/F- 2 go to zero (0).
Now as the input P, which has a phase difference with respect to P0 , crosses
zero (0) in the positive half cycle, the zero detector is activated, causing its
output to go high (1). This high input in tum toggles the J-K F/F-2, making
its output go high. This output (Q) of F/F-2 is connected to the clear input of
F/F-1 forcing the F/F-1 to reset. Hence the output ofF/F-1 goes to zero (0). The
AND gate is thus disabled, and the counter stops counting.
The number of pulses counted while enabling and disabling the AND gate
is in direct proportion to the phase difference, hence the disp lay unit gives a
direct readout of the phase difference between the two inputs having the same
frequency/
Copynghted mater al
NOTES ON
OSCILLOSCOPES
NOTES ON ................................................................................................................................ 1
OSCILLOSCOPES .................................................................................................................... 1
Oscilloscope ............................................................................................................................... 3
Analog and Digital ................................................................................................................. 3
Analog Oscilloscopes ............................................................................................................. 4
Cathode Ray Oscilloscope Principles .................................................................................... 5
Electron Gun ...................................................................................................................... 5
The Deflection System ....................................................................................................... 6
Displaying a Voltage Waveform........................................................................................ 9
Triggering......................................................................................................................... 10
Pulse Generator: ............................................................................................................... 12
Sweep Generator .............................................................................................................. 12
X-Y Operation.................................................................................................................. 14
External Triggering .......................................................................................................... 14
Digital Storage Oscilloscopes (DSO)................................................................................... 15
Measurement Techniques..................................................................................................... 16
Phase difference: .............................................................................................................. 17
Controls ................................................................................................................................ 20
Display Controls............................................................................................................... 20
Vertical Controls .............................................................................................................. 20
Position and Volts per Division Settings.......................................................................... 20
Horizontal Controls .......................................................................................................... 21
Input Coupling.................................................................................................................. 22
X-Y Button....................................................................................................................... 22
DUAL Button................................................................................................................... 22
Alternate and Chop Buttons ............................................................................................. 23
ADD Button ..................................................................................................................... 23
LEVEL and +/- Buttons ................................................................................................... 25
Appendix .............................................................................................................................. 27
2
Oscilloscope
In many applications, observing certain voltage waveforms in a circuit plays a crucial role in
understanding the operation of the circuit. For that purpose several measurement instruments
are used like voltmeter, ammeter, or the oscilloscope.
An oscilloscope (sometimes abbreviated as “scope”) is a voltage sensing electronic
instrument that is used to visualize certain voltage waveforms. An oscilloscope can display
the variation of a voltage waveform in time on the oscilloscope’s screen
Figure 1.
A probe is used to connect the oscilloscope to the circuit. Figure 1 shows an oscilloscope and
a probe connected to it.
_
Figure 2.
Figure 2 shows a typical probe. Oscilloscope shows the potential difference between the two
terminals of the probe. The terminal ending with a hook is usually connected to the node in
the circuit whose voltage is of interest. The other terminal is usually (but not always)
connected to the ground. The probes are attached to input channels of the oscilloscope. Most
oscilloscopes have at least two input channels and each channel can display a waveform on
the screen. Multiple channels are useful for comparing waveforms. For example, one can
observe the voltage waveforms at the input and the output terminals of a circuit
simultaneously, by using a two channel oscilloscope.
Oscilloscopes also come in analog and digital types. An analog oscilloscope works by directly
applying a voltage being measured to an electron beam moving across the oscilloscope
screen. The voltage deflects the beam up and down proportionally, tracing the waveform on
the screen. This gives an immediate picture of the waveform. In contrast, a digital
oscilloscope samples the waveform and uses an analog-to-digital converter (or ADC) to
3
convert the voltage being measured into digital information. It then uses this digital
information to reconstruct the waveform on the screen.
Analog Oscilloscopes
4
Cathode Ray Oscilloscope Principles
Figure 4 shows the structure, and the main components of a cathode ray tube (CRT). Figure 5
shows the face plane of the CRO screen.
Fluorescent
Coating Vertical
Vertical Deflection Horizontal Deflection
Plates Plates
CRO Screen Deflection
Plates
Electron Beam
Figure 4. Figure 5.
Electron beam generated by the electron gun first deflected by the deflection plates, and then
directed onto the fluorescent coating of the CRO screen, which produces a visible light spot
on the face plane of the oscilloscope screen.
A detailed representation of a CRT is given in Figure 6. The CRT is composed of two main
parts,
• Electron Gun
• Deflection System
e- Electron Beam
e- e-
Electron Gun
Electron gun provides a sharply focused electron beam directed toward the fluorescent-coated
screen. The thermally heated cathode emits electrons in many directions. The control grid
provides an axial direction for the electron beam and controls the number and speed of
electrons in the beam.
The momentum of the electrons determines the intensity, or brightness, of the light emitted
from the fluorescent coating due to the electron bombardment. Because electrons are
5
negatively charged, a repulsion force is created by applying a negative voltage to the control
grid, to adjust their number and speed. A more negative voltage results in less number of
electrons in the beam and hence decreased brightness of the beam spot.
Since the electron beam consists of many electrons, the beam tends to diverge. This is
because the similar (negative) charges on the electrons repulse each other. To compensate for
such repulsion forces, an adjustable electrostatic field is created between two cylindrical
anodes, called the focusing anodes. The variable positive voltage on the second anode
cylinder is therefore used to adjust the focus or sharpness of the bright spot.
Electron Beam
Electron Gun
CRO Screen
Deflection System
Figure 7.
As shown in Figure 7, the electron beam passes through the deflection plates. In reference to
the schematic diagram in Figure 8, a positive voltage applied to the Y input terminal causes
the electron beam to deflect vertically upward, due to attraction forces, while a negative
voltage applied to the Y input terminal causes the electron beam to deflect vertically
downward, due to repulsion forces. Similarly, a positive voltage applied to the X input
terminal will cause the electron beam to deflect horizontally toward the right, while a negative
voltage applied to the X input terminal will cause the electron beam to deflect horizontally
toward the left of the screen.
6
Vertical
Vy Deflection
CRO Screen Plates
Vx
Horizontal
Deflection
Plates
Figure 8.
The amount of vertical or horizontal deflection is directly proportional to the corresponding
applied voltage. When the electrons hit the screen, the phosphor emits light and a visible light
spot is seen on the screen.
Since the amount of deflection is proportional to the applied voltage, actually the voltages Vy
and Vx determine the coordinates of the bright spot created by the electron beam.
Example 1:
Suppose Vx = sin(t), Vy = cos(t) are applied to the horizontal and vertical deflection plates
respectively. Then the bright spot would follow a circular path on the CRO screen.
Vy = cos(t)
CRO Screen
Vx = sin(t)
Figure 9.
7
Example 2:
Vy(t)
Vy(t)
Figure 10-a.
Vs(t)
Vs(t)
t
Figure 10-b.
Vy(t)
Vy(t)
Vs(t)
Vs(t)
Figure 10-c.
8
In Figure 10-a, the input signal Vy(t) is applied to the vertical deflection plates, whereas the
horizontal deflection plates are connected to ground. It is assumed that the electron beam is
kept at the extreme left position when the horizontal deflection plates are connected to
ground. Under this configuration, the bright spot in the CRO screen will follow a vertical path
(will go up and down) at the extreme left position of the screen.
In Figure 10-b, the input signal Vs(t) is applied to the horizontal deflection plates, whereas the
vertical deflection plates are connected to ground. This time, the bright spot will travel from
extreme left to extreme right end of the screen and will stop there.
In Figure 10-c, the signals Vy(t) and Vs(t) are applied to the vertical and the horizontal
deflection plates respectively. This time the bright spot will follow a sinusoidal path, resulting
a visualization of the input signal Vy(t) on the CRO screen.
Actually the bright spot must follow the same path fast and repetitively (at least 30 times in a
second) so that the human eye can perceive the motion of the bright spot as a continuous
curve. Therefore, in order to display the waveform on the CRO screen for the example in
figure 10-c, the signals Vy(t) and Vs(t) should be applied to the vertical and the horizontal
deflection plates periodically and in synchronization. The next section discusses the details of
this procedure and depicts how CRO handles this problem.
s(t)
Extreme Right
Extreme Left
t
9
action is called blanking and prevents any reverse retrace (or shadow) as the beam is going
back to the extreme left-hand position. The time period including the trace time and the
flyback time is called the sweep period.
The period of the sawtooth waveform plays a crucial role in obtaining a steady waveform on
the CRO screen. The following section discusses the requirements on the period of the
sawtooth waveform and the need of a synchronization between the sawtooth waveform and
the input waveform.
Triggering
Vy (t)
CRO SCREEN
t
s(t) T
τ
Extreme Right
Extreme Left
Figure 12
Suppose the input Vy(t) and s(t) shown in Figure 12 are applied to the vertical and horizontal
deflection plates of the CRO respectively. Note that, At the beginning of each sweep cycle,
(i.e when the bright spot is at extreme left) Vy(t) gets exactly the same value (The points
indicated by red circles). Therefore the bright spot is following exactly the same path in each
sweep cycle. Thus, we can observe a steady waveform on CRO screen. Notice that the time
between the beginning of two consecutive sweep cycles is a multiple of input signal period.
(i.e. = n T. T, are shown on the figure 12, n is a positive integer.)
10
Yinput (t)
CRO SCREEN
t
S(t)
Extreme Right
Extreme Right Extreme Right
Extreme Left
Extreme Left Extreme Left Waveform is not steady!
t
Figure 12
For the given case, the bright spot is following different paths in different sweep cycles,
therefore we can not obtain a steady waveform on CRO screen.
In order to obtain stable and stationary waveform displays, the sawtooth signal should be
applied to the horizontal deflection plates, in synchronism with the waveform being
displayed. CRO handles this synchronization problem by using the following structure.
Vertical
Amplifier
CRO Display
~ Yinput
s(t) p(t)
Horizontal Sweep Pulse
Amplifier Generator Generator
Figure 13
Notice that, the voltage waveform which is to be displayed on the CRO screen (Yinput in this
case) is applied to the vertical amplifier. In the amplification stage, only the amplitude of the
input waveform is changed. After the amplification stage, the output of the vertical amplifier
is applied to the vertical deflection plates. Then in order to obtain a steady waveform on the
CRO screen, a sawtooth waveform having a period which is an integer multiple of the period
of the input voltage waveform should be applied to the horizontal deflection plates.
11
Yinput (t) Level
P(t)
t
S(t)
Pulse Generator:
The main function of the pulse generator (PG) (see figure 13.) is to produce periodical pulses
with a period of T, which is equal to the period of the input signal. For that purpose, the input
signal is compared to a certain voltage level (‘Level’ on the figure 14). Producing pulses each
time the input voltage is equal to that certain voltage level, may seem to result pulses which
are periodic with period T. But this is not the case. Notice that, the ‘Level’ intersects the input
signal more than once in one period. Therefore, one of the intersection points is neglected in
each period. The decision on which intersection point to neglect is made by inspecting the
slope of the signal at the intersection point.
The pulse generator produces pulses each time the input voltage level is same as the ‘Level’,
after checking the slope of the signal at that time instant. In the example given in figure 14,
there are two intersection points at each period, and the one with the negative slope (blue
points) are neglected. (The selected slope is positive for that case). Actually, the sign of the
slope can be selected by using the +/- button of the CRO. Also the voltage level that the signal
is being compared to, can be adjusted by using the level button on the oscilloscope.
Sweep Generator
The main function of the sweep generator is to produce one cycle of a sawtooth waveform,
when it receives a pulse at its input. If the sweep generator receives a trigger pulse during its
sweep cycle (i.e., during the trace period Tr), it will simply ignore the pulse and continue with
the completion of its sweep cycle.
Depending on the selected level and the slope of the input signal, the output of the pulse
generator will consist of narrow trigger pulses separated from each other by one period T.
Each time the input signal crosses a preselected level (and a preselected slope), the pulse
generator emits one narrow trigger pulse. The emitted pulse triggers the sweep generator to
begin producing one cycle of the sweep waveform; its duration is the trace period Tr. At the
12
end of each sweep cycle, the sweep generator stops its output and awaits the arrival of the
next trigger pulse before producing a new sweep cycle.
Notice that if the sweep generator receives a trigger pulse during its sweep cycle (i.e., during
the trace period Tr), it will simply ignore the pulse and continue with the completion of its
sweep cycle. The trigger pulse received after the completion of the trace period will initiate
the new sweep cycle. This allows the scope to display more than one cycle, of period T, of the
signal connected to its vertical deflection plates.
Level
Assume +/- button of CRO is
released. At these points, even
the input is equal to the ‘Level’,
the pulses are not generated,
because slope is (-)ve
t
Those pulses are ignored by the
Sweep generator, since they are
received during the sweep cycle.
t
CRO SCREEN
S(t)
Figure 15.
In the given example, first the voltage waveform to be displayed on the CRO screen is
compared with a voltage level. The blue and the red points show the intersection of the input
signal with the level. Assuming the positive slope is selected, pulse generator produces pulses
at each time the input signal is equal to ‘level’, and its slope is positive. The pulses generated
by the pulse generator, trigger the sweep generator, which produces one cycle of the sawtooh
waveform. The trace time of the sweep generator is adjusted by the time/div button which is
available on the front panel of the oscilloscope. The resulting sawtooth waveform is applied
to the horizontal deflection plates, which leads to a steady display of the input signal on the
oscilloscope screen. Notice that, at the beginning of each sweep period (when the bright spot
is at the extreme left), input signal voltage is equal to `level` and has a positive slope.
Therefore, the waveform shown on the CRO screen starts with a positive slope at the extreme
left and its value is equal to the ‘level’. One can change these settings by varying the ‘level’
control or the +/- button of the oscilloscope.
13
The whole process is called triggering because, obtaining a steady plot on the CRO screen can
only be achieved by producing pulses at the input of the Sweep Generator at the correct time
instances. (i.e. triggering the Sweep Generator at the correct time instances.)
X-Y Operation
When the variation of one voltage waveform, Vy(t), as a function of another, Vx(t),
eliminating the parameter time, t, is desired, X-Y mode of operation is used. In X-Y mode,
one signal is applied to the vertical deflection plates whereas the other signal is applied to the
horizontal deflection plates. The XY button on the front panel of the oscilloscope disconnects
the triggering signal from the horizontal deflection system, and connects the second input
signal instead. This process is done by using a switch shown as ‘X-Y button’ on the figure
below.
Vertical
Amplifier
CRO Display
~ Yinput
s(t) p(t)
Horizontal Sweep Pulse
Amplifier Generator Generator
~ Vexternal
Xinput ~
X-Y Button Ext. Button
Figure 16
External Triggering
Rather than the input signal itself, an external signal can also be used for triggering. For that
purpose multi-positional switch, which corresponds to Ext. Button of the CRO, should be set
to position 2, as shown in Figure 16. The external signal should satisfy certain conditions in
order to obtain a steady waveform on the CRO screen. Keeping in my that the period of the
sawtooth waveform, s(t), should be an integer multiple of the period of the input signal, can
you find the conditions needed on the frequency of the external triggering signal?
14
Digital Storage Oscilloscopes (DSO)
The concept behind the digital oscilloscope is somewhat different to an analogue scope.
Rather than processing the signals in an analogue fashion, the DSO converts them into a
digital format using an analogue to digital converter (ADC), then it stores the digital data in
the memory, and then processes the signals digitally, finally it converts the resulting signal in
a picture format to be displayed on the screen of the scope.
Since the waveform is stored in a digital format, the data can be processed either within the
oscilloscope itself, or even by a PC connected to it. One advantage of using the DSO is that
the stored data can be used to visualize or process the signal at any time. The analogue scopes
do not have memory therefore the signal can be displayed only instantaneously. The transient
parts of the signal (which may vanish even in milliseconds or microseconds) can not be
observed using an analogue oscilloscope.
The DSO’s are widely used in many applications in view of their flexibility and performance.
15
Measurement Techniques
The major concern in observing a signal on the oscilloscope screen is to make voltage and
time measurements. These measurements may be helpful in understanding the behavior of a
circuit component, or the circuit itself, depending on what you measure. Except for the X-Y
mode of operation, the oscilloscope displays the voltage value of the waveform as a function
of time. The oscilloscope screen is partitioned into the grids, which divides both the
horizontal axis(voltage) and the vertical axis(time) into divisions which will be helpful in
making the measurements. See Figure 17.
5
- 1
- 2
- 3
- 4
- 5
0 0. 1 0 . 2 0 . 3 0 . 4 0. 5 0. 6 0. 7 0. 8 0 . 9 1
Obviously one needs to know the time or the voltage values corresponding to each division, in
order to make accurate calculations. These values are determined by two variables namely the
time/div and the volt/div both of which can be adjusted from the relevant buttons available
on the front panel of the oscilloscope (see Figure 10-13). Also note that, the time/div button
controls the trace time of the sweep generator, whereas the volt/div button controls the `gain`
in the vertical amplifiers in the vertical deflection system.
Typical quantities, which are of primer interest when observing a signal with the scope, are
shown in Figure 18.
5
DC 1
Value
Ground 0
Peak
value
- 1
- 2
Period
- 3
Peak to peak
value
- 4
- 5
0 2 4 6 8 10 1 2 1 4 16 18 20
16
For the given figure, suppose that the variables volt/div and time/div are set to:
volt/div = 2Volts/div.
time/div = 1millisecond/div
Then the corresponding values shown on the figure are calculated to be;
Phase difference:
In some applications, one may need to monitor or compare two or more signals
simultaneously. A typical example can be the comparison of the input voltage with the output
voltage of a two-port (input and output ports) circuit. If the signals that are being monitored
have the same frequency, a time delay may occur between the signals (i.e. one signal may
lead the other or vice versa). Two waves that have the same frequency, have a phase
difference that is constant (independent of t). When the phase difference (modulo 2 ) is zero,
the waves are said to be in phase with each other. Otherwise, they are out of phase with each
other. If the phase difference is 180 degrees ( radians), then the two signals are said to be in
anti-phase. If the peak amplitudes of two anti-phase waves are equal, their sum is zero at all
values of time, t.
5
5
- 1
- 1
- 2
- 2
- 3
- 3
- 4
- 4
- 5
0 2 4 6 8 10 12 14 16 18 20
- 5
0 2 4 6 8 10 1 2 14 16 18 20
17
The phase difference is expressed in terms of radians or degrees. In Dual Mode of the
oscilloscope the phase difference can be calculated easily as follows.
Given the two signals having the same frequency, as shown in Figure 21,
5
3
∆T
2
T
- 1
- 2
- 3
Phase
- 4
- 5
Difference
0 2 4 6 8 1 0 1 2 1 4 1 6 1 8 2 0
define,
T = horizontal spacing of the peak values (or the zero crossings) of the two signals.
T = horizontal spacing for one period.
Note that, one has to specify the leading or the lagging signal in order to fully describe the
time difference between the two signals. In the figure above, the signal represented with
dashed curve leads the other. Suppose that the signal represented by the dashed curve is
connected to Channel I of the oscilloscope, and the other one is connected to Channel II. In
such a case Channel I is leading the Channel II with phase difference equal to , and Channel
II is lagging the Channel I with phase difference equal to . Determining the leading or the
lagging signal may be frustrating at first, but note that the dashed curve reaches its maximum
value before the other does.
The phase difference between the signals can also be determined in XY mode of the
oscilloscope. In the XY mode, the x-axis data is taken from one channel, y-axis data is taken
from the other. In that way, Channel I vs Channel II graph can be obtained, so that the
variation of a signal with respect to another can be observed. Figure 22 shows a typical graph
in XY mode, of two signals having a constant phase difference.
18
5
A B
0
- 1
- 2
- 3
- 4
- 5
- 5 - 4 - 3 - 2 - 1 0 1 2 3 4 5
B
One can show this relation by expressing one signal as, y (t ) = sin( wt ± θ ) and the other
2
C
signal as, x(t ) =sin( wt ) . Then consider the value of y(t) when x(t) is zero volts. It should be
2
noted that, the center of the ellipsoidal shape (sometimes circular or linear shapes) on the
screen should be at the origin of CRO unless any DC component is added to one of the
signals.
In XY mode, the leading or the lagging signal can not be determined. One has to switch to
DUAL mode in order to specify the leading signal.
5 5
5
5
4 4
4
4
3 3
3
2 2
1 1
0 0
-1 -1
-1
-1
-2 -2
-2
-2
-3 -3
-3
-3
-4 -4
-4
-4
-5 -5
- 5 - 4 -3 -2 -1 0 1 2 3 4 5 -5 - 4 -3 -2 - 1 0 1 2 3 4 5
-5
-5 - 4 -3 -2 - 1 0 1 2 3 4 5
-5
-5 - 4 -3 -2 -1 0 1 2 3 4 5
19
Controls
Display Controls
Display systems may vary between analog and digital oscilloscopes. Common controls
include:
• An intensity control to adjust the brightness of the waveform. As you increase the
sweep speed of an analog oscilloscope, you need to increase the intensity level.
• A focus control to adjust the sharpness of the waveform. Digital oscilloscopes may not
have a focus control.
• Other display controls may let you adjust the intensity of lights and turn on or off any
on-screen information (such as menus).
Vertical Controls
Vertical controls are used to position and scale the waveform vertically. Oscilloscopes also
have controls for setting the input coupling and other signal conditioning, described in this
section. Figure 24 shows the vertical controls of the DSO3062A
20
voltage you can display on the screen is the volts/div setting times the number of vertical
divisions.
Often the volts/div scale has either a variable gain or a fine gain control for scaling a
displayed signal to a certain number of divisions. Figure 25 shows the vertical controls of the
HM203-7 CRO.
Horizontal Controls
Horizontal controls are used to position and scale the waveform horizontally. Figure 26 and
27 show typical front panel for the horizontal controls.
The horizontal position control (x-pos.) is used to move the waveform from left and right to
exactly where you want it on the screen.
The time per division (time/div) setting lets you select the rate at which the waveform is
drawn across the screen (also known as the time base setting or sweep speed). This setting is a
scale factor. For example, if the setting is 1 ms, each horizontal division represents 1 ms and
the total screen width represents 10 ms (ten divisions). Changing the time/div setting lets you
look at longer or shorter time intervals of the input signal.
As with the vertical volts/div scale, the horizontal sec/div scale may have variable timing,
allowing you to set the horizontal time scale in between the discrete settings.
Also note that, the time/div button actually controls the trace time of sawtooth waveform in
the sweep generator. When sawtooth waveform is zero volt, the bright spot is at the extreme
left-hand position, and when it is maximum, the bright spot is at the extreme right position.
Therefore, the bright spot travels from extreme left to extreme right in a time equal to the
Trace time. Assume that the CRO screen is divided into N equal horizontal divisions. The
bright spot travels the N divisions in Tr seconds. Therefore each division corresponds to
21
(Tr/N) seconds. If the Trace time is changed, the corresponding time for each division is
changed. Time per division controls can be used to select the appropriate time/div (i.e., the
Trace time of the sawtooth waveform).
Input Coupling
Coupling means the method used to connect an electrical signal from one circuit to another.
In this case, the input coupling is the connection from your circuit to the oscilloscope. The
coupling can be set to DC, AC, or ground (GND). By setting the coupling control to AC, the
DC offset voltage is removed form the input waveform, so that you see the waveform
centered at zero volts. When DC coupling is selected, both AC and DC components of the
input waveform are passed to the oscilloscope. Figure 28 illustrates the difference. The signal
in Figure 28 is y (t ) = 3 + sin( wt ) where 3 Volts is DC component and sin(wt) is AC
component. By selecting AC coupling, DC component is eliminated and only the signal of
sin(wt) is shown on the screen (Figure 28-b). The AC coupling setting is useful when the
entire signal (alternating plus constant components) is too large for the volts/div setting.
5 5
4 4
3 3
2
2
1
1
0
0
- 1
- 1
- 2
- 2
- 3
- 3
- 4
- 4
- 5
- 5
0 2 4 6 8 10 1 2 14 16 1 8 2 0
0 2 4 6 8 10 1 2 1 4 1 6 18 2 0
Figure 28-a: 2V peak to peak sinusoidal Figure 28-b: 2V peak to peak sinusoidal
with 3 Volts offset, shown in DC mode. with 3 Volts offset, shown in AC mode.
The ground setting disconnects the input signal from the vertical system, which lets you see
where zero volts is on the screen. With grounded input coupling and auto trigger mode, you
see a horizontal line on the screen that represents zero volts. Switching from DC to ground
and back again is a handy way of measuring signal voltage levels with respect to ground.
X-Y Button
Most oscilloscopes have the capability of displaying a second channel signal along the X-axis
(instead of time). This is called XY mode. Pressing the X-Y button the oscilloscope is used in
XY mode. See Also (Notes on CRO)
DUAL Button
The oscilloscopes have the capability of displaying both channel signals on the screen at the
same time. This is called the Dual Mode. This mode is usually used to measure phase
difference between two signals which is explained in Phase difference part on page 17.
22
Alternate and Chop Buttons
On analog scopes, multiple channels are displayed using either an alternate or chop mode.
(Digital oscilloscopes do not normally use chop or alternate mode.)
Alternate mode draws each channel alternately - the oscilloscope completes one sweep on
channel 1, then one sweep on channel 2, a second sweep on channel 1, and so on. Use this
mode with medium- to high-speed signals, when the time/div scale is set to 0.5 ms or faster.
Alternate mode is available when only DUAL button is depressed.
Chop mode causes the oscilloscope to draw small parts of each signal by switching back and
forth between them. The switching rate is too fast for you to notice, so the waveform looks
whole. You typically use this mode with slow signals requiring sweep speeds of 1 ms per
division or less. Chop mode is available when both DUAL and ADD button are depressed.
Figure 29 shows the difference between the two modes. It is often useful to view the signal
both ways, to make sure you have the best view.
ADD Button
When ADD button is depressed, the signals of both channels are algebraically added and the
result is displayed on the screen. Volt/div scales of two channels should be the same in order
to appropriately see the summation of the signals. When the volt/div scales of the channels are
not the same, the signals are summed up as they are displayed on the screen (i.e. graphically).
Assume a signal 2 sin( wt ) is connected to Channel I and a signal 2 sin( wt ) is connected to
channel II. CH I is set to 2 volts/div (Figure 30-a) and CH II is set to 1 volt/div (Figure 30-b).
When the ADD button is depressed, the resulting signal on the screen is shown in Figure 31.
23
5 5
4
4
3
3
2
2
0
0
- 1
- 1
- 2
- 2
- 3
- 3
- 4
- 4
- 5
- 5
0 2 4 6 8 10 12 14 1 6 1 8 2 0
0 2 4 6 8 10 12 14 16 18 20
Figure 30-a: The first signal seen on the Figure 30-b: The second signal seen on the
oscilloscope with 2 volt/div scale. oscilloscope with 1 volt/div scale.
24
5
- 1
- 2
- 3
- 4
- 5
0 2 4 6 8 1 0 1 2 1 4 1 6 1 8 2 0
Figure 31: The sum of two signals in Figure 30-a and 30-b when ADD button is depressed.
INVERT Button
When the INVERT button of a channel is depressed, negative of the signal is displayed on
the CRO screen.
EXT Button
When the EXT button is depressed, the oscilloscope is used in external triggering mode.
External triggering is explained at Section External Triggering at page 14.
AT/NORM Button
Using the AT/NORM button you can switch between automatic trigger level selection (AT)
and manual trigger level selection (NORM). When the AT/NORM button is released, the
automatic trigger level is selected as zero volts, so that the value of the signal on the extreme
left of the screen is equal to zero. When the AT/NORM button is depressed, the user can
determine the trigger voltage level (the voltage on the extreme left) manually by adjusting
LEVEL knob.
25
5 5
4
4
3
3
2
2
1
1
0
0
- 1
- 1
- 2
- 2
- 3
- 3
- 4
- 4
- 5
- 5
0 2 4 6 8 1 0 1 2 1 4 1 6 1 8 2 0
0 2 4 6 8 10 1 2 14 16 1 8 2 0
Figure 32-a: The signal when Figure 32-b: The signal when AT/NORM
AT/NORM button is released. (LEVEL is button is depressed, LEVEL is set to 1 Volt
automatically set to 0 volt.) and +/- button is released.
5
5
4
4
- 1
- 1
- 2
- 2
- 3
- 3
- 4
- 4
- 5
0 2 4 6 8 10 12 1 4 1 6 18 20
- 5
0 2 4 6 8 10 1 2 1 4 16 18 20
Figure 32-c: The signal when AT/NORM Figure 32-d: The signal when
button is depressed, LEVEL is set to 1 Volt AT/NORM button is depressed, LEVEL is
and +/- button is depressed. set to 2 Volt.
26
Appendix
The front panel of the oscilloscopes DSO3062A and the HM203-7 CRO are shown in the
following figures respectively.
27
Figure A3: Schematic for the Front Panel of the DSO3062A.
28
Bridges 11
INTRODUCTION 11.1
A bridge circuit in its simplest fonn consists of a network of four resistance arms
fonni.ng a closed circuit, wi th a de source of current applied to two opposite junc-
tions and a current detector connected
tO the oU1er two junction~, as shown in A
Fig. I I. I. '•/ ':::-!._2
Bridge circuits are extensively
used for measuring component values
such as R, L and C. Since the bridge
circuit merely compares the value of
an unknown component with that of
an accurately known component (a
standard), its measurement accuracy
can be very high. This is b ecause the
readout of this comparison is based on
B
the null indication at bridge balance,
and is essentially independent of the Fig. Ill Wheatstone's bridge
characteristics of the null detector.
The measurement accuracy is therefore directly related to the accuracy of the
bridge componen£ and not to that of the null indicator used.
The basic de bridge is used for accurate measurement of resistance and is
called Wheatstone's bridge.
When SW 1 is closed, current flows and divides into the two arms at point
A , i.e. I 1 and I2 . The bridge is balanced when there is no current through the
galvanometer, or when the potential difference at points C and Dis equal, i.e. the
potential across the galvanometer is zero.
To obtain the bridge balance equation, we have from the Fig. 11 .1.
II Rl = 12 R2 (ll.l)
For the galvanometer current to be zero, tl1c following conditions should be
satisfied.
( 11.2)
{11.3)
Therefore E b =E ( R3 _ R4 )
a R, +Rl Rz +R.a
Thevenin's eq uivalent resistance can be deterrnillled by replacing the voltage
source E with its internal impedance R R
1 2
or otherwise short-circuited and cal- r - -"'11,..flr.....,.--vw v----,
culating the resistance looking into
R, =Internal
terminals a and b. Since the internal Resistance of Rs = 0 n
resistance is assumed to be very low, 8 b
the Source
we treat it as 0 n. Thevenin's equiva-
lent resistance circuit is shown in Fig. R3 ~
11 .3. FIQ. Il3 Thevenin's resistance
Bridges 325
The equivalent resistance of the circuit is R1//R3 in series with R2//R4 i.e.
R 111R3 + R/IR4•
I = E,~ ( 11.5)
8
R,~ + Rg
Example11.2 An rmbafm,ced
Whearstone bridge is given in Fig.
/l.5. Calculme the current through
the ga/vanomete1:
Solution The Thevenin's equivalent
voltage between a and b is the differ-
ence of voltages at these po·ints i.e.
£,1, = £ 0 - Eb = Eh- £,..
.. Eh=
,
£( R4
Rz + R4
_ R3 )
Rr + R3
B
Fig.II.S
E _ 6( 10 k 3.5 k )
th - 2.5 k + 10 k I k + 3.5 k
E,h = 6 (0.800- 0. 778}
£,h=O.I32V
Thevenio 's equivalent resistance is
Rr R3
R,,, = Rz R4 + --''-''-
R1 + R3 R2 + R4
I k X 3.5 k 2.5 k X 10 k
R,h = + --------
1 k + 3.5 k 2.5 k + 10 k
l26 Electronic lnsuumentotion
= 0.778k+2k
= 2.778 k R,h=2.n8k
The equivalent circuit connected along Ech -=- Rg =3000
with the galvanometer is as shown in Fig. o.132 v
11.6.
The current through the galvanometer is
given by Fig.ll.6 Equivalent circuit
I E,h 0.132 V = .
42 88
A
g = R," + Rg - 2.778 k + 0.3 k ~
11.23 Slightly Unbalanced Wheatstone's Bridge
If three of the four resistor in a bridge are equal toR and the fourth differs by 5%
or less, we can develop an approximate but accurate expression for Thevenin's
equivalent voltage and resistance.
Consider the circuit in Fig. 11.7.
The voltage at point a is
ExR ExR E
E =
n R +R
=--
2R 2
The voltage at point b is E
E _ R + D.r x E E(R + D.r)
I> - R + R + D.r 2R + D.r
Thevenin 's equivalent voltage
between a and b is the difference Fig.ll.7 Slighdy unbalanced Wheatstone's
between these voltages. bridge
Therefore E
ch
= E - E
a b
= E (<R + D.r) -
2R + D.r 2
.!.)
= E (2(R + D.r)- (2R + M))
2(2R + D.r)
= E (2R+2D.r-2R-M J
4R + 2D.r
- E ( 4R ~;D.r)
If D. r is 5% of R or less, D. r in the denominator can be neglected without
introducing appreciable error. Therefore, Thevenin 's voltage is
E = ExD.r =E(D.r )
th 4R 4R
If the detector is a 200 - 0 - 200 J.lA galvanometer, we sec that the pointer is
full scale for a 5% change in resistance.
Rd) R
-= - 1 (11.6)
R.b R2
and the usual balance equations for the bridge give the relationship
Rt
(Rx + R,b) = ~ (R3 +Rub) ( 11.7)
R,b Rt
but R.b + R,b = Ryand -R =R
ob 2
Reb R•
- + } = -+}
Rob R1
Rt +R2
= __,__.._
R2
Rl +R2
i.e.
Therefore
.-----tilt - - - - .
Ek
Hence R = RI R3 (J 1.8)
x R2
Equation (11.8) is the usual Wheatstone's
balance equation and it indicates that the
effect of the resistance of the connecting
leads from point a to point c has been
eliminated by connecting the galvanometer
to an iotennediate position, b.
The above principle fonns the basis of . , .
the construction of Kelvin's Double Bridge, fig.ll.ll Kelvl,ns double bridge
popularly known as Kelvin's Bridge. It is a Double bridge because it incorporates
a second set of ratio arms. Figure 11.11 shows a schematic diagram of Kelvin's
double bridge.
llO Bectronic Instrumentation
and E=
we get £ 11, = R
R2
R
(
xI R3 + Rx +
(a+ b)R,.
·
J (11.10)
1 + 2 a+b+ RY
1.e.
(a+b)R>. RR
1 1 hR1 R,, bR>,
Rx+ +R3 = - -· +R3 + · +--..__-
a+b+R>. R2 R2 (a+b+Ry) a+b+Ry
R = R1 R3 + bR1 Ry aRY
x R2 R2 (a+b+Ry) a+b+R>,
R = R1 R3
"
bR_,,
R2 + (a+h+R,.)
(RR
1
-
a)
b
2
R1 a
But - =
Therefore,
This is the usual equation for Kelvin' s bridge. lt indicates that the resistance of
the connecting lead RY' has no effect on the measurement, provided that the ratios
of the resistances of the two sets of ratio arms are equal. In a typical Kelvin 's
bridge the range of a resistance covered is I - 0.00001 Q (10 J..lohm) with an
accuracy of± 0.05% to ± 0.2%.
/
Standard
Resistors
9 Steps
of 0.001 0 each
Manganin Bar
with Sliding
Contact. 0.0011 0
fig.11.13 Practical kelvin's bridge
vapynghtoo ma!E.na
II
Bridges 333
l
E
Error
Signal
1
Fig. 11.14 Wheatstone's bridge error detector with resistance R, sensitive
to some physical parameters
Since R, varies by only a small amount, the amplitude of the error signal
is normally quite low. It is therefore amplified before being used for control
purposes.
.
:
'
I
.
:
• • • , • • • • L • • • ~- -- • • • , • • • •,
5 ' I 1 I I
••• 1 · · · - l . . . I I
I : I I I
E
Resistance
in
4 ---~----
I I
.'
--~~---r---~----·
1
'
'
I
'
'
I
- ' 1 - - - L. . . . . . . . ---~ - - -'l - - - -·
6V 3 1 I I I 1
kn : : : : :
2 ---1·---~--------~---1·---·
I I I I
Error I I I I I
Signal
e, t 1 --- ~- ---:----1·--·r---:----·
I :
I
I
I
1
: :
I
I
I
1
I
(0,0) 20 40 60 80 100
- Temperature ·c
(a) (b)
Ag.II.IS
Solution
(i) The value of R, when the bridge is balance is calculated as
R,. = R2 RJ = 5 k X 5 k 5 k.Q
R1 5k
334 E/eCIJ'Onic Instrumentation
e, = E ( R
1
; R
3
R
2
~ R..)
= 6 ( 5k 4.5 k )
5 k + 5k 5 k + 4.5 k
= 6 (0.5 - 0.4736)
= 6 (0.0263)
0.158 v
=
The error signal can also be determined by usmg the foJiowing
equation.
es = 0.150 v
r----e--llll - - - - - - ,
R3
Programmable
Resistors
Copyr!Qtllec. Male~ul
336 Electronic Instrumentation
another, or several others with completely different dimensions, and the desired
results calculated with the microprocessor.
(One such microprocessor-based instrument is the General Radio model
I658RLC digibridge.)
Such intelligent instruments represent a new era in impedance measuring
instruments. The following are some features of these instruments.
I. Automatically measures R, inductance L, capacitance C, dissipation
factor D and storage factors for inductors Q.
2. 0 .1% basic accuracy
3. Series or parallel measurement mode
4. Autoranging
5. No calibration required
6. Ten bins for component sorting/binning (equivalent, binary number)
7. Three test speeds
8. Three types of display-programmed bin limits, measured values or bin
number.
Most of these features are available because of the use of a microprocessor,
e.g. the component sorting/binning feature is achieved by programming the
microprocessor.
When using the instrument in this mode, bins are assigned a tolerance range.
When a component is measured, a digital readout (bin number) indicating the
proper bin for that component is displayed on the keyboard control panel.
ACBRIDGES 11.8
Impedances at AF or RF are commonly determined by means ofan ac Wheat~tone
bridge. The diagram of an ac bridge is given in Fig. 11.17. This bridge is similar to
a de bridge, except that the bridge arms are
impedances. The bridge is excited by an ac
source rather than de and the galvanometer
is replaced by a detector, such as a pair of
headphones, for detecting ac. When the
bridge is balanced,
zt
-=-
Zz
23 z4 Head
Phone
where Z 1, 2 2, Z3 and Z4 are the impedances of
the arms, and are vector complex quantities
that possess phase angles. It is thus necessary
to adj u:st both the magnitude and phase angles
of the impedance arms to achieve balance,
i.e. the bridge must be balanced for both the
Fig.ll.17 ac Wheatstone's bridge
reactan.ce and the resistive component.
vopynghloo ma!E.na
Bridges 337
and
R1
--~--
R2 c ~ C3RI [11.12(b)]
roCx ro C3 X R
2
Since R3 does not appear in the expression for C", as a variabie element it is an
obvious choice to eliminate any i11teraction between the two balance controls.
-~I' r1edf"1 · I
338 Electronic Instrumentation
= I Ok X WO X I 0-6 = 20 ~
50 k
The equivalent series circuit is sb:own in Fig. ll.l9.
500 kQ 201!f
. _ _b
Fig. 11.19
R 2850
Step 2: c" = -2 x C4 =--xo.s~F=0.7 1 25~F
. R1 2000
Step 3: Loss angle of the capacitor (a series RC circuit) is defined as the angle by
which current departs an exact quadrature from the applied voltage. •8' is the loss
angle of the capacitor and is given by tan 8.
tan 8 = XRx = roC:c"Dx = 2-'C D
"J x"x
,r
IOmH x 40 K
Step 2: =40mH
lOK
The equivalent series circuit is shown in Fig. 11.21 .
400kn 40mH
B --......JWVV---' L..._-b
Fig.l1.21
R.r + Jwlx = R2 R3
R2R3
( * + jOJC1 J
Rx + jOJLx = -- + jw C 1 R2 R3
R,
Equating real terms and imaginary tenms we have
R, R ,
Rx = - -- 3 andL_. = C 1 R2 R3 (11.15)
R,
Also Q= wLX =we; Rz R3 X R, =we, RJ
Rx R2 R3
Maxwell's bridge is limited to the measurement of low Q values (I - 10).
The measurement is independent of the excitation frequency. The scale of the
resistance can be calibrated to read inductance directly.
The Maxwell bridge using a fixed capacitor has the disadvantage that there is
an interaction between the resistance and reactance balances. This can be avoided
by varyimg the capacitances, instead of R2 and R3, to obtain a reactance balance.
However, the bridge can be made to read directly in Q.
vopynghloo mal£!m~
I
Bridges 341
fig. ll.23
Example 11.8 (b) The amrs of an ac J\llaxwe/1 's bridge are arranged as fol-
lows:
AB and BC are non-reac1ive resist01:~ of 100 Q each. DA a standard variable
reactor L 1ofresistance 32. 7 Q and CD consists ofa standard variable resistor
R in serie.~ with n coil ofun/..?ro,vn impedance Z. balance wasjound with Lt =
50 mH and Z = 1.36R- Find the Rand L of coil.
Detector
R , Rx+ z-
LX me
I
jR, +jml..x
. ·' R
I
I=
R2 R3
R 1 R., + ~ = R2 R3 (11.16)
I
and ~ =
.t ImL R (1 1.17)
m,C
Solving tor L, and Rx we have, Rx'"' ol L, C1 R1•
Substituting for Rx in Eq. ( 11.16)
2 LX
R1 (m R 1 C1 L,) + G;' = R2 R3
vopynghloo ma!E.nal
Bridges 343
. .1
(ff
2
RI cl L.. + cLX : R2 R3
I
Multiplying both sides by C 1 we get
at R~ q Lx+Lx = R2R3 c,
There 1iore, LX = I R2 2R3R2
C1
c 2 (11.18)
+W I I
Substituting for Lx in Eq. ( 11.17)
2 2
R = w C1 R1 R2 R3
(11.19)
.• 1+W 2 R2
I
c 2
I
The term w appears in the expression for both L.• and Rx · This indicates that
the bridge is frequency sensitive.
The Hay bridge is also used in the measurement of incremental inductance.
The inductance balance equation depends on the losses of the inductor (or Q) and
also on the operating frequency.
An inconvenient feature of this bridge is that the equation giving the balance
condition for inductance, contains the multiplier 1./( I + !/~). The inductance
balance thus depends on its Q and frequency. '
R R C
2 3 1
Therefore, L
.t
= --''--''---'":-
I+ (I/ Q)2
For a value of Q greaterthan I 0, the tem1 1 /~ will be smaller than 1/ 100 and
can be therefore neglected.
Therefore Lx = R2 R3 C 1, which is the same as Maxwell's equation. But for
inductors with a Q less than 10, the 1 /~ term cannot be neglected. Hence tbis
-bridge is not suited for measurements of coils having Q less than I0.
A commercial bridge measure from I )l H - I00 H with ± 2% error.
Fig.11.25
~ p r ted ma
344 E.lectronic Instrumentation
(J) z Rl R l R 3 Cl2
R .• = I+ (1)2 R2 Cl
I I
(3000)2 X 10 k X 2 k X I k X (I X 10- 6 ) 2
=
I+ (3000) 2 X (2 kl X (I X I0- 6 ) 2
180 X 103
=
1+36
= 4.86kQ
and from Eq. ( 11.18) we have,
L = R 2 R3 C I
x I + (1)2 R l2 e ll
10 k X I k X (I X 10- 6 )
= --~~:....:.;___.!.;,........;...;;_..:._...,.....,..
2 2
I + (3000) X (2 k) X
2
(I X I 0~)
I0 10
= = - = 0.27 = 270 mH
I +36 37
Therefore R.• = 4.86 k and Lx = 270 mH
Example 11.9 (b) Four arms of a Hay Bridge are arranged as follows:
AD is coil ofunknown impedance Z, DC is a non-inductive resistimce ofI kn,
CB is a non-inductive resistance of800 in series .with a standard capacitor n
of2 JlF. BA is a non-ilrductive resistan ce of 16500 !1, ifthe supply frequen cy
is 50 Hz. Calculate the value ofLand R ofcoil When the bridge is balanced.
Step 3: R = CJ R4 R2 R 3
(1)2
•
I+ (J) 2 c ~2 R 42
Example 11.9 (c) Find the unknown resiswnce and inducrance having tlze.fo/-
lowing bridge arms
C4 = I J.lF. R1 = RJ = R4 =1000 Q, (tJ = 314 radls
Solution To find R 1 and L1
Step 1: Given
R2 R3C4 R _ w2C] R4 R2 R3
I + w2 CJ R] ' - 1 + w2 CJ RJ
1
89 79
1+ (314)- :<(I J.lft x (I 000) 2
where
as
(
R -
X
_j_)
me,
= R2 (- j} +
R, (mCJ)
~ C,
CJ
Equating the real and imaginary tenns, we get
R2 C1
R =- - (11.20(a)]
_. CJ
and CX = -R, C) [ 11.20(b))
R2
The dial of capacitor C 1 can be calibrated directly to give the dissipation factor
at a particular frequency.
The dissipation factor D of a series RC circuit is defined as the cotangent of
the phase angle.
R
D = -X·' = mCX RX
X:
Also, Dis the reciprocal of the quality factor Q, i.e. D = 1/Q. D indicates the
quality of the capacitor.
Commercial units measure from I 00 pf- I J,lf, with ± 2% accuracy. The dial
of C3 is graduated in terms of direct readings for C.., if the resistance ratio is
maintained at a fixed value.
This bridge is widely used for testing small capacitors at low voltages with
very high precision.
The lower junction of the bridge is grounded. At the frequency nonnally used
on this bridge, the reactances of capacitor C3 and c;:, are much higher than the
resistances of R 1 and R2 • Hence, most ofthe voltage drops across C3 and Cx> and
very little across R1 and R2• Hence if the junction of R1 and R2 is grounded, the
detector is effectively at ground potential. This reduces any stray-capacitance
effect, and makes the bridge more stable.
Example 11.10 {b) A sample ofinsulation was placed in C/1'111 A 8 ofa Schering
bridge, when the bridge was balanced at a frequency of 50 Hz. the other arms
of the bridge were as follows
Arm BC- a non-inductive R of 100 Q
Arm CD- n noll-inductive R of300 Q i11 parallel with a capacitor of0.5 pF
Arm DA -a loss ji"ee capacitor of 100 pf
Determine the capacitance, egui1•alent series resista11ce a11d PF of the
iusulation in test arm AB
' ·. dma- I
348 £Jectronic lns[IUmentalion
Step 2: c, = R:R x C2 =
300
100
x
,
soo x w-'- = 166.6 pF
1000
= 2 X :1:-1-+ X 50 X-- X 0.5 X I 0-6
J(
= w- o.5 = o.o5
1x
Example II.ID (e) A capacitor is tested by a Schering bridge which .forms one
arm AB ofthe bridge. The other arms are
AD- a non-inductive resistance of I 00 .Q,
DC - a non-1-eactive resistance of300 Q in parallel with a capacitor of0.5 f.1F,
BC- a stm1dard loss fi"ee capacitor of1Q,OpF.
Tlte supplyfl-equency is 50 Hz. The bridge is balanced.
Calculate the capacitor value and /he power factor ofthe capacitor under test.
Example IUD (f) A sample Bakelite was tested by the bridge method
(Schering) at I I kV,50 Hz. Balance was obtained atthe jollowings values
AB - dielectric material under test in tlie form of a capacitor
BC- a siandard air capacitor of I 00 pf
CD -capacitor of0.6 p.F in paralle l with a /lOll-reactive resistance of300 n
DA- no~reactive resistance of100 Q
Ca/c;u/ate the capacitance and equivalent series resistance ofthe sample.
Step I : R = R2 x ~ = 100 x 0 6 10
· x -<i 6 MQ
X CJ 100 X 10-12
R. 300 -P
Step 2: C, = - x C3 = - x I00 x I0 - = 300 pF
· R2 100
= ~x5kQ =2ill
25
R 2ill
Step 2: CX = -R1 x c,, = - - x 0.25 "J::
Sill 1-'-'
2
~cpyr 1100"' a
Bridges 35 I
R _ R C
Therefore - 2 - - 1 +- 3 (11.21)
R4 R3 C1
l
and - -- = wC3 R 1 (11.22)
we, R3
cJ -;: __1 _
(I)= ---;=:====:=~
Jc,R,C3 R3
as w=2 7C/
l= 1 ( I 1.23)
· 21e Jc, R, c 3 R3
The two conditi\)ns for bridge balance, (1 1.21) and (11.23), result in an
expression determining the required resistance ratio R21R4 and another expression
determining the frequency of lhe applied voltage. If we satisfy Eq. (11.21) and
also excite the bridge with the frequency of Eq. ( 11.23}, the bridge will be
balanced. ·
In most Wien bridge circuits, the components are chosen such that R 1 =
R3 = R and C 1 = C3 = C. Equation ( 11.21) therefore reduces to Ri R4 = 2 and
Eq. (11.23) to/= 112rrRC, which is the general equation for the frequency of the
bridge-circuit.
The bridge is used for measuring frequency in the audio range. Resistances R1
and R3 can be ganged together to have identical values. Capacitors C 1 and C3 are
normally of fixed values.
The audio range is normally divided into 20 - 200 - 2 k - 20 kHz ranges.
In this case, the res-istances can be used for range changing and capacitors C 1
and C 3 -for' fine frequency control within the range. The bridge can also be used
for measuring capacitances. In that case, the frequency of operation must be
known.
The bridge is also used in a harmonic distortion analyzer, as a Notch filter, and
in audio frequency and radio frequency oscillators as a frequency determining
element.
An accuracy of0:5%- I% can be readily obtained using this bridge. Because
it is frequency sensitive, it is difficult to balance unless the waveform of the
applied voltage is purely sinusoidal.
f= I
2rr Js x 10-9 x 4. 7 x 103 x 1ox 10-9 x 1ox 1oJ
f= I
21C Js X 10- 10 X 4. 7
105
1= = 3.283 kHz
2TC Js X 4.7
RJ = R4
R2
(R, + w 2RI C2)
1 1
= 100 k ( 3 J k + I )
25 k . (15.7 1 k )2 X 3.1 k X {5.2 X I 0- 6 ) 2
= 12.41ill
= -=-2n-_J's===o==o=x=o=.5=J.l=:F=x=4==o=='o=x=t=J.L:==-
F
I
=--r============
2nJsoo x 400 x o.s x 10-
12
106
= -:-
2n- -/t:8700:=x=2:=:070
6
= 10 = 1000kHz = 1000 _ 0.398 kHz
2nx400 2x3.14x400 314x8
Step 2 : Also given,
R 4 C, R
-2+ - = -
RI c2 R3
400 0.5 J.lF 1000
- + = --
800 I J.lF R
0.5+0.5= 1000
R
R = IOOOQ
jRb
RbRd+j (J)LdRb- (J)C =RaRe
d
356 Electtonic lnsl/llmenta!ion
Therefore f= I (1 1.24)
· 21r ~Ld cd
The bridge can be used to measure unknown inductances or capacitances. The
losses Rd can be determined by keeping a fixed ratio Rj Rh and using a standard
variable resistance to obtain balance. If an inductance is being measured, a
standard capacitor is varied until balance is obtained. lf a capacitance is being
measured, a standard inductor is varied until balance is obtained. The operating
frequency of the generator must be known in
order to calculate the unknown quantity. Balance B
is indicated by the minimisation of sound in the
headphones.
A
MAXWELL-WIEN BRIDGE 11.17
As seen before, a positive phase angle of
inductive impedance can be compensated by the
negative phase angle of capacitive impedance, D
R,
Therefore, Z 1= - - - ' - - - -
1+ }ID c, R,
Z2 = R2, Z4 = R4 and Z 3 = R3 + jOJL3
Using balance condition, Z 1Z3 =Z2 Z4
Therefore, L3 = C1R2R4
H ence the unknown resistance R3 and unknown inductance L3 can be
determined
Example 11.11
Prom F ig. 11.3 1' R2 R~ 600 x 400
RIRJ = R2R,j, therefore R, "' T "' 1000 240 n
fig. ll.31(a)
c
oc oc
Cold Junction
c
ficure 4-35 Bridge-type thermocouple instrument.
~~
IC A
o-----~~-=--4
8 ..-
Current Coil
Ltne R e
o - - - - - - - -- - -- ·----4----'----<l __j
Figure 4-36 Diagram of an electrodynamometer wattme!er connected to mea!>ure
the power of a single-phase load .
Sec. 4-1 I .I, the electrodynamometer used as a voltmeter or an ammeter has the
fixed coils and the movable coil connected in series, thereby reacting to the effect
of the current squared. When used as a single-phase power meter, the coils are
connected in a different arrangement (see Fig. 4-36).
The fixed coils, or field coils, shown here as two separate elements, are
connected in series and carry the total line current (iJ. The movable coil. located
in the magnetic field of the fixed coils, is <::onnected in series with a current-
limiting resistor across the power line and carries a small current (ip). The instan-
taneous value of the current in the movable coil is ip = e/Rp, where e is the
instantaneous ·voltage across the power line, and Rp is the total resistance of the
movable coil and its series resistor. The deflection of the movable coil is propor-
tional to the product of these two currents, ir: and ip, and we can write for the
average deftection over one period:
1 rr e I (T (4-29)
9av = K T Jo i Rp dt = K2 T Jo ei dt
Line R Lood
Current Cotl
The watthour meter IS not often found in a laboratory situation but it is widely
used for the commercial measurement of electrical energy. In fact, it is evident
wherever a power company supplies the industrial or domestic consumer with
electrical energy. Figure 4-38 shows the elements of a single-phase watthour
meter in schematic form.
lI
1
Shaft
/
Magnet
Magnetic I
Core Currer.!
Cod
Lood
The power factor, by definition, is the cosine of the phase angle between voltage
and current, and power-factor measurements usually involve the determination of
this phase angle. This is demonstrated in the operation of the crossed-coil po wer-
factor meter. The instrument is basically an electrudynamomeler moverne.nt.
where the moving element consists of two coils, mounted on the same s haft but at
right angles to each other. The moving coils rotate in the magnetic field provided
by the field coil that carries the line current.
The connections for this meter in a single-phttSe circuit arc shown in the
circuit diagram of Fig. 4-40. The field coil is connected as usual in series with the
line and carries the line current. One coil of the movable elemenlisconnecred in
series with a resistor across the lines ~nd receives its current from the applied
. ------
potenttal difference. The second coiTOTll1Cmovable element is connected in
senes with an inductor across the lines. Since no control springs are used. the
balance position of the movable element depends on the resulting torque devel-
oped by the two crossed coils. When the movable element is in a balanced posi-
tion, the contribution to the total torque by each element must be equal but of
opposit~ sign. The developed torque in each coil is a function of the current
through the coil and therefore depends on the impedance of that coil circuit. The
torque is also proportional to the mutual inducta nce between each part of the
crossed coil and the stationary field coil. This mutua l inductance depends on the
angular position of the crossed-coil elements with respect to the position of the
stationary field coil. When the movable element is at balance, it can be shown that
Field Coil
Field Cot!
---
Load
its angular displacement is a funrlion of the pha~e angle between line current (field
coil) and line voltage (crossed coils). The indication of the pointer, which · is
connected to the movable element, is calibrated directly in terms of the phase
angle or power factor.
The polarized-vane power-faclor meter is shown in the construction sketch
of Fig. 4-41. This instrument is used primarily in three-phase power systems,
because its operating principle depends on the application of three-phase voJt-
aie. The outside coil is the potential coil, which is connertrd to the three phase
Moving Vane
INTRODUCTION! 4.1
The most commonJy used de meter is based on the fundamental principle of the
motor. The motor action is produced by the flow of a small amount of current
through a moving coil which is positioned in a permanent magnetic field. This
basic moving system, o ften called the D' Arsonval movement, is also referred to
as the basic meter.
Di fferent instrument fonns may be obtained by starting with the basic meter
movement and adding various elements, as follows.
1. Tbe bas ic meter movemem becomes a de instrument, measuring
(i) de current, by adding a shunt resistance, forming a microammeter, a
milliammeter or an ammeter.
(ii) de voltage, by adding a multiplier resistance, forming a milli
voltmeter, volt meter or kilovoltmeter.
(iii) resistance, by adding a battery and resistive network, forming an
ohm meter.
2. The basic meter movement becomes an ac instrument, measuring
(i) ac voltage or current, by adding a rectifier, forming a rectifier type
meter for power and audio frequencies.
(ii) RF voltage or current, by adding a thermocouple-type meter for RF.
(iii) Expanded scale for power line voltage, by adding a thermistor in a
resistive bridge network, forming an expanded scale (100 - 140 V)
ac meter for power line monitoring.
Copynqhted Mater al
80 Elearonic lnstrumenrarion
Example 4.1 Calculate the sensitivity ofa 200·p.A /neter movem; ,;t 1vhich
is to be used as a de voltmeter.
S= 1 =--~
(/Jid ) 200 f.lA
Therefore S = 5 knN
DC VOLTMETER 4.3
A basic 0 ' Arson val movement can be converted into a de voltmeter by adding
a series resistor known as multiplier, as shown in Fig. 4.1 . The function of the
multiplier is to limit the current through the
movement so that the current does not exceed Multiplier
+
the full scale deflection value. A de voltmeter
measures the potential difference between two
points in a de circuit or a circuit component. I
v
To measure the potential difference be·
tween two points in a de circuit or a circuit
component, a de voltmeter is always connect·
cd across them with the proper polarity.
The value of the multiplier required is Fig. 4.1 Basic de voltmeter
calculated as follows. Referring to Fig. 4.1,
lm = full scale deflection current of the movement (f;:,d)
R,. = internal resistance of movement
R, = multiplier resistance
V= full range voltage of the instrument
From the circuit of Fig. 4 .1
v
Therefore R_, =~- Rm
m
The multiplier limits the current through the movement, so as to not exceed
the value of the full scale deflection f;:wl·
The above equation is also used to further extend the range in DC voltmeter.
Example 4.2 (a) A basic D 'Arson val movement witlr a full scale deflection
of 50 J.1A and internal resis tance of500 f2 is used as a voltmete1: De/ermine rhe
value of tire multiplier resistance needed to measure a voltage range ofO- 10 V.
Solution Given
R = ~-R = IO -500
• Im m 50 "A
~
6
= 0.2 x L0 - 500=200k - 500
= 199.5kQ
Example 4.2 (b) Calculate tlze value of multiplier resistance on the 50V
range of a de voltmeter /hat uses a 500 pA meter movement with an internal
resistance of I kf2.
Solution
Step I: The sensitivity of 500 11A meter movement is given by
S = Ill.,= 11500 11A = 2 kQ/V.
Step 2: The value of the multiplier resistance can be calculated by
R_. = S x range - R,.
Rs = 2 kQ/V X 50 V - I kQ
= 100 kQ - I kQ = 99 kQ
__
number of resistors (multipliers) along
with a range switch to provide a greater
number of workabl e ranges. ...
Figure 4.2 shows a multirangc
voltmeter using a three position switch +
and three multipliers R 1, R2.• and R3 for / Rm
voltage values v,, v2, and v3.
Figure 4.2 can be further modified
to Fig. 4.3, which is a more practical
arrangement of the multipUer resistors Fig. 4.2 Multirange voluneter
of a multirange voltmeter.
ln this arrangement, the multipliers are connected in a series string, and the
range selector selects the appropriate amount of resistance required in series with
the movement.
Copynghte<l rnter al
82 Electronic Instrumentation
R R
v,
/ Rm
R, = -
v - Rm =
50
-<> - 500 = I x 106 - 500 = I 000 K - 500 = 999.5 k!2
lm 50 xtO
Case 3: For range 0- 100 V
Copy ed ...,ale
Voltmeters and Muhimeters 83
Example 4.5 Convert a 'basic D 'A rson val movement with an internal
resistance of 100 Q and aful{ scale deflection of 10 mA into a muliirange de
voltmeter with ranges from 0 - 5 V. 0 - 50 V and 0- 100 V.
R =~= 50
5k.Q
I [ ftc! 10 rnA
Therefore. R2 = R, - (R3 + R,.) = 5 k.Q - (400 .Q + I 00 .Q)
= 5 k.Q- 500 .Q = 4.5 K.Q
Step 3: For a I 00 V range (V1) position
R =~ = lOO = IOk.Q
r I fttl lOrnA
Therefore R 1 =R, - (R 2 + R3 + R,.) = 10 k.Q - (4.5 k.Q + 400 .Q + 100 .Q)
= l0k.Q - 5.Q = 5 k.Q
Hence it can be seen that R3 !bas a non-standard value.
Solution For a 10 V range (V4 position of switch), the total circuit resistance is
R =~= 10 =5 k.Q
' 2mA
Iftd
Therefore R4 = R1 - R,. = 5 k - 50 = 4950 .Q.
Copynghte<l Moler al
84 Electronic Instrumentation
R =-=
v 50
=25kQ
' I fs.t 2 rnA
R =~=
250
125 kQ
' l ftd 2mA
Therefore R1= R, - (R2 + R3 + R4 + Rno)
= 125 k- (25 k + 20 k + 4950 + 50)
= 125 k- 50 k
=75 kQ
Only the resistance R4 (low range multiplier) has a non-standard value.
Sensitivity The sensitivity or Ohms per Volt rating of a voltmeter is the ratio of
the total circuit resistance R, to the voltage range. Sensitivity is essentially the
reciprocal of the full scale deflection current of the basic movement. Therefore,
S = J/Ip d QJV.
The sensitivity · S' of the voltmeter bas the advantage that it can be used to
calculate the value of multiplier resistors in a de voltmeter. As.
Copyr.9nted mater al
Voltmeters •Ond Multimeters 85
, Example 4.8 Calculate the value ofmultip/ier resistance for the multiple
range de voltmeter circuit shown in Fig. 4.5 (a).
Copynghled mater al
86 Electronic Instrumentation
Solution
Step I : The sensitiVIty of 50 J.IA meter
Rs 1
movement is given by
S = Ill., = 1/ 50 J.IA = 20 ill!V. 10 v 30 v
The value of the multipHer resistance can be
calculated by 3V~ o-----'
Step 2: The value of the multiplier for 3 V range Fig. 4.5 (b)
Rs = S X range - R,
R, = 20 k.Q/V X 3V - I k.Q
= 60 k!l - I kQ = 59 k!l.
Step 3: The value of the multiplier resistance For 10 V range can be calculated by
R, = Sx range - Rm
Rs = 20 k.Q.N x lO V - I k.Q
= 200 k.Q - I kQ = 199 k.Q.
Step 4: The value of the multiplier resistance For 30V range can be calculated by
Rs = S x range -Rm
Rs = 20 illiV X 30 V - I kil
= 600 kn - I kQ = 599 kn
LOADING 4.6
When selecting a meter for a certain voltage measurement, it is important to
cons ider the sensitivity o f a de voltmeter. A low sensitivity mete1r may give a
correct reading when measuring voltages in a low resistance circuit, but it is
certain to produce unreliable readings in a high resistance circuit. A Voltmeter
when connected across two points in a highly resistive circuits, acts as a shunt for
that portion of the circuit, reducing the total equivalent resistance ofthat portion
as shown in Fig. 4.6. The meter then indicates a lower reading than what existed
before the meter was connected. This is called the loading effect of an instrument
and is caused mainly by low sensitivity instruments.
Solution Inspection of the circuit indicates that the voltage across the R2
resistance is
10 k
k + k X 100 V =50 V
10 10
This is the true voltage across R1 .
Case I
Using a voltmeter having a sensitivity of I 000 QJV.
It has a resistance of 11)00 x 50 = 50 kQ on its 50 V range.
Connecting the meter across R2 causes an equivalent parallel resistance given
by
R =10kx50k SOOM
8.33 kQ
ey 10 k +so k 60k
Now the voltage across the total combination is given by
R
V.= "" xV
R1 + R..,
8 33
V. = · k X I00 V = 45.43 V
I 10 k +8.33 k
Jm
88 Electronic Instrumentation
Case2
Using a vohmeter having a sensitivity of20,000 nN. Therefore it has a resistance
of
20,000 x 50 = 1000 k = I Mfl
This voltmeter when connected across R2 produces an equivalent parallel
resistance given by
9
R =IOkxiM_ 10 tOk=9.9kn
··q IOk+IM l.OIM 1.01
Solution (i) The voltage across the resistance Rb, without either meter connected,
is calculated using the voltage divider fonnula.
150
Therefore, VR = Sk x 30 = k - 5V
b 25 k + 5 k 30 k
(ii) Starting with meter I, having sensitivity S = I kflN
Therefore the total resistance it presents to the circuit
R111 = S X range = I kn!V X I 0 = I 0 kQ
I
The total resistance across R" is, Rb in parallel with meter resistance R,. 1
=5kx10k= 3 .33 kn
5k+IOk
Therefore, the voltage reading obtained with meter I using the voltage divider
equation is
R,q 3.33 k
VRb = X v= X 30 = 3.53 v
R,"' + Ra 3.33 k + 25 k
Therefore the voltage reading obtained with meter 2, using the voltage divider
equation is
VR = 4.88 k X 30 = 4.88 k X 30 = 4.9 V
b 25 k + 4.88 k 29.88 k
5 v -3.33 v
voltmeter
. I error = sv x I 00% =33.4%
5 V -4.9V
Similarly voltmeter 2 error = x 100% = 2%
SV
VRh = Rh X Jl = 5k X SO= 50 X 5 k SV
Ra + Rh 45 k + 5 k 50 k
On the 5 V range
R,. = S X range = 20 kfl X 5 V = I00 kfl
Copy• r od --,ate a
90 Electronic lnsuumenta!ion
5 V - 4.88V
The% error on the 10 V range = V x I00 =2.34%
5
On 30 V range
Rm = S x range = 20 illN x 30 V = 600 k
R = R, X Rb = 600 k X 5 k = 3000 k X I k = 4.95 k
•q Rm + Rb 600 k + 5 k 605 k
The voltmeter reading on the 30 V range
4 95
VRb = R,q XV= · k X 50= 4.95 V
R,q + R0 45 k + 4.95 k
Copyngl'led mater a!
Voltmeters ond Multimeters 91
Solution
Step 1: The current meter will be connected in to the circuit as shown in Fig 4.8
(a).
3V
(b)
Fig.4.8
Looking back into terminals x and y and using Thevinin's equivalent
resistance,
R = R + R2 x R3 = I k + I k x I k = l. 5 kQ
I I R2 +R3 lk
Therefore fm = 0.938 X J
The current thro the meter is 93.8% of the expected current, therefore the
meter current caused a 6.2% error due to effects of loading .
Copynghled mater al
I I
. .
3
DC AMMETER·· 3.1
The P?HMC galvanometer constitutes the basic movement of a de ammeter.
Since the coil winding of a basic movement is small and light, it can carry only
very smaLL curtt..'lltS. When large currents are to be measured, it is necessary to
bypass a major part of the current through a resistance called a shunt, as shown
io Fig. 3. I. The resistance of shunt can be calculated using conventional circuit
analysis.
Referring. to Fig. 3. I
R,. = h' temal resista nce o f the movement.
1, 1,"' shunt current
·•
I.
•.-
l, "'' full scale cleflcctitm current of the movement
l = ful i scale current of the ammeter+ shunt (i.e. total current)
' .
·. O'Arsonval
Movement
. _ _ _ _ _ _ _ _ _ _ __ L_ _ _ _ _ _ _ _ _ __J
'· Since the shtmt resistance is in parallel with the meter movement, the voltage
drop across the shunt and movement must be the same.
Therefore Vsh. = V,,
lsi: Rsh =- 1 R.,..
111
Copynghled mater al
Ammeters 67
Example 3.4 (a) Design an Aryton shunt (Fig. 3.4) to provide an ammeter
with a current range of0 - 1 mA, 10 mA, 50 mA and 100 mA. A D 'Arsonval
movement with an internal resistance ofI 00 Q andfull scale current of50 p.A
is used.
+ 1mA
SOmA
1OOmA 10mA R.
R3
+
t- D'Arsonval
Movement
R2
R1
-
Fig. 3,4(a) For Example 3.4(a)
(3.1)
ForO-lOrnA
(3.2)
ForO - SOmA
ForO-IOOrnA
99950 JlA (R1) =50 JlA ( 100 + R2 + R3 + R4) (3.4)
But R 1 + R2 + R3 = 5.26 - R4 • Substituting in Eq. 3.2, we have
9950 !lA (5.26- R4 ) = 50 JlA (I 00 + R4 )
9950 !lAx 5.26 - 99SO JlA x R4 = SOOO jl.A + SO JlA R4
(9950 JlA x S.26 - 5000 J.!.A) = 9950 JlA R4 + SO JlA R4
9950 JlA X S.26 - SOOO JlA 47377 JlA
Therefore R =
4 lOrnA 10
rnA = 4.737 n
R4 =4.74 Q
In Eq. 3.1, substituting for R4 we get
R 1 + R2 + R3 = 5.26 - 4.74 = 0.52
Therefore R 1 = 0.05263 Q
From Eq. 3.5, we have
R2 = 0.10526 - R 1 = 0.10526 - 0.05263 = 0.05263 Q
Hence the value of shunts are
R 1 = 0.05263 Q ;R2 = 0.05263 Q
R3 = 0.4147U; R4 = 4.74Q
Copyrghted Mater al
70 Electronic Instrumentation
Example 3.4 (b) Calculate the value _of the shunt resistors for the circuit
shown below.
Rm= 1kn
r - - - - { ;f } - - - - . ,
1A
10mA
Fig. 3.4(b) For Example 3.4(b)
R = Rm where n = I II,.
sh (n- I) ..
Step 2: When the meter is set on the I 00 rnA range, the resistance R, and Rc
provides the shunt.
The shunt can be found from the equation
R. =(R +R )=lm(Rm+R.,h)=100J.1A(IO. I +1000) = l.OIQ
slrz b c l 100 nlA
Step 3: The resistor which provides the shunt resistance on the lA range can be
found from the equation
100 J.lA (10. 1 + 1000) = 0.101 Q
IOOOmA
Copynghted mater al
Ammeters 71
=9.09!1
Hence Ra = 9.09 Q, Rb = 0.909 Q and Rc= 0.101 Q
Test
Meter Set Leads
to
Lowest External
Current Shunt
Range
Note that the range of the basic meter movement cannot be lowered. (For ex-
ample, if a I 00 flA movement with I00 scale division is used to measure I ~A,
the meter will deflect by only one division. Hence ranges lower than the basic
range are not practically possible.)
Copynghled mater al
Indicators and
Display Devices 2
INTRODUCTION 2.1
Analogue ammeters and voltmeters are classified together, since there is no
basic difference in their operating principles. The action of all ammeters and
voltmeters, except those of the electrostatic variety, depends upon a deflecting
torque produced by an electric current. In an ammeter this torque is produced
by the current to be measured, or by a definite fraction of it. In a voltmeter it is
produced by a current that is proportional to the voltage to be measured. Hence
both voltmeters and ammeters are essentially current measuring devices.
The essential requirements of a measuring instrument are (a) that its
introduction into the circuit where measurements are to be made, should not alter
the c ircuit conditions, and (b) the power consumed by it be small.
The induction principle is most generally used for Watt-hour meters. This
principle is not preferred for use in ammeters and voltmeters because of the
comparatively high cost and inaccuracy of the instrument.
3 E
moving coil system is often called the
D' Arsonval galvanometer.
The D'Arsonval movement shown
in Fig. 2.1 employs a spring-loaded
coil through which the measured cur- N s
rent flows. The coil (rotor) is in a nearly s
homogeneous field of a permanent mag- +
net and moves in a rotary fashion. The Fig. 2.1 D'Arsonval principle
amount of rotation is proportional to the
amount of current flowing through the coil. A pointer attached to the coil indi-
cates the position of the coil on a scale calibrated in terms of current or voltage.
It responds to de current only, and bas an almost linear calibration. The magnetic
shunt that varies the field strength is used for calibration.
Example 2.1 (a) A moving coil instrumem has the following data.
Number ofturns= 100
Width ofthe coil = 20 mm
Depth ofthe coil= 30 mm
Flll.x density in the gap = 0.1 Wb!nl
Calculate the deflecting torque when carlJ'ing a cumelll of 10 mA. Also
calculate the deflection, if the control spring constant is 2 x 1~ N m/degree.
I= 30 X 10-<>
0.1 X 30 X J0- 3 X 20 X 10-3 X 100
1= 30 X I0-6. = S rnA
0.1 X 600 X 10-6 X 100
Pointer
Magnet
Core
I
I
--1-- Permanent
Magnet
Coil
~~~1+--1-+- Control
Spring
Curve
2
[UnderdampedJ
Copyngrted matcnal
30 E{earonic Instrumentation
One Of the simplest methods is to attach an aluminium vane to the shaft of the
moving coil. As the coil rotates, the vane moves in an air chamber, the amount
of clearance between the chamber walls and the air vane effective IX controls the
degree of damping.
Some instruments use the principle of electromagnetic damping (Lenz's law), ·
where the movable coil is wound on a light aluminium frame. The rotation of the
coil in the magnetic field sets up a circulating current in the conduct.ive··frame,
causing a reiarding torque that opposes the motion of the coil.
A PMMC movement may also be damped by a resistor across the coil. When
the coil rotates in the magnetic field, a voltage is generated in the coil, which
circulates a current through it and the external resistance. Tills produces an
opposing or retarding torque that damps the motion. In any galvanometer, the
. value of. the external resistance that produces critical damping can be found.
This resistance is called critically damping external. resistance (CDRX). Most
voltmeter coils arc wound on metal fmmcs to provide Eh:ctru-Magm:tic damping. :
The metal frames constitute a short-circuit tum in a magnetic field. ·I
Ammeters coils, are however wound in a non-conductive frame, because the
coil turns are effectively shorted by the ammeter shunt. The coil itself provides
the EM damping.
If low frequency alternating current is applied to the movable coil, the ·
deflection of the pointer would be upscale for half the cycle of the input wave-
form and do\vnscale (in the opposite direction) for the next half. At power line
frequency (50 Hz) and above, the pointer cannot follow the rapid variations in
direction and quivers slightly around the zero mark, seeking the average value of
the ac (which equals zero). The PMMC instrument is therefore unsuitable for ac
measurements, !Jnless the current is rectified before reaching the coil.
Practical coil areas generally range from 0.5 - 2.5 cm2 •
The 1flux density
.
for modem instmments usually ranges from I 500 - 5000
Wb/cm-. ·
The power requirements ofD'Arsonval movements are quite small, typically
from 25 - 200 J.1W.
The accuracy ofthe instrument is generally of the order of2 - 5% of full scale
deflection.
The permanent magnet is made up of Alnico material.
Scale markings of basic de PMMC instruments are usually linearly spaced,
because the torque (and hence the poi'nter deflection) is directly proportional
to the coil current. The basic PMMC instrument is theretore a linear-reading
device . .
The advantages and disadvantages ofPMMC are as follows.
Advantages
I. They can be modified with the help of shunts and resistance to cover a
wide range of currents and voltages.
2. They display no hystersis.
3. Since operating fields of such insimmcnts arc very strong, they are not
significantly affected by stray magnetic fields.
Copyrghted Mater al
Indicators and Display Devices 3 I
Disadvantages
I. Some errors may set in due to ageing of control springs and the permanent
magnet.
2. Friction due to jewel-pivot suspension.
Pofnter
Spring
I...::::::::;:;=:::;;::;;::::=;::;:::.J Taut
Band
"'-cradle
Scale
->n--+--1--- Cradle
-+-- Ring Bar Magnet
~---1'--+-- Coil Assembly
·The current to be measured is passed through the coil, thereby energising it.
The interaction of the magnetic fields deflects the cradle to one side and moves
the pointer along the scale.
The movement of the cradle exerts a twisting Ioree on the steel bands. These
twisted bands supply the torque to return the pointer to zero, when no current
~..opyng~red mater al
32 Elearonic lnsrrumenlotion
Hows. There are no bearings, and there is a constant level of sensitivity throughout
the range of movement.
Taut band instruments have a higher sensitivity than those using pivots and
jewels. In addition taut band instruments are relatively insensitive to shock and
temperature and are capable of withstanding greater overloads than PMMC or
other types.
ELECTRODYNAMOMETER 2.4
The D' Arsonval movement responds to the average or de value of the current
flowing through the coil.
If ac current is sought to be measured, the current would flow through the
coil with positive and negative half cycles, and hence the driving torque would
be positive in one direction and negative in the other. [f the frequency of the ac
is very low, the pointer would swing back and forth around the zero point on the
meter scale.
At higher frequencies, the inertia of the coil is so great that the pointer does
not follow the rapid variations of the driving torque and vibrates around the zero
mark.
Therefore, to measure ac on a D' Arsonval movement, a rectifier has to be
used to produce a unidirectional torque. This rectifier converts ac into de and
the rectified current def. ects the coil. Another method is to use the heating effect
of ac current to produce an indication of ils magnitude. This is done using an
clcctrodynamometer (EDM).
An clectrodynamometcr is often used in accurate voltmeter and ammeters not
only at power line frequency but also at low AF range. The electrodynamometer
can be used by slightly modifying the PMMC movement. It may also serve as a
tmnsfcr instrument, because it can be calibrated on de and then used directly on
ac thereby equating ac and de measurements of voltage and current directly.
A movable coil is used to provide the magnetic field in an clcctrodynamometer,
instead of a permanent ma!,'llCt, as in the D' Arsonval movement. This movable
coil rotates within the magnetic field. The EDM uses the current under
measurement to produce the required field flux. A fixed coil, split into two equal
halves provides lhe magnetic field in which the movable coil rotates, as shown in
Fig. 2.6 (a). The coil halves arc connected in series with the moving coil and are
fed by the current being measured. The fixed coils are spaced far apan to allow
passage for the shaft of the movable coil. The movable coil carries a pointer,
which is balanced by counterweights. Its rotation is controlled by springs, similar
to those in aD' Arsonval movement.
The complete assembly is surrounded by a laminated shield to protect the
instrument from stray magnetic field which may affect its operation.
Damping is provided by aluminium air vanes moving in a sector shaped
chamber. (TI1e entire movement is very solid and rigidly constructed in order to
keep its mechanical dimensions stable, and calibration intact.)
The operation of the instrument may be understood from the expression for
the torque developed by a coil suspended in a magnetic field, i.e.
T=BxAxNxl
indicating that the torque which deflects the movable coil is directly proportional
to the coil constants (A and N), the strength of the magnetic field in which the coil
moves (B), and the current (I) flowing through the coil.
ln an EDM. the flux density (B) depends on the current through the fixed coil
and is therefore proportional to the deflection current (I). Since the coil constants
are fixed quantities for'any given meter, the developed torque becomes a function
of the current squared (l).
lf the EDM is used for de measurement, the square law can be noticed by the
crowding of the scale markings at low current values, progressively spreading at
higher current values.
For ac measurement, the develo~ed torque at any instant is proportional to
the instantaneous current squared (i ). The instantaneous values of;- are always
positive and torque pulsations are therefore produced.
The meter movement, however, cannot follow rapid variations of the torque
and take up a position in which the average torque is balanced by the torque of
the control springs. The meter deflection is therefore a function of the mean of
the squared current. The scale of the EDM is usually calibrated in terms of the
square root of the average current squared, and therefore reads the effective or
rms value of the ac.
The transfer properties of the EDM become apparent when we compare the
effective value of the aJt,e rnating current and the direct current in tenns of their
heating effect, or transfer of power.
· (If the EDM is calibrated with a direct current of 500 rnA and a mark is placed
on the scale to indicate this value, then that ac current which causes the pointer to
deflect to the same mark on the scale must have an nns value of 500 mA.)
The EDM has the disadvantage of high power consumption, due to its
construction. The current under measurement must not only pass through the
movable coil, but also provide the necessary field flux to get a sufficiently strong
magnetic field. Hence high mmf is required and the source must have a high
current and power.
In spite of this high power consumption the magnetic field is still weaker than
that of the D'Arsonval movement because there is no iron in the path, the entire
flux path consisting of air.
The EDM can be used to measure ac or de voltage or current, as shown in
Figs. 2.6 (a) and (b).
Typical values of EDM flux density are in the range of approximately 60
gauss as compared to the high flux densities (1000 - 4000 guass) of a good
D' Arsonval movement. The low flux density of the EDM affects the developed
torque and therefore the sensitivity of the instrument.
The addition of a series multiplier converts the basic EDM into a voltmeter
[Fig. 2.6 (b)] which can be used for ac and de measurements. The sensitivity of
Copynghted mater al
34 Elearonic Instrumentation
' - - - -- - - - - - - - - To Load
Toac
Source - - - - - - - - - - - - - - - - - - - Circuit
To ac To Load
"i;So"-'u"'rc,e,___ _ _......L_ _ _ _ _ _ _ _ _ ___;;,Circuit
The EDM movement is extensively used to measure power, both de and ac,
for any waveform of voltage and current.
An EDM used as a voltmeter or ammeter has the fixed coils and movable coil
connected in series, thereby reacting toP.
When an EDM is used as a single phase wattmeter, the coil arrangement is
different, as shown in Fig. 2. 7.
The fixed coils, shown in Fig. 2. 7 as separate elements, are connected in series
and carries the total line current. The movable coil located in the magnetic field
of the fixed coils is connected in series with a current-limiting resistor across the
power line. and carries a small current.
The deflection of the movable coil is proportional to the product of the
instantaneous value of current in the movable coil and the total line current. The
EDM wattmeter consumes some power for the maintenance of its magnetic field,
but this is usually small compared to the load power.
Copyr.9nted mater al
Indicators and Display Devices 35
t
Input
/
1 Current Coil
Copynghted mater al
l6 Electronic lnstrumentDtion
Control - - --
Spring
Field
Magnetic Coil
Poles Induced ----
by Field Flux
Moving Fixed
Vane Vane
Pointer
Control Spring - - -
Fixed Vane
Movable Vane
Copynghted mater al
38 Electronic lnstromentation
Adding a suitable multiplier converts the iron vane movement into a voltmeter;
adding a shunt produces different current ranges. When an iron vane movement is
used as an ac voltmeter, the frequency increases the impedance of the instrument
and therefore a lower reading is obtained for a given applied voltage. An iron
vane voltmeter should therefore always be calibrated at the frequency at which it
is to be used. The usual commercial instrument may be used within its accuracy
tolerance from 25-125 Hz.
INTRODUCTION
1.1 MEASUREMENTS
Measurements are made or measurement systems are set up for one or more of
the following functions:
I. To monitor processes and operations
2. To control processes and operations
3. To carry out some analysis
These functions are elaborated now.
1.1.1 Monitorln,g
Thermometers, barometers, anemometers, water, gas and electricity meters only
indicate certain quantities. Their readings do not perform any control function
in the normal sense. 1bese measurements are made for monitoring purposes only.
1.1.2 Control
The thermostat in a refrigerator or geyser determines the temperature of the
relevant environment and accordingly switches off or on the cooling or heating
mechanism to keep the temperature constant, i.e. to control the temperature. A
single system sometimes may require many controls. For example, an aircraft
needs controls from altimeters, gyroscopes, angle-of-attack sensors, thermo-
couples, accelerometers, etc.
Controlling a variable is rather an involved process and is therefore a subject
of study by itself. To get an idea of the steps involved, we will consider elements
of control in Chapter 15.
1.13 Analysis
Measurements are also made to
I. test the validity of predictions from theories,
2. build empiric.al models, i.e. relationships between parameters and
quantities associated with a problem, and
3. characterize materials, devices and components.
In general, these requirements may be called analysis.
1
Copynghtod mater at
2 INTROOucnON TO INSTRUMENTATION AND CONTROL
1.2 INSTRUMENTS
Trans-
Signal
condi-
Signal
lr3nS-
Display/
Recor-
~
ducer tioning mitter ding -
unit device -
Human
observer
Fig. 1.1 A generalized instrument.
1.2.1 T ransducer
A transducer senses the physical variable to be measured (i.e. measurand) and
converts it to a suitable signal, preferably an electrical one.
One point has to be noted in this context. All transducers extract some
energy from th.e measured medium which implies that the measurand is always
disturbed by the measurement system. Therefore, a perfect measurement is
theoretically impossible.
We will consider transducers in general in Chapter 5 and also study a few
representative transducers for measurement of a few non-electrical quantities in
Chapters 6-10.
Copyrgrtod matecal
2
STATIC CHARACTERISTICS OF
INSTRUMENTS
2.1 ACCURACY
Accuracy determines the closeness of an instrument reading to the true value of/
the measurand. Suppose, a known voltage of 200 V is being measured by a
voltmeter and the successive readings are 204, 205, 203, 203 and 205 volts. So,
the accuracy is about 2.5%. Here, though the repeatability of readings is not too
bad, the accuracy is low because the instrument may be having a large calibration
error. Hence, the accuracy can be improved upon by better calibration of the
instrument.
2.1.1 Precision
Precision is another term which is often used in the same connotation as the
accuracy. But in reality precision is different from accuracy. In the above example,
4
Copynghtod mater at
STATIC CHARACTER.ISTICS OF INSTRUMENTS s
the reading can be expressed as 204 :1: 1 V, which means precision is a little less
than 0.5% in this case.
Precision is, therefore, related to the repeatability of the instrument reading
and is a characteristic of the insttuinent itself. To improve the precision of an
insttuinent, its design and construction have to be improved upon.
Symbolically, therefore, if a denotes accuracy, p the precision and c the
calibration error, then a = p + c.
Precision of a meaSurement also depends on what is called the number of
significant figures. An example will perhaps make the point clear. Suppose the
resistance of a conductor is being measured by an analogue ohnuneter. The
ohmmeter indicates the true value, but the observer is unable to read the exact
value because of lack of graduation beyond a certain number of decimals. Thus,
though the instrument is showing the correct value, the precision of the
measurement depends upon the number of significant figures to which the ob-
server can read the value. And in an involved measurement where many meas-
urands are combined, the number of sigilificant figures plays a crucial role in
determining the precision of the ultimate measurement. We discuss below this
aspect in somewhat greater detail.
Example 2.1 Four capacitoxs of values 45.1, 3.22, 89.309 and 0.48 JlF are
connected in parallel. Fimd the value of the equivalent capacitor to the appropriate
number of significant figures.
2.2 SENSITIVITY
Sensitivity is defmed as the absolute ratio of the increment of the output signal
(or response) to that of the input signal (or measurand). Stated mathematically,
S : !iq.J!iq1 where q1 and q0 are the input and output quantities respectively.
Suppose in a mercury-in-glass thermometer the meniscus moves by I em
when the temperature changes by 1o•c. The sensitivity of the thennometer is
therefore 1 mmf'C.
If the functional relationship between the input quantity and the output
reading of an instrument is linear, we call it a linear instrument For example,
a mercury-in-glass thermometer is a linear instrument while a simple thennoc:ou-
ple arrangement for measuring temperature is nonlinear.
The sensitivity of a linear instrument is constant while that of a nonlinear
one varies from one range to another as will be evident from Fig. 2.2 .
llq,
In;put q1 Input q1
(a) (b)
Fl.g. 2.2 Sensitivity: (a) linear instrument where the sensitivity is constant over the
entire range; (b) nonlinear instrument where sensitivity varies from one range
to another.
STATIC CHARACTERISTICS OF INSTRUMENTS 7
q, q, q,
(a) (b) (c)
Fig. 2.3 Deviation from linearity: (a) oscillation with fiXed amplitude; (b) oscillation
with varying amplitude; (c) combined-type osciUation arowxl the best-fit straight
line.
1. The actual output of the instrument may oscillate with a constant amplitude
around the best-fit straight line. In this case, the nonlinearity is expressed
in terms of the amplitude (or maximum deviation). The amplitude is
calculated as the :1: of the full scale deflection (FSD).
2. The actual output of the instrument may oscillate around the best-fit
straight line, but the amplitude of oscillation varies with the input value.
Here, the nonlinearity is expressed as a function of the input value.
Actually, the slopes of lines connecting positive and negative deviations
are determined and the one with a higher deviation from the best-fit line
is used to express the per cent nonlinearity with respect to the input
value.
3. The actual output may oscillate with a fiXed amplitude around the best-
fit straight line over a certain range and then the amplitude may become
a function of the input over the rest. In that case, two computations arc
mad~ne for the fiXed amplitude part, expressed as :!:% of the FSD,
and another for the varying amplitude part, expressed as :!:% of the input
value. Nonlinearity is expressed in terms of the higher value.
Example 2.3 The output of a temperature transducer is recorded over its full-
scale range of 25°C as shown below:
CAlibration
temperature ( 0 C) 0.0 5.0 10.0 15.0 20.0 25.0
Output
reading ( 0 C) 0.0 5.0 9.8 14.8 19.9 25.0
Determine
(a) the static sensitivity of the device, and
(b) the maximum nonlinearity of the device.
Solution Let
q; be the calibration temperature in oc
q. be the output reading in oc
S be the sensitivity = luj/ luj;
D be the deviation from the calibration temperature
61 be the nonlinearity = 100D/FSD = 4D, since FSD = 25• c
Then, we have
q; /uj; q. luJ. s D 61
c·q c•q c·q c·q c•q (%)
0.0 0.0 0.0 0.0
5.0 5.0 5.0 5.0 1.0 0.0 0.0
10.0 5.0 9.8 4.8 0.96 -0.2 0.8
15.0 5.0 14.8 5.0 1.0 - 0.2 0.8
20.0 5.0 19.9 5.1 1.02 -0.1 0.4
25.0 5.0 25.0 5.1 1.02 0.0 0.0
Thus s = 0.96 and maximum nonlinearity a 0.8%.
2.3 REPRODUCIDILITY
2.4 DRIFT
Drift d~notes the change in the indicated reading of an instrument over time
when the value of the measurand remains constant. If there is no drift, the re-
producibility is 100%.
Several causes contribute to the dri~. Stray electromagnetic fields,
mechanical vibrations, changes in superincumbent temperature or pressure, Joule
heating of the components of the instrumen~ etc., are some of the causes. ln the
case of suspended coil PMMC (permanent magnet moving coil) instruments, the
release of internal strain of the suspension wire causes drift of the zero setting.
2.5.1 Hysteresis
Not all the energy put into a system while loading is recoverable upon unload-
Copyrgrted ~aJe a
STATIC CHARACTERISTICS OF INSTRUMENTS 9
ing. For example, a spring balance may show one set of readings when the
weight is increased in steps and another set of readings when the weight is de·
creased in steps. As a result, the plot of the pointer reading vs. weight may have
the appearance of Fig. 2.4.
q,
q,
Fig. 2.4 Hysteresis effects shown in an exaggerated way.
2.5.2 Threshold
Suppose an instrument is in its zero position, i.e. there is no input to it. If now
an input is gradually applied to it, the instrument will require some minimum
value of input before it shows any output (Fig. 2.5). This minimum input which
is necessary to activate an instrument to produce an output is termed its threshold.
q,
~; q,
Threshold
Fig. 2.5 Threshold effect.
2.5.3 Resolution
Over and above the threshold input, an instrument needs a minimum increment
in input to produce a perceptible output. This minimum necessary increment is
called the resolution of the instrument. Thus, resolution, which denotes the
smallest measurable change in input, is similar to sliding friction, while threshold
signifies the smallest initial input resembling the static friction.
Example 2.4 An analogue ammeter has a linear scale of 50 divisions. Its full-
scale reading is I 0 A and half a scale division can be read. What is the resolution
of the instrument? ·
Solution I scale division 10/50 A= 0.2 A. Thus, resolution= 1/2 scale division
= 0.2/2A = 0.1 A.
Example 2.5 The dead zone in a pyrometer is 0.125% of the span. The
instrument is calibrated from goo•c to 1800°C. What temperature change must
occur before it can be detected?
One may see from the tree that human errors can be subdivided into three
classes-gross errors, misuse and vbservational errors. We explain them below.
2.7.2 Misuse
A casual approach on the pan of the operator is the cause of this error. For ex-
ample, in electrical measurements, if the leads are not connected firmly, or an
ohmic contactt is not established, or the initial adjustment such as zero-checking
t1f materials contacted are metals, it is easy to establish an ohmic contact between them
through a proper cleaning of their surfaces. But if the contact is between a metal and a
semiconductor, it is necessary tOo consider their Fermi levels, or e·lse a rectifying contact
may result. For a discussion on this, see Solid State Electronic Devices, 4th ed., by
B.G. Streetman, Prentice-Hall of India (1995), pp. 187- 189.
Copyrghted ma!£!r al
STATIC CHARACTERISTICS OF INSTRUMENTS 11
is not done properly, or fo.r a microvolt order measurement proper care is not
taken to avoid thenno emfs arising out of junctions of dissimilar metals, etc.
errors will creep in. Alertness and perception on the part of the operator are the
only remedies for such errors.
As shown in Fig. 2.6, this error may have two possible origins-instrumental
and environmental.
2.8.1 Instrumental
The instrumental error, in turn, may originate from two different causes-inherent
shortcomings and loading effects.
/11here11t shorlcomillgs
As the name implies, this error creeps in owing to malfunctioning of the com-
ponents of instruments caused by ageing etc. For example, the spring of a galva-
nometer may become weak, thus changing its calibration. Therefore, to avoid
this error, the calibration of the instrument should be checked from time to time.
Loadi11g effects .
We have already mentioned that any measurement involves extraction of some
energy, however small, from the measured medium changing thereby the value
of the measurand from its pristine undisturbed state. This makes perfect
measurement theoretically impossible.
This phenomenon of extraction of energy by a measurement system from
the measured medium is known as the loading effect. We will now consider this
effect from the standpoint of electrical measurements.
Parallel or shunt connection: voltage measurement. Suppose the voltage
across the terminals A and B of the circuit in Fig. 2.7 is being measured in the
usual way. No sooner is the voltmeter connected to the terminals A and B, then
the circuit is changed and the value of £ 0 is altered. The following analysis will
make the point clear.
If z.is the true impedance between A and B, ZL is the impedance of the
Copynghtod mater at
12 INTRODOCTION TO INSTRUMENTATION AND CONTROL
-[F~!··f~
Z, i,
+
E, E, Z,
R, (a)
B
(b)
•
Fig. 2.7 Loading effect in voltllge measurement (a) schematic arrangement; (lb) Thcvenin
equivalent.
loading circuit (here, the voltmeter), E0 is the true voltage between A and B, and
EL is the voltage seen by the loading circuit, their values may be derived as
follows:
Th.erefore,
Thus,
Eo
EL = ---":::-
1+ ~ (2.1)
ZL
For the sake of simplicity, we made a de analysis, but it holds true for ac
as well. In fact, the voltage is modified both in magnitude and phase in the case
of ac. The point will be: clear from Example 2.9. In any case, it is clear from
this analysis that the instrument will give true result if ZL ~ oo and a Teasonably
accurate one if ZL > Z0 •
Series connection: curr ent measurement.. If AB in Fig. 2.8 is shorted and i 0
is the current flowing through the circuit, then i 0 a E0 /Z0 • When the current in
Z, A
~----;:t-----=->
E. T~___----!.,._::r-z,~
B
Fig. ,2.8 Loading effect in current measurement.
R Stcn Stcn
~ £, s s
-
0
-
.,.,
(a) (b)
R, 5 ten
~o~
(c)
Fig. 2.9 Example 2.6.
Solution Successive Thevenin equivalents of the circuit are shown in
Figs. 2.9(b) and (c). From·these we fmd
10
E. ; X 10 ; 5V
10 + 10
10 10
R = 10ldlll10ldl= X = 5ldl
I 10 + J0
Similarly, E. = 2.5 V,
Actual value of the current through R is
E 2-5
I = mA =-rnA= 100J.1A
• R. + 5 + 15 25
The ammeter will read
25
I, = E mA = · mA = 92.6J.lA
R0 + 5 + 15 + 2 27
For 99% accuracy,
Example 2.7 What is the true value of the voltage across the terminals A and
B (Fig. 2.10)? What would a voltmeter of 20 k.OIV sensitivity read on the
SO V and 10 V ranges?
500 k11
r-"""NV-~--A
. 20V S
8 _B
.___ _.._
"'
Fig. 2.10 Example 2.7.
Solwtion Current in the circuit, I= 20/(500 + 500) rnA.= 0.02 mA. Therefore,
£ 0 = 0.02 x w-3 x 500 x lol V = 10 V. But the voltmeter offers different load
resistances in its different ranges.
In the SOY range: The load resistance, RL = 20 x 1ol x 50 0 = 10 6 0 . Therefore,
10
= s.ov
1 '1-
250 X Jol
6
10
In the 10 v range: RL = 20 X I~ X 10 n=2 X lOS n. So,
10
EL = - - - - - . . . P< 4.4 v
250 X 1ol
1 :+-
2 X !OS
Note the magnitude of the error introduced through a wrong setting of the range
of the instrument.
Example 2.8 What percentage error may be expected in measuring the voltage
by the arrangement shown in Fig. 2.ll(a)?
~~~~~EI~:J~
soo 100 0 1000
(a) K. (b)
8§(c)
Fig. 2.11 Example 2.8.
Solution Computing the Thevenin equivalent of the first stage on the left, we
get
E'0 = E X 50 = El v
100 + 50 + 50 4
Thus the effective circuit becomes that shown in Fig. 2.1l{b). The Thevenin
equivalent of this circuit is
E~ - E' X 50 = Et X 50 = Et v
0
R:, + 100 + 50 + 100 4 300 24
The effective circuit now looks lilce Fig. 2.1 l{c). The voltllge developed across
the 1 kO resistance of the measuring insttument is
Eo-EM
E"
X 100 - ( 1 - ~~) X 100 = {1 - 0.9302) X 100
0
=6.98
Example 2.9 An oscilloscope having an input resistance of 1 Mn shunted by
a 50 pF capacitor is connected across a circuit having an effective resistance of
10 kO {Fig. 2.12). If the open circuit vollllge has 1.0 V peak for a sine wave,
what voltage will the oscilloscope indicate when the frequency is (i) 100 kHz
and (ii) 1 MHz?
8
Fig. 1.12 Example 2.9.
SollltiD11 {i) 100 kHz setting: The relevant quantities expressed by usual sym·
bols are:
X = _1_ = 1 = 10S 0
c 2rrjC 27r X lOS X 50 X 10-Jl 7r
z = RL ( -)"XC ) =
L RL- jXc
I - jtr
Eo I ]0
EL =
I+~
ZL
=
I+ jtr
10
=
I + (~J
= I -I.0j0.314
98
_ . _ •
- 0.91 - J0.286 - 0.9:>4 V L - 17.4
o
Xc=-n,
to•
Tr
Therefore,
2R1, ) =O
R, + RL
or
(2.5)
iri· the case of ac circuits, if z. = impedance of the source = R0 + jX., z. • =
complex conjugate of Z0 and ZL = impedance of the load = RL + jXL> it can be
shown that for maximum power transfer,
ZL = z.• = R 0 - jX0 (2.6)
Copyrgrtcd ma:eral
STATIC CHARACTERISTICS OF INSTRUMENTS 17
2. P,.,.., ~ E/14R0 •
3. Impedance matching is not critical because
(2.7)
Hence, for a 100/o deviation, i.e. RLIR. = 1.1, or, 0.9, PIP..... ,= I, which
me-ans that power transfer is almost 100% for this impedance mismatch.
For a 20% deviation, the corresponding figure is nearly 99% and even
for a I 000/o deviation, the power transfer is as much as 89%. Figure 2.13
offers a visual estimation of the amount of power transfer vis-a-vis the
impedance mismatch asjudged from the present theorem.
.. .. .....
....... .
..... 1\..
. . ,.,..
• 0 • • " .
.
•
R = / 2 (2R ) = E:
T o 2R
0
I
- ~ 50%.
2
Copynghlod mater at
18 INTRODUCTION TO INSTRUMENTATION AND CONTROL
z.
Solution ZSOUI« = 0.5 + jl n. Z.,.., = 1.5 + j4 n. Therefore, = Zsou""' + z.,...
= 2 + j5 n. Hence, for maximum power transfer, Z1oad = * c 2 - j5 n. z.
E. = 20 V. So, the transferred power= 202/(4 x 2) W = 50 W.
Example 2.11 In a series circuit, if E., R. and RL are the source voltage,
source resistance and load resistance respectively, and P and P.... are power
transferred to the load and the maximum power that can be transferred to the
load respectively, find the value of the ratio PIPnw in per cent when the source
resistance is 50% of the load resistance.
Example 2.12 A human nerve cell has an open circuit voltage of 80 mV and
it can deliver a current of 5 nA through a 6 Mil load. What is the maximum
power available from the cell?
(80 X JQ-))l
P.max = = 0.16nW
4 X JQ X 106
2.8.3 Environmental
Environmental factors, such as atmospheric pressure, temperature and humidity
affect many measurements changing certain parameters. Consider the simple
length measurement with the help of a scale. Plastic scales, now very common,
change their lengths in humid conditions, while metal scales, although enjoying
immunity from humidity, are affected by a change in temperature. Vibratioll5
caused by running machinery or vehicles play havoc in measurements involving
highly sensitive instruments such as electron microscopes. A lot of electromagnetic
instruments are affected by stray electromagnetic fields.
1
The remedy from such factors is, of course, controlling temperature and
humidity, using vibration-free mountings for instruments, and shielding
Copynghtod mater at
STATIC CHARACTEIUSTICS OF INSTRUMENTS 19
REVIEW QUESTIONS
2.1 What are the different errors encountered in measurements? Explain with
suitable examples.
[ET-6 Wl993]
2.2 Define and explain briefly the static performance parameters of instruments.
[ET-23 Wl993]
2.3 Explain, giving an example, as to why the input effective resistance of the
measuring instrument should be very high like that of a potentiometer or
a voltage·follower if we want to measure temperature by a thermocouple
accurately.
[ET-23 Sl994]
2.4 What are the various sources of gross, systematic and random errors in a
process of measurement? How are these errors minimized?
[ET· 6 Wl994]
2.5 What do you understand by the term 'instrument errors'?
(ET-23 Wl994]
2.6 What are the various sources of systematic errors? How do these errors
influence the accuracy of measurements?
[ET-6 Sl995]
2.7 Discuss the three major categories of errors usually encountered during
measurement.
[ET-6 Sl996]
2.8 Defme the term 'sensitivity' of an instrument.
[ET-6 Sl996]
2.9 The true value of a voltage is 100 V. The values indicated by a measuring
instrument are 104, 103, 105, 103 and 105 volts. Find the accuracy of the
measurement and the precision of the instrument.
[UoK 1993]
2.10 A voltmeter bas a uniform scale with I 00 divisions. The full-scale reading
is 5 V and !/5th of a division can be read. What is the resolution of the
instrument?
[UoK 1994)
2.11 The dead zone in a pyrometer is 0.125 of the span. The instrument is
calibrated from 800°C to ISOOOC. What temperature change must occur
before it is detected?
[UoK 1994)
2. 12 A current of3.12 A is flowing through a 53.655 n resistor. Find the value
of the voltage drop across the resistor to an appropriate number of significant
figures.
[UoK 1994)
2.13 List the desirable and undesirable static characteristics of instruments.
[UoK 1995)
2.14 Explain with an example the difference between accuracy and precision of
measurement. On what factors does precision depend? How can the
accuracy be improved upon?
[UoK 1995)
2. 15 Four capacitors of value$ 45.1 IJF,. 3.22 IJF, 89.309 ~tF and 0.48 !lf are
connected in parallel. Find the value of the equivalent capacitor to an
appropriate number of significant figures.
[UoK 1995)
2.16 lin a series circuit if £ 0 , R0 and RL are the source voltage, source resistance
and load resistance respectively, and P and Pmax are the power transferred
to the load and the maximum power that can be transferred to the load
respectively, find the value of the ratio PIPmv. in per cent when the source
resistance is 50% of the load resistance.
[UoK 1995)
2.17 Explain with examples the terms 'static sensitivity' , ' linearity', 'hysteresis'
and 'dead zone ' in an instrumentation system.
[UoK 1996)
2.18 Enumerate the various sources of error encountered in a measurement
system. [UoK 1996)
2.19 Choose the correct answers:
(a) The reliability of an itistrument refers to:
(i) measurement changes owing to temperature variation.
(ii) the degree to which repeatability continues to remain within the
specified limits.
(iii) tlte life of the instrument.
(iv) the extent to which the characteristics remain linear.
[ET-23 Wl993)
Copynghtod mater at
22 INTRODUCTION TO INSTRUMENTATION AND CONTROL
'·
Copyrgrted ~aJe a