LOD LOQ Determination

LOD and LOQ definitions are as follows: LOD: The lowest analyte concentration of a sample which can still be detected by the analysis method. LOQ: The lowes sample concentration which can be quantitatively detected with accuracy and an acceptable precision.

Uploaded by

harunsarip
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
853 views2 pages

LOD LOQ Determination

LOD and LOQ definitions are as follows: LOD: The lowest analyte concentration of a sample which can still be detected by the analysis method. LOQ: The lowes sample concentration which can be quantitatively detected with accuracy and an acceptable precision.

Uploaded by

harunsarip
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Home

Blog Search

Dadang Gatot Prasetyo's BLOG...

Thursday, September 17, 2009


Determination of LOD & LOQ of HPLC Analysis Method

Aboout Me...

Determining the LOD and LOQ based on the comparison of standard deviation (SD) of the peak area and the slope of calibration curve. LOD and LOQ definitions are as follows: LOD: The lowest analyte concentration of a sample which can still be detected by the analysis method but do not have to quantitate as an appropriate value. LOQ: The lowes sample concentration which can still be quantitatively detected with accuracy and an acceptable precision. The formula for the LOD and LOQ are as follows:

Dadang Gatot Prasetyo Technical Sevice Engineer Jakarta, Indonesia Email: [email protected] Member of CHROMmunity at http://chrommunity.chromacademy.com Yahoo Mesengger: dadanggp95 Skype: prasetyo.dadanggatot

RSS Subscription

Subscribe via RSS reader:


Posts RSS Comments RSS

Subscribe via Email Address:

Description: is the standard deviation and S is the slope of the calibration curve. Value of 3 and 10 in above equation in accordance with the analysis method provision of regulation maker in your country. Case example: Injection of caffeine standard solution runs on the HPLC, obtained caffeine calibration data, Which LOD and LOQ concentration value depending on the value of the standard solution series that we make as follows:

Recent Posts

Recent Comments

BlogRoll

Approximate wavelength (in vacuum) and frequency ranges for the various colors Logarithmic Relation Between Transmittance and Concenctration All Solvents are NOT Created Equal INTRODUCTION TO GC Cleaning the LCD Screen Time Constant/Smoothing/Filter on HPLC Detectors Determination of LOD & LOQ of HPLC Analysis Method

Table 1. Correlation between caffeine concentration and its area. Above calibration data will produce caffeine calibration curve as follow:

In the picture above, we have line equation: Y=66132X-3940, where the value of slope obtained from the equation is S=66132. Then calculate the standard deviation of Peak Area as follows:

Table 2. Deviation of measurement of caffeine peak area with linier regression. Yi is obtained from the regression equation Y=66132X-3940, for example: for X=25.075 ppm, then Yi=(66132 x 25.075) 3940 = 1654319.90. Standard deviation can be calculated by:

HOME

Copyright 2008 Vibrant Style. All rights reserved.

(Y-Yi)^2 value is obtained from table 2. = (49987352.84/3)^(1/2) = 4081.97 > (Standard Deviation) Thus:

Posted by Gatot - at 4:01 PM Reactions: 0 comments: Post a Comment