0% found this document useful (0 votes)
115 views4 pages

Proximitor Calibration Sheet

This document is a calibration sheet for a Proximitor device. It provides the displacement in mils, corresponding gap voltage readings, calculated scale factors in mV/mil and % deviation from the target scale factor of 200 mV/mil for displacements between 0-150 mils. It also includes calibration curves showing the incremental scale factor changes and the gap voltage versus displacement relationship.

Uploaded by

xabs
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
115 views4 pages

Proximitor Calibration Sheet

This document is a calibration sheet for a Proximitor device. It provides the displacement in mils, corresponding gap voltage readings, calculated scale factors in mV/mil and % deviation from the target scale factor of 200 mV/mil for displacements between 0-150 mils. It also includes calibration curves showing the incremental scale factor changes and the gap voltage versus displacement relationship.

Uploaded by

xabs
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

PROXIMITORCalibrationsheet

Customer :
Displacement GAP Scale Factor Scale Factor Plant :
(mils) (-Volts) (mV/mil) (%) Calibration date :
Product Service # :
0 0,69 -100,00%
Unit :
10 1,59 90,00 -55,00% Rack description :
Monitor pos.:
20 3,58 199,00 -0,50% Channel :
Probe location :
30 5,61 203,00 1,50% Tag # :

40 7,67 206,00 3,00% Probe P/N :


Probe S/N :
50 9,67 200,00 0,00% MOD y/n :
Extension Cable P/N :
60 11,71 204,00 2,00% Extension Cable S/N :
Proximitor P/N :
70 13,77 206,00 3,00% Proximitor S/N :
MOD y/n :
80 15,81 204,00 2,00%
Sensitivity: 200 mV/mil
90 17,82 201,00 0,50% Cold Gap: -Volts =

100 19,70 188,00 -6,00%

110 21,43 173,00 -13,50%

120 22,22 79,00 -60,50%

130 23,01 79,00 -60,50%

140 23,88 87,00 -56,50%

150 26,00 212,00 6,00%

Calibration Curves
Incremental Scale Factor

10,00%
8,00%
6,00%
4,00%
2,00%
in %

0,00%
-2,00% 0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150
-4,00%
-6,00%
-8,00%
-10,00%

30,00

25,00

20,00
Gap in -Volts

15,00

10,00

5,00
0,00
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150
mil