User Manual

MELAdem® 53
MELAdem® 53 C
Ion exchanger

EN
Dear customer,
We thank you for your confidence demonstrated by the purchase of this MELAG product.
As an owner-run and operated family concern founded in 1951, we have a long history of successful specialization in
hygiene products for practice-based use. Our focus on innovation, quality and the highest standards of operational
reliability has established MELAG as the world’s leading manufacturer in the instrument treatment and hygiene field.
You, our customer are justified in your demand for the best products, quality and reliability. Providing "competence in
hygiene" and "Quality – made in Germany", we guarantee that these demands will be met. Our certified quality
management system is subject to close monitoring: one instrument to this end is our annual multi-day audit conducted in
accordance with ISO 13485. This guarantees that all MELAG products are manufactured and tested in accordance with
strict quality criteria.

The MELAG management and team.


General notes

General notes
Please read this user manual carefully before commissioning the mixed-bed resin cartridge. The user manual includes
important safety information. The functionality and value-retention of this mixed-bed resin depends on the care accorded
to it.
Please store this user manual carefully and in close proximity to your mixed-bed resin cartridge. It represents a
component of the product.

About this manual


Symbols used
Symbol Explanation

Draws your attention to a situation, which if not avoided, can result in damage to the water
treatment unit.
Draws your attention to important information.

Formatting rules
Example Explanation

see Chapter 2 Reference to another text section or diagram within this manual.

ii
Contents

Contents
Chapter 1 – Performance specifications.......................................................................................... 5
Intended use .................................................................................................................................................................... 5
Mode of functioning ......................................................................................................................................................... 5

Chapter 2 – Device description ....................................................................................................... 6


Scope of delivery ............................................................................................................................................................. 6
Views of the device .......................................................................................................................................................... 7

Chapter 3 – Commissioning ............................................................................................................ 8


Removing from the packaging ......................................................................................................................................... 8
Requirements of the set-up location ................................................................................................................................ 8
Space requirements......................................................................................................................................................... 9
Requirements for connection to the water pipe ............................................................................................................... 9
Connecting MELAdem 53/MELAdem 53 C .................................................................................................................... 11
Commissioning .............................................................................................................................................................. 12
Installation example ....................................................................................................................................................... 13
Water distributor for connecting further devices ............................................................................................................ 17
Installing the water distributor ........................................................................................................................................ 17
Installation example ....................................................................................................................................................... 18
Combination with other MELAG devices ....................................................................................................................... 19

Chapter 4 – Maintenance .............................................................................................................. 20


Servicing intervals ......................................................................................................................................................... 20
Replacing the mixed-bed resin cartridge ....................................................................................................................... 20
Regenerating the exhausted mixed-bed resin cartridges............................................................................................... 21

Chapter 5 – Operating pauses ...................................................................................................... 22


Operating pauses .......................................................................................................................................................... 22
Transport and storage ................................................................................................................................................... 22

Technical data .............................................................................................................................. 23

Appendix A – Accessories ............................................................................................................ 24

Appendix B – Spare parts ............................................................................................................. 25

iii
Safety instructions

Safety instructions
When operating the mixed-bed resin cartridge, please observe the following safety instructions as well as
those contained in subsequent chapters.

Maintenance
 Use only original replacement parts for maintenance.

Leaks
 The water intake is to be closed upon the discovery of leaks. Check all hoses and hose
connections for leaks.

4
Chapter 1 – Performance specifications

Chapter 1 – Performance specifications

Intended use
This MELAdem 53/MELAdem 53 C mixed-bed resin cartridge works according to the ion exchange
procedure and produces demineralized (de-ionised) water. This procedure requires tap water of drinking
water quality. This ion exchanger does not provide low-germ water.
The MELAdem 53/MELAdem 53 C can be used to supply feeding water to washer-disinfectors and small
steam sterilizers. As an option, the MELAjet spray pistol can be connected as well as for the extraction of
demineralized water for other purposes.
This ion exchanger is conceived for use in the medical field such as clinics, medical and dental practices
as well as other medical institutions outside the patient area.

Mode of functioning
The mixed-bed resin cartridge is filled with a cation and anion resin which removes the salts dissolved in
water using the so-called adsorption process.
The mixed-bed resin is exhausted after absorbing a certain amount of salt and must be replaced and
regenerated (see Chapter 4 – Maintenance).
The capacity of a mixed-bed resin cartridge of a MELAdem 53/MELAdem 53 C depends on the hardness
of the local water supply (see Technical data).

5
Chapter 2 – Device description

Chapter 2 – Device description

Scope of delivery
Please check the scope of delivery before connecting the product.
Standard scope of delivery
 2x Mixed-bed resin cartridge
 User Manual
 Connecting bend with bleed valve for the DI water connection
 Bleed valve for decompression
 Black rubber seal, 3 mm (for water inflow distributor, Y-piece)
 2x Black rubber seal, 2 mm
 Water inflow hose, 2.5 m with a 180° arch and green flat seal
 Safety combination HD in accordance with EN 1717 with a mount and a green flat seal
 Water inflow hose 80 cm
 Sealable bag with cable tie
 Water inflow distributor, Y-piece
 Wrench (SW30)

Optionally
 Water distributor for the connection of multiple devices in parallel
 A water stop for the avoidance of water damage following leaks

6
Chapter 2 – Device description

Views of the device


1 5

7
2
8

Fig. 1: Views of the device MELAdem 53 (identical for MELAdem 53 C)

(1) Polypropylene head-ring


(2) Label for serial number and practice address
(3) Stainless steel mixed-bed resin cartridge
(4) Polypropylene foot ring
(5) Connection to the cold water pipe of the local drinking water supply (cold water IN)
(6) Ventilation screw
(7) Screw connection opening
(8) Connection to connecting bend DI water (pure water OUT)

7
Chapter 3 – Commissioning

Chapter 3 – Commissioning

Removing from the packaging


1. Remove the two mixed-bed resin cartridges included in the scope of delivery from the packaging.
2. Store one of the two mixed-bed resin cartridges included in the scope of delivery as a reserve in
accordance with the storage conditions, see Transport and storage [page 22].
3. Note the serial number of the mixed-bed resin cartridge on the label for control purposes. The
serial number is located on the type plate on the rear of the mixed-bed resin cartridge.

Fig. 2: Label for entering the serial number

Requirements of the set-up location

The following conditions must be given:


 The installation location is clean and frost-free.
 The mixed-bed resin cartridge is installed in a perpendicular position.
 The installation location is freely accessible for the installation, operation and regular replacement
of the mixed-bed resin cartridge.
 The hose connections must be freely accessible.

Increased requirements can be placed on the quality of the DI water (e.g. a low endotoxin content) for the
decontamination of certain medical devices such as ophthalmic instruments. Comply with the following:
 In such cases, an additional filter system is required for the decontamination of DI water.
 It is possible that the drinking water has been contaminated by the water installation (domestic
installation and pre-device periphery).
 Arrange for a check of the drinking water quality at the removal point or request a report (e.g. from the
building management) before setting up and installing the device.
 Further information is available from the corresponding trade associations and their publications. If in
doubt, contact your stockist or the pertinent professional association.

8
Chapter 3 – Commissioning

Space requirements
The space requirement of the mixed-bed resin cartridge corresponds to its dimensions plus a minimum of
25 cm for the safety combination EN 1717 and free access to the hose connections. If the washer-
disinfector / steam sterilizer is to be installed as an installation device, this should be placed in an adjacent
floor unit.

Fig. 3: Front and top view

PLEASE NOTE
The installation of a safety combination requires space of 25 cm above the mixed-bed resin
cartridge.

Requirements for connection to the water pipe


Installation of a cut-off valve with a back-flow preventer and a 3/4" external thread connection in the
proximity of the installation location. In order to provide a standard-conform connection for the mixed-bed
resin cartridge with back-flow preventer and pipe aerator independent of the on-site installation, we
recommend one of the following connections:

PLEASE NOTE
To prevent water damage, we recommend the use of a leak water detector e.g. the MELAG water
stop (art. no. 01056).

9
Chapter 3 – Commissioning

Connection to the cold water supply


The water flow pressure of the on-site tap should be a minimum of 1.5 bar and may not exceed 10 bar.

Alternative connection 1
If a separate water connection (nominal width DN15 with 1/2" collar) is present or will be installed, a water
tap with a back-flow preventer and a pipe aerator is installed.

(1) Tap with back-flow preventer and pipe


aerator (art. no. 37310)
(2) Cold water inflow hose

Fig. 4: Installation with back-flow preventer and pipe aerator

Alternative connection 2
If a cold water connection with an angle valve and a Ø 10 mm duct is present or will be installed, an
additional water tap with an integrated back-flow preventer and pipe aerator will be fitted directly to an
existing angle valve.

(1) Ø 10 mm cold water supply (to the mixer tap)


(2) Additional water tap with an integrated
back-flow preventer and pipe aerator
(3) Cold water inflow hose
(4) Existing angle valve

Fig. 5: Installation with an angle valve

Protection with a safety combination in accordance with EN 1717


The connection of consumers to the drinking water system must be performed in accordance with
EN 1717 so as to protect the drinking water supply from pollution. To this end, fit a safety combination of a
backflow preventer and a pipe aerator to the connection in accordance with EN 1717 part 4.

10
Chapter 3 – Commissioning

Connecting MELAdem 53/MELAdem 53 C


1. Unscrew the yellow caps from the connection fitting before connecting the hoses.

PLEASE NOTE
The yellow caps must be retained to seal the connection fitting during transport e.g. when
replacing the exhausted mixed-bed resin cartridge.

2. Fasten the caps in the sealable bag (included in the scope of delivery) to the head-ring of the
mixed-bed resin cartridge using the cable ties (see Fig. 6).

Open-end wrench

Fig. 6: Example illustration using the MELAdem 53


31.5cm
25 cm

Fig. 7 Connecting the mixed-bed rasin cartridge

(1) Cold water inflow hose connection of the washer-disinfector / steam sterilizer
(2) Water tap (on-site)
(3) Connecting bend with long inflow hose
(4) Safety combination HD consisting of a backflow preventer and a pipe aerator in accordance with
EN 1717
(5) Connection bend with short inflow hose (IN)

11
Chapter 3 – Commissioning

(6) Connection bend for DI water connection (OUT)


(7) Bleed valve for decompression of the mixed-bed resin cartridge
(8) Connection of the DI water inflow hose of the washer-disinfector / steam sterilizer

Proceed as follows to connect the mixed-bed resin cartridge (see Fig. 7):
1. Mount the safety combination HD (4) 25 cm over the mixed-bed resin cartridge on the wall in
accordance with EN 1717.
2. Connect the connecting bend of the long water inflow hose (3) to the safety combination HD (4)
(above). Then connect the other end to the drinking water supply or to the delivered Y-piece using
a wrench.
3. Connect the short water inflow hose (5) to the safety combination (4) (below) using the green flat
seal.
4. Connect the connecting bend of the short water inflow hose (5) with the 2 mm thick black rubber
seal to the connection fitting "IN" of the mixed-bed resin cartridge. Use the open-end wrench.
5. Connect the connecting bend for DI water (6) with the 2 mm thick black rubber seal to the
connection fitting "OUT". Use the open-end wrench.
6. Connect the DI water inflow hose of the washer-disinfector / steam sterilizer to the connecting
bend just mounted.

Commissioning
The mixed-bed resin cartridge must be de-aerated before initial commissioning und after the change as
followed (see Fig. 1):
1. Turn on the cold water tap.
2. Open the ventilation screw a little and wait until water emerges.

PLEASE NOTE
When releasing the ventilation screw, ensure that the screw connection opening does not work
loose.

3. Tighten the ventilation screw as soon as the first drops of water emerge from it.
4. Finally check the hose connections for leaks.
The mixed-bed resin cartridge is now ready to operate.

12
Chapter 3 – Commissioning

Installation example
Example 1 – MELAtherm 10 with MELAdem 53 C

(6) (1)
(7) (5)
60 cm (recommended)

(2)
(4)

(3)

Fig. 8: Installation example 1with MELAtherm 10 and the MELAdem 53 C

(1) Safety combination HD consisting of a backflow preventer and a pipe aerator in accordance with
EN 1717
(2) Bleed valve for decompression of the DI cartridge
(3) DI water inflow hose of the MELAtherm 10
(4) Cold water inflow hose of the MELAtherm 10
(5) Y-piece (included in the scope of delivery of the mixed-bed resin cartridge)
(6) Water tap (on-site)
(7) Effluent connection (on-site)

13
Chapter 3 – Commissioning

Example 2 – MELAtherm 10 with the MELAdem 53 1 and MELAjet

(2) (1)
(8)

(7)

(9)
(6)

(5) (3)

(4)

Fig. 9: Installation example 2 with MELAtherm 10, MELAdem 53 and MELAjet

(1) Safety combination HD consisting of a backflow preventer and a pipe aerator in accordance
with EN 1717
(2) MELAjet spray pistol
(3) Water branch for feed water supply (art. no. 37241)
(4) DI water inflow hose of the MELAtherm 10
(5) Cold water inflow hose of the MELAtherm 10
(6) Y-piece (included in the scope of delivery of the mixed-bed resin cartridge)
(7) Water tap (on-site)
(8) Water stop with cut-off valve and sensor (art. no. 01056), recommended for a permanent water
supply
(9) Effluent connection (on-site)

1 MELAdem 53 C can be connected as well.


14
Chapter 3 – Commissioning

Example 3 – MELAtherm 10 with Vacuklav 40 B+/44 B+ 2 and MELAdem 53 3

(1)

(2)

(13)
(12)

(11)
60 cm (empfohlen)

(4)
(8)
(3)
(10) (9) (7) (5)

(6)

Fig. 10: Installation example 3 with MELAtherm 10, Vacuklav 40 B+/44 B+ and MELAdem 53

(1) Safety combination HD consisting of a backflow preventer and a pipe aerator in accordance
with EN 1717 4
(2) Water stop with cut-off valve and sensor optional (art. no. 01056), recommended for a permanent
water supply
(3) Wall outlet (on-site)
(4) Water tap (on-site)
(5) Cold water inflow hose of the mixed-bed resin cartridge
(6) Effluent hose of the MELAtherm 10
(7) Cooling water inflow hose to the steam sterilizer
(8) Y-piece (included in the scope of delivery of the mixed-bed resin cartridge)
(9) Cold water inflow hose of the MELAtherm 10
(10) DI water inflow hose of the MELAtherm 10
(11) Water branch (art. no. 37241)
(12) Cooling water outflow 3/4"
(13) Steam sterilizer feed water connection

PLEASE NOTE
When using a Cliniklav 25, the mixed-bed resin cartridge must be connected to the steam
sterilizer reservoir.

2 Other MELAG steam sterilizers (with a permanent water supply) can be connected as well.
3 MELAdem 53 C can be connected as well.
4 Please observe the fitting height of the safety combination HD, see example 1.

15
Chapter 3 – Commissioning

Example 4 – MELAtherm 10 with Vacuklav 41 B+/43 B+ 5 and MELAdem 53 6

(2)
(1)
(12)

(11)

(3)

(4)
(10) (8)
(5)
(9) (7)

(6)

Fig. 11: Installation example 4 with MELAtherm 10, Vacuklav 41 B+/43 B+ and the MELAdem 53

(1) Safety combination HD consisting of a backflow preventer and a pipe aerator in accordance
with EN 1717 7
(2) Water stop with cut-off valve and sensor optional (art. no. 01056), recommended for a permanent
water supply
(3) Water tap (on-site)
(4) Cold water inflow hose of the MELAtherm 10
(5) Wall outlet (on-site)
(6) Effluent hose of the MELAtherm 10
(7) Y-piece (included in the scope of delivery of the mixed-bed resin cartridge)
(8) Cold water inflow hose of the mixed-bed resin cartridge
(9) DI water inflow hose of the MELAtherm 10
(10) Water branch for feed water supply (art. no. 37241)
(11) Steam sterilizer feed water connection
(12) One-way discharge

5 Other MELAG steam sterilizers (stand-alone) can be connected as well using a separate connection set.
6 MELAdem 53 C can be connected as well.
7 Please observe the fitting height of the safety combination HD, see example 1.

16
Chapter 3 – Commissioning

Water distributor for connecting further devices


MELAG recommends the MELAG water distributor for the connection of further devices (art. no. 69005).
The water distributor is connected to the output side of the mixed-bed resin cartridge on the connecting
bend (see Fig. 12). An additional washer - disinfector can be connected to the free connections e.g. a
MELAtherm 10 and max. three steam sterilizers.
The following tools are required to make the connection: Open-end wrench 16 mm; 27 mm; 30 mm;
32 mm AF; Allen key 6 mm.

(1) Sealing screw


(2) A straight screwed coupling with
O-Ring, for Ø 6 mm PE hose for
connecting a MELAG steam
7 sterilizer or the MELAjet spray
pistol.
(3) Restricted fitting G 3/4" – G1/4“
for connecting a washer-
disinfector
(4) Threadlocker LOCTITE 2701
(5) Hexagonal nut
(6) Connecting bend with bleed valve
(included in the scope of delivery
of the MELAdem 53)
(7) Distributor block

Fig. 12: Water distributor

Installing the water distributor


The water distributor must be fitted in accordance with the number of devices present in the practice.
Proceed as follows (see Fig. 12):
1. Screw the hexagonal nut (5) to the connecting bend (6).
2. Apply the threadlocker Loctite 2701 (4) to at least one of the connecting bend threads (6).
3. Screw the connecting bend (6) in the distributor block (7) in accordance with the local conditions.
4. Counter the screw thread of the connecting bend (6) with the hexagonal nut (5).
The full drying time of the threadlocker Loctite 2701 until commissioning of the mixed-bed resin
cartridge amounts to ten minutes.
5. Select the screw connections (1) and (3) which correspond to the housing which are to be
connected to the water distributor.
6. Unused openings are to be fitted with a sealing screw (1).
7. Finally, check all hose connections for leaks.

Fig. 13: Example illustration with MELAdem 53

17
Chapter 3 – Commissioning

Installation example
Example 5 – MELAtherm 10 with Vacuklav 41 B+/43 B+ 8, MELAquick 12+
and MELAdem 53
Autoklav MELAdem 53
Autoklav

(1)
(2) (4)
MELAtherm 10
Freier
(11) (3)
Anschluss für
Autoklav/ (14)
MELAtherm 10

(5)

(6)
(13) (7)
(12) (10) (9)
(11)
(8)

Fig. 14: Installation example 5 with MELAtherm 10, Vacuklav 41 B+/43B+, MELAquick 12+ and MELAdem 53

(1) One-way discharge


(2) Feed water connection of the steam sterilizer (Vacuklav 41 B+/43 B+)
(3) Safety combination HD consisting of a backflow preventer and a pipe aerator in accordance with
EN 1717 9
(4) Water stop with cut-off valve and sensor optional (art. no. 01056), recommended for a permanent
water supply
(5) Water tap (on-site)
(6) Y-piece (included in the scope of delivery of the mixed-bed resin cartridge)
(7) Wall outlet (on-site)
(8) Effluent hose of the MELAtherm 10
(9) Cold water inflow hose of the MELAtherm 10
(10) Cold water inflow hose of the MELAdem 53
(11) Waste water container MELAquick 12+
(12) DI water inflow hose of the MELAtherm 10
(13) Water distributor (art. no. 69005)
(14) Steam sterilizer feed water connection (MELAquick 12+)

8 Other MELAG steam sterilizers (stand-alone) can be connected as well using a separate connection set.
9 Please observe the fitting height of the safety combination HD, see example 1.
18
Chapter 3 – Commissioning

Combination with other MELAG devices


Alternative 1 Additionally-required components Art. no.

Water branch 37241

+ +
MELAdem 53/ MELAtherm 10 MELAjet
MELAdem 53 C (Evolution)
Alternative 2 Additionally-required components Art. no.
2x water inflow distributor, 37315
Y-piece10 or

+ + Water distributor set 69005

MELAdem 53 MELAtherm 10 MELAtherm 10


(Evolution) (Evolution)
Alternative 3 Additionally-required components Art. no.
Distributor water inflow, 37315
Y-piece 10
Water branch or 37241

+ + Water distributor set

5 m PE hose, Ø 6 mm
69005

28820
(running
meter)
MELAdem 53/ MELAtherm 10 steam sterilizer (fixed water supply)*
MELAdem 53 C (Evolution)
*Premium-Plus-Class, Pro-Class and other steam sterilizers with fixed water connection
Alternative 4 Additionally-required components Art. no.
Water branch or 37241

+ +
Water distributor set 69005

5 m PE hose, Ø 6 mm 28820
(running
meter)
MELAdem 53/ MELAtherm 10 steam sterilizer (Stand-Alone)
MELAdem 53 C (Evolution)
*Premium-Plus-Class, Pro-Class, Euroklav and other Stand-Alone steam sterilizers
Alternative 5 Additionally-required components Art. no.
Water distributor set 69005
5 m PE hose, Ø 6 mm 28820

+ +
max. three Depending on the additional devices (running
further to be connected, you will require meter)
devices further parts e.g. distributor, water
inflow, Y-piece or a water branch.

MELAdem 53 MELAtherm 10 (Evolution)

10 Required if only one on-site water connection is available.


19
Chapter 4 – Maintenance

Chapter 4 – Maintenance

Servicing intervals
Interval Measure Components
Daily Check for leaks and any damage Mixed-bed resin cartridge and its
connections
Every 6 months Check for leaks, swelling, jamming, Hoses and threaded connections
kinks or age-related brittleness PE hose MELAjet spray pistol (if
existing)
Every 6 years Replace all hoses MELAdem 53/53 C as well as MELAjet
spray pistol (if existing)
When required Regenerate mixed-bed resin Mixed-bed resin cartridge

Replacing the mixed-bed resin cartridge


MELAG devices check the conductivity of the DI water and issue a warning if the water quality is bad.
When connecting other devices (e.g. those from other manufacturers) not equipped with internal
conductivity measuring, the conductivity should be checked regularly using an external monitor.
As soon as device issues an error message triggered by an insufficient conductivity value, the mixed-bed
resin cartridge. Use the second (reserve) mixed-bed resin cartridge for the replacement. The replacement
can be performed independently or by your stockist or an authorized MELAG customer services agent.
Proceed in the following manner to replace the mixed-bed resin cartridge (see Fig. 7):
1. Turn off the cold water tap.
2. To do so, hold a small receptacle under the bleed valve on the connecting bend for DI water
(OUT) and open the bleed valve.
3. Wait until no further water leaves the bleed valve.
4. Shut the bleed valve
5. Unscrew the connecting bend of the cold water inflow hose (IN) and that of the DI water
connection fitting (OUT) using the wrench (cf. Fig. 15). Be prepared for an outflow of residual
water.

Fig. 15: Using the open-end wrench, unscrew the cold water inflow hose connecting bend (IN) and the
DI water connection fitting (OUT)

6. Empty the mixed-bed resin cartridge by placing it upside down in a small receptacle with a
capacity of approx. 400 ml. Wait until the end of the water discharge.

20
Chapter 4 – Maintenance

7. Screw on the yellow caps to the connection fitting of the cold water connection
(cold water IN) and the connection fitting DI water (OUT) (pure water OUT) of the exhausted
mixed-bed resin cartridge (cf. Fig. 15).
8. Pack the empty mixed-bed resin cartridge in its MELAG transport packaging.

PLEASE NOTE
Order the transport packaging with the aid of the "Order form for regeneration the
MELAdem 53/MELAdem 53 C".
On the first regeneration the transport packaging will be free of charge.

Fig. 16: Screw on the yellow caps to the connection fitting

9. Connect the new/regenerated mixed-bed resin cartridge as shown in Chapter 3 –


Commissioning (Fig. 7 [page 11]).

Regenerating the exhausted mixed-bed resin cartridges


To allow the mixed-bed resin cartridge to regenerate, please complete the regeneration form 11 and send it
to your stockist or depot (mail or fax). Please include a copy of the completed form when dispatching your
mixed-bed resin cartridge. MELAG will commission your stockist or depot to perform the regeneration.
The mixed-bed resin cartridge will be collected from your practice. Pack the exhausted mixed-bed resin
cartridge in the MELAG transport packaging and keep this ready for collection.
Your mixed-bed resin cartridge will be filled with new mixed-bed resin and returned to the practice 12.
After regenerating the cartridge, check the serial number on the name plate to ensure that it is the same
cartridge as you despatched.

Fig. 17: Aligning the serial number

Retain the MELAG transport packaging (one folded box, two foam inserts) in a safe and dry location.
These are required when returning the mixed-bed resin cartridge for regeneration.

11 You will find the regeneration form in the download area of the MELAG website.
12 This applies to Germany only. In all other areas, the regenerated mixed-bed resin cartridge will be
returned to your stockist.
21
Chapter 5 – Operating pauses

Chapter 5 – Operating pauses

Operating pauses
When making longer operating pauses, e.g. over-night or a weekend or during a holiday, switch off the
water shut-off valve or the central water shut-off valve, otherwise the insurance cover of the building can
be invalidated.

Transport and storage

NOTICE
Incorrect storage and transport can result in damage to the mixed-bed resin cartridges and
mixed-bed resin filling. Should the storage temperature be exceeded over the long-term,
this can result in a loss of quality of the mixed-bed resin.
 Comply with the following information.

 Empty the mixed-bed resin cartridge in accordance with the work steps outlined in Chapter 4 –
Maintenance [page 20].
 Retain the yellow caps for transport purposes.
 Please use exclusively MELAG transport packaging for its transport.
 Store and transport the mixed-bed resin cartridge in a temperature of between 2-20 °C.
 If stored in a cool (max. 20 °C) and dry atmosphere, the mixed-bed resin cartridge can be stored for at
least 18 months from the date of purchase.
 Protect the mixed-bed resin cartridge (contains water) against frost.

22
Technical data

Technical data

Model name MELAdem 53 MELAdem 53 C

Dimensions (D x H) 24 cm x 57 cm (approx. 61.5 cm 24 cm x 45 cm (approx. 49 cm


incl. connecting bend) incl. connecting bend)

Weight (with mixed resin) 18 kg (per mixed-bed resin 14 kg (per mixed-bed resin
cartridge) cartridge)

Amount of resin 20 l 15 l

Exterior connection G 3/4“ G 3/4“

Capacity per cartridge at 10 °dH: 2800 l at 10 °dH: 2100 l


(depends on the water at 20 °dH: 1200 l at 20 °dH: 960 l
hardness and the conductivity
of the local water) at 25 °dH: 700 l at 25 °dH: 650 l

Max. through-flow volume 800 l/h 800 l/h

Resulting water quality 1-5 µS/cm 1-5 µS/cm

Admissible entry temperature 5 °C - max. 40 °C 5 °C - max. 40 °C


of the water

Min. water pressure 1.5 bar 1.5 bar

Max. water pressure 10 bar 10 bar

23
Appendix A – Accessories

Appendix A – Accessories
Optional accessories Figure Art. no.
MELAG water stop valve 01056

MELAjet spray pistol 27300

PE hose, black, Ø 6 mm, running metre 28820

Water branch for feed water connection 37241

Tap with back-flow preventer and pipe aerator 37310

Water inflow distributor, Y-piece 37315

Tap with back-flow preventer and pipe aerator for fitting to 58130
an existing angle valve

Water distributor for the connection of multiple devices in 69005


parallel

24
Appendix B – Spare parts

Appendix B – Spare parts


Spare parts Figure Art. no.
Open-end wrench SW30 for MELAdem 53/ 15530
MELAdem 53 C

Water inflow hose EN 1717, 2.5 m 24930

Water inflow hose MELAdem 53/MELAdem 53 C, 0.8 m 24932

Rubber seal, 2 mm, black 52400


(connections to the mixed-bed resin cartridge)

Flat seal, green 54920


(to the cold water inflow hoses)

Rubber seal, 3 mm, black 56950


(on the hose of the safety combination)

Screw caps 3/4", yellow 58140

Connecting bend G3/4" with bleed valve 70405

25
BA_MELAdem53_53C_01038_01036_EN_v9.docx | Rev.: 9 – 20/0295 | Modifictaion date: 2020-02-01
MELAG Medizintechnik GmbH & Co. KG
Geneststraße 6-10
10829 Berlin
Germany
E-mail: [email protected]
Web: www.melag.com

Original instructions

Responsible for content: MELAG Medizintechnik GmbH & Co. KG


We reserve the right to technical alterations